Biometry for Premium IOLs 2016
Jack T. Holladay MD MSEE FACS Page 1 of 6
4/10/2016 JTH 1
Improvement Strategies: Biometry,
Topography, Power Calcs to
Ref Surprises
Jack T. Holladay, MD, MSEE, FACSClinical Professor of Ophthalmology
Baylor College of MedicineHouston, Tx 4/10/2016 JTH 2
The Promise of No Glasses
or Contact Lenses !
4/10/2016 JTH 7
Requirements
Accurate Biometry – Optical
(IOL Master or LenStar)
Accurate K’s- Repeatable
4th Generation Formula (WTW)
Personalized Lens Constant
Eliminate Corneal Astigmatism4/10/2016 JTH 8
4/10/2016 JTH 9
Subtractfrom Ascanmeasured
Axial Length~ 0.8 mm
Zeiss - IOL Master - 2000Zaldivar-Holladay JCRS May 2000
4/10/2016 JTH 10
J Cataract Refract Surg 2011; 37:2018–2027
Biometry for Premium IOLs 2016
Jack T. Holladay MD MSEE FACS Page 2 of 6
4/10/2016 JTH 11
Linear Regression to compensate for AVERAGE Index of Refraction in Long Eyes
Measured 36 34 mm
4/10/2016 JTH 12
Requirements
Accurate Biometry – IOL Master
Accurate K’s- Repeatable
4th Generation Formula (WTW)
Personalized Lens Constant
Eliminate Corneal Astigmatism
4/10/2016 JTH 13
IOL Power Calculations Pentacam can measure FRONT &
BACK SURFACE POWER
Can Calculate: Equivalent K-Reading (EKR)
65% Mean, Peak & Average
NET POWER4/10/2016 JTH 14
EKR Reports Keratometry value but adjusts
for Back Surface Power from Normal (Current IOL Formulas)
If corneal front surface is 7.5 mm (45 D), but if back surface -0.3 D > normal:
EKR = 45.0 – 0.3 = 44.7 D
Note: Net Power = 43.3 D
4/10/2016 JTH 15Holladay Report Equivalent K-Reading
EKR
4/10/2016 JTH 16
Biometry for Premium IOLs 2016
Jack T. Holladay MD MSEE FACS Page 3 of 6
4/10/2016 JTH 23
IOL Calcs – Abnormal Cornea(Use 65% MEAN EKR) Post Refractive Surgery
Post PKP
Keratoconus
Corneal Scar
Any Irregular Astigmatism4/10/2016 JTH 24
Use 65% Mean EKR( @ 4.5, 4 & 3 mm zones)
4/10/2016 JTH 25
Normal LASIK RK41 to 44 D3 D Range
36 to 41 D5 D Range
32 to 45 D13 D Range
4/10/2016 JTH 26
POST LASIK
4/10/2016 JTH 27
Post LASIK CALC Kmean = 39.8 D
Used 39.8 D => SEQ = +1.12 D
(+1.00 + 0.25 X 155 = 20/20)
65% mean = 38.8 D => +0.12 D
Use 65% mean K4/10/2016 JTH 30
Conclusions EKR – Use 65% Mean for all
IOL Calcs
Look @ smaller zones than 4.5 mm if pupil very small
(< 3.0 mm in dim light)
Biometry for Premium IOLs 2016
Jack T. Holladay MD MSEE FACS Page 4 of 6
4/10/2016 JTH 31
Requirements
Accurate Biometry – IOL Master
Accurate K’s- Repeatable
4th Generation Formula (WTW)
Personalized Lens Constant
Eliminate Corneal Astigmatism4/10/2016 JTH 32
Vergence Formula
IOLAL ELP
DPostRxV
K(Post R)ELP
1336 13361336
10001000
4/10/2016 JTH 33 4/10/2016 JTH 34
CONCLUSION: 9 EYESA
nter
ior
Seg
men
t Siz
e
Axial LengthShort Normal Long
Large
Normal
Small
Megalocornea Large Eye
+ axial Megalocornea Buphthalmos
hyperopia Megalocornea
+ axial myopia
axial hyperopia normal axial myopia
Small eye Microcornea
Nanophthalmia Microcornea + axial mopia
(20%)
(10%)(2%)
(2%)
(80%) (96%) (90%)
(0%)
(0%)
4/10/2016 JTH 35
Measurements taken forPredictors of ELP
Axial Length
Average K (Pre Ref)
Horizontal WTW
ACD
LT
Pre-opRefraction
Age4/10/2016 JTH 36
FORMULA PERFORMANCE
15 20 25 300
12
34
56
Axial Length (mm)Mea
n A
bsol
ute
Err
or (
D)
Holladay 1Hoffer QSRK/THolladay 2
N = 997
Biometry for Premium IOLs 2016
Jack T. Holladay MD MSEE FACS Page 5 of 6
4/10/2016 JTH 37
Requirements
Accurate Biometry – IOL Master
Accurate K’s- Repeatable
4th Generation Formula (WTW)
Personalized Lens Constant
Eliminate Corneal Astigmatism4/10/2016 JTH 38
Personalized Lens Constant Never use Manufacturer’s Constant
except to start 20 to 40 cases and continue Factors IOL Style Lens placement Post op medications Biometer, keratometer, …
4/10/2016 JTH 62
Requirements
Accurate Biometry – IOL Master
Accurate K’s- Repeatable
4th Generation Formula (WTW)
Personalized Lens Constant
Eliminate Corneal Astigmatism
TORIC IOL Calculations Commercial Calculators use a constant ratio (1.46) for
the corneal cylinder to the IOL cylinder Exact Calculation depends on IOL SEQ Power and ELP
… to correct 2D of corneal astigmatism 10 D IOL => 3.5 D Cylinder 22 D IOL => 2.9 D Cylinder 34 D IOL => 2.4 D Cylinder
A 1.1 D difference from 10 D to 34 D!
4/10/2016 JTH 65 4/10/2016 JTH 66
Toric Optimization
Biometry for Premium IOLs 2016
Jack T. Holladay MD MSEE FACS Page 6 of 6
4/10/2016 JTH 67 4/10/2016 JTH 68
PREOP 6 D Toric IOLLEF
4/10/2016 JTH 69
PREOP 6 D Toric IOL -- ODLEF
4/10/2016 JTH 70
PREOP 6 D Toric IOL -- OSLEF
4/10/2016 JTH 71
!Thank You!