IAEA code of practice for dosimetry in diagnostic radiology:
recommendations for hospitals and specific issues
Antti KosunenRadiation Metrology Laboratory-STUK
Radiation safety and dosimetry in medical applications
11‐12 May 2012
TTÜ, Ehitajate tee 5, Tallinn, Estonia
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 210.5.2012 / AK-STUK
• Review /introduction to contents of IAEA TRS 457 for clinical dosimetry
• Experiences of use the IAEA TRS 457(KAP meters, others)
• Remarks based on work at the SSDL of STUK
• Methods by TLD are not reviewed
Presentation
![Page 2: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for](https://reader030.fdocuments.us/reader030/viewer/2022011812/5e238a4f294c690a8d3a014a/html5/thumbnails/2.jpg)
12.05.2012
2
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 310.5.2012 / AK-STUK
http://www-pub.iaea.org/MTCD/publications/
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 410.5.2012 / AK-STUK
Dosimetry formalismDosimetry intrumentationQuantities and uncertainties
• General radiography• Fluoroscopy• Mammography• Computed tomography• Dental radiography
IAEA TRS 457, Clinical dosimetry
![Page 3: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for](https://reader030.fdocuments.us/reader030/viewer/2022011812/5e238a4f294c690a8d3a014a/html5/thumbnails/3.jpg)
12.05.2012
3
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 510.5.2012 / AK-STUK
• Dosimetry• phantom• patient
• Guidance for selection of patients for DRL
•Appendix VI. Application of patient dose measurements• Background• Guidance for DRLs• Guidance for organ dose conversion factors for examinations• Summary of using effective dose
IAEA TRS 457, Clinical dosimetry
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 610.5.2012 / AK-STUK
• Assumed that the x- ray equipment is properly maintainedaccording to QA program
• Assumed that the x ray tube output is linear relative to tube loading,if not complementary measurements are required
• HVLFor interpolations of calibration coefficientsFor conversions to organ doses (mammography)
Not necessrerily for every dose measurement, as information from Quality Control results exist
IAEA TRS 457, Clinical dosimetry
![Page 4: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for](https://reader030.fdocuments.us/reader030/viewer/2022011812/5e238a4f294c690a8d3a014a/html5/thumbnails/4.jpg)
12.05.2012
4
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 710.5.2012 / AK-STUK
• Examples and worksheets for measurements
• Dosimeters should be calibrated at the laboratory• Calibration interval:
• Based on national requirements,• Usually should not exceed two years
IAEA TRS 457, Clinical dosimetry
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 810.5.2012 / AK-STUK
( 2sd )
• Risk assessment : Absolute risk • Dose surveys
~20 %
• Relative risk• Interventional procedures • Quality assurance
measurements• Paediatric procedures
7 %
IAEA TRS 457, Accuracy requirements for dosimetryAt reference conditions of dosimetry
![Page 5: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for](https://reader030.fdocuments.us/reader030/viewer/2022011812/5e238a4f294c690a8d3a014a/html5/thumbnails/5.jpg)
12.05.2012
5
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 910.5.2012 / AK-STUK
Air kerma based formalism
IAEA TRS 457, Dosimetry formalism
Beam quality correction factor
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 1010.5.2012 / AK-STUK
IAEA TRS 457, Selection of instruments, clinical
Requirements of International Standards are followed:
Requirements for dosemeters at laboratory can be used also as a background information for clinical dosemeters
![Page 6: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for](https://reader030.fdocuments.us/reader030/viewer/2022011812/5e238a4f294c690a8d3a014a/html5/thumbnails/6.jpg)
12.05.2012
6
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 1110.5.2012 / AK-STUK
IAEA TRS 457, Measurement uncertainty (accuracy)
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 1210.5.2012 / AK-STUK
IAEA TRS 457, General radiology, dosimetry quantities
![Page 7: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for](https://reader030.fdocuments.us/reader030/viewer/2022011812/5e238a4f294c690a8d3a014a/html5/thumbnails/7.jpg)
12.05.2012
7
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 1310.5.2012 / AK-STUK
IAEA TRS 457, General radiology
Incident air kerma
Air kerma detector with isotropic response
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 1410.5.2012 / AK-STUK
IAEA TRS 457, General radiology, phantoms
![Page 8: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for](https://reader030.fdocuments.us/reader030/viewer/2022011812/5e238a4f294c690a8d3a014a/html5/thumbnails/8.jpg)
12.05.2012
8
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 1510.5.2012 / AK-STUK
IAEA TRS 457, Review on phantoms
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 1610.5.2012 / AK-STUK
IAEA TRS 457, General radiology, uncertainty
![Page 9: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for](https://reader030.fdocuments.us/reader030/viewer/2022011812/5e238a4f294c690a8d3a014a/html5/thumbnails/9.jpg)
12.05.2012
9
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 1710.5.2012 / AK-STUK
IAEA TRS 457, Fluoroscopy, dosimetry quantities
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 1810.5.2012 / AK-STUK
IAEA TRS 457, Fluoroscopy
•Anti- scatter grid in position if used•Focus to table top distance as used clinically•Phantom 20 cm, 30 cm for large patients•Collimation large, to phantom size•Exposure under ABC
Detector with back scatter measurement capability
![Page 10: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for](https://reader030.fdocuments.us/reader030/viewer/2022011812/5e238a4f294c690a8d3a014a/html5/thumbnails/10.jpg)
12.05.2012
10
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 1910.5.2012 / AK-STUK
IAEA TRS 457, Fluoroscopy
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 2010.5.2012 / AK-STUK
IAEA TRS 457, Fluoroscopy, uncertainties
![Page 11: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for](https://reader030.fdocuments.us/reader030/viewer/2022011812/5e238a4f294c690a8d3a014a/html5/thumbnails/11.jpg)
12.05.2012
11
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 2110.5.2012 / AK-STUK
IAEA TRS 457, Fluoroscopy, kerma -area product (PKA)
Toroi et.al. A tandem calibration method for kerma-area product meters, PMB 53, 2008
Tandem method withreference KAP meterBeam area method
Calibration of KAP meters in situ
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 2210.5.2012 / AK-STUK
5 mmAl total filtration
Toroi et.al. Effects of radiation quality on the calibration of kerma-area product meters in x-ray beams, PMB 53, 2008
IAEA TRS 457, Kerma -area product
Incident beam
Transmitted beam
KAP meter
![Page 12: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for](https://reader030.fdocuments.us/reader030/viewer/2022011812/5e238a4f294c690a8d3a014a/html5/thumbnails/12.jpg)
12.05.2012
12
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 2310.5.2012 / AK-STUK
IAEA TRS 457, Fluoroscopy, Kerma area- product
Beam area or tandem method
Over and under couch set-ups
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 2410.5.2012 / AK-STUK
IAEA TRS 457, KAP -meters
Toroi et.al. Effects of radiation quality on the calibration of kerma-area product meters in x-ray beams, PMB 53, 2008
![Page 13: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for](https://reader030.fdocuments.us/reader030/viewer/2022011812/5e238a4f294c690a8d3a014a/html5/thumbnails/13.jpg)
12.05.2012
13
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 2510.5.2012 / AK-STUK
IAEA TRS 457, Kerma -area product (KAP)
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 2610.5.2012 / AK-STUK
Toroi & Kosunen. The energy dependence of the response of a patient dose calibrator, PMB 54, 2009
IAEA TRS 457, KAP-meters
Large fields for reference KAP meter30 cm * 30 cm
50-150 kV
Patient Dose Calibrator PDC-1Radcal
![Page 14: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for](https://reader030.fdocuments.us/reader030/viewer/2022011812/5e238a4f294c690a8d3a014a/html5/thumbnails/14.jpg)
12.05.2012
14
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 2710.5.2012 / AK-STUK
KAP-meters, Set-up
80 kV, 4.8 mmAl, 10*10cm at 30 FDD
Toroi et.al. Calibration of kerma area product meters with a patient dose calibrator, IDOS proceedings, Vol2, IAEA, 2011http://www-pub.iaea.org/books/IAEABooks/8700/Standards-Applications-and-Quality-Assurance-in-Medical-Radiation-Dosimetry-IDOS-Proceedings-of-an-International-Symposium-Vienna-9-12-November-2010
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 2810.5.2012 / AK-STUK
IAEA TRS 457, KAP uncertainties
IEC 60580, 25 % (k=2)
In examinations with varying voltage, 25 %
For specific tube voltage and energy dependence considered: 6 % (k=2)
KAP indicators based on calculations need to be calibrated also
![Page 15: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for](https://reader030.fdocuments.us/reader030/viewer/2022011812/5e238a4f294c690a8d3a014a/html5/thumbnails/15.jpg)
12.05.2012
15
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 2910.5.2012 / AK-STUK
IAEA TRS 457, Mammography, dosimetry quantities
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 3010.5.2012 / AK-STUK
IAEA TRS 457, Mammography
Plane parallel IC withentrance window
(low energy dependence)
45 mm PMMA phantom
![Page 16: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for](https://reader030.fdocuments.us/reader030/viewer/2022011812/5e238a4f294c690a8d3a014a/html5/thumbnails/16.jpg)
12.05.2012
16
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 3110.5.2012 / AK-STUK
IAEA TRS 457, Mammography, HVL
HVL measurement
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 3210.5.2012 / AK-STUK
IAEA TRS 457, Mammography, glandular dose
![Page 17: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for](https://reader030.fdocuments.us/reader030/viewer/2022011812/5e238a4f294c690a8d3a014a/html5/thumbnails/17.jpg)
12.05.2012
17
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 3310.5.2012 / AK-STUK
IAEA TRS 457, Mammography, uncertainties
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 3410.5.2012 / AK-STUK
IAEA TRS 457, Computed tomography, dosimetry quantities
![Page 18: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for](https://reader030.fdocuments.us/reader030/viewer/2022011812/5e238a4f294c690a8d3a014a/html5/thumbnails/18.jpg)
12.05.2012
18
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 3510.5.2012 / AK-STUK
IAEA TRS 457, Computed tomography
CT air kerma indices
T, Nominal slice thicknessN, number of slices/detector rows per scan for MSCT/MDCT
In air or in a phantom
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 3610.5.2012 / AK-STUK
IAEA TRS 457, Computed tomography, phantoms
![Page 19: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for](https://reader030.fdocuments.us/reader030/viewer/2022011812/5e238a4f294c690a8d3a014a/html5/thumbnails/19.jpg)
12.05.2012
19
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 3710.5.2012 / AK-STUK
IAEA TRS 457, Dosimetry formalism
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 3810.5.2012 / AK-STUK
IAEA TRS 457, Computed tomography
![Page 20: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for](https://reader030.fdocuments.us/reader030/viewer/2022011812/5e238a4f294c690a8d3a014a/html5/thumbnails/20.jpg)
12.05.2012
20
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 3910.5.2012 / AK-STUK
IAEA TRS 457, Computed tomography
Air kerma lenght product
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 4010.5.2012 / AK-STUK
IAEA TRS 457, Computed tomography, kerma lenght product
![Page 21: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for](https://reader030.fdocuments.us/reader030/viewer/2022011812/5e238a4f294c690a8d3a014a/html5/thumbnails/21.jpg)
12.05.2012
21
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 4110.5.2012 / AK-STUK
AAPM, IAEA, Recent approach for wide CT beam dosimetry
Work by ICRU…?
2010 2011
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 4210.5.2012 / AK-STUK
IAEA HHS No5, Computed tomography
From :IAEA Human Health Report No:5, Status of computed tomography dosimetry for wide cone beam scanners, 2011
![Page 22: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for](https://reader030.fdocuments.us/reader030/viewer/2022011812/5e238a4f294c690a8d3a014a/html5/thumbnails/22.jpg)
12.05.2012
22
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 4310.5.2012 / AK-STUK
Computed tomography, kerma lenght product, STUK approach
Measurement of kerma lenght product for whole examination
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 4410.5.2012 / AK-STUK
IAEA TRS 457, Dental radiography, dosimetry quantities
![Page 23: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for](https://reader030.fdocuments.us/reader030/viewer/2022011812/5e238a4f294c690a8d3a014a/html5/thumbnails/23.jpg)
12.05.2012
23
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 4510.5.2012 / AK-STUK
IAEA TRS 457, Dental radiographyTRS include quidance for bitewiew and panoramic examinations
Uncertainties
Kerma lenght product 13 % - 6 %Kerma area product 13,2 % – 6,4 %
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 4610.5.2012 / AK-STUK
The development of advanced dosimetry techniques for diagnostic and interventional
radiologyCoordinated Research Project on Implementation of the
International Code of Practice on Dosimetry in Diagnostic Radiology TRS 457
IAEA TRS 457, Implementation and testingAdvanced techniques, 2010 -2013
![Page 24: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for](https://reader030.fdocuments.us/reader030/viewer/2022011812/5e238a4f294c690a8d3a014a/html5/thumbnails/24.jpg)
12.05.2012
24
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 4710.5.2012 / AK-STUK
Paediatrics - clinicalDevelopment of guidanceGeneral RadiographyfluoroscopyCTanalysis & uncertaintyInterpretation of clinical dataDevelopment of guidancedevelopment of questionnaireorgan dose + eff. Dose & riskSSDLPreparation for intercomparisonintercomparison -SSDL - KLPintercomparison -SSDL - KAPnew qualities [KAP & mammo]Farmer chambers with RQT
IAEA TRS 457, Implementation and testingAdvanced techniques, 2010 -2013
I/II
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 4810.5.2012 / AK-STUK
Advanced dosimetryDevelopment of guidanceMDCTcone beam CT vs CTmammographic tomosynthesisdental dosimetrySkin doseDevelopment of guidanceFilm characterisation SSDLhigh dose calibrationhigh dose patientslow dose calibrationlow dose patientsuncertainty
IAEA TRS 457, Implementation and testingAdvanced techniques, 2010 -2013
DICOMDevelopment of guidancesurvey of DICOM capabilityUncertainty formulationAssessment of clinical scenariosassessment of uncertainty budget from scenariosDevelopment of guidance
I/II
![Page 25: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for](https://reader030.fdocuments.us/reader030/viewer/2022011812/5e238a4f294c690a8d3a014a/html5/thumbnails/25.jpg)
12.05.2012
25
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 4910.5.2012 / AK-STUK
•IAEA TRS 457 contains a comprehensive description of dosimetry methods used for diagnostic radiology
•Two options for reference KAP are presented; beam area or use of a reference KAP meter
•Generally 13 – 15 % (k=2) level of uncertainty (KAP meters 25 %) can be achieved for measurements in clinical dosimetry. Accuracy below 10% requires use of reference class dosemetersand/or calibration of dosemeter for user specific radiation qualities.
Summary I/II
SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 5010.5.2012 / AK-STUK
For dosimetry of wide beam CT the IAEA TRS457 methods need to be complemented by more recent dosimetry guidance (AAPM, IAEA, others)
IAEA TRS 457 has limitations for wide beam CT dosimetry. More recent recommendations exist; AAPM (2010), IAEA (2011)
Advanced dosimetry techniques for diagnostic and interventional radiology are under investigation in the coordinated research project by IAEA.
Summary II/II