Improving Homogeneity in Breast Radiotherapy Plans
Maria CzerminskaYulia Lyatskaya
![Page 2: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/2.jpg)
Breast Plans
![Page 3: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/3.jpg)
Why homogeneity Matters?
Whole breast irradiation often leads to both acute and long term toxicities such as moist desquamation, pain, breast discomfort and breast hardnessMany studies shown that toxicities were associated with dose inhomogeneity (hot spots)
![Page 4: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/4.jpg)
Randomized trialsPignol et al358 patients were randomized in a multicenter double-blind clinical trial to either 2-dimentional treatment planning or IMRT planning with improved dose homogeneity. The incidence of moist desquamation in the IMRT group was 31.2% vs 47.8%, p=0.002
• Pignol JP, Olivotto I, Rakovitch E, et al. A multicenter randomized trial of breast intensity-modulated radiation therapy to reduce acute radiation dermatitis. J Clin Oncol. 2008 May 1;26(13):2085-92.
C
![Page 5: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/5.jpg)
Randomized Trials
Donovan et al306 patients were randomized to 2D or 3D IMRT. After 5 years 240 patients data was available for analysis. The 2D arm patients were 1.7 times more likely to have changes in breast appearance than IMRT group
. Donovan E, Bleakley N, Denholm E, et al. Randomised trial of standard 2D radiotherapy (RT) versus intensity modulated radiotherapy (IMRT) in patients prescribed breast radiotherapy. Radiother Oncol.2007 Mar;82(3):254-64.
RM
![Page 6: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/6.jpg)
Advantage of Dose Homogeneity
Breast V105% and V110% were significantly associated with increase in acute skin toxicity
V110% < 200cc: 31% grade >2 skin toxicityV110% > 200cc: 61% grade >2 skin toxicity
Vicini et.al. Int.J. Radiat Oncol Biol Phys 54: 1336-1344; 2002.
WB
![Page 7: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/7.jpg)
HarsoliaHarsolia et.alet.al. . Int.JInt.J. . RadiatRadiat OncolOncol BiolBiol PhysPhys 68: 137568: 1375--1380; 20071380; 2007
“The use of IMRT in the treatment of the whole breast results in a significant decrease in acute dermatitis, edema, and hyperpigmentation and a reduction in the development of chronic breast edema compared with conventional wedge-based RT.”
WB
![Page 8: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/8.jpg)
IMRT for breast treatment
What kind of IMRT?
Many beams: do we want many beams or two is enough?
Many beams with different angles may help with dose conformality, but will lead to higher doses in lung, heart and contralateral breastTangential beams provide best lung, heart and contralateral breast sparing
![Page 9: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/9.jpg)
Breast IMRT Planning
How to achieve optimal dose distribution?
Through complex fluence map optimization
![Page 10: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/10.jpg)
Fluence Optimization: Two Types of IMRT
Inverse plan IMRTFluence optimization is performed by planning software Commercial algorithms are not available for breast
Tangential beams onlyIncluding flash
Forward plan IMRTFluence optimization is performed by using field-in-field approach (iterative process of manual fluence optimization)More transparent, easier to understandMore appropriate for breast planning with 2-beam approach
![Page 11: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/11.jpg)
Forward vs Inverse IMRT
Inverse plan:Better conformality (+)High dose gradient (+)Larger volume of small dose due to multiple beam angles (-)Complex planning (-)
Forward plan:Conformality improvements are limited (-)Volume of small dose is not expanded (+)Dose homogeneity is improved (+)Transparent planning (+)
![Page 12: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/12.jpg)
IMRT approaches for breast
Hybrid techniques
Forward planned IMRT
![Page 13: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/13.jpg)
Hybrid Techniques
Mayo et al. (UMAss)
Two tangents and two IP-IMRT beams
Descovich et al. (UCSF)
Two tangents and two direct aperture optimized (DAO) IMRT beams
Faster planning, less dependent on planner’s ability, but need to contour the breast tissue
![Page 14: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/14.jpg)
Forward Planned Breast IMRT:
Memorial Sloan-Kettering Cancer CenterFox Chase HospitalWilliam Beaumont HospitalOther…
![Page 15: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/15.jpg)
C. Chui , L. Hong, and M. Hunt, et al. A simplified intensity moC. Chui , L. Hong, and M. Hunt, et al. A simplified intensity modulated radiation therapy technique for the breast, Med Phys 29 dulated radiation therapy technique for the breast, Med Phys 29 (2002), pp. 522(2002), pp. 522––529.529.
Simplified IMRT technique:
•Two tangential fields
•Midpoint of the segment intersecting the treatment volume for each pencil beam is determined
•The dose from the open field to this point is calculated
•The intensity of the pencil beam is determined as proportional to the inverse of the open field dose
Memorial Sloan-Kettering Cancer Center Technique
![Page 16: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/16.jpg)
Fox Chase Technique
Beamlets optimized based on the dose at the midplane
Monte Carlo method used for dose calculation
![Page 17: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/17.jpg)
William Beaumont Technique
MLC segments constructed based on BEV projections
MLCs conformed to isodosesurfaces in 5% increments (from 120% to 100%)
Additional MLC segments conforming to lung tissue were added to reduce transmission
“script” program was created within the planningsystem to automatically optimize the weights of theindividual sMLC segments.
![Page 18: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/18.jpg)
IMRT for breast treatment
Forward planning
Two tangential open fields and multiple subfields to achieve desired homogeneity
![Page 19: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/19.jpg)
Planning Strategy
- An open beam configuration is first calculated and evaluated.
- 4+ subfields per gantry angle are used to produce an optimal breast plan.
- No wedges.
![Page 20: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/20.jpg)
Subfields
- Generally have 1 lung block and 3 additional subfields per gantry angle
- Lung block is formed by fitting the MLC’s to the shape of the lung. Aids in lateral hot spots.
- Additional subfields are generated by manually fitting MLC’s to “hot” areas. Ex. 115%, 110%, etc…
![Page 21: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/21.jpg)
SubfieldsSub 1 Sub 2
Sub 3 Sub 4
Open 115%
110% 105%
![Page 22: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/22.jpg)
Weighting of Subfields
- Generally, the open beam portion receives ~ 80% of the dose while the subfields contribute ~20%.
- This makes FP IMRT similar to conventional treatment
- Minimizes effects of patient movement on target coverage
![Page 23: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/23.jpg)
Normalization
- Conventional breast plans are generally normalized to 97%
- Normalization for IMRT plans are based on coverage
![Page 24: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/24.jpg)
Alternative Strategy
![Page 25: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/25.jpg)
Create initial subfield; weight until just before 100% compromised.
Temporarily increase weight to produce 5 additional MUs.
Create opposing subfield. Conform MLCsto remaining 100% dose cloud.
Restore weighting of previous field by removing 5 MUs. Weight new field until just before 100% compromised.
![Page 26: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/26.jpg)
How Many Subfields
Patient Characteristics That Predict the Need for Intensity Modulated Radiation Therapy (IMRT) of Breast
C. Harrington, Y. Lyatskaya, M.Czerminska, J. Harris, R. CormackDana-Farber /Brigham and Women’s Cancer Center
![Page 27: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/27.jpg)
The plans of 100 patients were analyzed.
Variable values were (median [min- max]):•Separation: (20.9 [14.7-31.9] cm)•Size: (8 [4.0-15.3] cm)•Volume: ( 1660 [676-4151] cc )•Maximum open field dose: (116 [110.3-126.3] %)•Number of subfields: ( 6 [4-10])
Results
![Page 28: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/28.jpg)
One third of patients required seven or more subfields to achieve the planning method’s dose homogeneity goal.
A maximum open field dose over 120% predicts necessity for 8 or more subfields with high (0.9) specificity and selectivity
Results
![Page 29: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/29.jpg)
Can we always achieve a perfect homogeneity?
Two Tangents with 4 subfields from each side
![Page 30: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/30.jpg)
Too hot!Unacceptable hot spots
![Page 31: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/31.jpg)
Patients with Large Separations
31.7 cm
29.6 cm
![Page 32: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/32.jpg)
Oblique Fields TechniqueTangential field border
Oblique field order
![Page 33: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/33.jpg)
Oblique field Technique
![Page 34: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/34.jpg)
Improved homogeneityGlobal Dmax=110.5%
![Page 35: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/35.jpg)
1
2
3
4
![Page 36: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/36.jpg)
Coronal
![Page 37: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/37.jpg)
Sagittal
![Page 38: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/38.jpg)
Lung and Heart DVH
Lung
Heart Tangents
Tangents+obliques
![Page 39: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/39.jpg)
More reasons to improve homogeneity
Hypo-fractionation
•Canadian 266 cGy x 16
•Danish 267 cGy x 15
•British (RMH) 300 cGy x 13
![Page 40: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/40.jpg)
Fractionation
Randomized Trial of Breast Irradiation Schedules AfterLumpectomy for Women With Lymph Node-Negative
Breast CancerTimothy Whelan, Robert MacKenzie, Jim Julian, Mark Levine, Wendy Shelley,Laval Grimard, Barbara Lada, Himu Lukka, Francisco Perera, Anthony Fyles,
Ethan Laukkanen, Sunil Gulavita, Veronique Benk, Barbara Szechtman
Canadian Randomized Trial
Journal of the National Cancer Institute, Vol. 94, No. 15, August 7, 2002
![Page 41: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/41.jpg)
Canadian Study
Randomized trial with 1234 patients.
Women with invasive breast cancer who were treated by
lumpectomy and had pathologically clear resection margins
and negative axillary lymph nodes were randomly assigned
to receive whole breast irradiation of:
42.5 Gy in 16 fractions over 22 days (short arm)
50 Gy in 25 fractions over 35 days (long arm)
Whelan et. Al. Journal of the National Cancer Institute, Vol. 94, No. 15, August 7, 2002
![Page 42: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/42.jpg)
No difference in cosmetic outcome after 3 and 5 years
No difference in late radiation toxicitiesNo difference in disease free survival and
overall survival
Whelan et. Al. Journal of the National Cancer Institute, Vol. 94, No. 15, August 7, 2002
Results
% of Patients with no Toxicity
![Page 43: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/43.jpg)
Canadian Study
Conclusion:
•“Our results support the use of a shorter fractionation schedule for irradiation of women with lymph node-negative breast cancer treated by lumpectomy.
•The results are not applicable to women with very large breasts (i.e., with widths >25 cm)”
Whelan et. Al. Journal of the National Cancer Institute, Vol. 94, No. 15, August 7, 2002
![Page 44: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/44.jpg)
Canadian Fractionation at our Institution
266 cGy x 16 fractionsPatients over 60 with clear marginsNon-high grade diseaseNo chemoNegative nodesSmall separationHot spots<107%
![Page 45: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/45.jpg)
Example 1
•Right sided
•Separation 21.3 cm
•Energy 10 MV
•Tangents with subfields
![Page 46: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/46.jpg)
Axial
104%
![Page 47: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/47.jpg)
Coronal
![Page 48: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/48.jpg)
Sagittal
![Page 49: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/49.jpg)
Subfields
Medial
Lateral
![Page 50: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/50.jpg)
Example 2
•Right sided
•Separation 26.6 cm
•Energy 10 MV
•Tangents with subfields and oblique fields
![Page 51: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/51.jpg)
Oblique fields
![Page 52: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/52.jpg)
Dose Distribution
Global Hot Spot 107.2%
![Page 53: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/53.jpg)
DFCI/BWH breast cases with Canadian Fractionation in 2012
Patient Separation[cm]
Technique Energy [MV]
Hot Spot [%]
1 22.3 Tang+obliques 10 107.5 2 16.5 Tangents 6/10 108 3 21.5 Tang+obliques 10 107 4 26.6 Tang+obliques 10 107.2 5 24 Tang+obliques 10 106.3 6 21.5 Tangents 10 106.5 7 23.4 Tang+obliques 6/10 107.2 8 20.2 Tangents 6 107.6 9 17.3 Tangents 6 106.5 10 21.6 Tang+obliques 10 106.7 11 21.3 Tangents 10 105.9
![Page 54: Improving Homogeneity in Breast Radiotherapy Planschapter.aapm.org/NE/DOCUMENTS/Presentations/2012... · Randomized trials Pignol et al 358 patients were randomized in a multicenter](https://reader034.fdocuments.us/reader034/viewer/2022051717/5a6ff6ca7f8b9aac538b7db9/html5/thumbnails/54.jpg)
AcknowledgementsDana Farber/Brigham and Women’s
Treatment Planning Team:Cindy HancoxJohann BrownWee Pin Yeo
Nicholas CristianiBo Li
Tony OrlinaPhilip Selesnick
Coleman McDonoughSandra FranklinAntonio DamatoEmily NeubauerMandar Bhagwat