Hydraderm Lift®
“Double Ionisation”
OPERATING MANUAL
THE HYDRADERM LIFT CAN ONLY BE USED BY BEAUTY THERAPIST WHO HAVEFOLLOWED THE TRAINING COURSE
English
General characteristics- Manufacturer: ROS’S ESTETICA S.L.- Model: HYDRADERM LIFT - GUINOT- Classification: CLASS 1 EQUIPMENT (COSMETIC APPARATUS)
Mechanical characteristics- Size: L 420 mm / W 460 mm / H 215 mm- Weight: 7.5 kg
Electrical characteristics- Voltage: 230 V- Frequency: 50 Hz- T Type fuses: 500 mA / 250 V- Power: 25 W
Protection- The apparatus has 2 x 500 mA fuses / 250 V which protect against overloading
Application characteristics- Ionisation: Maximum current produces 0,5 mA per electrode- Stimulation: Low frequency current
(maximum frequency 165 Hz / maximum voltage 20 V)
Cleansing- After each use of the roller electrodes, the ball electrodes and the return handle, clean
with a damp towel using 70° ethyl alcohol and dry with a sterile cloth. The neon lightelectrode must also be cleaned after each use with 70° ethyl alcohol.
CAUTIONBefore using the HYDRADERM LIFT apparatus, you need to obtain authorisation fromclients using a cardiac stimulator as they run a certain risk caused by the interference ordeterioration of the cardiac stimulator.
In case of a breakdown, contact the nearest technical service or: ROS’S C/Del Molinot 46-48 Pol.Ind Cami Ral - 08860 Castelldefels (Spain) Tél : 00 34 93 42 66 017 - [email protected]
THE TECHNICAL DESCRIPTION OFTHE HYDRADERM LIFT EQUIPMENT
-1-
All Guinot-trained beauty therapists must attend a compulsory training course
on the HYDRADERM LIFT apparatus prior using it.
During this training they will be given a detailed training manual explaining how to use the HYDRADERM LIFT
apparatus and the associated skincare products.
-2-
THE HYDRADERM LIFT APPARATUS
A GALVANIC CURRENT
A constant strength of current, used to:
- Increase effectiveness of skin care cosmetics.- Moisturise the skin.
A STIMULATING CURRENT
Low frequency stimulating micro-current used to:
- Stimulate the contraction of the surface muscular fibres.
A HIGH FREQUENCY CURRENT
Alternative current used to:
- Strengthen the diffusion and the assimilation of skin care cosmetics.
THE ELECTRODES
To transmit the effect of the currents:
- Galvanic1 positive “Roller” electrode (+)1 negative “Roller” electrod (–)
- Stimulation2 stimulation “Ball” electrodes
- High Frequency1 High Frequency handle1 neon glass electrode 3
-3-
USINGTHE HYDRADERM LIFT APPARATUS
IONISATION ROLLER
1. Switch on the apparatus by pushing the button in the back.
Push the “ON” button; automatically the “Roller” indicator lights up (front panel), as well as the length of the treatment: 10 minutes.
Verify the connections with the help of the diagram below.
2. Put the potentiometer at level 5 using the + and – buttons located under “Galvanic”,then press the “MARCH” button to start the treatment (green light). Adjust the intensi-ty to skin sensitivity.
3. Move the “Roller” electrodes as indicated on the diagrams until the beep sounds.
FUS
EI0
-4-
STIMULATION
1. Push the “Mono Ball” button . Check the connections according to the diagram below:
2. Put the potentiometer at level 2 for the “Draining” step.
3. Put the potentiometer at level 5 for the “Stimulation” step, then increase dependingon the sensitivity of the client’s skin using the + and – buttons located under“Stimulation”.
4. Adjust the timer to 10 minutes. Then push the “MARCH” button to begin (greenlight).
5. Move the “Ball” electrodes ( ) around the areas indicated, in accordance with the diagrams, until the melody is obtained.
-5-
HIGH FREQUENCY OUTPUT
1. Push the “High Frequency” button. Adjust the strength of the potentiometer with the + and – buttons situated beneath the “High Frequency” sign: begin at 5 and gradually increase up to 10 (for sensitive skin do not go over 7).
Check the connections according to the diagram below:
2. Adjust the timer to five minutes, then press the “MARCH” button.
3. Move the “High Frequency” electrode over the gauze while exerting a little pressure (to avoid sparks), until the beep sounds.
Contraindications for using the device:- Pacemaker- Open cuts- Eczema- Metal plates and metal implants under the skin- Use of Vitamin A acid- Bruises- Do not use on thyroid area- Do not use the stimulation current in case of pregnancy
Please advise the client to check with a doctor prior to having the skincare treatment if any of the above apply.