Final Report · 1 Test Article Test Article Name Plain Medical Surgical Mask Manufacturer...

13
Sanitation & Environment Technology Institute, Soochow University Address: 199 Ren-Ai Road, Suzhou Industrial Park, Suzhou, Jiangsu 215123, P. R. China Website: www.sudatest.com E-mail: [email protected] Direct: +86 512 65880038 Free: 400 107 8828 Final Report Report Number: SDWH-M202000871-1(E) In Vitro Cytotoxicity Test of Plain Medical Surgical Mask According to ISO 10993-5: 2009 Agar Diffusion Method Sponsor: Shenzhonghai Medical Inc. Address: 1F/3F Building 4B 2F/3F Building 3, No.8, Longtian 3rd RoadLongtian CommunityPingshan DistrictShenzhenChina

Transcript of Final Report · 1 Test Article Test Article Name Plain Medical Surgical Mask Manufacturer...

Page 1: Final Report · 1 Test Article Test Article Name Plain Medical Surgical Mask Manufacturer Shenzhonghai Medical Inc. Address 1F/3F Building 4 È B 2F/3F Building 3, No.8, Longtian

Sanitation & Environment Technology Institute, Soochow University

Address: 199 Ren-Ai Road, Suzhou Industrial Park, Suzhou, Jiangsu 215123, P. R. China

Website: www.sudatest.com E-mail: [email protected]

Direct: +86 512 65880038 Free: 400 107 8828

Final Report Report Number: SDWH-M202000871-1(E)

In Vitro Cytotoxicity Test of

Plain Medical Surgical Mask According to ISO 10993-5: 2009

Agar Diffusion Method

Sponsor: Shenzhonghai Medical Inc.

Address:

1F/3F Building 4,B 2F/3F Building 3, No.8, Longtian 3rd

Road,Longtian Community,Pingshan District,Shenzhen,China

Page 2: Final Report · 1 Test Article Test Article Name Plain Medical Surgical Mask Manufacturer Shenzhonghai Medical Inc. Address 1F/3F Building 4 È B 2F/3F Building 3, No.8, Longtian

Sanitation & Environment Technology Institute, Soochow University Report No.: SDWH-M202000871-1(E)

Page 2 of 13

Content

Supplementary Explanation ....................................................................................... 3

Quality Assurance Statement ...................................................................................... 4

GLP Compliance Statement ........................................................................................ 5

Verification Dates ......................................................................................................... 5

Summary ....................................................................................................................... 6

Test Report .................................................................................................................... 7

1 Purpose.................................................................................................................... 7

2 Reference ................................................................................................................ 7

3 Compliance ............................................................................................................. 7

4 Identification of Test and Control Articles .......................................................... 7

4.1 Test Article .......................................................................................................................... 7

4.2 Control Article..................................................................................................................... 8 4.2.1 Negative Control ...................................................................................................... 8 4.2.2 Positive Control ........................................................................................................ 8

5 Equipment and Reagents ...................................................................................... 8

5.1 Equipment ........................................................................................................................... 8

5.2 Reagents .............................................................................................................................. 8

6 Identification of Test System ................................................................................. 8

7 Justification of Test System and Route of Administration ................................. 9

8 Experimental Design .............................................................................................. 9

8.1 Preparation of Extracts ........................................................................................................ 9 8.1.1 Pretreatment ............................................................................................................. 9 8.1.2 Sample and Control Preparation .............................................................................. 9

8.2 Experimental Procedure ...................................................................................................... 9

8.3 Results ............................................................................................................................... 10

8.4 Evaluation Criteria ............................................................................................................ 10

9 Conclusion ............................................................................................................ 10

10 Record Storage ..................................................................................................... 10

11 Confidentiality Agreement .................................................................................. 10

12 Deviation Statement ............................................................................................. 10

Annex 1 Results ....................................................................................................... 11

Annex 2 Photograph of Test Article ...................................................................... 12

Annex 3 Information Provided by Sponsor .......................................................... 13

Page 3: Final Report · 1 Test Article Test Article Name Plain Medical Surgical Mask Manufacturer Shenzhonghai Medical Inc. Address 1F/3F Building 4 È B 2F/3F Building 3, No.8, Longtian

Sanitation & Environment Technology Institute, Soochow University Report No.: SDWH-M202000871-1(E)

Page 3 of 13

Supplementary Explanation

(1) Please apply for rechecking within 15 days of receiving the report if there are any

objections.

(2) Any erasure or without special inspection and testing seal renders the report null

and void.

(3) The report is only valid when signed by the persons who edited, checked and

approved it.

(4) The results relate only to the articles tested.

(5) The report shall not be reproduced except in full without the written approval of

the institute.

Page 4: Final Report · 1 Test Article Test Article Name Plain Medical Surgical Mask Manufacturer Shenzhonghai Medical Inc. Address 1F/3F Building 4 È B 2F/3F Building 3, No.8, Longtian

Sanitation & Environment Technology Institute, Soochow University Report No.: SDWH-M202000871-1(E)

Quality Assurance Unit:

Quality Assurance Date

Page 4 of 13

Quality Assurance Statement

The Quality Assurance Unit inspected/audited this study in compliance with the

following GLP regulations:

Good Laboratory Practice (GLP) Regulation 21 CFR Part 58, U.S. Food and Drug

Administration (FDA). The laboratory is exempt from the following provisions: 21

CFR Part 58.105 Test and Control Article Characterization, and Part 58.113 Mixtures

of Articles with Carriers.

The Quality Assurance Unit conducted inspections on the following dates. The

findings were reported to the Study Director and to the Testing Facility Management.

The final report was reviewed by the Quality Assurance Unit. The final report

accurately describes the test methods in accordance with standard operating

procedures, and the results are consistent with raw data of non-clinical studies

conducted according to the study protocol.

Inspections Date of

Inspection

Date Reported to

Study Director

Date Reported to Testing

Facility Management.

Study Protocol 2020-03-31 2020-03-31

Study Procedure 2020-04-08 2020-04-08

Raw Data

Final Report

Page 5: Final Report · 1 Test Article Test Article Name Plain Medical Surgical Mask Manufacturer Shenzhonghai Medical Inc. Address 1F/3F Building 4 È B 2F/3F Building 3, No.8, Longtian

Sanitation & Environment Technology Institute, Soochow University Report No.: SDWH-M202000871-1(E)

Edited by:

Date

Reviewed by:

Study Director Date

Approved by:

Authorized Signatory Date

Sanitation & Environment Technology Institute, Soochow University

Page 5 of 13

GLP Compliance Statement

This study was fully in accordance with the technical requirements of the study

protocol.

This study was conducted in compliance with Good Laboratory Practice (GLP)

Regulation 21 CFR Part 58, U.S. Food and Drug Administration (FDA).

The laboratory is exempt from the following provisions: 21 CFR Part 58.105 Test and

Control Article Characterization, and Part 58.113 Mixtures of Articles with Carriers.

Verification Dates

Test Article Receipt 2020-03-26

Protocol Effective Date 2020-03-31

Technical Initiation Date 2020-03-31

Technical Completion Date 2020-04-10

Final Report Completion Date

Page 6: Final Report · 1 Test Article Test Article Name Plain Medical Surgical Mask Manufacturer Shenzhonghai Medical Inc. Address 1F/3F Building 4 È B 2F/3F Building 3, No.8, Longtian

Sanitation & Environment Technology Institute, Soochow University Report No.: SDWH-M202000871-1(E)

Page 6 of 13

Summary

1 Test Article

Test Article Name Plain Medical Surgical Mask

Manufacturer Shenzhonghai Medical Inc.

Address 1F/3F Building 4,B 2F/3F Building 3, No.8, Longtian 3rd Road,Longtian

Community,Pingshan District,Shenzhen,China

Model 20190019

Lot/Batch 201910210100

2 Main Reference

ISO 10993-5: 2009 Biological evaluation of Medical Devices — Part 5: Tests for In Vitro

Cytotoxicity

3 Test Method

Potential toxicity of test article was evaluated using Agar Diffusion Method in accordance with

ISO 10993-5: 2009 Biological evaluation of Medical Devices — Part 5: Tests for In Vitro

Cytotoxicity.

Study protocol number: SDWH-PROTOCOL-GLP-M202000871-1.

4 Conclusion

Under the conditions of this study, the test article Plain Medical Surgical Mask did not show

potential toxicity to L929 cells.

Page 7: Final Report · 1 Test Article Test Article Name Plain Medical Surgical Mask Manufacturer Shenzhonghai Medical Inc. Address 1F/3F Building 4 È B 2F/3F Building 3, No.8, Longtian

Sanitation & Environment Technology Institute, Soochow University Report No.: SDWH-M202000871-1(E)

Page 7 of 13

Test Report

1 Purpose

The purpose of the test is to determine the biological reactivity of a mammalian cell culture

(mouse fibroblast L929 cells) in response to the test article.

2 Reference

ISO 10993-5: 2009 Biological evaluation of Medical Devices — Part 5: Tests for In Vitro

Cytotoxicity

ISO 10993-12: 2012 Biological evaluation of Medical Devices — Part 12: Sample preparation and

reference materials

3 Compliance

Good Laboratory Practice Regulations, 21 CFR, Part 58.

ISO/IEC 17025:2005 General requirements for the competence of testing and calibration

laboratories (CNAS—CL01 Accreditation criteria for the competence of testing and calibration

laboratories) China National Accreditation Service for Conformity Assessment LABORATORY

ACCREDITATION CERTIFICATE Registration No. CNAS L2954.

RB/T 214—2017 Competence assessment for inspection body and laboratory mandatory

approval—General requirements for inspection body and laboratory Certification and

Accreditation Administration of the People's Republic of China INSPECTION BODY AND

LABORATORY MANDATORY APPROVAL Certificate No. CMA 180015144061.

4 Identification of Test and Control Articles

4.1 Test Article

Test Article Name Plain Medical Surgical Mask

Manufacturer Shenzhonghai Medical Inc.

Address 1F/3F Building 4,B 2F/3F Building 3, No.8, Longtian 3rd Road,

Longtian Community,Pingshan District,Shenzhen,China

Test Article Initial State Not Sterilized

CAS Code Not supplied by sponsor (N/S)

Model 20190019

Size L

Lot/Batch 201910210100

Test Article Material Elastic Strip

Packaging Material PE plastic bag

Physical State Solid

Color White

Density N/S

Stability N/S

Solubility N/S

Storage Condition Room Temperature

Intended Clinical Use It is intended to be worn by operating room personnel or other

healthcare personnel to protect both the patient and healthcare

personnel from transfer of microorganisms, body fluids, and particulate

material. This mask is intended to be used as a single use disposable

procedure mask.

Page 8: Final Report · 1 Test Article Test Article Name Plain Medical Surgical Mask Manufacturer Shenzhonghai Medical Inc. Address 1F/3F Building 4 È B 2F/3F Building 3, No.8, Longtian

Sanitation & Environment Technology Institute, Soochow University Report No.: SDWH-M202000871-1(E)

Page 8 of 13

The information about the test article was supplied by the sponsor wherever applicable.

The Sponsor is responsible for all test article characterization data as specified in the GLP

regulations.

4.2 Control Article

4.2.1 Negative Control

Negative Control Article Name: High Density Polyethylene

Manufacturer: U.S. Pharmacopeial Convention (USP)

Size: 3 Strips

Lot/ Batch#: K0M357

Physical State: Solid

Color: White

Stability: Stable at room temperature

Storage Conditions: Room temperature

4.2.2 Positive Control

Positive Control Article Name: Natural Latex Gloves

Storage Condition: Room temperature

5 Equipment and Reagents

5.1 Equipment

Equipment Name Equipment Number Calibration Expire

Autoclave SDWH2204 2020-04-16

Steel straight scale SDWH463 2020-07-29

Electronic Balance SDWH2601 2020-06-19

Electronic Balance SDWH230 2020-05-04

CO2 Incubator SDWH021 2020-04-16

Inverted microscope SDWH037 2020-05-04

Clean bench SDWH454 2020-05-06

Refrigerator SDWH448

5.2 Reagents

Reagent Name Manufacturer LOT

FBS CORNING 35081006

MEM HyClone AE29441978

Trypsin GiBco 2048080

Penicillin, Streptomycin sulfate GiBco 2108968

PBS CORNING 05418005

NRU Sigma 031M1744V

2X MEM GiBco 2106493

Agar Beyotime ST004E

6 Identification of Test System

L929 mouse fibroblast cells obtained from ATCC (American Type Culture Collection), USA.

Page 9: Final Report · 1 Test Article Test Article Name Plain Medical Surgical Mask Manufacturer Shenzhonghai Medical Inc. Address 1F/3F Building 4 È B 2F/3F Building 3, No.8, Longtian

Sanitation & Environment Technology Institute, Soochow University Report No.: SDWH-M202000871-1(E)

Page 9 of 13

7 Justification of Test System and Route of Administration

Historically, mouse fibroblast L929 cell has been used for cytotoxicity studies because of its

cytotoxic response sensitivity.

In vitro mammalian cell culture study has been used to historically evaluate the cytotoxicity of

biomaterial of medical device.

8 Experimental Design

8.1 Preparation of Extracts

8.1.1 Pretreatment

8.1.1.1 Sterilization for test samples

Autoclaving at 121°C for 30 min

8.1.1.2 Sterilization for control samples

Same as the test sample.

8.1.2 Sample and Control Preparation

Test group: The test article (see the image) will be aseptically clipped into a length of 10 mm,

saturate the sample with MEM;

Negative control group: Aseptically prepare 10 mm×10 mm negative control articles with a flat

surface.To ensure adequate contact with the agar overlay;

Positive control group: Aseptically prepare 10 mm×10 mm positive control articles with a flat

surface.To ensure adequate contact with the agar overlay.

8.2 Experimental Procedure

Aseptic procedures were used for handling cell cultures.

L929 cells were cultured in MEM medium (10% FBS, Penicillin 100 U/mL, Streptomycin sulfate

100 µg/mL) at 37°C in a humidified atmosphere of 5% CO2 and then digested by 0.5% trypsin

containing EDTA to obtain a 1.0105 cells/mL suspension.

The suspended cells were dispensed at 2mL per well in 6-well plate (Φ34.8 mm), and culture it in

cell incubator (5% CO2, 37°C,> 90% humidity) for 24 h.Cell morphology was evaluated to verify

that the monolayer was satisfactory.

After the cells grew to form a monolayer, original culture medium was discarded. The 6 well

plates were then added with 2 mL of agar medium (3% agar solution was mixed with 2xMEM

containing 20% FBS by equal volume). Place the prepared articles in each plate in contact with

the agar surface. Incubate all cultures for 48 h. Three replicates of each group were tested.

After 48 h incubation, mark the outline of the specimen on the bottom of the culture dish with a

permanent marker, and then remove the specimen. Add 2 mL of neutral red solution to each plate

and incubate for 1 h.Pour off the neutral red solution and examine each culture microscopically

under and around each control and test article. Assess changes in, for example, general

morphology, vacuolization, detachment, cell lysis and membrane integrity.

A useful way to grade test article is given in Table 1.

Table 1 Reactivity grades for Elution Test

Grade Reactivity Description of reactivity zone

0 None No detectable zone around or under specimen

1 Slight Some malformed or degenerated cells under specimen

2 Mild Zone limited to area under specimen

3 Moderate Zone extending specimen size up to 10 mm

4 Severe Zone extending farther than 10 mm beyond specimen

Page 10: Final Report · 1 Test Article Test Article Name Plain Medical Surgical Mask Manufacturer Shenzhonghai Medical Inc. Address 1F/3F Building 4 È B 2F/3F Building 3, No.8, Longtian

Sanitation & Environment Technology Institute, Soochow University Report No.: SDWH-M202000871-1(E)

Page 10 of 13

8.3 Results

The test article was placed on the agar surface to evaluate the cytotoxic properties with

growing-well L929 cell, after incubating at 37°C in 5% CO2 for 48 h to determine the potential

cytotoxicity. Intracytoplasmic granulation and cell lysis were observed. The agar diffusion test

method results showed that the reactivity grades of test article was 2 and the results of control

groups showed the test was valid.See Appendix I, table 2 and table 3 for specific results.

8.4 Evaluation Criteria

The achievement of a numerical grade greater than 2, based on Tables 1, was considered

cytotoxic.

9 Conclusion

Under the conditions of this study, the test article Plain Medical Surgical Mask did not show

potential toxicity to L929 cells.

10 Record Storage

All raw data pertaining to this study and a copy of the final report are to be retained in designated

SDWH archive.

11 Confidentiality Agreement

Statements of confidentiality were as agreed upon prior to study initiation.

12 Deviation Statement

There were no deviations from the approved study protocol which were judged to have any impact

on the validity of the data.

Page 11: Final Report · 1 Test Article Test Article Name Plain Medical Surgical Mask Manufacturer Shenzhonghai Medical Inc. Address 1F/3F Building 4 È B 2F/3F Building 3, No.8, Longtian

Sanitation & Environment Technology Institute, Soochow University Report No.: SDWH-M202000871-1(E)

Page 11 of 13

Annex 1 Results Table 2 Observation of the Cell morphology

Group Before

inoculation

Before treated

with extract 48h after treatment

Test article Discrete

intracytoplasmic

granules; no cell

lysis,no reduction

of cell growth

Discrete

intracytoplasmic

granules; no cell

lysis,no reduction

of cell growth

Zone limited to area under specimen

Negative

control

No detectable zone around or under

specimen.

Positive

control

Zone extending farther than 10 mm

beyond specimen.

Normal: Discrete intracytoplasmatic granules, no cell lysis, no reduction of cell growth.

Table3 Cell reactivity

Group Parallel 1 Parallel 2 Parallel3 Average Interpretation

Test Group 2 2 2 2 Mildly

cytotoxic

Negative

Control 0 0 0 0 None cytotoxic

Positive

Control 4 4 4 4

Severely

cytotoxic

Page 12: Final Report · 1 Test Article Test Article Name Plain Medical Surgical Mask Manufacturer Shenzhonghai Medical Inc. Address 1F/3F Building 4 È B 2F/3F Building 3, No.8, Longtian

Sanitation & Environment Technology Institute, Soochow University Report No.: SDWH-M202000871-1(E)

Page 12 of 13

Annex 2 Photograph of Test Article

Page 13: Final Report · 1 Test Article Test Article Name Plain Medical Surgical Mask Manufacturer Shenzhonghai Medical Inc. Address 1F/3F Building 4 È B 2F/3F Building 3, No.8, Longtian

Sanitation & Environment Technology Institute, Soochow University Report No.: SDWH-M202000871-1(E)

Page 13 of 13

Annex 3 Information Provided by Sponsor

1 Production Process

Not supplied by sponsor.

2 Other Information

Not supplied by sponsor.

————————————————————

End of Report