Optimization Goals

• Primary: Minimize time for the separationof a given number of peaks

• Secondary: Maximize the number of peaksseparated by a given column

Slide 6

Dial 1-816-650-0621 for e-Seminar Audio

HOW CAN THESE GOALS BE ACHIEVED?

Slide 7

Dial 1-816-650-0621 for e-Seminar Audio

Extreme Systems

Ultra-fast chromatogram (2 sec) obtained by using a fluidic logic gate injector

Ultra-fast chromatogram(< 1 sec) using cold trap injection device.

Complex Inlet systemsExtremely short columnsSmall internal diameter

Slide 8

Dial 1-816-650-0621 for e-Seminar Audio

Fast Temperature Programming

• Capillary GC column wrapped in a metal casing

• Resistively heats the capillary column

• Heats up to 20°C/sec

Slide 9

Dial 1-816-650-0621 for e-Seminar Audio

THERMEDICS EZFLASH™

Column: TCX-DB176m, 0.25mm I.D., 0.50µm

Carrier: Hydrogen @ 32 psigDetector: TEA in Nitrogen Mode, 850°CInjector: 300°COven: 80°C for 10 seconds

80-100°C at 120°C/min.100-160°C at 180°C/min160-280°C at 300°C/min.280-300°C at 60°C/min.

Compounds:1. Meperidine2. Methadone3. Benzotropine4. Clomipramine5. Verapamil

Drugs

Slide 10

Dial 1-816-650-0621 for e-Seminar Audio

Short Microbore Columns

10 m x 100 µm I.D., 02µmµ = 55 cm/sec H2

Inlet: Split 275:1, 275°CFID:350°C

70°C for 0.5 min30°/min to 120°C20°/min to200°C for 2 min

60m x 250 µm I.D., 025µmµ = 25 cm/sec He

Inlet: Split 150:1, 270°CFID:270°C

75°C for 8 min4°/min to 200°C200°C for 5 min

Western Spearmint Oil DB-Wax

Slide 11

Dial 1-816-650-0621 for e-Seminar Audio

Small Dimension Changes

Oven program:50°C for 0.5 min, 50-150°C at 25°/min, 150-260°c at 12°/min, 260-320°C at 15°/min

30 m x 0.25 mm I.D., 0.25 µm Hydrogen at 45 cm/secSplitless ECD

Pesticides DB-17ms

6 8 10 12 14 16 18

1 2 3

4

5

6

7 8 9 10

1112

13

14

1516

17

18

1920

6 7 8 9 10 11

1 23

45

6

7

8 91011

1213

14

1516

17

18

1920

20 m x 0.18 mm I.D., 0.18 µm Hydrogen at 45 cm/secSplitless ECDOven program:50°C for 0.5 min, 50-150°C at 35°/min, 150-260°C at 22°/min, 260-320°C at 25°/min

Slide 12

Dial 1-816-650-0621 for e-Seminar Audio

SAME GC SYSTEMOptimize the Method

20 40 60 80 100 120

0 20 40 60 80Time (min.)

Oven: 0°C for 15 min 0-50°C at 1°/min 50-130°C at 2°/min 130-270°C at 4°/minCarrier Gas: Helium at 24 cm/sec measured at 35°CInjector: Split 1:200, 250°CDetector: FID, 300°C

Carrier: H2, 24 psig, 31 cm/sOven: 35°C// 9.5 min// 13.3 °/min// 45°//

11 min// 1.4 °/min// 60°// 11min//2.7°/min// 220°// 3.6 min

Injector: Split 1:200, 0.2 µLDetector: FID @ 300°C

Column: DB-PETRO 100100m x 0.25 mm I.D., 0.5 µm

J&W P/N: 122-10A6

Slide 13

Dial 1-816-650-0621 for e-Seminar Audio

Questions to Ask

• What information do you need from your analysis?

• Do you have more baseline than you need betweenyour peaks?

• Do you need to resolve all of the components?

Slide 14

Dial 1-816-650-0621 for e-Seminar Audio

Factors Affecting Resolution

Rs = [(N)1/2/4] [k/(k+1)] [(α-1)/α)]

• Efficiency: N = theoretical plates

• Retention: k = retention factor

• Selectivity: α = separation factor

Slide 15

Dial 1-816-650-0621 for e-Seminar Audio

Variables for Shortening Run Times

• Stationary Phase

• Temperature Programming

• Carrier Gas: type and linear velocity

• Shorten Column Length

• Decrease Film Thickness

• Decrease Internal Diameter

Slide 16

Dial 1-816-650-0621 for e-Seminar Audio

Stationary phase and temperature changes

do not affect the Kc of all solutes equally

Slide 17

Dial 1-816-650-0621 for e-Seminar Audio

Distribution Constant

• Change in Kc affects retention

• Co-elution if solute Kc’s are equal

• Kc is determined by:

solute

stationary phase

temperature

K C =conc. of solute in mobile phaseconc. of solute in stationary phase

Slide 18

Dial 1-816-650-0621 for e-Seminar Audio

Start with the Right Phase

0 2 4 6 8 10 12 14

Time (min.)

0 1 2 3 4 5

Time (min.)

DB-115m x 0.32mm, 0.25µmOven:40°C for 2 min40-120°C at 5°C/min

DB-Wax15m, 0.32mm, 0.25µmOven:80-190°C at 20°C/min

Slide 19

Dial 1-816-650-0621 for e-Seminar Audio

Temperature Programming

Oven: 100°C for 0.8 min100° to 340°C at 20°/min340°C for 2.2 min

Carrier: Helium, 36 cm/sec at 100°Cconstant flow mode

Injection: 275°C, Splitless 0.8 minDetector: HP 5973 MSD,

300°C transfer line temp,full scan of m/z 50-500

Sample: 1.0 µL of a 20 ng/µL mixture

EPA Method 8082 AColumn: DB-XLB

30 m x 0.25 mm I.D., 0.25 µmJ&W P/N: 122-1232

Oven: 50°C for 0.5 min50° to 340°C at 10°/min340°C for 5 min

Carrier: Helium, 36 cm/s @ 100°Cconstant flow mode

Injection: 275°C, Splitless 0.5 minDetector: HP 5973 MSD,

300°C transfer line temp,full scan of m/z 50-500

Sample: 1.0 µL of a 20 ng/µL mixture

5.50 6.00 6.50 7.00 7.50 8.00 8.50 9.00 9.50 10.00 10.50 11.00 11.50 12.00 12.50

1

TMX

5 1831

5244

66

10187

110

151 153

141

137187

83

180170

206

DCB

14.00 15.00 16.00 17.00 18.00 19.00 20.00 21.00 22.00 23.00 24.00 25.00 26.00 27.00

1

TMX

5 18 31 52 4466

10187

151

110153

180

206

DCB

141

137170

187

83

Slide 20

Dial 1-816-650-0621 for e-Seminar Audio

Kc and Temperature

• Kc decreases with an increase in temperature

• Each solute’s Kc may change at it’s own rate

KC=conc. of solute in mobile phase

conc. of solute in stationary phase

Slide 21

Dial 1-816-650-0621 for e-Seminar Audio

DB-WAXTemperature Dependence

1

23

4

3

4

1,2

3

412

41

2,3

120°

100°

80°

70°

41 23

60°

1. α-Terpinene2. Dodecane3. Limonene4. 1,8-Cineole

Slide 22

Dial 1-816-650-0621 for e-Seminar Audio

Carrier GasType and Linear Velocity

Slide 23

Dial 1-816-650-0621 for e-Seminar Audio

Carrier Gas

10 20 30 40 50 60

0.25

0.50

0.75

1.00

u (cm/sec)

H

uoptOPGV

Van Deemter Curve

Slide 24

Dial 1-816-650-0621 for e-Seminar Audio

Carrier GasLinear Velocity

0 2 4 6 8 10 12Time (min.)

123

4

5

6 78

9

0 2 4 6 8 10 12Time (min.)

1 32

4

56

78

9

DB-1, 15 m x 0.25 mm I.D., 0.25 µm50°C for 2 min, 50-110°C at 5°/min

25 cm/sec 35 cm/secHelium

Slide 25

Dial 1-816-650-0621 for e-Seminar Audio

Carrier GasType

10 20 30 40 50 60

0.25

0.50

0.75

1.00

u (cm/sec)

H(mm)

He

H2

Van Deemter Curves:Hydrogen vs. Helium

Slide 26

Dial 1-816-650-0621 for e-Seminar Audio

Hydrogen vs. Helium

6.2 min.

R = 1.67 R = 1.65

Helium23.2 cm/sec

Hydrogen48 cm/sec

SE-5215m x 0.25mm150°C isothermal

Compounds:C17Pristane

13.0 min.

Slide 27

Dial 1-816-650-0621 for e-Seminar Audio

Optimize Column Dimensions

Slide 28

Dial 1-816-650-0621 for e-Seminar Audio

Column LengthResolution and Retention210°C isothermal

Resolution is proportional to square root of lengthIsothermal: Retention is proportional to lengthTemperature program: 1/3-1/2 of isothermal values

15 m

R=0.842.29 min

60 m

R=1.68

8.73 min

30 m

R=1.164.82 min

Slide 29

Dial 1-816-650-0621 for e-Seminar Audio

Decreasing Column Length

.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.00 50 1 00 1 50 2 00 2 50 3 00 3 50 4 00 4 50 5 00 5 50 6 00 6 50 7 00 7 50 8 00 8 50 9 00

1

2 3

4

5

1.0 2.0 3.0 4.0 5.0

1 2 3

4

5

6.0

BTEXCarrier: Helium, 36 cm/sec at 40°cOven : 40°C for 3 min, 5°/min to 100°C

1. Benzene2. Toluene3. Ethylbenzene4. m,p-Xylene5. o-Xylene

DB-530 m0.53 mm I.D., 0.5 µm

DB-515 m0.53 mm I.D., 0.5 µm

Slide 30

Dial 1-816-650-0621 for e-Seminar Audio

Decreasing Film Thickness

1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0

12

3

4

5

.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.0

1

2 3

4

5

BTEXCarrier: Helium, 36 cm/sec at 40°cOven : 40°C for 3 min, 5°/min to 100°C

DB-530 m, 0.53 mm I.D., 1.5 µm

1. Benzene2. Toluene3. Ethylbenzene4. m,p-Xylene5. o-Xylene

DB-530 m, 0.53 mm I.D., 0.5 µm

Slide 31

Dial 1-816-650-0621 for e-Seminar Audio

Effect of Film Thickness on Resolution

Rd fWhen solute k > 5

Rd fWhen solute k < 5

Slide 32

Dial 1-816-650-0621 for e-Seminar Audio

Column DiameterRetention: Same film thickness

Normalized to 0.25mm

I.D. (mm)Retention

Change (k)

0.18 1.39

0.20 1.25

0.25 1.00

0.32 0.78

0.53 0.47

Slide 33

Dial 1-816-650-0621 for e-Seminar Audio

Decrease DiameterEffect on Retention

0 5 10 15 20

0 2 4 6 8 10 12 14 16

DB-1, 30 m, 0.25 um80°C isothermal, He at 37 cm/secC10, C11, C12

0.25 mm, 0.25 um

0.32 mm, 0.25 um

21.81

16.06

Slide 34

Dial 1-816-650-0621 for e-Seminar Audio

Distribution ConstantKC

KC = kβ

β =r

2dfk =

t r’tm

Slide 35

Dial 1-816-650-0621 for e-Seminar Audio

Decrease DiameterPhase Ratio Held Constant

C7 - C20Carrier: Helium, 36 cm/sec at 40°COven : 60°C for 1 min, 20°/min to 300°C

0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 6.50 7.00 7.50 8.00 8.50 9.00 9.50 10.00 10.50 11.00 11.50

0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 6.50 7.00 7.50 8.00 8.50 9.00 9.50 10.00 10.50 11.00 11.50

DB-530 m, 0.53 mm I.D., 0.5 µm

DB-530 m, 0.45 mm I.D., 0.42 µmR = 2.03 RT = 11.76

R = 1.82 RT = 11.71

Slide 36

Dial 1-816-650-0621 for e-Seminar Audio

Column DiameterTheoretical Efficiency

k = 5

I.D. (mm) N/m

0.10 11905

0.18 6666

0.20 5941

0.25 4762

0.32 3717

0.53 2242

Slide 37

Dial 1-816-650-0621 for e-Seminar Audio

Decrease Diameter

• Small changes

- examples: 0.53 mm to 0.45 mm

0.25 mm to 0.20 mm

• Extreme changes

- Using short microbore columns: ≤ 0.1mm

Slide 38

Dial 1-816-650-0621 for e-Seminar Audio

.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.00 50 1 00 1 50 2 00 2 50 3 00 3 50 4 00 4 50 5 00 5 50 6 00 6 50 7 00 7 50 8 00 8 50 9 00

1

2 3

4

5

1 2

3

4

5

.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5

5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.0

Decrease DiameterAdjust Linear Velocity

BTEXCarrier: HeliumOven : 40°C for 3 min, 5°/min to 100°C

DB-530 m, 0.45 mm I.D., 0.42 µm

45.9 cm/sec

DB-530 m, 0.53 mm I.D., 0.5 µm

36 cm/sec

R = 3.33

R = 3.32

Slide 39

Dial 1-816-650-0621 for e-Seminar Audio

Decrease DiameterSmall Change - adjust conditions

Time(minutes)

2 4 6 8 10 12 14 16 18 20 22 24 26 28 30

1

2 3

4

5 6 7

89

10

11

12

14

1315

1617

19

20

21

22

23 24 26

27

29

28

30

3132

33

39

41

43

42

4547

5153

55

56

57

58

60

35 40 45 50 55

1

2 3

4

56

7

8

9 10

12

11

14

1315

16 1719

20

2123

22

24 2627

29

28

30

31

32

33

39

41

43

42

45

47

5153

55

56

57

58

60

5 10 15 20 25 30Time(minutes)

Column: DB-502.2 75m x 0.45mm x 2.55 µm,

Oven: 35°C for 6 min,35-200°C at 8°C/min200°C for 3.5 min

Carrier: Helium at 10 ml/minInjector: Purge and trap (O.I.A 4560)Trap: Tenax/Silica gel/CMSDesorb: 200°C for 0.6 minDetector B: ELCD (O.I.A 4420) with NiCat

reaction tube in the halogen mode

Column: DB-502.2 105m x 0.53mm x 3.0 µm,

Oven: 35°C for 10 min,35-200°C at 4°C/min200°C for 5 min

Carrier: Helium at 10 ml/minInjector: Purge and trap (O.I.A 4560)Trap: Tenax/Silica gel/CMSDesorb: 200°C for 0.6 minDetector B: ELCD (O.I.A 4420) with NiCat

reaction tube in the halogen mode

Slide 40

Dial 1-816-650-0621 for e-Seminar Audio

Extreme Diameter Changes

• Shorten column length to decrease run time

• Increase plates/meter by decreasing columndiameter

• For similar retention and selectivity keep thestationary phase and phase ratio

(β = r/df) the same.

Slide 41

Dial 1-816-650-0621 for e-Seminar Audio

Break

• For Questions and Answers

• Press *1 on Your Phone to

• Ask a Question

Slide 42

Dial 1-816-650-0621 for e-Seminar Audio

Considerations of using 0.1mm ID Columns

• Carrier Gas

• Temperature Program

• Injection efficiency

• Data system and detectors

• Working Range

Slide 43

Dial 1-816-650-0621 for e-Seminar Audio

Carrier Gas

• Hydrogen is the accepted carrier gas for fast GC analysis

• Fast optimal linear velocity

• Lower head pressure requirements

Slide 44

Dial 1-816-650-0621 for e-Seminar Audio

Van Deemter Curves:Hydrogen vs. Helium

10 20 30 40 50 60

0.25

0.50

0.75

1.00

u (cm/sec)

H(mm)

He

H2

Slide 45

Dial 1-816-650-0621 for e-Seminar Audio

TurpentineDB-WAX

1 4 5 6 72 3

1 4 5 62 3

1 4 52 3

10 m x 0.10 mm, 0.2 µm70° to 180°C at 15°/minH2, 30 cm/sec at 70°

10 m x 0.10 mm, 0.2 µm70° to 180°C at 15°/minH2, 65 cm/sec at 70°

10 m x 0.10 mm, 0.2 µm70° to 180°C at 15°/minH2, 95 cm/sec at 70°

Wh=1.08sBelow µ opt

Wh=0.72s

Wh=0.72s

µopt

OPGV

Slide 46

Dial 1-816-650-0621 for e-Seminar Audio

Pressure Considerations: 0.10 ID

Hydrogen

10m Velocity Split Ratio Split Flow Column Flow Head Pressure

µopt 65 300 204 0.68 33.5

OPGV 94 300 387 1.29 50.1

Helium

10m Velocity Split Ratio Split Flow Column Flow Head Pressure

µopt 41 300 153 0.51 43.3

OPGV 59 300 394 0.98 64.5

Slide 47

Dial 1-816-650-0621 for e-Seminar Audio

Kc and Temperature

• Kc of analytes must be maintained

• Temperature programs must be accurately scaledto maintain relative analyte retention

Slide 48

Dial 1-816-650-0621 for e-Seminar Audio

Method Translation Software

• Tool allowing GC methods to be translated to different conditions& maintain selectivity/resolution

• new column configuration

• different carrier gas

• faster separation

• Translates:- inlet Pressure, temp program, hold times

• Benefits- reduces methods development time- help assess if GC method compatible with HW

Slide 49

Dial 1-816-650-0621 for e-Seminar Audio

Method Translation Software

Slide 50

Dial 1-816-650-0621 for e-Seminar Audio

Temperature Programming

Temperature

30m, 0.25mm ID

10m, 0.1mm ID

Slide 51

Dial 1-816-650-0621 for e-Seminar Audio

Temperature Programming

30m, 0.25mm ID

Temperature

10m, 0.1mm ID

Temperature

Temperature program must be modified to give same temperatureof elution (i.e. faster ramps, shorter hold times)

Slide 52

Dial 1-816-650-0621 for e-Seminar Audio

Bergamot OilDB-WAX

4.2

Linalool

Linalyl acetate

2.6

Linalool

Linalyl acetate

10 20 30 40

2 4 6 8

1 2 3 4

30 m x 0.25 mm, 0.5 µm70° to 200°C at 3°/minH2, 45 cm/sec at 70°

10 m x 0.10 mm, 0.2 µm70° to 200°C at 10.4°/minH2, 65 cm/sec at 70°

10 m x 0.10 mm, 0.2 µm60° to 180°C at 30°/minH2, 65 cm/sec at 60°

Slide 53

Dial 1-816-650-0621 for e-Seminar Audio

Injector Efficiency

Injector Efficiency

Narrow columns generate narrow peaks

Injection band must be narrow to takeadvantage of the column efficiency

Slide 54

Dial 1-816-650-0621 for e-Seminar Audio

Injection Techniques

Split: high split ratio

Cold trapping

Slide 55

Dial 1-816-650-0621 for e-Seminar Audio

Data System Requirements

Narrow columns generate narrow peaks

0.5-5 second peak widths are common

Requires fast sample rate of detectors and data system

Slide 56

Dial 1-816-650-0621 for e-Seminar Audio

Detectors

FID - works well

NPD - works well

ECD - increase makeup, but very large dead volume

TCD - large volume issue

MS - scan speed

Area Counts Reduced

Scan rates- 50Hz for GC

12-20/sec scans for MS

Check your system

Slide 57

Dial 1-816-650-0621 for e-Seminar Audio

Working Range

• Column capacity is proportional to column diameter

• Column diameter will have little effect on detector sensitivity

W = Qs ÷ Qo

Qs = maximum column capacityQo = minimum amount that can be reliably detected

Slide 58

Dial 1-816-650-0621 for e-Seminar Audio

Working Range

0.53 0.32 0.25 0.10.53 0.32 0.25 0.10 DetectorMDQ

Diameter

400 ng

250ng

150 ng

40 ng

Values shown areapproximate columncapacities

Slide 59

Dial 1-816-650-0621 for e-Seminar Audio

Capacity: Effect on Resolution

0.5 1.0 1.5 2.0 2 4 6 8 10

0.5 1.0 1.5 2.0 2 4 6 8 10

1

2 3

4

1

2 3

4

1

23

4

1 4

2,3

10 m x 0.1 mm, 0.2 µm 30 m x 0.25 mm, 0.5 µm

1:1:1:1

1:100:1:1

Slide 60

Dial 1-816-650-0621 for e-Seminar Audio

Is Dilution the Solution?

12

3

4

123 4

0.5 1.0 1.5 2.0 0.5 1.0 1.5 2.0

1

2

34

0.5 1.0 1.5 2.00.5 1.0 1.5 2.0

1

2,3

4

10 m x 0.1 mm, 0.2 µm

1:1:1:1 0.01:0.01:0.01:0.01

0.01:1:0.01:0.011:100:1:1

Slide 61

Dial 1-816-650-0621 for e-Seminar Audio

Instrument Requirements

• High pressure capability

• Split Inlet

• Fast temperature ramping capability

• Fast detector scan or sample rate

Slide 62

Dial 1-816-650-0621 for e-Seminar Audio

Remember to ask the right questions

• What information do you need from your analysis?

• Do you have more baseline than you need betweenyour peaks?

• Do you need to resolve all of the components?

• Does your instrument have the necessary capabilities?

Slide 63

Dial 1-816-650-0621 for e-Seminar Audio

Remember the Variables for ShorteningAnalysis Time

• Stationary Phase

• Temperature Programming

• Carrier Gas: type and linear velocity

• Shorten Column Length

• Decrease Film Thickness

• Decrease Internal Diameter

Slide 64

Dial 1-816-650-0621 for e-Seminar Audio

Wrap-up E-Seminar Questions

Thank you for attending Agilent e-Seminars.