Helsinki T2 status
description
Transcript of Helsinki T2 status
Helsinki T2 status
• the 1st production GEM under tests (operational tests)
• the 2nd GEM not accepted due to persistent leakage current problems in two foils (all other parts finished) - wait for components for replacement (and for further GEMs)
• spare parts needed in future to avoid delays in the future
Risto Orava/ on behalf of Kari Kurvinen 8.11.2005
Characterisation of GEM foils:
A. leakage current measurementsB. visual inspectionC. optical scanning
Leakage current measurements of GEM foils• the four segments (A,B,C & D, see Fig.) of the GEM foils are measured separately by Picoamp /Voltage source device (Keithley 487).
• due to these observations, a graphical display (Labview) was added to the measurement system to detect & document instabilities as early as possible (the display was not yet available for the first production GEM).
• current limited by 100 M resistor (corresponding to the max current of 5 A at 500V).
• a foil is approved if the current stays at < 0.5 nA for 5+ min @500V
•leakage currents are measured three times during the assembly (the foil before and after framing, then after gluing into a stack).
36 measurements / detector
• so far 2 /12 foils found to be bad. These were observed to be temporarily short circuited during the second measurement! Both exhibited instabilities during the first measurement, still passed the criteria.
3 x 3 foils x 4
A bad GEM foil
TIFB-segment
C-segment
D-segment
A-segment is short circuited (60 – 150 k)
B,C & D segments OK
noise
HV ramp up
noise
noise
(Note: the time scale in the plots must be multiplied by two)
10 min0 min
6 min
6 min
0.5 nA
0.5 nA
0.5 nA
A ”good” GEM foil ?
(Note: the time scale in the plots must be multiplied by 2)
A-segment A-segment
in air
HV ramp up N2 flush started
N2 flush stopped
in N2
in N2
in N2
spark!
spark!
B-segment
C-segment
D-segment
> 5 min
> 10 min
noise
0.5 nA
0.5 nA
14 min 20 min
14 min
12 min
10 min
Visual inspection
• only the largest visible defects are observable• coordinates are recorded approximately• serious defects due to dust, stains or scratches are easy to see
number of recorded defects in the six latest foils:GEM 1 18, 12, 10 (top, middle, bottom)
GEM 2 10, 31, 40
General observation:the quality of GEM foils is good(i.e. is significantly improved from the past)
defect in the mask 0.5 mm
defect in the mask
2. Optical scanning of the GEM foils• scanning of the entire foil with a
resolution of 2400 (3200) dpi f easible• 0.5 (2) h scanning time per f oil• in a reflective or in a transparent mode
a) the most common defects recorded for the future reference (joint holes, stains, scratches)
b) blocked holes seen easily in the transparentmode (also partially blocked ones)
c) tracking of certain defects by sof tware f romthe picture fi le possible (e.g. by measuring the hole sizes)
Some improvements to the qualitychecking of the GEM manufacturing
1. Handling procedures of the separate detectorcomponents will be registered into a database
Optical scanning of GEM foils• scanning in mixed mode (transparent + reflective)•blue diffuser produces a colour contrast between the holes and copper surfaces1.8 million GEM holes
defects, dust, scratches,etc
scanned pictures can be analysed by using fitting algorithms to count anomalies and to register their coordinates
The worst foil so far: Blue curve shows all the seen objects; the red one only the ”roughly round ones” (e.g.scratches are filtered out). The number of defects (red ones > 80) was 240, while only 40 were found under visual inspection (see previous slide).
AIM: a quantitative measure of the quality of GEM foils