Exp 1 - Process Instrument

download Exp 1 - Process Instrument

of 8

Transcript of Exp 1 - Process Instrument

  • 8/18/2019 Exp 1 - Process Instrument

    1/8

    SUMMARY

    In this experiment, we use manometer, multimeter, a vacuum pump; a hand held

    terminal, a hand pump, an absolute pressure transmitter , a pressure gauge and pressure

    gauge transmitter as our device to measure the pressure. There are 2 experiments that had

     been held, the first experiment is pressure measurement and the second is calibration

    check.

    For the first experiment we have to run seven times to get the reading. For the

    first run the suggest pressure is 3 psig, and then 2! psig, 2 psig, "! psig, " psig, ! psig

    and lastl# psig. The value of the actual gauge pressure several differences with the

    suggested pressure because of the human error and using wrong techni$ue when handle

    the devices

    INTRODUCTION

    This %odel &%' 2 plant is a scaled(down real industr# process plant built on

    !ft x !ft steel platform, complete with its own dedicated control panel. The process

    e$uipment and process instrumentation are real industrial process t#pe. The plant is

    constructed in accordance to industrial process plant standards and practices, with fail(

    safe features.

    OBJECTIVES

    The ob)ective of the experiment was to evaluate and stud# measurement of 

     pressure * +bsolute &ressure Transmitter &T"-,auge &ressure Transmitter &T2- and

    &ressure auge &- / and To check the site calibration of the above pressure measuring

    instruments * 0and &ump, 1acuum &ump, %anometer, %ultimeter, 0and 0eld Terminal

    to configurecalibrate &T", &T2

  • 8/18/2019 Exp 1 - Process Instrument

    2/8

    RESULTS

    TABLE 1

    RUN I SET 1 SET 2 AVERAGE+ uggested &ressure, psig 3. 3. 3.

    4 +ctual auge &ressure, psig 3."! 3."" 3."3

    ' auge &ressure, psig 3. 3. 3.

    5 auge &ressure, psig 26.3 3." 26.7

    8 +bsolute &ressure, psia 93.: 99.! 99."!

    F4:

    RUN I I SET 1 SET 2 AVERAGE

    + uggested &ressure, psig 2!. 2!. 2!.

    4 +ctual auge &ressure, psig 2!.2: 2!.7 2!.96' auge &ressure, psig 2!. 2!. 2!.

    5 auge &ressure, psig 2!.3 2!.: 2!.!!

    8 +bsolute &ressure, psia 36.: 9.2 9.

    F42 I4 = 5I +t &ressure

    Indicator,

    &"2

    0 IF ( 8I +bsolute &ressure5eviation, psia

    .": .2 ."6

  • 8/18/2019 Exp 1 - Process Instrument

    3/8

    RUN I I I SET 1 SET 2 AVERAGE

    + uggested &ressure, psig 2. 2. 2.

    4 +ctual auge &ressure, psig 2.9 2.2> 2.33

    ' auge &ressure, psig 2. 2. 2.

    5 auge &ressure, psig 2. 2." 2."

    8 +bsolute &ressure, psia 39.9 39.9 39.9F4 3!.3

    " I4 = 'I auge

    &ressure

    5eviation, psig

    +t &ressure

    auge .9

    .9

    .2>

    .">

    .33

    .2:2 I4 = 5I +t &ressure

    Indicator,

    &"2

    0 IF ( 8I +bsolute &ressure

    5eviation, psia

    .7 .!> .>3

    RUN IV SET 1 SET 2 AVERAGE

    + uggested &ressure, psig "!. "!. "!.

    4 +ctual auge &ressure, psig "9.!6 "!.33 "9.6>

    ' auge &ressure, psig "!. "!. "!.

    5 auge &ressure, psig "!.9 "!.9 "!.!

    8 +bsolute &ressure, psia 26." 26.: 26.!

    F4

  • 8/18/2019 Exp 1 - Process Instrument

    4/8

    RUN V SET 1 SET 2 AVERAGE

    + uggested &ressure, psig ". ". ".

    4 +ctual auge &ressure, psig ".2 !.7! 7.:6

    ' auge &ressure, psig ". ". ".

    5 auge &ressure, psig ".2 6.6 ".!

    8 +bsolute &ressure, psia 29.7 29.3 29.!F4

    .26

    .""

    ."!

    .26

    2 I4 = 5I +t &ressure

    Indicator,&"2

    0 IF ( 8I +bsolute &ressure5eviation, psia

    .9 ."6 ."2

  • 8/18/2019 Exp 1 - Process Instrument

    5/8

    RUN VII SET 1 SET 2 AVERAGE

    + uggested &ressure, psig . . .

    4 +ctual auge &ressure, psig .: .96 .26

    ' auge &ressure, psig . . .

    5 auge &ressure, psig .2 .> .9

    8 +bsolute &ressure, psia "9.> "!. "9.:F4

  • 8/18/2019 Exp 1 - Process Instrument

    6/8

    TABLE 3: PT2 SITE CALIBRATION CHECK 

    RUN I PT2 Site Cali!ati"#

    P!"$%&t C"$e: '2(EAE21)2)G1111

    Se!ial N%*e!: +))3)),1(3

    Ma-i*%* $e.iati"# : )/),+ 0e!" I#i/ E!! )/)3

    Ta4 Na*e: PTI 0) 5 ')PSIA

    Cali!ati"# Ra#4e: (/3+ 6 7,

    A B C D E 8 G H

    I#9%t

    P!e%!e A99lie$;

    9i4

    at< "= Cali!ati"#

    Ra#4e

    O%t9%t

    P!e%!e;

    9i4

    Mea%!e$

    O%t9%t;

    *A

    T>e"!eti&al

    O%t9%t; *A

    < A&&%!a&?

    ( )

    m+

    5

     D

    mAC −  x "

    < De.iati"#

    ( )

    ( )"x

    9(2

    m+5

    mA

    C −

    De./

    Li*it

    Re*a!@

    &ass? If 5eviation @ A

    5ev. Cimit

    Fail ? If Dtherwise

    ("2.2:2.92 at 9.3@ ( ( 9. ( ( .3 6

    .3"!. at 2!.@ "9.: :. :. . . B .7! Pa

    "!.33. at !.@ 3." "2. "2. . . B .7! Pa

    3.39!. at 22.!@ 9!.2 ">. ">. . . B .7! Pa

    9!.3>. at ".@ !6.6 2. 2. . . B .7! Pa

  • 8/18/2019 Exp 1 - Process Instrument

    7/8

    DISCUSSION

    From Table ", we can see clearl# the average different between psia and psig is

    around "9.7 psia even though we do with the different suggested pressure. This is because

    of the value of atmosphere is constant "9.7 psia. Then the value that we obtained is notver# accurate because of the error, such as manufacturing tolerance, signal conditioning

    and displa# resolution. The important term that related the sensor and transmitter range of 

    instrument low and high value range-, span of instrument different high(low- and Eero

    of the instrument low value-.

    In this experiment we used a %anometer, a 0and &ump, %ultimeter and a 0and

    0eld Terminal as the devices to check the calibration of the pressure measuring

    instruments. It is to evaluate of pressure using +n +bsolute Transmitter &T"- and +n

    &ressure auge &-. In our experiment, there is a different value between actual gauge

     pressure and the gauge pressure reading from the measurement device. It could be cause

     b# the unstable pressure while the reading is taken. The readings of the hand pump were

    difficult to get because the pressure were not static therefore we have to be ver# alert to

    get an accurate reading

    CONCLUSION

    4ased on experiment we have conclude that how to evaluate and stud#

    measurement of pressure using the following +n +bsolute &ressure Transmitter, + auge

    &ressure transmitter and + &ressure auge. This experiment re$uire us how to use

    electronic and mechanical instrument to measure the pressure. e can conclude that the

    reading at pressure gauge and pressure indicator have the deviation. &ressure indicator 

    gives accurate reading than others.

    4esides that, this experiment need us to check the site calibration of the above

     pressure measuring instrument using the following + 0and &ump, + 1acuum &ump,

    %anometer, %ultimeter , and + 0and 0eld Terminal.

  • 8/18/2019 Exp 1 - Process Instrument

    8/8

    TUTORIAL

    1/ State >et>e! t>e 9!e%!e i >i4>e!l"e!e%al at t>e "tt"* "= t>e ta#@ =ille$

    it> a 4a t>a# at t>e t"9/

      The pressure is e$ual at the bottom of the tank filled with a gas than at the top because a  gas can be compressed or expanded to fit almost an# volume.

    2/ I= t>e &"#te#t "= a ta#@ >"l$i#4 ()*3 "= 4a a!e t!a#=e!!e$ t" a ta#@ it> ti&e a

    *%&> ."l%*e a#$ all "t>e! =a&t"! !e*ai# %#&>a#4e$; >at ill >a99e# t" 4a

    9!e%!e

    hen the value of the volume is decrease, the pressure is also decrease. hen the value

    of the volume is increase, the pressure is increase while the other factors remain

    unchanged. Therefore, doubling the volume of a gas cuts the pressure in half.

    3/ I= t>e 4a i# a ta#@ >eate$ a#$ all "t>e! =a&t"! !e*ai# %#&>a#4e$; >at ill

    >a99e# t" t>e 4a 9!e%!e

    If the gas in a tank heated and all other factors remain unchanged the gas pressure will

     become high and increase. This is because when the heat is transferred from fire to the

    tank it makes the atoms or molecules move activel# and directl# proportional to the

    temperature.

    (/ >at a!e t>e 9!e%!e *ea%!i#4 i#t!%*e#t %e$ i# t>i 9la#t

    The pressure measuring instruments used in this plant are; a %anometer +%8T8G

    HDF+ '&'(2 IJ5-, a %ultimeter %D58C 5%(36"+-, a pressure

    indicator and a pressure gauge.

    +/ Di&% !ie=l? t>e 9"ile e!!"! a==e&ti#4 t>e a&&%!a&? i# 9!e%!e *ea%!e*e#t

    a "e!.e$ i# t>e e-9e!i*e#t/

    • e did not control pressureslow down the valve properl# and affect the pressure

    reading.

    • 'areless.

    • The e$uipment is not working properl#.