Flexinivo ® FN 6 1 2 3 4 5 6 7 8 9 10 11 12 FN 6 gi010108 1 Adjustable point level switch Continuous level measurement system Technical information / Instruction manual page G Page Function G2 ------------------------------------------------------------------------------------------------------- Dimensions G3 ------------------------------------------------------------------------------------------------------- Technical data G4 ------------------------------------------------------------------------------------------------------- Electrical connection, switching logic G5 ------------------------------------------------------------------------------------------------------- Batch control or adjustable level detection G6 ------------------------------------------------------------------------------------------------------- Level measurement G8 ------------------------------------------------------------------------------------------------------- Manual motor operation G10 ------------------------------------------------------------------------------------------------------- Commissioning, mounting, safety instructions G11 Subject to technical change and price change. We assume no liability for typing errors. All dimensions in mm (inches). Different variations to those specified are possible. Please contact our technical consultants. All units of this information are CE - certified. Table of contents ING. ROLF HEUN | Meß- Prüf- Regeltechnik GmbH | Hufeisen 16 | 21218 Seevetal/Hittfeld Tel: 04105-5723-0 | Fax: 04105-5723-66 | [email protected] | www.heun-messtechnik.com

Transcript of Flexinivo FN 6 - heun-messtechnik.com€¦ · Technical information / Instruction manual ... Manual...

Page 1: Flexinivo FN 6 - heun-messtechnik.com€¦ · Technical information / Instruction manual ... Manual motor operation G10 ... page G . 1 ® FN 6 ...

Flexinivo® FN 6 1

2

3

4

5

6

7

8

9

10

11

12FN 6 gi010108 1

Adjustable point level switchContinuous level measurement systemTechnical information / Instruction manual

page G

Page

Function G2

-------------------------------------------------------------------------------------------------------

Dimensions G3

-------------------------------------------------------------------------------------------------------

Technical data G4

-------------------------------------------------------------------------------------------------------

Electrical connection, switching logic G5

-------------------------------------------------------------------------------------------------------

Batch control or adjustable level detection G6

-------------------------------------------------------------------------------------------------------

Level measurement G8

-------------------------------------------------------------------------------------------------------

Manual motor operation G10

-------------------------------------------------------------------------------------------------------

Commissioning, mounting, safety instructions G11

Subject to technical change and price change. We assume no liability for typing errors.

All dimensions in mm (inches). Different variations to those specified are possible.Please contact our technical consultants.

All units of this information are CE - certified.

Table of contents

ING. ROLF HEUN | Meß- Prüf- Regeltechnik GmbH | Hufeisen 16 | 21218 Seevetal/HittfeldTel: 04105-5723-0 | Fax: 04105-5723-66 | [email protected] | www.heun-messtechnik.com

Page 2: Flexinivo FN 6 - heun-messtechnik.com€¦ · Technical information / Instruction manual ... Manual motor operation G10 ... page G . 1 ® FN 6 ...

Flexinivo® FN 61

2

3

4

5

6

7

8

9

10

11

12FN 6gi0101082

Adjustable point level switchContinuous level measurement systemTechnical information / Instruction manual

page G

Function

The unit is designed to be connected to a PLC which controls the up and down movement of the sensor and evaluates the pulses from the incremental encoder. The PLC is not part of the delivery.

1. Batch control or adjustable level detection

The unit can be used to measure the presence or absence of bulk material on a predetermined level.

The most important steps of the measurement cycle are:

- Measurement starts always from the “upper stop position”. - The motor drives the sensor down. Fast mode can be used until the sensor is close to the desired level. The incremental encoder gives pulses (1 pulse per mm sensor movement). The pulses can be evaluated by the PLC. - When the sensor level is close to the desired level for batch control or full detection slow mode can be used to ensure accurate stop of the sensor. - The bulk material is filled into the container. - When the bulk material reaches the sensor, an output signal is actuated. - The sensor must now be driven up to the „upper stop position“ for reference purposes. - A new measurement can be started.

2. Level measurement

The unit can be used to measure the level of bulk material.

The function is as follows:

- Measurement starts always from the “upper stop position”. - The motor drives the sensor down. Fast or slow mode can be used depending on the desired accuracy. The Incremental encoder gives pulses (1 pulse per mm sensor movement). The pulses can be evaluated by the PLC. - When the sensor reaches material level, the motor is automatically stopped and an output signal is actuated. - The sensor must now be driven up to the „upper stop position“ for reference purposes . - A new measurement can be started.

ING. ROLF HEUN | Meß- Prüf- Regeltechnik GmbH | Hufeisen 16 | 21218 Seevetal/HittfeldTel: 04105-5723-0 | Fax: 04105-5723-66 | [email protected] | www.heun-messtechnik.com

Page 3: Flexinivo FN 6 - heun-messtechnik.com€¦ · Technical information / Instruction manual ... Manual motor operation G10 ... page G . 1 ® FN 6 ...

Flexinivo® FN 6 1

2

3

4

5

6

7

8

9

10

11

12FN 6 gi010108 3

Adjustable point level switchContinuous level measurement systemTechnical information / Instruction manual

page G

Dimensions

Sensor on „upper stop position“

Temperature-protection cover

ING. ROLF HEUN | Meß- Prüf- Regeltechnik GmbH | Hufeisen 16 | 21218 Seevetal/HittfeldTel: 04105-5723-0 | Fax: 04105-5723-66 | [email protected] | www.heun-messtechnik.com

Page 4: Flexinivo FN 6 - heun-messtechnik.com€¦ · Technical information / Instruction manual ... Manual motor operation G10 ... page G . 1 ® FN 6 ...

Flexinivo® FN 61

2

3

4

5

6

7

8

9

10

11

12FN 6gi0101084

Adjustable point level switchContinuous level measurement systemTechnical information / Instruction manual

page G

Mechanical data

Housing: aluminium RAL 5010 gentian blue

Enclosure: IP 66 to EN 60529

Process connection: flange similar to DN 100 PN16 aluminium, black

Overall weight: approx. 17kg Material in process: ribbon cable: PVC, high resistance vibrating fork: 1.4571 / 314 vibrating fork cover: PVC

Deviation of vertical mounting: max. 2° out of the vertical

Pressure connection: quick couppling including opposite part, for hose diameter 9mm; max. operation pressure: 0.2bar

Operating conditions

Incremental encoder: resolution: 1 pulse per mm sensor movement allover accuracy of measurement ca. 5mm

Accuracy of sensor: vibrating fork ca. 5 .. 20mm (depending on the application and the bulk material)

Measuring range: 600 .. max. 3800mm (see drawing) Sensor speed (motor): motor fast (up and downwards): ca. 80-180mm/s

motor slow (downwards): ca. 20-40mm/s

Process pressure: -0.3..+0.3 bar

Process and ambient 0°C .. 60°Ctemperature: -20°C .. 60°C with optional temperature-protection cover

technical data

Electrical data

Mains voltage: 230V 50-60Hz or 115V 50-60Hz both voltages +10% / -15%

Installed load: 130 VA (with heating)

Connection terminal: max. 2.5mm.

Cable entry: 2 x M25 x 1.5 + 1 closing element for cable diameter 9-14 mm 3x NPT 1/2“ conduit connection 3x NPT 3/4“ conduit connection

Incremental encoder: power supply: 10-30V DC, max. 70mA pulse output: A, B, N push-pull, max. 40mA load H-Level: > Supply voltage –2.5V L-Level: < 2.5V cable length: max. 100m

Signal outputs: „Vibrating fork signal », « Vibrating fork in upper stop position“: floating relais contact max. 250V AC, 2A, 500VA

Control Inputs: “motor up”, “motor down”, “motor fast/slow”: optocoupler 20-30V DC, max. 10mA each

Protection class: I

Heating: included, thermostat controlled

ING. ROLF HEUN | Meß- Prüf- Regeltechnik GmbH | Hufeisen 16 | 21218 Seevetal/HittfeldTel: 04105-5723-0 | Fax: 04105-5723-66 | [email protected] | www.heun-messtechnik.com

Page 5: Flexinivo FN 6 - heun-messtechnik.com€¦ · Technical information / Instruction manual ... Manual motor operation G10 ... page G . 1 ® FN 6 ...

Flexinivo® FN 6 1

2

3

4

5

6

7

8

9

10

11

12FN 6 gi010108 5

Adjustable point level switchContinuous level measurement systemTechnical information / Instruction manual

page G

Electrical connection / Switching logic

Inte

rnal

con

nect

ion

Heat

ing fast slow

down up

Mot

or fa

st

Mot

or s

low

Mot

or u

p/do

wn

Incrementalencoder

0V +24V (20..30V)Motor control input

from PLC

Power supply230V or 115V, obtain supply voltage on type plate

Powersupply

Pulseoutput

use shielded cable

Output Output

Vibr

atin

g fo

rksi

gnal

Vibr

atin

g fo

rk in

„up

per

stop

pos

ition

OV +10.

.30V

out A

out B

out N

Shie

ld

Pulse output diagram:

Shown when sensor moves upwards

When rotation of the incremental encoder changes direction the signal of A and B is inverted.

Switching logic:Vibrating fork signal

Switching logic:Vibrating fork in„upper stop position“

ING. ROLF HEUN | Meß- Prüf- Regeltechnik GmbH | Hufeisen 16 | 21218 Seevetal/HittfeldTel: 04105-5723-0 | Fax: 04105-5723-66 | [email protected] | www.heun-messtechnik.com

Page 6: Flexinivo FN 6 - heun-messtechnik.com€¦ · Technical information / Instruction manual ... Manual motor operation G10 ... page G . 1 ® FN 6 ...

Flexinivo® FN 61

2

3

4

5

6

7

8

9

10

11

12FN 6gi0101086

+

+

-85..250 VAC11..36 VDC24 VAC

0 V10..30 VDC

0 V

Adjustable point level switchContinuous level measurement systemTechnical information / Instruction manual

page G

+

+

-85..250 VAC11..36 VDC24 VAC

0 V10..30 VDC

0 V

Electrical connection for the pulse converter PAX I

Evaluation 0/4- 20 mA with pulse convertor PAX I

inte

rnal

con

nect

ion

heat

ing slow quick

off on

Mot

or q

uick

Mot

or s

low

Mot

or o

n/of

f

incremental encoder

0V +24V (20..30V)connection

for motor control

230V oder 115V, Check supply voltage on type plate

use sheathed cable

Output Output

Mea

sure

men

t sig

nal

vibr

atin

g se

nsor

Vibr

atio

nsso

nde

Vibr

atin

g se

nsor

Uppe

r sto

p

OV +10.

.30V

Outp

ut A

Outp

ut B

Outp

ut N

Shea

th

Reset with voltage10-30 VDC

Reset with potential free contact

alternative

Attention:Caution! Ensure that the counter is reset to zero, before starting a new measurement.

Current output0/4-20 mApotential free / aktivemax. 500 Ohm

PAX ICounter with analogue output card

alternativ

ING. ROLF HEUN | Meß- Prüf- Regeltechnik GmbH | Hufeisen 16 | 21218 Seevetal/HittfeldTel: 04105-5723-0 | Fax: 04105-5723-66 | [email protected] | www.heun-messtechnik.com

Page 7: Flexinivo FN 6 - heun-messtechnik.com€¦ · Technical information / Instruction manual ... Manual motor operation G10 ... page G . 1 ® FN 6 ...

Flexinivo® FN 6 1

2

3

4

5

6

7

8

9

10

11

12FN 6 gi010108 7

Adjustable point level switchContinuous level measurement systemTechnical information / Instruction manual

page G

Programming the pulse convertor PAX I

Programming the counter PAX I

1. Programming with either 0-20 mA or 4-20 mA (preset to 4-20 mA)

Keys DisplayPress PAR to start programming Pro

F1 ( press 8 times) Change from Pro to 8-ANA

PAR Type 4-20

To change to 0-20 mA press F2 change to type 0-20

Press F1 to return to 4-20 mA

To save press PAR first and then DSP actual measurement value

2. Programmierung Ablauflänge "L" - Wert Stromausgang

Upper display value AN-HI: Set to 000000 (This value must not be changed) 0 Impulse = 20 mA

Lower display value AN-LO: Can be set to 00xxxx xxxx Impulse = Run distance "L" in mm, which will give a current of 0/4 mA.

The lower display value can be set as follows:v

Keys DisplayPress PAR to activate Pro

F1 (press 8 times) Change from Pro to 8-AnA

PAR (press 3 times) Change from AN-LO to 001000

Use F1 or F2 to set the desired run length

To save press PAR first and then DSP

20 mA upper display valuet AN-HI 000000

0/4 mA lower display value AN-LO 00xxxx

ING. ROLF HEUN | Meß- Prüf- Regeltechnik GmbH | Hufeisen 16 | 21218 Seevetal/HittfeldTel: 04105-5723-0 | Fax: 04105-5723-66 | [email protected] | www.heun-messtechnik.com

Page 8: Flexinivo FN 6 - heun-messtechnik.com€¦ · Technical information / Instruction manual ... Manual motor operation G10 ... page G . 1 ® FN 6 ...

Flexinivo® FN 61

2

3

4

5

6

7

8

9

10

11

12FN 6gi0101088

Adjustable point level switchContinuous level measurement systemTechnical information / Instruction manual

page G

Batch control or adjustable level detection

down/up

fast/slow

L= 0

Vibrating fork in „upper stop position“

Terminals and diodes on pcb

Vibration forksignal

upper stopposition

0V +24V (20..30V)

Motor control inputs

Incremental encoder output

Vibrating fork driving downwards first fast mode and short before measuring point slow mode to bring vibrating fork precise to the desired level

fast mode

slow mode

Fast mode

L= 0

Slow mode

max. L ATTENTION: To avoid damaging the unit while driving down-wards, ensure that the vibrating fork stops at max. 3300 mm.

Vibrating fork works at the measuring level „L“, material can start to flow into the batch containerOpen valve

L= 0

max.L

ING. ROLF HEUN | Meß- Prüf- Regeltechnik GmbH | Hufeisen 16 | 21218 Seevetal/HittfeldTel: 04105-5723-0 | Fax: 04105-5723-66 | [email protected] | www.heun-messtechnik.com

Page 9: Flexinivo FN 6 - heun-messtechnik.com€¦ · Technical information / Instruction manual ... Manual motor operation G10 ... page G . 1 ® FN 6 ...

Flexinivo® FN 6 1

2

3

4

5

6

7

8

9

10

11

12FN 6 gi010108 9

Adjustable point level switchContinuous level measurement systemTechnical information / Instruction manual

page G

Batch control or adjustable level detection

Vibrating fork detecting material, level „L“ is reached

L= 0

Vibrating fork goes back up after completing the measurement

Fast mode

L= 0

Fast mode

max. L

Vibrating fork has reached the „upper stop position“

ATTENTION: Before starting next measurement, ensure that PLC counter is set to zero for reference

max. L

L=0

ING. ROLF HEUN | Meß- Prüf- Regeltechnik GmbH | Hufeisen 16 | 21218 Seevetal/HittfeldTel: 04105-5723-0 | Fax: 04105-5723-66 | [email protected] | www.heun-messtechnik.com

Page 10: Flexinivo FN 6 - heun-messtechnik.com€¦ · Technical information / Instruction manual ... Manual motor operation G10 ... page G . 1 ® FN 6 ...

Flexinivo® FN 61

2

3

4

5

6

7

8

9

10

11

12FN 6gi01010810

Adjustable point level switchContinuous level measurement systemTechnical information / Instruction manual

page G

Continuous level measurement

down/up

fast/slow

Vibrating fork in „upper stop position“

terminals and diodes on pcb

Vibration forksignal

upper stopposition

incremental encoder output

0V +24V (20..30V)

Motor control inputs

Vibrating fork driving downwards until fork is detecting (change between fast and slow mode is possible)

fast mode

slow mode

Fast mode

L= 0

Slow mode

max. L ATTENTION: To avoid damaging the unit while driving down-wards, ensure that vibrating fork stops at max. 3300 mm.

L= 0

max. L

When vibrating fork detects material the level „L“ is reached and the motor is stopped automatically

ING. ROLF HEUN | Meß- Prüf- Regeltechnik GmbH | Hufeisen 16 | 21218 Seevetal/HittfeldTel: 04105-5723-0 | Fax: 04105-5723-66 | [email protected] | www.heun-messtechnik.com

Page 11: Flexinivo FN 6 - heun-messtechnik.com€¦ · Technical information / Instruction manual ... Manual motor operation G10 ... page G . 1 ® FN 6 ...

Flexinivo® FN 6 1

2

3

4

5

6

7

8

9

10

11

12FN 6 gi010108 11

Adjustable point level switchContinuous level measurement systemTechnical information / Instruction manual

page G

Continuous level measurement

Vibrating fork driving upwards after finishing the measurement

Fast mode

L= 0

Vibrating fork reached the „upper stop position“

ATTENTION: Before starting next measurement, ensure that PLC counter is set to zero for new reference

L = 0

ING. ROLF HEUN | Meß- Prüf- Regeltechnik GmbH | Hufeisen 16 | 21218 Seevetal/HittfeldTel: 04105-5723-0 | Fax: 04105-5723-66 | [email protected] | www.heun-messtechnik.com

Page 12: Flexinivo FN 6 - heun-messtechnik.com€¦ · Technical information / Instruction manual ... Manual motor operation G10 ... page G . 1 ® FN 6 ...

Flexinivo® FN 61

2

3

4

5

6

7

8

9

10

11

12FN 6gi01010812

Adjustable point level switchContinuous level measurement systemTechnical information / Instruction manual

page G

Manual motor operation

Vibrating fork driving downwards while pushing the buttons

Vibrating fork driving upwardswhile pushing the button

push button

fast mode down fast mode up

slow mode down

push button

L= 0

max.L

L= 0

max.L

ING. ROLF HEUN | Meß- Prüf- Regeltechnik GmbH | Hufeisen 16 | 21218 Seevetal/HittfeldTel: 04105-5723-0 | Fax: 04105-5723-66 | [email protected] | www.heun-messtechnik.com

Page 13: Flexinivo FN 6 - heun-messtechnik.com€¦ · Technical information / Instruction manual ... Manual motor operation G10 ... page G . 1 ® FN 6 ...

Flexinivo® FN 6 1

2

3

4

5

6

7

8

9

10

11

12FN 6 gi010108 13

Adjustable point level switchContinuous level measurement systemTechnical information / Instruction manual

page G

Safety instructions / Commissioning / Mounting

Safety instructions

- Installation, maintenance and commissioning may be accomplished only by qualified technical personnel.

- The respectively valid installation instructions must be observed.

- For terminal connection of the device, the local regulations or VDE0100 (Regulations of German electrotechnical Engineers) must be observed.

- Use a fuse for the mains voltage (max. 6A).

- Provide protection for relay contacts and output transistors to protect the device against spikes with inductive loads.

- Compare the mains voltage applied with the specifications given on the name plate before switching the device on.

- Make sure that max. 8mm of the pigtails are bared (danger of contact with live parts).

- Make sure that the boots for protecting cable terminations are not longer than 8mm (danger of contact with live parts).

- Make sure that the screwed cable gland safely seals the cable and that it is tight (danger of water intrusion).

- A voltage-disconnecting switch must be provided near the device.

- In the case of a defect, the distribution voltage must automatically be cut off by a RCCB protection switch so as to protect the user of the device from indirect contact with dangerous electric tensions.

- In the case of inexpert handling or handling malpractice, the electric safety of the device cannot be guaranteed.

- Switch off the supply voltage before opening the device.

- Before opening the lid take care, that no dust deposits or whirlings are present.

Commissioning

Warning:In case of inexpert handling or handling malpractice the safety of the device cannot be guaranteed.

1. Connect the unit with mains voltage, evaluation units and steering units (see page G5).

2. Cable conduit fittings, which are not used, must be locked with a closing element.

3. Compare mains voltage and frequency with the on the type plate.

4. Switch mains voltage and PLC on.

5. Test the function of the unit, the PLC and the measurement functions. Ensure that the point level switch cannot be driven more than 3300mm downwards (to avoid damage of the unit).

6. The unit is now ready for work. Measurements can be started.

Mounting

The unit is mounted vertically with the flange on the silo. Avoid the point level switch to graze the socket (this causes wear of the cable).

The mounting position must be choosen carefully:

- cornices that might fall down may damage the point level switch or the rope. Observe proper distance from silo wall.

- filling of the silo might cover the point level switch with material (prevent measuring during filling or observe proper distance to infeed).

- upward and downward movement of the point level switch must not be obstructed, even if the point level switch oscillates; observe proper distance to stanchions and built- in fittings.

The electrical connections are made in accordance with the connection diagram. Make sure, that the cable in the screwed cable gland is seated tightly without fail.

Close both lids of the housing, to prevent entrance of water into the housing.

When the unit is used in the open, we recommend to use the temperature protection cover. It protects the unit against moisture, heat and cold. If the ambient temperature can drop to less than 0°C the use of a temperature-protection-cover is obligatory.

Take care that the sensor never drives through the socket into the „upper stop position“ to avoid damage to the unit (see relevant dimensions page G3).

Socket

Point level switch

Cornices

Distance