FUNCTION GENERATOR OPERATION MANUAL...function generator, which can directly generate Sine,...

12
1 FUNCTION GENERATOR OPERATION MANUAL INDEX 1. Introduction ………………………………………… ( 3) 2. Specifications ………………………………………… (3) 3. Working Theory ……………………………………… (7) 4. Structure ……………………………………………… (8) 5. Operation & Maintenance …………………………… (9) 6. Accessories ………………………………………… (14)

Transcript of FUNCTION GENERATOR OPERATION MANUAL...function generator, which can directly generate Sine,...

  • 1

    FUNCTION GENERATOR

    OPERATION MANUAL

    INDEX

    1. Introduction ………………………………………… (3)

    2. Specifications ………………………………………… (3)

    3. Working Theory ……………………………………… (7)

    4. Structure ……………………………………………… (8)

    5. Operation & Maintenance …………………………… (9)

    6. Accessories ………………………………………… (14)

  • 2

    This instrument is a highly stable, multipurpose, broad bandwidth

    function generator, which can directly generate Sine, Triangle,

    Square, Ramp, Pulse waveforms as well as Linear & Logarithm

    Sweep signals. Application of Single-chip-machine intelligently

    controls all of the functions, displays and indicators, which allow

    users to know about the working status of the instrument easily

    and accurately.

    1. Specifications

    1.1 Frequency Range: 0.6Hz-6MHz

    1.2 Waveform:

    Sine, Triangle, Square, Positive & Negative Ramp,

    Positive & Negative Pulse. Symmetry of Ramp and Pulse

    is variable within 80:20 (1Hz~100kHz)

    1.3 Sine Waveform:

    1) Distortion: ≤1% (0.1Hz~100kHz)

    2) Frequency Response: ≤±0.5dB(0.1Hz~100KHz);

    ≤±1dB(3MHz);

    ≤±2dB(6MHz);

    1.4 Rising/Falling Edge of Square Waveform: ≤50ns

  • 3

    1.5 TTL Output:

    1) Higher Level ≥2.4V, Lower Level ≤0.4V, capable

    of driving 20 TTL loads.

    2) Rising time: ≤30ns

    1.6 Output:

    1) Impedance: 50Ω±10%

    2) Output Amplitude: ≥20Vp-p (open circuit),

    3) Output Attenuator: 20dB, 40dB, 60dB

    4) DC Offset: 0 ~±10V adjustable

    5) Display Error of Amplitude: ±10% ±2 digits

    (output level higher that 1/10 of full scale)

    1.7 Power Output:

    1) Amplitude: ≥20p-p

    2) Output power: Max. 5W

    1.8 VCF Input:

    1) Max. Voltage: -5V~0V ±10%

    2) Max. Control rate: 1000:1

    3) Frequency: DC~1kHz

    1.9 Amplitude Modulation:

    1) Carrier Bandwidth:

    2) Modulation Freq.: 400Hz (INT), 0~1MHz (EXT)

    3) Modulation Depth: 0~85%

  • 4

    1.10 Frequency Modulation:

    1) Deviation: 0 ~ 10%

    2) Modulation Freq.: 400Hz(INT), 0 ~ 20kHz (EXT)

    3) Distortion: ≤ 1.5% ( fc=500kHz, fm=1kHz, 10%)

    1.11 Sweep

    1) Sweep mode: Linear, Logarithmic.

    2) Sweep time: 10ms~5s

    3) Sweep rate: 1000:1

    4) Output Amplitude: 7Vp-p

    5) Output Impedance: 600Ω

    1.12 Frequency Counter:

    1) Measuring Range: 1Hz-20MHz

    2) Input Impedance: 1MΩ / 20pF

    3) Sensitivity: 10mVrms

    4) Max Input: 150V AC+DC (with attenuator)

    5) Input attenuator: 20dB

    6) Measuring Error: not more than 3×10-5 ±1 digit

    1.13 Power Supply

    1) Voltage: AC 220V ± 10%

    2) Frequency: 50Hz±2Hz

  • 5

    1.14 Operating Ambient:

    1) Temperature: 0°C~40°C

    2) Humidity: ≤RH 90%

    3) Atmospheric Pressure: 86kPa~104kPa

    1.15 Dimension (mm): 290L × 260W × 90H

    1.16 Weight: 3kg

    2. Working Theory

    Figure 1

    2.1 Waveform Generator Circuit

    This segment is consisted of MAX038 Function Generator,

    Frequency & Duty Rate control circuits. Waveforms are

    selected by SCM.

  • 6

    2.2 SCM (Single-chip-machine) Circuit

    This segment is consisted of SCM 89C52, Panel Input,

    Frequency & Amplitude Display and Output & Indicator

    circuits. The main jobs are: Control waveform of output

    signal; Test and display the frequency of output or external

    input signal; Display amplitude of output waveform.

    2.3 Frequency Counting Circuit:

    It is consisted of Broad-banded Amplifier and

    Square-waveform Regulator, which external signal will go

    through.

    2.4 Power Amplifier

    In order to insure the stability and high slew rate, dual

    channel power amplifier is employed. The entire amplifier is

    with the characteristic of phase inversion.

    2.5 FM Mode

    When SWP/FM switch being pressed down and MOD

    switch is pulled up, a 1KHz modulation signal which

    generated by RC oscillator is applied to VCF end.

    2.6 Power Supply

    Four sets of power supply, ±23V/±15V/±5V/+5 are

    employed. ±23V is used for Amplifier. ±15V and ±5V are

    used for Waveform Generator. +5V is mainly for SCM circuit.

  • 7

    3. Structure

    This instrument has an all-metal frame and cover, plastic

    front panel with a elegant and small-sized appearance. Most

    of the components (including switches and knobs) are fixed

    on two PCBs. All adjustable components are located at an

    easy-to-adjust position. Only to unscrew the back panel and

    remove the cover up and down when the instrument need to

    be adjusted or repaired.

    4. Operation and Maintenance

    4.1 Symbols and functions in front panel and rear panel:

    See Table 1, Figure 2 and Figure 3.

    Figure 2

  • 8

    Table 1

    NO SYMBOL DESCRIPTION OPERATION

    1 POWER Power Switch Push down, power on

    2 △ Frequency Range

    Selection

    Press the key, the multiples of freq. will be changed

    from low to high. Collaborate with Knob “19” to set

    output frequency.

    3 ▽ Frequency Range

    Selection

    Press the key, the multiples of freq. will be changed

    from high to low. Collaborate with Knob “19” to set

    output frequency.

    4 Waveform Selection

    Press the key to change output waveform. Work

    with knob “17” and key “18”, polarity of ramp and

    pulse can be selected.

    5 DC OFFSET DC Offset Press button. Indicator lights when the function is

    activated.

    6 DC Offset Knob DC output can be adjusted with from

    -10V~+10V(Output waveform is 5Vp-p)

    7 MODE

    Linear/Logarithmic

    /External Sweep

    selection

    Sweep mode will change in sequence each time the

    button is pressed.

    8 SPEED Sweep Speed Adjust the knob to change sweep speed.

    9 Output Attenuator Press the button to set attenuation among 0, 20dB,

    40dB and 60dB. Indicators light respectively.

    10 OUTPUT Voltage Output Output terminal of waveforms, max. output 20Vp-p

    with impedance of 50Ω.

    11 AM/FM AM/FM Adjustment

    Activate modulation mode with button “24”. If the

    knob is Pull-out position, the instrument is in AM

    mode, modulation depth can b adjusted with the

    knob. If the knob is Push-in position, the instrument

    is FM mode, frequency deviation can be adjusted

    with the knob.

  • 9

    12 AMPLITUDE

    Output Amplitude

    Adjustment

    Waveform Inversion

    Work with button “9” to adjust amplitude of output

    waveforms.

    13 INPUT Counter Input

    Terminal

    To test the frequency of an external signal, connect

    the signal to this terminal.

    14

    EXT

    ATT20dB

    LPF

    Selector of

    INT/EXT signal,

    20dB Attenuator and

    Low-Pass Filter

    When external signal is selected and the signal is too

    strong, press the key, 20dB attenuator is activated

    with the indicator lit. Press the key again, LPF

    indicator lights(with attenuator, cutoff freq. 100kHz)

    About 20 seconds after no signal is detected, “0”

    will be displayed.

    15 WIDTH Sweep Width Use the knob to adjust sweep width when the

    instrument is in sweep mode.

    16 Amplitude Display

    Display peak to peak value of output signal (open

    circuit). If a 50Ω load is connected, actual value on

    the load is 1/2 of that is displayed. If desired output

    amplitude is lower than 1/10 of max. value when

    amplitude knob is fully turned, connecting an extra

    attenuator is advised. If the instrument is working in

    AM mode, displayed amplitude is an Un-cal value.

    17 SYMMETRY Symmetry Adjusting Symmetry can be adjusted from 20:80 to 80:20.

    18 Symmetry

    Activating Key

    Press the key to activate symmetry adjusting knob.

    19 Frequency

    Adjustment

    Turn the knob CW, output frequency is increased

    and vice versa.

    20 Frequency Display

    Display the frequency of output signal or external

    input signal. Indicator “GATE” blinks when counter

    is working well. If input frequency over 20MHz,

    indicator “OV.FL” will be on. Hz and kHz indicate

    what units of frequency would be at the moment

    when one of them turns on.

  • 10

    4.2 Symbols and functions in rear panel:

    Figure 3

    21 Power Socket AC 220V/0.5A fuse or AC110V/1A fuse.

    22 Line Voltage Selection 110V/220V switching.

    23 VCF IN/EXT MOD IN/EXT

    SWP IN Terminal

    a) Input terminal of Voltage-Control-Frequency,

    input voltage range 0~-5V.

    b) Input terminal of External Modulation signal.

    c) Input terminal of External Sweep signal.

    A sweep signal is generated while frequency

    vary along input control signal.

    24 MOD ON/OFF switch of AM/FM,

    INT/EXT modulation selection.

    25 TTL OUT Output terminal of TTL pulse,

    with impedance of 50Ω.

  • 11

    4.3 Service and Calibration

    This instrument can be operated continuously in proper

    ambient, but you are advised to calibrate it every three

    months. Calibration procedure is as:

    1) Sine Distortion Calibrating:

    Turn Amplitude Adjusting Knob to its max, select sine

    waveform output and set frequency at 1kHz. Connect this

    signal to a distortion meter. Adjust RP102 to have distortion

    match technical requirement.

    2) Output Amplitude Calibrating:

    Set the instrument in the same working status as 1), connect

    the signal to an oscilloscope. With observing peak-peak

    value of the output signal, adjust RP601 to have the output

    amplitude match technical requirement.

    3) Frequency Calibrating:

    Set frequency counter at “EXT”, connect a 10MHz signal of

    standard oscillator to the input terminal. Adjust C607 to get

    standard 9999.9kHz on the LED display. Set the amplitude

    of the standard signal to 100mVrms, adjust RP401 to have

    9999.9kHz stably displayed on LED.

  • 12

    4.4 Trouble Shooting

    Sufficient knowledge is needed when you are going to fix

    problems on this instrument. Locate the problem area in

    PCB by with observing the phenomenon, visual inspection

    is suggested in first place, and then check the PCB in static

    and dynamic mode. Replace the component which is found

    broken.

    5. Accessories

    Operation Manual 1copy

    Power Cord 1pc

    Test Lead(50Ω) 1pc

    BNC Cord 1pc

    Fuse (0.3A) 2pcs