TM 11-6625-1576-15 DEPARTMENT OF THE ARMY TECHNICAL MANUAL ORGANIZATIONAL, DS, GS, AND DEPOT MAINTENANCE MANUAL DISTORTION ANALYZER HEWLETT-PACKARD MODELS 333A AND 334A HEADQUARTERS, DEPARTMENT OF THE ARMY MAY 1967

description

HP-333A-334A

Transcript of HP-333A-334A

Page 1: HP-333A-334A

TM 11-6625-1576-15DEPARTMENT OF THE ARMY TECHNICAL MANUAL

ORGANIZATIONAL, DS, GS, AND DEPOTMAINTENANCE MANUAL

DISTORTION ANALYZERHEWLETT-PACKARD MODELS

333A AND 334A

HEADQUARTERS, DEPARTMENT OF THE ARMYMAY 1967

Page 2: HP-333A-334A

TM 11-6625-1576-15

WARNING

DANGEROUS VOLTAGES

EXIST IN THIS EQUIPMENT

Be careful when working on the power supply and

on the 115-volt ac line connections.

DO NOT TAKE CHANCES !

Page 3: HP-333A-334A

T h i s m a n u a l c o n t a i n s c o p y r i g h t e d m a t e r i a l o r i g i n a l l yprcpared by the Hewlet t -Packard Co.

TM11-6625-1576-15

TECHNICAL MANUAL ) HEADQUARTERS

) DEPARTMENT OF THE ARMY

N O . 1 1 - 6 6 2 5 - 1 5 7 6 - 5 ) Washington, D.C., 1 9 M a y 1 9 6 7

Organiza t ional , DS, GS, and Depot Maintenance

DISTORTION ANALYZER, HEWLETT-PACKARD MODELS 333A AND 334A

S e c t i o n Page

I GENERAL INFORMATION . . . . . . . . . 1-2.11-A.1 Scope . . . . . . . . . . . . . . . . .1-2.1l-A.2 Index of Publications. . . 1 - 2 . 1l-A.3 Forms and Records . . . . . . . . . . . 1 - 2 . 1l - 1 . D e s c r i p t i o n . . . . . . . . 1 - 31-6. Accessory Features . . . . . . . .1-31-8. Options Available . . . . . . . . .1-31-11. Modifications . . . . . . . . . . . . . . . . . . . . . . . . 1-3

Section PageII INSTALLATION . . . . . . . . . . . . . .2-1

2-1. Introduction . . . . . . . . . . . . . .2-12-3. Inspection . . . . . . . . . . . . . . 2-12-5. Power Requirements . . . . . . . . . . 2-12-7. Three-Conductor Power Cable . . . 2-12-10. Installation . . . . . . . . . . . . . . 2-12-12. Bench Installation . . . . . . . . . 2-12-14. Rack Installation . . . . . . . . . . 2-12-16. Repackaging for Shipment . . . . . . 2-1

Section PageIII OPERATING INSTRUCTIONS . . . . . . . 3-1

3-1. Introduction . . . . . . . . . . . . . 3-13-4. Controls and Indicators . . . . . . . 3-13-6. Adjustment of Mechanical Zero . . . 3-13-8. General Operating Information . . . 3-13-9. Input Connections . . . . . . . . . 3-13-11. Voltmeter Characteristics . . . . . 3-13-14. Use of Output Terminals . . . . . . 3-13-16. Operating Procedures . . . . . . . . 3-23-18. Distortion Measurement . . . . . . 3-2

Section PageIV THEORY OF OPERATION . . . . . . . . . . 4-1

4-1. Overall Description . . . . . . . . 4-14-3. Block Diagram Description . . . . . 4-14-4. Distortion Measuring Operation . . 4-14-6. Distortion Measurement in AM

. Carriers . . . . . . . . . . . . . 4-14-8. Voltmeter Operation . . . . . . . . 4-14-10. Schematic Theory . . . . . . . . . . 4-14-11. Impedance Converter Circuit . . . 4-14-14. Rejection Amplifier Circuit . . . . 4-24-36. High Pass Filter . . . . . . . . . . . 4-54-38. Meter Circuit . . . . . . . . . . . . 4-54-46. Power Supply Circuit . . . . . . . . . . 4-74-51. RF Detector Circuit (334A only) . l 4-7

Section PageV MAINTENANCE. . . . . . . . . . . . . 5-1

5-l. Introduction. . . . . . . . . . . 5-15-2. Test Equipment Required . . . . . 5-15-5. Performance Checks. . . . . . . . . 5-1

Section V (Cont ‘d)5-9. Fundamental Rejection Check . .5-10. Second Harmonic Accuracy Check5-11. Distortion Introduced by Instru-

ment Check and AutomaticControl Loop Operation . . . .

5-12. Frequency Calibration AccuracyCheck . . . . . . . . . . . .

5-13. Input Resistance Check . . . . . . 5-14. Input Shunt Capacitance Check . . 5-15. Minimum Input Level Check . . . 5-16. DC Isolation Check . . . . . . . . 5-17. Voltmeter Accuracy Check . . . . 5-18. High Pass Filter Check . . . . . . 5-19. Voltmeter Frequency Response

Check . . . . . . . . . . . . . . . . 5-20. Residual Noise Check . . . . . . . 5-21. AM Detector Check (Model 334A

only) . . . . . . . . . . . . . . .5-22. Adjustment and Calibration

Procedure. . . . . . . . . . . . . . . 5-25. Meter Mechanical Zero Set . . . . 5-27. Power Supply and Bias

Adjustments. . . . . . . . . . . 5-28. A3R16 and A3R30 Distortion

Adjust. . . . . . . . . . . . . . 5-29. Bridge Balance Adjustment (C3) .5-30. Voltmeter Gain Adjustments . . . 5-31. Voltmeter Frequency Response

Adjustment . . . . . . . . . . . 5-32. Sensitivity Switch Calibration . . .5-33. Troubleshooting Procedures . . . .5-39. Bottom Shield Removal . . . . . . 5-41. Servicing Etched Circuit Boards .5-44. Servicing Rotary Switches . . . .

Page 5-1 5-2

5-2

5-3 5-4 5-4 5-4 5-4 5-5 5-5

5-6 5-6

5-6

5-7 5-7

5-7

5-9 5-95-10

5-105-105-115-115-115-13

SectionVI SCHEMATIC DIAGRAMS

6-1. Schematic Diagrams

Page6-16-1

i i i

Page 4: HP-333A-334A
Page 5: HP-333A-334A

TM 11-6625-1576-15

Model 333A/334A Section 1Figure 1-1 and Table 1-1

Figure 1-1. Model 333A Distortion Analyzer

1-1

Table 1-1. Specifications

Page 6: HP-333A-334A

TM 11-6625-1576-15

Section I Model 333A/334ATable 1-1

Table 1-1. Specifications (Cont ‘d)

1-2

Page 7: HP-333A-334A

TM 11-6625-1576-15

SECTION I

GENERAL INFORMATION

1 - A . 1 S c o p e

T h i s m a n u a l i n c l u d e s i n s t a l l a t i o n a n d o p e r a t i o n i n s t r u c t i o n s a n d

c o v e r s o p e r a t o r ’ s , o r g a n i z a t i o n a l , d i r e c t s u p p o r t ( D S ) , g e n e r a l s u p p o r t

( G S ) , a n d d e p o t m a i n t e n a n c e . I t d e s c r i b e s H e w l e t t - P a c k a r d ( F e d e r a l s u p p o r t

C o d e 8 0 5 3 7 ) D i s t o r t i o n A n a l y z e r M o d e l s 3 3 3 A a n d 3 3 4 A ( f i g . l - l ) . A

b a s i c i s s u e i t e m s l i s t f o r t h i s e q u i p m n e n t i s n o t i n c l u d e d a s p a r t o ft h i s m a n u a l .

1 - A . 2 I n d e x o f P u b l i c a t i o n s

R e f e r t o t h e l a t e s t i s s u e o f D A P a m 3 1 0 - 4 t o d e t e r m i n e w h e t h e r t h e r e

a r e n e w e d i t i o n s , c h a n g e s , o r a d d i t i o n a l p u b l i c a t i o n s p e r t a i n i n g t o t h e

e q u i p m e n t . D A P a m 3 1 0 - 4 i s a n i n d e x o f c u r r e n t t e c h n i c a l m a n u a l s ,

t e c h n i c a l b u l l e t i n s , s u p p l y m a n u a l s ( t y p e s 7 , 8 , a n d 9 ) , s u p p l y b u l l e t i n s ,

l u b r i c a t i o n o r d e r s , a n d m o d i f i c a t i o n w o r k o r d e r s t h a t a r e a v a i l a b l e t h r o u g h

p u b l i c a t i o n s s u p p l y c h a n n e l s . T h e i n d e x l i s t s t h e i n d i v i d u a l p a r t s ( - 1 0 ,

- 2 0 , - 3 5 P , e t c ) a n d t h e l a t e s t c h a n g e s t o a n d r e v i s i o n s o f e a c h e q u i p m e n t

p u b l i c a t i o n .

l - A . 3 F o r m s a n d R e c o r d s

a . R e p o r t s o f M a i n t e n a n c e a n d U n s a t i s f a c t o r y E q u i p m e n t . U s e e q u i p -

m e n t f o r m s a n d r e c o r d s i n a c c o r d a n c e w i t h i n s t r u c t i o n s i n T M 3 8 - 7 5 0 .

b . R e p o r t o f D a m a g e d o r I m p r o p e r S h i p m e n t . F i l l o u t a n d f o r w a r d

D D F o r m 6 ( R e p o r t o f D a m a g e d o r I m p r o p e r S h i p m e n t ) a s p r e s c r i b e d i n

A R 7 0 0 - 5 8 ( A r m y ) , N A V S A N D A P u b l i c a t i o n 3 7 8 ( N a v y ) , a n d A F R ‘ 7 1 - 4

a i r F o r c e ) .

c . R e p o r t i n g o f E q u i p m e n t M a n u a l I m p r o v e m e n t s . R e p o r t o f e r r o r s ,

o m i s s i o n s , a n d r e c o m m e n d a t i o n s f o r i m p r o v i n g t h i s m a n u a l b y t h e

i n d i v i d u a l u s e r i s e n c o u r a g e d . R e p o r t s s h o u l d b e s u b m i t t e d o n D A F o r m

2 0 2 8 ( R e c o m m e n d e d C h a n g e s t o D A P u b l i c a t i o n s ) a n d f o r w a r d e d d i r e c t

t o C o m m a n d i n g G e n e r a l , U . S . A r m y E l e c t r o n i c s C o m m a n d ,

A T T N : A M S E L - M R - N M P - A D , F o r t M o n m o u t h , N e w J e r s e y 0 7 7 0 3 .

1 - 2 . 1

Page 8: HP-333A-334A
Page 9: HP-333A-334A

TM 11-6625-1576-15

Model 333A/334A Section IParagraphs l-l to l-13

1-1. DESCRIPTION.

1-2. The Hewlett-Packard ModelsDistortion Analyzers are solid state

333A and 334Ainstruments for

measuring distortion and ac voltages. The Models333A and 334A include two control loops that auto-matically tune both legs of a bridge circuit which re -jects the fundamental when the rejection circuit isinitially set within the range of the loops. A high im -pedance detector which operates from 500 Kc to greaterthan 65 Mc provides the capability” of monitoring thedistortion of the amplitude modulation on an rf carrier.

1-3. Distortion levels of O. 1% to 1OO% full scale aremeasured in seven ranges for any fundamental fre -quency of 5 cps to 600 Kc. Harmonics are indicated upto 3Mc. The high sensitivity of these instruments re-quires only O. 3 v rms for the 100% set level reference.The distortion characteristics can be monitored at theOUTPUT connectors with an oscilloscope, a true rmsvoltmeter, or a wave analyzer. The instruments arecapable of an isolation voltage of 400 volts above chassisground.

1-4. The voltmeter can be used separately for generalpurpose voltage and gain measurements. It has a fre -quency range of 5 cps to 3 Mc (20 cps to 500 Kc for 300pv range) and a voltage range of 300 pv to 300 v rmsfull scale.

1-5. The AM detector included in the Model 334A isa broadband dc restoring peak detector consisting of asemiconductor diode and filter circuit. AM distortionlevels as low as O. 3% can be measured on a 3 v to 8 vrms carrier modulated 30% in the standard broadcast

band, and lower than 1% distortion can be measuredat the same level of the carrier up to 65 Mc.

1-6. ACCESSORY FEATURES.

1-7. The accessory available with the 333A and 334ADistortion Analyzers is a voltage divider probe, -hp-Model No. 1000lA. The features of the probe are:

a. 10 megohms shunted by 10 pf, giving 10:1attenuation.

b. DC to 30 Mc bandwidth.

c. 2% division accuracy.

d. 600 v peak input.

e. 5 ns rise-time.

1-8. OPTION.

1-9. Option 0l is a standard -hp- Model 333A or 334Awith a special meter and meter amplifier, compen-sated to permit response to VU (volume units)characteristics.

1 - 1 1 . M o d i f i c a t i o n s

1 - 1 2 , S p e c i f i c a t i o n C 1 0 - 3 3 4 A i s a s t a n d a r d334A Dis tor t ion Analyzer modif ied by p lac ingthe INPUT and OUTPUT terminals on the rearp a n e l o f t h e i n s t r u m e n t . T h e r e a r t e r m i n a l sa r e b i n d i n g p o s t c o n n e c t o r s a n d a r e i np a r a l l e l w i t h t h e f r o n t p a n e l t e r m i n a l s .D u e t o t h e r e a r t e r m i n a l l e a d s , t h e s h u n tc a p a c i t a n c e o f t h e i n s t r u m e n t i s i n c r e a s e dby 20pF .

1-3

Page 10: HP-333A-334A
Page 11: HP-333A-334A

TM 11-6625-1576-15Model333A/334A Section II

Paragraphs 2-1 to 2-19

2-1. INTRODUCTION.

S E C T I O N I I

I N S T A L L A T I O N

2-2. This section contains information and instruc-tions necessary for the installation and shipping of theModels 333A/334A Distortion Analyzers. Includedare initial inspection procedures, power and groundingrequirements, installation information, and instruc -tions for repackaging for shipment.

2-3. INSPECTION.

2-4. This instrument was carefully inspected bothmechanically and electrically before shipment. Itshould be physically free of mars or scratches and inperfect electrical order upon receipt. To confirmthis, the instrument should be inspected for physicaldamage in transit. Also check for supplied accessories,and test the electrical performance of the instrumentusing the procedure outlined in Paragraph 5-5.

2-5. POWER REQUIREMENTS.

2-6. The Model 333A/334A can be operated from anyac source of 115 or 230 volts (+10%), at 50- 1000 cps.With the instrument disconnected from the ac powersource, move the slide (located on the rear panel) un-til the desired line voltage appears. The instrumentcan be battery operated by connecting two 28-50 Vbatteries (rated 80 milliamperes) to the battery ter-minal on the rear panel. Power dissipation is 10 wattsmaximum.

2-7. THREE-CONDUCTOR POWER CABLE.

2-8. To protect operating personnel, the NationalElectrical Manufacturers’ Association (NEMA) recom-mends that the instrument panel and cabinet be grounded.All Hewlett-Packard instruments are equipped with athree -conductor power cable, which when plugged in-to an appropriate receptacle, grounds the instrumentThe offset pin on the power cable three-prong connec-tor is the ground wire.

2-9. To preserve the protection feature when operat-ing the instrument from a two-contact outlet, use athree-prong to two-prong adapter and connect thegreen pigtail on the adapter to ground.

2-10. INSTALLATION.

2-11 The Model 333A/334A is fully transistorized;therefore, no special cooling is required. However,the instrument should not be operated where the am-bient temperature exceeds 55 C (191 F).

2-12. BENCH INSTALLATION.

2-1% The Model 333A/34A is shipped with plasticfeet and tilt stand in place, ready for use as a benchinstrument.

2-14. RACK INSTALLATION.

2-15. The Model 333A/334A may be rack mounted byusing the 5“ RackMount Kit (-hp- Part No. 5060-0775).Instructions for the conversion are included with thekit. The rack mount for the Model 333A/334A is anELA standard width of 19 inches. when mounted in arack using the rack mount kit, additional support atthe rear of the instrument should be provided if vibration or similar stress is likely.

2-16. REPACKAGING FOR SHIPMENT.

2-17. The following paragraphs contain a general guidefor repackaging of the instrument for shipment. Referto Paragraph 2-18 if the original container is to beused; 2-19 if it is not.

2-18. If original container is to be used, proceed asfollows:

a. Place instrument in original container if avail-able.

b. Ensure that container is well sealed with strongtape or metal bands.

2-19. If original container is not to be used, proceedas follows:

a. Wrap instrument in heavy paper or plastic be-fore placing in an inner container.

b. Place packing material around all sides of in-strument and protect panel face with cardboard strips.

c. Place instrument and inner container in a heavycarton or wooden box and seal with strong tape ormetal bands.

d. Mark shipping container with “DELICATEINSTRUMENT," "“RAGILE,” etc.

2-1

Page 12: HP-333A-334A

TM 11-6625-1576-15

Section III Model 333A/334AFigure 3-1

Figure 3-1. Front and Rear Panel Description

3 - 0

Page 13: HP-333A-334A

TM 11-6625-1576-15

Model 333A/334A Section IIIParagraphs-3-l to 3-15

SECTION III

OPERATING INSTRUCTIONS

3-1. INTRODUCTION.

3-2. The Models 333A and 334A Distortion Analyzersmeasure total harmonic distortion from 5 cps to 600Kc. Harmonics up to 3 Mc are included. The sharpelimination characteristics, >80 db, the low level ofinstrument induced distortion, and the meter accuracyof the 333A and the 334A result in accurate measure-ment of low level harmonic content in the input signal.

3-3. An RMS voltmeter is inherent in the 333A and334A, The voltmeter provides a full scale sensitivityof 300 u volts rms (residual noise <25 u volts). Thevoltmeter frequency range is from 5 cps to 3 Mc ex-cept on the 0. 0003 volt range, which is from 20 cps to500 Kc.

3-4. CONTROLS AND INDICATORS.

3-5. Figure 3-1 illustrates and describes the functionof all front and rear panel controls, connectors, andindlcators. The description of each component iskeyed to a drawing included within the figure.

3-6. ADJUSTMENTS OF MECHANICAL ZERO.

3-7. The procedure for adjustment of mechanicalzero is given in Section V, Paragraph 5-25.

3-8. GENERAL OPERATING INFORMATION.

3-9. INPUT CONNECTIONS.

3-10. Signal source can be connected to the 333A and334A through twisted pair leads or a shielded cablewith banana plug connectors. Keep all test leads asshort as possible to avoid extraneous pickup fromstray ac fields, When measuring low-level signals,battery operation is recommended to avoid groundloops. Another method for avoiding ground loops isby connecting only one instrument in a test setupdirectly to power line ground through a NEMA (three-prong) connector, Connect all other instruments tothe power source through a three-prong to two-prongadapter and leave the pigtail disconnected. Both the333A and 334A have a dc isolation of ±400 vdc fromthe external chassis with the shorting bar, (item 16 ,Figure 3-1), disconnected.

3-11. VOLTMETER CHARACTERISTICS.

9-12. The RMS VOLTS markings on the meter face arebased on the ratio between the average and effective(rms) values of a pure sine wave. The ratio of aver-age to effective values in a true sine wave is approxi-mately O. 9 to 1. When the meter is used to measurecomplex waves, the voltage indicated may not be therms value of the signal applied. This deviation ofmeter indication exists because the ratios of averageto effective values are usually not the same in a com -plexwave as in a sine wave. The amount of deviationdepends on magnitude and phase relation betweenharmonics and fundamental frequency of the signal

applied. Table 3-1 shows the deviation of the meterindication of a sine wave partly distorted by harmonic..As indicated in the table, harmonic content of lessthan approximately 10% results in very small errors.

Table 3-1

Effect of Harmonics on Voltage Measurements

Input Voltage True MeterCharacteristic RMS Value Indication

Fundamental = 100 100 100

Fundamental +10% 100. s 1002nd harmonic

Fundamental +20% 102 100-1022nd harmonic

Fundamental +50% 112 100-1102nd harmonic

Fundamental +10% 100.5 9 6 - 1 0 43rd harmonic

Fundamental +20% 102 94-1083rd harmonic

Fundamental +50% 112 90-1163rd harmonic

NOTE

This chart is universal in applicationsince time errors are inherent inall average-responding type voltage-measuring instruments.

3-13. In distortion measurements where the fundamen-tal frequency is suppressed and the remainder of thesignal is measured, the reading obtained on an average-responding meter may deviate from the true total rmsvalue. When residual wave contains many inharmoni-cally related sinusoids, the maximum error in the dis-tortion reading is about 11% low for distorilon levelsbelow 10%.

Measured Maximum Error TotalDistortion In Meter Indication Distortion

2. 5% +0. 11 X O. 025 = 0.025 +0. 0027 =0.00027 0.0277 or 2.8%

This example represents the maximum possible error,and in most cases the error is less. In distortionmeasurements, the reading of an average-respondingmeter is sufficiently close to the rms value to besatisfactory under most measurement conditions.

3-14. USE OF OUTPUT TERMINALS.

3-15. The OUTPUT terminals provide a O. 1 v rmsoutput for full scale meter deflection These ter-minals can be used to monitor the output signal withan oscilloscope, a true rms voltmeter, or a waveanalyzer. The combination of the distortion meterand oscilloscope provides more significant information

3-l

Page 14: HP-333A-334A

3-18. DISTORTION MEASUREMENT.

TM 11-6625-1576-15

Section IIIParagraphs 3-16 to 3-22

about the device under test than the expression of dis-tortion magnitude alone. Information obtained fromthe oscilloscope pattern is specific and reveals thenature of the distortion that sometimes occurs at suchlow levels that it is difficult to detect in the presenceof hum and noise. The impedance at the OUTPUTterminals is 2000 ohms, therefore, capacitive loadsgreater than 50 pf should ‘be avoided to maintain theaccuracy of meter readings.

3-16. OPERATING PROCEDURES.

3-17. The 333A and 334A Distortion Analyzers can beoperated from an ac power source ( 115/230 volt) or adc power source (+28 to +50 and -28 to -50 volt source).If a dc source is used, check the -25 v output. Ifnecessary, adjust the power supply according to theprocedures in Paragraph 5-27.

3- 19. DISTORTION MEASUREMENT IN PERCENT,AUTOMATIC MODE.

a. Position FUNCTION selector to SET LEVEL.

b. Position MODE switch to MANUAL.

c. If fundamental frequency is 1 Kc or greater,position HIGH PASS FILTER SWITCH to IN.

d. Rotate SENSITIVITY selector to MIN position.

NOTE

The bandwith of the SENSITIVITY se-lector is reduced in the two extremeCCW positions (positions used with aninput signal greater than 30 v).

e. Position METER RANGE selector to SET LEVEL100%).

f. Connect test leads from device under test toINPUT terminals.

REMOVE SHORTING STRAP BETWEENFLOATING GROUND ~ AND CHASSISGROUND (+) TERMINALS ON FRONTPANEL INPUT TERMINALS WHENMEASURING DISTORTION BETWEENTWO POINTS WHICH ARE BOTH ABOVEGROUND POTENTIAL.

g. With SENSITIVITY VERNIER control max. CCW,position SENSITIVITY selector for meter indicationgreater than 1/3 full scale.

NOTE

If unable to adjust for full scale deflec-tion which indicates input signal is be-low O. 3 volts, use manual mode andposition METER RANGE selector down-scale. Use this new position as the 100%SET LEVEL position, thus making thenext range 30%, etc.

h. Adjust SENSITIVITY VENIER control for fullscale deflection.

j. Position FRIQUENCY RANGE selector andfrequency dial to fundamental frequency of input signal.

3-2

Model 333A/334A

k. Position FUNCTION selector to DISTORTION,

m. Adjust METER RANGE selector and frequencydial vernier control for minimum meter indication.

n. Adjust COARSE and FINE BALANCE controlsfor further reduction of meter indication. Meter indi-cation must be less than 10%of SET LEVEL indication.

p. Position MODE switch to AUTOMATIC.

r. Observe percentage of distortion indicated onmeter. Meter indication is in conjunction with METERRANGE selector. For example, if meter indicates.4 and METER RANGE selector is on 1% position,distortion measured is 0.4%.

3-20.

a.

b.

NOTE

RMS voltage of input signal being ana-lyzed for distortion can be measuredby positioning FUNCTION selector toVOLTMETER position.

DISTORTION MEASUREMENT IN PERCENT,MANUAL MODE.

NOTE

In MANUAL mode the accuracy of dis -tortion measurements is affected byfrequency stability of the input signal.An inaccuracy in distortion indicationsoccure when the frequency drift of the in-put signal exceeds the bandwidth of therejection curve.

Perform steps a through n in Paragraph 3-19.

Repeat steps m and n until no further reductionin meter- indication can be obtained.

c. Observe percentage of distortion indicated onmeter.

3-21. DISTORTION MEASUREMENT IN DB,AUTOMATIC MODE .

a. Perform steps a through g of Paragraph 3-19.

b. Adjust SENSITIVITY VERNIER control for 0 dbmeter indication.

c. Perform steps j through 3-19.

d. Observe meter indication for distortion in db.

NOTE

Distortion in db is obtained by alge-braically adding meter indication todb indicated by METER RANGE se-lector: for example, If meter indi-cates -2 and METER RANGE selec-tor is on -20 db position, distortionmeasured is -22 db.

3-22. DISTORTION MEASUREMENT IN DB, MANUALMODE .

NOTE

Notes in Paragraphs 3-20 and 3-21apply.

a. Perform steps a through g of Paragraph 3-19.

b. Adjust SENSITIVITY VERNIER control for O dbmeter indication.

Page 15: HP-333A-334A

Model 333A/334A

c. Perform steps j through n of Paragraph 3-19.

d. Repeat steps m and n until no further reductionn meter indication can be obtained.

e. Observe meter indication for distortion in db.

3-23. DISTORTION MEASUREMENT OF AM RF

CARRIERS. (334A Only)

a. Set NORM - RF SET selector to RF SET.

OBSERVE MAXIMUM INPUT VOLTAGESAS INDICATED ON REAR PANEL.

b. Connect Input to RF INPUT on rear panel.

c. Adjust SENSITIVITY VERNIER control for O dbmeter indication.

d. Perform steps j through n of Paragraph 3-19.

e. Refer to Paragraph 3-20 for manual measure-ment in percent.

f. Refer to Paragraph 3-21 and 3-22 for automaticand manual measurement in db.

-24. VOLTMETER MODE.

NOTE

If DBM measurements are to be made,the DB markings on the METER RANGEswitch must each be lowered by 10.That is, the DB marking for the O. 3 vrange becomes -10 DBM, 1 v range be-comes O DBM, 3 v range becomes +10DBM etc. If the other DB markings areused, the DBM readings will be 10 DBMhigh.

TM 11-6625-1576-15

Section IIIParagraphs 3-23 to 3-27

3-25. The 333A and 334A perform as general purposeAC Voltmeters when the FUNCTION selector is setto VOLTMETER position.

NOTE

With the FUNCTION selector in VOLT-METER position, the SENSITIVITY se-lector is disabled.

a, Position METER RANGE selector to a rangeexceeding the value of the signal to be measured.

b. Connect signal to INPUT terminals.

c. Select a METER RANGE to give a reading asclose to full scale as possible and observe meterindication.

3-26. OUTPUT TERMINALS.

3-27. In VOLTMETER or SET LEVEL position of theFUNCTION selector, the 333A/334A can be used as alow distortion, high gain, wideband preamplifier. Aportion of the meter input (O. 1 v RMS output for fullscale meter deflection) is provided at the OUTPUTterminal J2. In the DISTORTION position, the distor-tion is provided for monitoring purposes.

DO NOT EXCEED THE VOLTAGESLISTED BELOW TO PREVENT BLOWINGFUSE F2:

VOLTMETER.= -1 V RANGEAND BELOW, AND DISTORTIONANALYZER, MODE-MAXIMUMSENSITIVITY.1. 300 V ABOVE 100 CPS2. 50 V ABOVE 1 KC

IF LOW FREQUENCIES ARE NOT TOBE MEASURED, Cl MAY BE REPLACEDWITH A SMALLER CAPACITOR, ANDTHE VOLTAGE LIMITS OF F2 MAY BERAISED ACCORDINGLY.

3-3

Page 16: HP-333A-334A

F i g u r e 4 - 1 . B l o c k D i a g r a m

T M 1 1 - 6 6 2 5 - 1 5 7 6 - 1 5

S e c t i o n I VF i g u r e 4 - 1 Model 333A/334A

Page 17: HP-333A-334A

TM 11-6625-1576-15

Model 333A/334A Section IVparagraphs 4-1 to 4-12

SECTION IV

THEORY OF

4-1. OVERALL DESCRIPTION.

4-2. Models 333A and 334A Distortion Analyzers in-clude an impedance converter, a rejection amplifier,a metering circuit, and a power supply, The Model334A also contains an AM detector. A block diagramof the instruments is shown in Figure 4-1. The im -pedance converter provides a low noise input circuitwith a high input impedance independent of sourceimpedance placed at the INPUT terminals. The re-jection amplifier rejects the fundamental frequency ofan input signal and passes the remaining frequencycomponents on to the metering circuit for measuringdistortion. The metering circuit provides visualindications of distortion and voltage levels on the frontpanel meter, M 1. The AM detector (Model 334A only)detects the modulating signal from the RF carrier andfilters any RF components from the modulating signalbefore it is applied to the impedance converter circuit.

4-3. BLOCK DIAGRAM DESCRIPTION.

4-4. DISTORTION MEASURING OPERATION.

4-5. For distortion measurement, the input signal isapplied to the impedance converter, Assembly A2,through the FUNCTION selector, S1, and the one meg-ohm attenuator, The one megohm attenuator, a voltagedivider network provides 50 db attenuation in 10 dbsteps. The desired level of attenuation is selected bythe SENSITIVITY selector, S2. The impedance con-verter provides an impedance conversion and unitygain between the instrument INPUT terminals and theinput of the rejection amplifier. The rejection ampli-fier consists of a preamplifier, a Wien bridge, and abridge amplifier. The SENSITIVITY VERNIER control,at the input of the preamplifier, provides a set levelsignal to obtain a full scale reading on the meter forany voltage level at the input of the instrument. Withthe FUNCTION selector in the SET LEVEL position, aground is applied in the Wien bridge circuit to allowa signal reference level to be set up on the meter.With the FUNCTION selector in the DISTORTION posi-tion, the Wien bridge is used as an interstate couplingnetwork between the preamplifier and bridge amplifier.The Wien bridge is tuned and balanced to reject thefundamental frequency of the applied input signal. Twoautomatic control loops consisting of two phase detec -tors, lamp drivers, lamps, and photocells provide finetuning and balance in the AUTOMATIC MODE. Theremaining frequency components are applied to thebridge amplifier and are measured as distortion bythe metering circuit. Negative feedback from thebridge amplifier to the preamplifier narrows the re-jection response of the Wien bridge. The output ofthe rejection amplifier is applied to the metering cir -cuit through the post-attenuator. The post-attenuatoris used to limit the input signal level applied to themetering circuit to 1 mv for full scale deflectlon.The metering circuit sensitivity is increased to 300uV for full scale deflection on the 300 pv range. Themetering circuit provides a visual indication of the

OPERATION

distortion level of the input signal. In addition to thevisual indication provided by the meter, the OUTPUTterminals provide a means of monitoring the distortioncomponents.

4-6. DISTORTION MEASUREMENT IN AMCARRIERS.

4-7. The Model 334A Distortion Analyzer contains anAM detector circuit for measuring envelope distortionin AM carriers. The input signal is applied to the in-put of the AM detector circuit where the modulatingsignal is recovered from the RF carrier. The signalis then applied to the impedance converter circuitthrough the one megohm attenuator and then throughthe same circuits previously described in the distor-tion measuring mode operation.

4-8. VOLTMETER OPERATION.

4-9. In the voltmeter mode of operation, the inputsignal is applied to the impedance converter circuitthrough the 1:1 and 1000:1 attenuator. The 1:1 atten-uation ratio is used in the .0003 to .3 VOLTS positionof the METER RANGE selector S3, and the 1000:1attenuation ratio is used in the 1 to 300 VOLTS posi-t ions. With the FUNCTION selector in the VOLT-METER position, the output of the impedance converterbypasses the rejection amplifier and is applied to themetering circuit through the post-attenuator (METERRANGE selector). Metering circuit sensitivity is in-creased from 1 mv for full scale deflection to 300 uvon the 300 uv range, as it was in the distortion mea-suring operation. The function of the post-attenuatorand metering circuit is the same for voltmeter opera-tion as for the distortion measuring operation.

4-10. SCHEMATIC THEORY.

4-11. IMPEDANCE CONVERTER CIRCUIT.

4-12. The input signal to the distortion analyzer isapplied to the impedance converter circuit (refer toFigure 6-2) through the 1:1 and 1000:1 attenuatorS3R12 in the voltmeter mode of operation and throughthe one megohm attenuator S2R1 through S2R6 in thedistortion mode of operation. Capacitive dividersS2C10 through S2C 10 in the attenuator keep the frequencyresponse flat. The impedance converter is a lowdistortion, high input impedance amplifier circuitwit h gain independent of the source impedance placedat the INPUT terminals. Instrument induced distor-tion of the signal being measured is minimized bykeeping the input impedance and the gain of the imped-ance converter linear. The input impedance is keptlinear by use of local positive feedback from the sourceof A2Ql to the gate of A2Q1 and to the protective diodesA2CR2 and A2CR3. Thus signals with large sourceimpedance can be measured accurately. Overall in-duced distortion is further minimized by a high openloop gain and 100% negative feedback. The high openloop gain is achieved by local positive feedback from

4-1

Page 18: HP-333A-334A

TM11-6625-1576-15

Section IVParagraphs 4-13 to 4-25 and Figure 4-2

the emitter of A2Q3 to the collector of A2Q2. Overallnegative feedback from the emitter circuit of A2Q4 tothe source of A2Q1 results in unity gain from the im-pedance converter.

4-13. The bias points of the transistors in the imped-ance converter are selected to minimize instrumentinduced distortion. A2Q 1, an extremely low noise,high impedance field effect transistor, is the majorcomponent that makes linearity of the Impedance con-verter independent of the signal source impedance.

4-14. REJECTION AMPLIFIER CIRCUIT.

4-15. The rejection amplifier circuit (see Figures 6-3and 6-6) consists of the preamplifier (A3Q1) thru A3Q3),the Wien bridge resistive leg and auto control loop(A5Q1 thru A5Q9 with associated lamp and photocell),the reactive leg and auto control loop (A5Q10 thruA5Q 18 with associated lamp and photocell), and thebridge amplifier (A3Q4 thru A3Q6).

4-16. PREAMPLIFIER CIRCUIT.

4-17. The signal from the impedance converter is ap-plied to the preamplifier, which is used during SETLEVEL and DISTORTION measuring operations. Neg-ative feedback from the junction of A3R1O and A3R11 isapplied to the junction of A3R2 and A3C2 to establishthe operating point for A3Q1. Negative feedback fromthe emitter of A3Q3 is applied to the emitter of A3Q1to stabilize the preamplifier. The preamplifier, likethe impedance converter, is designed for high openloop gain and low closed loop gain to minimize instru-ment induced distortion.

4-18. WIEN BRIDGE CIRCUIT.

4-19. In the distortion measuring operation the Wienbridge circuit is used as a rejection filter for thefundamental frequency of the input signal. With theFUNCTION selector, S1, in the DISTORTION position,the Wien bridge is connected as an interstage couplingnetwork between the preamplifier circuit and the bridgeamplifier circuit. The bridge is tuned to the fundamen-tal frequency of the input signal by setting the FRE-QUENCY RANGE selector, S4, for the applicable fre-quency range, and tuning the capacitors C4A throughC4D. The bridge circuit is balanced by adjusting theCOARSE balance control, R4, and the FINE balancecontrol, R5. In the AUTOMATIC MODE fine tuningand balancing are accomplished by photoelectric cellswhich are in the resistive and reactive legs of theWien bridge. The error signals for driving the photo-cells are derived by detecting the bridge output usingthe input signal as a reference.

4-20. When the Wien bridge is not tuned exactly to thefrequency to be nulled, a portion of the fundamentalfrequency will appear at the bridge output. The phaseof this signal depends on which leg of the bridge is nottuned, or on the relative errors in tuning if neither isset correctly. The magnitude of the signal is propor-tional to the magnitude of the tuning error of either orboth legs of the bridge.

4-2

Model 333A/334A

Figure 4-2. Bridge Waveforms

4-21. Figure 4-2a is a sinusoid input to the Wien bridge.If the resistive leg of the bridge iS slightly unbalanced,the output of the bridge is very small, but has thewaveform shown in Figure 4-2b and is in phase withthe input. As the resistive leg is tuned, the signalapproaches zero amplitude at null and then becomeslarger, but 180° out of phase, if the null position ispassed. When the resistive leg is correctly tunedand the reactive leg is tuned through null, a similarwaveform is produced, Figure 4-2c. The only differ-ence is that the reactive signal is 90° out of phase withthe resistive signal.

4-22. When the bridge output is detected using theinput signal as the reference, the error signals inphase or 180° out of phase with the reference developa voltage which is used to vary the resistance in theresistive leg of the bridge, to tune it to the correctnull position, Signals of the form in Figure 4-2c donot develop any voltage as the resistive detector isinsensitive to input differing from the reference by 90° .

4-23. In an independent, but similar control loop, thebridge input signal is shifted 90° and used as thereference signal for the detector. This detectordevelops control voltages to null the reactive leg ofthe bridge, but is insensitive to signals of the formin Figure 4-2b which are caused by small tuningerrors of the resistive branch.

4-24. The result is that the two control loops deriveinformation from a common source and develop twoindependent control signals for nulling the two legsof the bridge. These control voltages are used tovary the brilliance of lamps, which in turn causesresistance changes in photocells which form part ofthe Wien bridge.

4-25. Refer to Figure 4-3 for the phase relationshipof the bridge error signal and reference voltage at thebase of A5Q4. The shaded portions of the error sig-nals (b and c) indicate that part of the error signalwhich contributes to the dc lamp control voltage. Asindicated in d, any error signal that is 90° out ofphase with the reference does not affect the dc lampcontrol voltage.

Page 19: HP-333A-334A

TM 11-6625-1576-15

Model 333A/334A Section IVParagraphs 4-26 to 4-27 and Figure 4-3

4-26. The operation of the reactive branch control loopis similar to that of the resistive branch. The phasedelay circuit (Figure 6-6), A5Q15, A5Q16, S4AF andS4C1 through S4C5, shifts the reference voltage 90º,as shown in Figure 4-3f. This makes the detectorA5Q12 sensitive to components of the bridge errorsignal that are 90° out of phase (g and h). The outputof the lamp driver, Q14, controls the brilliance ofA6DS2, which varies the resistance of A6V2 throughA6V5 to tune the branches of the reactive leg. DeckAR of the FREQUENCY RANGE switch, S4, switchesA5R56 in parallel with A5R55 on the top three frequencyranges. A6DS2 will become brighter, and lower the

resistance of A6V2 through A6V5, making variationin resistance less than on the two lower ranges.However, less variation in resistance is needed totune the leg, because the impedance in the reactive legbecomes progressively less as the higher frequencyranges are selected.

4-27. Any error signal that is not an integral multipleof 90 is the result of the reactive leg of the bridgebeing detuned, and the resistive leg being unbalanced..For example, an error signal that is 45º out of phase(Figure 4-3e and j) will result in outputs from bothdetectors to tune the bridge and reject the fundamental.

Figure 4-3. Reference and Error Phase Relationship4-3

Page 20: HP-333A-334A

TM 11-6625-1576-15

Section IVParagraphs 4-28 to 4-35 and Figures 4-4 to 4-5

Figure 4-4. Wien Bridge Circuit and Rejection Characteristics

4-28. When the bridge circuit is tuned and balanced,the voltage and phase of the fundamental, which appearsat junction of the series reactive leg (S4R1, 3, 5, 7,or 9 and C4A/B) and the shunt reactive leg (S4R11,13, 15, 17, or 19, and C4C/D), is the same as at themidpoint of the resistive leg (A3R12 and A3R14). Whenthese two voltages are equal and in phase, the funda-mental frequency will not appear at the drain of thefield effect transistor A3Q4. For frequencies otherthan the fundamental, the reactive leg of the Wienbridge offers various degrees of attenuation and phaseshift which cause a voltage at the output points of thebridge. This difference voltage between the reactiveleg and resistive leg is amplified by A3Q4, A3Q5, andA3Q6. Figure 4-4 illustrates a typical Wien bridgecircuit and the rejection characteristics for it.

4-29. The Wien bridge circuit is designed to cover acontinuous frequency range of over a decade for eachposition of the FREQUENCY RANGE selector, S4. S4provides course tuning of the reactive leg by changingthe bridge circuit constants in five steps at 1 decadeper step. For the automatic control loop, the referencevoltage is taken from R6 at the input to the rejectionamplifier and applied to the buffer amplifier A5Q7.The reference voltage is amplified and clipped by A5Q8and A5Q9 and coupled to the detector A5Q4. The outputof the metering circuit, which contains the fundamentalfrequency if either leg of the bridge is untuned, is ap-plied to the buffer amplifier A5Q1. It is amplified byA5Q2 and A5Q3 and coupled to the detector A5Q4.

4-30. Refer to Figure 4-5 simplified partial schematicfor detector operation. The discussion is applicableto both resistive and reactive detector circuits.

4-31. The signals from the error amplifier, (A5Q2and A5Q3) will be equal and of opposite phase, and willcancel out each other when the detector, A5Q4 is off.However, when the positive half of the referencesquare wave gates A5Q4 on, the signal from the col-lector of A5Q3 will be shorted to ground. Thus thesignal from the collector of A5Q2 will be coupledthrough the filter network to the base of AQ5. If thesignal from A5Q2 is in phase with the reference, the

4-4

positive half of the signal will be passed, and if it isout of phase, the negative half will be passed.

4-32. The normal working voltage at TP3 is between0 and -1 volt. The dc output of the filter networkcauses the voltage at TP3 to go in a positive direction(toward zero) for in phase error signals, and in anegative direction (toward -1 v) for out of phase errorsignals. The change in base voltage is then amplifiedby A5Q5 and lamp driver A5Q6. This will change thebrilliance of lamp A6DS1, which will vary the resis-tance of A6V1 in the direction necessary to balancethe resistive leg of the bridge.

Figure 4-5. Auto Control Loop Detector

4-33. When the FUNCTION selector is set to the VOLT-METER or SET LEVEL position, the junction of theseries and shunt reactive branches of the Wien bridgeis connected to circuit ground through R19 by S1BFwhich disables the frequency rejection characteristicof the bridge circuit. With the bridge circuit disabled,the rejection amplifier circuit provides one db of gainfor the fundamental frequency and the harmonics. Inthe SET LEVEL operation, this signal is used to es-tablish the SET LEVEL reference.

4-34. BRIDGE AMPLIFIER CIRCUIT.

4-35. The bridge amplifier circuit consists of threestages of amplification, A3Q4 through A3Q6. The

Page 21: HP-333A-334A

TM 11-6625-1576-15

Section IVParagraphs 4-36 to 4-45 and Figure 4-6

Figure 4-6. Rejection Amplifier Block Diagram and Typical Frequency Rejection Characteristic

first stage of amplification, A3Q4, is a field effecttransistor which amplifies the difference signal betweenthe gate and the source. The field effect transistoris selected for maximum noise performance with thehigh impedances of the Wien bridge circuit. The signalfrom the drain is applied to the two stage feedback am-plifier A3Q5 and A3Q6. The output of A3Q6 is coupledto the meter circuit by the post attenuator S3R1 throughS3R11. Negative feedback from the output of the bridgeamplifier is applied to the preamplifier circuit to nar-row the frequency rejection characteristic. It can benoted from the rejection characteristic (refer to Figure4-4) for the bridge that the rejection of harmonic volt-ages is not constant. Typically the second harmonicis attenuated several db more than the third harmonicand the third more than the fourth. The result of thenegative feedback is illustrated by the rejection charac-teristic shown in dashed lines on the attenuation and

phase characteristic of Figure 4-4. Figure 4-6 showsa simplified block diagram of the rejection amplifierwith the typical frequency-rejection characteristic.Refer to Figure 4-7, Bandwidth Versus Null Depth forfurther detail on the rejection characteristic.

4-36. HIGH PASS FILTER.

4-37. The HIGH PASS FILTER (see Figure 6-3) isnormally used when the fundamental of the input signalis greater than 1 Kc. In the voltmeter mode of opera-tion, the filter is not used. In the SET LEVEL andDISTORTION position of the FUNCTION switch thefilter presents >50 db attenuation to 50 or 60 cyclehum components, but offers no attenuation to frequen-cies over 1 Kc. The filter assembly, A7, consists ofA7C1, A7C2, and A7L1. The filter can be insertedor bypassed by the HIGH PASS FILTER switch, S9.

4-38. METER CIRCUIT.

4-39. The meter circuit (refer to Figure 6-4) consistsof the post attenuator, the meter amplifier circuit,and the meter rectifier circuit.

4-40. POST ATTENUATOR.

4-41. The post attenuator, S3R1 through S3R11, is aseries of resistive networks which attenuate the inputsignal in 10 db steps. The attenuator is used in con-junction with either the input sensitivity attenuator or

the 1000:1 attenuator to limit the signal level to themeter amplifier to 1 mv for full scale deflection onall ranges from 1 mv to 300 v full scale. The metercircuit sensitivity is increased to 300 uv for full scaledeflection on the 300 uV range by switching resistorsA2R29 and A2R30 into the calibration network. Resis-tor A2R41 and capacitor A2C29 are also switched into

the calibration network on the 300 Pv range to extendthe passband of the amplifier.

4-42. METER AMPLIFIER CIRCUIT.

4-43. The meter amplifier circuit consists of a fivestage amplifier circuit, A2Q5 through A2Q9, whichdevelops the current for full scale meter deflection.Negative dc feedback from the emitter circuit ofA2Q8 is applied to the base of A2Q5 to stabilize thedc operating point of the meter amplifier circuit andto minimize the tendency for dc drift due to ambienttemperature changes. A2R51 and A3CR8 are electric-ally in the circuit only when the meter circuit is over-loaded. When the voltage on the emitter of A2Q9 be-comes abnormally large during an overload, A2CR8breaks down and provides a lower resistance chargingpath for A2C15 which reduces the transient recoverytime of the meter circuit. Negative ac feedback isapplied from the collector circuit of A2Q9 to the emit-ter circuit of A2Q5. This feedback is used to ensureflat frequency response, to improve linearity, and toreduce the effect of variation of transistor parameterswith environmental changes. In this manner, thecalibration of the instrument is made dependent onhigh quality passive components.

4-44, METER RECTIFIER CIRCUIT.

4-45. The meter rectifier is connected in a bridge typeof configuration with a diode in each upper branch anda dc milliammeter connected across the midpoints ofthe bridge. The simplified meter rectifier is illu-strated in Figure 4-8. The generator represented byA2Q5 through A2Q9 with the internal impedance RO

provides the meter, M1, with current for full scaledeflection and develops a voltage across the calibrationnetwork which closes the ac feedback loop. CapacitorsA2C27 and A2C28 are used as coupling capacitors forthe ac feedback loop, output signal to the OUTPUTconnector, and the bridge error signal to the input ofthe automatic fine tuning loops. The mechanical inertiaof the meter and A2C26 prevents the meter from re-sponding to individual current pulses. Therefore,the meter indication corresponds to the average valueof the current pulses rather than the peak value. Themeter is calibrated to indicate the rms value of a sinewave. Resistor A2R45 impresses a fixed bias acrossdiodes A2CR6 and A2CR7 (biasing them close to thebarrier voltage) to make the meter circuit responselinear to large variations in signal amplitude. Thelinearity of this type of circuit is also increased byincluding the meter circuit in the overall feedbackloop.

4-5

Page 22: HP-333A-334A

TM 11-6625-1576-15

Section IVFigure 4-7

Model 333A/334A

Figure 4-7. Bandwidth Versus Null Depth

4 - 6

Page 23: HP-333A-334A

TM 11-6625-1576-15

model 333A/334A Section IVParagraphs 4-46 to 4-52 and Figure 4-8

Figure 4-8. Simplified Metering Circuit

4-46. POWER SUPPLY CIRCUIT.

4-47. The power supply circuit (refer to Figure 6-5)consists of a +25 voIt series regulated supply and a-25 volt series regulated supply which is the referencesupply for the +25 volt supply.

4-48. The -25 volt regulated supply is of the conven-tional series regulator type. The amplifier A1Q5 isused to increase the loop gain of the circuit, thus im-proving voltage regulation. The positive feedback ap-plied to the junction of A1R1l and A1R12 is used tofurther improve the line frequency suppression of thecircuit.

4-49. The +25 volt regulated supply is of the conven-tional series regulator type and operates the same asthe -25 volt regulated supply.

4-50. Diodes A1CR5 and A1CR6 are coupling and pro-tection diodes for external battery supplies. The diodesprotect the series regulator circuits from applicationof incorrect polarity at the battery input terminals.The diodes also protect external batteries from beingcharged when the ac power is being used with batteriesconnected to the battery terminals.

4-51. RF DETECTOR CIRCUIT. (334A only)

4-52. The RF detector circuit consists of a rectifierA4CR1 and filter circuit shown on the schematic dia-gram of Figure 6-2. The RF signal is applied to thecircuit through the RF INPUT connector on the rearpanel. The rectifier diode A4CR1 recovers the modu-lating signal from the RF carrier and the filter circuitremoves any RF components before the signal is ap -pIied to the impedance converter circuit through theNORM -RF DET switch, S7.

4-7

Page 24: HP-333A-334A

TM 11-6625-1576-15

Section VTable 5-1

Model 333A/334A

Table 5-1. Test Equipment Required

5 - 0

Page 25: HP-333A-334A

TM 11-6625-1576-15

Model 333A/334A Section VParagraphs 5-l to 5-9 and Figure 5-1

SECTION VMAINTENANCE

5-1. INTRODUCTION. c. Set Test Oscillator (-hp - Model 651A) frequencvto 5 KC and adjust amplitude for indication of 1.0 volt

5-2. This section contains maintenance and serviceinformation for the 333A and 334A Distortion Analyz -ers. Included are Performance Checks, Adjustmentand Calibration Procedures, and TroubleshootingTechniques.5-3. TEST EQUIPMENT REQUIRED.

5-4. Test equipment used in the calibration of the 333Aand 334A is given in Table 5-1, Test Equipment Re-quired. This table lists the type of equipment to beused, required characteristics, and recommendedcommercially available test equipment.5-5. PERFORMANCE CHECKS.

5-6. The Performance Checks are in-cabinet proce-dures that can be used to verify instrument perform-ance, These procedures can be used for periodicmaintenance, to check specifications after a repair,or for incoming quality control inspection.

5-7. Performance Checks for both the 333A and 334Aare provided. The performance checks are applicableto both instruments except where noted in paragraphheading.

5-8. The Performance Checks are performed with theac power cord connected to nominal line voltage (115v/230 V) 50 to 1000 cps, MODE SWITCH to MANUAL,HIGH PASS FILTER to OUT, and NORM RF DET switchto NORM, unless otherwise specified. Selector posi-tions for the SENSITIVITY Selector will be referredto as follows: MIN=Position 1, next step-Position 2,etc, to Max= Position 6.

5-9. FUNDAMENTAL REJECTION CHECK

a.

b.

S e t

Connect 33lA/332A as shown in Figure 5-1.

Set Distortion Analyzer controls as follows:

FUNCTION Selector . . . . . . VOLTMETERMETER RANGE Selector . . . . . . . 1 VOLTFREQUENCY RANGE Selector . . . . . . X1OOFrequency Dial . . . . . . . . . . . . . . . 50f i l t e r ( W h i t e I n s t r L a b M o d e l2640) . . . . . . . . . . . . . . . . . . . . . . . . .5Kc

on Distortion Analyzer meter.

d. Switch Distortion Analyzer FUNCTION to SETLEVEL. Switch METER RANGE to O. 3 VOLTS. AdjustSENSITIVITY controls for full scale meter indication.

e. Set Wave Analyzer controls (-hp - Model 302A)as follows:

SCALE VALUE . . . . . . . . . . RELATIVEMAX INPUT VOLTAGE . . . . . . . . ...1RANGE . . . . . . . . . . . . . . 0 DECIBELSMODE SELECTOR . . . . . . . . . .NORMAL

f. Adjust Wave Analyzer FREQUENCY controlsfor maximum meter reading (approximately 1 Kc).

g. Adjust Wave Analyzer REF ADJUST for O dbmeter indication.

h.

NOTE

If the range of the REF ADJUST controlis insufficient to set meter to O db refer-ence, adjust the Distortion AnalyzerSENSITIVITY VERNIER control slightlyto set reference.

Switch Distortion Analyzer FUNCTION to DIS-TORTION. Adjust BALANCE controls and FREQUEN-CY dial for null indication on Distortion Analyzermeter. Reduce Distortion Analyzer METER RANGEselector setting as necessary to maintain deflectionon meter scale.

j. After achieving null indication on DistortionAnalyzer meter, observe Wave Analyzer mete r andcontinue to adjust for a null indication. Reduce WaveAnalyzer RANGE setting as necessary to maintaindeflection on meter scale.

NOTE

The Distortion Analyzer OUTPUT is a Oto O. 1 volt signal representing a percent-age of full scale, regardless of RANGE

5-1

Figure 5-1. Test Setup for Fundamental Rejection Check

Page 26: HP-333A-334A

TM 11-6625-1576-15

Section VParagraphs 5-10 to 5-11, Figure 5-2 and Table 5-2

setting. Each decrease of the RANGEswitch represents 10 db fundamental re-jection. Therefore, the total fundamentalrejection is the sum of the Wave Analyzerreading and the 333A/334A indication.

k. The Distortion Analyzer METER RANGE settingplus the Wave Analyzer RANGE setting plus the twometer indications shall total more than -80 db.

5-10. SECOND HARMONIC ACCURACY CHECK.

a. Connect Test Oscillator (-hp-Model 651A) 600Ωoutput to Distortion Analyzer *

b. Set Distortion Analyzer controls as follows:

FUNCTION Selector . . . . . . . SET LEVELFREQUENCY RANGE . . . . . . . . . . . X1Frequency Dial . . . . . . . . . . . . . . .20METER RANGE Selector . . . . . . . . . 0 DB

c. Set Test Oscillator controls (-hp- Model 651A)as follows:

FREQUENCY RANGE . . . . . . . . . . . X10FREQUENCY Dial. . . . . . . . . . . . . 2OUTPUT ATTENUATOR . . . . . 1.0 VOLTSAMPLITUDE . . . . . . . . . . . . . . l Volt

d. Adjust Distortion Analyzer SENSITIVITY con-trols for a meter reading of O db.

e. Switch Distortion Analyzer FUNCTION selectorto DISTORTION. Adjust Frequency dial and BALANCEcontrols for a null indication on meter.

f. Switch Distortion Analyzer FUNCTION selectorto SET LEVEL.

g. Adjust Test Oscillator frequency to 40 cps. Ad-just AMPLITUDE control for a O db indication on theDistortion Analyzer meter.

h. Switch Distortion Analyzer FUNCTION selectorto DISTORTION. The meter reading shall not changemore than ±0. 6 db.

Model 333A/334A

j. Repeat steps a through h at the frequency settingsindicated in Table 5-2. The meter readings shallchange within the limits specified.

Table 5-2. Second Harmonic Accuracy Check

5-11.

a.

b.

c.

DISTORTION INTRODUCED BY INSTRUMENTCHECK AND AUTOMATIC CONTROL LOOPOPERATION.

Connect 333A/334A as shown in Figure 5-2.

Set Distortion Analyzer controls as follows:

FUNCTION Selector . . . . . . . SET LEVELSENSITIVITY Selector . Position 1 step CCW

from Full CW positionSENSITIVITY VERNIER Control . . full CCWMETER RANGE Selector . . . . . . . . . . 0 dbFREQUENCY RANGE Selector . . . . . . . XlFrequency Dial . . . . . . . . . . . . . . .5

Set oscillator for approximately 1 volt output at5 cycles.

d. Set filter box for 5 cycles.

e. Adjust oscillator amplitude for an indication of+2 db on the Distortion Analyzer meter.

f. Switch Distortion Analyzer FUNCTION selectorto DISTORTION. Adjust frequency dial and BALANCEcontrols for a null meter indication. (If reading is inlower 1/3 of meter scale, decrease METER RANGEselector setting. )

g. The meter indication at “null” shall be at least-8 db on the -60 db METER RANGE which is equiva-lent to -70 db. Note reading.

Figure 5-2. Instrument Induced Distorition and Automatic Control Loop Test Setup

5 - 2

Page 27: HP-333A-334A

Model 333A/334A

h. Set METER RANGE to O. 01 volt RANGE and off-set frequency dial to a lower reading so that metercads full scale. Set MODE switch to AUTOMATICid note distortion level. Distortion level should be

within +3 -O db of manually nulled reading.

j. Set MODE switch to MANUAL (and meter rangeto o. 01). Offset frequency dial past null to a higherdial reading so that meter reads full scale. ReturnMODE switch to automatic. Distortion reading shouldbe within +3 -O db of manually nulled reading obtainedin step g of this paragraph.

k. Set MODE switch to MANUAL and adjust fre -quency dial for null. Adjust COURSE BALANCE Con-trol CW with METER RANGE set at 0. 01 so that meterreads full scale. Return MODE Switch to AUTOMATIC.Distortion reading should be within +3 -0 db of manuallynulled reading obtained in step g of this paragraph.

m. Set MODE switch to MANUAL and METERRANGE to O. 01. Adjust COARSE BALANCE ControlCCW so that meter reads +2 db. Set Mode switch toAutomatic. Distortion reading should be within +3 -Oof manually nulled reading obtained in step g ofthis paragraph.

n. Repeat steps b through m with controls set asindicated in Table 5-3. Except in steps j thru m useMETER RANGE setting of O. 03 to obtain +2 db readingwhen detuning Frequency Dial and COARSE BALANCEControl to verify automatic control loop operation.

Table 5-3. Distortion Introduced By Instrument Check

5-12. FREQUENCY CALIBRATION ACCURACYCHECK.

b.

c.

TM 11-6625-1576-15

Section VParagraph 5-12, Table 5-3, and Figure 5-3

NOTE

From 5 cps to 10 cps the FREQUENCYdial may be as much as 3% low. In thistest the dial is held constant, and the in-put frequency is varied and monitored.If the dial is low, the input frequency atnull will be high. If the period of the in-put frequency is measured, it will be lowwhen the dial is low. From 200 Kc to 600Kc the dial may be as much as 8% high,Consequently, a low frequency reading atnull would indicate that the dial is readinghigh.

Set Distortion Analyzer controls as follows:

FUNCTION Selector . . . . . . . SET LEVELMETER RANGE Selector . . . . . . . . 0 DBSENSITIVITY Selector . . . . . . . . . . MIN.FREQUENCY RANGE Selector . . . . . .XlFrequency Dial . . . . . . . . . . . . . . . .5

Set Test Oscillator controls (-hp- Model 203A)as follows:

FREQUENCY RANGE . . . . . . . . . . . X1FREQUENCY DIAL . . . . . . . . . . . . . . 5OUTPUT ATTENUATOR . . . . . . 1.0 VOLT

d. Set Electronic Counter controls (-hp - Model5532A) as follows:

SENSITIVITY . . . . . . . . . . 3 VOLTS RMSFunction Switch. . . . . 1 PERIOD AVERAGEDDISPLAY . . . . . . . . . . . . . . Fu1l CCW

e. Adjust Test Oscillator AMPLITUDE control fora full scale indication on the Distortion Analyzer meter.

f. Switch Distortion Analyzer FUNCTION selectorto DISTORTION.

g. Adjust Test Oscillator FREQUENCY DIAL fora null indication on the Distortion Analyzer meter.(If reading is in lower 1/3 of meter scale, decreaseMETER RANGE selector setting, )

h. Adjust Distortion Analyzer BALANCE controlsfor a null indication on the meter. Repeat steps g and

a. Connect 333A/334A as shown in Figure 5-3.

Figure 5.-3. Test Setup for Frequency Calibration Accuracy Check

5-3

h until a null is reached.

Page 28: HP-333A-334A

TM11-6625-1576-15

Section vParagraphs 5-13 to 5-16 and Table 5-4

Model 333A/334A

j. The Electronic Counter shall indicate the periodof 5 cps -3%, i. e. 194 to 200 msec.

k. Repeat steps b through h with controls set asindicated in Table 5-4. The Electronic Counter shallindicate the Test Oscillator output frequencies withinthe limits indicated.

5-13. INPUT RESISTANCE CHECK

a. Connect Test Oscillator (-hp- Model 651A) 600Ω output to Distortion Analyzer INPUT terminals.

b. Set Distortion Analyzer controls as follows:FUNCTION Selector . . . . . . VOLTMETERSENSITIVITY Selector . . . . . . . . . MAX.SENSITIVITY VERNIER Control . . . . . MIN.METER RANGE Selector. . . . . . . 1 VOLT

c. Set Test Oscillator controls as follows:

FREQUENCY RANGE . . . . . . . . . . X10FREQUENCY Dial . . . . . . . . . . . . . 10OUTPUT ATTENUATOR . . . . . . 1.0 VOLT

d. Adjust Test Oscillator AMPLITUDE control foran indication of 1.0 volt (Ein) rms on the DistortionAnalyzer meter.

e. Connect a 100 K Ω (Ra) ±1. 0%, 1/2 watt, fixedcarbon film resistor in series with the DistortionAnalyzer INPUT. Note the Distortion Analyzer meterreading (Eo).

f. Calculate the Distortion Analyzer input resis-tance using the following formula:

g. The input resistance shall be 1 M Ω ±5%.

h. Switch FUNCTION selector to Distortion andcalculate the input resistance in this position. It shallbe 1 M Ω ±5%.

5-14.

a.

b.

INPUT SHUNT CAPACITANCE CHECK.

Set Distortion Analyzer controls as follows:

FUNCTION Selector . . . . . . . VOLTMETERMETER RANGE Selector . . . . . . . 1 VOLT

Connect an L - C meter to the 333A/334A andmeasure the input capacitance.

c. The L - C meter shall indicate less than 30 pf.

d. Switch the Distortion Analyzer on the 0. 3 rangeand measure capacitance. Meter shall indicate lessthan 60 pf.

5-4

e. Set Distortion Analyzer controls as follows:.FUNCTION Selector . . . . . .DISTORTIONSENSITIVITY Selector . . . . . . . . . . MIN.METER RANGE Selector . . . . . . . VOLTS

f. Measure Capacitance at each SENSITIVITYselector setting of the Distortion Analyzer. The L - Cmeter shall indicate less than 60 pf on each of thesesettings.

5-15. MINIMUM INPUT LEVEL CHECK.

a. Connect Test Oscillator (-hp- Model 651A) 600Ω output to Distortion Analyzer INPUT terminals l

Terminate test oscillator with600 Ω ±1% 1/2 w resistor.

b.

c.

d.

Set Distortion Analyzer controls as follows:

FUNCTION Selector . . . . . . . VOLTMETERMETER RANGE Selector . . . . . . .3 VOLTSENSITIVITY Selector . . . . . . . . . MIN.SENSITIVITY VERNIER . . . . . . . . , CCW

Set Test Oscillator for 20 cps.

Adjust Test Oscillator amplitude for a DistortionAnalyzer- meter indication O. 3 volts.

e. Switch Distortion Analyzer FUNCTION selectorto SET LEVEL.

f. Switch SENSITIVITY selector to MAX andVERNIER to full CW. The SENSITIVITY controls shallhave sufficient range to give a full scale meter reading.

5-16.

a.

b.

DC ISOLATION CHECK.

Connect 333A/334A as shown in Figure 5-4.

REMOVE SHORTING BARS BETWEENPOWER LINE GROUND TERMINALS ONDISTORTION ANALYZER INPUT TERMI-NALS AND FUNCTION GENERATOR OUT-PUT TERMINALS.

Set Distortion Analyzer controls as follows:

FUNCTION Selector . . . . . .VOLTMETERMETER RANGE Selector . . . . . . . 1 VOLT

c. Apply ac power to dc power supply and set for400 v. Set Power SuppIy controls but do not apply dcto the Distortion Analyzer.

d. Set Function Generator for 1 Kc and adjust theamplitude control for an indication of 0. 9 on the Dis-tortion Analyzer meter.

Page 29: HP-333A-334A

TM 11-6625-1576-15

Model 333A/334A Section VParagraphs 5-17 to 5-18 and Figures 5-4 to 5-5

Figure 5-4. DC Isolation Check Test Setup

e. Apply the 400 v dc to the Distortion Analyzer. g. Check the Distortion Analyzer voltmeter track-There shall be no change in the indication on the Dis - ing at 0. 1 volt increments from 0. 1 volt to 1 volt. Thetortion Analyzer meter, or any indication on the am- voltmeter tracking accuracy shall be within ± 1%.meter of the power supply. h. Repeat steps d through g with the variable Line

5-17. VOLTMETER ACCURACY CHECK

a. Connect Voltmeter Calibrator (-hp- Model 738B)and Variable Line Transformer (Superior Type UC1M)to Distortion Analyzer as shown in Figure 5-5. Re-move shorting bar between power line ground and cir-cuit ground terminals.

b, Set Variable Line Transformer for 115 v output.

c. Set Distortion Analyzer FUNCTION Selector toVOLTMETER.

d. Set voltmeter calibrator for 400 cps output.

e. Check the Distortion Analyzer voltmeter fullscale readings on all ranges against the appropriaterms input voltages from the voltmeter calibrator. Thevoltmeter accuracy shall be within ±2%,

f. Set the Distortion Analyzer METER RANGEselector to 1 VOLT range.

Transformer set to 105 v & 125 v.

5-18. HIGH PASS FILTER CHECK

a.b.

Connect the 333A/334A as shown in Figure 5-6.

Set Distortion Analyzer controls as follows:

FUNCTION Selector. . . . . . . SET LEVELMETER RANGE Selector . . . . SET LEVELHIGH PASS FILTER switch . . . . . . . InSENSITIVITY Selector . . . . . . . . Position 5SENSITIVITY VERNIER . . . . . . . CCW

c. Adjust frequency response test set to 5Kc andset output amplitude to obtain a zero db indication onDistortion Analyzer.

d. Adjust frequency response test set meter to setlevel reference.

e. Set frequency response test set to 1 Kc and adjustoutput amplitude so that test set meter reads set leveI.

Figure 5-5. Voltmeter Accuracy check Test Setup

5 - 5

Page 30: HP-333A-334A

TM 11-6625-1576-15

Section V Model 333A/334AParagraphs 5-19 to 5-21, Figure 5-6 and Table 5-5

f. Reading on Distortion Analyzer shall be withinO. 5 db of zero db setting.

g. Set frequency response test set to 60 cps andadjust the output amplitude so that test set meter readsset level.

h. Switch Distortion Analyzer Meter Range to 0. 003volt range,

j. Reading should be > -40 db.

5-19, VOLTMETER FREQUENCY RESPONSE CHECK.

a. Connect Distortion Analyzer to test equipmentas shown in Figure 5-6.

b. Set Distortion Analyzer controls as follows:

FUNCTION Selector . . . . . . . .VOLTMETERMETER RANGE Selector. . . . . .0.01 VOLTS

c. Set the Variable Line Transformer output to 115 v.

d. Adjust the Oscillator for an indication of O. 9 at400 cps on the Distortion Analyzer meter.

e. Adjust the Frequency Response Test Set METERSET control to SET LEVEL indication on the meter.

f. Switch the Oscillator RANGE switch to X 1 andset the Frequency Dial to 5.

g. Readjust the Oscillator AMPLITUDE controluntil the Frequency Response Test Set meter indicatesSET LEVEL.

h. The Distortion Analyzer meter shall indicatebetween 0.855 and 0.945 (±5%).

j. Set the Oscillator to the frequencies listed inTable 5-5. Repeat step g after each setting. The Dis -tortion Analyzer meter shall indicate 0.9 ± the toler-ances indicated.

k. Switch the Frequency Response Test Set RANGESELECTOR to the 1-3 Mc position. Set the FREQ.TUNING dial to the frequencies listed in Table 5-5.Adjust the Frequency Response Test Set AMPLITUDEcontrol until the meter indicates SET LEVEL aftereach frequency setting. The Distortlon Analyzer metershall indicate 0.9 ± the tolerances indicated.

5-6

n. Repeat steps b through k with the Variable LineTransformer set to 105 v and 125 v.

Table 5-5. Voltmeter Frequency Response Check

5-20. RESIDUAL NOISE CHECK

a. Connect a shielded 600 Ω resistor across theDistortion Analyzer INPUT terminals. (See Figure5-7 for details on constructing shielded load. ) Securethe shorting bar between the power line ground and cir-cuit ground terminals.

b. Set Distortion controls χοντρολσ as follows:

FUNCTION Selector . . . . . . . VOLTMETERMETER RANGE Selector . . . . 0.0003 VOLTS

c. The meter shall indicate less than 25 µ volts.

d. Remove the 600 Ω resistor. Connect a shielded100 K ohm resistor across the INPUT terminals. (SeeFigure 5-7 for details on constructing shielded load. )

e. The meter shall indicate less than 30 µ volts.

5-21. AM DETECTOR CHECK (Model 334A only).

a. Connect Signal Generator (hp- Model 606A) 50RF OUTPUT to Distortion Analyzer RF INPUT.

b. Set Distortion Analyzer controls as follows:

FUNCTION Selector . . . . . . . SET LEVEL NORM-RF DET Switch . . . . . . . RF DETMETER RANGE Selector . . . . . . . . . 0 DBFREQUENCY RANGE Selector . . . . . X100FREQUENCY Dial . . . . . . . . . . . . . 10

Page 31: HP-333A-334A

TM 11-6625-1576-15

Model 333A/334A Section VParagraphs 5-23 to 5-27, Figure 5-7 and Table 5-6

No.

1

2

3

4

4

5

6

7

8

Description

Connector, male

Connector, male w/insulator

Lug, terminal 90°

Resistor, 100 K Ω, 1/2 W, 5%metal film

Resistor, 600 K Ω, 1/4 w, O. 5%metal film

Washer, int. lock

Spacer, 6-32 threaded

Shield

Screw, bind. head, 6-32 x 1/4 in

-hp- Part No.

1251-0174

1251-0175

0360-0042

0758-0053

0757-1037

2190-0007

0380-0058

1251-1073

2470-0001#

Figure 5-7. Shielded Load Assembly

c. Set Signal Generator controls as follows:

FREQUENCY. . . . . . . . . . . . . . . l McMODULATION SELECTOR . . . . . . . EXTATTENUATOR/VERNIER . . . . . 3.O VOLTSModulate Signal Generator output 30% with a 1000cps signal using a 203A Function Generator orcomparable oscillator.

d. Adjust Distortion Analyzer SENSITIVITY Select-or and VERNIER control for 0 DB reference.

e. Switch FUNCTION selector to DISTORTION. Ad-just BALANCE controls and Frequency Dial for nullindication.

f. Distortion shall be more than -40 db down.5-22. ADJUSTMENT AND CALIBRATION

PROCEDURE.

5-23. The following is a complete adjustment and cali-bration procedure for the 333A and 334A. Beforeproceeding with these adjustments, the PerformanceChecks should be performed to determine whetheradjustments are necessary, If your instrument doesnot meet the test limits specified in the followingsteps, refer to Troubleshooting Procedure, Paragraph5-33. Adjustment locations are shown in Figure 5-8.5-24. The Adjustment and Calibration Procedure isperformed with the ac power cord connected to nomi-nal line voltage ( 115 v/230 v) 50 to 1000 cps, MODE

switch to MANUAL, HIGH PASS FILTER to OUT, andNORM RF DET switch to NORM, unless otherwisespecified.

5-25. METER MECHANICAL ZERO SET.

5-26. The meter is properly zero-set when the pointerrests over the zero calibration mark on the meterscale and the meter is in its normal operating environ-ment and turned off, Adjust the zero-set, if necessary,as follows:

a. Rotate mechancial zero-adjustment screw clock-wise until meter pointer is to the left of zero and mov-ing upscale toward zero.

b. Continue to rotate adjustment screw clockwise;stop when pointer is exactly at zero. If the pointerovershoots zero, repeat step a.

c. When pointer is exactly on zero, rotate adjust-ment screw approximately 2 degrees counterclockwise.This is enough to free the zero adjustment screw fromthe meter suspension. If pointer moves during thisstep, repeat steps a through c.

5-27. POWER SUPPLY AND BIAS ADJUSTMENTS

a. Refer to Table 5-6 for the power supply and biasmeasurement points, adjustments, voltage limits andline regulation specifications.

Table 5-6. Power Supply Adjustments

* Referenced to -25 volt supply; -25 volt supply must be adjusted first.* * This voltage has been adjusted for optimum distortion performance. Do not readjust unless A2Q1 is changed.‡ This voltage has been adjusted for optimum distortion performance, Do not readjust unless A3Q4 is changed.

5-7

Page 32: HP-333A-334A

TM11-6625-1576-15

Section V Model 333A/334AFigure 5-8

.

Figure 5-8 Component and Adjustment Location5-8

Page 33: HP-333A-334A

Model 333A/334A

b. Remove top and bottom covers.

c. Secure the shorting bar between the power lineground and circuit ground INPUT terminals.

d. Connect common lead of VTVM (-hp- Model412A) to either INPUT ground terminals.

e. Perform the measurements and adjustmentslisted in Table 5-6.

DO NOT ALLOW PROBE TO SHORT TP1TO GROUND. IF THIS HAPPENS, A2Q4MAY BE DESTROYED.

5-28. A3R16 AND A3R30 DISTORTION ADJUST.

a. Connect test equipment to the 333A/334A asshown in Figure 5-9.

b. Set Distortion Analyzer controls as follows:

A3R16 ADJ A3R90 ADJ

FUNCTION Selector . . SET LEVEL SET LEVELMETER RANGE Switch . SET LEVEL SET LEVEL

Range RangeFREQUENCY RANGE . . . . . . X1K . . . . X100

SelectorFREQUENCY Dial. . . . . . . . . . 5 . . . . . . 50SENSITIVITY Selector . . . Position 5 . . Position 5SENSITIVITY VERNIER . . . . CCW . . . . CCW

c. Adjust Filter Set to 5 Kc.

d. Set Function Generator for 5 Kc and adjust theoutput amplitude for +2 db reading on the DistortionAnalyzer.

e. Switch Distortion Analyzer to DISTORTION andobtain best null (minimum of 40 db down from +2 dbreference). Decrease METER RANGE switch asnecessary.

f. Set MODE switch to Automatic and METERRANGE switch to 0.0003 range and adjust A3R16 forminimum distortion reading (minimum of 70 db downfrom +2 db reading on the Distortion Analyzer).

NOTE

When adjusting A3R16, the voltage atA3TP2 must remain within the limits of+ 19.9 to +20. 5 volts.

TM 11-6625-1576-15

Section VParagraphs 5-28 to 5-29 and Figure 5-9

g. Repeat steps b through f, adjusting A3R30 in-stead of A3R16.

5-29. BRIDGE BALANCE ADJUSTMENT (C3).

MAINTAIN THE LEAD DRESS TO C3 ANDTO THE TUNING CAPACITOR, C4. ANYCHANGE IN LEAD DRESS WILL CAUSE ACHANGE IN CAPACITANCE.

a. Connect the Test Oscillator to the DistortionAnalyzer.

b. Set the Distortion Analyzer controls as follows:

FUNCTION Selector . . . . . . . SET LEVELMETER RANGE Switch . . SET LEVEL RangeFREQUENCY RANGE Selector . . . . . X1KSENSITIVITY Selector . . . . . . .Position 5

c. Turn the slot on trimmer capacitor C3 parallelwith the side casting of the Distortion Analyzer.

d. Set the Test Oscillator to 5 Kc and Adjust theamplitude to give an indication of +2. 0 db on the Dis -tortion Analyzer.

e. Set the Distortion Analyzer FUNCTION selectorto DISTORTION. Adjust the Frequency dial and BAL-ANCE controls for the best null. (Null must be at least90 db down from +2. 0 db reference. ) After the Dis-tortion Analyzer is nulled, the COARSE BALANCEcontrol should be near the center of its extremes. ifit is near either end, there will not be enough range onthe BALANCE control at other frequencies.

f. Position the METER RANGE switch and FUNC-TION selector to SET LEVEL

g. Set the Test Oscillator to 60 Kc and adjust theamplitude to give an indication of +2. O db on the Dis-tortion Analyzer.

h. Set the Distortion Analyzer FUNCTION selectorto DISTORTION and set the FREQUENCY RANGEswitch to X10K.

j. Without disturbing the BALANCE controls, ad-just the Frequency dial and C3 for best null. (Nullmust be at least 40 db down from +2. 0 db reference.

Figure 5-9. A3R16 and A3R30 Distortion Adjust Test Setup

5-9

Page 34: HP-333A-334A

TM 11-6625-1576-15

Section VParagraphs 5-30 to 5-32

Model 333A/334A

1)

2)

3)

5-30.

a.

NOTE

If the 60 on the Frequency dial is not underthe cursor after step j is completed, referto Figure 7-1 and perform the followingsteps:

Remove set screws (#2) from knob on Frequencydial, and then remove knob (#31). Do not changeC4 setting.

Loosen retaining screws on Frequency dial (#3),and slip dial until 60 is under cursor, beingcareful not to disturb C4 setting.

Tighten retaining screws and replace knob andset screws.

VOLTMETER GAIN ADJUSTMENTS.

Connect Voltmeter Calibrator (-hp- Model738BR) to Distortion Analyzer as shown in Figure 5-5.Remove shorting bar between earth ground and floatingground terminals.

b. Set Distortion Analyzer controls as follows:

FUNCTION Selector . . . . . . . VOLTMETERMETER RANGE Selector . . . .0.001 VOLTS

c. Set Voltmeter Calibrator to 400 ˜ RMS, 1.0millivolt output.

d. Adjust A2R34 (400 cps CAL. 2) for a meter in-dication of 1.0.

e. Set Voltmeter Calibrator to 400 ~ RMS, O. 3millivolt output. Set Distortion Analyzer METERRANGE selector to 0.0003 VOLTS.

f. The meter shall indicate 3.0 ± 1/3 of a minordivision (±O. 55%). If necessary, change A2R30 to setreading within these limits,

NOTE

If transistors are replaced in the meteramplifier, it may be necessary to changeA2R30. Typical values for this resistorare between 140-260 ohms. Increasingresistance decreases the meter readingand vice versa.

g. Perform the Voltmeter Accuracy Check of Para-graph 5-17. The meter shall indicate within thelimits specified.

5-31. VOLTMETER FREQUENCY RESPONSEADJUSTMENT.

a. Connect the test equipment to the 333A/334A asshown in Figure 5-6.

NOTE

Range switch cover must be in place be-fore performing these adjustments.

b. Set the Distortion Analyzer controls as follows:

FUNCTION Selector . . . . . . . VOLTMETERMETER RANGE Selector . . . 0.0003 VOLTS

c. Set the frequency response test set for 400 cpsad adjust the output amplitude for a Distortion Ana-lyzer meter indication of 0.9 on the 1.0 scale.

5-10

d. Adjust the frequency response test set metercontrol for a set level indication

e. Change frequency response test set frequency t'20 cps.

f. Readjust the amplitude until the frequency re-sponse test set meter indicates set level.

g. The Distortion Analyzer meter shall indicate0.9 ± 5%.

NOTE

If reading exceeds these limits, changeA2C25. Typical value for this capacitoris between 50 and 200 µf. Increasing ca-pacitance reduces the meter reading, andvice versa.

h. Switch Distortion Analyzer METER RANGEselector to O. 001 VOLTS, Repeat steps c and d.

j. Change frequency response test set to 5 cps.Repeat step f.

k. The Distortion Analyzer meter shall indicate0.9 ± 5%.

m. Switch the Distortion Analyzer METER RANGEselector to O. 01 VOLTS. Repeat steps c and d.

n. Adjust frequency response test set frequencyfor 2Mc and adjust the output amplitude until the fre-quency response test set indicates set level.

p. Adjust A2R39 (GAIN ADJUST) for a DistortionAnalyzer meter indication of O. 9 on the 1.0 scale.

q. Change the frequency response test set frequen-cy to 3 Mc. Readjust the output amplitude until thefrequency response test set indicates set level.

r. Adjust A2C21 (3 Mc CAL trimmer) for a Dis-tortion Analyzer meter Indication of O. 9 on the 1.0scale.

NOTE

The frequency response adjustments madethus far interact, It will be necessary torepeat steps n thru r until all adjustmentsare within specifications.

s. Switch Distortion Analyzer METER RANGE toO. 3 VOLTS. Repeat steps c and d.

t. Change the frequency response test set frequen-cy to 2.5 Mc. Readjust the output amplitude until themeter indicates set level. Adjust S3C6 (VOLTAGECAL at 2.5 Mc O. 3 v RANGE trimmer) for a Distor-tion Analyzer meter indication of O. 9 on the 1.0 scale.

v. Switch the Distortion Analyzer METER RANGEselector to 3 VOLTS. Repeat steps c and d.

w. Change the frequency response test set frequen-cy to 200 Kc. Repeat step f.

x. Adjust S3C1 (200 Kc CAL) for a Distortion Ana-lyzer indication of O. 9 on the 1.0 scale.

5-32. SENSITIVITY SWITCH CALIBRATION.

a. Connect instruments as shown in Figure 5-6.

Page 35: HP-333A-334A

Model 333A/334A

b.

c.

NOTE

Selector positions for theselector will be referred

SENSITIVITYto as follows:

MIN = Position 1, next step = Position2, etc. , to MAX = Position 6.

Set Distortion Analyzer controls as follows:

FUNCTION Selector . . . . . . . SET LEVELSENSITIVITY Selector . . . . . .Position 5SENSITIVITY VERNIER Control . . . . . MAXMETER RANGE Selector . . . . 0.01 VOLTS

Set Frequency Response Test Set controls (-hp-Model 739AR)as follows:

RANGE SELECTOR . . . . . . . . EXTERNALOUTPUT ATTENUATOR . . . . . . . . 0.03

d. Set Oscillator (-hp- Model 200SR) frequency to400 cps. Adjust AMPLITUDE control for an indica-tion on Distortion Analyzer meter of 0. 9 on the 1. 0)scale.

e. Adjust Frequency Response Test Set METERSET for SET LEVEL indication.

f. Change Oscillator frequency to 100 Kc. Read-just Oscillator AMPLITUDE control for SET LEVELindication on Frequency Response Test Set meter.

g. Adjust S2C 5 for a meter indication of O. 9 on the1.0 scale.

h. Repeat steps d through f, calibrating sensitivityat the settings indicated in Table 5-7.

NOTE

When switching the SENSITIVITY selector,be careful not to move SENSITIVITY VERN-IER control from MAX position.

Table 5-7. Sensitivity Switch Calibration

ISENSITIVITY

ITest Set

IAdjust

Switch ATTENUATOR for 009 IPosition 5 0.03 S2C 5Position 4 0.1 S2C 4Position 3 0.3 S2C 3Position 2 1.0 S2C2Position 1 3.0 S2C 1

5-33. TROUBLESHOOTING PROCEDURES.

5-34. This section contains procedures designed to as-sist in the isolation of malfunctions. These proceduresare based on a systematic analysis of the instrumentcircuits in an effort to localize the problem. Theseoperations should be undertaken only after it has beenestablished that the difficulty can not be eliminated bythe Adjustment and Calibration Procedures. An in-vestigation should also be made to insure that thetrouble is not a result of conditions external to the333A/334A.

THE 333A/334A CONTAINS VERY HIGHIMPEDANCE, HIGH FREQUENCY CIR-CUITS. CONTAMINATION OF THESWITCHES, CIRCUIT BOARDS OR TUN-ING CAPACITORS WILL CAUSE HIGH

TM 11-6625-1576-15 Section V

Paragraphs 5-33 to 5-42 and Table 5-7

IMPEDANCE LEAKAGE PATHS AND SUB-SEQUENT DETERIORATION OF THE PER-FORMANCE OF THE INSTRUMENT.AVOID TOUCHING ANY OF THESE CIR-CUITS WITH THE BARE FINGER, ASSKIN OILS ARE EXTREMELY CONTAMI-NATING. IF HANDLING IS NECESSARY,WEAR CLEAN COTTON OR RUBBERGLOVES. DO NOT USE A PENCIL TOTRACE CIRCUITS IN THE INSTRUMENT.GRAPHITE PENCIL LEAD IS AN EX-TREMELY GOOD CONDUCTOR AND MNACCIDENTLY INTRODUCED PATH OFTHIS TYPE IS SOMETIMES DIFFICULTTO LOCATE. TO AVOID SURFACE CON-TAMINATION OF A PRINTED CIRCUITOR SWITCH, CLEAN WITH A WEAK SO-LUTION OF WARM WATER AND MILD DE-TERGENT AFTER REPAIR. RINSE THOR-OUGHLY WITH CLEAN WATER ANDALLOW IT TO DRY COMPLETELY BE-FORE OPERATING. DO NOT USE ALCO-HOL OR ANY OTHER CLEANING SOLU-TION EXCEPT DETERGENT AND WATER.DO NOT APPLY ANY COMMERCIALMOISTURE SEALING SPRAY TO THEBOARDS. APPLICATION OF THESEAGENTS WILL CAUSE LEAKAGE PATHSAND SUBSEQUENT DETERIORATION TOTHE OPERATION OF THE INSTRUMENT.

5-35. Conduct a visual check of the 333A/334A forpossible burned or loose components, loose connec-tions, or any other obvious condition which mightsuggest a source of trouble.

5-36. Table 5-8 contains a summary of the front panelsymptoms and internal adjustment symptoms thatmight be encountered. It should be used in initialefforts to select a starting point for troubleshootingprocedures.

5-37. Table 5-9, in conjunction with Figure 5-10, con-tains procedures which may be used as a guide inisolating malfunctions. The steps in Table 5-9 de-scribe the normal conditions which should be encoun-tered during the checks (circled numbers N ) inFigure 5-10.

5-38. The checks outlined in Table 5-9 are not designedto measure all the circuit parameters, rather to local-ize a malfunction. Therefore, it is quite possible thatadditional measurements will be required to complete-ly isolate a problem. Circuit parameters may varyslightly between instruments; therefore, it should notbe necessary to precisely duplicate the values given.

5-39, BOTTOM SHIELD REMOVAL.

5-40. The bottom shield is attached by seven screws.Access to six is obtained by removing the bottomcover. Access to the other is obtained by removingthe left (as viewed from the front of instrument) rearside cover.

5-41. SERVICING ETCHED CIRCUIT BOARDS.

5-42. The Model 333A has six-etched circuit boardsand the Model 334A, seven. Use caution when re-moving them to avoid damaging mounted components.

5-11

Page 36: HP-333A-334A

TM 11-6625-1576-15

Section VTable 5-8

Model 333A/334A

5-12

Page 37: HP-333A-334A

TM 11-6625-1576-15

Section VParagraph 5-43 to 5-45

The -hp- Part Number for the assembly is marked onthe circuit board to identify it.

5-43. The etched circuit boards are of the plated-through type. The electrical connection between thetwo sides of the board is made by a layer of metalplated through the component hole. When working onthese boards, observe the following rides:

a. Use a low-heat (25 to 30 watts) small-tip sol-dering iron, and a small diameter rosin core solder.

b. Remove circuit components by placing the sol-dering iron on the component lead on either side of theboard, and pulling up on the lead. If a component isobviously damaged, clip leads off as close to the com-ponent as possible and then remove leads with a sol-dering iron.

EXCESSIVE HEAT CAN CAUSE THE CIR-CUIT AND BOARD TO SEPARATE, ORCAUSE DAMAGE TO THE COMPONENTS.

c. Clean component lead hole by heating the holewith the iron and inserting a wooden toothpick Re-move the toothpick after the solder has cooled andinsert the new component lead

d. To replace components, shape new leads andinsert them in lead holes. Reheat with soldering ironand add a small amount of new solder as required toinsure a good electrical connection.

e. Clean excessive flux from the connection andadjoining area

5-44. SERVICING ROTARY SWITCHES.

5-45. The 333A/334A has five rotary type switchesFUNCTION, SENSITIVITY, FREQUENCY RANGE,MODE, and METER RANGE. When working on theseswitches, observe the following rules:

a. Use a low-heat (25 to 50 watts) small-tip sol-dering iron and a small diameter rosin core solder.

b. When replacing components, attempt to dressthem as nearly to their original alignment as possible.

c. Clean excessive flux from the connection andadjoining area.

d. After cleaning the switch, apply a light coat oflubriplate to the switch detent balls. DO NOT applylubricant to the switch contacts or allow lubricant tocontaminate components.

5-13

Page 38: HP-333A-334A

TM 11-6625-1576-15

S e c t i o n VT a b l e 5 - 9

Model 333A/334A

5-14

Page 39: HP-333A-334A

Figure 5-10. Troubleshooting Logic

5-15

S e c t i o n V

Figure 5-10

T M 1 1 - 6 6 2 5 - 1 5 7 6 - 1 5

Page 40: HP-333A-334A

Figure 5-11 Frequency Tuning Assembly

5-16

Model 33A/334ASection VTable 5-10

TM 11-6625-1576-15

Page 41: HP-333A-334A

TM11-6625-1576-15

Section VIParagraphs 6-1 to 6-2

SECTION VI

SCHEMATICS

6-1. SCHEMATIC DIAGRAMS. Main signal paths and significant feedback paths areidentified (refer to notes on the schematic diagrams).

6-2. This section contains the schematic diagram, An A5 board component location diagram is included,wiring data, and A5 board component location diagram because the A5 components are not identified on thefor the 33A and 334A Distortion Analyzers. The sche- prlnted circuit boards. A partial interconnectingmatic diagram illustrate the circuits contained within diagram is also included to assist in reconnectingeach assembly as well as the assembly interconnections. pc boards.

6-1

Page 42: HP-333A-334A

6-2

Figure 6-1. P/O Internal Wiring Data

Section VIFigure 6-1

Model 333A/334A

TM 11-6625-1576-15

Page 43: HP-333A-334A

By Order of the Secretary of the Army:

Offical:KENNETH G. WICKHAM

Major General, United States Army,The Adjutant General.

HAROLD K. JOHNSON,General, United States Army,

Chief of Staff.

Distribution:Active Army :

USAMB (1)USACDCEC (1)USACDCCEA (1) TOAD (6)USACDCCEA : LEAD (8)

Ft Huachuca (1)

NG : None.USAR: None.For explanation of abbreviations used, see AR 320-50.

* U. S. GOVERNMENT PRINTING OFFICE : 1990 0 - 262-912 (30346)