Exp 1 - Process Instrument
Transcript of Exp 1 - Process Instrument
-
8/18/2019 Exp 1 - Process Instrument
1/8
SUMMARY
In this experiment, we use manometer, multimeter, a vacuum pump; a hand held
terminal, a hand pump, an absolute pressure transmitter , a pressure gauge and pressure
gauge transmitter as our device to measure the pressure. There are 2 experiments that had
been held, the first experiment is pressure measurement and the second is calibration
check.
For the first experiment we have to run seven times to get the reading. For the
first run the suggest pressure is 3 psig, and then 2! psig, 2 psig, "! psig, " psig, ! psig
and lastl# psig. The value of the actual gauge pressure several differences with the
suggested pressure because of the human error and using wrong techni$ue when handle
the devices
INTRODUCTION
This %odel &%' 2 plant is a scaled(down real industr# process plant built on
!ft x !ft steel platform, complete with its own dedicated control panel. The process
e$uipment and process instrumentation are real industrial process t#pe. The plant is
constructed in accordance to industrial process plant standards and practices, with fail(
safe features.
OBJECTIVES
The ob)ective of the experiment was to evaluate and stud# measurement of
pressure * +bsolute &ressure Transmitter &T"-,auge &ressure Transmitter &T2- and
&ressure auge &- / and To check the site calibration of the above pressure measuring
instruments * 0and &ump, 1acuum &ump, %anometer, %ultimeter, 0and 0eld Terminal
to configurecalibrate &T", &T2
-
8/18/2019 Exp 1 - Process Instrument
2/8
RESULTS
TABLE 1
RUN I SET 1 SET 2 AVERAGE+ uggested &ressure, psig 3. 3. 3.
4 +ctual auge &ressure, psig 3."! 3."" 3."3
' auge &ressure, psig 3. 3. 3.
5 auge &ressure, psig 26.3 3." 26.7
8 +bsolute &ressure, psia 93.: 99.! 99."!
F4:
RUN I I SET 1 SET 2 AVERAGE
+ uggested &ressure, psig 2!. 2!. 2!.
4 +ctual auge &ressure, psig 2!.2: 2!.7 2!.96' auge &ressure, psig 2!. 2!. 2!.
5 auge &ressure, psig 2!.3 2!.: 2!.!!
8 +bsolute &ressure, psia 36.: 9.2 9.
F42 I4 = 5I +t &ressure
Indicator,
&"2
0 IF ( 8I +bsolute &ressure5eviation, psia
.": .2 ."6
-
8/18/2019 Exp 1 - Process Instrument
3/8
RUN I I I SET 1 SET 2 AVERAGE
+ uggested &ressure, psig 2. 2. 2.
4 +ctual auge &ressure, psig 2.9 2.2> 2.33
' auge &ressure, psig 2. 2. 2.
5 auge &ressure, psig 2. 2." 2."
8 +bsolute &ressure, psia 39.9 39.9 39.9F4 3!.3
" I4 = 'I auge
&ressure
5eviation, psig
+t &ressure
auge .9
.9
.2>
.">
.33
.2:2 I4 = 5I +t &ressure
Indicator,
&"2
0 IF ( 8I +bsolute &ressure
5eviation, psia
.7 .!> .>3
RUN IV SET 1 SET 2 AVERAGE
+ uggested &ressure, psig "!. "!. "!.
4 +ctual auge &ressure, psig "9.!6 "!.33 "9.6>
' auge &ressure, psig "!. "!. "!.
5 auge &ressure, psig "!.9 "!.9 "!.!
8 +bsolute &ressure, psia 26." 26.: 26.!
F4
-
8/18/2019 Exp 1 - Process Instrument
4/8
RUN V SET 1 SET 2 AVERAGE
+ uggested &ressure, psig ". ". ".
4 +ctual auge &ressure, psig ".2 !.7! 7.:6
' auge &ressure, psig ". ". ".
5 auge &ressure, psig ".2 6.6 ".!
8 +bsolute &ressure, psia 29.7 29.3 29.!F4
.26
.""
."!
.26
2 I4 = 5I +t &ressure
Indicator,&"2
0 IF ( 8I +bsolute &ressure5eviation, psia
.9 ."6 ."2
-
8/18/2019 Exp 1 - Process Instrument
5/8
RUN VII SET 1 SET 2 AVERAGE
+ uggested &ressure, psig . . .
4 +ctual auge &ressure, psig .: .96 .26
' auge &ressure, psig . . .
5 auge &ressure, psig .2 .> .9
8 +bsolute &ressure, psia "9.> "!. "9.:F4
-
8/18/2019 Exp 1 - Process Instrument
6/8
TABLE 3: PT2 SITE CALIBRATION CHECK
RUN I PT2 Site Cali!ati"#
P!"$%&t C"$e: '2(EAE21)2)G1111
Se!ial N%*e!: +))3)),1(3
Ma-i*%* $e.iati"# : )/),+ 0e!" I#i/ E!! )/)3
Ta4 Na*e: PTI 0) 5 ')PSIA
Cali!ati"# Ra#4e: (/3+ 6 7,
A B C D E 8 G H
I#9%t
P!e%!e A99lie$;
9i4
at< "= Cali!ati"#
Ra#4e
O%t9%t
P!e%!e;
9i4
Mea%!e$
O%t9%t;
*A
T>e"!eti&al
O%t9%t; *A
< A&&%!a&?
( )
m+
5
D
mAC − x "
< De.iati"#
( )
( )"x
9(2
m+5
mA
C −
De./
Li*it
Re*a!@
&ass? If 5eviation @ A
5ev. Cimit
Fail ? If Dtherwise
("2.2:2.92 at 9.3@ ( ( 9. ( ( .3 6
.3"!. at 2!.@ "9.: :. :. . . B .7! Pa
"!.33. at !.@ 3." "2. "2. . . B .7! Pa
3.39!. at 22.!@ 9!.2 ">. ">. . . B .7! Pa
9!.3>. at ".@ !6.6 2. 2. . . B .7! Pa
-
8/18/2019 Exp 1 - Process Instrument
7/8
DISCUSSION
From Table ", we can see clearl# the average different between psia and psig is
around "9.7 psia even though we do with the different suggested pressure. This is because
of the value of atmosphere is constant "9.7 psia. Then the value that we obtained is notver# accurate because of the error, such as manufacturing tolerance, signal conditioning
and displa# resolution. The important term that related the sensor and transmitter range of
instrument low and high value range-, span of instrument different high(low- and Eero
of the instrument low value-.
In this experiment we used a %anometer, a 0and &ump, %ultimeter and a 0and
0eld Terminal as the devices to check the calibration of the pressure measuring
instruments. It is to evaluate of pressure using +n +bsolute Transmitter &T"- and +n
&ressure auge &-. In our experiment, there is a different value between actual gauge
pressure and the gauge pressure reading from the measurement device. It could be cause
b# the unstable pressure while the reading is taken. The readings of the hand pump were
difficult to get because the pressure were not static therefore we have to be ver# alert to
get an accurate reading
CONCLUSION
4ased on experiment we have conclude that how to evaluate and stud#
measurement of pressure using the following +n +bsolute &ressure Transmitter, + auge
&ressure transmitter and + &ressure auge. This experiment re$uire us how to use
electronic and mechanical instrument to measure the pressure. e can conclude that the
reading at pressure gauge and pressure indicator have the deviation. &ressure indicator
gives accurate reading than others.
4esides that, this experiment need us to check the site calibration of the above
pressure measuring instrument using the following + 0and &ump, + 1acuum &ump,
%anometer, %ultimeter , and + 0and 0eld Terminal.
-
8/18/2019 Exp 1 - Process Instrument
8/8
TUTORIAL
1/ State >et>e! t>e 9!e%!e i >i4>e!l"e!e%al at t>e "tt"* "= t>e ta#@ =ille$
it> a 4a t>a# at t>e t"9/
The pressure is e$ual at the bottom of the tank filled with a gas than at the top because a gas can be compressed or expanded to fit almost an# volume.
2/ I= t>e &"#te#t "= a ta#@ >"l$i#4 ()*3 "= 4a a!e t!a#=e!!e$ t" a ta#@ it> ti&e a
*%&> ."l%*e a#$ all "t>e! =a&t"! !e*ai# %#&>a#4e$; >at ill >a99e# t" 4a
9!e%!e
hen the value of the volume is decrease, the pressure is also decrease. hen the value
of the volume is increase, the pressure is increase while the other factors remain
unchanged. Therefore, doubling the volume of a gas cuts the pressure in half.
3/ I= t>e 4a i# a ta#@ >eate$ a#$ all "t>e! =a&t"! !e*ai# %#&>a#4e$; >at ill
>a99e# t" t>e 4a 9!e%!e
If the gas in a tank heated and all other factors remain unchanged the gas pressure will
become high and increase. This is because when the heat is transferred from fire to the
tank it makes the atoms or molecules move activel# and directl# proportional to the
temperature.
(/ >at a!e t>e 9!e%!e *ea%!i#4 i#t!%*e#t %e$ i# t>i 9la#t
The pressure measuring instruments used in this plant are; a %anometer +%8T8G
HDF+ '&'(2 IJ5-, a %ultimeter %D58C 5%(36"+-, a pressure
indicator and a pressure gauge.
+/ Di&% !ie=l? t>e 9"ile e!!"! a==e&ti#4 t>e a&&%!a&? i# 9!e%!e *ea%!e*e#t
a "e!.e$ i# t>e e-9e!i*e#t/
• e did not control pressureslow down the valve properl# and affect the pressure
reading.
• 'areless.
• The e$uipment is not working properl#.