Vacuum gauges for the fine and high vacuum

Karl Jousten, PTB, Berlin

1. Measurement of vacuum pressures and the calibration chain

2. Overview of measurement principles and gauge types

3. Direct gauges, indirectly measuring gauges

4. Accuracy of vacuum gauges

Page 2: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

2

Measurement of vacuum pressure

The definition of pressure p:

It follows one of the measurement principles.

Fine and high vacuum gauges

AFp = Mass Time

Length

Page 3: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

3

Measurement of vacuum pressure

AFp =

Fine and high vacuum gauges

Gas pressure p

p ≈ 0

Balance asforce meter

F

A

This is a traceable instrument usable as primary standard

Page 4: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

4

Errors and uncertainties

Fine and high vacuum gauges

Error: A wrong reading of a gauge. Adeviation from a true value defined by the SI units.

Uncertainty: The possible range by which a reading may not reflect the true value defined by the SI units.

True value

Indicated value

Error/deviation

Possible true value Indicated value

Uncertainty

Page 5: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

5

Measurement of vacuum pressure

AFp =

Fine and high vacuum gauges

Gas pressure p

p ≈ 0

Balance asforce meter

F

A

Page 6: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

6

The calibration chain

ordinary vacuum gauge

working standard

secondary standard

primary standard

Fine and high vacuum gauges

Errors and uncertainties

Reliability increases

Uncertainty increases

Page 7: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

7

Traceability and primary standardsFully developed primary standards partly developed standards

Fine and high vacuum gauges

Page 8: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

8

Best accuracy available

Relative uncertainties of pressures in primary standards

2E-021E-02

4E-032E-03

3E-04

1E-04

3E-05

7E-063E-06

1E-06

1E-05

1E-04

1E-03

1E-02

1E-01

1E+00

1E-10 1E-08 1E-06 1E-04 1E-02 1E+00 1E+02 1E+04 1E+06

p in Pa

Rel

ativ

e U

ncer

tain

ties

(k=2

)

Mercury ManometerContinuous expansion Series expansion

p atm

Traceability and primary standards

Page 9: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

9

Measurement principles and gauges

Fine and high vacuum gauges

1E-04

1E-03

1E-02

1E-01

1E+00

1E-12 1E-10 1E-08 1E-06 1E-04 1E-02 1E+00 1E+02 1E+04pressure in Pa

Mea

sure

men

t unc

erta

inty

AFp =

Heat conductivity

Viscosity

Ionisation

Ionisation and Suppression of X-ray

and ESD

Piezo PiezoresistiveMembraneMechanical Resonance Capacitance

Piston/Cyl U-tube

Gas independent

Total pressure

Gas dependent-D

ifficult t

o inter

pret

Page 10: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

10

Measurement principles and gauges

Fine and high vacuum gauges

Page 11: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

11

Measurement principles and gauges

Fine and high vacuum gauges

Page 12: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

12

Measurement principles and gauges

Fine and high vacuum gauges

Gas inlet

hgAFp Δ== ρ

The mercury U-tube exists since Torricelli (1644). It is still the most accurate vacuum gauge > 100 Pa (1 mbar)!

Page 13: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

13

Measurement principles and gauges

AFp =

Fine and high vacuum gauges

Gas

The rotating piston gauge

Gap: 0.2µm

Page 14: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

14

Measurement principles and gauges

Fine and high vacuum gauges

M

A

P1 P2

Z

R

Raum 1 Raum 2

Mechanical gauges

3 Groups:

1. Ref.side patm and contains meas.dev.

2. Ref.side p=0, meas.dev. on test side (1)

3. Ref.side p=0 and contain meas.dev.

Page 15: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

15

Measurement principles and gauges

Fine and high vacuum gauges

Mechanical gauges: Bourdon gauges

Page 16: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

16

Measurement principles and gauges

Fine and high vacuum gauges

Mechanical gauges: Bourdon gauges2 Photo cells

Light source

Amplifier

Precise

Resistor

Mirror

Coils

Quartz

spiral

Page 17: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

17

Measurement principles and gauges

Fine and high vacuum gauges

Mechanical gauges: Bourdon gauges

Page 18: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

18

Measurement principles and gauges

Fine and high vacuum gauges

Piezoeffect used by membrane

Page 19: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

19

Measurement principles and gauges

Fine and high vacuum gauges

Piezoresistive effect

Material: Silicon (MEMS)

Page 20: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

20

Measurement principles and gauges

Fine and high vacuum gauges

Capacitance diaphragm gauge

Sensitivity of deflection: 0.4 nm!

Membrane (INVAR, Ceramic): as low as 25 µm.

Two improve zero stability: 2 capacitors plus thermostated housing

Page 21: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

21

Measurement principles and gauges

Fine and high vacuum gauges

Electrical block diagram ofcapacitancediaphragm gauge

Page 22: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

22

Measurement principles and gauges

Fine and high vacuum gauges

Thermal transpiration effect

1

2

1

2TT

pp

=

2

1

1

2TT

nn

=

Cross section

Page 23: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

23

Measurement principles and gauges

Fine and high vacuum gauges

Thermal transpiration effect

036.1296318

=

Page 24: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

24

Measurement principles and gauges

Fine and high vacuum gauges

Resonance Silicon Gauges

Designed by MEMS

Page 25: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

25Fine and high vacuum gauges

Page 26: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

26

Measurement principles and gauges

Fine and high vacuum gauges

Heat conductivity through a gas

Page 27: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

27

Measurement principles and gauges

Uncertainties due to the physical principle of measurementExample: Pirani gauge

Fine and high vacuum gauges

Radiation and conduction

Heat transport by gas

convection

Page 28: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

28

Measurement principles and gauges

Fine and high vacuum gauges

Electrical circuit for Pirani gauge

Constant temperature

Constant heating voltage, current, or power

Page 29: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

29

Measurement principles and gauges

Mikro Pirani (MEMS manufactured) by MKS

Fine and high vacuum gauges

Heated sheet 60°C

MEMS: higher Knudsen number, no convection

Page 30: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

30

Measurement principles and gauges

Correction factor for helium for 4 different Piranigauges

Fine and high vacuum gauges

0,00

0,20

0,40

0,60

0,80

1,00

1,20

1,40

1E-03 1E-02 1E-01 1E+00 1E+01 1E+02p in mbar

Cor

rect

ion

fact

or C

FH

e/N

2

VM1 (Pfeiffer)VM2 (Thyracont)VM3 (MKS)VM4 (Leybold)Mean

Helium

MEMS Pirani

Page 31: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

31

Measurement principles and gauges

Experimental standard deviations of repeat calibrations for 4 different Pirani gauges at various pressures

Fine and high vacuum gauges

# Gauge s in %0,05 mbar 3 mbar 30 mbar

1 Pfeiffer TPR 280 0,19 0,13 0,092 Thyracont VSP52 0,06 0,35 3,303 MKS 925C 0,10 0,12 0,194 Leybold TTR91 0,03 0,09 0,12

Page 32: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

32

Measurement principles and gauges

Thermocouple gauge

Fine and high vacuum gauges

Page 33: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

33

Measurement principles

Fine and high vacuum gauges

Viscosity

( )⎟⎟⎠

⎞⎜⎜⎝

⎛−⎟

⎠⎞

⎜⎝⎛⋅= ωωω

σρπ

πRDd

mkTp

&

208

Page 34: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

34

Measurement principles and gauges

Sprinningrotor gauge

Fine and high vacuum gauges

Page 35: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

35

Measurement principles and gauges

Sprinning rotor gauge

Fine and high vacuum gauges

1,85E-07

1,90E-07

1,95E-07

2,00E-07

2,05E-07

2,10E-07

2,15E-07

404 406 408 410 412 414 416

Rotor frequency f

DC

R-S

igna

l

s-1

s-1

Residual drag vs. frequency of rotor

Page 36: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

36

Measurement principles and gauges

Sprinning rotor gauge

•No gas consumption (e.g. by ionization)

•No dissociation (hot cathode)

•Low outgassing rate

•Predictable reading

•High accuracy

•High long-term stability

Fine and high vacuum gauges

Page 37: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

37

Measurement principles

Fine and high vacuum gauges

Ionisation

Page 38: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

38

Measurement principles

Fine and high vacuum gauges

Fine vacuum IGIonization gauges for different vacuum ranges

Page 39: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

39

Measurement principles

Fine and high vacuum gauges

Ionisation gauges for fine vacuum

Page 40: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

40

Previous investigations showed that TDLAS is applicable forvacuum measurement:

CO, mid-infrared (5 µm), resolution down to 10-5 Pa, high accuracy.

Partial Absorpion, Reference Predicted from ch2

0.0

0.5

1.0

1.5

2.0

2.5

35 40 45 50 55 60 65 70 75 80

time (ms)

Inte

nsi

ty (

a.u

.)

( ) ( ) LnSeII 10

1 Φ−−− = λλ

Fine and high vacuum gauges

Measurement principles

⎟⎟⎠

⎞⎜⎜⎝

−Φ= −

−− )()(ln

)()( 1

10

10

1 λλ

λλ II

LTSkTp

Page 41: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

41

How accurate are vacuum gauges ?

Reasons for inaccuracies of vacuum gauges

General Example

Uncertainties due to calibration chain

Has the vacuum gauge been ever calibrated? Against what standard?

Uncertainties due to installation Pressure at gauge position may not reflect the pressure where the experiment takes place.

Uncertainties due to operation Outgassing of an ion gauge may falsify an outgassing rate measurement.

Inaccuracies caused by the physical principle of measurement

Thermal conductivity or ion gauge is used, but gas mixture is not (accurately) known.

Uncertainties caused by the device itself

See Table 2.

Fine and high vacuum gauges

Page 42: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

42

How accurate are vacuum gauges ?

Reasons for inaccuracies

Gas species dependence:

Real total pressure only for force/area measuring gauges and > 100 Pa (1 mbar)! Below 100 Pa consider the thermal transpiration effect.

Spinning rotor gauges: Use a weighted mean mass, if approximate relative composition is known.

Thermal conductivity gauges and ionisation gauges : Scaling factors are available, but do have high uncertainties.

Fine and high vacuum gauges

2

1⎟⎠⎞

⎜⎝⎛∑==

n

iiieff mam ∑ =

=

n

iia

11

∑==

n

iiieff CFaCF

1

Page 43: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

43

How accurate are vacuum gauges ?Uncertainties due to the vacuum gauge itself

Fine and high vacuum gauges

General ExamplesOffset measurement residual drag in SRG, zeroing of Pirani gauge,

X-ray- and ESD-effect for ion gauges

Offset instability (drift) Offset drifts with environmental temperature (Piroutte effect in SRG), bridge is no more balanced with time

Resolution Number of digits shown

Influences of environment (mainly temperature) Enclosure temperature of Pirani changes varies, thermal transpiration effect changes in CDG, amplifier changes amplification

Non-Linearity Ion gauge (sensitivity changes with pressure)

Integration time (scatter of data), repeatibility Same signal at repeat measurements? Integration time in SRG, in picoammeter with ion gauge.

Reproducibility (stability of calibration constant) Calibration constants change with time.

Hysteresis Mechanical gauges (up, down measurement)

Prior usage, cleanliness Surfaces change, accommodation coefficients change, secondary yield changes

Page 44: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

44

How accurate are vacuum gauges ?

Fine and high vacuum gauges

Table: Relative measurement uncertainty of commercially available vacuum gauges.

Gauge type Measurement range in Pa

Normal uncertainty

Optimum range in Pa

Lowest uncertainty

Piston gauges 10...105 102...105 10-4... 10-5

Quartz-Bourdon-manometer 103...105 103...105 3x10-4... 2x10-4

Resonance silicon gauges 10 ... 105 0.003... 0.0005 100 ... 105 2x10-4... 5x10-5

Mechanical vacuum gauge 102 ... 105 0.1 ...0.01

Membrane vacuum gauge 102 ... 105 0.1 ...0.01

Piezo 102 ... 105 1 ...0.01

Thermocouple gauge 10-1 ... 102 1... 0.3

Pirani gauges 10-1... 104 1 ... 0.1 1 ... 100 0.02 ... 0.01

Capacitance diaphragm gauges 10-4 ...105 0.1... 0.003 10-1 ... 105 0.006... 0.001

Spinning rotor gauges 10-5 ...10 0.1 ... 0.007 10-3 ...10-1 0.006...0.004

Penning gauges 10-7 ... 1 0.5 ... 0.2 10-5 ... 1 0.3...0.1

Magnetron gauges 10-8 ... 1 1 ...0.1 10-6 ... 1 0.1...0.02

Ionisation gauges (Emission cathodes) 10-10 ... 10-2 1...0.05 10-8 ... 10-2 0.2...0.02

Page 45: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

45

How accurate are vacuum gauges ?

Fine and high vacuum gauges

Lowest relative uncertainties for vacuum gauges and primary standards

Errors > 100 % (error factor > 1) are possible.

1

0,2

0,05 0,05

0,0060,0040,004 0,003

0,0020,001

0,1

0,010,02

0,010,02

1E-06

1E-05

1E-04

1E-03

1E-02

1E-01

1E+00

1E-10 1E-08 1E-06 1E-04 1E-02 1E+00 1E+02 1E+04 1E+06

p in Pa

Rel

ativ

e U

ncer

tain

ties

(k=2

)

Mercury Manometer

Continuous expansion Series expansion

p atm

Ionisation gaugeMechanical

Piston

QBS

CDGSRG

Pirani

Page 46: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

46

Todays commercial gauges

Fine and high vacuum gauges

Pirani gauge

Old classical gauges:

Gauge head + controller

Today: Active gauges or transmitter (all in one)

or

Digital gauges (digital output via interface)

Page 47: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

47

Todays commercial gauges

Fine and high vacuum gauges

Profibus ConverterTransmitter gauge plus

Page 48: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

48

Todays commercial gauges

Fine and high vacuum gauges

Commercial „active“ Line vacuum gauges

Page 49: How accurate are vacuum gaugescas.web.cern.ch/sites/cas.web.cern.ch/files/lectures/platjadaro-2006/... · Vacuum gauges for the fine and high vacuum Karl Jousten, PTB, Berlin 1. Measurement

CAS_2006

Fine and high vacuum gauges

We have discussed:

Metrological system - primary standards- calibration chain

Measurement principles and gauges

Direct, indirect measuring gauges

Sources of uncertainties with

values from 0.001% up to 100% or factor