Hemostasis Test Validation, Performance and …...Hemostasis Test Validation, Performance and...

32
Hemostasis Test Validation, Performance and Reference Intervals Richard A. Marlar, Ph.D. Pathology and Laboratory Medicine Oklahoma City VA Medical Center University of Oklahoma Health Sciences Center

Transcript of Hemostasis Test Validation, Performance and …...Hemostasis Test Validation, Performance and...

Page 1: Hemostasis Test Validation, Performance and …...Hemostasis Test Validation, Performance and Reference Intervals Richard A. Marlar, Ph.D. Pathology and Laboratory Medicine Oklahoma

Hemostasis Test Validation, Performance and

Reference Intervals

Richard A. Marlar, Ph.D.Pathology and Laboratory MedicineOklahoma City VA Medical Center

University of Oklahoma Health Sciences Center

Page 2: Hemostasis Test Validation, Performance and …...Hemostasis Test Validation, Performance and Reference Intervals Richard A. Marlar, Ph.D. Pathology and Laboratory Medicine Oklahoma

DISCLOSURE Richard A. Marlar, PhD

Relevant Financial RelationshipsNone

Off Label UsageNone

Page 3: Hemostasis Test Validation, Performance and …...Hemostasis Test Validation, Performance and Reference Intervals Richard A. Marlar, Ph.D. Pathology and Laboratory Medicine Oklahoma

Chemistry and Hematology Lab Values

Results expressed as amounts or participlesAlbumin 3-4 g/dLSodium 135-145 mEq/LRBC 4.3-5.9 x106/ mm3

International StandardsSeamless values between methodsExcellent reproducibility

Page 4: Hemostasis Test Validation, Performance and …...Hemostasis Test Validation, Performance and Reference Intervals Richard A. Marlar, Ph.D. Pathology and Laboratory Medicine Oklahoma

Chemistry and Hematology Testing Accuracy

Page 5: Hemostasis Test Validation, Performance and …...Hemostasis Test Validation, Performance and Reference Intervals Richard A. Marlar, Ph.D. Pathology and Laboratory Medicine Oklahoma

Coagulation Lab Values

Results expressed as arbitrary valuesPTT 32 secFactor VIII 34%FVIII Inhibitor 12 Bethesda units

Few International StandardsValues vary between methods

Heparin Therapeutic range:Reagent A 65-85 secReagent B 85-120 sec

Moderate reproducibility

Page 6: Hemostasis Test Validation, Performance and …...Hemostasis Test Validation, Performance and Reference Intervals Richard A. Marlar, Ph.D. Pathology and Laboratory Medicine Oklahoma

Coagulation Testing Accuracy

Page 7: Hemostasis Test Validation, Performance and …...Hemostasis Test Validation, Performance and Reference Intervals Richard A. Marlar, Ph.D. Pathology and Laboratory Medicine Oklahoma

, Performance and Reference

Intervals Outline

Validation of New AssayComponents required for validationProtocol for validation

Assay Method PerformanceInternal QC performanceExternal Proficiency Testing

Reference IntervalsBasic principlesCommon methods

Page 8: Hemostasis Test Validation, Performance and …...Hemostasis Test Validation, Performance and Reference Intervals Richard A. Marlar, Ph.D. Pathology and Laboratory Medicine Oklahoma

Hemostasis Test Validation

Page 9: Hemostasis Test Validation, Performance and …...Hemostasis Test Validation, Performance and Reference Intervals Richard A. Marlar, Ph.D. Pathology and Laboratory Medicine Oklahoma

Bottom-line for Hemostasis Testing

All aspects of hemostasis testing leads to Post Medical Decision MakingTesting parameters should be based on “Evidence Based Medicine”, however Tradition is more commonMust compromise between statistical number and minimal number for evaluation of a test

Page 10: Hemostasis Test Validation, Performance and …...Hemostasis Test Validation, Performance and Reference Intervals Richard A. Marlar, Ph.D. Pathology and Laboratory Medicine Oklahoma

Coagulation Laboratory Assays and Procedures

Evaluation of a pathophysiologic condition

Assist in the diagnosis of the cause of bleeding or thrombosisMonitoring Therapy

To have clinical value as an assay:Total error must be low to reflect the biological condition

Page 11: Hemostasis Test Validation, Performance and …...Hemostasis Test Validation, Performance and Reference Intervals Richard A. Marlar, Ph.D. Pathology and Laboratory Medicine Oklahoma

New Test Implementation and Validation

Range of tests:FDA approvedAnalyte Specific Reagent (ASR)Research Use Only (RUO)“Home Brew”

All new tests must be systematically validated and regulated

Level of evaluation variesAll new tests must be validatedAll tests must meet a defined level of reliabilityAll tests must continue to be validated on a regular basis

Page 12: Hemostasis Test Validation, Performance and …...Hemostasis Test Validation, Performance and Reference Intervals Richard A. Marlar, Ph.D. Pathology and Laboratory Medicine Oklahoma

Coagulation Laboratory Analysis Assay Error

For an analysis to have value:Total error must be low to differentiate between normal and abnormal

General types of error:Individual biological variabilityPre-Analytical variabilityAnalytical variability in test performanceStatistical errorInterfering substances

Page 13: Hemostasis Test Validation, Performance and …...Hemostasis Test Validation, Performance and Reference Intervals Richard A. Marlar, Ph.D. Pathology and Laboratory Medicine Oklahoma

New Test Implementation and Validation

Assay Validation ProtocolPurpose of validating is to insure the generation of high quality data for accurate diagnosis of diseaseTo ensure a proper validation, a pre-approved protocol must be usedProtocol must include:

All administrative aspects of testStatistical validation of assayAll test parameters evaluated

Page 14: Hemostasis Test Validation, Performance and …...Hemostasis Test Validation, Performance and Reference Intervals Richard A. Marlar, Ph.D. Pathology and Laboratory Medicine Oklahoma

Coagulation Test Implementation and Validation Protocol

Responsibilities

Ultimate responsibility:Laboratory DirectorDepartmental Director

Specific responsibilities:Department SupervisorLaboratory Information Manager (LIM)Quality Assurance Technologist

Page 15: Hemostasis Test Validation, Performance and …...Hemostasis Test Validation, Performance and Reference Intervals Richard A. Marlar, Ph.D. Pathology and Laboratory Medicine Oklahoma

Coagulation Test Implementation and Validation Protocol

Supervisor’s Responsibilities

Existence of patentsWritten procedure

Proper formatNot final draftUsually provided by manufacturer

Slide #1

Page 16: Hemostasis Test Validation, Performance and …...Hemostasis Test Validation, Performance and Reference Intervals Richard A. Marlar, Ph.D. Pathology and Laboratory Medicine Oklahoma

Coagulation Test Implementation and Validation Protocol

Supervisor’s ResponsibilitiesStatistical Validation study

Varies depending on type of testIncluded Parameters

AccuracyPrecisionAnalytical Sensitivity and SpecificityInterferencesReference Range and AMRLimits of DetectionCorrelation

Slide #2

Page 17: Hemostasis Test Validation, Performance and …...Hemostasis Test Validation, Performance and Reference Intervals Richard A. Marlar, Ph.D. Pathology and Laboratory Medicine Oklahoma

Coagulation Test Implementation and Validation Protocol

Supervisor’s Responsibilities

Quality Control procedureControl materialsFrequencyCriteria for acceptability

MSDS InformationComplete test information for LISNecessary information to cliniciansCost analysis and test charge

Slide #3

Page 18: Hemostasis Test Validation, Performance and …...Hemostasis Test Validation, Performance and Reference Intervals Richard A. Marlar, Ph.D. Pathology and Laboratory Medicine Oklahoma

Coagulation Test Implementation and Validation Protocol

Supervisor’s Responsibilities

Training Plan:Training methods and materialsList of staff requiring trainingDocumentation of completionDocumentation of competency assessmentNew employee training checklist

Slide #4

Page 19: Hemostasis Test Validation, Performance and …...Hemostasis Test Validation, Performance and Reference Intervals Richard A. Marlar, Ph.D. Pathology and Laboratory Medicine Oklahoma

Coagulation Test Implementation and Validation Protocol

LIM’s Responsibilities

Add test information to LIS filesVerify report format

Lab Information SystemHospital Information System

Establish CPT coding and other coding necessary for cost recovery

Page 20: Hemostasis Test Validation, Performance and …...Hemostasis Test Validation, Performance and Reference Intervals Richard A. Marlar, Ph.D. Pathology and Laboratory Medicine Oklahoma

Coagulation Test Implementation and Validation Protocol

QA Technologist’s Responsibilities

Add procedure to Document Inventory LogAssign document number

Add signed procedure to SOP manualAdd to Specimen Collection manualAdd to Activity Menu

For accrediting organizationImplement Proficiency Testing procedure

Page 21: Hemostasis Test Validation, Performance and …...Hemostasis Test Validation, Performance and Reference Intervals Richard A. Marlar, Ph.D. Pathology and Laboratory Medicine Oklahoma

Validation Parameters Coagulation's Dilemma

Validation varies on type of testFDA approvedResearch Use Only“Home Brew”

AccuracyPrecisionInterferencesReference RangeAMRLimits of DetectionCorrelation

Page 22: Hemostasis Test Validation, Performance and …...Hemostasis Test Validation, Performance and Reference Intervals Richard A. Marlar, Ph.D. Pathology and Laboratory Medicine Oklahoma

Hemostasis Test Performance

Page 23: Hemostasis Test Validation, Performance and …...Hemostasis Test Validation, Performance and Reference Intervals Richard A. Marlar, Ph.D. Pathology and Laboratory Medicine Oklahoma

Performance of Hemostasis Tests

Necessary to ensure accurate results over timeMust have on-going evaluation of the methodMust be able to compare results

From validation until current runInternal QC:

On-going periodic analysisRequired by regulatory agenciesSet of rules to determine acceptance or failure

External Quality AssessmentPeriodic comparison with other labsBest measure of accuracySet of rules to determine acceptance or failure

Page 24: Hemostasis Test Validation, Performance and …...Hemostasis Test Validation, Performance and Reference Intervals Richard A. Marlar, Ph.D. Pathology and Laboratory Medicine Oklahoma

Reference Interval in Coagulation and Hemostasis

INTERPRETATION OF COAGULATION DATA IS A COMPARATIVE DECISION-MAKING PROCESS!Compare patient result to reference intervalReference interval set up for:

“Normal” or healthyPhysiologic conditionsPathologic conditions

Therapeutic reference interval is determined in similar manner

Page 25: Hemostasis Test Validation, Performance and …...Hemostasis Test Validation, Performance and Reference Intervals Richard A. Marlar, Ph.D. Pathology and Laboratory Medicine Oklahoma

Establishing Reference Intervals for Coagulation Assays

Ref Interval Criteria

New Analyte Determination is difficultLack data on physiology, pathology and medical aspects

New Method Determination somewhat easierPhysiology and medical aspects known

Transference Easier to determineMust use caution in establishing

Page 26: Hemostasis Test Validation, Performance and …...Hemostasis Test Validation, Performance and Reference Intervals Richard A. Marlar, Ph.D. Pathology and Laboratory Medicine Oklahoma

Use of “Transference” for Establishing Reference Interval

Applying previously established ranges (by lab, others or literature) to new assay methods.Methods to accept transference:

Use of inspection and subjectivity of method procedure to determine similar rangeValidation of range with small number of reference samples

Methods not well established for coagulation assays

Page 27: Hemostasis Test Validation, Performance and …...Hemostasis Test Validation, Performance and Reference Intervals Richard A. Marlar, Ph.D. Pathology and Laboratory Medicine Oklahoma

Validation of Coagulation Tests Summary

Validation of a coagulation test can be difficultSet up protocol for validation:

Protocol written Approved before starting

Protocol to Include:Administrative issuesValidation with statistical analysis

Page 28: Hemostasis Test Validation, Performance and …...Hemostasis Test Validation, Performance and Reference Intervals Richard A. Marlar, Ph.D. Pathology and Laboratory Medicine Oklahoma

Performance Criteria of Coagulation Tests

Summary

Continued evaluation of assay Internal Quality Control

Periodic testing of same sampleMaintain precision and accuracy over timeEstablished rules for acceptability

External Quality ControlMeasure of accuracy and precision

Page 29: Hemostasis Test Validation, Performance and …...Hemostasis Test Validation, Performance and Reference Intervals Richard A. Marlar, Ph.D. Pathology and Laboratory Medicine Oklahoma

Reference Intervals Determinations for Coagulation Tests

SummaryReference interval is critical for clinical decision makingDifferent types of reference intervals:

NormalPhysiologicPathologicTherapeutic

Reference interval determinations:New analyteNew methodTransference

Page 30: Hemostasis Test Validation, Performance and …...Hemostasis Test Validation, Performance and Reference Intervals Richard A. Marlar, Ph.D. Pathology and Laboratory Medicine Oklahoma

Reference Literature for Validation, Performance and Ref Intervals

Guidelines for Validation of Assays:CLSI Evaluation Documents for Validation Studies (EP05-EP21)

Guidelines for Reference Intervals:CLSI Document on Reference Intervals in the Clinical Laboratory (C28-A2)

Guidelines for Specific Coagulation Tests:CLSI Document on PT and APTT (H47-A2)Numerous CLSI Documents on Coagulation Testing (H series)

Page 31: Hemostasis Test Validation, Performance and …...Hemostasis Test Validation, Performance and Reference Intervals Richard A. Marlar, Ph.D. Pathology and Laboratory Medicine Oklahoma

Questions?

Page 32: Hemostasis Test Validation, Performance and …...Hemostasis Test Validation, Performance and Reference Intervals Richard A. Marlar, Ph.D. Pathology and Laboratory Medicine Oklahoma

Testing Validation Parameters Coagulation's Compromise

Optimum Practical MeasureAccuracy 100 10-20 +5%Precision 10 within

10 between5 within runs

5 between runs%CV

Interferences 25 of each ~5 of eachReference Range 100 normals 20-40 normals Mean +2 SD

AMR 50-100 20-40Limits of Detection 25

undetectable5-10

undetectable+3 SD of

zeroCorrelation 200 20-25 R > 0.90

Slope ~1.0