Hercules™ Suspension System Installation & …„¢ Suspension System Installation &...

28
Hercules™ Suspension System Installation & Specifications AADCO Medical Inc “Ergonomic Solutions for X-ray Safety” AADCO Medical, Inc. PO Box 410 Catamount Commercial Park, Route 66 Randolph, VT 05060 Tel: 802-728-3400 Fax: 802-728-3107 Email: [email protected] Web: www.aadcomed.com AADCO Medical, Asia Pacific, Ltd. Unit B 6/F Kai Yip Factory 15-17 Sam Chuk St. Diamond Hill Kowloon, Hong Kong Tel: 852-2543-4286 Fax: 852-2543-4221 Email: [email protected]

Transcript of Hercules™ Suspension System Installation & …„¢ Suspension System Installation &...

Page 1: Hercules™ Suspension System Installation & …„¢ Suspension System Installation & Specifications AADCO Medical Inc “Ergonomic Solutions for X-ray Safety” AADCO Medical, Inc.

Hercules™ Suspension SystemInstallation & Specifications

AADCO Medical Inc“Ergonomic Solutions for X-ray Safety”

AADCO Medical, Inc.PO Box 410

Catamount Commercial Park, Route 66Randolph, VT 05060

Tel: 802-728-3400Fax: 802-728-3107

Email: [email protected]: www.aadcomed.com

AADCO Medical, Asia Pacific, Ltd.Unit B 6/F Kai Yip Factory

15-17 Sam Chuk St. Diamond HillKowloon, Hong Kong

Tel: 852-2543-4286Fax: 852-2543-4221

Email: [email protected]

Page 2: Hercules™ Suspension System Installation & …„¢ Suspension System Installation & Specifications AADCO Medical Inc “Ergonomic Solutions for X-ray Safety” AADCO Medical, Inc.

AADCO Medical Inc.“Ergonomic Solutions for X-ray Safety”

ASSEMBLY AND OPERATING INSTRUCTIONS

Proper assembly involves several factors. Following them will produce a safe work environ-ment.

Please follow the assembly directions carefully. Any modification to the structure of the sus-pension system or its parts may void the listings of this product. Poorly performed assem-bly or negligence may void the Warranty.

WARNING

These procedures concerning mounting the suspension system to a solid or concrete ceiling assume that professional architectural or engineering services have been consulted. The installer should ensure that the solid ceiling or concrete is structurally sufficient for the mounting of this type of device. The procedures outlined below are the manufacturers recommendations but are in no way a substitute for the advice of a qualified professional architect or engineer. These mounting instruc-tions should be kept together with the respective lamp operating instructions for future reference.

Page 3: Hercules™ Suspension System Installation & …„¢ Suspension System Installation & Specifications AADCO Medical Inc “Ergonomic Solutions for X-ray Safety” AADCO Medical, Inc.

AADCO Medical Inc.“Ergonomic Solutions for X-ray Safety”

1

Table of Contents

1.......................Configuration Overview 1-2 and 3-Arm Hercules™

2....................... Preparatory Ceiling Work2.1....................Hercules Suspension Column to Concrete Ceiling2.1.1.................Attaching Ceiling Anchorage Plate to Solid Ceiling2.1.2.................Hercules Column Installation - to a Unistrut Framework2.2.................... Installation - Ceiling Track2.2.1.................Track Installation - Across Unistrut Channels2.2.2.................Track Installation - Parallel to Unistrut Channels

3....................... Installation of the Hercules Suspension System including Extension Arms3.1.................... Installation of Column Base and Suspension Tube to Ceiling

4....................... Installation of the Spring Arms4.1.................... Installation of Heavy Lifter Spring Arms (OS-11) 4.2.................... Installation of Medium Weight Spring and Extension Arms (OS-10)4.2.1.................Adjusting the Spring Tension of the Medium Weight Arm (OS-10)4.2.2.................Removing Equipment from the Arm4.3.................... Installation of Heavy Lifter Spring Arms with wire chase (OS-12)

5.......................Mounting Equipment5.1....................Mounting a Lamp Head5.2....................Mounting Equipment Onto the OS-11 & OS-12 Heavy Lifter Arms5.2.1.................Adjusting the Spring Tension of the Heavy Lifter Arm5.3....................Mounting a Protective Window5.3.1.................Center Mount, Series 596, 745, and 765 Protective Windows5.4.................... Frame Mount, series 545, 645, 445, 568, 434, 534, and 665 Protective Windows

5.6....................Adjustment of the Heavy Lifter Extension Arm

6.......................Track Mounted Lamp Power Source Cable Drape Track

7....................... Electrical Connections7.1.................... Electrical Connections - Ceiling Mounted Lamps7.2....................Column Wiring7.3.................... Electrical Connections - Track Systems with Lamp7.4.................... Lamp Wiring Diagram7.4.1.................Wiring Diagram for ErgoLite® Halogen Procedure Lamp OSP-5117.4.2.................Wiring Diagram for ErgoLite® Halogen Procedure Lamp OSP-511X7.4.3.................Wiring Diagram for ErgoLite® LED Procedure Lamp OSP-511LX7.4.4.................Wiring Diagram for Three-Head LED Surgical Lamp OSPL-37.4.5.................Wiring Diagram for Five-Head LED Surgical Lamp OSPL-5

8....................... Inspection and Maintenance

9....................... Specifications

Hercules Manual Rev A.1 Jul 2012

Page 4: Hercules™ Suspension System Installation & …„¢ Suspension System Installation & Specifications AADCO Medical Inc “Ergonomic Solutions for X-ray Safety” AADCO Medical, Inc.

AADCO Medical Inc.“Ergonomic Solutions for X-ray Safety”

2

Copyright

Copyright© 2011 by AADCO Medical, Inc. All rights reserved. No part of this manual may be reproduced in any form without prior written permission of AADCO Medical.

Disclaimer

Improper installation or use of the overhead systems described in this manual, could result in personal injury and prop-erty damage. AADCO Medical, Inc., disclaims liability for any modifications or attachments to other equipment which are

not in conformity with the specifications and information contained within this manual. Because of various site condi-tions, user preferences and other unknown factors, these installation instructions are for guidance only and should not be

considered appropriate in every case.

Important

Important information The electrical or mechanical instructions, and other information referred to in this manual, should be carried out by suitably qualified engineers who have sole responsibility for this work. Carefully read the information contained in this manual before beginning to assemble and install the equipment. Failure to read and understand the

warnings and instructions could lead to serious injury and property damage.

WARNING

Collision of the Radiation Shields or Lamps with other equipment (i.e. C-Arm) will damage the shield & lamp and may cause them to fail.

Applicability

This manual applies to AADCO Medical, Inc. Rayshield® Hercules Suspension System and

Ergolite Lamps. AADCO Medical reserves the right to change specifications and contents of this manual without obliga-tion.

If any problems are encountered while installing any AADCO Medical product, contact the Tech Support Department at:

AADCO Medical, Inc. PO Box 410 Catamount Commercial Park, Route 66 Randolph, Vermont 05060

Phone: 802-728-3400 Fax: 802-728-3107 Email: [email protected] World Wide Web: http://www.aadcomed.com

Trademarks RayShield® is the registered trademark of AADCO Medical, Inc ErgoLite® is the registered trademark of AADCO Medical, Inc.

MRI-Safe® is the registered trademark of AADCO Medical, Inc. Hercules™ is the trademark of AADCO Medical, Inc.

Unistrut® is the registered trademark of Unistrut Corporation. Loctite® is the registered trademark of Loctite Corporation.

X-RAY EQUIPMENTWITH RESPECT TO ELECTRIC SHOCK, FIRE MECHANICAL

AND OTHER SPECIFIED HAZARDS ONLY IN ACCORDANCE WITH UL 60601-1,

CSA/CN C22.2 NO. 601.1-M-90, IEC 60601-2-32

55-BB

!

X-RAY EQUIPMENT

Questions

or

Comments

Page 5: Hercules™ Suspension System Installation & …„¢ Suspension System Installation & Specifications AADCO Medical Inc “Ergonomic Solutions for X-ray Safety” AADCO Medical, Inc.

AADCO Medical Inc.“Ergonomic Solutions for X-ray Safety”

3

1. Configuration Overview: 1, 2, and 3-Arm Hercules

Hercules with OS-11 Heavy Lifter

Hercules with OS-11 Heavy Lifter and OS-12 Heavy lifter with wire chase

Hercules with OS-10 Medium Weight

Hercules with OS-12 Heavy Lifter

with wire chase

Page 6: Hercules™ Suspension System Installation & …„¢ Suspension System Installation & Specifications AADCO Medical Inc “Ergonomic Solutions for X-ray Safety” AADCO Medical, Inc.

AADCO Medical Inc.“Ergonomic Solutions for X-ray Safety”

4

Hercules with OS-11 Heavy Lifter, OS-12 Heavy Lifter with wire chaseand OS-10 Medium Weight

Hercules with OS-11 Heavy Lifter and OS-10 Medium Weight

Hercules with OS-12 Heavy Lifter with wire chase and OS-10 Medium Weight

Page 7: Hercules™ Suspension System Installation & …„¢ Suspension System Installation & Specifications AADCO Medical Inc “Ergonomic Solutions for X-ray Safety” AADCO Medical, Inc.

AADCO Medical Inc.“Ergonomic Solutions for X-ray Safety”

5

2. Preparatory Ceiling Work

WARNINGThese procedures concerning mounting the Hercules to a solid or concrete ceiling assume that professional architectural or engineering services have been consulted. The installer should ensure that the solid ceiling or concrete is structurally sufficient for the mounting of this type of device. The procedures outlined below are the manufacturers recommendations but are in no way a substitute for the advice of a qualified professional architect or engineer. These mounting instructions should be kept together with the respective shield operat-ing instructions for future reference.

The equipment’s weight and the tilting as a result of the articulated arm(s) will require thorough work, in par-ticular the use of an authorized hammer drill and compliance with all drill tolerances.

Anchorage plate or intermediate flanges should only be attached to a ceiling of concrete strength classifica-tion B 25 or higher. The dowel anchor must be inserted completely into the concrete in case Heraklith or other lightweight construction lining is present. Correspondingly longer threaded bolts will be required for attaching the ceiling anchorage plate to accommodate the lining.

Furthermore, it is imperative that neither the base mount nor the attachment parts come into contact with the solid ceiling’s reinforcement parts.

2.1 Hercules Suspension Column to Concrete CeilingsThe Hercules Suspension system can be mounted fixed to a solid ceiling or Unistrut framework, or it can be mounted mobile to a ceiling mounted track. The Hercules Suspension Systems comes with a 140mm square base mount.

This square base mount has holes to accept (4) M12 bolts evenly spaced at 150 mm diameter which supports the vertical column tube.

When attaching this Hercules Suspension System to a solid ceiling or Unistrut framework, it is absolutely mandatory to use an anchorage plate (OSP-736NF sold separately) to sup-port and adjust the heavy system. This anchorage plate must be firmly anchored flat and level.

It is imperative that neither the Base Mount nor the attachment parts come in contact with the ceilings reinforcement parts.

The anchorage plate can easily be attached to ceilings of appropriate concrete strength, using dowel anchors M10 or safety dowels M8.

However, the forces which are created in addition to the equipments weight caused by the tilting force of the significantly protruding articulated arms will require thorough drill work, especially the use of an authorized hammer drill and compliance with drill toler-ances.

The base mount and vertical column need to be vertically adjusted so the arms won’t drift.

2.1.1 Attaching Ceiling Anchorage Plate to Solid CeilingDrill (drill bit diameter 12) at least 100 mm deep using an authorized hammer drill. Use ceiling anchorage plate as template (an assistant might be required). Push safety dowels through drill holes of ceiling anchorage plate so as to create a close fit with the washers.

Thoroughly tighten screws using a torque wrench 18.4 ft/pds (25 Nm).

Lightweight construction linings with a maximum thickness of 30 mm can be spanned using a safety dowel. Longer dowels will be required for a thickness exceeding 30 mm.

Use an intermediate flange to span the spacing between the solid ceiling and the intermediate ceiling (preferably for spacings larger than 400 mm).

The intermediate flange should be flush with the intermediate ceiling’s lower surface.

Page 8: Hercules™ Suspension System Installation & …„¢ Suspension System Installation & Specifications AADCO Medical Inc “Ergonomic Solutions for X-ray Safety” AADCO Medical, Inc.

AADCO Medical Inc.“Ergonomic Solutions for X-ray Safety”

6

These procedures concerning a Unistrut framework assume that professional architectural or engineering services have been consulted. The installer should ensure that the framework is securely fixed to the main structure of the building and should consult the Unistrut manufacturers’ installation instructions for proper mounting information.

Anchorage Plate:A steel plate 350mm x 350mm x 6mm thick with a 13mm diameter hole in each corner, 260mm apart, & 1 central 60mm diameter hole, plus 4 x 12mm diameter holes to suit the ceiling column base fixing holes. Four M12 x 40mm long cap screws, washers & locknuts are included which are used for attaching the ceiling column to the plate.

WARNING: The Hercules Column Base flange must be bolted to a flat and level surface.

Step 1 Determine that existing or newly installed unistrut channels, concete ceiling and/or other supporting elements can support the applied load and that professional architectural or engineering services have been consulted to make this determination.

Step 2Align the holes in the upper flange of the ceiling column with the optional anchorage plate and use the supplied M12 x 40mm long cap screws to bolt together. Apply Loctite to the threads.

Step 3Bolt the anchorage plate, or a suitable alternative, to a Unistrut framework so that the distance between the plate and any suspended ceiling is no greater that 35cm.

Step 4Use a carpenter’s level to ensure that the column is vertical.

Cable for lamp

Unistrut channel

Ceiling column

Anchorage plate

Suspended ceiling

Anchorage plateCable for lamp

Unistrut framework

Suspended ceiling

Max 35cm

Concrete Ceiling

Anchorage Plate below ceiling

Anchorage Plate

above ceiling

Cable for lamp

Note

Optional item

2.1.2 Hercules Column Installation - to a Unistrut Framework

Page 9: Hercules™ Suspension System Installation & …„¢ Suspension System Installation & Specifications AADCO Medical Inc “Ergonomic Solutions for X-ray Safety” AADCO Medical, Inc.

AADCO Medical Inc.“Ergonomic Solutions for X-ray Safety”

7

2.2. Installation - Ceiling Track

2.2.1 Track Installation - Across Unistrut Channels

NOTE: With Unistrut channels across the track the following applies:

■ The track should be suspended from a minimum of 4 Unistrut channels.

■ The track must not extend beyond the last bolt point on each end.

■ The track may be mounted across the channels at a maximum angle of 30°.

Step 1 Bolt the track to the channels using 4 bolts with lock washer per side. If Power Source Cable Drape Track (OSP-747NF) is also being used, refer to section 6 before proceeding.

A hardware set comprises one each of the following: 3⁄8-16 hex Flathead screw, lock washer and Unistrut nut.

3⁄8-16 setscrew/washer/nutCeiling Track

Unistrut channel across track

Carriage

Hardware attachment points

S-270 Track

S-260 Track

Ceiling Track

Unistrut channel across track

Carriage3⁄8-16 setscrew/washer/nut

The Hercules Suspension System can be mounted on a ceiling track, OS-260NF Regular Ceiling Track, for 1 or 2 arm configurations where the combined equipment load is less than 114.6 lbs (52 Kg) or OS-270NF Heavy load Track for 2 or 3 arm configurations where the combined equipment load is more than 114.6 lbs. (52 Kg) up to 213.8 lbs (97 Kg).

When mounting the Hercules Suspension system on the OS-270NF Heavy Load Track a 4 to 6 bolt adaptor plate is re-quired.

WARNING: These procedures concerning a Unistrut framework assume that professional architectural or engineering services have been consulted. Therefore, these instructions are only guidelines for installation and are in no meant to supplant or substitute for professional architectural or engineering services. In all cases, the installer should ensure that the framework is securely fixed to the main structure of the building and should consult the Unistrut manufacturers’ installation instructions for proper mounting information. These installation procedures assume the track is mounted to Unistrut or similar channels. With suspended ceilings or in rooms with high ceil-ings, a framework that is fixed to the structure of the building should support these channels. Suitably qualified engineers should carry out the installation of the track.

Page 10: Hercules™ Suspension System Installation & …„¢ Suspension System Installation & Specifications AADCO Medical Inc “Ergonomic Solutions for X-ray Safety” AADCO Medical, Inc.

AADCO Medical Inc.“Ergonomic Solutions for X-ray Safety”

8

2.2.2. Track Installation - Parallel to Unistrut ChannelsNOTE: With Unistrut channels parallel to the track the following applies:

■ The track should be suspended from 2 parallel 2.5 meter long (or longer) Unistrut channels set either 7" or 13" apart.

■ The track must be attached to the two Unistrut channels by 8 hardware sets (4 per side).

■ The track offers the optional use of hardware attachment points that are either 7 inches on center or 13 inches on center.

Step 1 Bolt the track to the two channels using 4 hardware sets per side. If Power Source Cable Drape Track (PS-747) is also be-ing used refer to section 7 before proceeding.

3⁄8-16 setscrew/washer/nutCeiling Track

Parallel Unistrut channels

Hardware attachment points

S-260 Track

S-270 Track

Ceiling Track

Parallel Unistrut channels

3⁄8-16 setscrew/washer/nut

Page 11: Hercules™ Suspension System Installation & …„¢ Suspension System Installation & Specifications AADCO Medical Inc “Ergonomic Solutions for X-ray Safety” AADCO Medical, Inc.

AADCO Medical Inc.“Ergonomic Solutions for X-ray Safety”

9

3. Installation of the Hercules Column including Extension Arms

NOTE: Due to the weight of this unit, it is preferable to have 2 people perform this installation.

If the column has more than 1 extension arm, it is suggested that the arms be taped or strapped together to limit move-ment. Lift the column into place, while paying attention to the power cord to keep it from being pinched by using the channel provided in the base mount. The cord and channel should be aligned with the strain relief connector on the trol-ley. Install the supplied M12 x 40mm hex bolts with flat washers and lock washers. Apply Loctite to the threads.

Warning:Do not connect power supply until all equipment is installed and adjusted.

Transit Locking Knob Warning: Do Not release (pull) the transit locking knob on the bottom of the spring arm until the

equipment is installed and secured. The arm is loaded with a high-tension spring and will spring up with great force.

Page 12: Hercules™ Suspension System Installation & …„¢ Suspension System Installation & Specifications AADCO Medical Inc “Ergonomic Solutions for X-ray Safety” AADCO Medical, Inc.

AADCO Medical Inc.“Ergonomic Solutions for X-ray Safety”

10

3.1 Installation of Column Base and Suspension Tube to Ceiling

Before installing the column base and suspension tube, all preparatory ceiling work must be complete and anchorage plate must be securely mounted flat and level.

Two persons are needed to lift and secure the column as-sembly.

Lift the column into place and secure it using the supplied M12 button head screws (A) and M12 flat washer (B) up thru the column and plate. Secure it using M12 flat washer (B), M12 lock washer (C), and nut (D) as shown in the dia-gram on the left. Start all 4 before tightening with a torque wrench to 18.4 ft/lbs (25 Nm).

The column and suspension tube must be vertical (plumb)

The equipment’s arms will not remain in place and may turn away and will require excessive braking should the suspen-sion tube be adjusted not completely vertical.

This vertical attacment is extremenly important.

B

AB

CD

Page 13: Hercules™ Suspension System Installation & …„¢ Suspension System Installation & Specifications AADCO Medical Inc “Ergonomic Solutions for X-ray Safety” AADCO Medical, Inc.

AADCO Medical Inc.“Ergonomic Solutions for X-ray Safety”

11

4.1 Installation of the Heavy Lifter Spring Arms

Warning:The spring arms are factory-preadjusted for the accompanying equipment. Note that the spring arm should be installed and secured to the extension arm before any lamp, shield,

or other equipment is attached.

Make sure the power is cut off.

1. Detach the cover (D) at the end of the extension arm.

2. Remove circlip (C) and washer (B) from spring arm. The spacer (E), and the large equalizing disc (A) will stay on the spring arm axis.

3. Push spring arm from below into extension arm, push washer (B) and securing ring (circlip) (C) in the correct sequence onto the connecting axis. Please note that the securing ring must completely snap into the groove of the coupling trunnion.

4. Replace cover on extension arm.

The disassembly of the spring arm is performed in reverse order.

D

C

B

Friction Adjustment Screw

A

E

Spring Arm Axis

Transit Locking Knob

Warning:Never allow unloaded spring arm to bounce up (i.e. after equipment has been removed).

4. Insatallation of the Spring Arms

Page 14: Hercules™ Suspension System Installation & …„¢ Suspension System Installation & Specifications AADCO Medical Inc “Ergonomic Solutions for X-ray Safety” AADCO Medical, Inc.

AADCO Medical Inc.“Ergonomic Solutions for X-ray Safety”

12

4.2 Installation of the Medium Weight Spring and Extension Arms

Warning: Never allow unloaded spring arm to bounce up (i.e. after lamp or equipment has been removed).

A

B

C

D

E

The spring arms are factory pre-adjusted for the accompanying lamp or equipment. Note that the spring and extension arm should be installed and secured to the column before any lamp, shield or other equipment is attached.

*** Make sure the power is cut off.***

1. Remove the screw (E) at the cover at the end of the extension arm and remove the cover (D) including its attached rotary electrical connector.

2. Remove circlip (C) and washer (B) from the bottom of the column. The large equalizing disc (A) should stay on the column axis.

3. Using 2 people, lift the spring/extension arm from below and push it into place, While holding it in place, install the washer (B) and circlip (C) in the correct sequence. Please note that the circlip (C) must completely snap into the groove of the axis. Once the circlip and arm are secured, replace the cover (D) with rotary connector up into the extension arm and secure the cover with the screw (E).

The disassembly of the arm is done in the reverse order.

Page 15: Hercules™ Suspension System Installation & …„¢ Suspension System Installation & Specifications AADCO Medical Inc “Ergonomic Solutions for X-ray Safety” AADCO Medical, Inc.

AADCO Medical Inc.“Ergonomic Solutions for X-ray Safety”

13

4.2.1 Adjusting the Spring Tension of the Medium Weight ArmAfter completing the installation work or when modifying the load, the spring force may need to be adjusted. If the spring force is too low, in other words the spring arm sinks to the lower stop point, the spring force will need to be in-creased.

1. Locate the spring adjustment slot in the side of the springarm. Raise or lower the equipment to reveal the adjustment wheel with holes (see photo).

2. Insert a 5mm adjustment rod into one of the holes and rotate the adjustment wheel upward (CW).

3. Remove adjustment rod and check spring tension by raising and lowering the equipment.

4. If the spring arm still sinks, repeat spring adjustment until desired support is achieved.

If the spring force is too high, in other words the spring arm rises to the upper stop point, the spring force will need to be de-creased.

1. Locate the spring adjustment slot in the side of the spring-arm. Raise or lower the equipment to reveal the adjustment wheel with holes (see photo).

2. Insert a 5mm adjustment rod into one of the holes and rotate the adjustment wheel downward (CCW).

3. Remove adjustment rod and check spring tension by raising and lowering the equipment.

4. If the spring arm still rises, repeat spring adjustment until desired support is achieved.

4.2.2 Removing Equipment from the Arm

1. CAUTION: The lower arm (6) must not rise higher than Position B (horizontal) shown on the diagram check carefully. Before removing any equipment, secure (push) the transit locking knob up (in), then lift the arm until it stops against the transit locking knob.

2. Remove screw (2).3. Lift the safety sleeve (1).4. a) CAUTION: Ensure that the arm (6) is in Position B when the

equipment is removed. b) Hold on firmly to the item of equipment. c) Remove safety segment key (4) and separate the equipment

from the arm (6). If necessary, turn the equipment slightly to make this easier.

5. Replace the safety segment key (4) and re-fit the safety cover (1) with screw (2).

NOTE: We recommend that removal of equipment from the arm be carried out by two people.

Adjustment Wheel with Holes

5mm Adjustment Rod

Page 16: Hercules™ Suspension System Installation & …„¢ Suspension System Installation & Specifications AADCO Medical Inc “Ergonomic Solutions for X-ray Safety” AADCO Medical, Inc.

AADCO Medical Inc.“Ergonomic Solutions for X-ray Safety”

14

4.3 Installation of the Heavy Lifter Spring Arms with wire chaseWarning:

The spring arms are factory-preadjusted for the accompanying equipment. Note that the spring arm should be installed and secured to the extension arm

before any lamp, shield, or other equipment is attached.

Make sure the power is cut off.

1. Using two persons, lift spring arm to insert the Link (F) into clevis (A)

2. Slide the Link (F) into Clevis (A) and then insert Pin (B) through Clevis (A) and Link (F). Be sure locate the notch in the Pin to align with the hole in the Link (F).

3. Carefully install shoulder bolt (C) into Link (F) and secure Pin (B)

4. Push washer (D) and securing ring (circlip) (E) in the correct sequence onto the connecting Pin (B). Please note that the securing ring must completely snap into the groove of the pin.

5. .

The disassembly of the spring arm is performed in reverse order.

AB

C

DE

F

Warning:Never allow unloaded spring arm to bounce up (i.e. after equipment has been removed).

Page 17: Hercules™ Suspension System Installation & …„¢ Suspension System Installation & Specifications AADCO Medical Inc “Ergonomic Solutions for X-ray Safety” AADCO Medical, Inc.

AADCO Medical Inc.“Ergonomic Solutions for X-ray Safety”

15

The OS-10 mediumweight Electric wired arm can support lamps weighing up to 54.4 lbs (27Kg).

1. Remove and retain the s countersunk screw (2) from the collar (1) located at the end of the spring arm.

2. Lift the collar (1) and remove the Safety segment Key (4) from the slot in the spring arm. Apply a light coat of lithium or automotive bearing grease to the groove in the arm-coupling journal (12) at the top of the lamp head.

3. Insert this coupling journal (12) up into the bore of the spring arm, aligning the groove in the journal with the slot in the arm. Some maneuvering may be necessary to properly align the internal electric socket.

4. Insert the safety segment key (4) into the slot so that it engages with the groove.

5. Slide to collar (1) down and align the screw holes on collar and connector (11).

6. Apply Loctite to the screw threads and replace the silver screw (2).

Once this installation is complete, the factory preset spring tension should now support the equipment.

The Transit Locking knob may be released (pulled) to allow the arm to be moved up past the horizontal position. The factory set spring tension should support the equipment and hold it in any position.

5.1 Mounting A Lamp Head

Caution The Non Ferrous ErgoLite® Lamp (Catalog # OSP-511NF) must be wired through an external transformer.

Safety

Segment Key

4

12

1

2

3

11

4

Warning: Do Not release (pull) the transit locking knob on the bottom of the spring arm until the equipment is installed

and secured. The arm is loaded with a high-tension spring and will spring up with great force.

5. Mounting Equipment

Mounting equipment onto the OS 10 Lightweight Electric wired arm

Page 18: Hercules™ Suspension System Installation & …„¢ Suspension System Installation & Specifications AADCO Medical Inc “Ergonomic Solutions for X-ray Safety” AADCO Medical, Inc.

AADCO Medical Inc.“Ergonomic Solutions for X-ray Safety”

16

5.2 Mounting Equipment Onto the OS-11 & OS-12 Heavy Lifter Arms

The OS-11 & OS-12 Heavy lifter arm can support a many types of equipment including Radiation shields, injector mounts, lamps or monitors up to 99 lbs (45Kg). These mounts (sold separately) may require 2 persons to lift and secure onto the arm.

The mounting screws to attach this equipment (3 M6x 15mm flathead hexsocket screws) will be supplied with the equipment mounts. Remove and save these 3 screws from the equipment post. Lift the equipment and insert the equip-ment post into the coupling. Rotate the equipment post to align with the screw hole in the coupling sleeve and install the saved screws. All 3 screws should now be installed and tightened. Any equipment not sold by AADCO Medical Inc, (monitors, injectors, ect.) should now be installed per manufactures specifications.Once this installation is complete, the factory preset spring tension should now support the equipment.The Transit Locking knob may be released (pulled) to allow the arm to be moved up past the horizontal position. The fac-tory set spring tension should support the equipment and hold it in any position.

Warning:Do Not release (pull) the transit locking knob on the bottom of the spring arm until the equipment is installed

and secured. The arm is loaded with a high-tension spring and will spring up with great force.

Page 19: Hercules™ Suspension System Installation & …„¢ Suspension System Installation & Specifications AADCO Medical Inc “Ergonomic Solutions for X-ray Safety” AADCO Medical, Inc.

AADCO Medical Inc.“Ergonomic Solutions for X-ray Safety”

17

5.2.1 Adjusting the Spring Tension of the Heavy Lifter Arm

After completing the installation work or when modifying the load, the spring force may need to be adjusted.

If the spring force is too low, in other words the spring arm sinks to the lower stop point, the spring force will need to be increased.

1. Locate the spring adjustment slot in the side of the springarm. Raise or lower the equipment to reveal the adjustment wheel with holes (see photo).

2. Insert a 4mm Allen wrench or 5mm rod into one of the holes and rotate the adjustment wheel downward.

3. Remove 4mm Allen wrench or 5mm rod and check spring tension by raising and lowering the equipment.

4. If the spring arm still sinks, repeat spring adjustment until desired support is achieved.

If the spring force is too high, in other words the spring arm rises to the upper stop point, the spring force will need to be decreased.

1. Locate the spring adjustment slot in the side of the springarm. Raise or lower the equipment to reveal the adjustment wheel with holes (see photo).

2. Insert a 4mm Allen wrench or 5mm rod into one of the holes and rotate the adjustment wheel upward.

3. Remove 4mm Allen wrench or 5mm rod and check spring tension by raising and lowering the equipment.

4. If the spring arm still rises, repeat spring adjustment until desired support is achieved.

Spring Adjustment Slot

Page 20: Hercules™ Suspension System Installation & …„¢ Suspension System Installation & Specifications AADCO Medical Inc “Ergonomic Solutions for X-ray Safety” AADCO Medical, Inc.

AADCO Medical Inc.“Ergonomic Solutions for X-ray Safety”

18

5.3.1 Center Mount, Series 596, 745 and 765 Protective Windows

Step 1

If the clear lead acrylic window has a corner cut-out, determine if it is normally positioned bottom left or bottom right. The central ball & socket joint allows the window to move in any direction and the central vertical support can turn 360° in the Type 2000 arm, so the cut-out can be positioned at any corner. However, backward tilt is restricted by the position of the central support. Maximum forward tilt at the top of the window is achieved when the support is nearest to the practitioner and the window is on the other side.

Step 2

Place the 45mm od. hub of the ball & socket joint in the central hole in the clear lead acrylic window. This joint is located at the lower end of the vertical support. Attach the window-retaining disc to the joint with 4 - M4 x 20mm long counter-sunk allen head screws. Apply Loctite to the threads.

Step 3

Refer to Section 5.2.1 Adjusting the Spring Tension of the Heavy Lifter Arm

5.3 Mounting a Protective Window

WARNING The height adjustable spring arm is under tension and may spring up with considerable force.

5.4 Frame Mount, Series 545, 645, 445, 568, 434, 534, & 665 Protective WindowsThese are shipped fully assembled and ready to install. Refer to Section 5.3.

Coupling

Center vertical support

Ball & socket joint

Retaining disk

Hub

Page 21: Hercules™ Suspension System Installation & …„¢ Suspension System Installation & Specifications AADCO Medical Inc “Ergonomic Solutions for X-ray Safety” AADCO Medical, Inc.

AADCO Medical Inc.“Ergonomic Solutions for X-ray Safety”

19

Friction Adjustment Screw

5.6 Adjustment of the Heavy Lifter Extension Arm

Adjusting the Brakes

The Heavy Lifter extension arm has two hex socket screws at the seat for the central axis and two hex socket screws at the seat for the spring arm with which a brake that is located between the needle bearings of the extension arm can be adjusted.

It is recommended to completely open the brakes to check the vertical adjustment of the suspension tube to avoid any angular deviations that might exist.

The Friction adjustment Screws must be released during the installation of the spring arms.

If the suspension tube is adjusted in a manner so that the extension arm and spring arm are stationary in any position, the brakes can be lightly tightened in order to avoid inadvertent movement of the arms during positioning.

Friction Adjustment Screw

Page 22: Hercules™ Suspension System Installation & …„¢ Suspension System Installation & Specifications AADCO Medical Inc “Ergonomic Solutions for X-ray Safety” AADCO Medical, Inc.

AADCO Medical Inc.“Ergonomic Solutions for X-ray Safety”

20

The power source cable drape system is track mounted in a 5 foot cable drape track that is factory attached to the equipment track.

Step 1

In the process of mounting the equipment track, select a side for the cable drape track.

Step 2

Snugly tighten the equipment track bolts.

Step 3

Install tie-wraps through the slot on each of the three drape trolleys (Photo 2).

Step 4

Install power source cable, using tie wraps (Photo 3), draped with adequate spacing to allow free movement along the entire equipment track, without cable tension or strain of any type. When positioned, tighten the tie-wrap snugly and cut off excess material.

Cable track (optional)

Cable cord gripSupply cable

6. Track Mounted Lamp Power Source Cable Drape Track

Drape Trolley

Drape Track

Photo 2Photo 3

Page 23: Hercules™ Suspension System Installation & …„¢ Suspension System Installation & Specifications AADCO Medical Inc “Ergonomic Solutions for X-ray Safety” AADCO Medical, Inc.

AADCO Medical Inc.“Ergonomic Solutions for X-ray Safety”

21

WARNING ELECTRICAL SURGES

Power off the on-site power supply and secure it from being switched on again.

NOTEThe supply network must be secured (fuse settings) for 230V@16A or 120V@15A, 12V@15A or 24V@15A, re-

spectively. A complete (all-pole) disconnection is required for the connection on-site, with clearance between the contacts of 3mm or more.

WARNINGELECTRICAL SURGES

A ceiling anchorage which was not correctly grounded might cause electrical surges in the case of failure. Check the premounted ground connection and connect on-site ground conductors.

7.1 Electrical Connections - Ceiling Mounted LampsThe supply cable for columns attached to a concrete ceiling should be connected via a cut out in the edge of the flange. The supply cable for columns attached to the optional anchorage plate should be connected via a central hole in the plate and via a 20 od. side hole located below the column flange. The cable emanating from the center of the column should also protrude from the 20 od. side hole and be connected to the terminal.

WARNING:Special care must be taken to ensure that the AC line are correctly connected. The lamps must be connected to an electrical supply in accordance with accepted local electrical standards and regulations. Turn off the line power before installing or preforming maintenance. Lamps are supplied for low voltage connection:

a) Low voltage lamps require a separate external transformer. Requirements are stated on the lamphead.

Caution:Before connecting to the DC line Supply, check the lamphead label for voltage requirements.

7 Electrical Connections

7.2 Column Wiring

Page 24: Hercules™ Suspension System Installation & …„¢ Suspension System Installation & Specifications AADCO Medical Inc “Ergonomic Solutions for X-ray Safety” AADCO Medical, Inc.

AADCO Medical Inc.“Ergonomic Solutions for X-ray Safety”

22

7.3 Electrical Connections - Track Systems with Lamp

NOTE

The following steps must be performed by a qualified electrician in accordance to local electrical standards and regulations.

The movable carriage in the ceiling track is equipped with an electrical supply cable cord grip with a strain relief feature. The supply cable furnished with the Hercules suspension system must be threaded thru this cord grip when mounted to the ceiling track assembly.

This will require the dis-assembly of the plug (see diagram). Remove and retain the plug from the supply cord.

Remove and retain the 2 M6x12 socket head countersunk screws from the lower face of the carriage and remove a 43mm x 57mm plate to reveal a cavity. Install the cord grip into 43mm x 57mm plate, and then thread the cord thru the cord grip and plate. (see diagram). Next slide the cord grip and plate along the supply cord until contact with the carriage. Reinstall the 2 M6x12 socket head countersunk screws, and tighten the cord grip to secure the cord.

Next reinstall the medical grade cord plug (see diagram).

Once all equipment is installed the medical grade plug and cord may be plugged into a fused 120V@15 outlet.

7.4. Lamp Wiring Diagram

7.4.1. Wiring Diagram for ErgoLite® Halogen Procedure Lamp OSP-511

Power Requirements:The Lamp has an input voltage requirement of 24VAC. This voltage is to be provided by an external power supply. Please refer to lamp mounting instructions Section 4.1. Mounting a Lamp Head.

Black 120V

Black Yellow

Yellow 25.2V

PS-511T External transformer

Primary Secondary

PS-511T External Transformer is a UL recognized component, File # E87140, Model # 273-1512B

7.4.2. Wiring Diagram for ErgoLite® Halogen Procedure Lamp OSP-511X

Power Requirements:The Lamp has an input voltage requirement of 120VAC. The Lamp has an internal transformer to convert to 24V. Please refer to lamp mounting instructions Section 4.1. Mounting a Lamp Head.

Hot Wire

Neutral Wire

Ground Wire

Fuse 5A250V

Switch 10A125 VAC

Lamp 24V50W

ChassisGround

Power Supply 2A24 VAC

Hot Wire

Neutral Wire

Ground Wire

Fuse 5A250V

Switch 10A125 VAC

ChassisGround

Power Source120 VAC

Lamp 24V50W

Electronic Transformer120 VAC Primary23.5 VAC Secondary60 Watt 2.6 Amps

Page 25: Hercules™ Suspension System Installation & …„¢ Suspension System Installation & Specifications AADCO Medical Inc “Ergonomic Solutions for X-ray Safety” AADCO Medical, Inc.

AADCO Medical Inc.“Ergonomic Solutions for X-ray Safety”

23

7.4.3. Wiring Diagram for ErgoLite® LED Procedure Lamp OSP-511LXPower Requirements:The Lamp has an input voltage requirement of 24VDC 2.5A. This voltage is to be provided by an external power supply. Please refer to lamp mounting instructions Section 4.1. Mounting a Lamp Head.

7.4.4. Wiring Diagram for Three-Head LED Surgical Lamp OSPL-3Power Requirements:The Lamp has an input voltage requirement of 24VDC 10A. This voltage is to be provided by an external power supply. Please refer to lamp mounting instructions Section 4.1. Mounting a Lamp Head.

7.4.5. Wiring Diagram for Five-Head LED Surgical Lamp OSPL-5Power Requirements:The Lamp has an input voltage requirement of 24VDC 12.5A. This voltage is to be provided by an external power supply. Please refer to lamp mounting instructions Section 4.1. Mounting a Lamp Head.

Hot Wire

Neutral Wire

Ground Wire

Fuse 3AG10A 32V

Switch 10A125 VAC

LED Array

ChassisGround

DimmerControl

PowerSupply

24VDC 5A

100-240VAC50-60Hz Input

ChassisGround

Hot Wire

Neutral Wire

Ground Wire

Fuse 3AG15A 32V

Switch

LED Array

ChassisGround

DimmerControl

PowerSupply

24VDC 10A

100-240VAC50-60Hz Input

+

-

G

N

L

Hot Wire

Neutral Wire

Ground Wire

Fuse 3AG15A 32V

Switch

LED Array

ChassisGround

DimmerControl

PowerSupply

24VDC 12.5A

100-240VAC50-60Hz Input

+

-

G

N

L

Page 26: Hercules™ Suspension System Installation & …„¢ Suspension System Installation & Specifications AADCO Medical Inc “Ergonomic Solutions for X-ray Safety” AADCO Medical, Inc.

AADCO Medical Inc.“Ergonomic Solutions for X-ray Safety”

24

8. Inspection & Maintenance

Inspection

After initial installation and periodically thereafter:

■ Inspect all screws/bolts to ensure that they are tightened.

■ Inspect all parts for signs of wear or damage.

■ Ensure that cabling to a lamp is not damaged.

■ Check the total system to ensure that it moves smoothly.

Cleaning

The suspension parts are provided with high quality baked enamel coating. They can be cleaned and disinfected using agents normally used in operating rooms.

The lamp head and clear plastic lens can be kept clean by wiping with a soft viscose sponge. After drying with a non fluffing cloth the lens should be wiped with an anti-static agent. This largely prevents dust deposits as a result of electro-static charging. Do not use scouring agents or use disinfectant containing more than 20% alcohol.

Protective Windows fitted to radiation shields contain lead acrylic. Small marks on the surface of lead acrylic can be re-moved with be fine acrylic polish. Use isopropyl alcohol on a soft cloth to remove marks.

Maintenance

The radiation shields and surgical lamps described in this manual are constructed in such a way that all mechanical elements, bearings and joints only need periodic inspection. However, preventative maintenance procedures are recommended where the articulated joints are greased annually with a light coat of a lithium based automotive grease (or equivalent).

Caution:If any worn or damaged parts are

found do not continue to use them.

WARNING Turn off the power to any fitted lamp before cleaning or performing maintenance. Engage (push) the transit lock button into the spring arm and raise against the stop before removing a protective window or lamp. (See Sections 4 & 5) The suspension system has joints and pivots, which may create pinch points. Keep body parts away when these are in motion.

Page 27: Hercules™ Suspension System Installation & …„¢ Suspension System Installation & Specifications AADCO Medical Inc “Ergonomic Solutions for X-ray Safety” AADCO Medical, Inc.

AADCO Medical Inc.“Ergonomic Solutions for X-ray Safety”

25

Specifications

DimensionsOverall Heavy lifting Arm Reach 68 inchesOverall Heavy lift with large wire chase 75 inchesOverall Light lifting arm reach 62 inchesStandard Column Height 28.38 inchesOther Dimensions, See Diagram on right.

Standard System Weights Single Light lifting arm configuration 85 lbsSingle Heavy lifting arm configuration 100 lbs.Dual Arm Configuration, 1 Light lifting & 1 Heavy lifting arm. 115 lbs.Triple Arm Configuration, 1 Light lifting & 2 Heavy lifting arms 145 lbs.

Total System Load CapacitySingle Light lifting arm configuration 15 lbs.Single Heavy lifting arm configuration 99 lbs.Dual Arm Configuration, 1 Light lifting & 1 Heavy lifting arm. 114 lbs.Triple Arm Configuration, 1 Light lifting & 2 Heavy lifting arms 213 lbs.

Individual Arm Counterbalance CapacityHeavy lifting Arm up to 99 lbs.Heavy lifting Arm with large wiring channel up to 98 lbs.Light lifting Arm 15.4 lbs.

9. Specifications

Page 28: Hercules™ Suspension System Installation & …„¢ Suspension System Installation & Specifications AADCO Medical Inc “Ergonomic Solutions for X-ray Safety” AADCO Medical, Inc.

AADCO Medical, Inc.PO Box 410

Catamount Commercial Park, Route 66Randolph, VT 05060

Tel: 802-728-3400Fax: 802-728-3107

Email: [email protected]: www.aadcomed.com

AADCO Medical, Asia Pacific, Ltd.Unit B 6/F Kai Yip Factory

15-17 Sam Chuk St. Diamond HillKowloon, Hong Kong

Tel: 852-2543-4286Fax: 852-2543-4221

Email: [email protected]

AADCO Medical Inc“Ergonomic Solutions for X-ray Safety”

Hercules Manual Rev A.1 Jul 2012