Hardware Update John Jayne 5 th AMS User’s Meeting 10/11/04

Transcript of Hardware Update - cires1.colorado.educires1.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... ·...

Page 1: Hardware Update - cires1.colorado.educires1.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · Cryopump Temp Test 1 - Pump from vent condition Cryopump on from start Test 2 - Cryopump

Hardware Update

John Jayne5th AMS User’s Meeting

10/11/04

Page 2: Hardware Update - cires1.colorado.educires1.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · Cryopump Temp Test 1 - Pump from vent condition Cryopump on from start Test 2 - Cryopump

Hardware • TOF – AMS• Light Scattering module• Short chamber• Beam Width Probe• High pressure lens for larger size particles• Rigid critical aperture holder • Cryopump• New Vaporizer…

Page 3: Hardware Update - cires1.colorado.educires1.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · Cryopump Temp Test 1 - Pump from vent condition Cryopump on from start Test 2 - Cryopump
Page 4: Hardware Update - cires1.colorado.educires1.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · Cryopump Temp Test 1 - Pump from vent condition Cryopump on from start Test 2 - Cryopump

18 mm

450 mm

403 mm

378 mm

140 mmPivot point

Channel skimmer1 mm ID x 25.4 L

Channel aperture3.8mm ID x 20 mm L

Channel aperture3.8mm ID x 10 mm L

0.15”(3.8mm) OD Heater

Distances and Aperturesfor 255-xxx AMS Chamber

395 mm

178 mm

Chopper

353 mmBeam Probe

Aug. 2003

New chamber is 10 cm shorter, series 215-xxx

Page 5: Hardware Update - cires1.colorado.educires1.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · Cryopump Temp Test 1 - Pump from vent condition Cryopump on from start Test 2 - Cryopump

Lens For larger particle transmission~0.3 – 3 umUnit: mm

d1=0.75 d2=0.65 d3=0.55 d4=0.45 d5=0.37 d6=0.31 d7=0.25

2.5

15

0.2

0.25

1.0

20D

10D

0.4

240

2

Sevenlenses

Sevenlenses

Freeflight

Targetplate

Particleinjection

Gasinlet

Schreiner at al. (1999) Zhang et al. (2002)

Page 6: Hardware Update - cires1.colorado.educires1.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · Cryopump Temp Test 1 - Pump from vent condition Cryopump on from start Test 2 - Cryopump

VCO Style Critical Aperture

Holder

Page 7: Hardware Update - cires1.colorado.educires1.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · Cryopump Temp Test 1 - Pump from vent condition Cryopump on from start Test 2 - Cryopump

The orientation of the pinhole is important

“Pimple” effect

Page 8: Hardware Update - cires1.colorado.educires1.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · Cryopump Temp Test 1 - Pump from vent condition Cryopump on from start Test 2 - Cryopump

Possible Design for Improved Critical Aperture Holder

Page 9: Hardware Update - cires1.colorado.educires1.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · Cryopump Temp Test 1 - Pump from vent condition Cryopump on from start Test 2 - Cryopump

Current HTP w/Critical Aperture and Valve Body

Page 10: Hardware Update - cires1.colorado.educires1.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · Cryopump Temp Test 1 - Pump from vent condition Cryopump on from start Test 2 - Cryopump

Current Standard Lens w/Critical Aperture and Valve Body

Page 11: Hardware Update - cires1.colorado.educires1.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · Cryopump Temp Test 1 - Pump from vent condition Cryopump on from start Test 2 - Cryopump

Ricor MicroStarTM Cryopumpfast pump out / background reduction in ARI Aerosol Mass

Spectrometer ionizer chamberhttp://www.ricor.com

•1000 Ls-1 pumping speed for water• 17 kg (38 lbs)• 400 Watts start-up at 50 VDC

(~200W normal load)

• Cryoshiled surrounds ionizer• Mounted between P5 (V301) and chamber

Page 12: Hardware Update - cires1.colorado.educires1.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · Cryopump Temp Test 1 - Pump from vent condition Cryopump on from start Test 2 - Cryopump

Comparison of Mass Spec with and without Cryopump

102

103

104

105

106

107

Ion

Rat

e (H

z)

100908070605040302010AMU

Normal Background Background with Ricor Pump

120K

Water vapor drops ~15x; organics drop between 2 and 10x

Page 13: Hardware Update - cires1.colorado.educires1.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · Cryopump Temp Test 1 - Pump from vent condition Cryopump on from start Test 2 - Cryopump

Ricor Microstar Cryopumpperformance in AMS

102

103

104

105

106

107

Ion

Sig

nal (

Hz)

6:20 PM9/18/2003

6:45 PM

dat

9:00 AM9/24/2003

10:00 AM

dat

400

350

300

250

200

150

100

50

Cryopum

p Temp, K

10:00 AM9/26/2003

12:00 PM

dat

Water (m18) Organic (m55+m57) Cryopump Temp

Test 1 - Pump from vent condition Cryopump on from start

Test 2 - Cryopump on after 4 days of 'normal' pumping

Test 4 - Ionizer off overnight, turbos on Cryopump on later

Ionizer on

Cryopump on

Page 14: Hardware Update - cires1.colorado.educires1.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · Cryopump Temp Test 1 - Pump from vent condition Cryopump on from start Test 2 - Cryopump

Current Vaporizer Design

Nov. 17, 2000John Jayne

Scale 1" = .25"

Aerodyne Research, Inc.45 manning Road Billerica, MA 01821(978)663-9500 (x233)[email protected]

Note A

Notes:

A- Two heater leads.

B- make one heater using 80% dense molybdenum and one heater using 50% dense molybdenum (radius of curvature on cone angles < ~ .015)

One lead attached to ID of thermally isolating support

0.285" 0.150"

60 DegNote B

0.15

0"

0.400"

Thermally isolating support

Thermocouple

Page 15: Hardware Update - cires1.colorado.educires1.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · Cryopump Temp Test 1 - Pump from vent condition Cryopump on from start Test 2 - Cryopump

30 Deg

0.137"

20 Deg

0.198"

40 Deg

0.084"

Oven Angle

1) 180 - 2) 3) 180 - 54) 5

N relative to beam axis

Specular reflectionsN incoming on beam axis

N 1st reflectionN 2nd reflection

N 3rd reflection5) 180 - 9N 4th reflection6) 9N 5th reflection

25 Deg

0.157"

0.15

0"

Minimum No. of Collisions length20 4 .19825 3 .15730 3 .13740 2 .084

Oven Angle

Cone Shaped Oven Heater Design to maximize partic le collisions and minimize cone length

Page 16: Hardware Update - cires1.colorado.educires1.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · Cryopump Temp Test 1 - Pump from vent condition Cryopump on from start Test 2 - Cryopump

Particle trapping type vaporizer

Svane et al, 2004Pettersson Sweden

• Will effect temporal profile of vapor plume and resolution of particle size measurement.• Will be physically larger, or will reduce collection angle.• Temperature gradients…

Page 17: Hardware Update - cires1.colorado.educires1.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · Cryopump Temp Test 1 - Pump from vent condition Cryopump on from start Test 2 - Cryopump

Digital interface for control electronics??

-eliminate toggle switches-enable data logging of pumps stats and vaporizer,

etc.-enable computer control of pumps, vaporizer,

chopper, etc.

Page 18: Hardware Update - cires1.colorado.educires1.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · Cryopump Temp Test 1 - Pump from vent condition Cryopump on from start Test 2 - Cryopump

Discussion of hardware failures

Wire fatigue problem on chopper assembly

Page 19: Hardware Update - cires1.colorado.educires1.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · Cryopump Temp Test 1 - Pump from vent condition Cryopump on from start Test 2 - Cryopump
Page 20: Hardware Update - cires1.colorado.educires1.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · Cryopump Temp Test 1 - Pump from vent condition Cryopump on from start Test 2 - Cryopump

Air Beam and Ionization Efficiency

• Ionization efficiency, IE (ions/molecule)– Sample a particle of known size (no. of

molecules) and measure the number of ions produced. Typically a few ions generated from 106 molecules.

• Air Beam, AB (Hz, ions / second)– The signal at N2 (or O2), use signal strength

as a measure of instrument sensitivity.Typically a few million ions per second.

Page 21: Hardware Update - cires1.colorado.educires1.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · Cryopump Temp Test 1 - Pump from vent condition Cryopump on from start Test 2 - Cryopump

Variability in Ionization Efficiency and Air Beam

• Position of filament with respect to slit in ion reference region. IE/AB constant

• Multiplier collection efficiency. IE/AB constant• Pressure in TOF chamber (small internal leaks

can have a measurable effect on AB intensity). IE/AB larger

• Positioning of vaporizer inside ionizer volume IE/AB smaller

• Machining imperfections in skimmer channel. AB could be off-axis. IE/AB larger

Page 22: Hardware Update - cires1.colorado.educires1.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · Cryopump Temp Test 1 - Pump from vent condition Cryopump on from start Test 2 - Cryopump

Summary of AMS Performance at Time of DeliveryHigh efficiency ionizer

High performance quad

12x106

8

4

0

Air B

eam (H

z)

Cal

tech

BC

-Lab

AR

I IU

MIS

TU

C-T

oohe

yS

UN

YA

SU

CE

HM

PI

AR

I II

U. T

okyo

UM

IST

Flig

htN

OA

AE

nv C

anU

C-J

imen

ezJu

lich

DO

E -

G1

NIE

SJC

AP

Uta

hU

MR

Wyo

mm

ing

Mai

nz S

ampl

ing

DO

E G

1S

AN

YU

JAR

IE

nvC

anII

Mai

nz-T

OF

UN

HN

OA

A P

ME

LC

U A

stro

Bio

TO

FA

RI T

OF

IIIN

OA

A-P

3

12x10-6

8

4

0

Ioni

zatio

n E

ffici

ency

2.0x10-12

1.5

1.0

0.5

0.0

IE/A

B

Cal

tech

BC

-Lab

AR

I IU

MIS

TU

C-T

oohe

yS

UN

YA

SU

CE

HM

PI

AR

I II

U. T

okyo

UM

IST

Flig

htN

OA

AE

nv C

anU

C-J

imen

ezJu

lich

DO

E -

G1

NIE

SJC

AP

Uta

hU

MR

Wyo

mm

ing

Mai

nz S

ampl

ing

DO

E G

1S

AN

YU

JAR

IE

nvC

anII

Mai

nz-T

OF

UN

HN

OA

A P

ME

LC

U A

stro

Bio

TO

FA

RI T

OF

IIIN

OA

A-P

3

Group

Average Value = 0.77(+/-0.30) x 10-12

AMS_PerformanceSummary

Page 23: Hardware Update - cires1.colorado.educires1.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · Cryopump Temp Test 1 - Pump from vent condition Cryopump on from start Test 2 - Cryopump

CH3O+ = 31

Page 24: Hardware Update - cires1.colorado.educires1.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · Cryopump Temp Test 1 - Pump from vent condition Cryopump on from start Test 2 - Cryopump

Converging nozzle to improve collection of micron size particles

A fabrication challenge…

•Two nozzles have been made (not quite to spec).• Measurement results not as expected, no improvement in large particle transmission

Page 25: Hardware Update - cires1.colorado.educires1.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · Cryopump Temp Test 1 - Pump from vent condition Cryopump on from start Test 2 - Cryopump

Adjustable Inlet with Micrometer heads