Experimental Methods - Information and Library...

22
Chapter 2 Experimental Methods This chapter gives the description of experimental methods employed in the preparation and characterization of Li 2 O–Al 2 O 3 ZrO 2 –Sio 2 glasses. The details of the apparatus used and the techniques adopted for measuring electrical properties, optical absorption, ESR and photo luminescence spectra are described in detail in this chapter.

Transcript of Experimental Methods - Information and Library...

Page 1: Experimental Methods - Information and Library …shodhganga.inflibnet.ac.in/bitstream/10603/8317/10/10...78 Experimental Methods 2.1 Introduction In this chapter the detailed description

Chapter 2

Experimental Methods

This chapter gives the description of experimental methods

employed in the preparation and characterization of Li2O–Al2O3–

ZrO2–Sio2 glasses. The details of the apparatus used and the

techniques adopted for measuring electrical properties, optical

absorption, ESR and photo luminescence spectra are described in

detail in this chapter.

Page 2: Experimental Methods - Information and Library …shodhganga.inflibnet.ac.in/bitstream/10603/8317/10/10...78 Experimental Methods 2.1 Introduction In this chapter the detailed description

78

Experimental Methods

2.1 Introduction

In this chapter the detailed description of the methods used in the

preparation of pure as well as transition metal ions and rare earth ions doped

glasses and glasses are presented. Various measurement techniques employed

for characterization of the samples are also discussed. The description of the

apparatus used and detailed procedure adopted for studying dielectric

properties, Infrared spectra, optical absorption, electron spin resonance and

luminescence spectra of transition metal ions and rare earth ions doped Li2O–

ZrO2–SiO2 glasses are also included.

2.2 Glass preparation

2.2.1 Composition of the glass

Within the glass forming region of Li2O–Al2O3-ZrO2–SiO2 system the

following compositions are chosen for the present study. The details of the

compositions are:

7. ((30-x) Li2O-10 Al2O3-5ZrO2-55 Si2O2: x TiO2 (0 � x � 5)

8. (30-x) Li2O-10 Al2O3-5ZrO2-55 Si2O2: x WO3 (0 � x � 5)

9. 30 Li2O-9Al2O3/Y2O3/Sc2O3-5 ZrO2-55 SiO2: 1 Pr2O3/Er2O3

Page 3: Experimental Methods - Information and Library …shodhganga.inflibnet.ac.in/bitstream/10603/8317/10/10...78 Experimental Methods 2.1 Introduction In this chapter the detailed description

79

2.2.2 Methods of preparation of glasses

The glasses used for the present study are prepared by the melting and

quenching techniques [1-3]. The starting materials used for the preparation of

the present glasses were analytical grade reagents (99.9 % pure) of Li2CO3,

Al2O3, Y2O3, Sc2O3, ZrO2, SiO2, TiO2, WO3, Pr6O11, Er2O3 (Metall, Chaina).

The compounds of required compositions were thoroughly mixed in an agate

mortar and melted in a platinum crucible. The furnace used was a PID

temperature controlled furnace (Fig. 2.1). The glasses were melted at about

1450 - 1500 oC for an hour till a bubble free liquid was formed. The resultant

melt was poured on a rectangular brass mould (containing smooth polished

inner surface) held at room temperature. The samples were subsequently

annealed at 350 oC in another furnace.

The glasses were then ground and optically polished. The approximate

final dimensions of the glasses used for studying the electrical and optical

properties are 1 cm x 1 cm x 0.2 cm. For dielectric measurements thin coating

of silver paint was applied on either side of the glasses, to serve as electrodes.

Page 4: Experimental Methods - Information and Library …shodhganga.inflibnet.ac.in/bitstream/10603/8317/10/10...78 Experimental Methods 2.1 Introduction In this chapter the detailed description

80

The photographs of some of Li2O–ZrO2–SiO2 glass samples are shown below.

Fig. 2.2 Photographs of some of the pre polished Li2O–ZrO2–SiO2 glasses

(c)

Fig. 2.1 (a) PID controller and the photographs of (b) Melting furnace (c) Annealing furnace.

(a)

(b)

Er3+ doped glasses

Pr3+ doped glasses

Page 5: Experimental Methods - Information and Library …shodhganga.inflibnet.ac.in/bitstream/10603/8317/10/10...78 Experimental Methods 2.1 Introduction In this chapter the detailed description

81

2.3 Characterization of the samples

2.3.1 X–Ray diffraction

The crystalline phases if any in the glasses samples were checked by X-

ray diffraction spectra recorded on Rigaku D/Max ULTIMA III X-ray

diffractometer (Fig. 2.3) with CuKα radiation. (Fig. 2.4) represents the

schematic diagram of general X-ray diffractometer.

Fig. 2.3 Rigaku D/Max ULTIMA III X-ray diffractometer

Fig. 2.4 Schematic representation of X-ray diffraction

Page 6: Experimental Methods - Information and Library …shodhganga.inflibnet.ac.in/bitstream/10603/8317/10/10...78 Experimental Methods 2.1 Introduction In this chapter the detailed description

82

Glassy or amorphous materials do not have a long-range atomic order, i.e.,

atoms are arranged randomly. Therefore, a diffraction pattern containing sharp

peaks is not expected as observed in crystalline materials.

2.3.2 Physical parameters

The density (d) of the glasses was determined by the standard principle

of Archimedes’ using o-xylene (99.99 % pure) as the buoyant liquid. A direct

reading balance (capacity 100 gm, readability 0.1 mg) was used for weighing.

The bulk glass was suspended on a very thin copper strand that was set in the

immersion liquid container; the density of the samples was determined by

weighing them in the liquid and in air. From the measured values of density d

and calculated average molecular weight M , various physical parameters such

as metal ion concentration Ni, mean ion separation Ri, which are useful for

understanding the physical properties of these glasses were evaluated using

standard formulae [4, 5].

The refractive index (nd) of the glass samples was measured (at λ =

589.3 nm) at room temperature using Abbe refractometer with monobromo

naphthalene as the contact layer between the glass and the refractometer prism.

Page 7: Experimental Methods - Information and Library …shodhganga.inflibnet.ac.in/bitstream/10603/8317/10/10...78 Experimental Methods 2.1 Introduction In this chapter the detailed description

83

The average errors in these physical parameters were calculated and

given below.

Density, d (g/cm3) ±0.0001

Dopant ion concentration, Ni (1020/cm3) ±0.01

Inter-ionic distance of dopant ions, Ri (Å) ±0.01

Polaron radius, Rp(Å) ±0.01

Field strength, Fi (1015 cm-2) ±0. 01

Refractive index ±0. 001

2.3.3 Differential scanning calorimetric studies Thermal analysis of the samples was carried out by Netzsch

Simultaneous DSC/TG Thermal Analyzer STA409C (Fig. 2.6(a)) with 32-bit

controller to determine the glass transition temperature and crystalline peaks.

High temperature furnace together with a sample carrier suitable for Cp

measurements and Al2O3 crucibles were used. Apparatus was calibrated both

Fig. 2.5 Abbe refractometer (Model NAR-4T)

Page 8: Experimental Methods - Information and Library …shodhganga.inflibnet.ac.in/bitstream/10603/8317/10/10...78 Experimental Methods 2.1 Introduction In this chapter the detailed description

84

for temperature and for sensitivity with melting temperatures and melting

enthalpies of the pure metals: Ga, In, Sn, Zn, Al, Ag, Au. All the recordings

were carried out in argon (of 5N pure) atmosphere to prevent samples from

oxidation. Heating rate was 10 C/min in the temperature range 30–1300 C.

The glass transition temperature and other glass forming ability parameters are

evaluated to an accuracy of ± 1oC. The schematic sketch of Netzsch

Simultaneous DSC/TG Thermal Analyzer STA409C is also presented in Fig.

2.6(b).

Fig. 2.6(b) The sketch of Netzsch Simultaneous DSC/TG Thermal Analyzer STA409C.

Fig. 2.6(a) Netzsch Simultaneous DSC/TG Thermal Analyzer STA409C.

Page 9: Experimental Methods - Information and Library …shodhganga.inflibnet.ac.in/bitstream/10603/8317/10/10...78 Experimental Methods 2.1 Introduction In this chapter the detailed description

85

2.4 Analytical techniques

2.4.1 Dielectric measurements

The dielectric measurements of the samples used in the present

investigation were taken on HP 4263B LCR Meter in the frequency range 102 -

105 Hz [6] in the temperature range 30 – 300 oC.

2.4.1 (a) HP 4263B LCR Meter

Capacitance measurements were made with HP 4263B LCR Meter. The

HP 4263B LCR Meter is a fully automatic high performance test instrument

used to measure a wide range of parameters. The measurement speed of the

instrument is 29 ms; it has a very test quick recovery and has got built in

comparator. It can measure the capacitance to a high accuracy and wide

measurement range. The two measurement display sections, capacitance (C)

and dissipation factor (D) provide the direct read out of the parameters. The

instrument has got two modes of measurement viz., normal mode and average

mode; in normal mode operation, the instrument performs measurements at

very high speed whereas in the average mode it has short, medium or long

range rate of average of the measurement results. The second mode has been

adopted in the present measurements; show a high reliable and repeatable

value than in the normal mode of the measurement. The accuracy in the

measurement of dielectric constant is ~ 0.001 and that in loss is ~10-4.

Page 10: Experimental Methods - Information and Library …shodhganga.inflibnet.ac.in/bitstream/10603/8317/10/10...78 Experimental Methods 2.1 Introduction In this chapter the detailed description

86

2.4.1 (b) Sample holders

Two different types of sample holders have been used for room and high

temperature dielectric measurements:

i) For measurements at room temperature a laboratory made sample

holder with teflan base and micrometer arrangement (Fig. 2.8) was used for

dielectric loss measurements.

ii) For the temperature variation of the dielectric properties at a fixed

frequency the sample holder shown in Fig. 2.9) has been used. The sample was

held in between two silver electrodes and the holder was heated by a 200 W

nichrome heater. The heating was controlled by adjusting the current with a

variac. For good thermal stability, cold water was circulated round the jacket.

The readings were recorded after the sample had attained the steady

temperature. A chromel – alumel thermo couple was attached to the sample

Fig. 2.7 HP 4263B LCR Meter

Page 11: Experimental Methods - Information and Library …shodhganga.inflibnet.ac.in/bitstream/10603/8317/10/10...78 Experimental Methods 2.1 Introduction In this chapter the detailed description

87

holder very close to the sample. The thermo emf developed across the junction

was measured using digital PID temperature controller.

2.4.2 Spectroscopic properties

Under the spectroscopic properties (a) optical absorption, (b) electron

spin resonance spectra (c) infrared transmission spectra, Raman spectra (d)

luminescence spectra were studied.

2.4.2 (a) Optical absorption spectra

The optical absorption spectra of the glasses were recorded at room temperature in the spectral wavelength range covering 300–2100 nm with a spectral resolution of 0.1 nm using JASCO Model V-670 UV–vis–NIR spectrophotometer (Fig. 2.10(a)).

Fig. 2.8Laboratory made dielectric sample holder used at room temperature.

Fig. 2.9 High temperature sample holder.

Page 12: Experimental Methods - Information and Library …shodhganga.inflibnet.ac.in/bitstream/10603/8317/10/10...78 Experimental Methods 2.1 Introduction In this chapter the detailed description

88

A schematic diagram of the optical spectrophotometer arrangement in

the equipment is shown in Fig. 2.10(b). The light beam emitted from the light source

is reflected by the mirror M1 and directed into the monochromator. Deuterium lamp

D2 is used as a light source from 200 nm to light source switching wavelength (whose

initial value is 350.5 nm) and halogen lamp W1 from light source switching

wavelength 1100 nm, which are interchanged automatically according to the

wavelength range. The light beam coming from the monochromator is passed

through the stray-light cut off filter F, reflected by mirror M2 and then split by

the half mirror M3 into the sample and reference beams. Each beam passes

through the respective cell to photo diode detector.

Fig. 2.10(a) Jasco Model V-670 UV-vis-NIR spectrophotometer

Page 13: Experimental Methods - Information and Library …shodhganga.inflibnet.ac.in/bitstream/10603/8317/10/10...78 Experimental Methods 2.1 Introduction In this chapter the detailed description

89

Fig. 2.11 shows the electrical system. In the electrical system of the

spectrophotometer, the main control element is a microcomputer CPU which

controls the light source lighting, light source switching, filter switching,

wavelength scan, CRT display, keyboard and printer.

The sample and the reference beams are detected by photodiode

detectors. These beams are then logarithmically converted and their difference

is obtained by a differential amplifier. Some amount of the signal is then added

so that the signal zero and level zero may correspond to ABS zero. The signal

Fig. 2.10(b) Schematic diagram of the optical system of the spectrophotometer.

Page 14: Experimental Methods - Information and Library …shodhganga.inflibnet.ac.in/bitstream/10603/8317/10/10...78 Experimental Methods 2.1 Introduction In this chapter the detailed description

90

passes through the amplifiers 1 and 2. Amplifier 1 serves to finely adjust the

signal level and amplifier 2 serves to apply a GAIN up to 10-fold to the signal.

Then, after A/D-converter the signal is read by the CPU. The resolving power

of the instrument is 0.1nm.

Amplifier Sample beam

Bus line

I/O port

Printer

Printer driver Video RAM

CRT controller

CPU

RO

M

Dat

a R

AM

Lamp External I/O

Wavelength scan

Lamp switching

Filter Clock Key board

A/D

Switch

D/A

1 2

3

4

Detector Log amplifier

Differential amplifier

Amplifier (x 10, x 1)

Amplifier

CRT

2.4.2 (b) Electron spin resonance spectra

The electron spin resonance spectra of Li2O–Al2O3-ZrO2–SiO2 glasses

doped with titanium ions were recorded at liquid nitrogen temperature using

E11Z Varian X-band (ν = 9.5 GHz) JEOL JES-TES100 X-band ESR

spectrometer (Fig. 2.12) of 100 kHz field modulation that works.

Fig. 2.11 Schematic diagram of electrical system of the spectrophotometer.

Page 15: Experimental Methods - Information and Library …shodhganga.inflibnet.ac.in/bitstream/10603/8317/10/10...78 Experimental Methods 2.1 Introduction In this chapter the detailed description

91

The principle of microwave spectrometer used for Electron Spin

Resonance and the schematic diagram of the JOEL ESR spectrometer are

shown in Figs. 2.13 & 2.14. The arrangement consists of a microwave unit (a

pre amplifier, a Gunn oscillator and an automatic frequency control circuit), a

cavity resonator, a detector to measure the variations of the microwave power, a

modulating system for the amplification of the signal, a recording unit, an

electromagnet to produce uniform magnetic field and an electronic system to

measure the variations in magnetic field.

Fig. 2.12 JEOL ESR spectrometer

Fig. 2.13 Principle of a microwave spectrometer used for recording ESR signal.

Circulator

Sample Electromagnet

N S

Phase shifter

Detector Oscillator Amplifier

Cavity resonator

Field Sweep

Page 16: Experimental Methods - Information and Library …shodhganga.inflibnet.ac.in/bitstream/10603/8317/10/10...78 Experimental Methods 2.1 Introduction In this chapter the detailed description

92

The Gunn diode oscillator generates microwaves over a frequency range

of 8.8 to 9.6 GHz (microwave X-band). The automatic frequency control circuit

is used to match the frequency of the Gunn oscillator with the resonant

frequency of the cavity resonator. When ESR is excited, microwaves from the

resonator are reflected and enter the balance mixer that is made up of magic T

and a crystal mount. The waves are then detected and amplified by the pre

amplifier of the microwave unit. The magnetic field produced by the

electromagnet operated at 200 V, 2 kVA can be varied with a field sweep of ±

2500 gauss. DC magnetic field modulation is used to reduce the noise

contribution. The fine powder of the sample is placed at the middle of the

cavity resonator such that it is at the middle of the poles of the electromagnet.

Fig. 2.14 Schematic diagram of the JEOL ESR spectrometer.

Sample cavity

Mod

. coi

ls

Recorder Filter circuit

Amplifier

Modulation power supply

100 kHz 80 Hz

2nd Derivative

switch controls Preamplifier Microwave unit

AFC circuit

Gunn Oscillator power supply

Power supply

Sweep voltage excitation generator

Linear field sweep unit

Hall element excitation

power supply

Oscilloscope

Det

ecto

r

Electromagnet

Page 17: Experimental Methods - Information and Library …shodhganga.inflibnet.ac.in/bitstream/10603/8317/10/10...78 Experimental Methods 2.1 Introduction In this chapter the detailed description

93

The sample is subjected the microwave magnetic field of a constant frequency

which is perpendicular to the external dc magnetic field. When the resonance

condition is satisfied, the Q value of the cavity resonator changes and this

variation is detected, amplified and recorded as a function of the field. A typical

ESR absorption and first derivative signals are shown in Fig. 2.15.

Fig. 2.15 (a) ESR absorption curve (b) ESR differential curve

P ESR absorption signal

Signal width

(a) Magnetic field

H

Modulation width: III (b) Differential curve

H

III < ∆H

HP/dP

Page 18: Experimental Methods - Information and Library …shodhganga.inflibnet.ac.in/bitstream/10603/8317/10/10...78 Experimental Methods 2.1 Introduction In this chapter the detailed description

94

2.4.2 (c) Infrared transmission spectra

Infrared transmission spectra were recorded on a (Fig. 2.16) with a

resolution of 0.1 cm−1 in the spectral range 400–2000 cm−1 using potassium

bromide pellets (300 mg) containing pulverized sample (1.5 mg). These pellets

were pressed in a vacuum die at ∼680MPa. The spectra were recorded on Jasco-

FT/IR-5300 spectrophotometer.

2.4.2 (d) Raman spectra

The Raman spectra of this glasses were recorded with an NIR excitation

line (1064 nm) with a resolution of 0.1 cm-1 using a Bio-Rad Raman

spectrometer Model FTS 175C (Fig. 2.17) equipped with an FT Raman

supplementary accessory working in a back-scattering geometry system.

Fig 2.16 JASCO-FT/IR–5300 spectrophotometer

Page 19: Experimental Methods - Information and Library …shodhganga.inflibnet.ac.in/bitstream/10603/8317/10/10...78 Experimental Methods 2.1 Introduction In this chapter the detailed description

95

2.4.2 (e) Photoluminescence spectra

The photoluminescence spectra of the samples were recorded at room

temperature on a Photon Technology International (PTI) Spectrofluorometer

(Fig. 2.18). This instrument contains autocalibrated quadrascopic

monochrometer for wavelength selection and quadracentric sample

compartment. The light source is high intensity continuous xenon lamp with

high sensitivity TE–cooled InGaAs detector with lock-in amplifier and chopper

for noise suppression and an additional emission mono with a 600 groove

grating blazed at 1.2 microns. The system provides unmatched NIR

luminescence recording capability from 500 nm–2.2 µm. The spectral

resolution is 0.1 nm.

Fig. 2.17 Bio-Rad Raman spectrometer Model FTS 175C

Page 20: Experimental Methods - Information and Library …shodhganga.inflibnet.ac.in/bitstream/10603/8317/10/10...78 Experimental Methods 2.1 Introduction In this chapter the detailed description

96

The brief sketch of the methods employed in recording the absorption,

excitation and photoluminescence spectra are shown in Fig 2.19.

The fluorescence decay curves were recorded for rare earth ions doped

glasses and glasses using Jobinyvon spectrofluorolog-3 with pulsed xenon lamp

of 450 W.

Fig. 2.18 Photon Technology International fluorescence spectrophotometer

Page 21: Experimental Methods - Information and Library …shodhganga.inflibnet.ac.in/bitstream/10603/8317/10/10...78 Experimental Methods 2.1 Introduction In this chapter the detailed description

97

In general it is to be pointed out the data have been taken with the

several Li2O–ZrO2–SiO2 glasses and glasses doped with different transition

metal oxides and rare earth oxides are found to be reproducible with in the

accuracies mentioned.

Fig. 2.19 Brief sketch of the methods adopted in recording (a) absorption (b) excitation and (c) photoluminescence spectra.

S

C λexe

C Excitation

a M

Jc J

b

M

P λ

UV or M

Spectral composition of luminescence

Luminescence

c M

Luminescence intensity with different λexe

S

S

Page 22: Experimental Methods - Information and Library …shodhganga.inflibnet.ac.in/bitstream/10603/8317/10/10...78 Experimental Methods 2.1 Introduction In this chapter the detailed description

98

References

1. James E Shelby, Introduction to Science and Technology of Glass Materials (Royal Society of Chemistry, 2005). 2. W. Vogel, Glass Chemistry (Springer Verlag, Berlin 1994) 3. S.R. Elliot, Physics of Amorphous Materials (Longman, Essex 1990). 4. M.J. Weber, R.A. Saroyan and R.C. Ropp, J. Non-Crst. Solids 4 (1981) 137. 5. M.J. Weber, J. Chem. Phys. 48 (1968) 4774. 6. Hewlett-Packard Company, Hyogo, Japan. HP 4263 B LCR Meter

Instruction Manual. 7. M. Piaescki, I.V. Kityk, P. Bragiel, L.R.P. Kassab, D.M. da Silva, Mater.

Lett. 61 (2007) 2943