George Segall, M.D.
VA Palo AltoHealth Care System
Stanford University
PET Myocardial Perfusion Imaging: Protocols and
Procedures
![Page 2: George Segall, M.D. Stanford University PET Myocardial Perfusion Imaging: Protocols and Procedures.](https://reader035.fdocuments.us/reader035/viewer/2022081505/551c0957550346ad4f8b501b/html5/thumbnails/2.jpg)
PET Perfusion Agents
Rubidium-82 (82Rb) Chloride 75 sec 3.15
Nitrogen-13 (13N) Ammonia 10 min1.19
O-15 (15O) Water 110 sec1.72
Copper-62 (62Cu) PTSM 9.8 min2.94
Potassium-38 (38K) Chloride 7.6 min 2.70
Carbon-11 (11C) Butanol 20 min0.96
Radioisotope Form Half-life MeV
Diffusible tracers
PTSM, pyruvaldehyde methylthiosemicarbazone
![Page 3: George Segall, M.D. Stanford University PET Myocardial Perfusion Imaging: Protocols and Procedures.](https://reader035.fdocuments.us/reader035/viewer/2022081505/551c0957550346ad4f8b501b/html5/thumbnails/3.jpg)
Medicare Coverage Policy
1995Rb-82 PET for myocardial perfusion
1998F-18 deoxyglucose PET and SPECT for myocardial
viability
2003N-13 ammonia for myocardial perfusion
![Page 4: George Segall, M.D. Stanford University PET Myocardial Perfusion Imaging: Protocols and Procedures.](https://reader035.fdocuments.us/reader035/viewer/2022081505/551c0957550346ad4f8b501b/html5/thumbnails/4.jpg)
Medicare Coverage Policy for Rb-82 and N-13 ammonia
PET whether at rest alone, or rest with stress, is approved when PET is
• performed in place of SPECT, or
• used following SPECT when results are equivocal, technically uninterpretable, or discordant with clinical care
SPECT may not be used following an inconclusive PET
![Page 5: George Segall, M.D. Stanford University PET Myocardial Perfusion Imaging: Protocols and Procedures.](https://reader035.fdocuments.us/reader035/viewer/2022081505/551c0957550346ad4f8b501b/html5/thumbnails/5.jpg)
RAM License
Strontium-82 200 mCi
Rubidium-82 200 mCi
Strontium-85 1000 mCi
![Page 6: George Segall, M.D. Stanford University PET Myocardial Perfusion Imaging: Protocols and Procedures.](https://reader035.fdocuments.us/reader035/viewer/2022081505/551c0957550346ad4f8b501b/html5/thumbnails/6.jpg)
Rb-82 Infusion System
![Page 7: George Segall, M.D. Stanford University PET Myocardial Perfusion Imaging: Protocols and Procedures.](https://reader035.fdocuments.us/reader035/viewer/2022081505/551c0957550346ad4f8b501b/html5/thumbnails/7.jpg)
Relative Advantages
Rb-82 N-13 ammonia
Availability Excellent Limited
Logistics Easy Difficult
Cost Fixed Variable
![Page 8: George Segall, M.D. Stanford University PET Myocardial Perfusion Imaging: Protocols and Procedures.](https://reader035.fdocuments.us/reader035/viewer/2022081505/551c0957550346ad4f8b501b/html5/thumbnails/8.jpg)
Decision Analysis
Occasional Imaging and
On-site cyclotron
N-13 ammonia
Daily Imaging and
PET/CT capacity
Rb-82 chloride
![Page 9: George Segall, M.D. Stanford University PET Myocardial Perfusion Imaging: Protocols and Procedures.](https://reader035.fdocuments.us/reader035/viewer/2022081505/551c0957550346ad4f8b501b/html5/thumbnails/9.jpg)
Decision Analysis
Flexibility Exercise
N-13 ammoniaAdenosine
DipyridamoleDobutamine
Convenience Adenosine
Rb-82 chloride
Dipyridamole
Regadenoson
Regadenoson
![Page 10: George Segall, M.D. Stanford University PET Myocardial Perfusion Imaging: Protocols and Procedures.](https://reader035.fdocuments.us/reader035/viewer/2022081505/551c0957550346ad4f8b501b/html5/thumbnails/10.jpg)
Scheduling
N-13 ammonia
scheduled > 1D in advance
• problematic cases
inconclusive SPECT
high BMI
Rb-82 chloride
scheduled day of exam
• ED and inpatients
√ √
√
![Page 11: George Segall, M.D. Stanford University PET Myocardial Perfusion Imaging: Protocols and Procedures.](https://reader035.fdocuments.us/reader035/viewer/2022081505/551c0957550346ad4f8b501b/html5/thumbnails/11.jpg)
Emission protocol
N-13 ammonia
Rb-82 chloride
• 20-30 mCi
• 2D gated emission
• start 10 min p injection
• 40-60 mCi
• 2D gated emission
• start 90 sec p injection
![Page 12: George Segall, M.D. Stanford University PET Myocardial Perfusion Imaging: Protocols and Procedures.](https://reader035.fdocuments.us/reader035/viewer/2022081505/551c0957550346ad4f8b501b/html5/thumbnails/12.jpg)
2D vs 3D Gated Rb-82 PETKnesaurek. BMC Nucl Med 2007;7:4
16 normal volunteers had resting Rb-82 on a BGO PET camera
Low dose (20 mCi) 3D acquisition compared to high dose (60 mCi) 2D acquisition
![Page 13: George Segall, M.D. Stanford University PET Myocardial Perfusion Imaging: Protocols and Procedures.](https://reader035.fdocuments.us/reader035/viewer/2022081505/551c0957550346ad4f8b501b/html5/thumbnails/13.jpg)
Emission Reconstruction
Rb-82 chloride
N-13 ammonia
• reconstruction filter: Butterworth 0.4 roll-off / power 10
• post reconstruction: axial smoothing
• reconstruction filter: Butterworth 0.40 roll-off / power 10
• no post reconstruction smoothing
![Page 14: George Segall, M.D. Stanford University PET Myocardial Perfusion Imaging: Protocols and Procedures.](https://reader035.fdocuments.us/reader035/viewer/2022081505/551c0957550346ad4f8b501b/html5/thumbnails/14.jpg)
CT protocol
• 140 keV
• 30-210 mAs (noise index 25.0)
• pitch 0.5
• ungated
![Page 15: George Segall, M.D. Stanford University PET Myocardial Perfusion Imaging: Protocols and Procedures.](https://reader035.fdocuments.us/reader035/viewer/2022081505/551c0957550346ad4f8b501b/html5/thumbnails/15.jpg)
PET Perfusion Agents
•Rb-82 Chloride and N-13 ammonia are both extractable agents which passively diffuse across cell membranes
•Both tracers have high first-pass myocardial extraction fractions, which are inversely related to flow is a non-linear manner.
![Page 16: George Segall, M.D. Stanford University PET Myocardial Perfusion Imaging: Protocols and Procedures.](https://reader035.fdocuments.us/reader035/viewer/2022081505/551c0957550346ad4f8b501b/html5/thumbnails/16.jpg)
Rb-82 chloride
•Rb-82 is rapidly cleared from the circulation
Tang. Phys Med Biol 2009;54:3161-3171
![Page 17: George Segall, M.D. Stanford University PET Myocardial Perfusion Imaging: Protocols and Procedures.](https://reader035.fdocuments.us/reader035/viewer/2022081505/551c0957550346ad4f8b501b/html5/thumbnails/17.jpg)
N-13 ammonia
•N13 is rapidly cleared from the circulation
![Page 18: George Segall, M.D. Stanford University PET Myocardial Perfusion Imaging: Protocols and Procedures.](https://reader035.fdocuments.us/reader035/viewer/2022081505/551c0957550346ad4f8b501b/html5/thumbnails/18.jpg)
N-13 ammonia•N-13 ammonia is trapped in the myocardium from the
action of glutamate dehydrogenase1 and glutamine synthetase2
13NH313NH4
+
glutamineurea
-ketoglutarate
glutamate
NH3
glutamine1 2
13NH3
![Page 19: George Segall, M.D. Stanford University PET Myocardial Perfusion Imaging: Protocols and Procedures.](https://reader035.fdocuments.us/reader035/viewer/2022081505/551c0957550346ad4f8b501b/html5/thumbnails/19.jpg)
PET Perfusion Agents
Rb-82 Chloride N-13 ammonia
Relationship to CBF
Schelbert Hr, Schwaiger M. In: PET and Autoradiography. Eds: Phelps ME, Mazziotta JC, Schelbert H 1986
Bergman SR. In: PET of the Heart. Eds: Bergmann SR, Sobel BE. 1992
![Page 20: George Segall, M.D. Stanford University PET Myocardial Perfusion Imaging: Protocols and Procedures.](https://reader035.fdocuments.us/reader035/viewer/2022081505/551c0957550346ad4f8b501b/html5/thumbnails/20.jpg)
Work Flow
0 5 10 15 20 25 30 35 40 45 50 55 60
inje
ct
scan
Pers
inje
ct
scan
inje
ct scan
Persin
ject sca
n
N-13 ammonia20-30 mCi
Rb-82 40-60 mCi
min
CT
CT
![Page 21: George Segall, M.D. Stanford University PET Myocardial Perfusion Imaging: Protocols and Procedures.](https://reader035.fdocuments.us/reader035/viewer/2022081505/551c0957550346ad4f8b501b/html5/thumbnails/21.jpg)
Rb-82 chloride
0 5 10 15 20 25 30 35 40 45 50 55 60
inje
ctscan
Pers
inje
ct
scan
inje
ct
min
CT
Aden
inje
ct
scan
scan
CT
![Page 22: George Segall, M.D. Stanford University PET Myocardial Perfusion Imaging: Protocols and Procedures.](https://reader035.fdocuments.us/reader035/viewer/2022081505/551c0957550346ad4f8b501b/html5/thumbnails/22.jpg)
Rb-82 chloride
0 5 10 15 20 25 30 35 40 45 50 55 60
inje
ctscan
Pers
inje
ct
scan
inje
ct
min
CT
Reg
inje
ct
scan
scan
CT
![Page 23: George Segall, M.D. Stanford University PET Myocardial Perfusion Imaging: Protocols and Procedures.](https://reader035.fdocuments.us/reader035/viewer/2022081505/551c0957550346ad4f8b501b/html5/thumbnails/23.jpg)
N-13 ammonia
0 5 10 15 20 25 30 35 40 45 50 55 60
inje
ct
scan
Pers
min
CT inje
ct
scan
scan
CT
60 65 70 75 80 85 90
Ex scan
CT
0 5 10 15 20 25
inje
ct
inje
ct
min•••
![Page 24: George Segall, M.D. Stanford University PET Myocardial Perfusion Imaging: Protocols and Procedures.](https://reader035.fdocuments.us/reader035/viewer/2022081505/551c0957550346ad4f8b501b/html5/thumbnails/24.jpg)
Cyclotron starts production 23 minutes prior to calibration time
Injection times
0 5 10 15 20 25 30 35 40 45 50 55 60
Ex
RegAd
Per
Dob
= inject
![Page 25: George Segall, M.D. Stanford University PET Myocardial Perfusion Imaging: Protocols and Procedures.](https://reader035.fdocuments.us/reader035/viewer/2022081505/551c0957550346ad4f8b501b/html5/thumbnails/25.jpg)
Nor
mal
Rb-
82 p
erfu
sion
stu
dy
![Page 26: George Segall, M.D. Stanford University PET Myocardial Perfusion Imaging: Protocols and Procedures.](https://reader035.fdocuments.us/reader035/viewer/2022081505/551c0957550346ad4f8b501b/html5/thumbnails/26.jpg)
Nor
mal
Rb-
82 p
erfu
sion
stu
dy
![Page 27: George Segall, M.D. Stanford University PET Myocardial Perfusion Imaging: Protocols and Procedures.](https://reader035.fdocuments.us/reader035/viewer/2022081505/551c0957550346ad4f8b501b/html5/thumbnails/27.jpg)
Nor
mal
Rb-
82 p
erfu
sion
stu
dy
![Page 28: George Segall, M.D. Stanford University PET Myocardial Perfusion Imaging: Protocols and Procedures.](https://reader035.fdocuments.us/reader035/viewer/2022081505/551c0957550346ad4f8b501b/html5/thumbnails/28.jpg)
Nor
mal
Rb-
82 p
erfu
sion
stu
dy
![Page 29: George Segall, M.D. Stanford University PET Myocardial Perfusion Imaging: Protocols and Procedures.](https://reader035.fdocuments.us/reader035/viewer/2022081505/551c0957550346ad4f8b501b/html5/thumbnails/29.jpg)
Abn
orm
al R
b-82
per
fusi
on s
tudy
![Page 30: George Segall, M.D. Stanford University PET Myocardial Perfusion Imaging: Protocols and Procedures.](https://reader035.fdocuments.us/reader035/viewer/2022081505/551c0957550346ad4f8b501b/html5/thumbnails/30.jpg)
Abn
orm
al R
b-82
per
fusi
on s
tudy
![Page 31: George Segall, M.D. Stanford University PET Myocardial Perfusion Imaging: Protocols and Procedures.](https://reader035.fdocuments.us/reader035/viewer/2022081505/551c0957550346ad4f8b501b/html5/thumbnails/31.jpg)
Nor
mal
N-1
3 pe
rfus
ion
stud
y
![Page 32: George Segall, M.D. Stanford University PET Myocardial Perfusion Imaging: Protocols and Procedures.](https://reader035.fdocuments.us/reader035/viewer/2022081505/551c0957550346ad4f8b501b/html5/thumbnails/32.jpg)
Nor
mal
N-1
3 pe
rfus
ion
stud
y
![Page 33: George Segall, M.D. Stanford University PET Myocardial Perfusion Imaging: Protocols and Procedures.](https://reader035.fdocuments.us/reader035/viewer/2022081505/551c0957550346ad4f8b501b/html5/thumbnails/33.jpg)
Abn
orm
al N
-13
perf
usio
n st
udy
72 year old man with peripheral vascular disease.
Coronary arteriography:•LAD: 60% ostial and 80% mid vessel stenoses•LCX: 90% proximal stenosis and occluded OM•RCA: 90% ostial stenosis
![Page 34: George Segall, M.D. Stanford University PET Myocardial Perfusion Imaging: Protocols and Procedures.](https://reader035.fdocuments.us/reader035/viewer/2022081505/551c0957550346ad4f8b501b/html5/thumbnails/34.jpg)
Summary
• PET myocardial perfusion protocols are simple and easy to incorporate in most practices
• Rb-82 and N-13 ammonia are qualitatively equivalent
![Page 35: George Segall, M.D. Stanford University PET Myocardial Perfusion Imaging: Protocols and Procedures.](https://reader035.fdocuments.us/reader035/viewer/2022081505/551c0957550346ad4f8b501b/html5/thumbnails/35.jpg)
Summary
• High throughput, and low cost per patient makes Rb-82 PET practical in a high volume practice
• NH3-ammonia PET is better suited for occasional use in patients with inconclusive SPECT, or as the test of choice in problematic patients
![Page 36: George Segall, M.D. Stanford University PET Myocardial Perfusion Imaging: Protocols and Procedures.](https://reader035.fdocuments.us/reader035/viewer/2022081505/551c0957550346ad4f8b501b/html5/thumbnails/36.jpg)