GMC SECUTEST SIII+

58
1 Produktvorstellung – SECUTEST SIII+ Dieter Feulner Electrical safety Test for appliances Farhaan Ahmed

Transcript of GMC SECUTEST SIII+

Page 1: GMC SECUTEST SIII+

1Produktvorstellung – SECUTEST SIII+ Dieter Feulner

Electrical safety Test for appliances

Farhaan Ahmed

Page 2: GMC SECUTEST SIII+

2Produktvorstellung – SECUTEST SIII+ Dieter Feulner

Agenda

Secutest SIII General setup

Measurement

Software

Page 3: GMC SECUTEST SIII+

3Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+general

SECUTEST SIII

Page 4: GMC SECUTEST SIII+

4Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+general

Universal test instrument for testing the electrical safety of portable electrical equipment in commercial and medical applications.

In order to evaluate electrical safety, tests are performed for protective conductor connections, insulation resistance and leakage currents.

Page 5: GMC SECUTEST SIII+

5Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+general

Automatic adaptation to either 110 V / 60 Hz or 230 V / 50 Hz mains power

Automatic recognition of safety class (I, II) and mains connection errors

Menu-driven test sequences: fully automated or manual

Safety for the user thanks to integrated personal protection - suitable for persons with basic electro-technical instruction

Page 6: GMC SECUTEST SIII+

6Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+general

Convenient memory and report generating functions (optional)

Data interface for PC, printer and barcodes

Report generation with printer module (PSI module), DA-II printer adapter or PC software

User interface in various languages: English, French, German, Italian, Spanish, Czech, Dutch

Parallel test sockets for devices under test without mains plug

Page 7: GMC SECUTEST SIII+

7Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+general

Patient leakage current: AC and DC components measured separately

Equivalent leakage current measurement with 230 V test voltage

Function test with power analysis

Temperature measurement

Current measurement (with optional clip-on ammeter)

Voltage and resistance measurement

Page 8: GMC SECUTEST SIII+

8Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+general

Dimensions: 292 x 130 x 243 mm, weight: 4.0 e.q. 4.5 kg

Including automatic detection and indication of whether or not the probe is connected to PE

Including direct read-out of measurement data via the RS 232 interface (direct printing)

Including calibration certificate in accordance with DKD

Page 9: GMC SECUTEST SIII+

9Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+Tester schematic

Push-buttons (left and right) for releasingthe handle from its snap-in position

Test socket forconnecting the device under test

Push-buttons(left and right)for releasing the lid

Lid

Compartment for probe and accessories

Carrying handle and tilt stand

OptionalSECUTEST PSIalphanumeric keypad,printer & memory

OptionApplied parts for measuringMedical devices

Optional service socket for connecting a PC or printer

Optional additional unit for high voltage test

Page 10: GMC SECUTEST SIII+

10Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ front plate

Jacks for Live / Neutral /PE conductor

Jacks for connectingthe probe

Function selectorswitch

Scroll key formenu and parameterselection

LCD-displaygraphical withbacklight

Signal lamp for mainsconnection error

Key for startingtest sequence

Help key

Key switchingtest voltage to the test soecket

Signal lamp forthe function test

Jacks forapplied parts

Page 11: GMC SECUTEST SIII+

11Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ rotary switch

Page 12: GMC SECUTEST SIII+

12Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ Principle work with cursor

Page 13: GMC SECUTEST SIII+

13Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ Setup

General Setup

Page 14: GMC SECUTEST SIII+

14Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ Setup

Save the settings

Change the limits

Make settings for all test sequences (standard’s)

Look at the stored test result

Information about the software of the tester

Page 15: GMC SECUTEST SIII+

15Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ Limits

Set backgound ligth time

Duration for each test in a sequence

Nominal Voltage for leakage current calculation

Test of short circuit in mains

Wait after power on to start measurement

Do not ask any user input after starting the test

Page 16: GMC SECUTEST SIII+

16Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ Test Sequence

Test stops if measurement is out of limit range

For PSI only

Intrinsic error of tester is included in test result

Warning if IT system is present

Internal beeper

Detectiong of PE connection

Measured value is printed via RS232 immediately

Page 17: GMC SECUTEST SIII+

17Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ Protocols

Save the test result in a protocol

Print the test result via RS232

Show all stored protocols(and select one)

Print all stored protocols via RS232

Delete all stored protocols

Page 18: GMC SECUTEST SIII+

18Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ Service

Information about internal Software of the tester

Release of Software, Customer

Set time and date

Page 19: GMC SECUTEST SIII+

19Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ AUX

Measurements

Page 20: GMC SECUTEST SIII+

20Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ R-PE

Test current can be set to 200 mA and (10) 25 A

Page 21: GMC SECUTEST SIII+

21Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ ISO / HV passive leakage

Measuring the insulation resistance

Measuring the protective leakage current with an alternative method

For Medical equipment only

High voltage – Insulation test (with DC voltage)

Page 22: GMC SECUTEST SIII+

22Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ active Leakage current

Protective earth current

Touch current (current from accessible parts to ground – measured with the probe)

For medical equipment only

Alternative measurement for protective earth current

For medical devices only (overall leakage current)

Page 23: GMC SECUTEST SIII+

23Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ Function Test

Power supply of device is measured

Page 24: GMC SECUTEST SIII+

24Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ Multimeter

Using the Tester like a Multimeter

Voltage measurement with the probe (against PE)

Voltage measurement withtest leads

(Low) Resistance measurements

Page 25: GMC SECUTEST SIII+

25Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ AUX

Measurements with auxiliary

Current measurement with a clamp

Protective earth with a clamp

Temperature measurement with PT100 or PT 1000

Page 26: GMC SECUTEST SIII+

26Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ AUTO

Auto mode

Predefined test sequences

Page 27: GMC SECUTEST SIII+

27Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ AUTO

Test Sequence IEC 60335

For routine test

Page 28: GMC SECUTEST SIII+

28Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ AUTO 60335 START

Select :

standard

Connection of the device

Type of device

ID-Numer of device

Change settings for this test sequence

Page 29: GMC SECUTEST SIII+

29Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ AUTO 60335 SETUP

Select :

standard

Connection of the device

Type of device

ID-Numer of device

Change settings for this test sequence

Page 30: GMC SECUTEST SIII+

30Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ AUTO 60335 CONFIGURE

Select :

standard

Connection of the device

Type of device

ID-Numer of device

Change settings for this test sequence

Page 31: GMC SECUTEST SIII+

31Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ AUTO 60335 PE

Select :

standard

Connection of the device

Type of device

ID-Numer of device

Change settings for this test sequence

Page 32: GMC SECUTEST SIII+

32Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ AUTO 60335 HV

Select :

standard

Connection of the device

Type of device

ID-Numer of device

Change settings for this test sequence

Page 33: GMC SECUTEST SIII+

33Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ AUTO 60335 Leakage Current

Measurement (type of leakage current)

Measurement Value

Test Voltage

Measurment conditions

Page 34: GMC SECUTEST SIII+

34Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ AUTO WIK Start screen

The test starts automatically if device is switched on

Push return button to change the Type

Page 35: GMC SECUTEST SIII+

35Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ AUTO WIK 60335

The test starts automatically if device is switched on

Push return button to change the Type

Page 36: GMC SECUTEST SIII+

36Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ AUTO WIK 60335 SETUP

The test starts automatically if device is switched on

Push return button to change the Type

Page 37: GMC SECUTEST SIII+

37Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ AUTO WIK 60335 Configure

The test starts automatically if device is switched on

Push return button to change the Type

Page 38: GMC SECUTEST SIII+

38Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ AUTO WIK ground continuity

The test starts automatically if device is switched on

Push return button to change the Type

Page 39: GMC SECUTEST SIII+

39Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ AUTO WIK HV

The test starts automatically if device is switched on

Push return button to change the Type

Page 40: GMC SECUTEST SIII+

40Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ AUTO WIK LEAKAGE Current

The test starts automatically if device is switched on

Push return button to change the Type

Page 41: GMC SECUTEST SIII+

41Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ Service and Repair

Test Sequence DIN VDE 0701

For Maintenance and Repair

Page 42: GMC SECUTEST SIII+

42Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ AUTO 0701

Select :

standard

Connection of the device

Type of device

ID-Numer of device

Change settings for this test sequence

Page 43: GMC SECUTEST SIII+

43Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ AUTO 0701 visual

Select :

standard

Connection of the device

Type of device

ID-Numer of device

Change settings for this test sequence

Page 44: GMC SECUTEST SIII+

44Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ AUTO 0701 PE

Select :

standard

Connection of the device

Type of device

ID-Numer of device

Change settings for this test sequence

Page 45: GMC SECUTEST SIII+

45Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ AUTO 0701 RISO

Select :

standard

Connection of the device

Type of device

ID-Numer of device

Change settings for this test sequence

Page 46: GMC SECUTEST SIII+

46Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ AUTO 0701 Leakage

Leakage current active or passive, depending on type of device

Devices with elctronic switches shall be measured active in both directions of mains supply (Differential + Touch current)

Page 47: GMC SECUTEST SIII+

47Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ AUTO 0701 result

Select :

standard

Connection of the device

Type of device

ID-Numer of device

Change settings for this test sequence

Page 48: GMC SECUTEST SIII+

48Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ AUTO 0701 Function Test

Select :

standard

Connection of the device

Type of device

ID-Numer of device

Change settings for this test sequence

Page 49: GMC SECUTEST SIII+

49Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ Testing of 3-phase devices

For three-phase devices use the AT3-III:

• fully automatic test sequence

• high voltage test (optional)

• residual current measurement

Page 50: GMC SECUTEST SIII+

50Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+

Software

Page 51: GMC SECUTEST SIII+

51Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ SECUTEST (P)SI

Enter customer information.

Enter repair informationenter further informaton of the appliance.

PRINT STOREor

Activates the „PSI-menue“

Page 52: GMC SECUTEST SIII+

52Produktvorstellung – SECUTEST SIII+ Dieter Feulner

internal Test Report using free software Z853

Page 53: GMC SECUTEST SIII+

53Produktvorstellung – SECUTEST SIII+ Dieter Feulner

internal Test Report using free software Z853

Page 54: GMC SECUTEST SIII+

54Produktvorstellung – SECUTEST SIII+ Dieter Feulner

Print direct

Page 55: GMC SECUTEST SIII+

55Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ Software and documentation

Test ProtocolTesting of the Electrical Safety according to IEC601.1

Order number: Customer:aaaaAenggleisfff95081 erlangen

Contractor:GMC-Instruments

Device: defibrilatorccc Manufacturer:draegerType: oem Protection class: Inventory no.: aaa Isolated applied parts: Type Test according to IEC 601.1 Date of test: 01.02.01Visual inspection:[ ] Protective conductor (only for class I equipment)[ ] Casing and mechanical parts[ ] Insulating parts[ ] Terminal connectors including plugs[ ] Markings[ ] OtherMeasurement Meas. value Limit value OK

Protective conductor continuity +0.063Ohm <0.200OhmInsulation resistance >+310.0MOh >2.000MOhmDifferential currentEarth leakage current +0.016mA <0.500mAEarth leakage current single fault condition +0.037mA <1.000mAHousing leakage current +000.5µA <100.0µAHousing leakage current single fault condition +022.8µA <0.500mAPatient leakage current AC +000.4µA <0.100mAPatient leakage current AC single fault condition +001.2µA <0.500mAPatient current DC +000.2µA <10.00µAPatient current DC single fault condition +000.1µA <050.0µAPatient current with mains on applied parts +000.4µA <05.00mAAuxiliary patient current ACAuxiliary patient current AC single fault conditionAuxiliary patient current DCAuxiliary patient current DC single fault conditionDielectric strength test [ ]Functional test[ ] Functional and safety testDate of next test : Test equipment used Type:

SECUTEST 0751/601Manufacturer:GOSSEN-METRAWATT

Signature:Testing person:

_____________________________________________________Place: Date 10.03.02

Data-management andReport generation

with MS Office

PC.doc-word

Page 56: GMC SECUTEST SIII+

56Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ Software and documentation

Maintenance and servicemanagement program

PS3

Page 57: GMC SECUTEST SIII+

57Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+ summary

•Easy to carry (lightwight, compact)

•fully automatically test sequence, even for 3-phase devices

•excellent price-performance ratio by customer-defined configuration

•Easy to handle, online-help•Leakage current AC and DC•High voltage test up to 6kV (DC)•0,2; 10 or 25 Ampere test current•Function test with power measurement•Internal memory up to 125 protocols

Page 58: GMC SECUTEST SIII+

58Produktvorstellung – SECUTEST SIII+ Dieter Feulner

SECUTEST SIII|+

Thanks