BEIS(Boost efficiency IS)-High sensitivity ion source for GC-MS/MS-
MS Business Unit, Analytical & Measuring Instruments Division
Market trend of GC-MS/MS In the past, Dioxins in food had been analyzed by GC-HRMS
Recently, GC-MS/MS was also confirmed as official method (EU589/2014)
The market will grow in Europe, Asia and Africa where export food to Europe
Trend of regulation for POPs analysis by GC-MS/MS
China・Dioxins in food and feed sample・Dioxins in environmental sample
Taipei・Dioxins in food and feed sample・Dioxins in environmental sample
EU・Dioxins in food and feed sample・Dioxins in environmental sample
Not published yet2
Background
Confidential
Market trend of Dioxin analysis (in EU) ML is getting lower and lower
3
Background
Confidential
Food and FeedMaximum level
Conc. in sample Conc. in vial
Pigs fat 1,00 pg/g fat 0.06 pg/uL
Vegetable oils 0,75 pg/g fat 0.04 pg/uL
Foods for infants and young children 0,10 pg/g wet weight 0.04 pg/uL
About competitors situation Release new ion source for Dioxin analysis and improving spec. (IDL)
Make the technical document for Dioxin analysis
TSQ9000
Company Model Spec.
SHIMADZU TQ8050NX OFN:IDL < 0.5fg
Thermo TSQ9000 OFN:IDL < 0.4fg
Agilent 7010B OFN:IDL < 0.5fg
Improve the structure of ion source Diameter of electron gun was optimized
Ionization efficiency was much improved
4
Features of BEIS(Boost efficiency IS)
Confidential
e-
e-e- e
-e-e
-
e-
e-e-
e-
e-
e-
e-
e-
e-
e-e-
e-
e-
e-
e-
e-
e-e
-
e-e-e-e-
e-
e-
e-
e-
e-
e-
e-
e-
Improve
Previous NEW!!
1. Boost up sensitivity of your analysis
2. Realize high reliable analysis for Dioxins
Increase the sensitivity for 3 – 4 times
5
Boost up sensitivity of your analysis
Confidential
Analysis result of DXNs Analysis result of OFN
Company Model Spec.
SHIMADZU TQ8050NX OFN:IDL ≦ 0.3fg
Thermo TSQ9000 OFN:IDL < 0.4fg
Agilent 7010B OFN:IDL < 0.5fg
3.900 3.925 3.950
1000
2000
3000
4000
5000
6000
7000
8000
Octafluoronaphthalene (1 fg)
Comparison of spec. (IDL)
1,2,3,7,8-Pentachlorodibenzo-p-dioxin (0.05 pg/uL)
20.0 21.00.0
0.5
1.0
1.5
2.0
2.5(x1,000)
353.90>290.90355.90>292.90
20.0 21.00.0
0.5
1.0
1.5
2.0
2.5(x1,000)
353.90>290.90355.90>292.90
BEIS Prev.
4 times
15.0 16.00.0
1.0
2.0
3.0
4.0(x1,000)
305.90>242.90303.90>240.90
15.0 16.00.0
1.0
2.0
3.0
4.0(x1,000)
305.90>242.90303.90>240.90
2,3,7,8-Tetrachlorodibenzofuran (0.05 pg/uL)
BEIS Prev.
Confidential
6
Realize high sensitivity analysis even if use low emission current
Realize high reliable analysis for Dioxins
36.5 37.0 37.5 38.00.00
0.25
0.50
0.75
1.00
1.25
1.50(x10,000)
443.80>380.80441.80>378.80
28.0 28.5 29.0 29.50.0
1.0
2.0
3.0
4.0
5.0
6.0
(x1,000)
391.80>328.90389.80>326.90
27.0 27.5 28.0 28.50.0
1.0
2.0
3.0
4.0
5.0
6.0
(x1,000)
391.80>328.90389.80>326.90
20.5 21.0 21.5 22.00.00
0.25
0.50
0.75
1.00
(x10,000)
337.90>274.90339.90>276.90
36.0 36.5 37.0 37.50.00
0.25
0.50
0.75
1.00
1.25
1.50(x10,000)
443.80>380.80441.80>378.80
15.5 16.0 16.5 17.00.00
0.25
0.50
0.75
1.00
1.25
(x10,000)
305.90>242.90303.90>240.90
16.5 17.0 17.5 18.00.00
0.25
0.50
0.75
1.00
1.25
(x10,000)
305.90>242.90303.90>240.90
2,3,7,8-Tetrachlorodibenzofuran
BEIS with 150uA Prev. with 300uA
2 times !
1,2,3,7,8,9-Hexachlorodibenzo-p-dioxin
BEIS Prev.
1,2,3,7,8-Pentachlorodibenzofuran Octachlorodibenzofuran
BEIS Prev.BEIS Prev.
19.5 20.0 20.5 21.00.00
0.25
0.50
0.75
1.00
(x10,000)
337.90>274.90339.90>276.90
Keep the sensitivity over than 500 injection in Dioxin analysis
7
Realize high reliable analysis for Dioxins
Confidential
0.00
0.20
0.40
0.60
0.80
1.00
1.20
1.40
1.60
1.80
2.00
0 50 100 150 200 250 300 350 400 450 500
2,3,7,8-TeCDD 1,2,3,7,8-PeCDD 1,2,3,4,7,8-HxCDD 1,2,3,6,7,8-HxCDD 1,2,3,7,8,9-HxCDD 1,2,3,4,6,7,8-HpCDD OCDD
Transition of the sensitivity of STD (0.05 pg/uL)
STDTest
sample
Ratio (
-)Are
a o
f each I
.N.
/ Are
a o
f First
Inje
ction
Transition of Peak Area (PCDD)
Sales Strategy
Proposes high reliable analysis for Dioxins Promotes to customers who want to start or replace GC-HRMS for Dioxins analysis
Sales targets
BEISDioxin method package
Counterworks against competitor’s proposals
・Usability such as AART function
・Cover many compounds such as
PCBs, BFRs
Confidential
・High sensitivity at low conc.
・High durability (enables
analyze hundreds of sample)
Europe:18 unit Taiwan:4 unit China:***Depending on the regualtion
8
Lifetime of Filament
9
Confidential
Note
e-e-e-e-
e- e-e-e-e- e
-
e-
e-e-
e-
e-
0.00
0.20
0.40
0.60
0.80
1.00
1.20
1.40
0 100 200 300 400 500 600
2,3,7,8-TeCDD 1,2,3,7,8-PeCDD OCDD
Injection number 537
Ratio (
-)Are
a o
f each I
.N.
/ Are
a o
f First
Inje
ction
Transition of Peak Area (PCDD)
Filament make much
more electron than
previous when use the
same emission current
Lifetime of filament may be shorter
・BEIS is suitable for POPs analysis especially Dioxins.
・For other applications (such as pesticides in food), the sensitivity is enough even if use
current ion source. We should recommend current ion source because of filament lifetime.