    Guidelines for Use of the Analytical HPLCs (Agilent 1100 Series)

    Compiled by Jake Wiener. Justice.I. Sample Preparation

    Never use EtOAc as an eluant or to dissolve your sample when using chiral columns

    1. For chiral columns: dissolve your sample (5 mg/1 mL solvent) in ethanol,isopropanol, and/or hexanes ONLY. Other solvents will destroy the chiral

    columns.

    2. For the achiral columns: dissolve your sample (5 mg/1 mL solvent) in any solventthat will dissolve your sample.

    3. Place this solution in a small sample vial, and place the vial in any position on theautosampling tray.

    II. Sign Up Rules

    Always sign the logbook before running a sample

    1. If you insert runs ahead of other people's runs, let ALL of those people know sothat they can plan accordingly.

    2. Generally, ee runs have priority over racemic assay runs, though MacMillan usersalso have priority over other groups. If you bump a user from another group, let

    them know.

    III. Running a Sample

    1. At Chemstation on the PC, go to the Online instance of either instrument 1 or 2,depending on where your sample is.

    2. Select sequence table from the sequence pull-down menu.3. Enter the information for your run, including the location of your sample, the

    name of your sample, the method you wish to run, and the datafile where you willstore the data. Enter a 1 under the inj/vial column.

    4. Methods are typically in the format of percent additive in hexane over time (e.g,6EOH40 is 6% EtOH in hexanes for 40 minutes). You should check each method

    to make sure it involves the correct solvent percentages and time by choosingLoad Method from the method pull-down menu. Do this also to modify an

    existing method for a new solvent composition or time, and then usesave as to

    give this modified method a new name.

    5. If you wish automatically to switch columns during a sequence, use a line with amethod labeled POSX where X = the position of the column you wish to use

    (see below). This line does not require a vial or sample name but you must stillwrite 1 in the inj/vial column.

    6. Before injecting a sample for a run of a certain polarity, the column must beequilibrated for 20 minutes with that solvent system. You can do this by inserting

    a line with no sample name, no datafile, 1 inj/vial, and no vial number with a

    method ending in E (for equilibration), as in 4IPAE, which is a 4% iPrOH

    equilibration run of 20 minutes (these methods do not print.) Alternately, you canselect a wait time in the sequence parameters, as discussed below.

  • 7/28/2019 HPLC Agilent Guideline

    2/3

    7. Follow steps 3, 4, 5, and 6 for all runs you wish to have in the sequence.8. For chiral columns: the last line of a sequence is 10_IPA which stores the

    column under 10% isopropanol. All chiral columns must be stored under 10%

    IPA when they are not in use. Note: If you use multiple columns then you must

    store EACH column under 10_IPA. That is, before the POSX method, you need a

    10_IPA. If you suspect that your sample may require additional time to come offthe column, use 10% IPA for a longer amount of time.

    9. For achiral columns: the last line of a sequence is 100HEX20 which stores thecolumn under hexanes. All achiral columns must be stored under hexanes when

    they are not in use. Note: If you use multiple columns then you must storeEACH column under its appropriate storage solvent. That is, before the POSX

    method, you need a 100HEX20 for the achiral columns. If you suspect that your

    compound may require additional time to come off the column, use a more polar

    method to wash and then store under 100HEX20.10.Click OK.11.Select sequence parameters from the sequence menu and fill in the appropriate

    information. Make sure that the box on the right side under shutdown and nextto the post sequence cmnd/macro is checked. Then select theshutdown.macro from the pull down list of choices. If you want the machine to

    wait for a certain amount of time after loading a solvent system but before

    injecting (to equilibrate) then enter a wait time. Click OK. Then click save

    sequence as from the sequence menu. Name your sequence.12.Check that the solvent bottles are full.13.On the main chemstation screen, turn on the HPLC using the on button on the

    bottom right of the picture of the system.14.Choose run sequence from the run control menu.15.Check the pressure to make sure the system is functioning properly (see pressure

    limits below.)

    IV. Solvent Refilling Rules

    1. After your runs finish, refill all solvent reservoir bottles with filtered, HPLC gradesolvent. This kind of solvent is located in the fire cabinet in the wooden cabinet

    next to the HPLC. If you use the last of (or almost the last of) the solvent in the

    bottle, get a new bottle and filter the solvent using the filter tool located in the

    drawer under the HPLC (see the operator to learn how to filter the solvent.)Return bottles to the fire cabinet when done.

    2.

    When you refill the hexane solvent reservoir bottle, make sure the computerthinks there are 150 mL less than there really are. If you fill the reservoir all the

    way up, enter into the computer the maximum volume (0.8 L.)3. No EtOAc on the chiral columns.

    V. Pressure Limits

    Acceptable pressures are less than 55 bar for the SIL, SIL-RX, OD-H, and OJ columns.

  • 7/28/2019 HPLC Agilent Guideline

    3/3

    Acceptable pressures are less than 35 bar for the AS and AD columns. If the pressure is

    too high, open the bypass valve (black finger tight valve near the top of the HPLC) andthen tell the HPLC operator.

    VI. Using the Column Changer

    Two Options:

    1. Use a normal sequence or method

    Press the left button on the front of the column changer to switch between columns. Set

    the Position number according to the column you want to run. Position 1 is a bypassline that is used to flush the system, especially after using the Silica column and before

    using the chiral columns.

    Position 1 = Bypass

    Position 2 = AD + GPosition 3 = ODH + G

    Position 4 = OJ + G

    Position 5 = AS + G

    Position 6 = Silica

    2. Use a sequence containing automated column switching

    Column selection is controlled by a method, with each column being selected by its ownmethod. The column-selecting methods are given names corresponding to the Position

    number of each column, as above (pos1.m, pos2.m, etc.). The method to switch a column

    takes about 15 seconds to run and the column-selecting methods set the solvent flow rate

    to 0.00 mL/min. Insert a column-selecting method before the actual methods of asequence that you wish to run on that column. For a column-selecting method in a

    sequence, set the vial as blank and inj/vial to 1. In a multi-column sequence, make sure

    to wash each column with 10_IPA before moving on to the next column.

    If you have set up a wait time in the sequence parameters, loading the method that

    switches columns will invoke that wait time, in addition to the wait time that will occur

    when the next solvent-method loads; thus you may wish to equlibrate as in section III,part 6 above and use 0 for the wait time.

    IMPORTANT: If you use the silica column with EtOAc, flush the system with some

    non-EtOAc solvent system for 30 minutes by using the pos1.m method (the bypassmethod), or by manually setting the changer to position 1 with the right button on

    the front of the changer, ESPECIALLY before using the chiral columns.