INTRODUCTIONTechnique for counting and
examining microscopic particles.
Particles are suspended in fluid.
Very powerful tool for analyzing multiple parameters of cells in a heterogeneous population
![Page 4: flow cytometry presentation](https://reader036.fdocuments.us/reader036/viewer/2022081511/587252e31a28ab852f8b79c5/html5/thumbnails/4.jpg)
![Page 5: flow cytometry presentation](https://reader036.fdocuments.us/reader036/viewer/2022081511/587252e31a28ab852f8b79c5/html5/thumbnails/5.jpg)
FACSCalibur
1-2 Lasers, 3-4 Colors
FACSCanto II
2-3 Lasers, 6-8 Colors Up to 8 Colors
LSR II
Up to 7 Lasers, 18 Colors
LSRFortessa
Up to 5 Lasers, 18 Colors
![Page 6: flow cytometry presentation](https://reader036.fdocuments.us/reader036/viewer/2022081511/587252e31a28ab852f8b79c5/html5/thumbnails/6.jpg)
Generaly:
Each cell is subjected to a laser beam.
The light reflected from each cell is captured.
Information is then interpreted statistically by a software
![Page 7: flow cytometry presentation](https://reader036.fdocuments.us/reader036/viewer/2022081511/587252e31a28ab852f8b79c5/html5/thumbnails/7.jpg)
HISTORY
Mark Fulwyler was the inventor of the forerunner of the modern flow cytometer.
This was developed in 1965
First fluorescence based flow cytometer was developed in 1986by Wolfgang Gohde.
![Page 8: flow cytometry presentation](https://reader036.fdocuments.us/reader036/viewer/2022081511/587252e31a28ab852f8b79c5/html5/thumbnails/8.jpg)
HISTORYThe flow cytometer was originally called
pulse cytophotometry.
In 1988, the name was officially changed to “flow cytometry” at the Conference of the American Engineering Foundation in Pensacola, Florida.
![Page 9: flow cytometry presentation](https://reader036.fdocuments.us/reader036/viewer/2022081511/587252e31a28ab852f8b79c5/html5/thumbnails/9.jpg)
THE INSTRUMENT
Modern flow cytometers can analyze several thousand particles every second.
It can also actively separate and isolate particles having specified property.
![Page 10: flow cytometry presentation](https://reader036.fdocuments.us/reader036/viewer/2022081511/587252e31a28ab852f8b79c5/html5/thumbnails/10.jpg)
COMPONENTSLiquid stream system: carries and aligns cells so that they pass
through a single file.
Measuring system: measure the impedence.
optical system: lamps, high power and low power lasers
Detector and analogue system
computer
![Page 11: flow cytometry presentation](https://reader036.fdocuments.us/reader036/viewer/2022081511/587252e31a28ab852f8b79c5/html5/thumbnails/11.jpg)
PRINCIPLEBeam of light of single wavelength is directed onto
hydrodynamically focused stream of liquid
Detectors are placed where the stream passed through the
light beam.
The light scattered by the cells are detected.
The amount of light scattered is measured.
This is analysed and results are interpreted.
![Page 12: flow cytometry presentation](https://reader036.fdocuments.us/reader036/viewer/2022081511/587252e31a28ab852f8b79c5/html5/thumbnails/12.jpg)
HYDRODYNAMIC FOCUSINGCells to be analyzed are suspended in liquid
and forced through a small aperture.A tube is used through which the sheath fluid
is pumped.Cells along with the fluid are forced through
this narrow aperture.Cells move in a single file or line.Laser hits each cell and data from each cell
can be read.
![Page 13: flow cytometry presentation](https://reader036.fdocuments.us/reader036/viewer/2022081511/587252e31a28ab852f8b79c5/html5/thumbnails/13.jpg)
1. Cells to be analyzed are suspended in liquid and forced through a small aperture.
2. A tube is used through which the sheath fluid is pumped.
3. Cells along with the fluid are forced through this narrow aperture.
4. Cells move in a single file or line.
5. Laser hits each cell and data from each cell can be read.
HYDRODYNAMIC FOCUSING
![Page 14: flow cytometry presentation](https://reader036.fdocuments.us/reader036/viewer/2022081511/587252e31a28ab852f8b79c5/html5/thumbnails/14.jpg)
SCATTERINGCells pass through the laser.
Light gets refracted or scattered in all angles.
2 types of scatter are analyzed:
Forward scatter: scatter in the forward direction
Side scatter: light scattered in very large angles.
![Page 15: flow cytometry presentation](https://reader036.fdocuments.us/reader036/viewer/2022081511/587252e31a28ab852f8b79c5/html5/thumbnails/15.jpg)
Principles of Flow Cytometer
Immunobiology Janeway et. al.
![Page 16: flow cytometry presentation](https://reader036.fdocuments.us/reader036/viewer/2022081511/587252e31a28ab852f8b79c5/html5/thumbnails/16.jpg)
FORWARD SCATTERForward scatter is the light that is scattered by the cell in
the forward direction.
Magnitude is proportional to the size of the cell.
The detector converts intensity of light into voltage or an electric pulse.
![Page 17: flow cytometry presentation](https://reader036.fdocuments.us/reader036/viewer/2022081511/587252e31a28ab852f8b79c5/html5/thumbnails/17.jpg)
![Page 18: flow cytometry presentation](https://reader036.fdocuments.us/reader036/viewer/2022081511/587252e31a28ab852f8b79c5/html5/thumbnails/18.jpg)
![Page 19: flow cytometry presentation](https://reader036.fdocuments.us/reader036/viewer/2022081511/587252e31a28ab852f8b79c5/html5/thumbnails/19.jpg)
FLOURESCENCE IN FLOW CYTOMETRY
Most common method of studying cellular characteristics.
Antibodies are tagged with a flourophore.
The antibody binds to cells.
When laser hits the flourophore, the molecule gets excited.
Signal is emitted which can be detected.
![Page 20: flow cytometry presentation](https://reader036.fdocuments.us/reader036/viewer/2022081511/587252e31a28ab852f8b79c5/html5/thumbnails/20.jpg)
Energy State of Fluorescence during Excitation and
Emission
http://probes.invitrogen.com/resources/education/tutorials/4Intro_Flow/player.html
![Page 21: flow cytometry presentation](https://reader036.fdocuments.us/reader036/viewer/2022081511/587252e31a28ab852f8b79c5/html5/thumbnails/21.jpg)
Commonly Used Fluorochromes for FACS Analysis
Fluorochromes Emission Max (nm)
Fluorescein isothiocyanate (FITC)
519 (green)
488
R-phycoerythrin (R-PE) 578 (orange-yellow) 488Propidium iodide (PI) 620 (red) 488
PerCP 678 (red) 488APC 660 (red) 633
Laser (nm)
APC-Cy7 785 (infra red) 633PerCP-Cy5.5 695 (far red) 488
![Page 22: flow cytometry presentation](https://reader036.fdocuments.us/reader036/viewer/2022081511/587252e31a28ab852f8b79c5/html5/thumbnails/22.jpg)
![Page 23: flow cytometry presentation](https://reader036.fdocuments.us/reader036/viewer/2022081511/587252e31a28ab852f8b79c5/html5/thumbnails/23.jpg)
DETECTION OF FLOURESCENCEFluorescent t light is travels the same path as
side scatter.The light is directed through a series of
filters and mirrors.Particular wavelength of light is detected by
the appropriate detector.
![Page 24: flow cytometry presentation](https://reader036.fdocuments.us/reader036/viewer/2022081511/587252e31a28ab852f8b79c5/html5/thumbnails/24.jpg)
![Page 25: flow cytometry presentation](https://reader036.fdocuments.us/reader036/viewer/2022081511/587252e31a28ab852f8b79c5/html5/thumbnails/25.jpg)
TWO COLOUR EXPERIMENTCells are labeled with two different fluorophores.The fluorophores must have compatible spectra.
Eg: AlexaFluor 488 ans Phycoerithrin (PE) have compatible spectra.
Both have an excitation peak at 480- 520 nm.The excitation peaks are far away so that discrete emission can be collected
![Page 26: flow cytometry presentation](https://reader036.fdocuments.us/reader036/viewer/2022081511/587252e31a28ab852f8b79c5/html5/thumbnails/26.jpg)
FLUORESCENCE ACTIVATED CELL SORTER(FACS)
It is a specialized form of cell sorting.
Provides a method for sorting heterogeneous mixture of biological material like cells based on light scattering and fluorescent characteristics.
![Page 27: flow cytometry presentation](https://reader036.fdocuments.us/reader036/viewer/2022081511/587252e31a28ab852f8b79c5/html5/thumbnails/27.jpg)
![Page 28: flow cytometry presentation](https://reader036.fdocuments.us/reader036/viewer/2022081511/587252e31a28ab852f8b79c5/html5/thumbnails/28.jpg)
![Page 29: flow cytometry presentation](https://reader036.fdocuments.us/reader036/viewer/2022081511/587252e31a28ab852f8b79c5/html5/thumbnails/29.jpg)
MEASURABLE PARAMETERSVolume and morphological complexity of cellsCell pigments such as chlorophyllCell cycle analysis, cell kineticsChromosomal analysisProtein modificationAntigensEnzymatic activityMembrane fluidity
![Page 30: flow cytometry presentation](https://reader036.fdocuments.us/reader036/viewer/2022081511/587252e31a28ab852f8b79c5/html5/thumbnails/30.jpg)
Terminology of Flow Cytometer
Operation• Laser excitation• PMT• Optical filtersData analysis• Electronic compensation
Curr. Protocol. Immunol. 2002 Chap. 5
![Page 31: flow cytometry presentation](https://reader036.fdocuments.us/reader036/viewer/2022081511/587252e31a28ab852f8b79c5/html5/thumbnails/31.jpg)
Laser Excitation
•488nm Blue Laser–Basic laser which is equipped in almost all Flow Cytometers–Generates FSC/SSC–Fluorochromes excited:
•FITC, Alexa488, GFP, CFSE•PE and PE tandems•PI•PerCP and PerCP tandems, 7-AAD
![Page 32: flow cytometry presentation](https://reader036.fdocuments.us/reader036/viewer/2022081511/587252e31a28ab852f8b79c5/html5/thumbnails/32.jpg)
Optical Filters Light Transmittance through Longpass, Shortpass
(Cutoff) and Bandpass (Center point) filters
www.bdbiosciences.com
![Page 33: flow cytometry presentation](https://reader036.fdocuments.us/reader036/viewer/2022081511/587252e31a28ab852f8b79c5/html5/thumbnails/33.jpg)
Emissions from fluorescent dyes bound to individual cells are detected by photomultiplier tubes (PMTs) which convert and multiply light signals (analog) as much as 100 million times into electronic signals (digital(
What is a Photomultiplier Tube (PMT)?
http://micro.magnet.fsu.edu/primer/digitalimaging/concepts/photomultipliers.html
![Page 34: flow cytometry presentation](https://reader036.fdocuments.us/reader036/viewer/2022081511/587252e31a28ab852f8b79c5/html5/thumbnails/34.jpg)
FSCLaser
FACS Instruments Generate Three Types of Data
Forward scatter (FSC) Approximate cell size
Side scatter (SSC) Cell complexity or granularity
Fluorescence To investigate cell structure and function
SSC
SSC
![Page 35: flow cytometry presentation](https://reader036.fdocuments.us/reader036/viewer/2022081511/587252e31a28ab852f8b79c5/html5/thumbnails/35.jpg)
Excitation and Emission
Use the maximum excitation wavelengths to determine lasers that can be used to excite the fluorochrome
Use the maximum emission wavelengths to determine filters and PMTs that can be used to measure the signal
![Page 36: flow cytometry presentation](https://reader036.fdocuments.us/reader036/viewer/2022081511/587252e31a28ab852f8b79c5/html5/thumbnails/36.jpg)
In order to properly analyze multicolor
flow cytometry experiments it is
necessary to employ a mechanism
called color compensation.
Specialized circuitry in the flow
cytometer is used to subtract a portion
of one detector's signal from another,
leaving only the desired signal. In the
above example, region A represents
unwanted FITC fluorescence appearing
in the FL2 detector.