IHE Radiology Mammography Acquisition Workflow - from a RIS perspective
Antje Schroeder – IHE Radiology Technical Committee Co-Chair (Siemens Healthcare)July 25, 2013
Background• Mammography Image Profile has been well accepted in the
Mammography field– Consistent creation and display of 2D Mammography Images
• Mammography Acquisition Workflow has been developed to address specific workflow needs/exception in screening Mammography– Rejection of images for quality reasons to eliminate low quality images from
review– Correction of image view and laterality to avoid additional exposures or
incorrect reporting– Procedure Updates performed at the modality. • Profile has been well accepted with Modality and PACS vendors,
however so far there is a lack of RIS support
Mammography Acquisition Workflow • Use cases
• Supplemental Views• Finish exam with supplemental views in a different room• Conversion of a procedure at modality• Screening recall – Patient returns for subsequent exam• Screening recall – Technical reasons• Mark image rejected for quality reasons• View correction
• Use of coded values for scheduling and documentation purposes
RIS is essential to allow complete workflow and a consistent display at review station
Value Proposition
• Reduce Errors and enhance Patient Care • Prevents additional exposure due to incorrect labeling • Prevents review of bad quality images• Prevents delays in patient care by only making relevant
information available at the point of care. • Improve Throughput
• Reduces staff time wasted identifying and correcting errors (in a coordinated fashion) among the HIS, RIS, PACS and modality
• Re-use functionality defined in other profiles
Actors and Transactions
Relationship to other Profiles• Scheduled Workflow
• Basic Workflow model of SWF is extended:– Handling of exceptions in acquisition workflow (additional image
acquisition, procedure updates at modality, changes of acquisition system, procedure conversion)
– Use of Protocol and Reason codes to drive workflow – Handling for varying scheduled and performing procedure codes– Procedure updates in order to keep participating actors aligned to allow
proper display of modified studies• Mammography Image• Consistent creation and display of Mammography Images
• Image Object Change Management• Re-use mechanism (Rad-66 – Image Rejection Note Stored) to
– flag bad quality images in order to prevent display – Correct incorrectly labeled images
• Key Image Note• Re-use of manifest (Key Object Selection Document)
Convert procedure
Convert diagnostic procedure: unilateral bilateral• Radiologist reviews images of left breast while patient still present.• Radiologist decides that images of right breast are needed, too.
Change procedure directly at Modality, no canceling/rescheduling at RIS necessary Based on MPPS RIS detects procedure change and updates Image Manager/
Archive and Order Placer to use correct information for subsequent workflow steps.
Modality Archive WorkstationDSS/OF
Store images of R breast
MPPS in progressMPPS completed
Query/ Retrieve imagesfor bilateral diagnostic exam
Procedure update: unilateral bilateralMPPSforwarded
Finish acquisition in different acquisition room
Supplemental images for one procedure acquired at a second Modality• Radiologist decides to take additional images / convert a procedure, while patient
is waiting outside acquisition room• Original room is occupied, so the acquisition continues in different room
DSS/OFF needs to provide worklist for either a COMPLETED or a DISCONTINUED exam
Avoid unscheduled acquisition with manual reconciliation
Archive WorkstationDSS/ OF
Query/ Retrieveimages from both Modalitiesas part of the same Study
Modality 2Store images
MPPS in progressMPPS completed
MPPS forwardedQuery MWL
Query MWL(patient-based)
MPPS in progress/ completedStore images
Modality 1
Recall – patient returns for subsequent exam
Screening recall• Radiologist assesses acquired images incomplete (BIRADS 0).• Patient has left and needs to come back for additional imaging.
DSS/OF schedules a new procedure (as part of the screening order) with– a specific Reason for Requested Procedure code value– a reference to the original study (Referenced Study Sequence)
link between the original and the recall study Regular acquisition at Modality Image Display presents both studies as one based on reason code and study
reference
MPPS forwarded
Modality Archive WorkstationDSS/ OF
Store images (recall study)
MPPS
Query/ Retrieveimages from original and recall study
Query MWL (on recall)
Mammography Acquisition WorkflowUse of codes: DSS/OF, Modality, Archive
• Use of coded values for Scheduling and driving modality behavior:• RIS provides Requested Procedure, Reason for Requested procedure and
scheduled protocols to enable selection of modality settings and support technologies
Documentation what was performed:• enable displays to present images in adequate layout, • enable tracking of performed work and compose charges
RIS needs to provide this information to enable all subsequent workflow steps
Coding Scheme Code Value* Code Meaning Definition
DCM 111590 Recall for technical reasons Patient returns for additional images to improve the quality of the current exam.
DCM 111591 Recall for imaging findings Patient returns for additional images to clarify findings from the current exam.
DCM 111416 Follow-up at short interval from prior study
The prior study recommended a follow-up breast imaging exam in 1 to 11 months (generally in 6 months).
SRT R-408C3 Diagnostic The exam is done in order to answer a specific diagnostic question.
* Note: This table is not shown completely (see MAWF TI Profile text: Table 4.5-6)
What does this mean for the RIS
• RIS is an important workflow driver in MAWF:• Provide procedure codes to drive subsequent acquisition and
documentation steps• Detect and communicate procedure changes• Schedule recall-studies with references to initial study and
appropriate recall reason codes• Provide work items already in use
Next Step
• Review Trial Implementation version of MAWF:– http://www.ihe.net/Technical_Framework/
index.cfm#radiology• Sign-up to participate in IHE Connectathon in
January 2014
Thank you
Modality or QA workstation Archive WorkstationDSS/ OF
Store rejection KOS
MPPS (Modality) orCreator PPS
PPS forwarded
B) Rejected images are hidden or displayed
A) Rejected images are hiddenB) Rejected images are provided
AvailabilityNotification
A) Rejected images not part of query result
Q/R
Reject images of insufficient quality
Mark images as rejected for quality reasons• After completed acquisition, images are considered of insufficient quality• Modality or Evidence Creator create a Key Object Selection (KOS) object
– Document Title “Rejected for Quality Reasons”– Document Title Modifier is valued (DICOM PS 3.16, CID 7011)– References to all rejected images
• Image Archive is configurable to provide or hide rejected images and KOS• Image Display is configurable to present or hide rejected images and KOS
Correct view labels of images
View, laterality or protocol are corrected• After completed acquisition, incorrect view labeling is detected • Modality or Evidence Creator create corrected images
– Copy most values except e.g. View Codes, View Modifiers, Laterality, Protocol– Correct references from “For Presentation” to “For Processing” images– Evidence Creator records its identifying data (Contributing Equipment Seq.)
• Modality or Evidence Creator create a Key Object Selection (KOS) object– Document Title “Rejected for Patient Safety Reasons” ( DICOM CP)– References all incorrect images (For Presentation & For Processing)
• Image Archive & Image Display hide KOS and the incorrect images
PPS forwarded
Modality or QA workstation Archive WorkstationDSS/ OF
Store rejection KOSStore corrected image(s)
MPPS (Modality) orCreator PPS
Q/R: incorrect imagesare not part of query resultIncorrect images
are hidden at Q/R
AvailabilityNotification
Supplemental views to a completed exam
Supplemental implant displacement image after screening• Append case, where Modality fills protocol codes in a new MPPS.
Modality supports Assisted Acquisition Protocol Setting.– Procedure unchanged, e.g. “Screening bilateral”Note: Compared to SWF, Modality always fills in protocol code(s).
• Dept. Scheduler/ Order Filler may use protocol codes, e.g. notify user• Image Display selects presentation layout based on procedure/ protocol/ reason
codes
Modality Archive WorkstationDSS/ OF
Store suppl. images
MPPS in progressMPPS completed
MPPS forwarded
Query/ Retrievescreening and implant images
Actors and Transactions