HEARING AID HEURISTICS ANALYSIS
• Objective: To conduct usability and human factors testing on hearing aid device to assess if the required specifications are met.
• Specifications :Spec # Source Specification
Direction
Units of measure Marginal Ideal Notes DRI
S9
CN2, CN4, CN11
Range of adult ear size accommodated
max percentile25th to 75th
10th to 90th
Use ANSUR database to obtain dimensions corresponding to theses percentiles
ISE
S10
CN4, CN6, CN10, CN11
Weight of earpiece min g <15 <12marginal values will depend on ergonomic info
ISE
S14 CN3
Percent of surveyed people who identify a picture of the device as something other than a hearing aid.
max percent >60 >80team should conduct surveys
ID
S15 CN3
Percent of surveyed hard of hearing people who prefer the form of the new device to standard behind the ear hearing aids
max percent >50 >75team should conduct surveys
ID
S16 CN3
Percent of surveyed hearing people who would use the device for Bluetooth or music listening
max percent >50 >70team should conduct surveys
ID
S17 CN11
Percent of surveyed people who feel the device is comfortable to wear
max percent >60 >80team should conduct surveys
ID
Specifications 9, 10• Range of adult ear size accommodated
• Final design will be measured to ensure dimensions fall with in the 5%-95% percentile
• Weight of earpiece• Final prototype will be weighed to ensure it’s weight is below 15
grams
SPECIFICATIONS 14-17 Two Surveys:• Small Group • Imagine-RIT
Small Group Survey• Small heuristics study of 5-6 students between the ages
of 18-25• Evaluators can be either Hearing or Hard of Hearing
students• Incentive for participating in survey will be granted to
evaluators (Javas Tokens)• Data collected will be confidential
Procedure• Conduct one on one interview with each participant for a
maximum of 20 minutes• Each participate will be ask to wear a non-functional
prototype and perform a set of tasks such as walking with device, adjusting volume etc.
• A Likert system will be used to rank satisfaction of each question
ImagineRIT Survey • ImagineRIT visitors who are at least 18 or older will be
allowed to participate in this survey • Each participate will be ask to wear a non-functional
prototype and perform a set of tasks such as adjusting volume/profiles etc.
• A Likert system will be used to rank their response for each question.
Survey Questions• Each question will reference a function on the device to
evaluate basic usability standards:• Flexibility and efficiency of use• Aesthetic and minimalist design • User Control and freedom • Recognition rather than recall
Sample Questions
Please mark how satisfied you are with the hearing aid device for each situation:• Overall comfort/fit
• Very satisfied, Satisfied, Neutral, Dissatisfied, Very Dissatisfied
• Ease of adjusting volume• Very satisfied, Satisfied, Neutral, Dissatisfied, Very Dissatisfied
• Visibility of hearing aid• Very satisfied, Satisfied, Neutral, Dissatisfied, Very
Dissatisfied
Statistics For Analyzing Data
Ordinal data:• α= 0.05• Frequency Distribution/Cross Tabulation• Kendall tau B or C• Chi-square test