Lyra MP-Reader, Centrifuge and Reader for Microplates
Ordering InformationCatalog No. Description
854002 Lyra MP-Reader
009980 PC Unit Lyra MP-ReaderCentrifuge and Reader for Microplates
Capacity
2 microplates
Performance
Centrifuge time 80 secAgitation time (2 phases) 23 sec/100 secReading time (2 microplates) 20 secCycle time (identification, centrifugation, agitation and reading of 2 microplates) <7 min
Technical Data (Centrifuge)
Speed centrifuge 1,000 rpm Speed centrifuge tolerance ± 5Speed agitation 940-310 rpmSpeed agitation tolerance ± 10RTF 101 g
Electrical Data
Voltage 100-250 VAC Power fluctuation ± 10%Transient over-voltages category II Frequency 50-60 HzPower consumption max 110 VA
Dimensions
Width 590 mmHeight 430 mmDepth 620 mmWeight 38 kg
Environmental Conditions
Operating temperature +5°C to +55°C(41°F to 131°F)
Operating humidity, non-condensing 5% rh to 80% rhAltitude (max) 2,000 m
Main Features
Positive identification
CCD camera Complete traceability of tests, results and operator
Bi-directional CE compliant, according to IVD Directive 98/79/EC
Little bench space needed
Bio-Rad Laboratories I M M U N O H E M A T O L O G Y
The Complete Solution for Safe Transfusion
Centrifugation, agitation, reading and interpretation
Clinical Diagnostics Group
DiaMed1785 Cressier FR Switzerlandwww.diamed.com
© Bio-Rad H010123 06/2010
Bio-Rad Laboratories is a multinational manufacturer and distributorof clinical diagnostics and life science research products.
Printed in France with soy-based inks on paper from sustainably managed forests
The Lyra MP-Reader integrates three functions in a singleinstrument: centrifugation, agitation and reading. It is theoptimal instrument if you require smart processing of yourmicroplates.
Thanks to its multi-functionality, Lyra will make your manual work easier and provide you with higher securityand traceability. This compact instrument can be usedconnected to a pipetting automate or as an extension of an existing system.
Lyra MP-Reader
Bio-Rad Laboratories I M M U N O H E M A T O L O G Y
Supreme Security
• The unique barcode on each microplate allows fulltraceability, including patient identification, test protocolsperformed, reading and interpretation of each microplate,and also prevents errors such as the use of incorrect or expired microplates.
Objective and Standardized
• The proven Lyra software guarantees objective andstandardized reading and interpretation independent of the user assessment.
Flexible Data Management
• Through the use of the integrated software, a result/interpretation can be displayed, printed and sent to a host computer.
• The database allows you to archive data, group or searchit using different criteria, thereby ensuring the integrity of allocated results.
Saving Time is Simple
• Simply load 2 microplates and start the cycle.Centrifugation, agitation and reading of the microplateswill be performed automatically. Results of all 192 wellsare available within less than 7 minutes.
Customize or Extend
• Lyra can be connected to any ID- / MP-Pipettor, or be used stand-alone. This enables you to customizeyour laboratory according to your current and futurerequirements.
Validation and Control
• A high resolution CCD camera guarantees clear imagesthat a user can examine in order to validate or evenmodify* results and interpretations reported by Lyra.
* Depending on the authorization level with audit train in database.
Centrifuge and Reader for Microplates
Lyra MP-ReaderCentrifuge and Reader for Microplates
1 Sample identification or patient name
2 Test code and test name
3 Barcode and type of microplate
4 Picture of microplate-wells
5 Result of each reaction*
6 Interpretation of result*
7 Position of test on microplate
8 Confirm result/interpretation with one click
9 Enter operator remarks
* Result/interpretation by Lyra. If authorized, the user can modify this (audit trail of modification in database).
321
4
7
5
6
8 9
Microplate carousel
Patient results
Loading of a microplate