Improving Non-dispersive Infrared (NDIR) Sensors … › images › documents › CIEJoint2019 ›...
Transcript of Improving Non-dispersive Infrared (NDIR) Sensors … › images › documents › CIEJoint2019 ›...
Improving Non-dispersive Infrared (NDIR) Sensors for Greenhouse Gas Monitoring
Howard W. Yoon
1. Motivation
2. Basic concepts, technical approach
3. NIST design for NDIR and comparisons to Picarro CRDS
4. Conclusion and future work
Outline
Keeling curve (from Mauna Loa)
• Measurements using non-dispersive infrared (NDIR) sensor
• Calibrated daily using standard gas bottles
• CO2 concentrations traceable to manometric techniques
http://scrippsco2.ucsd.edu/
400
320
1960 2020
Historical CO2 levels and projection
http://scrippsco2.ucsd.edu/
(2002 CO2 emissions map by Jesse Allen, based on data from the Vulcan Project.)
Ultimate goal of real-time CO2 monitoring using distributed sensor networks
1. In order to reduce CO2 emissions, we need to measure both sources and sinks.
2. Denser sensor grid is needed.
3. Constant calibrations using standard gases are expensive.
4. Better sensors are needed.
Use of commercial NDIR sensors
Use of a commercial NDIR at the heart of the NOAA Tall Tower analysis system
Picarro CRDS system Limited to NIR where lasers exist
Technical Approach
• Gas concentrations measured using Beer-Lambert law • Use of broad-band lamp sources • Dual-band filters (on/off resonance wavelength) • Typically use pyroelectric detectors with modulated source
Licor 820
Beer-Lambert law
𝐴 = 𝜀𝑙𝑐 A absorbance
Molar absorptivity
l path length of the sample
c concentration
Need to adjust path length for desired sensitivity (we use a 60 cm path length)
Calibrations using standard gas bottles (traceable to gravimetric or manometric standards)
Why now?
150 mm focal length ZnSe lens 50 mm diameter
ISP Optics
Temperature-stabilized
Pyroelectric
Detector with
preamp
Temperature-stabilized
6 mm diameter
copper
Aperture
100 mm focal length ZnSe lens 25.4 mm diameter
Edmunds Optics
50 mm focal length ZnSe lens 25 mm diameter
Edmunds Optics
Temperature-stabilized
tube (TE cooler)
Laser
Pointer
Foam
Insulation
Temperature-stabilized
tube (TE cooler)
Development of the Ambient Radiation Thermometer 8 um to 14 um filter ZnSe lenses Pyroelectric detector
Internal view of the ART
H. W. Yoon, et al. "Improvements in the design of thermal-infrared radiation thermometers and sensors," Opt. Express 27, 14246-14259 (2019)
DN-45 filed May 2019
Repeatability of < 2 mK over 5 days
< 1 mK noise-equivalent temperatures close to ambient temperatures
30 40 50 60 7021.98
21.99
22.00
22.01
22.02
22.03
Tem
per
atu
re,
deg
C
Time (h)
ART
SPRT
Resolution of < 1 mK temperatures at room temperatures
1. Traceable to standard gases 2. Signal of NDIR depends upon
• Internal set temperatures • Detector responsivity • Source stability • Source frequency • Light-pipe stability of transmittance • Humidity
Modification of the ART design for NDIR
Temperature-stabilized
Pyroelectric
Detector with
preamp
Temperature-stabilized
6 mm diameter
copper
Aperture
100 mm focal length ZnSe lens 25.4 mm diameter
Edmunds Optics
50 mm focal length ZnSe lens 25 mm diameter
Edmunds Optics
Temperature-stabilized
tube (TE cooler)
Laser
Pointer
Foam
Insulation
Temperature-stabilized
tube (TE cooler)
Temperature-stabilized
tube (TE cooler)
Foam
Insulation
Temperature-stabilized
tube (TE cooler)
Temperature-stabilized
Pyroelectric
Detector with
preamp
Temperature-stabilized
6 mm diameter
copper
Aperture
100 mm focal length ZnSe lens 25.4 mm diameter
Edmunds Optics
50 mm focal length ZnSe lens 25 mm diameter
Edmunds Optics
Temperature-stabilized
tube (TE cooler)
Laser
Pointer
Foam
Insulation
Temperature-stabilized
tube (TE cooler)
Temperature-stabilized
tube (TE cooler)
Foam
Insulation
Temperature-stabilized
tube (TE cooler)
Schematic of NDIR sensor
Tungsten lamp (5 V 0.06 A) with lifetime of 100,000 hours (11 years of continuous operation) Pulsed with sine-wave signal from a function generator at 3.9 Hz
Copper, straight- tubing light pipe (0.5 inch diameter) 60 cm long
CaF lenses inside a temperature-stabilized tube
One of two filters at 3.9 um (off) or at 4.26 um (on)
ZnSe windowed TE-cooled pyroelectric detector with preamplifier Lockin detection system used
Sealed using sapphire windows at both end of tube
Upper view of setup
Light-pipe effect of tubing
Regular copper straight pipe works well Will also use ultra-polished SS tube and Au-coated Cu pipe
Comparisons to Picarro cavity ring-down spectroscopy (CRDS) instrument
Picarro CRDS 0.05 ppm resolution 0.5 ppm drift over 1 month $30k/unit NDIR sensor 0.15 ppm resolution 1.0 ppm over 3 months (TBD) $2k/unit (TBD)
Off-resonance signal stability is 0.2 ppm of CO2 for hours
Stability of off-resonance signal at 3.6 um
0 20 40 60 80 100
0.09626
0.09627
0.09628
0.09629
0.09630
0.09631
NDIR 7 Hz off resonance 3_6 um.opj
Sig
nal
(V
)
Time (min)
50 uV or 1 ppm
NDIR CRDS comparison 1 (30 hour duration)
5825 5830 5835 5840 5845 5850 5855
360
370
380
390
400
410
28 30 32 34 36 38 40 42 44 46 48 50 52 54 56 58
-0.0705
-0.0700
-0.0695
-0.0690
-0.0685
-0.0680
Pic
arro
, C
O2 c
onc.
ppm
Time, hours
Picarro
ART NDIR Picarro comparison with Sapphire windows 60 cm tube heated to 40 deg C and N2 purge 09_01_19.opj
ND
IR
NDIRScaling NDIR vertical axis to have it overlay on Picarro (using Picarro to calibrate the NDIR)
Short-term instability in the Picarro signal due to fluctuations in nitrogen gas purge
Deviation in NDIR due to system drift of ~2 ppm in 12 hours
NDIR CRDS comparison 2 5874 5876 5878 5880 5882 5884 5886 5888
370
380
390
400
410
420
430
76 78 80 82 84 86 88 90 92
-0.0702
-0.0700
-0.0698
-0.0696
-0.0694
-0.0692
-0.0690
-0.0688
-0.0686
-0.0684
-0.0682
-0.0680
-0.0678
-0.0676
pic
arro
picarro
nd
ir
Time, hours
ndir
NDIR sub ppm stability and resolution
5906 5907 5908 5909 5910 5911 5912
372
374
376
378
380
382
384
386
5906 5907 5908 5909 5910 5911 5912
-0.0698
-0.0697
-0.0696
-0.0695
-0.0694
-0.0693
-0.0692P
icar
ro,
CO
2 c
on
c. p
pm
Picarro
ND
IR
Time, hours
NDIR
Constructing a dual-band system
4.26 um 180 nm 3.06 um 160 nm
Lessons learned
Signal of NDIR depends upon 1. Internal set temperatures (stabilize
temperatures of both tube and detector assemblies)
2. Detector responsivity (use interference filters on sapphire substrates and protected pyroelectric detectors)
3. Source stability (use long-life, low-power lamps)
4. Source frequency (use a function generator as a source with sine-wave output)
5. Light-pipe stability of transmittance (use gold-coated Cu pipe or SS pipe)
6. Humidity (condition gas before analysis)
1. NDIR sensors can be improved with better pyroelectric detectors and an actively stabilized detector compartment
2. The improved designs can be competitive with CRDS over short periods (12 hours)
3. The improved design can measure CO2 with 0.2 ppm noise
4. We expect to achieve 1 ppm stability over 3 months with the dual-band detector and a thermally stable and structurally stable setup
5. Other gas species between 1 um to 14 um can be measured with lower uncertainties with the improved design
Conclusion