Installing the Dionex PeakNet®-PA System

For PeakNet-PA Version 5.21

© 2002 Dionex Corporation

Document No. 031832Revision 01June 2002

Page 2: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

©2002 by Dionex CorporationAll rights reserved worldwide.Printed in the United States of America.

This publication is protected by federal copyright law. No part of this publication may be copied or distributed, transmitted, transcribed, stored in a retrieval system, or transmitted into any human or computer language, in any form or by any means, electronic, mechanical, magnetic, manual, or otherwise, or disclosed to third parties without the express written permission of Dionex Corporation, 1228 Titan Way, Sunnyvale, California 94088-3603 U.S.A.

DISCLAIMER OF WARRANTY AND LIMITED WARRANTYTHIS PUBLICATION IS PROVIDED “AS IS” WITHOUT WARRANTY OF ANY KIND. DIONEX CORPORATION DOES NOT WARRANT, GUARANTEE, OR MAKE ANY EXPRESS OR IMPLIED REPRESENTATIONS REGARDING THE USE, OR THE RESULTS OF THE USE, OF THIS PUBLICATION IN TERMS OF CORRECTNESS, ACCURACY, RELIABILITY, CURRENTNESS, OR OTHERWISE. FURTHER, DIONEX CORPORATION RESERVES THE RIGHT TO REVISE THIS PUBLICATION AND TO MAKE CHANGES FROM TIME TO TIME IN THE CONTENT HEREINOF WITHOUT OBLIGATION OF DIONEX CORPORATION TO NOTIFY ANY PERSON OR ORGANIZATION OF SUCH REVISION OR CHANGES.

TRADEMARKSDX-LAN™ is a trademark and PeakNet® is a registered trademark of Dionex Corporation.Wonderware InTouch® is a registered trademark of Wonderware Corporation.Pentium® is a registered trademark of Intel Corporation.Windows™ is a trademark and Windows NT® is a registered trademark of Microsoft Corporation.

PRINTING HISTORYRevision 01, June 2002

Page 3: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Contents

Doc. 031832-01 6/02 i

Installing the PeakNet-PA System

1 PeakNet-PA System Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

2 PeakNet-PA Moduleware-BIOS Compatibility . . . . . . . . . . . . . . . . . . . . . 2

3 Installation Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

4 Installing the Computer DX-LAN Card . . . . . . . . . . . . . . . . . . . . . . . . . . . 44.1 Identifying the DX-LAN Card Type . . . . . . . . . . . . . . . . . . . . . 44.2 Installing the DX-LAN Card in the Computer . . . . . . . . . . . . . . 5

5 Connecting the DX-LAN Network . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65.1 Connecting a 10BASE-T DX-LAN Network (Overview) . . . . . 65.2 Connecting a BNC DX-LAN Network (Overview) . . . . . . . . . . 95.3 DX-LAN Network Connection Examples . . . . . . . . . . . . . . . . 11

6 Installing the DX-LAN Card Device Driver . . . . . . . . . . . . . . . . . . . . . . . 16

7 Installing PeakNet-PA Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 197.1 Installing PeakNet-PA Programs . . . . . . . . . . . . . . . . . . . . . . . 197.2 Installing Wonderware InTouch (Runtime) . . . . . . . . . . . . . . . 217.3 Installation of the Runtime License . . . . . . . . . . . . . . . . . . . . . 26

8 Setting Up the Database . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27

9 Setting Up InTouch to Start Automatically . . . . . . . . . . . . . . . . . . . . . . . . 30

10 Setting the Time and Date Formats . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34

11 Selecting a Video Resolution (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . 35

Page 4: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Installing the PeakNet-PA System

ii Doc. 031832-01 6/02

Page 5: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Doc. 031832-01 6/02 1

Installing the PeakNet-PA System

NOTE When you purchase your PeakNet-PA software, allnecessary software and computer hardware are installedat the factory. These instructions are for referenceshould you need them at a later time.

1 PeakNet-PA System Requirements

• IBM-compatible computer, Pentium® 133 MHz minimum

• 256 MB RAM minimum

• Hard disk drive with minimum of 20 MB of free space for PeakNet-PA program files and data files

• 3.5-inch high-density (1.44 MB) diskette drive

• CD-ROM drive

• Windows™-compatible monitor and video adapter; Super VGA 800 x 600 with 256 colors recommended

• Mouse supported by Windows

• (Optional) Printer supported by Windows (parallel interface recommended). Color printer recommended for printing trend plots

• Microsoft Windows NT®, Version 4.0—installed before beginning the PeakNet-PA system installation

• Database program that supports ODBC (for example, Microsoft Access)

Page 6: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Installing the PeakNet-PA System

2 Doc. 031832-01 6/02

2 PeakNet-PA Moduleware-BIOS Compatibility

The Moduleware and BIOS of all installed Dionex devices must be compatible with PeakNet-PA. Incompatibilities will cause system communication failures.

NOTE When running PeakNet-PA (release 5.21 or higher), theBIOS version for all devices must be 30.00 or higher.

Before downloading new Moduleware to a device, check the following table to verify that the Moduleware version number supports PeakNet-PA (release 5.21 or higher).

How to Identify the Current Versions of Moduleware and BIOS

• Most devices display their Moduleware and BIOS versions at power-up.

• On many devices, you can also select the POWER-UP SCREEN from the DIAGNOSTIC MENU to display the Moduleware and BIOS versions.

• On the UI20 Universal Interface, press the front panel Status button to display the Moduleware and BIOS versions.

• From the PeakNet-PA Configuration Editor, select Moduleware from the Configure menu to display the Moduleware version for any module.

Module Moduleware Version Module Moduleware Version

AD20 3.00 or higher EG40 2.25 or higher

AD25 1.00 or higher GP40 3.00 or higher

AS50 1.08 or higher GP50 3.00 or higher

CC80 5.26 or higher IC20 3.00 or higher

CD20 3.00 or higher IC25 1.00 or higher

CD25 1.00 or higher IP20 3.00 or higher

DX-120 3.01 or higher IP25 3.00 or higher

ED40 3.00 or higher UI20 3.01 or higher

ED50 1.00 or higher

Table 1. Moduleware Versions Required for PeakNet-PA (Release 5.21 or Higher) Support

Page 7: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Installation Overview

Doc. 031832-01 6/02 3

3 Installation Overview

Installation of the PeakNet-PA system consists of the following steps:

• Install the DX-LAN™ card in the PeakNet-PA workstation

• Connect the DX-LAN network

• Install the DX-LAN card device driver

• Install PeakNet-PA software

• Set up the database

• Set up Wonderware InTouch® to start automatically

• Set the time and date formats

• (Optional) Select a video resolution

Page 8: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Installing the PeakNet-PA System

4 Doc. 031832-01 6/02

4 Installing the Computer DX-LAN Card

4.1 Identifying the DX-LAN Card TypeDionex ships PeakNet-PA systems with the following types of DX-LAN cards:

• ISA Plug-and-Play card (P/N 052351)

• PCI Plug-and-Play card (P/N 052350)

• PCI Plug-and-Play card for 10BASE-T only (P/N 056908)

Before proceeding, check the card's part number and Figure 1 to determine which type of card you have.

Windows NT must be installed on the PeakNet-PA workstation beforethe card is installed.

Figure 1. Outside of DX-LAN Card Brackets

ISA Plug-and-Play PCI Plug-and-Play(for 10BASE-T only)

PCI Plug-and-Play

Page 9: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Installing the Computer DX-LAN Card

Doc. 031832-01 6/02 5

4.2 Installing the DX-LAN Card in the Computer

1. Turn off the computer and disconnect its power cord.

2. Remove the computer cover.

3. Locate an empty expansion slot into which the DX-LAN card will fit. Use the type of slot appropriate for the card (either ISA or PCI). If several slots are empty, leave as much space as possible around each installed card to ensure adequate air circulation.

4. Unscrew and remove the metal filler bracket that covers the rear panel opening for the selected expansion slot. Save the screw.

5. Carefully slide the DX-LAN card into the slot, aligning the card's metal mounting bracket with the expansion slot opening on the rear panel.

6. Align the card's bottom edge connectors with the connectors on the motherboard. Firmly push the card into the motherboard connectors until the card is fully seated.

7. Use the saved screw to secure the card's metal mounting bracket to the expansion slot.

8. Replace the computer cover and any cables you removed.

To prevent electrostatic discharge from damaging the DX-LAN card'selectronic components, use an electrostatic grounding strap duringthe installation. If one is not available, do the following: beforeremoving the card from its anti-static bag, touch the computer's metalframe to discharge any static electricity from your body.

DO NOT TURN ON THE COMPUTER until after the DX-LAN network isconnected (see Section 5) and the power is turned on to at least oneof the DX-800 analyzer modules. This ensures that correctcommunication is established between the DX-LAN card and the DX-800 Analyzer modules.

Page 10: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Installing the PeakNet-PA System

6 Doc. 031832-01 6/02

5 Connecting the DX-LAN Network

The DX-LAN network consists of a PeakNet-PA workstation and the devices in each DX-800 analyzer channel. For example, a network might consist of a workstation, a component controller, a detector, a pump, and an EG40-PA eluent generator.

Each device (including the workstation) must be equipped with a DX-LAN interface card, and either a BNC connector or a 10BASE-T RJ-45 (telephone-style) connector. The channel’s AE80 enclosure is equipped with either two BNC connectors (In and Out) or one 10BASE-T RJ-45 connector.

A DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure varies, depending on the type(s) of DX-LAN devices on the network.

• For an overview of how to connect a 10BASE-T network, see Section 5.1.

• For an overview of how to connect a BNC network, see Section 5.2.

• For examples of various network configurations, see Section 5.3.

5.1 Connecting a 10BASE-T DX-LAN Network (Overview)

Required Parts

Part Part Number Quantity

10BASE-T DX-LAN cables

960279–10 ft (3 m)960281–6 ft (1.8 m)960286–1 ft (0.3 m)

1 cable for each 10BASE-T network device

10BASE-T “combo” Ethernet hub

057398 1 or more, depending on the network configuration

Table 2. Parts Required for 10BASE-T Network Connections

The 10BASE-T DX-LAN cable is a Category 5 unshielded twisted-paircable. Do not substitute a cable of an inferior grade. Using anincorrect cable will cause a device to lose communication with thehost computer.

When using 10BASE-T cabling, you must install a hub. If you simplyplug the 10BASE-T cable from a device into the connector on thecomputer’s DX-LAN card, the connection will not work.

Page 11: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Connecting the DX-LAN Network

Doc. 031832-01 6/02 7

10BASE-T Network Connections for Networks with a Single Hub

1. Verify that a DX-LAN interface card is installed in each device to be connected. For complete installation instructions for any Dionex device, refer to the individual product manual.

2. Plug a 10BASE-T DX-LAN cable into a 10BASE-T port on the Ethernet hub.

3. Plug the other end of the cable into the 10BASE-T DX-LAN connector on the rear panel of a 10BASE-T device.

4. Repeat Steps 2 and 3 for the remaining 10BASE-T devices.

5. Set the Normal/Uplink push button on the hub to Normal.

Figure 2 shows example connections for a 10BASE-T network with one hub.

Figure 2. Example 10BASE-T DX-LAN Network: Connecting to One Hub

8 7 6 5 4 3 2 1

10BASE-T Device

10BASE-T Device

10BASE-T Device

(10BASE-T Device)10BASE-T Cable

Hub

10BASE-T Cables

Workstation

Page 12: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Installing the PeakNet-PA System

8 Doc. 031832-01 6/02

10BASE-T Network Connections for Networks with Multiple Hubs

1. Verify that a DX-LAN interface card is installed in each device to be connected. For complete installation instructions for any Dionex device, refer to the individual product manual.

2. Set the Normal/Uplink push button on the first Ethernet hub to Normal and set the push buttons on all other hubs to Uplink.

3. Using a 10BASE-T DX-LAN cable, connect port 8 on the second hub to any port on the first hub. Repeat for other hubs in the network. The Normal/Uplink push button affects port 8 only.

4. Using a 10BASE-T DX-LAN cable, connect the 10BASE-T network devices to ports 1 through 7 on the second and subsequent hubs.

Figure 3 shows example connections for a 10BASE-T network with multiple hubs.

Figure 3. Example 10BASE-T DX-LAN Network: Connecting to Multiple Hubs

Normal/Uplinkbutton =Normal

Hub 1

8 7 6 5 4 3 2 1

Hub 2 . . .

8 7 6 5 4 3 2 1

Hub x

8 7 6 5 4 3 2 1

Normal/Uplinkbutton =Uplink

Normal/Uplinkbutton =Uplink

10BASE-TDevice

10BASE-TDevice

10BASE-TDevice

10BASE-TDevice

(10BASE-T Device)

Workstation

Page 13: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Connecting the DX-LAN Network

Doc. 031832-01 6/02 9

5.2 Connecting a BNC DX-LAN Network (Overview)

1. Verify that a DX-LAN interface card is installed in each device to be connected. For complete installation instructions for any Dionex device, refer to the individual product manual.

2. Attach a BNC tee connector to the BNC connector on each device and on the BNC port on the Ethernet hub, if used.

a. Twist the tee onto the BNC connector until the locking pins that protrude from either side of the BNC connector are fully engaged in the slots on the tee connector.

b. Pull firmly on the tee connector to verify that it cannot move.

Part Part Number Quantity

BNC tee connector 921914 1 for each BNC network device including the 10BASE-T “combo” Ethernet hub, if present

Coaxial DX-LAN cable

960406 (3 ft)960404 (6 ft)960405 (15 ft)046016 (50 ft)

Enough to connect each device in a single line

50 ohm terminator resistor plugs

921034 2; one for each end of the network

Table 3. Parts Required for BNC Network Connections

The coaxial DX-LAN cable is a 50 ohm coaxial impedance cable. (Fiftyohm cables are imprinted with “RG-58U.”) Do not substitute aninferior cable, such as a 75 ohm television coaxial cable. Failure touse the correct cable or to lock it into place on the BNC tee connectorwill cause the device to lose communication with PeakNet-PA.

A 50 ohm terminator resistor plug must be installed at each end of theBNC DX-LAN network. Before beginning operation, verify that bothends of the DX-LAN have terminator resistor plugs installed.

Page 14: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Installing the PeakNet-PA System

10 Doc. 031832-01 6/02

3. Connect a coaxial DX-LAN cable to the first device.

a. Slide the metal sleeve on the end of the DX-LAN cable over the BNC tee connector.

b. Twist the metal sleeve onto the tee connector until the locking pins on the tee are fully engaged.

c. Pull on the end of the cable to verify that it cannot move.

4. Plug the free end of the coaxial DX-LAN cable into one of the following:

• The BNC tee connector on another device, or

• The BNC tee connector on the Ethernet hub (see Figure 4).

5. Repeat Steps 3 and 4 for each device to be connected.

6. Install 50 ohm terminator resistor plugs in the BNC tee connectors at each end of the BNC DX-LAN network.

Figure 4 shows connections for an example BNC DX-LAN network.

Figure 4. Example BNC DX-LAN Network

BNC Device

First BNC Device on Network

Last BNC Device on Network

Coaxial CableCoaxial Cable

10BASE-T Hub

TeeTee

or

Tee

Tee

Terminator Terminator

TerminatorNOTES:

Link the devices in a single line in any order.Install the PeakNet-PA workstationanywhere in the network.Install a terminator plug at each end of the network.

Page 15: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Connecting the DX-LAN Network

Doc. 031832-01 6/02 11

5.3 DX-LAN Network Connection ExamplesExample 1 (Figure 5) shows the DX-LAN connections for a dual-channel DX-800 with the following configuration:

• The PeakNet-PA workstation and AE80 enclosures have 10BASE-T (RJ-45) connectors

• In Channel 1, all devices have RJ-45 connectors

• In Channel 2, the CC80, pump, and detector have BNC connectors; the EG40-PA eluent generator has an RJ-45 connector

Example 2 (Figure 6) shows the DX-LAN connections for a dual-channel DX-800 with the following configuration:

• The PeakNet-PA workstation has a 10BASE-T (RJ-45) connector

• The AE80 enclosures have BNC DX-LAN connectors (In and Out)

• In Channel 1 and Channel 2, the CC80s, pumps, and detectors have BNC connectors; the EG40-PA eluent generators have RJ-45 connectors

Example 3 (Figure 7) shows the DX-LAN connections for a single-channel DX-800 with the following configuration:

• All devices have 10BASE-T (RJ-45) connectors

Example 4 (Figure 8) shows the DX-LAN connections for a dual-channel DX-800 with the following configuration:

• The PeakNet-PA workstation and AE80 enclosures have BNC connectors

• In Channel 1 and Channel 2, the CC80s, pumps, and detectors have BNC connectors; the EG40-PA eluent generators have RJ-45 connectors

Page 16: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Installing the PeakNet-PA System

12 Doc. 031832-01 6/02

Figure 5. DX-LAN Network Connections:Example 1

Pump10BASE-T

Workstation

Normal/Uplink=Normal

Hub

8 7 6 5 4 3 2 1

10BASE-T

Hub

8 7 6 5 4 3 2 1

CC8010BASE-T

Detector10BASE-T

EG40-PA10BASE-T

PumpBNC

Hub

8 7 6 5 4 3 2 1

CC80BNC

DetectorBNC

EG40-PA10BASE-T

Channel 1 AE80 Enclosure

Normal/Uplink =Uplink

BNC Tee

BNC Tee

BNC Tee

BNC Terminator

BNC Terminator

RJ-45 RJ-45

RJ-45

RJ-45

RJ-45

RJ-45

RJ-45 RJ-45

Channel 2 AE80 Enclosure

Normal/Uplink =Uplink

NOTES

Use 10BASE-T DX-LAN Category 5 unshielded twisted-pair cables for all connections between RJ-45 connectors on the hub and the 10BASE-T devices.

Use BNC DX-LAN 50 ohm coaxial impedance cables (“RG-58U”) for all connections between BNC devices (including the BNC connector on the hub).

Connect a 50 ohm terminating resistor at each end of the BNC DX-LAN network.

Page 17: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Connecting the DX-LAN Network

Doc. 031832-01 6/02 13

Figure 6. DX-LAN Network Connections:Example 2

Hub

7 6 5 4 3 2 1 BNC Terminator

BNC Terminator

PumpBNC

Hub

8 7 6 5 4 3 2 1

CC80BNC

DetectorBNC

EG40-PA10BASE-T

Normal/Uplink = Normal

BNC Tee

BNC Tee

BNC Tee

BNC IN

BNCOUT

PumpBNC

Hub

8 7 6 5 4 3 2 1

CC80BNC

DetectorBNC

EG40-PA10BASE-T

BNC Tee

BNC Tee

BNC Tee

Normal/Uplink= Normal

BNC IN

BNCOUT

NOTES

Use 10BASE-T DX-LAN Category 5 unshielded twisted-pair cables for all connections between RJ-45 connectors on the hub and the 10BASE-T devices.

Use BNC DX-LAN 50 ohm coaxial impedance cables ( ) for all connections between BNC devices (including the BNC connector on the hub).

Connect a 50 ohm terminating resistor at each end of the BNC DX-LAN network.

“RG-58U”

Workstation

Normal/Uplink=Normal 8

10BASE-T RJ-45

Channel 1 AE80 Enclosure

Channel 2 AE80 Enclosure

RJ-45 RJ-45

Page 18: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Installing the PeakNet-PA System

14 Doc. 031832-01 6/02

Figure 7. DX-LAN Network Connections:Example 3

NOTE

Use 10BASE-T DX-LAN Category 5 unshielded twisted-pair cables for all connections between RJ-45 connectors on the hub and the 10BASE-T devices.

Pump10BASE-T

Workstation10BASE-T

Hub

8 7 6 5 4 3 2 1

CC8010BASE-T

Detector10BASE-T

EG40-PA10BASE-T

Channel 1 AE80 Enclosure

Normal/Uplink =Normal

RJ-45

RJ-45

RJ-45

RJ-45

RJ-45

RJ-45

Page 19: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Connecting the DX-LAN Network

Doc. 031832-01 6/02 15

Figure 8. DX-LAN Network Connections:Example 4

BNC Terminator

PumpBNC

Hub

8 7 6 5 4 3 2 1

CC80BNC

DetectorBNC

EG40-PA10BASE-T

Normal/Uplink = Normal

BNC Tee

BNC Tee

BNC Tee

BNC IN

BNCOUT

PumpBNC

Hub

8 7 6 5 4 3 2 1

CC80BNC

DetectorBNC

EG40-PA10BASE-T

BNC Tee

BNC Tee

BNC Tee

Normal/Uplink= Normal

BNC IN

BNCOUT

NOTES

Use BNC DX-LAN 50 ohm coaxial impedance cables ( ) for all connections between BNC devices (including the BNC connector on the hub).

Connect a 50 ohm terminating resistor at each end of the BNC DX-LAN network.

Use 10BASE-T DX-LAN Category 5 unshielded twisted-pair cables for all connections between the RJ-45 connectors on the hub and the 10BASE-T devices.

“RG-58U”

WorkstationBNC

Channel 1 AE80 Enclosure

Channel 2 AE80 Enclosure

RJ-45 RJ-45

BNC Terminator

Page 20: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Installing the PeakNet-PA System

16 Doc. 031832-01 6/02

6 Installing the DX-LAN Card Device Driver

1. After installing the DX-LAN card in the computer (see Section 4) and cabling the DX-LAN network (see Section 5), turn on at least one of the DX-800 Analyzer modules. Then, turn on the computer.

2. Log onto Windows NT as an administrator.

3. Open the Control Panel and double-click the Network icon.

• If the Network dialog box appears, go on to Step 4.

• If the Network Configuration message box appears, go on to “Installing Windows NT Networking and the DX-LAN Card Driver” on page 18.

4. In the Network dialog box, select the Adapters tab.

5. Click the Add button.

The Select Network Adapter dialog box appears and displays a list of network adapters.

6. Click the Have Disk button.

7. The Insert Disk dialog box appears. The default disk drive is A. If your disk drive is B, select the B drive.

8. Dionex provides a disk with drivers in the PeakNet-PA installation kit. Insert the disk into the drive and click OK.

9. In the list of network adapters, select the entry for the adapter you installed and click OK.

Do not install PeakNet-PA software until after both the DX-LANcard and the driver are installed.

In order to install the DX-LAN card driver and PeakNet-PA, youmust be logged onto NT with administrator privileges. Ifnecessary, ask the Windows NT administrator to grant youadministrator privileges temporarily. Administrator privileges arenot required to run PeakNet-PA.

Page 21: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Installing the DX-LAN Card Device Driver

Doc. 031832-01 6/02 17

10. When Windows NT has finished copying the required files, select the Bindings tab.

11. In the Show Bindings for box, select All Adapters.

12. In the list of adapters, double-click the entry for the adapter you installed

This displays a list of protocols for the adapter.

13. Disable all protocol entries except NetBEUI. To do this, select each entry and click the Disable button.

14. Click OK to close the dialog box.

15. When the system asks whether to restart Windows now, click Yes.

Page 22: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Installing the PeakNet-PA System

18 Doc. 031832-01 6/02

Installing Windows NT Networking and the DX-LAN Card Driver

1. Follow these instructions only if Windows NT Networking has not been installed on your computer and the Network Configuration message box appears after you double-click the Network icon in the Control Panel (see Section 6, Step 3). Click Yes to install Windows NT Networking now.

The Network Setup Wizard appears.

2. Accept the default setting, Wired to the network, and click Next.

3. Click the Select from list button.

The Select Network Adapter dialog box appears.

4. Click the Have Disk button.

5. The Insert Disk dialog box appears. The default disk drive is A. If your disk drive is B, select the B drive.

6. Dionex provides a disk with drivers in the PeakNet-PA installation kit. Insert the disk into the drive and click OK.

7. The Select OEM Option dialog box appears. Select the entry for the adapter you installed and click OK.

The selected adapter will appear in the Networks Adapter list in the Network Setup Wizard.

8. Click Next.

The Network Protocols list appears.

9. Select the NetBEUI protocol and clear all other options.

10. Continue following the Network Setup Wizard. Accept default settings for all the remaining steps.

11. When prompted, insert the Windows NT CD-ROM, enter the CD-ROM drive letter, and click OK.

12. When the dialog box appears informing you there is more than one hardware bus, accept the default settings and click OK.

13. Continue following the Network Setup Wizard. When you are finished, click Yes to restart the computer.

Page 23: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Installing PeakNet-PA Software

Doc. 031832-01 6/02 19

7 Installing PeakNet-PA Software

The PeakNet-PA software installation consists of three phases:

• Installation of Dionex PeakNet-PA programs

• Installation of Wonderware InTouch

• Installation of the InTouch Runtime license

7.1 Installing PeakNet-PA Programs1. Make sure that the drive on which you plan to install PeakNet-PA has

at least 20 MB of free space.

2. Insert the PeakNet-PA CD-ROM into the CD-ROM drive.

3. Click the Start button on the Windows taskbar and select Run.

4. Type [drive]:\SETUP and click OK.

For example, if the CD-ROM drive is drive D, enter D:\SETUP.

5. SETUP will now guide you through the installation procedure; follow the on-screen instructions as they appear. When asked to select components to install, verify that the Example Database (Microsoft Access format) option is selected.

When PeakNet-PA installation is complete, the following dialog box appears.

Page 24: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Installing the PeakNet-PA System

20 Doc. 031832-01 6/02

6. Click Finish and go on to Section 7.2.

Page 25: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Installing PeakNet-PA Software

Doc. 031832-01 6/02 21

7.2 Installing Wonderware InTouch (Runtime)1. Insert the InTouch 6.0b (P/N 06-649) CD-ROM into the CD-ROM

drive.

2. Click the Start button on the Windows taskbar and select Run.

3. Type [drive]:\SETUP (for example, D:\SETUP) and click OK.

4. When the following dialog box appears, select Next> to install InTouch in the default directory, C:\InTouch.32.

Page 26: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Installing the PeakNet-PA System

22 Doc. 031832-01 6/02

5. In the next dialog box, select the following for installation:

• InTouch 6.0b (the default selection)

• InTouch 6.0b SQL Access

Page 27: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Installing PeakNet-PA Software

Doc. 031832-01 6/02 23

6. Select Next> to go to the following dialog box. Select the Runtime Only option. This option installs the files necessary to run InTouch with PeakNet-PA. The other two options are not used with PeakNet-PA.

Page 28: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Installing the PeakNet-PA System

24 Doc. 031832-01 6/02

7. Select Next> to go to the following dialog box and verify that the following options are entered:

• Program Folders: InTouch for Windows NT

• Existing Folders: Administrative Tools (Common)

Page 29: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Installing PeakNet-PA Software

Doc. 031832-01 6/02 25

8. Select Next> and then review the selected settings.

9. Select Next> to continue the installation. When installation is finished, check the View the Release Notes now box and click Finish.

Page 30: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Installing the PeakNet-PA System

26 Doc. 031832-01 6/02

7.3 Installation of the Runtime LicenseBefore you can use InTouch, a runtime “soft key” license must be installed. The license is installed from the E&M-InTouch-Runtime-2000Tag/100Window disk (P/N 01-797). The disk is serialized and can be installed only once.

1. Insert the Runtime license disk into the disk drive.

2. Click the Start button on the Windows taskbar and select Run.

3. Type [drive]:\SETUP (for example, A:\SETUP) and click OK.

4. The setup program will now guide you through the installation procedure; follow the on-screen instructions as they appear.

5. When the following dialog box appears, select InTouch in the Components list.

6. Continue with the setup procedure. When you are finished, review the release notes.

Page 31: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Setting Up the Database

Doc. 031832-01 6/02 27

8 Setting Up the Database

After each run, PeakNet-PA automatically exports sample data to an ODBC-supported database. The database information is then reported to the InTouch program where it is displayed in the Sample Results and the Component Trending screens. The data is also used for printing scheduled reports.

For correct operation of PeakNet-PA, a database program that supports ODBC (for example, Microsoft Access) must be installed on the PeakNet-PA workstation. In addition, PeakNet-PA must be set up as a data source for the database.

If Access is not already installed on the workstation, run the Microsoft Office setup and select the following options:

• Microsoft Access

• Converters, Filters, and Data Access

If you plan to use another ODBC-supported database, follow the installation instructions supplied with that database.

After the database and ODBC driver are installed, an ODBC icon will appear in the Windows Control Panel.

Optimization Note:

The PeakNet-PA database file (peaknet-pa.mdb) is copied to the c:\peaknet\database folder during PeakNet-PA setup. This database file can become quite large over time. If a drive with more free space than the C: drive is available, move peaknet-pa.mdb to the alternate drive before completing the following data source setup procedure.

For example, if the computer has a C: drive with 1GB of free space and a D: drive with 6GB of free space, move peaknet-pa.mdb to a folder on the D: drive (for example, D:\PNdatabase).

You can also specify an alternate drive for storing data files. See “MainMenu Program - Selecting Options” in the online Help for details.

Page 32: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Installing the PeakNet-PA System

28 Doc. 031832-01 6/02

Setting Up PeakNet-PA as a Data Source

The following instructions describe the setup procedure required to use the PeakNet-PA database with Microsoft Access. To setup other database programs, use these steps as a guideline and also refer to the Microsoft ODBC Help.

1. Open the Windows Control Panel. (Click the Start button on the taskbar, select Settings, and click Control Panel.)

2. In the Control Panel, double-click the ODBC icon.

NOTE If the ODBC icon does not appear, rerun the MicrosoftAccess setup to install the ODBC driver.

The ODBC Data Source Administrator dialog box appears.

3. Click the Add button.

The Create New Data Source dialog box appears.

Page 33: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Setting Up the Database

Doc. 031832-01 6/02 29

4. Select the Microsoft Access Driver (*.mdb) and click Finish.

The ODBC Microsoft Access Setup dialog box appears.

5. In the Data Source Name box, enter PeakNet-PA.

6. In the Description box, enter For the PeakNet-PA database, or any other desired description.

7. Under Database, click the Select button.

The Select Database dialog box appears.

Page 34: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Installing the PeakNet-PA System

30 Doc. 031832-01 6/02

8. Navigate to the location of the PeakNet-PA database. The default location is c:\peaknet\database. In the example shown in the previous figure, the database was moved to an alternate drive and folder (d:\peaknet\database). See “Optimization Note:” on page 27.

NOTE The Example Database (Microsoft Access format)option must be selected during PeakNet-PA setup (seeSection 7.1). If this option was not selected and thedatabase (peaknet-pa.mdb) was not installed, rerunPeakNet-PA setup, selecting only the database option.

9. Select peaknet-pa.mdb. This is the default database file provided with PeakNet-PA.

10. Click OK.

9 Setting Up InTouch to Start Automatically

Follow these steps to set up InTouch to start automatically when the PeakNet-PA Analyze program starts.

1. Start InTouch by clicking the Start button on the Windows taskbar and selecting Programs>InTouch for Windows NT>InTouch.

You can also start InTouch from Windows Explorer by double-clicking INTOUCH.EXE in the InTouch.32 directory.

The Wonderware welcome page appears.

2. Select Continue.

Page 35: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Setting Up InTouch to Start Automatically

Doc. 031832-01 6/02 31

The Application Directory List dialog box appears.

3. Select Search and use the drives and the directory lists to find the \DIONEXWW directory.

4. Click OK.

Page 36: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Installing the PeakNet-PA System

32 Doc. 031832-01 6/02

The \DIONEXWW directory appears at the top of the InTouch Application Directories list.

5. Click Done.

The InTouch Analyzer Status screen will now start automatically when the Analyze program starts.

6. Confirm that you are able to access the InTouch Analyzer Status screen by highlighting C:\DIONEXWW\ in the InTouch dialog box and then selecting the Window Viewer icon (on the right side of the dialog box).

Page 37: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Setting Up InTouch to Start Automatically

Doc. 031832-01 6/02 33

7. When prompted to run AnalyzeDX.exe (the Analyze program), click Cancel. Close the InTouch dialog box.

Window Viewer Icon

Page 38: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Installing the PeakNet-PA System

34 Doc. 031832-01 6/02

10 Setting the Time and Date Formats

To ensure a correct (and unique) time stamp for each injection, the following time and date formats must be selected for the PC:

• The (GMT) Casablanca, Monrovia time zone (it does not use daylight savings time)

• The HH:mm:ss (24-hour) time format

• The MM/dd/yy date format

1. To select the time zone, open the Control Panel and double-click the Date/Time icon; select the Time Zone tab and select (GMT) Casablanca, Monrovia.

2. Select the Date & Time tab and set the current date and time.

3. To select the 24-hour time format, go to the Control Panel and double-click the Regional Settings icon. Select the Time tab and select HH:mm:ss from the Time Style list.

4. To set the date format, select the Date tab in the Regional Settings dialog box and select MM/dd/yy from the Short date style list.

Page 39: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Selecting a Video Resolution (Optional)

Doc. 031832-01 6/02 35

11 Selecting a Video Resolution (Optional)

Dionex recommends 800 x 600 video resolution for PeakNet-PA. In addition, the InTouch application is optimized to run at that resolution. However, if your computer is set to a different resolution and you want InTouch to run at that resolution, follow the steps below to convert InTouch to your computer’s resolution:

1. Start InTouch by clicking the Start button on the Windows taskbar and selecting Programs>InTouch for Windows NT>InTouch; or, from Windows NT Explorer, double-click INTOUCH.EXE in the InTouch.32 directory.

The InTouch dialog box appears.

2. Click the button at the bottom of the dialog box that depicts a computer.

The Node Configuration dialog box appears.

Click to change the video resolution

Page 40: For PeakNet-PA Version 5 - labservice.co.krA DX-LAN network can include 10BASE-T devices only, BNC devices only, or a combination of 10BASE-T and BNC devices. The installation procedure

Installing the PeakNet-PA System

36 Doc. 031832-01 6/02

3. Select the Resolution tab.

4. Select the Allow WindowViewer to dynamically change resolution check box

5. Select the Convert to screen video resolution option.

6. Click OK.

The InTouch dialog box will be displayed.

7. Select Cancel.