Electro-Magnetic Field Analysis Electro-Magnetic Field Analysis andand
Faraday Cage DesignFaraday Cage Design
![Page 2: Faraday Cage Design for PD Test Setup](https://reader033.fdocuments.us/reader033/viewer/2022061214/547ed50db37959492b8b5578/html5/thumbnails/2.jpg)
Faraday cage design process
Identifying and characterizing the magnetic and electric field sources through measurements
Calculating the cage aperture based on frequency composition of the EMI and shield thickness based on field strength
Simulating the test setup (source, measuring equipment and faraday cage) and evaluating shielding effectiveness
Selection of the most cost effective and least disruptive field reduction method
![Page 3: Faraday Cage Design for PD Test Setup](https://reader033.fdocuments.us/reader033/viewer/2022061214/547ed50db37959492b8b5578/html5/thumbnails/3.jpg)
Magnetic Field Magnetic Field AnalysisAnalysis
![Page 4: Faraday Cage Design for PD Test Setup](https://reader033.fdocuments.us/reader033/viewer/2022061214/547ed50db37959492b8b5578/html5/thumbnails/4.jpg)
Faraday Cage Model Enclosing PD Measuring Equipment
Current Cycling Test Equipment (1200A)
Faraday Cage
PD Measuring Equipment
![Page 5: Faraday Cage Design for PD Test Setup](https://reader033.fdocuments.us/reader033/viewer/2022061214/547ed50db37959492b8b5578/html5/thumbnails/5.jpg)
B-Field In Actual Lab Conditions
The B-Field is very low and any thin cage can give the required shielding. Considering the worst scenario, wherein the current cycling loop is in proximity (10 cm) of PD equipment further evaluation is carried out
The current cycling loop is about 6m away from PD measuring equipment
![Page 6: Faraday Cage Design for PD Test Setup](https://reader033.fdocuments.us/reader033/viewer/2022061214/547ed50db37959492b8b5578/html5/thumbnails/6.jpg)
Flux density on PD measuring equipment
Case A: Without faraday cage Case B: With faraday cage
Case A Case B[Cage shown for reference]
![Page 7: Faraday Cage Design for PD Test Setup](https://reader033.fdocuments.us/reader033/viewer/2022061214/547ed50db37959492b8b5578/html5/thumbnails/7.jpg)
Flux density on the impinging phase of measuring equipment
Case A
Case B
[Cage shown for reference]
![Page 8: Faraday Cage Design for PD Test Setup](https://reader033.fdocuments.us/reader033/viewer/2022061214/547ed50db37959492b8b5578/html5/thumbnails/8.jpg)
Flux density along the cross -section of faraday cage enclosing PD measuring equipment
Case A
Case B
![Page 9: Faraday Cage Design for PD Test Setup](https://reader033.fdocuments.us/reader033/viewer/2022061214/547ed50db37959492b8b5578/html5/thumbnails/9.jpg)
Case A
Case BOuter Surface
Flux density on the nearest face of faraday cage
Case BInner Surface
![Page 10: Faraday Cage Design for PD Test Setup](https://reader033.fdocuments.us/reader033/viewer/2022061214/547ed50db37959492b8b5578/html5/thumbnails/10.jpg)
Flux density (Tesla)
Case A Case B
On Measuring Equipment
0.5 0.003
On Faraday cage (Outer face)
0.12 0.48
Magnetic Field Analysis
![Page 11: Faraday Cage Design for PD Test Setup](https://reader033.fdocuments.us/reader033/viewer/2022061214/547ed50db37959492b8b5578/html5/thumbnails/11.jpg)
Inferences
Considering maximum radio or PD discharge interference frequency of 3 GHz the mesh size shall be smaller than 5cm
Considering the worst scenario the flux density on faraday cage for Case1 is 0.5 Tesla. The use of steel mesh (µr-1000 and thickness 2mm) as in Case2 brought down the B-value from 0.5 Tesla to less than 3mTesla (required limit 7-50 mTesla)
The skin depth evaluation is not considered as the field is static and hence flux distribution is based on ratio of air to mesh permeability alone
Theoretical calculations are in agreement with simulation results
Ref. Std: IEEE Std C95.6™-2002
![Page 12: Faraday Cage Design for PD Test Setup](https://reader033.fdocuments.us/reader033/viewer/2022061214/547ed50db37959492b8b5578/html5/thumbnails/12.jpg)
Electric Field AnalysisElectric Field Analysis
![Page 13: Faraday Cage Design for PD Test Setup](https://reader033.fdocuments.us/reader033/viewer/2022061214/547ed50db37959492b8b5578/html5/thumbnails/13.jpg)
Faraday Cage Model Enclosing PD Measuring Equipment
Current Cycling Test Equipment 1 kV
Faraday Cage
PD Measuring Equipment
![Page 14: Faraday Cage Design for PD Test Setup](https://reader033.fdocuments.us/reader033/viewer/2022061214/547ed50db37959492b8b5578/html5/thumbnails/14.jpg)
Case A Case B
Electric stress on PD measuring equipment
Case A: Without faraday cage Case B: With faraday cage
Emax- 0.15 kV/m Emax- 0.00004 kV/m
![Page 15: Faraday Cage Design for PD Test Setup](https://reader033.fdocuments.us/reader033/viewer/2022061214/547ed50db37959492b8b5578/html5/thumbnails/15.jpg)
Case A
Case B
Electric field distribution
![Page 16: Faraday Cage Design for PD Test Setup](https://reader033.fdocuments.us/reader033/viewer/2022061214/547ed50db37959492b8b5578/html5/thumbnails/16.jpg)
Electric Stress (kV/m)
Case A Case B
On measuring Equipment
0.15 0.00004
On Faraday cage 0.02 0.08
Electric Field Analysis
![Page 17: Faraday Cage Design for PD Test Setup](https://reader033.fdocuments.us/reader033/viewer/2022061214/547ed50db37959492b8b5578/html5/thumbnails/17.jpg)
Inferences
The maximum electric field is 0.08 kV/m on faraday cage (worst scenario case) which is well with in the required limit (5 kV/m)
The electric and magnetic fields in the laboratory are low. The effect of electric field is much lower than magnetic field. Hence, the proposed shield design is based on minimum requirements and magnetic field.
The steel mesh of µr-1000 and thickness 2mm shall provide the required shielding
Ref. Std: IEEE Std C95.6™