Four Point Probe Procedure for Pro4 using Keithley.

15
Four Point Probe Procedure for Pro4 using Keithley

Transcript of Four Point Probe Procedure for Pro4 using Keithley.

Four Point ProbeProcedure for Pro4 using Keithley

Overview

• What is Four Point Probing• How the system works• Pro 4 Set Up• Simple Calculations behind Four Point Probing• Procedure for using Pro4

What is Four Point Probing

• Four Point Probing is a method for measuring the resistivity of a substance.

• Impurity concentrations can be estimated from the resistivity

Resistivity vs Sheet Resistance

• Bulk or volume resistivity (r) is measured in ohms-cm

• Independent of sample size or shape

• Sheet resistance (rs) is measured in ohms-per-square

• Can be used to measure the value a resistor in a IC

Pro-4 Set Up

Pro-4 probing station from LUCAS LABS

with 4 point probe head

KEITHLEY 2400 power/source meter

Computer with Pro4 software and

interface

The 4 point probing setup consists of 3 key components

Probing Station

Source Meter

Pro-4 Software

The 4 point probing setup can measure resistivity or the thickness of a film. But, either one has to be known.

Resistivity Probe Stand

Contact Lever

Probe head electrical connection

Probe Head

Mounting Chuck (Aluminum base with

Teflon surface

How the system works • Current is passed through the two outer probes

• Voltage is measured between the two inner probes

• Read and record both current and voltage values from the

Keithley source meter

• Sheet Resistance is measured using (V/I) and k

• V = volts, I = Amps (convert current reading to amps)

• k=constant factor = to 4.53 when the wafer diameter is

much greater than the probe spacing – typical for wafers

• Sheet resistance (rs) = (k)(V/I)= ohms/square

For the bulk resistivity of a wafer

• The thickness of the wafer/film

must be known – use calipers or

micrometer block to measure the

wafer thickness

• Convert caliper reading in mm to

um (microns)

• Resistivity of wafer will be shown

on the computer screen

• There is a second k factor but for

our work this k factor is not a

factor and can be ignored

(typically >.995)

To measure the thickness of a waferUse a caliper or the micrometer stage shown below

To turn on, press and hold the right buttonTo turn off, press and hold both buttonsThe reading on the screen is in mm – a reading of .400mm = 400 microns (um)

Procedure for using the Pro-4

Enter thickness in microns (um) as measured

Choose either single or multiple readings per wafer. Load the wafer under the 4 point probe head and lower the head using the handle

Procedure for using the Pro-4

Procedure for using the Pro-4

Lower the probe heading using the contact lever until

the computer screen shows a measurement taking place. Once completed, the screen will show the resistivity and

V/I

Record resistivity in ohm-cm as shown on the screen

Saving the data

• After the measurement is

completed, the resistivity at each

location will be displayed on the left

hand side of the screen.

• When all the points are tested, the

data can be saved and read using

excel

• The Pro-4 can be used to measure resistivity or the thickness. But, either

one has to be known.

• Typically the thickness of the wafer can be measured

• The # of points to be tested and the shape of the sample can be selected.

• A single point or multiple points on the sample can be tested to obtain the

average resistivity.

• The resistivity is automatically displayed when the thickness is entered.

• Additional information is also displayed including V/I

Summary

Sample Wafer CalculationsUsing voltage and current measurements

• A current of 1.0 mA is passed through the wafer and a voltage reading of 0.030 v is noted. I = 1.0 MA = .001 amp

• V/I = .030 v/.001A = 30 ohm• rs = (V/I) k = (.030/.001)(4.53) = (30)(4.53) =

135.9 ohms/square• The wafer is measured as 0.40 mm = .04 cm• r = (135.9)(.04cm) = 5.43 ohm-cm