Experiences commissioning the Toshiba Aquilion ONE … commissioning the Toshib… · Experiences...
Transcript of Experiences commissioning the Toshiba Aquilion ONE … commissioning the Toshib… · Experiences...
![Page 1: Experiences commissioning the Toshiba Aquilion ONE … commissioning the Toshib… · Experiences commissioning the Toshiba Aquilion ONE 320 detector row CT scanner N Weir, N Bate,](https://reader033.fdocuments.us/reader033/viewer/2022050917/5ab11a327f8b9ac66c8c1815/html5/thumbnails/1.jpg)
Experiences commissioning the Toshiba Aquilion ONE 320 detector row CT scannerdetector row CT scanner
N Weir, N Bate, L Nichol
Department of Medical PhysicsRoyal Infirmary of Edinburgh
NHS Lothian
CTUG 2009 Belfast
![Page 2: Experiences commissioning the Toshiba Aquilion ONE … commissioning the Toshib… · Experiences commissioning the Toshiba Aquilion ONE 320 detector row CT scanner N Weir, N Bate,](https://reader033.fdocuments.us/reader033/viewer/2022050917/5ab11a327f8b9ac66c8c1815/html5/thumbnails/2.jpg)
The Clinical Research Imaging Centre (CRIC), Edinburgh
• 320 MDCT: Toshiba Aquilion ONE
• 3T MRI: Siemens Verio• PET-CT: Siemens Biograph • PET-CT: Siemens Biograph
mCT X (128 slice CT)
CTUG 2009 Belfast
![Page 3: Experiences commissioning the Toshiba Aquilion ONE … commissioning the Toshib… · Experiences commissioning the Toshiba Aquilion ONE 320 detector row CT scanner N Weir, N Bate,](https://reader033.fdocuments.us/reader033/viewer/2022050917/5ab11a327f8b9ac66c8c1815/html5/thumbnails/3.jpg)
The Clinical Research Imaging Centre (CRIC), Edinburgh
CTUG 2009 Belfast
![Page 4: Experiences commissioning the Toshiba Aquilion ONE … commissioning the Toshib… · Experiences commissioning the Toshiba Aquilion ONE 320 detector row CT scanner N Weir, N Bate,](https://reader033.fdocuments.us/reader033/viewer/2022050917/5ab11a327f8b9ac66c8c1815/html5/thumbnails/4.jpg)
Toshiba Aquilion ONE: Scanner StatsDetector rows 320x0.5mm
Total detector width (z) at isocentre
16 cm
Min rot time 0.35s
Dose reduction SureExposure3D and QDS
Scan modes Sequence, spiral, volume, dynamic volume
CTUG 2009 Belfast
Spiral mode 64x0.5mm or 32x0.5mm, no gantry tilt
Volume mode Static or stepped table, 80-320 detector rows selectable640 slice mode ([email protected]) possible
3 beam shaping filters Small: <= 240 mmMedium: > 240 mm,
<= 400 mmLarge: > 400 mm
Tohisba Medical Systems
![Page 5: Experiences commissioning the Toshiba Aquilion ONE … commissioning the Toshib… · Experiences commissioning the Toshiba Aquilion ONE 320 detector row CT scanner N Weir, N Bate,](https://reader033.fdocuments.us/reader033/viewer/2022050917/5ab11a327f8b9ac66c8c1815/html5/thumbnails/5.jpg)
Cone beam reconstruction: ConeXact
CTUG 2009 BelfastTohisba Medical Systems
![Page 6: Experiences commissioning the Toshiba Aquilion ONE … commissioning the Toshib… · Experiences commissioning the Toshiba Aquilion ONE 320 detector row CT scanner N Weir, N Bate,](https://reader033.fdocuments.us/reader033/viewer/2022050917/5ab11a327f8b9ac66c8c1815/html5/thumbnails/6.jpg)
Clinical imaging at 320slice
CTUG 2009 Belfast
Tohisba Medical Systems
![Page 7: Experiences commissioning the Toshiba Aquilion ONE … commissioning the Toshib… · Experiences commissioning the Toshiba Aquilion ONE 320 detector row CT scanner N Weir, N Bate,](https://reader033.fdocuments.us/reader033/viewer/2022050917/5ab11a327f8b9ac66c8c1815/html5/thumbnails/7.jpg)
Clinical imaging at 320slice
CTUG 2009 Belfast
Tohisba Medical Systems
![Page 8: Experiences commissioning the Toshiba Aquilion ONE … commissioning the Toshib… · Experiences commissioning the Toshiba Aquilion ONE 320 detector row CT scanner N Weir, N Bate,](https://reader033.fdocuments.us/reader033/viewer/2022050917/5ab11a327f8b9ac66c8c1815/html5/thumbnails/8.jpg)
Clinical imaging at 320slice
CTUG 2009 BelfastTohisba Medical Systems
![Page 9: Experiences commissioning the Toshiba Aquilion ONE … commissioning the Toshib… · Experiences commissioning the Toshiba Aquilion ONE 320 detector row CT scanner N Weir, N Bate,](https://reader033.fdocuments.us/reader033/viewer/2022050917/5ab11a327f8b9ac66c8c1815/html5/thumbnails/9.jpg)
Volume mode dosimetry: What to measure?
• Mori et al., Med Phys, 32, 2005:Integration length at least 300 mm required to represent >90% of
integral dose within beams up to 138mm
• Scan through small detector (e.g. Farmer / RTI) ?
CTUG 2009 Belfast
• Scan through small detector (e.g. Farmer / RTI) ?
![Page 10: Experiences commissioning the Toshiba Aquilion ONE … commissioning the Toshib… · Experiences commissioning the Toshiba Aquilion ONE 320 detector row CT scanner N Weir, N Bate,](https://reader033.fdocuments.us/reader033/viewer/2022050917/5ab11a327f8b9ac66c8c1815/html5/thumbnails/10.jpg)
Volume mode dosimetry: What to measure?
• Good agreement between CTDI300w, CTDI600w• D100w numerically ≈ CTDI300w
CTUG 2009 Belfast
![Page 11: Experiences commissioning the Toshiba Aquilion ONE … commissioning the Toshib… · Experiences commissioning the Toshiba Aquilion ONE 320 detector row CT scanner N Weir, N Bate,](https://reader033.fdocuments.us/reader033/viewer/2022050917/5ab11a327f8b9ac66c8c1815/html5/thumbnails/11.jpg)
Volume mode dosimetry: D100
Cone beam40 - 160 mm @ isocentre
100 mm Ionisation
CTUG 2009 Belfast
Ionisation chamber
z
![Page 12: Experiences commissioning the Toshiba Aquilion ONE … commissioning the Toshib… · Experiences commissioning the Toshiba Aquilion ONE 320 detector row CT scanner N Weir, N Bate,](https://reader033.fdocuments.us/reader033/viewer/2022050917/5ab11a327f8b9ac66c8c1815/html5/thumbnails/12.jpg)
Volume mode dosimetry: D100 resultsCTDI300 CTDI300
Filter kV D100 CF Edin (calc) Geleijns et al %diffAir L 80 9.2 1.02 9.4 10.1 -7.3
L 100 15.8 1.04 16.4 17.8 -7.7L 120 23.7 1.05 24.9 26.9 -7.5L 135 30.4 1.05 32.0 34.7 -7.9M 120 23.3 1.05 24.5 26.8 -8.7S 120 26.7 1.05 28.0 29.5 -5.0
Body L 80 3.0 1.02 3.0 3.3 -7.6
CTUG 2009 Belfast
Body L 80 3.0 1.02 3.0 3.3 -7.6L 100 6.0 1.04 6.2 6.6 -5.6L 120 9.7 1.08 10.4 11.1 -6.0L 135 13.0 1.06 13.8 14.9 -7.5M 80 2.7 0.98 2.6 2.9 -10.3M 100 5.2 1.02 5.3 5.9 -9.7M 120 8.6 1.04 9.0 9.8 -8.5M 135 11.7 1.06 12.4 13.2 -6.4
Head S 80 6.4 1.08 6.9 7.2 -3.7S 100 11.8 1.10 13.0 13.6 -4.8S 120 18.5 1.11 20.5 21.5 -4.5S 135 23.9 1.10 26.3 28.1 -6.3
![Page 13: Experiences commissioning the Toshiba Aquilion ONE … commissioning the Toshib… · Experiences commissioning the Toshiba Aquilion ONE 320 detector row CT scanner N Weir, N Bate,](https://reader033.fdocuments.us/reader033/viewer/2022050917/5ab11a327f8b9ac66c8c1815/html5/thumbnails/13.jpg)
Volume mode dosimetry: Beam profile
Geleijns et al.
Edin
CTUG 2009 Belfast
FWHM 18 cm
FWHM 17 cm
![Page 14: Experiences commissioning the Toshiba Aquilion ONE … commissioning the Toshib… · Experiences commissioning the Toshiba Aquilion ONE 320 detector row CT scanner N Weir, N Bate,](https://reader033.fdocuments.us/reader033/viewer/2022050917/5ab11a327f8b9ac66c8c1815/html5/thumbnails/14.jpg)
Measuring CTDI300
CTUG 2009 Belfast
![Page 15: Experiences commissioning the Toshiba Aquilion ONE … commissioning the Toshib… · Experiences commissioning the Toshiba Aquilion ONE 320 detector row CT scanner N Weir, N Bate,](https://reader033.fdocuments.us/reader033/viewer/2022050917/5ab11a327f8b9ac66c8c1815/html5/thumbnails/15.jpg)
CTDI300 measurements: Results
CTDI300 CTDI300Filter kV Edin (meas) Geleijns et al %diff
Air L 80 9.9 10.1 -2.4L 100 17.0 17.8 -4.7L 120 25.4 26.9 -5.4L 135 32.7 34.7 -5.6M 120 25.3 26.8 -5.5
CTUG 2009 Belfast
S 120 29.1 29.5 -1.4
Body L 80 3.2 3.3 -3.8L 100 6.4 6.6 -3.1L 120 10.5 11.1 -5.7L 135 14.1 14.9 -5.2M 120 9.4 2.9 -3.8
![Page 16: Experiences commissioning the Toshiba Aquilion ONE … commissioning the Toshib… · Experiences commissioning the Toshiba Aquilion ONE 320 detector row CT scanner N Weir, N Bate,](https://reader033.fdocuments.us/reader033/viewer/2022050917/5ab11a327f8b9ac66c8c1815/html5/thumbnails/16.jpg)
CTDI300 measurements: Results
Perisinakis et al. PMB 52, 2007:“The ratio of (CTDIw / CTDIair) determined using the optimum [chamber] length was found to be independent of beam collimation”
CTUG 2009 Belfast
of beam collimation”
![Page 17: Experiences commissioning the Toshiba Aquilion ONE … commissioning the Toshib… · Experiences commissioning the Toshiba Aquilion ONE 320 detector row CT scanner N Weir, N Bate,](https://reader033.fdocuments.us/reader033/viewer/2022050917/5ab11a327f8b9ac66c8c1815/html5/thumbnails/17.jpg)
CTDI300 measurements: Results
CTDI300 CTDI300Filter kV Edin (meas) Geleijns et al %diff
Air L 80 9.9 10.1 -2.4L 100 17.0 17.8 -4.7L 120 25.4 26.9 -5.4L 135 32.7 34.7 -5.6M 120 25.3 26.8 -5.5
(CTDIair)Edin
CTUG 2009 Belfast
S 120 29.1 29.5 -1.4
Body L 80 3.2 3.3 -3.8L 100 6.4 6.6 -3.1L 120 10.5 11.1 -5.7L 135 14.1 14.9 -5.2M 120 9.4 2.9 -3.8
(CTDIair)Edin
*(CTDIw/CTDIair)Geleijns %diff3.2 1.46.3 -1.6
10.5 0.314.1 -0.59.3 -1.8
![Page 18: Experiences commissioning the Toshiba Aquilion ONE … commissioning the Toshib… · Experiences commissioning the Toshiba Aquilion ONE 320 detector row CT scanner N Weir, N Bate,](https://reader033.fdocuments.us/reader033/viewer/2022050917/5ab11a327f8b9ac66c8c1815/html5/thumbnails/18.jpg)
Volume mode dosimetry: Scanner displayed values
On Aquilion ONE, Toshiba define:• Extended CTDIvol -> CTDIvol.e• Extended DLP -> DLP.e
CTDIvol.e = CTDIw’ x R (R = no rotations)
CTUG 2009 Belfast
Where CTDIw’ uses:CTDI100’ = -50
+50∫ D(z) / min{NxT, 100mm} dz
i.e. for a volume scan with collimation ≥ 100 mmCTDI100’ ≡ D100
and,DLP.e = CTDIvol.e x N x T
![Page 19: Experiences commissioning the Toshiba Aquilion ONE … commissioning the Toshib… · Experiences commissioning the Toshiba Aquilion ONE 320 detector row CT scanner N Weir, N Bate,](https://reader033.fdocuments.us/reader033/viewer/2022050917/5ab11a327f8b9ac66c8c1815/html5/thumbnails/19.jpg)
Image quality - Phantoms
CTUG 2009 Belfast
![Page 20: Experiences commissioning the Toshiba Aquilion ONE … commissioning the Toshib… · Experiences commissioning the Toshiba Aquilion ONE 320 detector row CT scanner N Weir, N Bate,](https://reader033.fdocuments.us/reader033/viewer/2022050917/5ab11a327f8b9ac66c8c1815/html5/thumbnails/20.jpg)
24,000 images….24,000 images….
CTUG 2009 Belfast
![Page 21: Experiences commissioning the Toshiba Aquilion ONE … commissioning the Toshib… · Experiences commissioning the Toshiba Aquilion ONE 320 detector row CT scanner N Weir, N Bate,](https://reader033.fdocuments.us/reader033/viewer/2022050917/5ab11a327f8b9ac66c8c1815/html5/thumbnails/21.jpg)
Image quality results – Interslice noise
Interslice Noise (32cm Water phantom)
25
30
35
40
CTUG 2009 Belfast
0
5
10
15
20
-80 -60 -40 -20 0 20 40 60 80
z (mm)
std
. d
ev.
![Page 22: Experiences commissioning the Toshiba Aquilion ONE … commissioning the Toshib… · Experiences commissioning the Toshiba Aquilion ONE 320 detector row CT scanner N Weir, N Bate,](https://reader033.fdocuments.us/reader033/viewer/2022050917/5ab11a327f8b9ac66c8c1815/html5/thumbnails/22.jpg)
Image quality results –Uniformity through volume:Body phantom
CTUG 2009 Belfast
![Page 23: Experiences commissioning the Toshiba Aquilion ONE … commissioning the Toshib… · Experiences commissioning the Toshiba Aquilion ONE 320 detector row CT scanner N Weir, N Bate,](https://reader033.fdocuments.us/reader033/viewer/2022050917/5ab11a327f8b9ac66c8c1815/html5/thumbnails/23.jpg)
Image quality results – Body phantom:Uniformity in z
CTUG 2009 Belfast
![Page 24: Experiences commissioning the Toshiba Aquilion ONE … commissioning the Toshib… · Experiences commissioning the Toshiba Aquilion ONE 320 detector row CT scanner N Weir, N Bate,](https://reader033.fdocuments.us/reader033/viewer/2022050917/5ab11a327f8b9ac66c8c1815/html5/thumbnails/24.jpg)
Image quality results – Spatial resolution
CTUG 2009 Belfast
![Page 25: Experiences commissioning the Toshiba Aquilion ONE … commissioning the Toshib… · Experiences commissioning the Toshiba Aquilion ONE 320 detector row CT scanner N Weir, N Bate,](https://reader033.fdocuments.us/reader033/viewer/2022050917/5ab11a327f8b9ac66c8c1815/html5/thumbnails/25.jpg)
Image quality results – Spatial resolution
MTF variation with position in beam
0.6
0.7
0.8
0.9
1 MTF50
MTF10
CTUG 2009 Belfast
0
0.1
0.2
0.3
0.4
0.5
0.6
-60 -40 -20 0 20 40 60
position within beam
lp/m
m
![Page 26: Experiences commissioning the Toshiba Aquilion ONE … commissioning the Toshib… · Experiences commissioning the Toshiba Aquilion ONE 320 detector row CT scanner N Weir, N Bate,](https://reader033.fdocuments.us/reader033/viewer/2022050917/5ab11a327f8b9ac66c8c1815/html5/thumbnails/26.jpg)
-40 mm
Image quality results – Slice thickness thro’ volume
0 mm +40 mm
Bead ramp profiles
140
160
-40mm
CTUG 2009 Belfast
0
20
40
60
80
100
120
0 10 20 30 40 50 60 70 80pixel
PV
-40mm
0 mm
+40 mm
![Page 27: Experiences commissioning the Toshiba Aquilion ONE … commissioning the Toshib… · Experiences commissioning the Toshiba Aquilion ONE 320 detector row CT scanner N Weir, N Bate,](https://reader033.fdocuments.us/reader033/viewer/2022050917/5ab11a327f8b9ac66c8c1815/html5/thumbnails/27.jpg)
Conclusions
• Modified methodology for dosimetry and image quality testing required for wide beam system
• Largely possible with existing equipment, manufacturer supplied phantomsmanufacturer supplied phantoms
• Automated image analysis essential
CTUG 2009 Belfast
![Page 28: Experiences commissioning the Toshiba Aquilion ONE … commissioning the Toshib… · Experiences commissioning the Toshiba Aquilion ONE 320 detector row CT scanner N Weir, N Bate,](https://reader033.fdocuments.us/reader033/viewer/2022050917/5ab11a327f8b9ac66c8c1815/html5/thumbnails/28.jpg)
Acknowledgements
• Toshiba Medical Systems• IQWorks Project• CRIC CT Radiographers: David Brian, Tessa Smith• Dundee Medical Physics Department
CTUG 2009 Belfast
![Page 29: Experiences commissioning the Toshiba Aquilion ONE … commissioning the Toshib… · Experiences commissioning the Toshiba Aquilion ONE 320 detector row CT scanner N Weir, N Bate,](https://reader033.fdocuments.us/reader033/viewer/2022050917/5ab11a327f8b9ac66c8c1815/html5/thumbnails/29.jpg)
CTUG 2009 Belfast
Bye!
Tohisba Medical Systems