FMT-XCT – Workpackage 2 advancement
description
Transcript of FMT-XCT – Workpackage 2 advancement
FMT-XCT – Workpackage 2 advancement
Marco Brambilla Véronique Rebuffel
Markus MronzMarek Karolczak
FMT-XCT Year 2 meeting 26.03.2010 – Heraklion, Crete
Outline
• Work-package 2 advancement summary• X-ray CT system design• Dual-energy X-ray protocol• Scattered radiation estimation• Absorbed dose estimation• Conclusions• (XPCI proposal)
M.Brambilla - FMT-XCT year 2 Meeting – 26.03.2010, Crete
WP 2 Objectives
• provide adequate accommodation of the optical components,
• eliminate X-ray interference with optical components,
• offer improved contrast between organs as is important for optimal use of XCT information as priors in the FMT inversion procedure (WP4).
Hybrid imaging system for small animal
XCT FMT
XCT module should:
Enhanced contrast XCT strategies:
• Dual-energy XCT
• Use of X-ray contrast agents
No Name M.M.
2 CEA-LETI (lead) 51.7
7 CT-Imaging 6
5 FIHGM 18
6 UZH 1
1 HMGU 1
M.Brambilla - FMT-XCT year 2 Meeting – 26.03.2010, Crete
WP 2 Deliverables: current stateDeliverables:- D2.1 (mo.1): XCT design - completed
Report “WP2: XCT design”, 11/2008, issued by CEA-LETI and CT-Imagingdistributed to the Consortium.
- D2.2 & 2.3 (mo.15): Dual energy prototype & software - completedPrototype and software demonstrated during Training session (06/2009).Report ready to be delivered.
- D2.4 (mo.18): Preliminary technical specifications - completedBased on LETI lab bench. Report ready to be delivered
- D2.5 (mo.21): Measurements of scattered energy - partially completedBased on LETI lab bench only. Draft report prepared.
- D2.6 (mo.24): Optimal contrast-enhancing strategy - delayed- D2.7 (mo.24): Optimized system design - almost completed
Update from D2.1
M.Brambilla - FMT-XCT year 2 Meeting – 26.03.2010, Crete
D2.1
MM
D2.4
D2.5
D2.2,
2.3
D2.6,
2.7
Milestone (n°5) : selection of an appropriate XCT technology for FMT-XCT system (by the executive committee)Month 18: preliminary recommendation / Month 24: official decision
WP 2 Problems
CEA-LETI bench is different from final prototype• more intense x-rays source• different environment
→ Repeat optimization process for FMT-XCT system
→ Repeat measurements (dose, contrast) on FMT-XCT system
Final prototype available only end of May 2010• Defect delivered gantry motor (still waiting !!!)• Problem with the PLC (re-design of the PCB of the PLC)• Time to construct the X-Ray shielding
→ Integrate gantry motor when delivered and finalize the prototype
→ Integration of the FMT chain (more than initially planned) has started at CT-Imaging
CEA-LETI bench not convenient for extensive tests using living mice• vertical position, fixation by forelegs• successive CTs, (Dual energy + with contrast agent), very difficult on a living
mouse
→ Tests planned for D2.6 require a more easily usable bench
M.Brambilla - FMT-XCT year 2 Meeting – 26.03.2010, Crete
WP2 Corrective actions and revised planning
M.Brambilla - FMT-XCT year 2 Meeting – 26.03.2010, Crete
Final prototype available
@ CT-Imaging
Mo.25 (04/2010)
D 2.5D 2.4 v2
Final prototype @ HMGU
D 2.6D 2.7 v2
M5
Scattering measurementsAdjust DE protocol
Mo.27 (06/2010)
TBDMo 24 (Now)
on test benches on final prototype at CTI on final prototype at HMGU
Comparison of contrast enhancement strategies
D2.5 (mo.21): Measurements of scattered energy → Update current preliminary report
D2.7 (mo.24): Optimized system design → Update current report integrating last changes
D2.6 (mo.24): Optimal contrast-enhancing strategy →
Comparison of on specified living mice.
Refine data analysis procedure and comparison protocolPossible experimental approaches:
→ DE experiments at CEA-LETI, contrast agent ones at FIHGM
→ All experiments at FIHGM
→ All experiments using the final prototype (required)
• Single energy CT• Dual energy CT • Contrast agent CT
Dimensions: 145 cm x 140 cm x 103 cm
Back view Front view
WP2 (2.7) / WP5 (5.3) / Milestone 8XCT final design, gantry development and shielding
M.Brambilla - FMT-XCT year 2 Meeting – 26.03.2010, Crete
WP2 (2.7) / WP5 (5.3) / Milestone 8XCT final design, gantry development and shielding
M.Brambilla - FMT-XCT year 2 Meeting – 26.03.2010, Crete
XCT 1) X-Ray tube: Oxford UltraBright 80W, 90kV, 13-20µm
2) Detector: Hamamatsu C7942, Csl scintil. + CMOS photo array, 2400 x 2400 pix, 9 f/s @ 4x4 binning
3) Beam modification: collimator, shutter, prefilter for dual energy mode
FMT 4) CCD camera: Princeton Instr. Pixis 1024
5) Filter wheel: FRM 65 (Owis)
6) Scattered X-Ray protection: CCD shutter and shielding
7) Laser positioning: KDT 105 and optical collimator
8) Stray light shielding: PEAK tube with openings for optical path
1
2
3
45
6
7
8
WP2 (2.7) / WP5 (5.3) / Milestone 8XCT final design, gantry development and shielding
M.Brambilla - FMT-XCT year 2 Meeting – 26.03.2010, Crete
XCT acquistion PC
FMT acquistion PC
FMT Laser stage control
PLC Master
Animal bed
Gantry motor
Gantry position
InterlockIO‘s
PLC Slave
X-Ray tube
Illumination
X-Ray shutter
CCD shutter
X-Ray prefilter
CDD filter wheel
Detector CCD camera
Network
Block diagram
of the main components
Task 2.2: CT bench at CEA-LETI
Versatile tool, in order to help Dual Energy development and validation.
M.Brambilla - FMT-XCT year 2 Meeting – 26.03.2010, Crete
• Results:
• LE:• Generator: 40kV-6mA• Filter: 60um Sn• 2.4 mAs
• HE:• Generator: 70kV-2mA• Filter: 100um Pb• 0.8 mAs
Dual-energy protocol for FMT-XCT
Energy couple determination
• Multi-parametric problem
• x-rays generator kV and mA
• x-rays filters
• duration of irradiation for LE and HE acquisitions
• By means of a simulation software:
• simulated all the dual energy chain trying different combinations of the parameters
• find the combination with best figure of merit
• Tried different FOMs and different levels of simulation accuracy
• Validate results with phantom and animal study
M.Brambilla - FMT-XCT year 2 Meeting – 26.03.2010, Crete
DE protocol validation (1/2)
Observations:
• Little contrast between organs
• Contrast between soft tissues/adipose tissue
• High level of noise (mainly in HE)
• Same type of contrast in LE/HE
M.Brambilla - FMT-XCT year 2 Meeting – 26.03.2010, Crete
DE protocol validation (2/2)
Mou
se 1
Mou
se 2
Mou
se 3
to distinguish adipose from other soft tissues
Feasible:
Easy:
to distinguish bones from soft tissues
Very difficult:
to identify organs
High energy does not add significant information
(unless very high dose@ both energies)
M.Brambilla - FMT-XCT year 2 Meeting – 26.03.2010, Crete
Beam hardening estimation
• Stronger attenuation of lower part of the spectrum
→ (in radiographs) attenuation not linear any more against object thickness→ (in tomography) cupping artifacts
Radiography: → good linearity at both LE and HE
LE Tomography: → Negligible cupping
M.Brambilla - FMT-XCT year 2 Meeting – 26.03.2010, Crete
→ Very limited beam hardening influence
Absorbed dose
Low energy High energy
Dose rate [mGy/min]Single radiography
(300ms exposure) [mGy]
Full tomographic scan (3’20’’ duration)
[mGy]Dose rate [mGy/min]
Single radiography (300ms exposure)
[mGy]
Full tomographic scan (3’20’’ duration)
[mGy]
Abdomen 93.93 0.47 312.8 14.78 0.07 49.0Chest 65.40 0.33 217.8 10.25 0.05 34.1Mean 77.56 0.39 258.3 12.16 0.06 40.5
lIIIIa exp100
M.Brambilla - FMT-XCT year 2 Meeting – 26.03.2010, Crete
Scattered X-ray measure for CCD shielding
Scattered field depends on the bench
→ Not possible to estimate shielding geometry
→ Possible to estimate shielding thickness
mmcmcm
I
I
kVevPbt e
45.2245.049.24
)10log(
log)70,(
1
1
6
0
tkVevPbI
I e )70,(exp0
M.Brambilla - FMT-XCT year 2 Meeting – 26.03.2010, Crete
Parameter optimization for final prototype
• Optimization by simulation using datasheet parameters• Simulation prarameters:
• source:• Oxford Instr. UltraBright u-focus• 13um focal spot size• 20-90kV anodic voltage range• 2mA maximum anodic current• W anode
• detector:• Hamamatsu C7942 (same as CEA-LETI bench)
• geometry:• source-obj. dist.: 20cm• obj.-detector dist.: 20cm
Optimal configurationLow energy:
• 35kV• 2mA• 50um tin filter
High energy:• 70kV• 1.25mA• 50um lead filter
M.Brambilla - FMT-XCT year 2 Meeting – 26.03.2010, Crete
• Better contrast than CEA-LETI bench• To be verified experimentally on final prototype !
M.Brambilla - FMT-XCT year 2 Meeting – 26.03.2010, Crete
Conclusions and perspectives
• Prototype construction almost completed• Prototype testing planned• Dual energy protocol developed and tested (to be confirmed)
• Most relevant information (and dose) in low energy data
• Preliminary scattered radiation measurements performed• Optimal contrast enhancement strategy not yet defined
Conclusions
Perspectives
• Scarce information contribution from high energy data• Dual energy protocol not suggested• Low energy acquisition only
• We propose at no additional man-months:• Different X-ray contrast enhancement technique investigation→ X-rays phase contrast imaging
M.Brambilla - FMT-XCT year 2 Meeting – 26.03.2010, Crete
Thank you
WP 2 Remaining and corrective actions
- D2.5 (mo.21): Measurements of scattered energy - Month 27
- X-ray measurements on the final prototype (first experiments done by CT-Imaging)
- Optimization of the CCD camera shielding
- D2.7 (mo.24): Optimized system design - Month 27
- Update of the prepared report integrating last changes
- D2.6 (mo.24): Optimal contrast-enhancing strategy - Month 30 to be discussed
Comparison of on specified living mice.
Possible experimental approaches:
-> X-ray experiments at CEA-LETI, contrast agent ones at FIHGM
-> All experiments at FIHGM
-> All experiments using the final prototype (required)
Refine data analysis procedure and comparison protocol
• Single energy CT• Dual energy CT • Contrast agent CT
M.Brambilla - FMT-XCT year 2 Meeting – 26.03.2010, Crete