Hydra Lift Operating Manual

7
Hydraderm Lift ® “Double Ionisation” OPERATING MANUAL THE HYDRADERM LIFT CAN ONLY BE USED BY BEAUTY THERAPIST WHO HAVEFOLLOWED THE TRAINING COURSE English

Transcript of Hydra Lift Operating Manual

Page 1: Hydra Lift Operating Manual

Hydraderm Lift®

“Double Ionisation”

OPERATING MANUAL

THE HYDRADERM LIFT CAN ONLY BE USED BY BEAUTY THERAPIST WHO HAVEFOLLOWED THE TRAINING COURSE

English

Page 2: Hydra Lift Operating Manual

General characteristics- Manufacturer: ROS’S ESTETICA S.L.- Model: HYDRADERM LIFT - GUINOT- Classification: CLASS 1 EQUIPMENT (COSMETIC APPARATUS)

Mechanical characteristics- Size: L 420 mm / W 460 mm / H 215 mm- Weight: 7.5 kg

Electrical characteristics- Voltage: 230 V- Frequency: 50 Hz- T Type fuses: 500 mA / 250 V- Power: 25 W

Protection- The apparatus has 2 x 500 mA fuses / 250 V which protect against overloading

Application characteristics- Ionisation: Maximum current produces 0,5 mA per electrode- Stimulation: Low frequency current

(maximum frequency 165 Hz / maximum voltage 20 V)

Cleansing- After each use of the roller electrodes, the ball electrodes and the return handle, clean

with a damp towel using 70° ethyl alcohol and dry with a sterile cloth. The neon lightelectrode must also be cleaned after each use with 70° ethyl alcohol.

CAUTIONBefore using the HYDRADERM LIFT apparatus, you need to obtain authorisation fromclients using a cardiac stimulator as they run a certain risk caused by the interference ordeterioration of the cardiac stimulator.

In case of a breakdown, contact the nearest technical service or: ROS’S C/Del Molinot 46-48 Pol.Ind Cami Ral - 08860 Castelldefels (Spain) Tél : 00 34 93 42 66 017 - [email protected]

THE TECHNICAL DESCRIPTION OFTHE HYDRADERM LIFT EQUIPMENT

-1-

Page 3: Hydra Lift Operating Manual

All Guinot-trained beauty therapists must attend a compulsory training course

on the HYDRADERM LIFT apparatus prior using it.

During this training they will be given a detailed training manual explaining how to use the HYDRADERM LIFT

apparatus and the associated skincare products.

-2-

Page 4: Hydra Lift Operating Manual

THE HYDRADERM LIFT APPARATUS

A GALVANIC CURRENT

A constant strength of current, used to:

- Increase effectiveness of skin care cosmetics.- Moisturise the skin.

A STIMULATING CURRENT

Low frequency stimulating micro-current used to:

- Stimulate the contraction of the surface muscular fibres.

A HIGH FREQUENCY CURRENT

Alternative current used to:

- Strengthen the diffusion and the assimilation of skin care cosmetics.

THE ELECTRODES

To transmit the effect of the currents:

- Galvanic1 positive “Roller” electrode (+)1 negative “Roller” electrod (–)

- Stimulation2 stimulation “Ball” electrodes

- High Frequency1 High Frequency handle1 neon glass electrode 3

-3-

Page 5: Hydra Lift Operating Manual

USINGTHE HYDRADERM LIFT APPARATUS

IONISATION ROLLER

1. Switch on the apparatus by pushing the button in the back.

Push the “ON” button; automatically the “Roller” indicator lights up (front panel), as well as the length of the treatment: 10 minutes.

Verify the connections with the help of the diagram below.

2. Put the potentiometer at level 5 using the + and – buttons located under “Galvanic”,then press the “MARCH” button to start the treatment (green light). Adjust the intensi-ty to skin sensitivity.

3. Move the “Roller” electrodes as indicated on the diagrams until the beep sounds.

FUS

EI0

-4-

Page 6: Hydra Lift Operating Manual

STIMULATION

1. Push the “Mono Ball” button . Check the connections according to the diagram below:

2. Put the potentiometer at level 2 for the “Draining” step.

3. Put the potentiometer at level 5 for the “Stimulation” step, then increase dependingon the sensitivity of the client’s skin using the + and – buttons located under“Stimulation”.

4. Adjust the timer to 10 minutes. Then push the “MARCH” button to begin (greenlight).

5. Move the “Ball” electrodes ( ) around the areas indicated, in accordance with the diagrams, until the melody is obtained.

-5-

Page 7: Hydra Lift Operating Manual

HIGH FREQUENCY OUTPUT

1. Push the “High Frequency” button. Adjust the strength of the potentiometer with the + and – buttons situated beneath the “High Frequency” sign: begin at 5 and gradually increase up to 10 (for sensitive skin do not go over 7).

Check the connections according to the diagram below:

2. Adjust the timer to five minutes, then press the “MARCH” button.

3. Move the “High Frequency” electrode over the gauze while exerting a little pressure (to avoid sparks), until the beep sounds.

Contraindications for using the device:- Pacemaker- Open cuts- Eczema- Metal plates and metal implants under the skin- Use of Vitamin A acid- Bruises- Do not use on thyroid area- Do not use the stimulation current in case of pregnancy

Please advise the client to check with a doctor prior to having the skincare treatment if any of the above apply.

-6-