Hydrogen Cyanide HCN No.12M 50 100 200 300 400 600 800 ppm n=1 12M 1 ppm2 pump strokesYellow Red Temperature 0 to 40 (32 to 104 ) correction not used Relative humidity 0 to 90 % correction not used 10 % for 50 to 200 ppm, 5 % for 200 to 800 ppm10 tubes for 10 tests 36 months Detecting limit : Colour change : Operating conditions : Relative standard deviation : Tube quantity and number of tests per box : Shelf life : Performance 17 to 50 ppm 50 to 800 ppm 800 to 2400 ppm 2200 mL1100 mL1/250 mL1/3 1 3 2 min 1 min 30 sec Measuring range Number of pump strokes Correction factor Sampling time 2HCN HgCl2 Hg CN2 2HCl HCl Base Chloride Reaction principle Possible coexisting substances and their interferences Hydrogen sulphide 500 ppm Red Sulphur dioxide 500 ppm Substance Concentration Interference Changes colour by itself to Chlorine, hydrogen sulphideup to 500 ppm, and nitric acid are trapped in the pretreatmentpale blue and yellowlayers. If the pretreatment reagent is entirely consumed, a higher reading will be given. Permeation tube method Calibration gas generation Scale

Transcript of Hydrogen Cyanide HCN No - gastec.co.jp

Page 1: Hydrogen Cyanide HCN No - gastec.co.jp

Hydrogen Cyanide HCN No.12M

50 100

200

300

400

600

800

ppm

n=1

12M

1 ppm(2 pump strokes)Yellow → RedTemperature 0 to 40 ℃ (32 to 104 ℉) correction not usedRelative humidity 0 to 90 % correction not used10 %(for 50 to 200 ppm), 5 %(for 200 to 800 ppm)10 tubes for 10 tests36 months

Detecting limit :Colour change :Operating conditions :

Relative standard deviation :Tube quantity and number of tests per box :Shelf life :

Performance

17 to 50 ppm 50 to 800 ppm 800 to 2400 ppm 2(200 mL) 1(100 mL) 1/2(50 mL) 1/3 1 3 2 min 1 min 30 sec

Measuring rangeNumber of pump strokesCorrection factorSampling time

2HCN + HgCl2 → Hg(CN)2 + 2HClHCl + Base → Chloride

Reaction principle

Possible coexisting substances and their interferences

Hydrogen sulphide ≧ 500 ppm + RedSulphur dioxide ≧ 500 ppm +

Substance Concentration Interference Changes colour by itself to

Chlorine, hydrogen sulphide(up to 500 ppm), and nitric acid are trapped in the pretreatment(pale blue and yellow)layers. If the pretreatment reagent is entirely consumed, a higher reading will be given.

Permeation tube methodCalibration gas generation

Scale