Hematology Analyzer 23Hematology Analyzer 23 (A new generation 3-diff hematology analyzer)...
Transcript of Hematology Analyzer 23Hematology Analyzer 23 (A new generation 3-diff hematology analyzer)...
@zyb10
Hematology Analyzer 23
(A new generation 3-diff hematology analyzer)
www.zybio.com Distributed by: Malaysian Diagnostics Corporation Sdn. Bhd.
Distributed by:
Highlight
Easy operation and maintenance
Intuitive navigation menu
10.4-inch large touch screen
Cost saving
Two reagents, lyse inside the analyzer
Low reagent consumption
Good performance in capillary blood mode
WB mode: 1 0 µL
PD mode: 20 µL
Applicable for pediatric dept.
- -
Compact design
Small size for space saving
The fastest throughput
70 samples/h
Distributed by:
General Feature
• One key to aspirate probe cleanser for deep cleaning in whole device
• Multi-language, such as Chinese, English, Spanish, Portuguese and etc.
• Freely shift parameter unit, including China, International, UK, French, Netherlands and etc.
• Longer shelf life up to 24 months for Diluent and Lyse
• Cyanide free measurement for HGB, stable and environment-friendly liquid
• Simplified and modern outer packaging stem from Swiss design
• Reagent inventory monitored by two ways including software and hardware components
• Automatically loaded lot number and expiry information through a barcode scanner
,,
@zyb10
r
Distributed by:
Performance & Specification
Performance
Parameters
WBC(109/L)
RBC(1012/L)
HGB(g/dL)
HCT(%)/MCV(fL)
PLT(109/L)
Specification
Principle
Parameter
Throughput
Sample volume
Reagent
Calibration
QC
Input
Output
Data storage
Interface
Power requirement
Working environment
Weight
Dimension(mm)
Background
:;;0.2
:;;0.02
:;;1
HCT,;;0.5
,;;5
Precision
Detection Limits CV
(7.00-15.00) ,;;2.0%
(4.00-6.9) :;;2,5%
(3.50-6.50) ,;;1.5%
(11-18) :;;1,5%
MCV:(70-110) ,;;0.5%
(100-149) ,;;5.0%
(150-500) ,;;4.0%
Impedance method for WBC, RBC, PL T counting
Cyanide free reagent for HGB test
Carry-over Linearity range
:;;0.5% (0-300.00)
:;;0.5% (0-8.00)
:;;0,5% (0-25)
I HCT:(0-67)
,;;1.0% (0-4000)
21 Parameters: WBC, Lym#, Mid#, Gran#, Lym%, Mid%, Gran%, RBC, HGB, HCT,
MCV, MCH, MCHC, ROW-CV, ROW-SD, PLT, MPV, POW, PCT, P-LCR, P-LCC
3 Histograms: WBC, RBC, PL T
70 samples per hour
Whole blood: 10 µI
Pre-diluted: 20 µI
2 reagents+ 1 maintenance cleanser
(Diluent, Lyse, Probe cleanser)
Manual calibration
Auto calibration: calibrator, fresh blood
L-J,X-B
10.4-inch touch screen
Mouse, keyboard (optional)
Built-in recorder, external printer (optional)
;:,,40,000
LAN port, HL7 protocol, support bilateral LIS
4 USB ports for external printer, external barcode scanner, keyboard, etc.
1 OOV-240V~, 50/60Hz, < 200W
15°C-35°C
18Kg
300(W)X400(D)X41 O(H)
Distributed by:
@zyb10
Manufactured by :
Zybio Inc.
Address: Floor 1 to Floor 4, Building 30, No.6 of Taikang Road,
Block C of Jianqiao Industrial Park, Dadukou District, Chongqing, China 400082
Tel: +86-23 6865 5509 Fax: +86-23 6869 9779
Email: [email protected] Website: www.zybio.com
Hematology
www.zybio.com
Distributed by :
Malaysian Diagnostics Corporation Sdn. Bhd.
Address: No.1, Jalan Serindit 2, Bandar Puchong Jaya, 47100 Puchong,
Selangor, Malaysia.
Tel: +6 03-5882 7011