![Page 1: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/1.jpg)
MAC
HER
EY-N
AGEL
Handling interferences ̶ Experience report
Dr. Christian Prokisch, 11.08.2019Collaborative Efforts to Improve Environmental Monitoring
![Page 2: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/2.jpg)
Agenda
Company Nutrient Removal P‐Tests
N‐Tests
Tox Tests
Green(er) COD
![Page 3: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/3.jpg)
CompanyMN Water Analysis
![Page 4: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/4.jpg)
4
MN today
Company
• 4th Generation family owned• More than 650 employees• More than 25.000 products• Turnover 115 Mio. €
Subsidiaries Distributors
![Page 5: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/5.jpg)
5
Business units
Company
1911 1959 1961 1970 1989
![Page 6: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/6.jpg)
6
Get acceptance for COD tube tests
Company – MN Water AnalysisWhy are we here?
• Market share NANOCOLOR®
• Germany: ca. 40%• US: 0%
• Get experience with US regulations
![Page 7: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/7.jpg)
7
Company – MN Water Analysis
• NANOCOLOR® VIS II• Revolutionary user
experience with 10 inch HD touch screen display
• Intuitive, icon‐based menu guidance
• Integrated turbidity control
Outstanding photometers
![Page 8: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/8.jpg)
8
Company – MN Water Analysis
• NANOCOLOR® VARIO 4• Touch screen with intuitive
menu navigation• Lockable protective lids and
safety covers• Calibrated sensors to control
digestion temperature optional
Competitive heating blocks
![Page 9: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/9.jpg)
Nutrient removal
![Page 10: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/10.jpg)
10
Nutrient removal
• Nitrogen and phosphorus are the primary causes of cultural eutrophication
• Approximately 25% of all water body impairments are due to nutrient‐related causes
• Oxygen depletion• Algal growth• Ammonia
• More stringent effluent limits lead to • P‐removal• N‐removal
EPA Fact Sheet
![Page 11: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/11.jpg)
P‐Test
![Page 12: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/12.jpg)
12
Phosphate acc. to EPA 365.3
P‐Test
PO43‐ + 12 MoO4
2‐ +
24 H+ + 3 NH4+
(NH4)3[P(Mo3O10)4] +
12 H2O
Phosphor
molybdenum blue
Reduction
→→
![Page 13: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/13.jpg)
13
Dissolved ortho‐Phosphate
P‐Test
O ‐O – P – O‐
O‐
ortho‐Phosphat
![Page 14: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/14.jpg)
14
Total ortho‐phosphate
P‐Test
![Page 15: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/15.jpg)
15
Phosphate acc. to EPA 365.3
P‐Test
• For total phosphate, suspended solids contribute to total amount
• Make sure to draw a representative sample
• Carefully homogenize sample
![Page 16: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/16.jpg)
16
“Report results in mg/L P or mg/L PO4‐P”
P‐Test
Know the difference…
![Page 17: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/17.jpg)
17
Testing NO3‐ and PO4
3‐ in paralle
P‐Test
Danger of carry‐over
Phosphate test WITHOUT changing the pipette tip
Wrong results
Nitrate tests contain phosphoric acid
→ →
![Page 18: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/18.jpg)
18
When dealing with e.g. paper industry discharges
P‐Test
Be aware of possible silica interference
+ Silica > 150 mg/L
Phosphate test Wrong results
→
![Page 19: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/19.jpg)
N‐Test
![Page 20: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/20.jpg)
20
N‐Test
• Total N seems to be the target valuefor eutrophication
• Available EPA methods (examples)• NH3 acc. To EPA 350.1• Total Kjeldahl acc. to HACH 10242• NO3
‐ acc. to HACH 10206• NO2
‐ acc. to EPA 353.2• Not available EPA method
• Total N
Total N vs. Parameters 40 CFR Part 136
![Page 21: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/21.jpg)
21
Total N is already part of an approved test
N‐TestHow to detect Total N
TKN NO3‐Approved N‐methods:
TN NO3‐‐ + NO3
‐
org.N + NH3+ + NO3
‐What is total N?
What you then do
+
HACH TKN method: TKN = TN ‐ NO3‐
![Page 22: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/22.jpg)
22
Total N testing
N‐Test
Norg
NinorgNO3
‐ Determination
![Page 23: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/23.jpg)
23
Digestion
N‐Test
![Page 24: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/24.jpg)
24
Determination
N‐Test
JUST invert!
![Page 25: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/25.jpg)
25
Do not shake
N‐Test
• Sample may develop turbidities
![Page 26: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/26.jpg)
26
N‐Test
No!
Filtration possible
![Page 27: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/27.jpg)
27
Test for Chloride
N‐Test
Yes!• Chloride determination with QUANTOFIX© Chloride• Dilute• Chloride elimination cartridges
![Page 28: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/28.jpg)
Tox Tests Inhibitors for denitrifaction
![Page 29: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/29.jpg)
29
Processes
Tox Tests for denitrification
NH4+
NO2‐
NO3‐
N2
Ammonium Oxidation
Nitrite Oxidation
Denitrification (Reduction)
Inhibitors
A‐Tox
N‐Tox
![Page 30: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/30.jpg)
30
How Tox‐tests work
Tox Tests for denitrification
FeedOxygen
Nothing happens
Nothing happens
Feed
Oxygen
![Page 31: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/31.jpg)
31
Tox Tests for denitrificationControl and sample
Inhibitors
Control
Inhibited sample
ConsumptionControl: 4 O2
ConsumptionSample: 3 O2
Inhibition = ∆O2 : ConsumptionControl x 100 = [1 : 4] x 100 = 25 %
∆O2 = 1
![Page 32: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/32.jpg)
32
How the test is done
Tox Tests for denitrification
• Use oxygen probe• Measure
• Control• Sample
• Calculate inhibition from differenceSyringe
Oxygen Probe
Reaction vessel
![Page 33: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/33.jpg)
33
ISO 9509 vs. BioFix® Tox Tests
Tox Tests for denitrification
• Only for POTW with nutrient removal• Easily detects nitrification‐inhibitors• Lab test, result in 12 minutes
![Page 34: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/34.jpg)
Green(er) COD?
![Page 35: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/35.jpg)
35
Green(er) COD?
• Tube tests for chemical oxygendemand
• Principle• Cr2O7
2‐ + “Dirt” 2 Cr3+
• NANOCOLOR® COD HR 1500• NANOCOLOR® COD LR 150
Method summary
Cr2O72‐, 436 nm
e.g. COD 160
Cr3+, 620 nme.g. COD 1500
![Page 36: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/36.jpg)
36
Green(er) COD?
• Cr2O72‐ is actice reagent
• Test is based on oxidation potential• So far not replacable for real samples
• Hg2+ masks Chloride• allows use of Silver as catalyst• Prevents false to high results caused by Cl‐
‐ Oxidation
More detail
Ag+
Cr2O72‐ + “Dirt” + Cl‐ + Hg2+ Cr3+ +HgCl2
![Page 37: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/37.jpg)
Cr2O72‐ + “Dirt” 2 Cr3+
37
Green(er) COD?Mercury is used to overcome Chloride interference
Ag+
Cr2O72‐ + Cl‐ 2 Cr3+ (false too high result)
Ag+ + Cl‐ AgCl (turbidity)
Cr2O72‐ + 2 Cl‐ + Hg2+ Cr2O7
2‐ +HgCl2
Ag+ + 2Cl‐ + Hg2+ Ag+ + HgCl2
Chloride interference Part I
Chloride interference Part II
![Page 38: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/38.jpg)
38
Green(er) COD?
• POTW inflow• Expected “Chloride error” at 800 ppm
COD: approx. 3%• Most substances easy to crack
• POTW effluent• Expected “Chloride Error” at 40 ppm
COD: approx. 60%• Ag+ catalyst needed to crack
refractary substances
Hg‐free COD Part I: Chloride Compensation
Use test strip to measure Chloride and substract from result
Cr2O72‐ + Cl‐ + Dirt Cr3+Ag+
![Page 39: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/39.jpg)
39
Green(er) COD?
• POTW inflow• Filtration not suitable
because loss of particles
• POTW effluent• Filtration no problem (little
particles in effluent)• Ag+ catalyst available to
crack refractary substances
Hg‐free COD Part II: Chloride Removal
Cr2O72‐ + Cl‐ + Dirt Cr3+Ag+
Use cartridge to remove Chloride(involves some kind of filtration)
![Page 40: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/40.jpg)
• Chloride compensation
• Inflow: reasonably OK
• Outflow: useless
• Cl- teststrip would need improvement
• Chloride removal
• Inflow: useless
• Outflow: reasonably OK (expected)
• Handling would need improvement
40
Hg‐free COD could work – if the demand was there…
Green(er) COD?Result
![Page 41: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/41.jpg)
Summary
![Page 42: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/42.jpg)
42
Summary
• P‐Tests• Take care when Nitrate is tested at the
same time• N‐Test
• Total Nitrogen test is almost there• Tox‐Tests for denitrification
• Easy detection of inhibitors the reducePOTW performance
• Green(er) COD• Improved Mercury‐free test could be
available when needed
![Page 43: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/43.jpg)
MAC
HER
EY-N
AGEL
Thank you for your attention!
Dr. Christian Prokisch I cprokisch@mn‐net.com I +49‐2421‐969‐166
![Page 44: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian](https://reader035.fdocuments.us/reader035/viewer/2022062403/610ed415d2a3066c8400d569/html5/thumbnails/44.jpg)
Image credits©© Erwin Wodicka (2, 19‐27), SG‐design (10), Zauberhut (15), Darknightsky (20), Arnd Drifte (21) , oriori (25), sirirak (27), KaeslerMedia (29) , sirirak (40) / Fotolia