Density Meter Calibration
David AakreTechnologist
![Page 2: Frac Density Meter Calibration](https://reader036.fdocuments.us/reader036/viewer/2022081421/586f8e541a28ab54768b7211/html5/thumbnails/2.jpg)
Density Transmitter/ Meter/ Radiation Source (Cesium 137)
![Page 3: Frac Density Meter Calibration](https://reader036.fdocuments.us/reader036/viewer/2022081421/586f8e541a28ab54768b7211/html5/thumbnails/3.jpg)
Meter Box/ Thermo
![Page 4: Frac Density Meter Calibration](https://reader036.fdocuments.us/reader036/viewer/2022081421/586f8e541a28ab54768b7211/html5/thumbnails/4.jpg)
Thermo Meter –Set Up • Two Point Calibration on Air and Water • DAC codes are used to Calibrate• Two different procedures to calibrate In-line and Blender. Procedures
are two pages long and contain cal- points for different size Iron.• Simulate 4 MA- Exit Set Up, Press 8, down Arrow. Right arrow to
accept• Simulate 20 Ma- Exit Set Up, Press 5, down arrow. Right arrow to
accept.• Clear Simulation- Exit Set Up, press 9, down arrow, right arrow to
accept.• Calibrate on Base Fluid- Exit Set Up, press 121, down arrow, right
arrow to accept. Take 120 seconds to cal.
![Page 5: Frac Density Meter Calibration](https://reader036.fdocuments.us/reader036/viewer/2022081421/586f8e541a28ab54768b7211/html5/thumbnails/5.jpg)
Record The Results of Thermo Calibration So You Can Recover Meter
• SGDO MEMORY LOSS RECOVERY PROCEDURE
• MUST HAVE LAST CALIBRATION RECORDED: WATER STANDARDIZATION VALUE AND SLOPE CORRECTION VALUE
• 1. Erase memory by doing a “74” command. (“Exit Setup”, “74”, ↓ , → , → )• 2. Input oilfield setup. (“Exit Setup”, “1016”, ↓ , → , → )• 3. Input pipe ID. (“Exit Setup”, “048003”, ↓ , “??.??”, ↓ , “Exit Setup”)• 4. Run Standardization cycle. (“Exit Setup”, “121”, ↓ , → ) (Ignore what is in the pipe)• 5. This was done to set command “flags” in memory.• 6. Input the “water” value recorded during the last calibration.• 7. “Exit Setup”, “128003”, ↓ , “?.???E?”, ↓ , “Exit Setup”• 8. Input the “slope correction“ value recorded from the last calibration.• 9. “Exit Setup”, “049003”, ↓ , “?.???”, ↓ , “Exit Setup”• The Meter Should be Restored to last Calibration. So Easy a Green Hat Can Do it. lol
![Page 6: Frac Density Meter Calibration](https://reader036.fdocuments.us/reader036/viewer/2022081421/586f8e541a28ab54768b7211/html5/thumbnails/6.jpg)
Lime Density Meter/ Thermo -Denso
![Page 7: Frac Density Meter Calibration](https://reader036.fdocuments.us/reader036/viewer/2022081421/586f8e541a28ab54768b7211/html5/thumbnails/7.jpg)
Lime Calibration Screen Calibrate on Air. Fill Detector with Water Cal on H20
![Page 8: Frac Density Meter Calibration](https://reader036.fdocuments.us/reader036/viewer/2022081421/586f8e541a28ab54768b7211/html5/thumbnails/8.jpg)
Lime PPA Calibration Screen/ 11 cal points
![Page 9: Frac Density Meter Calibration](https://reader036.fdocuments.us/reader036/viewer/2022081421/586f8e541a28ab54768b7211/html5/thumbnails/9.jpg)
Source/ Transmitter( Pre-Amp)
![Page 10: Frac Density Meter Calibration](https://reader036.fdocuments.us/reader036/viewer/2022081421/586f8e541a28ab54768b7211/html5/thumbnails/10.jpg)
Thermo- Pre Amp Set UP
![Page 11: Frac Density Meter Calibration](https://reader036.fdocuments.us/reader036/viewer/2022081421/586f8e541a28ab54768b7211/html5/thumbnails/11.jpg)
BERTHOLD LB 444 Frac Densometer
![Page 12: Frac Density Meter Calibration](https://reader036.fdocuments.us/reader036/viewer/2022081421/586f8e541a28ab54768b7211/html5/thumbnails/12.jpg)
Test Density System with Frac Slurry or Cement
• Fracturing Chemicals • Water • Biocides • Clay Control • Polymers • pH Control • Fluid Loss Additives • Conductivity Enhancers • Gel Stabilizers • Gel Breakers • Surfactants • Crosslinkers
![Page 13: Frac Density Meter Calibration](https://reader036.fdocuments.us/reader036/viewer/2022081421/586f8e541a28ab54768b7211/html5/thumbnails/13.jpg)
Frac Slurries Mixture• Gel Mixture:• Buffer- weak acid• Crosslinker- increases viscosity• Beryllium• Guar- Gelling Agent• Breaker: Oxides, Acids, Enzymes, Breaks gel down in about 30 minutes• Potassium Chloride• Sand- 20/40 specific gravity 2.65• Water PH 7 or 8• Make sure sand is suspended in mixture- does not fall out
• Slick Water:• Carbopol 940- High viscosity clear gel- Hair Gels• Ethylene Glycol• Caustic soda• Water PH 7 or 8• Sand-20/40 specific gravity 2.65
![Page 14: Frac Density Meter Calibration](https://reader036.fdocuments.us/reader036/viewer/2022081421/586f8e541a28ab54768b7211/html5/thumbnails/14.jpg)
Measure Density of Your Mixtures
![Page 15: Frac Density Meter Calibration](https://reader036.fdocuments.us/reader036/viewer/2022081421/586f8e541a28ab54768b7211/html5/thumbnails/15.jpg)
Densometer Testing/ Calibration
• Mix five gallon mixtures of slurry using different Densities.
• Measure Density of Mixtures with AccuPYC 1340.• Pour Mixture in densometer and shut in. Add
100 to 150 psi of pressure using a compressor.• Take and average of five meter readings and
compare to know Density(AccuPYC). Can also use this process to calibrate a densometer.
![Page 16: Frac Density Meter Calibration](https://reader036.fdocuments.us/reader036/viewer/2022081421/586f8e541a28ab54768b7211/html5/thumbnails/16.jpg)
DAVID AAKRE TECHNOLOGIST
Practice makes perfect!!!