IAEA code of practice for dosimetry in diagnostic radiology: recommendations for...

25
12.05.2012 1 SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALEN RADIATION AND NUCLEAR SAFETY AUTHORITY IAEA code of practice for dosimetry in diagnostic radiology: recommendations for hospitals and specific issues Antti Kosunen Radiation Metrology Laboratory-STUK Radiation safety and dosimetry in medical applications 1112 May 2012 TTÜ, Ehitajate tee 5, Tallinn, Estonia SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALEN RADIATION AND NUCLEAR SAFETY AUTHORITY 2 10.5.2012 / AK-STUK • Review /introduction to contents of IAEA TRS 457 for clinical dosimetry • Experiences of use the IAEA TRS 457 (KAP meters, others) • Remarks based on work at the SSDL of STUK • Methods by TLD are not reviewed Presentation

Transcript of IAEA code of practice for dosimetry in diagnostic radiology: recommendations for...

Page 1: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for

12.05.2012

1

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY

IAEA code of practice for dosimetry in diagnostic radiology:

recommendations for hospitals and specific issues

Antti KosunenRadiation Metrology Laboratory-STUK

Radiation safety and dosimetry in medical applications

11‐12 May 2012

TTÜ, Ehitajate tee 5, Tallinn, Estonia

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 210.5.2012 / AK-STUK

• Review /introduction to contents of IAEA TRS 457 for clinical dosimetry

• Experiences of use the IAEA TRS 457(KAP meters, others)

• Remarks based on work at the SSDL of STUK

• Methods by TLD are not reviewed

Presentation

Page 2: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for

12.05.2012

2

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 310.5.2012 / AK-STUK

http://www-pub.iaea.org/MTCD/publications/

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 410.5.2012 / AK-STUK

Dosimetry formalismDosimetry intrumentationQuantities and uncertainties

• General radiography• Fluoroscopy• Mammography• Computed tomography• Dental radiography

IAEA TRS 457, Clinical dosimetry

Page 3: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for

12.05.2012

3

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 510.5.2012 / AK-STUK

• Dosimetry• phantom• patient

• Guidance for selection of patients for DRL

•Appendix VI. Application of patient dose measurements• Background• Guidance for DRLs• Guidance for organ dose conversion factors for examinations• Summary of using effective dose

IAEA TRS 457, Clinical dosimetry

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 610.5.2012 / AK-STUK

• Assumed that the x- ray equipment is properly maintainedaccording to QA program

• Assumed that the x ray tube output is linear relative to tube loading,if not complementary measurements are required

• HVLFor interpolations of calibration coefficientsFor conversions to organ doses (mammography)

Not necessrerily for every dose measurement, as information from Quality Control results exist

IAEA TRS 457, Clinical dosimetry

Page 4: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for

12.05.2012

4

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 710.5.2012 / AK-STUK

• Examples and worksheets for measurements

• Dosimeters should be calibrated at the laboratory• Calibration interval:

• Based on national requirements,• Usually should not exceed two years

IAEA TRS 457, Clinical dosimetry

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 810.5.2012 / AK-STUK

( 2sd )

• Risk assessment : Absolute risk • Dose surveys

~20 %

• Relative risk• Interventional procedures • Quality assurance

measurements• Paediatric procedures

7 %

IAEA TRS 457, Accuracy requirements for dosimetryAt reference conditions of dosimetry

Page 5: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for

12.05.2012

5

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 910.5.2012 / AK-STUK

Air kerma based formalism

IAEA TRS 457, Dosimetry formalism

Beam quality correction factor

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 1010.5.2012 / AK-STUK

IAEA TRS 457, Selection of instruments, clinical

Requirements of International Standards are followed:

Requirements for dosemeters at laboratory can be used also as a background information for clinical dosemeters

Page 6: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for

12.05.2012

6

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 1110.5.2012 / AK-STUK

IAEA TRS 457, Measurement uncertainty (accuracy)

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 1210.5.2012 / AK-STUK

IAEA TRS 457, General radiology, dosimetry quantities

Page 7: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for

12.05.2012

7

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 1310.5.2012 / AK-STUK

IAEA TRS 457, General radiology

Incident air kerma

Air kerma detector with isotropic response

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 1410.5.2012 / AK-STUK

IAEA TRS 457, General radiology, phantoms

Page 8: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for

12.05.2012

8

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 1510.5.2012 / AK-STUK

IAEA TRS 457, Review on phantoms

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 1610.5.2012 / AK-STUK

IAEA TRS 457, General radiology, uncertainty

Page 9: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for

12.05.2012

9

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 1710.5.2012 / AK-STUK

IAEA TRS 457, Fluoroscopy, dosimetry quantities

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 1810.5.2012 / AK-STUK

IAEA TRS 457, Fluoroscopy

•Anti- scatter grid in position if used•Focus to table top distance as used clinically•Phantom 20 cm, 30 cm for large patients•Collimation large, to phantom size•Exposure under ABC

Detector with back scatter measurement capability

Page 10: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for

12.05.2012

10

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 1910.5.2012 / AK-STUK

IAEA TRS 457, Fluoroscopy

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 2010.5.2012 / AK-STUK

IAEA TRS 457, Fluoroscopy, uncertainties

Page 11: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for

12.05.2012

11

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 2110.5.2012 / AK-STUK

IAEA TRS 457, Fluoroscopy, kerma -area product (PKA)

Toroi et.al. A tandem calibration method for kerma-area product meters, PMB 53, 2008

Tandem method withreference KAP meterBeam area method

Calibration of KAP meters in situ

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 2210.5.2012 / AK-STUK

5 mmAl total filtration

Toroi et.al. Effects of radiation quality on the calibration of kerma-area product meters in x-ray beams, PMB 53, 2008

IAEA TRS 457, Kerma -area product

Incident beam

Transmitted beam

KAP meter

Page 12: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for

12.05.2012

12

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 2310.5.2012 / AK-STUK

IAEA TRS 457, Fluoroscopy, Kerma area- product

Beam area or tandem method

Over and under couch set-ups

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 2410.5.2012 / AK-STUK

IAEA TRS 457, KAP -meters

Toroi et.al. Effects of radiation quality on the calibration of kerma-area product meters in x-ray beams, PMB 53, 2008

Page 13: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for

12.05.2012

13

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 2510.5.2012 / AK-STUK

IAEA TRS 457, Kerma -area product (KAP)

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 2610.5.2012 / AK-STUK

Toroi & Kosunen. The energy dependence of the response of a patient dose calibrator, PMB 54, 2009

IAEA TRS 457, KAP-meters

Large fields for reference KAP meter30 cm * 30 cm

50-150 kV

Patient Dose Calibrator PDC-1Radcal

Page 14: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for

12.05.2012

14

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 2710.5.2012 / AK-STUK

KAP-meters, Set-up

80 kV, 4.8 mmAl, 10*10cm at 30 FDD

Toroi et.al. Calibration of kerma area product meters with a patient dose calibrator, IDOS proceedings, Vol2, IAEA, 2011http://www-pub.iaea.org/books/IAEABooks/8700/Standards-Applications-and-Quality-Assurance-in-Medical-Radiation-Dosimetry-IDOS-Proceedings-of-an-International-Symposium-Vienna-9-12-November-2010

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 2810.5.2012 / AK-STUK

IAEA TRS 457, KAP uncertainties

IEC 60580, 25 % (k=2)

In examinations with varying voltage, 25 %

For specific tube voltage and energy dependence considered: 6 % (k=2)

KAP indicators based on calculations need to be calibrated also

Page 15: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for

12.05.2012

15

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 2910.5.2012 / AK-STUK

IAEA TRS 457, Mammography, dosimetry quantities

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 3010.5.2012 / AK-STUK

IAEA TRS 457, Mammography

Plane parallel IC withentrance window

(low energy dependence)

45 mm PMMA phantom

Page 16: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for

12.05.2012

16

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 3110.5.2012 / AK-STUK

IAEA TRS 457, Mammography, HVL

HVL measurement

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 3210.5.2012 / AK-STUK

IAEA TRS 457, Mammography, glandular dose

Page 17: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for

12.05.2012

17

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 3310.5.2012 / AK-STUK

IAEA TRS 457, Mammography, uncertainties

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 3410.5.2012 / AK-STUK

IAEA TRS 457, Computed tomography, dosimetry quantities

Page 18: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for

12.05.2012

18

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 3510.5.2012 / AK-STUK

IAEA TRS 457, Computed tomography

CT air kerma indices

T, Nominal slice thicknessN, number of slices/detector rows per scan for MSCT/MDCT

In air or in a phantom

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 3610.5.2012 / AK-STUK

IAEA TRS 457, Computed tomography, phantoms

Page 19: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for

12.05.2012

19

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 3710.5.2012 / AK-STUK

IAEA TRS 457, Dosimetry formalism

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 3810.5.2012 / AK-STUK

IAEA TRS 457, Computed tomography

Page 20: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for

12.05.2012

20

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 3910.5.2012 / AK-STUK

IAEA TRS 457, Computed tomography

Air kerma lenght product

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 4010.5.2012 / AK-STUK

IAEA TRS 457, Computed tomography, kerma lenght product

Page 21: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for

12.05.2012

21

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 4110.5.2012 / AK-STUK

AAPM, IAEA, Recent approach for wide CT beam dosimetry

Work by ICRU…?

2010 2011

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 4210.5.2012 / AK-STUK

IAEA HHS No5, Computed tomography

From :IAEA Human Health Report No:5, Status of computed tomography dosimetry for wide cone beam scanners, 2011

Page 22: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for

12.05.2012

22

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 4310.5.2012 / AK-STUK

Computed tomography, kerma lenght product, STUK approach

Measurement of kerma lenght product for whole examination

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 4410.5.2012 / AK-STUK

IAEA TRS 457, Dental radiography, dosimetry quantities

Page 23: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for

12.05.2012

23

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 4510.5.2012 / AK-STUK

IAEA TRS 457, Dental radiographyTRS include quidance for bitewiew and panoramic examinations

Uncertainties

Kerma lenght product 13 % - 6 %Kerma area product 13,2 % – 6,4 %

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 4610.5.2012 / AK-STUK

The development of advanced dosimetry techniques for diagnostic and interventional

radiologyCoordinated Research Project on Implementation of the

International Code of Practice on Dosimetry in Diagnostic Radiology TRS 457

IAEA TRS 457, Implementation and testingAdvanced techniques, 2010 -2013

Page 24: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for

12.05.2012

24

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 4710.5.2012 / AK-STUK

Paediatrics - clinicalDevelopment of guidanceGeneral RadiographyfluoroscopyCTanalysis & uncertaintyInterpretation of clinical dataDevelopment of guidancedevelopment of questionnaireorgan dose + eff. Dose & riskSSDLPreparation for intercomparisonintercomparison -SSDL - KLPintercomparison -SSDL - KAPnew qualities [KAP & mammo]Farmer chambers with RQT

IAEA TRS 457, Implementation and testingAdvanced techniques, 2010 -2013

I/II

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 4810.5.2012 / AK-STUK

Advanced dosimetryDevelopment of guidanceMDCTcone beam CT vs CTmammographic tomosynthesisdental dosimetrySkin doseDevelopment of guidanceFilm characterisation SSDLhigh dose calibrationhigh dose patientslow dose calibrationlow dose patientsuncertainty

IAEA TRS 457, Implementation and testingAdvanced techniques, 2010 -2013

DICOMDevelopment of guidancesurvey of DICOM capabilityUncertainty formulationAssessment of clinical scenariosassessment of uncertainty budget from scenariosDevelopment of guidance

I/II

Page 25: IAEA code of practice for dosimetry in diagnostic radiology: recommendations for ...yhisope.cb.ttu.ee/images/docs/RadiationSafety/AK... · 2014-09-15 · IAEA code of practice for

12.05.2012

25

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 4910.5.2012 / AK-STUK

•IAEA TRS 457 contains a comprehensive description of dosimetry methods used for diagnostic radiology

•Two options for reference KAP are presented; beam area or use of a reference KAP meter

•Generally 13 – 15 % (k=2) level of uncertainty (KAP meters 25 %) can be achieved for measurements in clinical dosimetry. Accuracy below 10% requires use of reference class dosemetersand/or calibration of dosemeter for user specific radiation qualities.

Summary I/II

SÄTEILYTURVAKESKUS • STRÅLSÄKERHETSCENTRALENRADIATION AND NUCLEAR SAFETY AUTHORITY 5010.5.2012 / AK-STUK

For dosimetry of wide beam CT the IAEA TRS457 methods need to be complemented by more recent dosimetry guidance (AAPM, IAEA, others)

IAEA TRS 457 has limitations for wide beam CT dosimetry. More recent recommendations exist; AAPM (2010), IAEA (2011)

Advanced dosimetry techniques for diagnostic and interventional radiology are under investigation in the coordinated research project by IAEA.

Summary II/II