GC sem1 2014
-
Upload
abre-groenewald -
Category
Documents
-
view
12 -
download
0
description
Transcript of GC sem1 2014
MSL935004A Maintain instruments and equipment
Gas Chromatograph
Contents: Quiz on last lesson
GC and its components -Review of basic Cleaning and maintenance of main components Trouble shootings Calculations involved in GC
Basic review of GC-8A Shimadzu
What are Applications of GC?
• separation volatile and thermally stable organic compounds
• quantitative & qualitative
• Operates at high temperature usually 5-10 oC higher than highest bp of mix
Important to know that:
Carrier Gas (Mobile phase): • Essential in moving analyte through column • High Ultra Purity (less then 10ppm impurities)
• Restricted choice
• Choice of N2 , Ar, He, Air, H2,
• Storage • Inert tubing
Comparison of gases
Gas Application Comments
He General carrier or make-up gas Excellent but expensive Can not use with ECD unless with make up gas
N2 General make-up gas Carrier gas (packed column)
Cheap Not good for capillary as carrier gas (gives long run times)
H2 Carrier gas (capillary columns) Combustion gas (FID, NPD & FPD)
Highly explosive (oven trip sensor & careful venting) Best carrier gas (capillary)
Air Combustion gas (FID, NPD & FPD) Cheap and readily available
Injector: • Via micro-syringe • Sample not large for optimum column efficiency • Slow Injection of large sample causes band
broadening (loss of resolution) • Temperature should be about 50°C higher than the
boiling point of the least volatile component of the sample
Packed -Inert solid supported material (glass or stainless steel coil of 1-5m length and 5mm diameter) filled with stationary phase or coated with liquid stationary phase. Carbowax- a brand name for polyethylene glycol
Column (Stationary phase):
Capillary-
stationary phase is coated on inert solid tube Common polymers use as stationary phases are polysilioxanes
• Polar or Non-polar stationary phase
• Temp must be controlled
Detector: Detector Selectivity Detectability
Flame ionization (FID)
Most organic compounds 100 pg
Thermal conductivity (TCD)
Universal 1 ng
Electron capture (ECD)
Halides, nitrates, nitriles, peroxides,
anhydrides, organometallics
50 fg
Nitrogen-phosphorus Nitrogen, phosphorus
10 pg
Flame photometric (FPD)
Sulphur, phosphorus, tin, boron, arsenic,
germanium, selenium, chromium
100 pg
Photo-ionization (PID)
Aliphatics, aromatics, ketones, esters,
aldehydes, amines, heterocyclics,
organosulphurs, some organometallics
2 pg
Maintenance of GC
Preventive maintenance of your instruments will results in peak performance. Frequency of maintenance schedule will depends on how often you are using the instrument, if a GC is running multiple samples in an 8.0 hrs day, then you would be required to perform:
Daily: check of leakage, performance of standards, run calibration curve and gas pressure.
Weekly: check gas pressure cylinders, replace if necessary.
Change septum, injector liner, perform manual leak test.
Monthly: Check gas leak from primary gas supply to GC.
Otherwise depending on how many samples are tested/day, how clean the samples are? Condition of syringe (auto sampler), environmental factors (temperature) will need to be considered to allow you in adopting a suitable schedule.
Septa • High temp. septa are recommended to prevent injector
and column contamination. • Septum should be changed 30-50 manual injection and
75-100 for auto sampler injection. • If septa leak, air will get inside column and it will rapidly
damage the column. • When changing septa, column temp. should be reduced
to 35oC • Sign of leakage or bleeding septa (ghost peak or poor
reproducibility.
Injection Technique:
• Sample injection should be smooth and rapid, with quick removal of the syringe after injection, in order to avoid peak broadening.
• Injection volume is major issue, auto sampler is the solution. In manual injection try to rinsed and flushing the syringe 6 times.
Column • Check for a sample problem by injecting a reference standard. • If you get a good chromatogram, the problem most likely is sample related. If the chromatogram is not satisfactory, the problem probably is column or instrument related. • Check for a column problem by replacing the column with a duplicate column, one known to provide good results under proper conditions. If results are good, the problem is related to the original column. If the symptom persists, the problem is related to the instrument. • Incorrect installation of column is one source of leakage • It is important to turn oven off and wait for 10-15 minute to cool,
then turn carrier gas off (to prevent oxidation of column) before changing or installing a column.
A dirty detector creates noisy baselines.
• If the TCD is displaying problems such as a wandering baseline, increased noise level, or changes in response on a checkout chromatogram, it is probably contaminated with deposits from such things as column bleed or dirty samples.
• The TCD is cleaned by a process known as bakeout. Bakeout should be performed only after you have confirmed that the carrier gas and the flow system
components are leak and contaminant free.
• must turn off the TCD and cap the detector column fitting to prevent irreparable damage to the filament caused by oxygen entering the detector.
1. Turn the detector off.
2. Remove the column from the detector and cap the detector column fitting.
3. Set the reference gas flow rate between 20 and 30 mL/min. Set the detector temperature to 400oC.
4. Allow thermal cleaning to continue for several hours. Then cool the system to normal operating temperatures.
Trouble shootings
Please see GC resource on Moodle
What does tM represent?
The time between sample injection and an analyte peak reaching a detector at the end
of the column is termed the retention time, tR
Basic Definitions & Calculations
Term
tm
tr
t’r
Capacity factor, k’
Separation factor,
Definition/equation
-Retention time of unrestrained peak
-Retention time a component
-Adjusted retention time t’r = tr - tm
𝑘′ = 𝑡′
𝑟
𝑡𝑚
=𝑡𝑟−𝑡𝑚
𝑡𝑚
( ideal 1-5)
α= k ‘2 / k ‘1
Term
Resolution, R
Definition/equation
Measure of separation between two peaks:
𝑅𝑠 =2(𝑡𝑟2 − 𝑡𝑟1)
𝑊1 + 𝑊2
𝑊1 = 𝑏𝑎𝑠𝑒𝑤𝑖𝑑𝑡ℎ 𝑝𝑒𝑎𝑘 1
W2= 𝑏𝑎𝑠𝑒𝑤𝑖𝑑𝑡ℎ 𝑝𝑒𝑎𝑘 2
Term
Peak Asymmetry factor, AF
Number of theoretical plate, N
Height Equivalent to a Theoretical Plate,
Definition/equation
𝐴𝐹 =𝐵
𝐴𝑎𝑡 𝑊𝑎. 1
Where A = leading peak halfwidth
B= tailing peak halfwidth
N = 5.45 (tr/W1/2)2
where w1/2 is the peak width at half-height
HETP=L/N
where L is the length of the column
Sensitivity:
Slope of calibration plot (external stand)
LOD= 3 𝑥 𝑁𝑜𝑖𝑠𝑒 𝑝∓𝑝
𝑠𝑒𝑛𝑠𝑖𝑡𝑖𝑣𝑖𝑡𝑦
LOQ = 10 𝑥 𝑁𝑜𝑖𝑠𝑒 𝑝∓𝑝
𝑠𝑒𝑛𝑠𝑖𝑡𝑖𝑣𝑖𝑡𝑦
System Suitability Test
The following criteria should be used to determine system suitability
Injection of standard solution in 6 replicate
Relative standard deviation in peak area and retention time should be evaluated
RSD should be Peak Area (2%) and retention time (1%)