Gas Chromatography
Modern Techniques in Heterogeneous Catalysis Research
Prof. Neil M. Schweitzer
Updated 2-4-16
• General technique used to separate gas into pure components based on interactions with a stationary phase in a column
Gas Chromatography
• Why use GC?
– Easy to quantitate
• High Sensitivity
– High resolution
– High reproducibility
– Very Fast!- The Essential Chromatography & Spectroscopy Guide, GC and GC/MS, Agilent Technologies
Gas Chromatography
Oven - L. Polite, Practical Gas Chromatography
These are the general parts of aGC, and the focus of ourdiscussion…
GC Injection Ports
• What is the purpose of the Injection port?
– Introduce the sample onto the column - Temperature
– Control the flow of gases through the column - Pressure
• Many exist, but most common is the split/splitless injector
- L. Polite, Practical Gas Chromatography
There is no flow controller on the column… flowrate through column is controlled by the head pressure of the column based on its characteristics (diameter and length)
Keeps air from a leaky septum out of the column
Valve controls percent of analyte vented based on assigned split ratio
If GC can’t maintain column pressure, always check septum first!
Fixed volume of gas is injectedinto the GC inlet when the 6-wayvalve turns back and forth
On-line Gas Injection
• Injection amount determined from Ideal-gas law
– Highly reproducible
• Accomplished using a 6-way valve and a fixed sample loop
Columns• How do columns separate gas components?
• What influences gas retention time?
– Keq
– Temperature
– Flowrate
Assuming fast mass transport, theanalyte in the gas phase and thatadsorbed on/in the stationaryphase quickly reach equilibrium
The analyte in the mobile phasespasses through the small columnvolume to the next volume
The analyte continues to travelthrough the entire length of thecolumn to the detector in this way,propelled by the carrier gas in themobile phase
At 𝑡 = 0, a plug of analyte gas isinjected into the first small volumeof the column
The mobile phase and stationaryphase come to equilibrium again
A carrier gas (the mobile phase)continually passes through thecolumn containing a stationaryphase (black and white layer)
0 2dL
time
a) b) c) d) e) f)
dL 0 2dLdL
t = 0
ColumnsLiquid Columns:
Solid Columns:
Porous Layer Open Tubular
- Agilent: The Essential Chromatography and Spectroscopy Catalogue
Analyte absorbs INTO the stationary phase
Analyte adsorbs ONTO the stationary phase
Detectors• Many to select from, each with various selectivities and sensitivities
- L. Polite, Practical Gas Chromatography
Thermal Conductivity Detector(Universal Detector)
Mass-spec Detector
Nitrogen PhosphorousDetector
Flame Photometric Detector(Phosphorous and Sulfur)
Infrared Detector(IR sensitive)
Photoionization Detector(UV-vis sensitive)
Flame Ionization Detector(Hydrocarbons)
Flame Ionization Detector
• Cons:
– Destructive
– Can’t detect O2, CO2, CO, etc..
• Pros:
– High sensitivity to hydrocarbons
– Signal roughly proportional to number of C atoms
- L. Polite, Practical Gas Chromatography
Analyte exits column and mixes with hydrogen
Analyte enters hydrogen flame, combusting and producing carbon cations
Carbon cations are detected by an electrometer
Thermal Conductivity Detector300 K 400 K 500 K 600 K
--- Air 26.2 33.3 39.7 45.7
Ar Argon 17.9 22.6 26.8 30.6
H2 Hydrogen 186.9 230.4 - -
H2O Water 18.7 27.1 35.7 47.1
H2S Hydrogen sulfide 14.6 20.5 26.4 32.4
NH3 Ammonia 24.4 37.4 51.6 66.8
He Helium 156.7 190.6 222.3 252.4 8
Kr Krypton 9.5 12.3 14.8 17.1
NO Nitric oxide 25.9 33.1 39.6 46.2
N2 Nitrogen 26.0 32.3 38.3 44.0
N2O Nitrous oxide 17.4 26.0 34.1 41.8
Ne Neon 49.8 60.3 69.9 78.7
O2 Oxygen 26.3 33.7 41.0 48.1
O2S Sulfur dioxide 9.6 14.3 20.0 25.6
CO Carbon monoxide 25.0 32.3 39.2 45.7
CO2 Carbon dioxide 16.8 25.1 33.5 41.6
CHCl3 Trichloromethane 7.5 11.1 15.1 -
CH4 Methane 34.1 49.1 66.5 84.1
CH4O Methanol - 26.2 38.6 53.0
C2H2 Acetylene 21.4 33.3 45.4 56.8
C2H4 Ethylene 20.5 34.6 49.9 68.6
C2H6 Ethane 21.3 35.4 52.2 70.5
C2H6O Ethanol 14.4 25.8 38.4 53.2
C3H6O Acetone 11.5 20.2 30.6 42.7
C3H8 Propane 18.0 30.6 45.5 61.9
C4H10 Butane 16.4 28.4 43.0 59.1
C5H12 Pentane 14.4 24.9 37.8 52.7
C6H14 Hexane - 23.4 35.4 48.7
• Pros:
– Non-destructive
– Universal detector
• N2, O2, CO2, CO, Ar, ect.
• Cons:
– Identifies a property of gases, not the gases themselves…
– Relatively low sensitivity compared to other detectors
• Detecting H2? Use any inert carrier…
• Detecting anything else? Use a He carrier…
TCD detects the change in the thermal conductivity of a gas mixture (the carrier gas plus eluted analyte) relative to a reference gas (the carrier gas alone)
NOTE: The schematic above represents a general TCD configuration, it does not represent an Agilent TCD configuration
We can quantify this seemingly abstract term, and use this understanding to develop optimum methods!
Method Development
• What makes this an ideal Chromatograph?– High retention
– High selectivity
– High efficiency
HIGH RESOLUTION!
• Always start with an established method!
Method Development
𝑅𝑆 =𝛼 − 1
𝛼×
𝑘
1 + 𝑘×
𝑁
4
Master Resolution Equation:
Selectivity CapacityFactor
Efficiency
- L. Polite, Practical Gas Chromatography
THINK BACK:
• What influences gas retention time?
– Keq
– Temperature
– Flowrate 0 2dLdL
Alpha is only a function of the stationary phase used, based on the equilibrium between the stationary phase and the analyte
Alpha measures the retention of two analytes relative to each other in the column and is independent of temperature and flowrate!
Method Development
𝑅𝑆 =𝛼 − 1
𝛼×
𝑘
1 + 𝑘×
𝑁
4
Master Resolution Equation:
Selectivity CapacityFactor
Efficiency
- L. Polite, Practical Gas Chromatography
THINK BACK:
• What influences gas retention time?
– Keq
– Temperature
– Flowrate 0 2dLdL
tR: total time analyte spends in the column
t0: time analyte spends in the mobile phase
t’R: time analyte spends in the stationary phase
At higher temperatures, equilibrium (regardless of analyte or stationary phase) will favor the gas phase, and gases will elute at shorter retention times because the analyte spends less time in the stationary phase
Essentially, the analyte should spend sufficient time in the stationary phase to achieve adequate separation of analytes
Method Development
𝑅𝑆 =𝛼 − 1
𝛼×
𝑘
1 + 𝑘×
𝑁
4
Master Resolution Equation:
Selectivity CapacityFactor
Efficiency
• Efficiency Increases with:– Longer column
– Smaller Diameter
– Thinner film
– Optimized Flowrate
- L. Polite, Practical Gas Chromatography
THINK BACK:
• What influences gas retention time?
– Keq
– Temperature
– Flowrate 0 2dLdL
Van Deemter Plot
U (cm/s)Take home message… higher efficiency leads to sharper peaks!
• Goals: develop a method that optimizes resolution and time
– Select a column with high Selectivity and high Efficiency
– Optimize temperature program to optimize resolution and minimize time
• Start with a broad temperature ramp
– Optimize flowrate to optimize resolution and minimize time
• Start with an ideal flowrate (~25 cm/s)
Method Development
Acknowledgments:Dr. Lee Polite, Axion Labs: www.axionlabs.com** We highly recommend Practical Gas Chromatography, an