IMRT Delivery System QA IMRT Dose DeliveryIMRT Dose Delivery • Acceptance testing •...

9
1 IMRT Delivery System QA Thomas LoSasso, PhD Memorial Sloan Kettering Cancer Center IMRT Dose Delivery Acceptance testing Commissioning Quality assurance Verification QA Static IMRT Leaf affects borders only affects dose within target position 1-2 mm OK 0.2 mm OK Why: specific tests for IMRT? Accuracy of leaf positioning (gaps) Gap error Dose error 0.0 5.0 10.0 15.0 20.0 0 1 2 3 4 5 Nominal gap (cm) % Dose error Range of gap width 2.0 1.0 0.5 0.2 Gap error (mm) Acceptance testing MLC alignment MLC leaf / gap calibration Timing errors Beam characteristics at small MU A60 A31 B30 B1 B60 B31 A30 A1 MLC Alignment A60 A31 B30 B1 A1 A30 B31 B60 Film 1 Film 2 Field B 90° Field B 270° Field A 90° Field A 90°

Transcript of IMRT Delivery System QA IMRT Dose DeliveryIMRT Dose Delivery • Acceptance testing •...

Page 1: IMRT Delivery System QA IMRT Dose DeliveryIMRT Dose Delivery • Acceptance testing • Commissioning • Quality assurance • Verification Q A Static IMRT Leaf affects borders only

1

IMRT Delivery System QA

Thomas LoSasso, PhD

Memorial Sloan Kettering Cancer Center

IMRT Dose Delivery

• Acceptance testing • Commissioning • Quality assurance • Verification

QA

Static IMRT Leaf affects borders only affects dose within target position 1-2 mm OK 0.2 mm OK

Why: specific tests for IMRT?

Accuracy of leaf positioning (gaps)

Gap error Dose error

0.0

5.0

10.0

15.0

20.0

0 1 2 3 4 5

Nominal gap (cm)

% D

ose

erro

rRange of gap width

2.01.0

0.50.2

Gap error (mm)

Acceptance testing

• MLC alignment • MLC leaf / gap calibration • Timing errors • Beam characteristics at small MU

A60

A31B30

B1

B60

B31A30

A1

MLC Alignment

A60

A31B30

B1

A1

A30B31

B60

Film 1

Film 2

Field B90°

Field B270°

Field A90°

Field A90°

Page 2: IMRT Delivery System QA IMRT Dose DeliveryIMRT Dose Delivery • Acceptance testing • Commissioning • Quality assurance • Verification Q A Static IMRT Leaf affects borders only

2

Film 1

Film 20.5 mm misalignments

perfect alignment

Doubly - exposed films MLCtable.txt corrections

00.20.40.60.8

11.21.41.61.8

-10 -8 -6 -4 -2 0 2 4 6 8 10

Distance from central axis (cm)

Corr

ectio

n (m

m)

"precise"

Varian

MLCtablemeasurement

0.0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1.0

-20 -10 0 10 20

Distance from axis (cm)

Gap

at i

soce

nter

upperlower

Gap measurement - mechanical DMLC - Symmetry and Flatness

0.90

0.95

1.00

1.05

-15 -10 -5 0 5 10 15

Distance from axis (cm)

Rela

tive

outp

ut

open5mm5mm/open

12 cm

14 cm

0.90

0.95

1.00

1.05

-15 -10 -5 0 5 10 15

Distance from axis (cm)

Rela

tive

outp

ut

transmissiongap only

0.90

0.95

1.00

1.05

-15 -10 -5 0 5 10 15

Distance from axis (cm)

Rela

tive

outp

ut

open5mm5mm/open

DMLC - Symmetry and Flatness

y = 0.009945x + 0.017685R2 = 0.999977

0.00

0.10

0.20

0.30

-4 -2 0 2 4 6 8 10 12 14 16 18 20

1.80

6 MV

y = 0.009954x + 0.018705R2 = 0.999984

0.00

0.10

0.20

0.30

-4 -2 0 2 4 6 8 10 12 14 16 18 20

1.90

15 MV

Gap correction - measurement

Rel

ativ

e ou

tput

Nominal gap (mm)

Page 3: IMRT Delivery System QA IMRT Dose DeliveryIMRT Dose Delivery • Acceptance testing • Commissioning • Quality assurance • Verification Q A Static IMRT Leaf affects borders only

3

1.4

1.5

1.6

1.7

1.8

1.9

2

2.1

2.2

2.3

CAX 5cm off 10cm offSeries4 Series5 Series6

445

6X 15X

444

6X 15X

Phelps

6X 15X

442

6X

245

6X 15X

+/- 1%

15 MV 6 MV

Mark 1 Mark 2 Millenium

Effe

ctiv

e ga

p of

fset

(mm

)

Effective gap offset variation

1.0

1.2

1.4

1.6

1.8

2.0

-25 -20 -15 -10 -5 0 5 10 15 20 25

Distance from axis (cm)

Eff g

ap o

ffset

(mm

@ is

o)

Millennium 120

2–4% Dose

Off-axis variation - "cupping"

with “cupping”correction

with individual leafcorrections

Off-axis variation - corrected

1.0

1.2

1.4

1.6

1.8

2.0

-25 -20 -15 -10 -5 0 5 10 15 20 25

Distance from axis (cm)

Eff g

ap o

ffset

~1% dose

1.0

1.2

1.4

1.6

1.8

2.0

-25 -20 -15 -10 -5 0 5 10 15 20 25

Distance from axis (cm)

Eff g

ap o

ffset

S = 0.047 mm

S = 0.027 mm

Frequency of deviations (%) # GlobalField Room Angle logged - prescribed leaf positions @ iso (mm) checks RMS

<0.05 0.05-1.0 1.0-2.0 2.0-3.0 3.0-4.0

LAO 245 0 49 50 0.5 0.04 12528 0.26225 50 50 0.5 0.04 12528 0.26

445 0 52 48 0.5 0.02 12492 0.26225 48 51 0.6 0.01 12510 0.26

LPO 245 0 59 40 0.9 0.09 0.03 9918 0.26285 59 40 1.0 0.09 0.04 9918 0.28

445 0 61 38 0.9 0.10 9810 0.22285 58 41 1.0 0.10 9810 0.26

PA 245 0 50 49 0.4 0.01 0.01 18228 0.140 50 50 0.4 0.01 0.01 18198 0.14

445 0 52 47 0.4 0.02 18180 0.140 52 47 0.4 0.01 18144 0.14

RPO 245 0 62 36 1.4 0.10 0.03 10116 0.2475 63 36 1.4 0.11 0.03 10116 0.24

445 0 63 35 1.4 0.12 10152 0.2275 63 35 1.4 0.12 10062 0.22

RAO 245 0 47 53 0.3 0.02 14490 0.14135 48 52 0.3 0.01 14508 0.14

445 0 49 50 0.3 0.01 14526 0.14135 50 49 0.3 0.01 14436 0.14

# checks = # moving leaves * BOT / 55 msec

Dose differences (log-dva)Overlay

Log file Leaf sequence file dose distribution

Inf

Post

Sup

Ant

jaws

2 cm

Page 4: IMRT Delivery System QA IMRT Dose DeliveryIMRT Dose Delivery • Acceptance testing • Commissioning • Quality assurance • Verification Q A Static IMRT Leaf affects borders only

4

Delivered with dose mode at 400MU/min(CL2300, 6MV, 100 SSD, 1.5 cm depth)

-100

-50

0

50

100

150

200

Diff

eren

ce (%

)

# Segment0.25 MU/seg

4 MU/seg

1 MU/seg

16 MU/seg 25 MU/seg

-15-10

-505

1015

4 MU/seg

16 MU/seg 25 MU/seg

Courtesy of Ping Xia, UCSF

Timing error – SMLC (Varian)

400 mu/min

Courtesy of Gary Ezzell and the Journal of Applied Clinical Medical Physics.

Timing error – SMLC (Varian)

14 MU8X

28 MU4X

110 MU 55 MU2X

110 MU

Dose differences (log-dva)Overlay

-2.0-1.5-1.0-0.50.00.51.01.52.0

0 10 20 30 40 50 60 70 80 90 100MU (%)

Leaf

erro

r (lo

gged

- pr

escr

ibed

)

14B

14A

-2.0-1.5-1.0-0.50.00.51.01.52.0

-50 -40 -30 -20 -10 0 10 20 30 40 50Distance from axis (mm)

Leaf

erro

r (lo

gged

- pr

escr

ibed

)

14A

14B

150 MU

Timing / communication errors

-2.0-1.5-1.0-0.50.00.51.01.52.0

0 10 20 30 40 50 60 70 80 90 100MU (%)

Leaf

erro

r (lo

gged

- pr

escr

ibed

)

14B

14A

-2.0-1.5-1.0-0.50.00.51.01.52.0

0 10 20 30 40 50 60 70 80 90 100MU (%)

Leaf

erro

r (lo

gged

- pr

escr

ibed

)

14B

14A

-2.0-1.5-1.0-0.50.0

0.51.01.52.0

-50 -40 -30 -20 -10 0 10 20 30 40 50Distance from axis (mm)

Leaf

erro

r (lo

gged

- pr

escr

ibed

)

14A

14B

-2.0-1.5-1.0-0.50.00.51.01.52.0

-50 -40 -30 -20 -10 0 10 20 30 40 50Distance from axis (mm)

Leaf

erro

r (lo

gged

- pr

escr

ibed

)

14A

14B

60 MU150 MU

Timing / communication errors

Overlayplanlog

Difference(log - plan)

Log file Leaf sequence file dose distribution

150 MU prescribed / 60 MU delivered

A

GF

ECD BA

GF

ECD B

Comparison of prescribed and delivered doses

Page 5: IMRT Delivery System QA IMRT Dose DeliveryIMRT Dose Delivery • Acceptance testing • Commissioning • Quality assurance • Verification Q A Static IMRT Leaf affects borders only

5

0.85

0.90

0.95

1.00

1.05

1.10

1.15

0 30 60 90 120 150MU delivered

Rel

ativ

e do

se (c

orre

cted

to 1

50

MU

)

ABCDEFG

PrescribedMU = 150

MU decrease dose errors increase Timing errorIs there an analogous problem for DMLC?

• Trailing leaves (B-side) compensate for the leading leaves (A-side)

• Dose (MU) error = position error (B) – position errorleaf speed (B) leaf speed (A)

• However, some leaf sequencers actively utilize the beam hold-off to control leaf speed

Beam Stability: Flatness, Symmetry

• Stability of flatness and symmetry affects dose rate for small fields directed off the central axis.

• For an open 20x20cm2 field, measure profiles for irradiations ranging from 1 to 100 MU.

– Sun Nuclear Profiler (46 diodes, 10 profiles/sec).

• Flatness is +/-3% if more than 5MU delivered.• Symmetry +/-3% if more than 4MU delivered.

Courtesy of M. Sharpe, William Beaumont Hospital

Mark 2 (442)

0

0.01

0.02

0.03

0.04

-10 -5 0 5 10

ABaverage

Millenium (444)

0

0.01

0.02

0.03

0.04

-10 -5 0 5 10

ABaverage

Mark 2 (245)

0

0.01

0.02

0.03

0.04

-10 -5 0 5 10

ABaverage

Mark 1 (445)

0

0.01

0.02

0.03

0.04

-10 -5 0 5 10

ABaverage

Transmission profiles across leaves

Distance from axis (cm)

Tran

smis

sion

Parallel

Perpendicular

X

Y

Respiratory gating

Leaf motion

Splitting large fields

Page 6: IMRT Delivery System QA IMRT Dose DeliveryIMRT Dose Delivery • Acceptance testing • Commissioning • Quality assurance • Verification Q A Static IMRT Leaf affects borders only

6

Routine QA

MLC QAMLC specificMechanical stabilityAccuracy of leaf positioning (gaps)

VerificationPatient / field specificTx plan is transferred and delivered correctly

Quality assurance

Motor fatigue Film pattern (semi-weekly)

MLC calibration Dosimetry (monthly)

Backlash Dosimetry (monthly)

Grease Alcohol (semi-annually)

0

10

20

30

40

50

60

70

Jan-92

Jan-93

Jan-94

Jan-95

Jan-96

Jan-97

Jan-98

Jan-99

Jan-00

Jan-01

Jan-02

Jan-03

Date

# 445 245 442

Room

42

70

21

1st MLC TxMLC installed

1st dMLC Tx

Motors replacedMotor replacement history MLC motors replaced445 245 442B A B A B A

26 26 2625 25 2524 2 24 2423 23 2322 22 2221 21 2120 20 2019 19 1918 18 1817 3 17 2 1716 2 16 2 1615 5 15 1514 14 1413 13 1312 12 1211 2 2 11 1110 2 10 109 2 9 98 2 8 87 7 76 6 65 5 54 2 4 43 2 3 32 2 21 1 1

6a Field Size Defined by MLC at 100 cm SAD by Light Field: Date: __________ Initials: __________ Tolerance: 0.5 mm

Digital setting (cm) A carriage B carriage A leaves B leaves 20.0 16.0 12.0 8.0 4.0 0.5 -4.0 -8.0 -12.0 -16.0

Note: If leaf positions exceed tolerance, reinitialize and recheck. If this does

not correct the problem, verify that the correction parameters in files MLCTABLE.TXT and MLCXCAL.TXT have not changed.

4 cm

Leaf / carriage positioning vs off-axis distance

Page 7: IMRT Delivery System QA IMRT Dose DeliveryIMRT Dose Delivery • Acceptance testing • Commissioning • Quality assurance • Verification Q A Static IMRT Leaf affects borders only

7

← - 0.5 mm

→ + 0.5 mm

← - 0.2 mm

→ + 0.2 mm

errors introducedLeaf positioning

1 mm bands 17b DMLC Output (Gap width) Check

15 MV x-rays, File: DMLC_QA2.dva, Dose = 100 MU, Dose rate = 240 MU/min SAD: 100cm, Depth = 3.1cm, Backup = 10cm, C.S. = 10x10cm2 Tolerance: (The ratios should be within 2% (< 0.2 mm gap) of the annual calibration.)

Measure the output for the scanned field before and after reinitializing the MLC. Measure the output for a 10x10cm2 reference field size defined by the jaws (MLC retracted). Calculate the ratio of the scanned field to the reference field outputs.

Reference10x10

DMLC - before reinitialization

DMLC - after reinitialization

Reference 10x10

Average Annual Ratio

17c Beam Holdoff Functionality

Date: Initials

Same setup as 17b, Dose = 5 MU Record the output and the Tx time. Check if the holdoff is being applied. Reading Functioning

DMLC Output vs Time

0.158

0.160

0.162

0.164

0.166

0.168

Jan-98 Jan-99 Jan-00 Jan-01 Jan-02 Jan-03 Jan-04Date

DM

LC /

10x1

0

445 245 442 444 241441 244

0.2 mm ~1%

17d Output vs Gantry and Collimator Angles

Date: Initials

15 MV x-rays, File: DMLC_QA2.dva, Dose = 100 MU, Dose rate = 240 MU/min SAD: 100cm, Spokas in air w/buildup, C.S. = 10x10cm2 Tolerance: (The ratios should be within 3% (< 0.2 mm gap) of each other.) Position a cylindrical chamber in air at isocenter.

For each gantry angle, measure the open field and the scanned field outputs. At gantry angles of 90° and 270°, also measure output for collimator angles of 90°. Calculate the ratio of the scanned field to the reference field outputs.

Collimator angle

Gantry Reference DMLC DMLC / Reference angle 0° 0° 90° 0° 90°

0° 90°

270° 180° 0°

0.16

0.165

0.17

0.175

0.18

1998 1999 2000 2001 2002

Millennium (444)

0.16

0.165

0.17

0.175

0.18

1998 1999 2000 2001 2002

0-090-0270-0180-090-90270-90

Millennium (241)

± 0.2 mm(± 1%)

DMLC output vs gantry and collimator angle

Year

DM

LC /

10x1

0

0.16

0.165

0.17

0.175

0.18

1998 1999 2000 2001 2002

Mark 2 (245)

0.16

0.165

0.17

0.175

0.18

1998 1999 2000 2001 2002

Mark 1 (445)

DMLC output vs gantry angle(diode array)

0.97

0.98

0.99

1.00

1.01

1.02

1.03

-15 -10 -5 0 5 10 15

Distance from axis (cm)

Rela

tive

outp

ut

0-0

90-0

270-0

0.2 mm

Page 8: IMRT Delivery System QA IMRT Dose DeliveryIMRT Dose Delivery • Acceptance testing • Commissioning • Quality assurance • Verification Q A Static IMRT Leaf affects borders only

8

Summary1. Acceptance testing

• MLC capable of IMRT delivery

2. Commissioning• MLC modeled accurately in TPS

3. Routine MLC QA• leaves and carriages - mechanics

4. Patient specific QA• independent verification

IMRTDose Accuracy

Overlay Difference

PlanFilm Film - Plan

2 cm

Improved calculations needed for IMRT fields

~25%

2 cm

Film - Plan

Dosimetry (film – plan)

T&G Source MLC scatter

Corrections (plan – plan)

Lung

PAfield

x10

Tongue and Groove

-1 0 1 2

Page 9: IMRT Delivery System QA IMRT Dose DeliveryIMRT Dose Delivery • Acceptance testing • Commissioning • Quality assurance • Verification Q A Static IMRT Leaf affects borders only

9

Tongue and groove

01020304050607080

-7 -6 -5 -4 -3 -2 -1 0 1 2 3 4 5

Distance from CAX (cm)

Rel

ativ

e do

se

-1 cm 0 cm 1 cm 2 cm

DMLC output vs gantry and collimator angle

0.950.960.970.980.991.001.011.021.031.041.05

1998 1999 2000 2001 2002

DMLC

/ 10

x10

(nor

m to

0-0

)

0-0 180-090-0 90-90270-0 270-90Mark 1 (445)

adju

sted

bea

rings

repl

aced

bea

rings

adju

sted

driv

e be

lts

adju

sted

bea

rings

adju

sted

bea

rings

adju

sted

bea

rings