Haze Measurements Haze Measurements in Beerin Beer
Ellen Norman
![Page 2: Haze Measurements in Beer - National Physical Laboratoryresource.npl.co.uk/optical_radiation/ormclub/member/content/jun07/... · • Can be an indicator of process ... Hach Radiometer](https://reader036.fdocuments.us/reader036/viewer/2022062306/5a75b2267f8b9a9c548cbc55/html5/thumbnails/2.jpg)
science with service
Haze in beer
• Definitions• Haze types in beer • Challenges to harmonised
measurements• Current solutions
![Page 3: Haze Measurements in Beer - National Physical Laboratoryresource.npl.co.uk/optical_radiation/ormclub/member/content/jun07/... · • Can be an indicator of process ... Hach Radiometer](https://reader036.fdocuments.us/reader036/viewer/2022062306/5a75b2267f8b9a9c548cbc55/html5/thumbnails/3.jpg)
science with service
Haze in beer
• Key quality parameter• One of the first things a consumer
makes a judgement on• Can be an indicator of process
efficiency• Measured under a variety of
conditions
![Page 4: Haze Measurements in Beer - National Physical Laboratoryresource.npl.co.uk/optical_radiation/ormclub/member/content/jun07/... · • Can be an indicator of process ... Hach Radiometer](https://reader036.fdocuments.us/reader036/viewer/2022062306/5a75b2267f8b9a9c548cbc55/html5/thumbnails/4.jpg)
science with service
Haze definitions
• EBC = European Brewery Convention
• ASBC = American Society of Brewing Chemists
• FNU = Formazine Nephelometric Unit
![Page 5: Haze Measurements in Beer - National Physical Laboratoryresource.npl.co.uk/optical_radiation/ormclub/member/content/jun07/... · • Can be an indicator of process ... Hach Radiometer](https://reader036.fdocuments.us/reader036/viewer/2022062306/5a75b2267f8b9a9c548cbc55/html5/thumbnails/5.jpg)
science with service
Conversions
1 EBC 1 FNU/NTU 1 ASBC
1 0.25 0.014 EBC
4 1 0.057 FNU/NTU
70 17.5 1 ASBC
![Page 6: Haze Measurements in Beer - National Physical Laboratoryresource.npl.co.uk/optical_radiation/ormclub/member/content/jun07/... · • Can be an indicator of process ... Hach Radiometer](https://reader036.fdocuments.us/reader036/viewer/2022062306/5a75b2267f8b9a9c548cbc55/html5/thumbnails/6.jpg)
science with service
Descriptive scale
Grade EBC ASBC
Brilliant 0.0 to 0.5 0.0 to 34.5
Almost brilliant 0.5 to 1.0 34.5 to 69
Very Slightly Hazy / Cast
1.0 to 2.0 69 to 138
Slightly Hazy 2.0 to 4.0 138 to 276
Hazy 4.0 to 8.0 276 to 552
Very hazy > 8.0 > 552
![Page 7: Haze Measurements in Beer - National Physical Laboratoryresource.npl.co.uk/optical_radiation/ormclub/member/content/jun07/... · • Can be an indicator of process ... Hach Radiometer](https://reader036.fdocuments.us/reader036/viewer/2022062306/5a75b2267f8b9a9c548cbc55/html5/thumbnails/7.jpg)
science with service
Challenges to harmonisation
• Different manufacturers • Angles of measurement • Wavelength• Angles of aperture• Standards for calibration
![Page 8: Haze Measurements in Beer - National Physical Laboratoryresource.npl.co.uk/optical_radiation/ormclub/member/content/jun07/... · • Can be an indicator of process ... Hach Radiometer](https://reader036.fdocuments.us/reader036/viewer/2022062306/5a75b2267f8b9a9c548cbc55/html5/thumbnails/8.jpg)
science with service
Recommended methods
• EBC–Measuring wavelength not specified
• ASBC–580 nm
• MEBAK–650 ± 30 nm
![Page 9: Haze Measurements in Beer - National Physical Laboratoryresource.npl.co.uk/optical_radiation/ormclub/member/content/jun07/... · • Can be an indicator of process ... Hach Radiometer](https://reader036.fdocuments.us/reader036/viewer/2022062306/5a75b2267f8b9a9c548cbc55/html5/thumbnails/9.jpg)
science with service
Other method issues
• Degassing• Temperature
–Chill haze
• Forced ageing• Laboratory• In-line
![Page 10: Haze Measurements in Beer - National Physical Laboratoryresource.npl.co.uk/optical_radiation/ormclub/member/content/jun07/... · • Can be an indicator of process ... Hach Radiometer](https://reader036.fdocuments.us/reader036/viewer/2022062306/5a75b2267f8b9a9c548cbc55/html5/thumbnails/10.jpg)
science with service
Sample preparation
60%19%
10%3%
8%
No degasTossingSonicationWarmingOthers
![Page 11: Haze Measurements in Beer - National Physical Laboratoryresource.npl.co.uk/optical_radiation/ormclub/member/content/jun07/... · • Can be an indicator of process ... Hach Radiometer](https://reader036.fdocuments.us/reader036/viewer/2022062306/5a75b2267f8b9a9c548cbc55/html5/thumbnails/11.jpg)
science with service
Cuvette vs. bottle
Product Cuvette
∅ 60 mm
Brown bottle
∅ 60.5 mm
Green bottle
∅ 59.8 mm
Formazin 0.5 EBC 0.51 0.49 0.50
Formazin 1.0 EBC 0.99 0.98 0.98
Formazin 2.5 EBC 2.50 2.48 2.48
Pils ( colour 10 EBC) 0.28 0.25 0.33
Dark beer (colour 110 EBC) 2.93 3.13 2.97
Brau. Ind. 10/2000, 568-571.
![Page 12: Haze Measurements in Beer - National Physical Laboratoryresource.npl.co.uk/optical_radiation/ormclub/member/content/jun07/... · • Can be an indicator of process ... Hach Radiometer](https://reader036.fdocuments.us/reader036/viewer/2022062306/5a75b2267f8b9a9c548cbc55/html5/thumbnails/12.jpg)
science with service
Types of haze meter used - 1
0
5
10
15
20
MonitekHaffmans VosLG AutoHachRadiometerSigristDr LangeOther
![Page 13: Haze Measurements in Beer - National Physical Laboratoryresource.npl.co.uk/optical_radiation/ormclub/member/content/jun07/... · • Can be an indicator of process ... Hach Radiometer](https://reader036.fdocuments.us/reader036/viewer/2022062306/5a75b2267f8b9a9c548cbc55/html5/thumbnails/13.jpg)
science with service
Types of haze meter used - 2
0
5
10
15
20
MonitekHaffmans VosLG AutoHachRadiometerSigristDr LangeOther
![Page 14: Haze Measurements in Beer - National Physical Laboratoryresource.npl.co.uk/optical_radiation/ormclub/member/content/jun07/... · • Can be an indicator of process ... Hach Radiometer](https://reader036.fdocuments.us/reader036/viewer/2022062306/5a75b2267f8b9a9c548cbc55/html5/thumbnails/14.jpg)
science with service
Variation in measurementsLaboratory turbidity equipment
Formazin in cuvette Beer (Pils) in cuvette
0.5 EBC 1.0 EBC 2.5 EBC + 1.0 EBC
+ 2.0 EBC
VOS ROTA 90 0.50 1.00 2.50 0.28 1.20 2.02
VOS 4000 0.49 0.98 2.46 0.46 1.22 1.65
Radiometer (green light)
0.48 0.90 2.15 0.35 1.00 1.50
Sigrist (green light)
0.48 0.94 2.45 0.51 1.35 1.93
Dr. Lange 0.62 1.08 2.58 0.26 1.46 2.29
Brau. Ind. 10/2000, 568-571.
![Page 15: Haze Measurements in Beer - National Physical Laboratoryresource.npl.co.uk/optical_radiation/ormclub/member/content/jun07/... · • Can be an indicator of process ... Hach Radiometer](https://reader036.fdocuments.us/reader036/viewer/2022062306/5a75b2267f8b9a9c548cbc55/html5/thumbnails/15.jpg)
science with service
Measurement angle - 1
0
10
20
30
40
50
60
13°
90°
![Page 16: Haze Measurements in Beer - National Physical Laboratoryresource.npl.co.uk/optical_radiation/ormclub/member/content/jun07/... · • Can be an indicator of process ... Hach Radiometer](https://reader036.fdocuments.us/reader036/viewer/2022062306/5a75b2267f8b9a9c548cbc55/html5/thumbnails/16.jpg)
science with service
Measurement angle - 2
17
45
2
90/25°90°13°
![Page 17: Haze Measurements in Beer - National Physical Laboratoryresource.npl.co.uk/optical_radiation/ormclub/member/content/jun07/... · • Can be an indicator of process ... Hach Radiometer](https://reader036.fdocuments.us/reader036/viewer/2022062306/5a75b2267f8b9a9c548cbc55/html5/thumbnails/17.jpg)
science with service
Calibration standards
56
81 1
FormazinPolymerDiatomaceous earthStandard glass
![Page 18: Haze Measurements in Beer - National Physical Laboratoryresource.npl.co.uk/optical_radiation/ormclub/member/content/jun07/... · • Can be an indicator of process ... Hach Radiometer](https://reader036.fdocuments.us/reader036/viewer/2022062306/5a75b2267f8b9a9c548cbc55/html5/thumbnails/18.jpg)
science with service
Causes of haze in beer
• ‘Natural’ hazes–precipitates & colloids
• Process aids–Poor quality– Incorrect use
• Contaminants–Foreign bodies–Microbial infection
![Page 19: Haze Measurements in Beer - National Physical Laboratoryresource.npl.co.uk/optical_radiation/ormclub/member/content/jun07/... · • Can be an indicator of process ... Hach Radiometer](https://reader036.fdocuments.us/reader036/viewer/2022062306/5a75b2267f8b9a9c548cbc55/html5/thumbnails/19.jpg)
science with service
Protein hazes
• Protein-PGA• Protein-polyphenol• Chill haze: Tannoids and protein• Permanent polymerisation• Protein-pentosan• Protein-carbohydrate• Protein-mannose
![Page 20: Haze Measurements in Beer - National Physical Laboratoryresource.npl.co.uk/optical_radiation/ormclub/member/content/jun07/... · • Can be an indicator of process ... Hach Radiometer](https://reader036.fdocuments.us/reader036/viewer/2022062306/5a75b2267f8b9a9c548cbc55/html5/thumbnails/20.jpg)
science with service
Large Proteinaceous Skins
![Page 21: Haze Measurements in Beer - National Physical Laboratoryresource.npl.co.uk/optical_radiation/ormclub/member/content/jun07/... · • Can be an indicator of process ... Hach Radiometer](https://reader036.fdocuments.us/reader036/viewer/2022062306/5a75b2267f8b9a9c548cbc55/html5/thumbnails/21.jpg)
science with service
Protein-polyphenol
![Page 22: Haze Measurements in Beer - National Physical Laboratoryresource.npl.co.uk/optical_radiation/ormclub/member/content/jun07/... · • Can be an indicator of process ... Hach Radiometer](https://reader036.fdocuments.us/reader036/viewer/2022062306/5a75b2267f8b9a9c548cbc55/html5/thumbnails/22.jpg)
science with service
Carbohydrate hazes
• Beta Glucan–Chill haze–Permanant particles
• Starch, dextrins • Pentosans
![Page 23: Haze Measurements in Beer - National Physical Laboratoryresource.npl.co.uk/optical_radiation/ormclub/member/content/jun07/... · • Can be an indicator of process ... Hach Radiometer](https://reader036.fdocuments.us/reader036/viewer/2022062306/5a75b2267f8b9a9c548cbc55/html5/thumbnails/23.jpg)
science with service
Sub-visible carbohydrates
![Page 24: Haze Measurements in Beer - National Physical Laboratoryresource.npl.co.uk/optical_radiation/ormclub/member/content/jun07/... · • Can be an indicator of process ... Hach Radiometer](https://reader036.fdocuments.us/reader036/viewer/2022062306/5a75b2267f8b9a9c548cbc55/html5/thumbnails/24.jpg)
science with service
Other ‘natural’ hazes
• Hop constituents–Polyphenols– iso-alpha acids, reduced hop acids
• Calcium oxalate • Glycerol phosphate monomers
![Page 25: Haze Measurements in Beer - National Physical Laboratoryresource.npl.co.uk/optical_radiation/ormclub/member/content/jun07/... · • Can be an indicator of process ... Hach Radiometer](https://reader036.fdocuments.us/reader036/viewer/2022062306/5a75b2267f8b9a9c548cbc55/html5/thumbnails/25.jpg)
science with service
Process aids
• Diatomaceous earth & clay• Polyamides• Silica, and silicates• Cellulose fibres• Finings• Alginates & their esters (PGA)
![Page 26: Haze Measurements in Beer - National Physical Laboratoryresource.npl.co.uk/optical_radiation/ormclub/member/content/jun07/... · • Can be an indicator of process ... Hach Radiometer](https://reader036.fdocuments.us/reader036/viewer/2022062306/5a75b2267f8b9a9c548cbc55/html5/thumbnails/26.jpg)
science with service
Process aids
![Page 27: Haze Measurements in Beer - National Physical Laboratoryresource.npl.co.uk/optical_radiation/ormclub/member/content/jun07/... · • Can be an indicator of process ... Hach Radiometer](https://reader036.fdocuments.us/reader036/viewer/2022062306/5a75b2267f8b9a9c548cbc55/html5/thumbnails/27.jpg)
science with service
Contaminants
• Man-made• Detergents; oil; can lacquer; glass;
silicates and phosphates.
• Microbiological hazes• yeast: wild or brewers • bacteria: cocci or rods• moulds: mycelium/filaments• algae & fungi
![Page 28: Haze Measurements in Beer - National Physical Laboratoryresource.npl.co.uk/optical_radiation/ormclub/member/content/jun07/... · • Can be an indicator of process ... Hach Radiometer](https://reader036.fdocuments.us/reader036/viewer/2022062306/5a75b2267f8b9a9c548cbc55/html5/thumbnails/28.jpg)
science with service
Beer Colour
• Some red and darker beers can give artificially high results
• Only on certain instruments–Light wavelength is key factor
• Most brewers set specifications accordingly
• A ratio measurement compensates influence of beer colour
![Page 29: Haze Measurements in Beer - National Physical Laboratoryresource.npl.co.uk/optical_radiation/ormclub/member/content/jun07/... · • Can be an indicator of process ... Hach Radiometer](https://reader036.fdocuments.us/reader036/viewer/2022062306/5a75b2267f8b9a9c548cbc55/html5/thumbnails/29.jpg)
science with service
Summary
• Haze in beer is a key quality parameter
• There is no international standard for the analysis
• Ring trials and collaboration are beginning to harmonise methods
• Ultimately individual brewers work with what is best for them