Home Brew X Band Wavemeter

download Home Brew X Band Wavemeter

of 4

Transcript of Home Brew X Band Wavemeter

  • 7/25/2019 Home Brew X Band Wavemeter

    1/4

    M

    Olberg WISNN

    Loretta Road

    M A 2154

    Home Brew

    X Band

    Waveme

    ndispensable test gear

    e need to know the op-

    erating frequency of an

    l in th e

    t with an adjustable cav-

    it y wavemeter and a detec-

    sees this energy and con-

    resonance indicator.

    The wavemeter must

    Photo

    A.

    Cavity wavemeter

    for X-band.

    have some sort of dial

    which can be calibrated in

    frequency. The accuracy of

    this calibration will depend

    upon many factors, such as

    the temperature stability of

    the cavity, the resettability

    of the adjusting system, and

    the exactness o f the calibra-

    tion reference. There are

    others, but these are the

    three which wil l concern us

    while constructing a suit-

    able unit. The other prob-

    lems will become evident

    as we move along.

    A cavity wavemeter

    i s

    a

    parallel resonant circuit

    (see Fig.

    1 .

    You have used

    these devices as traps in the

    output of your transmitters

    on the lower frequencies

    and in many other circuits

    of your equipment. The op-

    eration of a microwave uni t

    isn't different from those.

    The Q of the circuit in

    Fig. 1 Parallel resonant

    equivalent circuit of wrve-

    meter.

    for microwave mavens

    which i t is used and the

    Q

    of

    the wavemeter itself make

    the difference for a cali-

    brated frequency refer-

    ence. A microwave cavity

    such as the device we are

    about to construct can

    have a working

    Q

    of nearly

    3000 The working

    Q

    will

    depend upon the loading of

    the cavity and several other

    factors.

    An important limitation

    to recognize is that the

    device we are making will

    put you within the band

    limits, but it is no t an abso-

    lute frequency meter.

    A cavity wavemeter cir-

    cuit diagram might look

    like the ci rcu it shown i n Fig.

    2

    A difference in the circui t

    when using waveguide for

    the transmission line will

    be, for example, that the

    coupling loop in Fig. 1 is a

    C A V I T Y

    7 5

    A V E G U I D E

    Fig. 2. Cavity wavemeter cir-

    cu it diagram.

    one-turn link to the tank

    c o ~ l n the microwave con-

    f~gu ra t~one wil l use, th~si s

    simply a small hole In the

    E-plane wall of the wave-

    guide The diameter of the ,

    hole determ~nes ow much

    coupli ng there i s to the tank

    clrcuit, wh~ ch,n our case, S

    1

    the cavlty The d~ameter f

    the hole controls coupling

    and many other factors lt

    must be placed carefully so

    that the m atch to the trans-

    mission l ~ n e ill not be up-

    set

    1

    Now, let us examine the

    construction of a cavity

    wavemeter fo r X-band. (See

    Fig. 3. The cavity we will

    construct is, in microwave

    parlance, a right circular

    cylinder cavity operating n

    the

    TE,,,

    mode. It is mount-

    ed on the E-plane of an

    X-band waveguide, off cen-

    ter so that the center of the

    cavity will sit directly cen-

    tered over a hole in the

    waveguide. This hole is

    called an iris hole. The re-

    mainder of the cavity

    bot

    tom, which hangs over the

    waveguide, is covered with

    a brass half-circle soldered

    in place. This is shown in

    73 Magazine March,

    1981

  • 7/25/2019 Home Brew X Band Wavemeter

    2/4

    IOP V IE W, E - P L A N E

    ?G -52 /U WAVEGUIDE 112

    X

    I

    ( OUT S IDE D IM E NS IONS )

    S E E DE T A IL

    m

    \ U G - 3 9 / U F L A N G ES ( 2 R E P ' D I

    DE T A IL ON CY L INDE R

    TOP PLATE AND

    CAVITY OVER IRIS HOLE

    T HRE A DE D COL L E T

    B R A S S

    fMENSIONS ARE IN INCHES

    DO NOT SC 4L E DRAWING

    1/2 COVER PLATE

    FOR CAVITY BOTTOM

    1/16 1 . BRASS SHEET

    DRILL AND TAP

    m f 1 / 4 - 3 6

    3 / 8 - 2 4 T H O

    TO FORM COL LET- SAW CUT PLAC ES 120 '

    APART TO TOP OF PLATE (AF TER TAPPING

    g

    ,

    i

    ?6

    A ND T HRE A DING)

    D R I L L A ND T AP F OR 4 - 4 0 X 1/41 LONG SCREW

    C3AE, , E$20' APAR T TO MAT CH HOL ES ON

    ---TIGHT F IT INSIDE 1 3 8 10 WATER P IPE (COPPER)

    114 ln DIA . BRASS R OD

    1 -3 HD

    B

    D I AL B b R i E L

    3 / 8

    in.

    0.0

    WATER

    CAVITY

    DRIL L 3 P L A CE S P L A S T IC T UB E

    I 2 O 0 A P A R T FO R 4 - 4 0 F L A T

    HE A D S CRE W S , COUNT E RS INK

    i i /H

    F ,

    t

    1/16 A A:4IA

    TUNING PLATE

    PLASTIC DISC

    DE T A IL S HOW ING CA V IT Y

    S OL DE R E D T O W A V E GUIDE DR IL L 1 14 D IA E X A CT

    E ND V IE W

    v

    Fig 3. Construction details for cavity wavemeter.

    -the photos and drawings ac-

    companying this article.

    Tuning of the cavi ty is ac-

    length. This i s done by mov-

    ing a circular plate longi-

    d t o the circular plate

    es as the device t o

    t into the cavi ty. The

    g wi ll

    as a collet- threaded

    is particula r part of

    t care since i t di-

    i s

    used to

    In the design shown, a

    s used. I t i s larger,

    more easily-and for me to

    read without my glasses.

    The disk is constructed

    from a piece of flat brass at

    least 1/16 thick. Dimen-

    sions of the disk shown in

    the drawings should be fol-

    lowed carefully . As you wil l

    note, the dimensions allow

    only .03-inch spacing be-

    tween the wall and the disk

    edge. This means that the

    drive mechanism must also

    be placed in the center of

    the drive holder. In each

    case, i t is best to have your

    friendly machinist fabricate

    these pieces on a lathe. This

    means that the top cover,

    the disk, and disk holder all

    must be centered exactly or

    the disk w il l rub on the cav-

    ity walls.

    The disk must have some

    absorbing material added

    to the back o f it. This is nec-

    essary because the back of

    the cavity

    hat is behind

    the disk-otherwise can

    support resonances which

    wll be coupled weakly from

    the waveguide and wil l pro-

    duce unwanted responses

    which can mislead the user.

    The absorbers may be con-

    structed from a disk made

    of a piece of attenuator

    card. If you have access to

    powdered iron material,

    paper can be shaped

    around the periphery of the

    disk to make a form. Mix

    the powdered iron particles

    with some coil dope and

    pour in enough to make a

    level surface about 1/16

    thick. Be sure that the sur-

    face is held level while the

    mix is hardening. (Pow-

    dered iron can be obtained

    from cup core rnanufactur-

    ers or, if you are really am-

    bitious, you can make your

    own simply by crushing a

    cup core with a steel rol ler.

    Make sure that the particles

    are very small.) After the

    mix has hardened, be sure

    to remove all of the paper

    and glue from the. edge of

    the disk.

    Next, construct the wave-

    guide section by mounting

    the two flanges, one on

    either end of the guide. This

    is done by cutt ing the wave-

    guide ends square and then

    removing all burrs from

    them with a smooth file.

    Each end wil l have a flange

    soldered in place. This is

    done simply by sliding the

    flanges on each end of the

    guide, after first applying a

    small amount of soldering

    paste (flux) to each end.

    Now heat the entire assem-

    bly until the flux runs and

    turns the copper waveguide

    a deeper color. Apply just

    enough solder so that i t wil l

    run around the joint. When

    this operation is completed,

    lap the ends of the flanges

    against a piece of very fine

    emery paper. Now inspect

    the joint to be sure that

    there are no gaps in the

    solder around the face of

    the joint.

    Next, measure the dis-

    tance between the rear

    butts of the flange, and in

    the center of this distance,

    using a square, draw a line

    across the E-plane of the

    waveguide.

    On one side of the guide

    lying on this line, place a

    centerpunch mark at the

    distance from the outside

    wall shown in the mechani-

    cal assembly drawing, Fig.

    3.

    Now drill the hole also

    described in this detail.

    Deburr this hole inside and

    out of the guide so that i t is

    smooth. This is the iris

    7

    Magazine March, 1981 6

  • 7/25/2019 Home Brew X Band Wavemeter

    3/4

    Fig 4. Test setup for calibration, as described in the text.

    couplin g hole, and it should

    be perfectly round with no

    burrs or raised pieces stick-

    irig down into the wave-

    guide. The deburring job is

    easy on the to p of the guide

    and can be done with an-

    other drill bit held by hand

    and turned to li ft off the

    burr. The inside of the

    waveguide may be done

    with a very small bladed

    knife such as an X-actom

    scalpel. This can also be

    done fro m the top of the

    waveguide by inserting the

    point into the hole against

    the burr.

    Clean a piece of copper

    pip e (shown in the assembly

    drawing) as the cavity cyl-

    inder.

    t

    should be free fro m

    all burrs on each end and

    polished clean with Dutch

    Cleanser or a similar pot-

    scouring compound, partic-

    ularly on the inside of the

    cylinder. O n one end of the

    cylinder, using a machinist's

    square and try, set the cen-

    ter so that a pair of marks

    can be scribed across the

    open face of the cylinder

    and on the end of its side

    walls.

    Using f lu x here, too,

    solder the half-circle of

    brass in place so that it lies

    on the t wo scribe marlts just

    made t o the bot tom edges

    of the cylinder. Now the

    bottom of the cavity will be

    half closed. Use acid-base

    so lder ing f lux fo r th is

    operation.

    The cavity may now be

    mount ed to the waveguide

    section by al igning the

    scribe marks on the cavity

    ou ts ide wa l ls w i th the

    scribe line made on the

    face o f t he waveguide. The

    half circle of brass wi ll b ut t

    against one side wall of the

    wavegu ide p r ov id in g a

    solder surface. Using

    a

    C-clamp and a small, flat

    piece of wood, clamp the

    cavity in place, placing the

    wood under the clamp so

    that it does not mar the top

    of the cavity. Adjust the

    cavity so that i t aligns with

    the marks on its walls and

    the waveguide. Make sure

    that the half-circle butts

    tigh tly against the side wall

    of the waveguide. Now ap-

    ply soldering paste to the

    bott om of the cavity and to

    the waveguide surfaces

    that make contact wit h the

    cavity parts. Heat and sol-

    der, using just enough

    solder so tha t it wil l run

    when the assembly is hot

    enough. Let the assembly

    coo l thorough ly be fo re

    removing the clamp.

    When the soldering is

    completed, the waveguide

    section should have the

    cavity mounted off-center

    on the guide with the iris

    hole off-center with the di-

    ameter of the cavity (see

    Fig. 3 details). The cavity

    should be in line with the

    scribe marks previously

    described as the alignment

    lines. If this is not the case,

    but the alignment marks

    a r e o n l y a v er y f e w

    thousands of an inch off , do

    not rework to im~rovehe

    alignment; this small devia-

    tion will not degrade the

    operation of. the wavemeter

    too badly. Inspect the in-

    side of the cavity where it

    mates with the waveguide

    and be sure that solder has

    c o m p l e t e d t h e r un ';

    around the entire cavity

    surface that is in contact

    with the waveguide.

    When all soldering is

    comple te o n the waveguide

    and cavity body, wash off

    paste or acid will degrade

    the future performance of

    the instrument when i t

    causes corrosion.

    The most importa nt parts

    of the assembly are at-

    tempted next, These parts

    are shown in the rnechani..

    cal drawings as parts A

    tlirough and should be

    asse~r>bIedn this order. Be-

    fo re assembly, i t yo .)desire

    to silver-plate ail parts of

    the cavity and waveguide,

    some improvement of the

    cavity wi ll result and the

    ~er fo rmanceof the wave-

    meter wi ll be bette r. Silver-

    plating can be dispensed

    with, however, since the

    unit wi ll perform very well

    without it.)

    Now the t uning disk and

    lead screw can be assem-

    bled into the bearing and

    coll et assembly. To d o this,

    place a lock nu t of the com-

    pression type (as found on

    most screwdr iver -adjust

    locking pots) on the

    col-

    let-see A in Fig. 3. The

    loc king nut is a device

    which has inside tapered

    threads, 3/8-24, and is abo ut

    114-inch thick. It will com-

    press the collet against the

    lead screw threads causing

    it t o move tightly in the col-

    let bearing. Put the nut on

    so that it is loose.

    Now, from the opposite

    side o f the bearing pla te (A),

    insert the threaded end of

    the brass l/Cinch rod and

    screw it up through the col-

    let until about an inch of it

    is exposed. This item is

    labeled in Fig.

    3.

    Now in-

    stall piece C, the tuning

    disk, on the end of the

    brass-threaded rod. Gently

    hea t this piece and solder

    it

    to the threaded rod, taking

    care not to get solder in the

    fine threads. When the

    piece is cooled, wash of f al l

    solder paste. Now, on the

    back of the tuning plate,

    add the powdered Iron as

    described earlier.

    When the three pieces

    have been assembled, in-

    sert the tuning plate into

    Photo B Wavemeter parts and bo ttom view o f the cavity,

    all soldering paste wit h very

    the cavity and align the top

    showing ho w i t is offset

    from

    the waveguide. hot water. Any traces of the p la te so th at the three

    7 agazine March

    1981

  • 7/25/2019 Home Brew X Band Wavemeter

    4/4

    screw holes in the cavity

    wiiil

    top mate with the

    th;-.e threaded holes fo r the

    three 4-40 flat-head screws,

    which can be found on the

    s : d e of the top plate (A)

    F ess the top plate down

    Flrrnly into the c av ~t y nd

    r~ st al l he screws to hold

    l plate firm

    Now screw in the tuning

    p la te by ad jus t ing the

    ~h re ade d od: see that ~

    moves in and out of the

    cavi ty easily, and adjust the

    collet lock nut so that the

    rod moves firmly but not

    :lard. Now put a dab of

    Super Cluen on the lock

    nut, taking care not to get it

    in the cracks of the collet

    and on the lead screw

    threads. This will lock the

    collet lock nut in place.

    Now the last piece,

    D,

    can be assembled. Simply

    glue the plastic disk to the

    to p of the 1-112-inch diame-

    ter tube. This tube can be

    made from any available

    plastic or metal material. It

    fastens t o the extended end

    of the threaded brass rod

    wi th tw o 1/2-36 nuts. Place

    one nut on the end of the

    rod and engage enough

    threads t o al low about 114

    to protrude through the as-

    sembled dial barrel

    (D).

    Then place another nut

    ith the same threads on

    he r od and tighten unti l i t is

    rotate. Now you have a

    micrometer barrel dial and

    the cavity . Cover the out-

    wall wi th

    on marks can be added.

    To calibrate the wave-

    wi ll need a signal

    e w hich can be tuned

    d the band on

    cluded. The source can be

    your uncalibrated X-band

    transmitter, a signal genera-

    tor, or s imply a klystron and

    power supply. A calibrated

    wavemeter and an attenua-

    tor ending in a crystal de-

    tector mounted in a wave-

    guide adapter make up the

    waveguide components re-

    quired for calibration. A

    meter t o register the crystal

    cur rent wi ll serve as the re-

    action indicator. Fig. 4

    shows how to connect these

    devices for calibration.

    Let's assume you are us-

    ing a klystron and power

    supply to do the job. Be

    sure that the klystron is os-

    cillating in a mode that will

    not stop or squeg during

    the operation. Once i t is os-

    cillating , adjust the attenu-

    ator t o provide a three-

    quarter scale reading on the

    current meter. Set the kly-

    stron so that it is in the

    range of 9.8 CHz t o12 CHz

    by adjusting its frequency

    control. Now adjust the cal-

    ibrated wavemeter (which

    we will call the reference

    wavemeter from now on)

    unt il you see a pronounced

    dip in the output-current in-

    dicating meter. By adjust-

    ing the frequency control

    on the klystron and the ref-

    erence wavemeter, you

    should be able to set

    UD

    the

    starting calibration fre-

    quency of 9.8 CHz. When

    you have this frequency

    and vou are sure of it. de-

    tune the reference wave-

    meter. Now tune the device

    you have just constructed

    unt il y?u get a similar reac-

    tion on the output-current

    meter. The dial barrel will

    be quit e near the top of the

    wavemeter cavity. Now

    simply make a pencil or pen

    mark on the calibration dial

    paper, using the moving

    barrel edge as a guide for

    your pencil.

    Computer

    or

    Electronics

    USINESS

    INVESTMENT

    W LYH

    s

    looking to

    buy

    a small electronics business

    based

    in

    the southeastern Uni ted States.

    The business should be:

    Well.managed

    Profitable

    Have sales between

    $500,000 and $5,000,000

    @Andhave growth potential

    Be sure to contac t me wi th your lead:

    Stew Siebens

    1401 North Western Avenue

    Lake Forest, Illinois 60045

    Te le~hone312) 295-2010

    side. If you decide to make

    the calibration steps at a

    finer set of divisions after

    you have done a coarse cal-

    ibration, do not remove the

    coarse steps; they can be

    your check points t o check

    out how well the device re-

    sets.

    The drawings showing

    the mechanical construc-

    tion of the wavemeter are

    supplemented by photo-

    graphs so that you can veri-

    fy how the assembly goes

    together. It is not difficu lt,

    and most amateurs who

    have had experience with

    hand tools should be able

    to complete the unit satis-

    factorily. It is hoped that

    the meter will be helpful to

    those amateurs who want

    to know that they are within

    your power-output indica-

    tor, read off your frequen-

    cy, and then detune the

    wavemeter. You may leave

    it in the line since it will

    cause very lit tl e loss to your

    output.

    Materials for this device

    were obtained from Lec-

    tronic Research Labs, Inc.,

    Atlantic and Ferry Ave.,

    Camden, New Jersey.Al l of

    the materials for waveguide

    components are listed in

    their catalog. There are

    numerous other sources of

    material for this band and

    others, bu t this catalog

    should be in the hands of a ll

    microwave enthusiasts.

    All letters with questions

    regarding this and other ar-

    ticles I have written wil l be

    answered if the writer in-

    Continue the calibrat ion the X-band confines. cludes an SASE. Al l tele-

    process, sett ing the fre- To use the wavemeter in phone calls also wi ll be an-

    quency changes in 50-MHz your rig, simply place it in

    swered, but please remem-

    steps, for a start. Put the t he wa ve gu id e ci r cu it , ber what your time zone is

    coarse calibrations on one

    which is the output toward so that I don't get your call

    side of the calibrat ion line the antenna, then tune it in the early hours of the

    and fine steps on the other

    until you get a reaction in m0r ning .B

    73

    agaz ine

    arch, 1981

    7