With Tent Screw, primary vertical and horizontal formation is possible without placing the implantBy utilizing the Cover Screw, fixation of membrane to the Tent Screw is done with easeSecondary stabilization of the membrane is completed with Fixing ScrewWith Tent Screw space for bone formation on the vast defect area can be maintainedBone formation utilizing Spacer, Membrane and Cover Screw after placing Implant
By utilizing Neo GBR kit, Tent Screw, Cover Screw and Fixing Screw, vertical and horizontal bone level can be augmented.
bone level augmentation for vertical, horizontal and wide area maintaining space for bone formation to ease the fixing of membrane
screw to secondarily stabilize the membrane
Connect stopper to Drill Drill at least 3mm depending onthe bone density
Fix Tent Screw with Dual Hex Driver after drilling with 1.5
Stopper connectedto Drill
Stopper connectedto Drill
(For Contra angle) (For Handle)
Neo GBR kit’s Tent Screw is a Temporary Screw that works as a prop to procure space to enable vertical and horizontal bone formation on large defected area
Neo GBR kit’s Fixing Screw secondarily stations Membrane and maintains the desirable volume of bone formation
1. Insert 3mm or more into the bone and needs primary stability of 15~25Ncm2. Tent Screw could break when 30Ncm is exceeded3. Insert to the level of bone grafting
[ While fastening Fixing Screw ]
[ While fastening Tent Screw ]
Fix Tent Screw with Dual Hex Driver after drilling with 1.3
Fix Tent Screw with Dual Hex Driver after drilling with 1.0
Insert by Self Tapping
Fixate Fixing Screw with PhillipsDriver after drilling with 1.3
Fixate Fixing Screw with PhillipsDriver after drilling with 1.0
Self Tapping Self Tapping with 5~7mmScrew
Insert Tent Screw leaving space for augmentation (15~25Ncm)
Decortificated surroundings afterTent Screw insertion
Choose appropriate CTi-mem Tighten Fixing Screw in to FixingHole
Create matching hole for CoverScrew on CTi-mem with tweezers
to stabilize and submerge Tent Screw, Spacer and Membrane
to fixate Membrane using Cover Screw after connecting Spacer to Fixture directly
Tent Screw connectedto dual Hex Driver 1.6Hex
Cover Screw connectedto dual Hex Driver 1.0Hex
Fixing Screw connected to Phillips Driver
Prior to Bone grafting, to avoid bone material from entering Tent Screw, insert material after tightening Cover Screw. Check for correspondence with Cover Screw Hole
After removing Cover Screw apply CTi-mem, fixate Tent Screw on CTi-mem with Cover Screw(10~15Ncm)
Adapt the remaining part of CTi-mem to the tissue and cover up