Group 3: Microfluidic System for Galvanotaxis Measurements Team Members: Arunan Skandarajah and Devin Henson Advisors: Dr. Janetopoulos, Dr. John Wikswo, and Dr. Paul King

Transcript of Group 3: Microfluidic System for Galvanotaxis Measurements Team Members: Arunan Skandarajah and...

Page 1: Group 3: Microfluidic System for Galvanotaxis Measurements Team Members: Arunan Skandarajah and Devin Henson Advisors: Dr. Janetopoulos, Dr. John Wikswo,

Group 3:Microfluidic System for

Galvanotaxis MeasurementsTeam Members:

Arunan Skandarajah and Devin HensonAdvisors:

Dr. Janetopoulos, Dr. John Wikswo,and Dr. Paul King

Page 2: Group 3: Microfluidic System for Galvanotaxis Measurements Team Members: Arunan Skandarajah and Devin Henson Advisors: Dr. Janetopoulos, Dr. John Wikswo,

Physiological Presence of Electrical Fields

• Normal development and maintenance processes result in endogenous electric fields1,2 – Established in wounds- ~.4-2 V/cm– Generated across gland and organ walls– Present during fetal organization

• Result from ion flow across barriers

Page 3: Group 3: Microfluidic System for Galvanotaxis Measurements Team Members: Arunan Skandarajah and Devin Henson Advisors: Dr. Janetopoulos, Dr. John Wikswo,

Cells Respond to Electric Fields• Endothelial cells in wounds can be

manipulated to speed or slow healing1

• Neuronal cells grow along electrically generated paths1

• Metastasizing cells will respond in the opposite manner of less malignant cells3

• Our model organism, Dictyostelium discoideum, also undergoes electrotaxis at voltages from 2-20V/cm4

Page 4: Group 3: Microfluidic System for Galvanotaxis Measurements Team Members: Arunan Skandarajah and Devin Henson Advisors: Dr. Janetopoulos, Dr. John Wikswo,

Problem StatementHow can we make experimentation easier than is currently possible?

Issues with current technology1. Difficult to fit bulky system on microscope stage2. Exposed and electrified liquid dangerous to experimenters3. Voltages as high as 500 V must be applied across device. 4. Excessive media and reagents required

Forrester, 2007 (2)

Page 5: Group 3: Microfluidic System for Galvanotaxis Measurements Team Members: Arunan Skandarajah and Devin Henson Advisors: Dr. Janetopoulos, Dr. John Wikswo,

Solution Process• Utilize Microfabrication to

– Reduce size of overall device to the scale of centimeters

• Allow for easy use on microscope stage• Reduce the required applied voltage to ~50 V by

reducing field size• Reduce the volume of cells and media required

– Create a closed design• Perfuse Media through Device

– Control of ion and pH gradients– Avoid problems with agar

• Eliminate voltage drop resulting from long bridges

Page 6: Group 3: Microfluidic System for Galvanotaxis Measurements Team Members: Arunan Skandarajah and Devin Henson Advisors: Dr. Janetopoulos, Dr. John Wikswo,

Microfluidic Design

Experimental Cell Area

Page 7: Group 3: Microfluidic System for Galvanotaxis Measurements Team Members: Arunan Skandarajah and Devin Henson Advisors: Dr. Janetopoulos, Dr. John Wikswo,

Resistance Distribution Calculations

• DB Buffer = 552.49 Ω*cm• Conductivity electrode meter from the Cliffel Lab in

SC5516

• 2% agar = 650 Ω*cm with considerable variation

• Voltage drop due to electrolysis of water prevents precise measurement

Page 8: Group 3: Microfluidic System for Galvanotaxis Measurements Team Members: Arunan Skandarajah and Devin Henson Advisors: Dr. Janetopoulos, Dr. John Wikswo,

Current Status• Microfabrication and AutoCAD training complete• Prototypes produced• Cell successfully cultured and seeded into devices• Cell motility captured, but stated rates of movement have

not been matched• Must establish baseline cell response to evaluate device

before proceeding furtherCells seeded in device, imaged at 32x using Zeiss Axiovert microscope and QImaging camera

Page 9: Group 3: Microfluidic System for Galvanotaxis Measurements Team Members: Arunan Skandarajah and Devin Henson Advisors: Dr. Janetopoulos, Dr. John Wikswo,

Current Status• Two macro-scale devices for establishing cell

response

1. Coverslip method- replica from literature

- easy to seed and clean

- does not resolve any targeted problems

2. Microfluidic- “halfway”

- lower voltage, better setup

- more difficult to clean and seed

PDMS

Mcrofluidic channel

Agar bridge access holes

glass slide

Coverslip roof

Agar bridge contact area

Page 10: Group 3: Microfluidic System for Galvanotaxis Measurements Team Members: Arunan Skandarajah and Devin Henson Advisors: Dr. Janetopoulos, Dr. John Wikswo,

Preliminary Cell Motility Data

• Cells are moving at approximately 1 μm/min• This value is 1/3 of that reported in literature4

• Poor gas exchange?

Page 11: Group 3: Microfluidic System for Galvanotaxis Measurements Team Members: Arunan Skandarajah and Devin Henson Advisors: Dr. Janetopoulos, Dr. John Wikswo,

Recent Developments

• Reached Zhao to discuss protocol for macro-scale device– Reduce chamber length for better gas exchange (1

cm and 5 mm lengths)– Possible visit to UC-Davis

• Received films for fabrication of microfluidic designs

• Need to confirm new microfluidic devices work similarly to previous iterations for controlling pH and ion gradients– Find optimal flow rates and cell seeding methods

Page 12: Group 3: Microfluidic System for Galvanotaxis Measurements Team Members: Arunan Skandarajah and Devin Henson Advisors: Dr. Janetopoulos, Dr. John Wikswo,

Immediate Tasks

• Try Zhao’s suggestions with shorter chamber lengths

• Characterize flow profile and pH gradients in new microfluidics

• Edit Kevin Seale’s code for our cell type and phase contrast imaging

• Understand baseline response of cells in electric field

• Select among prototypes for optimal microfluidic channel dimensions

Page 13: Group 3: Microfluidic System for Galvanotaxis Measurements Team Members: Arunan Skandarajah and Devin Henson Advisors: Dr. Janetopoulos, Dr. John Wikswo,

ImageJ Cell Tracking Software

• Requires 8-bit images– QCapture capable of capturing in 8-bit– Previously captured 12-bit images are

convertible to 8-bit no lost images

• Obtained code for finding and tracking cells (Kevin Seale, SyBBURE)

• Other macros for specific cell tracking downloadable online

Page 14: Group 3: Microfluidic System for Galvanotaxis Measurements Team Members: Arunan Skandarajah and Devin Henson Advisors: Dr. Janetopoulos, Dr. John Wikswo,

Future Direction

• Utilize optimal microfluidic design to replicate cell response from literature

• Improvements in using our device– Lower applied voltage necessary– Controlled pH and ion gradients without agar– Reduced overall device size and reagent

consumption

• Quantify data– Program ImageJ to obtain cell displacement– Find directedness and trajectory rates

Page 15: Group 3: Microfluidic System for Galvanotaxis Measurements Team Members: Arunan Skandarajah and Devin Henson Advisors: Dr. Janetopoulos, Dr. John Wikswo,

Works Cited

1. Nuccitelli, R., A role for endogenous electric fields in wound healing. Current Topics in Developmental Biology, 2003. 58: p. 1-26.

2. Forrester, J.V., Lois, N., Zhao, M., McCaig, C. The spark of life: the role of electric fields in regulating cell behaviour using the eye as a model system, Ophthal. Res. 39:4-16, 2007.

3. Pu, J., McCaig, C.D., Cao, L., Zhao, Z., Segall, J.E., Zhao, M. EGF receptor signaling is essential for electric-field-directed migration of breast cancer cells, J. Cell Sci. 120:3395-3403, 2007

4. Zhao, M., Jin, T., McCaig, C.D., Forrester, J.V., Devreotes, P.N. Genetic analysis of the role of G protein-coupled receptor signaling in electrotaxis, J. Cell Biol. 157:921-927, 2002.

.