Gastro-Jejunal Feeding Tube Guide Wire for Infants Dr. Riad Rahhal, Director, Pediatric Gastroenterology Fellowship Program University of Iowa Children's Hospital Department of Pediatrics Arunan Arivalagan, Ahmad El-Hattab, Sean Ephraim, Ryan Hoeman Biomedical Engineering, University of Iowa, Iowa City, IA Toby Donjakowski, Medical Device Specialist

Transcript of guidewirefinalpresentationapr12-13106767798515-phpapp01-110714155500-phpapp01 (1)

Page 1: guidewirefinalpresentationapr12-13106767798515-phpapp01-110714155500-phpapp01 (1)

Gastro-Jejunal Feeding Tube Guide Wire for InfantsDr. Riad Rahhal, Director, Pediatric Gastroenterology Fellowship ProgramUniversity of Iowa Children's HospitalDepartment of Pediatrics

Arunan Arivalagan, Ahmad El-Hattab, Sean Ephraim, Ryan HoemanBiomedical Engineering, University of Iowa, Iowa City, IA

Toby Donjakowski, Medical Device Specialist

Page 2: guidewirefinalpresentationapr12-13106767798515-phpapp01-110714155500-phpapp01 (1)

Gastro-jejunal procedure

What is it?

Placement of feeding tube into the patient’s jejunum

Addresses infants that cannot swallow

Page 3: guidewirefinalpresentationapr12-13106767798515-phpapp01-110714155500-phpapp01 (1)

Gastro-jejunal procedure

Accommodations for infants:

Alternative to nasogastric/ nasoentric procedure

Page 4: guidewirefinalpresentationapr12-13106767798515-phpapp01-110714155500-phpapp01 (1)

Surgical procedure steps

1. Endoscope placement

Page 5: guidewirefinalpresentationapr12-13106767798515-phpapp01-110714155500-phpapp01 (1)

Surgical procedure steps

1. Endoscope placement

2. Guidewire placement

Page 6: guidewirefinalpresentationapr12-13106767798515-phpapp01-110714155500-phpapp01 (1)

Surgical procedure steps

1. Endoscope placement

2. Guidewire placement

3. Endoscope removal

Page 7: guidewirefinalpresentationapr12-13106767798515-phpapp01-110714155500-phpapp01 (1)

Surgical procedure steps

1. Endoscope placement

2. Guidewire placement

3. Endoscope removal4. Placement of

feeding tube and removal of guidewire

Page 8: guidewirefinalpresentationapr12-13106767798515-phpapp01-110714155500-phpapp01 (1)

Final result

Page 9: guidewirefinalpresentationapr12-13106767798515-phpapp01-110714155500-phpapp01 (1)

Kimberly-Clark feeding tube

Page 10: guidewirefinalpresentationapr12-13106767798515-phpapp01-110714155500-phpapp01 (1)

Savary-Gilliard guidewire

Page 11: guidewirefinalpresentationapr12-13106767798515-phpapp01-110714155500-phpapp01 (1)

Problems

Upon guidewire removal:

Uncoiling of tip

Page 12: guidewirefinalpresentationapr12-13106767798515-phpapp01-110714155500-phpapp01 (1)

Problems

Upon guidewire removal:

Uncoiling of tip

Damage to feeding tube valve

Page 13: guidewirefinalpresentationapr12-13106767798515-phpapp01-110714155500-phpapp01 (1)

Objectives

main objective:

design a guidewire for use with GJ feeding tube insertion

Page 14: guidewirefinalpresentationapr12-13106767798515-phpapp01-110714155500-phpapp01 (1)

ObjectivesMaintain

Flexibility Sterilizability Biocompatibility

Implement Better interaction with feeding tube Reusability Low cost

Page 15: guidewirefinalpresentationapr12-13106767798515-phpapp01-110714155500-phpapp01 (1)

Our solution

Small, constant diameter▪ Unlike Savary-

Gilliard guidewire

Page 16: guidewirefinalpresentationapr12-13106767798515-phpapp01-110714155500-phpapp01 (1)

Our solution

Small, constant diameter

Reinforcing brace▪ Demonstration vide

o

Page 17: guidewirefinalpresentationapr12-13106767798515-phpapp01-110714155500-phpapp01 (1)

Advantages

Constant diameter▪ Minimal valve/coil issues

Reinforcing brace▪ More tensile strength

Biocompatible▪ Nitinol, stainless steel

Reusable

prevent uncoilin

g

Page 18: guidewirefinalpresentationapr12-13106767798515-phpapp01-110714155500-phpapp01 (1)

Manufacturing modification

ORIGINAL MODIFIED

Page 19: guidewirefinalpresentationapr12-13106767798515-phpapp01-110714155500-phpapp01 (1)

Manufacturing modification

Page 20: guidewirefinalpresentationapr12-13106767798515-phpapp01-110714155500-phpapp01 (1)

Manufacturing modification

Page 21: guidewirefinalpresentationapr12-13106767798515-phpapp01-110714155500-phpapp01 (1)

Our Prototype

Specifications Stainless steel tail

and coil Nitinol brace Uniform diameter our

prototypeSavary-Gilliard

Page 22: guidewirefinalpresentationapr12-13106767798515-phpapp01-110714155500-phpapp01 (1)

Comparison

OUR PROTOTYPE

Body diameter 0.040 inches

Coil diameter 0.041 inches

Coil reinforcement Nitinol brace

SAVARY-GILLIARD

Body diameter 0.031 inches

Coil diameter 0.070 inches

No coil reinforcement

nearlyconstant

Page 23: guidewirefinalpresentationapr12-13106767798515-phpapp01-110714155500-phpapp01 (1)

Force Testing

Force vs. position▪ Testing video

Page 24: guidewirefinalpresentationapr12-13106767798515-phpapp01-110714155500-phpapp01 (1)

Test Results and Discussion

1 2 30.001.002.003.004.005.006.007.008.009.00

10.00Static Frictional Force

SavaryProtoype

Coil Position

Forc

e (N

)

Page 25: guidewirefinalpresentationapr12-13106767798515-phpapp01-110714155500-phpapp01 (1)

Looking Ahead

More testing Biocompatibility Paper work

Confidentiality agreements patenting FDA approval

Page 26: guidewirefinalpresentationapr12-13106767798515-phpapp01-110714155500-phpapp01 (1)