FACSAria SORP Standard Operation Protocol Basic Operation SORP Standard... · 2016-01-08 ·...
Transcript of FACSAria SORP Standard Operation Protocol Basic Operation SORP Standard... · 2016-01-08 ·...
Faculty Core Facility, Li Ka Shing Faculty of Medicine
BD FACSAria SORP Operation Manual
1
FACSAria SORP Standard Operation Protocol
Basic Operation
1. Checking Lasers Status
a. Click on BD Coherent Connection program to ensure the laser(s) are
“ON”. Lasers warm up duration ~ 30 minutes.
b. Turn off the lasers that are not required for analysis and/or sorting.
If only channels of FITC, PE, and APC are required; the UV and
Violet lasers can be switched off.
Faculty Core Facility, Li Ka Shing Faculty of Medicine
BD FACSAria SORP Operation Manual
2
Power of Blue laser is NOT optional. It is ON, together with the
equipment.
Switch off UV laser if it is not required. UV laser excites at
~355nm, it may cause DNA damage during sorting experiment.
Make sure the laser output of UV laser is set at 5.0 before laser is
turned “ON” or “OFF”. You may refer to the following steps.
To switch OFF UV laser,
Output must first be reduced to 5.0 mW before turning off
the laser
Faculty Core Facility, Li Ka Shing Faculty of Medicine
BD FACSAria SORP Operation Manual
3
2. BD FACSDiva Software Log In
a. Log into FACSDiva software with your own login name and
password. Please contact the administrator to establish a new user
account.
b. When connected, click ‘Use CST Settings’ when a mismatch was
located.
Faculty Core Facility, Li Ka Shing Faculty of Medicine
BD FACSAria SORP Operation Manual
4
3. For Sorting: Stream Break-Off and Droplet Formation (100µm
nozzle)
a. Turn ON the Stream and allow time for it to
stabilize.
b. Adjust the Frequency (Freq) between 27 and 30
to identify the HIGHEST break-off, where the
End-portion of the breaking stream can still be
observed.
c. Adjust the Amplitude (Ampl) to introduce DROP 1
measured between 120 and 180; but repeat the
search using a different Frequency if the finishing
Amplitude value is greater than 20.
d. Fine tune the Amplitude value to stabilize the
GAP measured between 8 and 12.
e. Ensure that the satellite droplet has completely
merged with the trailing drop (before
disappearing).
f. Match the input number with the measured value
in DROP 1.
g. Allow the stream to stabilize for a minimum of 3
minutes before the activation of Sweet Spot.
h. Ensure the stream has been stable for at least 3
minutes and that the Sweet Spot is turned ‘ON’.
Faculty Core Facility, Li Ka Shing Faculty of Medicine
BD FACSAria SORP Operation Manual
5
Drops Too FEW Drops Too MANY Break-off Too SOON
Solution: Increase Freq Solution: Decrease Freq Solution: Decrease Ampl
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
Gap Too NARROW Gap Too WIDE Break-off Too LATE
Solution: Increase Ampl Solution: Decrease Ampl Solution: Increase Ampl
Faculty Core Facility, Li Ka Shing Faculty of Medicine
BD FACSAria SORP Operation Manual
6
4. For sorting: ACCUDROP Drop Delay Test (100µm nozzle)
a. Turn on Voltage Test Sort move voltage sliders to appropriate
positions appeared similar to below.
DO NOT touch the deflection plates when the plate voltage is
ON!!! (When the voltage warning light illuminates, indicating that the
plates are charged at 12,000-volt).
b. Adjust the Micrometer Dial to obtain evenly bright bead spots for the
four side streams.
Core Stream Bead Spots TOO BRIGHT Bead Spots UNEVENLY BRIGHT
c. Use the existing well-prepared Accudrop bead solution in the fridge.
You are allowed to prepare fresh stock only if Accudrop solution has
been used up (1 drop Accudrop beads + 350µL FACSFlow solution).
d. Go to Browser Window Experiment New Experiment
Accudrop Drop Delay OK.
Faculty Core Facility, Li Ka Shing Faculty of Medicine
BD FACSAria SORP Operation Manual
7
e. Click “+” of Specimen_001, and set the tube pointer onto Tube_001.
f. Make sure the flow rate is 1.00 followed by loading sample into the
Bulk Injection Chamber.
g. Adjust the flow rate to achieve ~1200 events per second; if failed,
increase the flow rate or prepare suspension at a higher concentration.
h. Go to Browser Window ‘Create a new Sort Layout’.
i. Select ‘2 Tube’ in Device ‘Initial’ in Precision right click on
the Left Position and select ‘Add > P1’.
Faculty Core Facility, Li Ka Shing Faculty of Medicine
BD FACSAria SORP Operation Manual
8
j. Click Voltage and Optical Filter on the Side Stream Window Sort
on the Sort Layout Window.
k. Click Cancel on the Confirm Prompt.
l. Adjust the voltage slider and make sure the left side stream is
contained in the box.
m. Adjust the Drop Delay value (either or ) to achieve ~100%
intensity in the left side stream.
Faculty Core Facility, Li Ka Shing Faculty of Medicine
BD FACSAria SORP Operation Manual
9
n. When proximate to 100% intensity, change to ‘Fine Tune’.
o. Optimize the drop delay again (either or ) until the left side stream
intensity is consistently > 95%.
Note: Two percentage values must add up to 100 %.
p. Turn off the Voltage and Optical Filter.
q. Click Sort to stop sorting; select Cancel when asked if preferred to
save the report. Unload the tube.
Faculty Core Facility, Li Ka Shing Faculty of Medicine
BD FACSAria SORP Operation Manual
10
5. Sample Collecting Tube(s) Alignment (FOR SORTING)
a. For sorting, lift up the lid and insert the mock tubes into appropriate
collection device. Holders are available for 1.5mL microtubes,
12x75mm tubes, and 15mL Falcon tubes.
b. To keep the collecting tubes cool throughout the sorting process,
collection device should be connected with tubings directed from the
water bath.
c. Click Voltage Test Sort Waste Drawer Open the sort
block door Adjust the slider of the required side stream and
locate the position for target cells collection.
d. Turn off Voltage Test Sort Waste Drawer Close sort block
door when alignment has been completed remove the mock tubes
and replace new collection tubes with fresh media.
Faculty Core Facility, Li Ka Shing Faculty of Medicine
BD FACSAria SORP Operation Manual
11
6. Creating A New Experiment
a. Browser Window ‘New Folder’ ‘New Experiment’.
b. Select ‘Cytometer Settings’ ‘Parameters’ ‘Delete’ the
unnecessary parameters on the Inspector Window.
c. FSC (measures Cell Size) and SSC (measures Cell Granularity) are
MUST for all kind of analysis; their ‘Area’, ‘Height’ and ‘Width’ but
‘Log’ boxes should not be ticked. ‘Log’ and ‘Area’ boxes of
fluorescence channels must be ticked except for cell cycle and/or
DNA analysis.
Faculty Core Facility, Li Ka Shing Faculty of Medicine
BD FACSAria SORP Operation Manual
12
d. Choose Experiment Experiment Layout and define labels for
each parameter. Select the column of fluorescence channel and enter a
label in the Quick Entry Label field.
e. Select ‘New Specimen’ expands the Specimen to show Tube 001.
Highlight the tube with the Tube Pointer.
f. Select ‘Dot Plot or Histogram’ move the cursor onto the blank
worksheet.
Faculty Core Facility, Li Ka Shing Faculty of Medicine
BD FACSAria SORP Operation Manual
13
g. Right click on a plot ‘Duplicate’ to create another plot of the same
type with identical axis.
h. Select each individual axis, and opt from a list the preferred parameter.
Faculty Core Facility, Li Ka Shing Faculty of Medicine
BD FACSAria SORP Operation Manual
14
i. Below shows a template of plots used in routine analysis.
Dot Plot Dot Plot Dot Plot
Dot Plot Dot Plot Dot Plot
Histogram Histogram Histogram
Faculty Core Facility, Li Ka Shing Faculty of Medicine
BD FACSAria SORP Operation Manual
15
j. Right click on the plot Show Population Hierarchy
Right click on the plot Create Statistics View right click on the
statistics view Edit Statistics View
k. Select Statistics tab tick mean of FSC-A and FSC-H tick the
mean of the ‘A’rea of the parameters click OK
Faculty Core Facility, Li Ka Shing Faculty of Medicine
BD FACSAria SORP Operation Manual
16
7. Procedures for sample acquisition
a. Put your sample tube onto the loading dock. Ensure that the flow rate
is 1.0. Press ‘Load’ on Acquisition Dashboard.
b. Identify the population of interest by adjusting the voltage of FSC and
SSC on Parameters Press Restart to accelerate the changes
Faculty Core Facility, Li Ka Shing Faculty of Medicine
BD FACSAria SORP Operation Manual
17
c. Cytometer LaserAdjust the FSC Area Scaling until the mean of
FSC-A and FSC-H are APPROXIMATELY SAME.
Return to Parameters tab and finely re-adjust voltage of FSC and
SSC.
d. Adjust voltage of each of the fluorescent channels; preferably the
peak of population of interest is greater than ZERO but less than 102.
e. Press ‘Unload’ on Acquisition Dashboard to unload the current
sample tube.
f. Repeat above steps with the positive control sample tubes. Adjust the
voltage of corresponding channels if their signal peaks are outside the
limit of the histograms.
Faculty Core Facility, Li Ka Shing Faculty of Medicine
BD FACSAria SORP Operation Manual
18
8. Creating Gates
a. Set the current tube pointer to the following tube and repeat step l, m,
o and p with the Unstained / Negative Control tube.
b. Gate the cells of interest according to the following sequence by using:
Autopolygon
Polygon Quadrant Interval Snap-to
Rectangle
Highlight P3 population under P2
Create a polygon gate (P1) to
gate the main population on
the FSC-A vs SSC-A dot plot
Highlight P1 population in “Population Hierarchy’
Create a polygon gate (P2)
around the singlet event on the
FSC-H vs FSC-W dot plot
Highlight P2 population under P1
Create a polygon gate (P3)
around the singlet event on
the SSC-H vs SSC-W dot plot
Faculty Core Facility, Li Ka Shing Faculty of Medicine
BD FACSAria SORP Operation Manual
19
c. To define fluorescence positive signal, interval gate (P4, P5,…) beyond
negative peak of fluorescent channels in histogram plot of unstained
samples can be created. For over multiple fluorescence channels,
quadrant gate could be created to define single/double positive signals
(Q1; Q2; Q3; Q4….).
*P1 is the children of All Events and the parent of P2 population; P2
population is the children of P1 population and the parent of P3 population
and the grandparent of P4, P5, & P6 populations. On the hierarchy table,
you should highlight the P1 population when a gate for P2 population is
drawn, the P2 population when a gate for P3 population is drawn, and the
P3 population when gates for P4, P5, & P6 populations are drawn.
Faculty Core Facility, Li Ka Shing Faculty of Medicine
BD FACSAria SORP Operation Manual
20
d. Acquisition Dashboard click Next Tube to create a slot for new
sample tube / to assign to the following sample tube slot.
e. Ensure the Acquisition Setup has been arranged as follows:
f. Acquisition Dashboard click Record Data to record entire sample set.
Faculty Core Facility, Li Ka Shing Faculty of Medicine
BD FACSAria SORP Operation Manual
21
9. Setting up a Sorting Experiment
a. Ensure that the flow rate is 1.0 before loading the sample tube.
b. Browser Window ‘Create a new Sort Layout’.
c. Determine the most appropriate options from Device and Precision
on the Sort Layout Window.
(i) For 2 - 12x75mm tubes sorting, select ‘2 Tube’ device
(ii) For 4 – 15mL Falcon tubes sorting, select ‘4 Tube’ device
(iii) To achieve the highest yield sorting, choose ‘Yield’
(iv) To achieve the highest purity sorting, choose ‘Purity’ in ‘2
Tube’ sorting and ‘4-way Purity’ in ‘4 Tube’ sorting.
(v) To achieve single cell sorting, choose ‘Single Cell’
Faculty Core Facility, Li Ka Shing Faculty of Medicine
BD FACSAria SORP Operation Manual
22
d. Right click Far Left / Left / Right / Far Right Position ‘Add’
Gate of Interest Sort
e. Click OK on the Confirm Prompt.
f. Acquire data for the entire sample set. In each run, record all data by
clicking ‘Record Data’.
g. Constantly check for the sufficient level of samples inside the loading
chamber and plenty of room inside the collecting tubes for sorted
samples. Unload the sample tube and rinse the tubing with
FACSRinse® solution and Milli-Q water for five minutes each, if the
threshold rate fluctuates high and low, and/or the stream is unstable.
Faculty Core Facility, Li Ka Shing Faculty of Medicine
BD FACSAria SORP Operation Manual
23
Notes – During the process of Sorting
1. To help achieve an optimal result, users are advised to pay attention to
the following items:
Sample: 5 tubes of 1mL sample, rather than 1 tube of 5mL sample
Threshold Rate: Stable and no fluctuation
Sort Rate: Stable and no fluctuation
Stream: Stable and no sudden movement
Side Stream: Stable and no fanning
Sample Tube: Sufficient sample level to prevent clogging
Collecting Tube: Plenty of room for collection
2. If required to change the collecting tubes ‘Pause’ retrieve the
fully filled collection tubes and replace with new tubes ‘OK’ on
the Confirm Prompt ‘Resume’
3. Upon completion, click ‘Sort’ ‘OK’ ‘Unload’ the sample tube.
10. Cleaning the System
a. To 3 x 5mL tubes add 2 ml of FACSClean® Solution, FACSRinse®
Solution, and Milli-Q H2O.
b. Load each tube into the loading chamber.
c. Increase the flow rate to 11.0 and let each solution run for 5 minutes.
d. Repeat the same with FACSRinse® Solution for 10 min, if Propidium
Iodide (PI) is used.
Faculty Core Facility, Li Ka Shing Faculty of Medicine
BD FACSAria SORP Operation Manual
24
11. Export FCS Data / Experiment
a. To save Experiment, right click on the ‘Experiment’ ‘Export’
‘Experiment’ ‘Browse’ to choose the destination folder.
b. Create a new folder and rename, then click ‘Export’ ‘OK’.
c. To save FCS files, right click ‘Specimen’ ‘Export’ ‘FCS
Files’.
Faculty Core Facility, Li Ka Shing Faculty of Medicine
BD FACSAria SORP Operation Manual
25
d. Select FCS 3.0 ‘OK’ ‘Browse’ to choose folder destination.
e. Create a new folder and rename ‘Choose Directory’ ‘Save’.
f. Delete experiment after having FCS data / experiment exported.
12. Log Out
a. To log out of FACSDiva software, go to File Log Out.