ISWS CR 12 Loan c.2 ASTIA Doc. No. AD-152596 AFCRC-TR-58-246 ILLINOIS STATE WATER SURVEY METEOROLOGIC LABORATORY at the UNIVERSITY OF ILLINOIS Urbana, Illinois EVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955 —March 31, 1958 The research reported in this document has been sponsored by the Geophysics Research Directorate of the Air Force Cambridge Research Center, Air Research and Development Command, Cambridge, Massachusetts.

Transcript of Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD...

Page 1: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

ISWS CR 12 Loan c.2

ASTIA Doc. No. AD-152596 AFCRC-TR-58-246

ILLINOIS STATE WATER SURVEY

METEOROLOGIC LABORATORY at the

UNIVERSITY OF ILLINOIS Urbana, Illinois

EVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR

by

K. E. Wilk

FINAL REPORT on

Contract No. AF 19(604)-]395

March 16, 1955 —March 31, 1958

The research reported in this document has been sponsored by the Geophysics Research Directorate of the Air Force Cambridge Research Center, Air Research and Development Command, Cambridge, Massachusetts.

Page 2: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

ASTIA Doc. No. AD-152596 AFCRC-TR-58-246

Illinois State Water Survey Meteorologic Laboratory

at the University of Illinois

Urbana, Illinois

Evaluation of the AN/APQ-39(XA-3) Cloud Detector Radar

by K. E. Wilk

FINAL REPORT on

Contract No. AF 19(604)-1395 March 16, 1955 - March 31, 1958

G. E. Stout Project Director Wo Co Ackermann

Chief

The research reported in this document has been sponsored by the Geophysics Research Directorate of the Air Force Cambridge Research Center, Air Research and Development Command,, Cambridge, Massachusetts.

Page 3: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

TABLE OF CONTENTS

Page ABSTRACT 1 ACKNOWLEDGMENTS 1 INTRODUCTION 2

History 2 Evaluation Site 3 Radar Installation . . 4 Radar Characteristics 8 Beam Collimation . .8

THEORETICAL DETECTION CAPABILITIES 9 Minimum Detectable Reflectivity 10 Cloud Reflectivity .12

EMPIRICAL DETECTION CAPABILITIES 17

Control Data . . . . .18 Radar Data Collection . 19 Radar Data Analysis 20 Cloud Detectability 21 Detection of Precipitation 25 Cloud Differentiation . . . . 26 Invisible Targets 29

SYNOPTIC INTERPRETATION . 31 REFERENCES . . . . . . . . . . . . . . . . 3 9

Page 4: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

LIST OF ILLUSTRATIOKS Figure Page

1 Average Monthly Occurrence of Cyclone Centers, Illinois, 1949-1955 . . . . .. ... 5

2 Average Monthly Hourly Cloud Occurrences, Springfield, Illinois, 1949-1955 . . . . . . . 5

3 AN/APQ-39 Radar Installation . . . . . . . . . 6 4 AN/APQ-39 Radar (Modified) . . . . . . . . . 7 5 Attenuation by Water Vapor and Oxygen

(after Ryde) .............................. 13 6 Attenuation in Cloud (after Ryde) . . . . . . 13 7 Low Cloud and Associated Radar Echo . . . . . 27 8 Middle Cloud and Associated Radar Echo . . . . 28 9 High Cloud and Angels and Associated

Radar Echoes ...............................30

10 Air Masses and Fronts and Associated RADAR ECHOES................................32

11-13 Radar and Synoptic Composite . . . . . . . . . 35-37

LIST OF TABLES Table Page 1 Cloud Reflectivity for 0.86 cm Radar . . . . . 15 2 Determination of Maximum Detectable Range . . 16 3 Control Data Collection . . . . . . . . . . . 18 4 Skew-T, Log P Diagram Cloud Analysis . . . . . 19 5 Cloud Detectability . . . . . . . . . . . . . 22 6 Measurements of Cloud Bases and Tops . . . . . 24

7 Characteristics of Radar Echoes Associated with Variotis "winter Air Masses and Frontal Zone .......................................33

Page 5: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

ABSTRACT This report is an evaluation of a ground based AN/APQ-39

(XA-3) airborne radar as a cloud indicator during the winter months in the Midwest. Data on clouds, precipitation, and other meteorological parameters were collected by means of a modified AN/APQ-39(XA-3) radar, meteorological instruments, and cloud photography. The observed data and radar characteristics are analyzed to determine the empirical detection capabilities, theoretical detection capabilities, and the synoptic interpre­tations of the radar echoes. The results indicate the AN/APQ-39 (XA-3) radar is capable of detecting 77 percent of the observed clouds. Cloud genera usually can be identified from the con­figuration of the radar return. The radar also provides limited knowledge of the existence and character of precipitation and of the air mass structure.

ACKNOWLEDGMENTS This report was written under the direction of William C.

Ackermann, Chief of the Illinois State Water Survey. Research was accomplished under the general guidance of the Project Director, Glenn E. Stout, Head, Meteorology Section.

The author wishes to express appreciation to Donald Staggs, Project Engineer, for the installation, maintenance and modifi­cation of the radar. Special credit is also due Jack E. Taylor, Meteorological Aide, for assistance in data collection and analysis.

Page 6: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

-2-

Credit is due Ralph J. Donaldson, Jr., Geophysics Research Directorate, Air Force Cambridge Research Center and R. B. Leasure, Wright Air Development Center, for suggestions and guidance during the study.

INTRODUCTION The problem of obtaining measurements of the vertical

distribution of cloud and moisture through the troposphere confronts both the research and the operational meteorologist. Visual point observations are necessarily very subjective, and reinforce the need for better instrumentation. The ceiling light and ceilometer provide limited, but objective measurements of the lower limit of the cloud distribution. However, the determination of the vertical dimension relies mainly upon the analysis of occasional Rawinsonde data and observations from transitory aircraft.

The application of vertically pointing K-band radar as a cloud detector has been investigated in recent years. Excellent evaluations of 1.25 cm wavelength radar have been made by Plank, Atlas, and Paulsen, and Leasure and Thompson.(2)

This report follows the format of investigation of these studies but as applicable to 0.86 cm radar, which will hereafter be referred to as the APQ-39 radar.

History The contract originated on March 16, 1955 as an in-flight

Page 7: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

-3-

evaluation program of the APQ-39 airborne radar. After detailed planning, subsequent instrumentation and installation, final flight tests were attempted. In February 1956 the aircraft crashed with complete loss of equipment and personnel.

In September 1956, the initial program was revised by Supplemental Agreement No. 1 and in April 1957, by Supplemental Agreement Mo. 2. The latter revision directed the contractor to evaluate the APQ-39 radar as a ground-based unit.

The APQ-39 radar was transported from Wright Air Develop­ment Center to the University of Illinois Airport in Movember 1956. Installation of the equipment at the new location was completed in February 1957. The period from February 1957 to August 1957 was devoted to trouble shooting and the correction of numerous equipment malfunctions. Routine operation was attempted on several occasions, but equipment failures repeat­edly interrupted data collection. The repetition of specific malfunctions indicated the need for a major change in equipment design. Personnel at Wright Air Development Center concurred and, in October 1957, an additional antenna was acquired. The modification to a twin antenna system and elimination of the duplexer was completed in December 1957. Routine data collec­tion was carried on until February 28, 1958.

Evaluation Site The evaluation of the APQ-39 radar was conducted at the

Illinois State Water Survey Meteorologic Laboratory which is

Page 8: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

-4-

located at the University of Illinois Airport (CMI). The site is five miles south of Champaign, 75 miles east-northeast of Springfield, and 83 miles southeast of Peoria. The surrounding area is primarily rural with no significant variations in terrain elevation.

The Gulf of Mexico provides the principal source of mois­ture advected over Illinois. The average monthly occurrence of cyclone centers in the state, as shown in Figure 1, reaches a maximum in January with a secondary peak in March. The average cloudiness as reported by ground observers,(3) shows the relative monthly distribution of the major cloud types, Figure 2.

Radar Installation The radar was installed in a "window" mounting between the

TPS-10 and CPS-9 towers, as shown in Figure 3. The enclosed antenna housing in the enlarged portion of Figure 3 constituted the original installation before modification to a twin antenna system. After removal of the duplexer, it became the receiver unit. The open antenna on the left surmounts the remoted transmitter unit.

The interior installation of the radar components is shown in Figure 4. The cables running to the opening in the upper left of Figure 4c lead to the receiver, Figure 4a, and the transmitter, Figure 4b.

Page 9: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

FIG. 1 AVERAGE MONTHLY OCCURRENCE OF CYCLONE CENTERS ILLINOIS, 1949-1955

FIG. 2 AVERAGE MONTHLY HOURLY CLOUD OCCURRENCES SPRINGFIELD, ILLINOIS 1949-1955

Page 10: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

FIG. 3 AN/APQ-39 RADAR INSTALLATION A. CPS-9 Radar Tower, B. TPS-10 Radar Tower, C. AN/APQ Radar Set — Insert, C-1. Transmitter, C-2. Receiver

Page 11: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

FIG. 4 AN/APQ-39 RADAR (MODIFIED)

Page 12: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

-8-

Radar Characteristics The principal characteristics of the APQ-39 radar are as

follows: Value

1. Frequency (mean) 34.5 Kmc 2. Power Output

Peak 9.18 kw* Average 5.01 watts

3. Pulse Width 0.8 µ sec. 4. Pulse Repetition Rate 683 pps 5. Receiver Bandwidth 1.5 mcs 6. Receiver I-F Frequency 30 mcs 7. Receiver Sensitivity -100 dbm 8. Beam Width 0.6 degrees 9. Wavelength 0.86 cm

Beam Collimation Repeated failures of the crystal, TR tube, and ATR tube

suggested the elimination of the duplexer and resulted in the decision to use a twin antenna installation.

The use of separate antennas for transmitting and re­ceiving required approximate beam coincidence. Theoretical considerations based on results obtained by Robbiani and Swingle, emphasized the need for variable antenna tilt so as to hold the signal loss to a prescribed minimum. For example, conceding a received signal loss of one db, it would be necessary to employ three antenna tilt angles in order to cover a height interval of 840 to 60,000 feet. However, predetermination of the desired height subdivisions was very difficult since the vertical distribution of the targets was unknown. Also, in many

*The peak power output was measured with the AN/UPM-14(XN-14) Test Set and directional couplero Values as high as 16 kw have been measured by Aerial Reconnaissance Laboratory, Wright Air Development Center.

Page 13: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

-9-

instances it was necessary to sample widely spaced intervals in the vertical depending on the synoptic conditionso Therefore, in the APQ-39 evaluation, a compromise was established between height intervals and received signal loss. Two antenna tilt angles were selected. On "high" tilt a one db loss existed between 3200 and 30,000 feet. On "low" tilt a one db and two db loss occurred at the upper (3200 feet) and lower (1200 feet) extremes, respectively. The acceptance of a two db loss at 1200 feet seemed justified considering that this low level afforded less range and intervening target attenuation. Also, the precipitation type target common to the extreme lower levels has a high threshold of detectability. Antenna tilt was checked empirically using optimum detection of cloud and precipitation targets at the two predetermined levels (approx­imately 2300 and 6700 feet) corresponding to zero received signal loss.

THEORETICAL DETECTION CAPABILITIES Before considering the empirical data analysis, it is

useful to gain an a priori understanding of the theoretical aspects of the APQ-39 radar capabilities. Although numerous assumptions must be made in the following evaluation, the role of individual radar characteristics and the importance of moisture variation becomes more evident.

Also, since many of the values of the radar and

Page 14: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

-10-

atmospheric parameters used exceed the practical accuracy obtainable, it is obvious that the numerical conclusions cannot be taken as absolute, but only as relative.

Minimum Detectable Reflectivity The basic radar - range equation of Austin(5) is used

in the following to define the minimum detectable reflectivity of the 0.86 cm radar.

where: , average echo power, watts Pt, peak transmitted power = 9.l8 x 103 watts G , antenna gain = 9.2 x 10 λ , wave length = 0.86 cm Ø, θ, effective beam widths = 0.60 degrees h , pulse length = 240 meters η , reflectivity per unit volume, cm-1

κ , transmission factor R , target range, km

evaluating C: C = 6.lxl0-16ptG2λ2 Øθh = 3.03

solving for η :

converting E to µ seconds and to milliwatts:

Page 15: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

-11-

taking logarithms: 10 log η = -10 log C1 + 20 log R µ sec + S + A

where S, measured receiver sensitivity = -100 dbm A, attenuation

inserting values for C, and S: 10 logη = -151.28 + 20 log R + A

which expresses the minimum detectable reflectivity as a function of range.

Determination of A: at any one range,

A = Awv + Acld + Ap where: Awv is attenuation by atmospheric water vapor (Wwv ),

«

Acld is attenuation by intervening cloud (W c l d), Ap is attenuation by precipitation (Wp).

Values of Wwv applicable to various types of Midwest air masses are approximated as follows:

Water vapor content Air mass average to 400 mb Pc, winter, polar Canadian 0.5 g/m3

Pc, summer, polar Canadian 3.6 g/m3

Npp, winter, modified maritime polar 1.7 g/m3

Tg, winter, Gulf 7.5 g/m3

Tg, summer, Gulf 12.7 g/m3

Values of cloud liquid water content are obtained from Diem's cloud drop samples.

Page 16: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

-12-

Cloud type Average water content Cumulus 0.50 g/m3 Stratocumulus 0.13 g/m3

Stratus 0.19 g/m3

Altostratus - Altocumulus 0.22 g/m3

Nimbo stratus 0.31 g/m3

Figures 5 and 6, which are based upon work by Ryde,(7)

express Wwv and Wcld in terms of attenuation at 0.86 cm wave­length.

Haddock(8) has determined Ap as a function of Wp where Wp is expressed in surface rainfall rate. The relationship has an approximate value of 0.07 db/1000 feet/mm/hour and is assumed applicable to the freezing level.

Thus, minimum detectable reflectivity values can be calculated assuming various ranges and various synoptic con­ditions, and compared to the reflectivity from various cloud types.

Cloud Reflectivity Rayleigh scattering and Diem's cloud drop data are used

to determine cloud reflectivity. For Rayleigh scattering,

where: η1 = reflectivity, (cm )

m = refractive index -1

Page 17: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

FIG. 5 ATTENUATION BY WATER VAPOR AND OXYGEN (AFTER RYDE) ONE WAY ATTENUATION AT λ = 0.86cm.

FIG. 6 ATTENUATION IN CLOUD (AFTER RYDE) ONE WAY ATTENUATION AT λ = 0.86cm.

Page 18: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

- 1 4 -

λ = wavelength, (cm) Z = function of the number and diameter of the

particles, (µ 6/m3). The quantity (m2-l)2/(m2+2)2 has a value of 0.89

at λ = 0.86 cm. evaluating: η1 = 4.94 x 10-28 Z (for water) Bartnoff and Atlas (9) have found that,

where: G, a dimensionless factor, has a preferred value of 1.35 , density of particles (g/cm3) has a value of 1 for water

do, median volume diameter (µ) W, liquid water content, (g/m3)

10 , constant for converting Z to units of µ6 /m3

then: η1 = 12.85 x 10 - 1 6 do

3W

and: 10 log η1 = -l48.9 + 30 log do + 10 log W

which expresses cloud reflectivity as a function of median volume diameter and liquid water content.

Using Diem's cloud drop samples, the cloud reflectivities were computed. Table 1 lists values of 10 log η1 for various cloud types.

Page 19: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

-15-

TABLE I

CLOUD REFLECTIVITY FOR 0.86 cm RADAR

Cloud Type 30 log 10 log 10 log η1 3 db

1. Stratus 34.5 -7.2 -121.6 2. Altostratus- 34.5 -6.6 -121.0

Altocumulus 3. Stratocumulus 31.4 -8.9 -126.4 4. Cumulus1 35.4 -4.4 -117.9 5. Cumulus2 35.9 -3.0 -116.0 6. Nimbostratus 39.0 -5.1 -115.0

Individual cloud detectability can be expressed as a function of maximum detectable range assuming various air masses and vertical moisture distributions. Table 2 illustrates the importance of the parameters in the determination of maximum detectable range.

For example, assume a synoptic situation consisting of a semi-tropical air mass in summer with a single As-ac cloud layer from 14,000 to 16,000 feet. As can be seen from Table I, the cloud reflectivity would be -121.0 db. The minimum radar detectable signal at a range of 15,000 feet (approximately 28.4 micro-seconds) would require a reflectivity of -122.2 db neglecting attenuation. Considering the attenuation by the water vapor (Figure 5) of the air mass whose average liquid

10 log microns g/m

Page 20: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

TABLE 2 DETERMINATION-OF MAXIMUM DETECTABLE RANGE

Air Mass 1. Pc, polar Canadian, winter and summer 2. Npp, modified mP, central U. S., winter 3. Tg, mT, Gulf, winter and summer

Wwv

1. Pc, winter, 0.5 g/m3 Pc, summer, 3.6 g/m3

2. Npp, winter, 1.7 g/m3 (Average from SFC to 400 mb) 3. Tg, winter, 7.5 g/m3, Tg, summer, 12.7 g/m3

W c l d

Diem's cloud samples, g/m3

Cu 0.36; Cu 0.50; Sc 0.13; St 0.19; As-Ac 0.22; Ns 0.31

W P Rain, mm/hr

A wv Figure 5 (one way attenuation)

Acld Figure 6 (one way attenuation)

AP Rain, 0.07 db/1000 ft/mm/hr to ht≥OCC (one way attenuation) A A w v + A c l d + A p

Maximum Detectable Range

10 log = -151.28 + 20 log R sec + A

Page 21: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

-17-

water content is 12.7 g per cubic meter, the required reflectivity would then become -121.1 db and therefore the cloud would be detectable. However, if a layer of stratocumulus were present from 1500 to 5500 feet, an additional factor of attenuation due to the intervening cloud would increase the required reflec­tivity by 0.2 db which would result in the upper layer becoming nond et ec table.

Although the individual attenuation effects caused by water vapor, intervening cloud, and precipitation are extremely small, the cumulative effect on a target which is already near the threshold of detectability can be critical. Obviously, the range factor is by far the most important parameter to consider. However, one must also consider the intervening environment of the target in evaluating the capabilities and limitations of the cloud detection radar.

EMPIRICAL DETECTION CAPABILITIES

Control Data The evaluation of the APQ-39 radar was accomplished by

integrating the various control data into target definition. Analysis of the radar echo required the knowledge of several basic parameters concerning the target and its environment. Table 3 lists the control data, the source, and the order of acceptance of the source,,

Page 22: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

-18-

TABLE 3

CONTROL DATA COLLECTION SOURCE

Skew-T Dome Pilot Service "A" and Pibal Diagram Camera Report Local Observations

PARAMETERS Freezing Level 1 2

Temperature Inversion 1 2

Air Mass Identity 1

Cloud Temperature 2 1

Cloud Top 2 1 Cloud Base 2 4 1 3 Cloud Amount 4 2 3 1 Cloud Type 3 2 1

Cloud types and amounts were extracted at 30-minute intervals from the dome camera film or from local observations. Cloud heights were obtained from local observations. Pilot reports and balloon observations vere used when considered representative of the conditions in the local area. The two daily Radiosonde observations from the Weather Bureau station

Page 23: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

-19-

at Peoria were independently analyzed and the data applied for a period of one to six hours, depending on the stability of the synoptic conditions. The skew-T, log P diagram cloud analysis was held as objective as possible, using the pre­selected criteria listed in Table 4.

TABLE 4

SKSW-T, LOG P DIAGRAM CLOUD ANALYSIS

Cloud Type Cloud Base Cloud Top Cloud Amount Low and Middle 3°C temp, dew pt. 3°C temp, dew Temp, dew pt. Stratified spread pt. spread spread

o o Low Convective - Level of 0 C to 3 C Convective Condensation-Level decrease in overcast

positive area 3°C to 5°C broken

Middle Level of Free Top of positive 5°C to 7°C Convective Convection area or 3°C temp, scattered

dew pt. spread High Stratified Average of height Height of and Convective of -40°C isotherm tropopause minus

and 90% humidity- 2000 feet contrail curve

Radar Data Collection Since the APQ-39 was designed as an airborne radar, airborne

application of results obtained in a ground-based evaluation is limited. However, to simulate airborne presentation, data were collected simultaneously with two continuous strip cameras. One camera employed a film speed of approximately one inch per fifteen

Page 24: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

-20-

minutes with a fixed vertical scale of 50,000 feet. The second camera operated at a film speed of approximately one and one-half inches per minute with a variable vertical scale. The presenta­tion on the slow moving film approximates that of the fast camera in an aircraft flying 22.5 times the target velocity. Also, for analysis purposes, the dual data collection permitted examination of the echo both in detail and composite form.

Radar Data Analysis The radar presentation was analyzed at 30-minute intervals.

Individual echoes, or layers of echo, were examined for measure­ments of bases, tops, horizontal continuity, and vertical character. Additional information, such as slope, uniformity of bases and tops, and sharpness of boundary, was extracted as an average, applicable to the total period of operation.

A total of 34 days of APQ-39 radar data was analyzed in this manner. The choice of 30-^inute intervals for data extrac­tion was for two reasons. First, numerous equipment malfunctions resulted in sporadic data collection until the second antenna installation in December 1957. Consequently, the treatment of days as separate cases with singular observations was not possible. Secondly, the wide variations of the targets in time and space seemed to justify, if not necessitate, the use of a short time interval. Therefore, each 33-minute radar observation was assumed to be independent and evaluated as a single case.

Page 25: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

-21-

Cloud Detectability Table 5 summarizes the 561 cases investigated for the

nine types of targets observed. It was found that 77 percent of all observed clouds were detectable. Since most of the cases represent wintertime conditions, the results may not be applicable to the other three seasons. The detectability of cumulus, for example, was definitely biased by the large per­centage occurrence of associated precipitation.

The number of detectable cases in maritime and continental air reflects the comparative number of occurrences of the cloud in the air masses. As would be expected, the majority of the observed total cloudiness is associated with the maritime con­dition. However, the magnitude of the contrast is somewhat unreal. The high wintertime occurrence of low cloud in con­tinental air resulted in the surface observer's inability to see and identify any middle or high cloud which might have been present.

The slight difference in the percentage detectability in the two types of air masses is not significant. Although one might suspect that maritime air would require a higher minimum detectable signal because of the increase.in attenuation due to water vapor.

The effect of target temperature appears to be indicative of the variance expected between winter and summer conditions. Both stratocumulus and altocumulus-altostratus exhibit better detectability at lower temperatures. However, temperature

Page 26: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

TABLE 5 CLOUD DETECTAMLm

CLOUD TIPE LOW "Stratus 106 0 100 21 85 100 100 9* +0 - 33* 9* 42* 27* 53* 28 26 Stratocumulus 102 25 80 66 36 85 73 27 29* -6 +1 26 25* 27 53 34 61 12 9 Cumulue 10 9 53 7 3 50 60 22 17 +5 - 23 13 22 45 - 51 6 60 Niabostratus 17 0 100 12 5 100 100 21 - -5 - 97 21 118 - 99 6 35 TOTAL 235 34 87 106 129 85 90 20 28 -2 45 19 20 65 - 66 52 19

MIDLLK 1 Altocumulus 30 28 52 19 11 43 79 120 90 -18 - '38 - 120 158 125 149 5 9 Altostratua 38 5 88 30 8 85 100 108 - - - 108 - 130 189 4 9 Altocumulus- 71 13 85 64 7 88 64 98 74 -11 -1 12 4 98 110 115 151 17 20 Altostratua TOTAL 139 46 75 113 26 74 79 109 82 -15 25 4 109 134 123 162 26 14

HIGH Cirrus 22 35 39 27 5 55 63 304 289 -45 -44 22 51 304 326 213 253 Cirrostratus 36 14 72 30 6 68 100 218 230 -41 -40 93 94 218 311 221 251 TOTAL 58 49 54 57 11 38 33 261 260 -43 -42 56 73 261 319 217 252

TOTAL ALL CLOUDS 432 129 77 276 166 75 87

*feet x 102

Page 27: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

-23-

observations of the other cloud types were too limited to support of deny this relationship or to suggest a liminal value.

There is little apparent relationship between target thick­ness and detectability. This is understandable when one considers the pulse length of the APQ-39 radar. Assuming a filled beam volume, the minimum thickness of a threshold target would be approximately 480 feet. Cloud layers and measurements which approached this condition were too infrequent to illustrate the effect of thickness. On the other hand, one would suspect that the drop size would be directly related to the cloud thickness. This factor is apparent in the case of cumulus in which the mean thickness of the detectable cases is 1000 feet greater than that of the non-detectable cases.

The comparison between estimated cloud bases and tops and radar echo bases and tops does not directly represent deviation between target and indicated target dimensions. The cloud parameters were observed or in some cases, forecasted, using the criteria listed previously. Therefore the evaluation is related mainly to the observational technique involved.

The limited number of balloon and aircraft measurements with associated radar data are listed separately in Table 6. On many occasions the two observations were widely separated in space depending on the drift of the balloon and the height of the target. Since radar echo base fluctuations of 500 feet per minute were not uncommon, the probable deviation of

Page 28: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

- 2 4 -

measurement is within 500 feet. Therefore, it appears that the integrated echo base is a more representative measurement than the point balloon observation.

TABLE 6

MEASUREMENT OF CLOUD BASES AND TOPS

CLOUD BASE MEASUREMENTS BALLOON RADAR ft. x 102 ft. x 102

l 47 150 144 130 114 130

81 90 48 45 38 35 27 30 26 35 20 15

CLOUD TOP MEASUREMENTS AIRCRAFT RADAR

f t . x 102 f t . x 102

240 270 220 230 90 80 70 40 63 50 45 26 34 35 20 15

The echo-cloud top relationship shows more deviation. However, there was great time and space separation in the data. Also, the small number of observations for comparison prevents the assigning of absolute values to the base and top deviations. Further evaluation should be made on an operational basis where numerous detailed aircraft reports are available.

Page 29: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

-25-

Detection of Precipitation The APQ-39 radar is extremely sensitive to the detection

of surface precipitation. Signal return is visible when rain­fall rates are below the resolution of a standard recording raingage. Rainfall rates of one mm/hr and greater produce a saturated signal. The return from snow is somewhat weaker with the minimum detectable rate quite variable.

The effect of rain attenuation becomes noticeable when the rate exceeds five mm/hr. In one case, the detectable top lowered from 20,000 feet to 15,000 feet. Naturally, this example cannot be extrapolated since the target detectability above the precipitation is important. Although high rainfall rates were not experienced, it can be assumed that the effect would be very detrimental to upper cloud detection. In this respect, in-flight radar operation would be more favorable for cloud detection than would a ground-based installation. The precipitation normally would constitute the terminal target and therefore reduce the consequences of the attenuation. In either case, the important problem is to distinguish between cloud base and precipitation. The differentiation is usually impossible. However, it is feasible in certain cases where cloud is detected before the start of precipitation. By reducing the receiver gain until the target is just visible, the encounter of rain becomes evident due to the saturated condition of its signal.

Page 30: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

-26-

Also, precipitation during the wintertime generally will display greater uniformity in stratified conditions. The separation in low level shower activity is normally impossible. The only indicator is the echo's proximity to the surface.

Cloud Differentiation Examples of seven of the basic cloud genera and their re­

spective radar presentations are illustrated in Figures 7 through 9.

Figure 7a show's a layer of stratocumulus with its typical undulating and rounded elements as portrayed by the dome camera. The associated radar echo reflects its gentle, rolling action with a stratified texture arranged in constrained vertical developments. Stratocumulus can be differentiated from the wintertime cumulus example in Figure 7b by the minute cellular texture of the cumulus. Echo from wintertime cumulus often exhibits a "no return" or weak layer, as shown in this example between 5000 and 7500 feet. The cloud base and light pre­cipitation are evident to 5000 feet with the top strongly out­lined from 7500 to approximately 9000 feet. The echo from stratus in Figure 7c is easily identified by the height, uniformity, and the continuous nature of the base and top.

Figure 8 illustrates the characteristic patterns of middle clouds. In Figures 8a and 8b note the similarity of the alto­cumulus and altostratus echo to the stratocumulus and stratus echo in Figures 7a and 7c, respectively. The combined middle

Page 31: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

FIG. 7 LOW CLOUD AND ASSOCIATED RADAR ECHO

Page 32: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

FIG. 8 MIDDLE CLOUD AND ASSOCIATED RADAR ECHO

Page 33: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

-29-

cloud example in Figure 8c incorporates the uniformity of the altostratus with the texture of the altocumulus.

The high cloud examples in Figure 9 are more difficult to differentiate. The cirrostratus uniformity in Figure 9a is slightly more evident when compared to the more cellular nature of the cirrocumulus in Figure 9b.

It appears that normal cloud observation criteria, e.g. height, structure, and texture, are applicable to echo inter­pretation. Trained observers should experience little dif­ficulty in identifying cloud genera.

However, no attempt should be made to determine cloud species or varieties. These classifications are based on luminance, transparency, and the arrangement of microscopic elements, and are not apparently analogous to any echo characteristics.

Invisible Targets Occasionally, the APQ-39 detects targets apparently invis­

ible to a ground observer. These echoes, normally referred to as "angels", occur at the lower levels and seldom exceed 7000 feet. They have been found to be more prevalent in early afternoon, but have been detected at all hours. There is no apparent association with the prevailing cloud. "Angels" have been detected on days with low cloud as well as on clear days. Figure 9c illustrates a clear sky example of angel occurrence. No theoretical or empirical explanation as to the nature

Page 34: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

FIG. 9 HIGH CLOUD AND ANGELS AND ASSOCIATED RADAR ECHOES

Page 35: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

-31-

or cause of this phenomena is offered in this evaluation.

SYNOPTIC INTERPRETATION The ability of the APQ-39 radar to detect 77 percent of

the total cloud encountered during operations suggests the possibility of radar identification of synoptic patterns. The recognition of air masses, frontal zones, and cyclones is feasible whenever the characteristics are reflected in the associated cloud structure. Boucher, in investigation of

(10) 1.25 cm radar, nas found that 85 percent of wintertime occurrences of precyclonic, cold front or trough, or post-cyclonic synoptic patterns could be readily classified. Data collected with the APQ-39 radar exhibits a similar classifica­tion. Table 7 illustrates the echo characteristics which are typical of well-defined winter air masses and frontal zones. Examples of these are shown in Figure 10. Figure 10a illus­trates the limited, well-defined echo in Continental Polar air. Figure 10b shows the mixed stratified and cellular condition typical of Maritime Polar air. The continuous stratified echo experienced in Maritime Tropical air is illustrated in Figure 10c.

Figure 10d shows the well-defined base and indefinite top characteristic associated with echo in the proximity of a cold front. The echo in Figure 10e reflects the rapid approach of a weak warm front with the overrunning pictured at half-hour intervals as uniform stratified echo lowering and gradually

Page 36: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

FIG. 10 AIRMASSES AND FRONTS AND ASSOCIATED RADAR ECHOES

Page 37: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

TABLE 7 CHARACTERISTICS OF RADAR ECHOES

ASSOCIATED WITH VARIOUS WINTER AIR MASSES AND FRONTAL ZONES Air Mass Frontal

zones Character of echo

Average depth of echo

Continuity of echo

Echo Boundary-

Remarks

Maritime tropical

Stratified 10,000' Continuous Indef­inite

Echo exhibits uniform bands

Maritime polar

Stratified -cellular

10,000' Continuous Indef­inite

Generally stratified with embedded cells

Continen­tal polar

Cellular 10,000' Broken Sharp Echo appears almost coherent clusters of cell

Warm Stratified 20,000' Continuous Indef­inite top, sharp base

Initially broken multilayers, thick­ening and lowering with top streamers continually merging with main layer. Base remains ragged until precip. begins

Cold Cellular 20,000' Broken Indef­inite top, sharp base

Non-uniform echo heavy but short in duration

Occlu­sion

Stratified -cellular

30,000' Broken Indef­inite

Great vertical variation of stratified echo

Page 38: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

thickening. The last illustration, Figure 10f shows the heavy cirrus overrunning, altocumulus, and lower dense stratus which were associated with an occlusion.

An example of the types in composite form is illustrated in the analysis of the intense cyclone of February 26, 27, and 28, 1958. Continuous radar operation throughout the period afforded a complete cross-section of the transgressing air-masses. Representative portions of the APQ-39 radar data are shown in the perspective drawing of the cyclone in Figures 11, 12 and 13. Available TPS-10 and CPS-9 radar data are included in the illustrations to complete the radar perspective.

The low pressure cell progressed from Missouri across Illinois and into Michigan. The first composite shows the occlusion approaching the radar site with the pre-frontal shower activity vividly illustrated by the radar presentation. The second composite illustrates the transition zone near the cyclone center. The two radar recordings show the thin stratified condition changing to a post-cyclonic condition as the low center moved northeastward. The third composite (Figure 13) presents the cross section in the northwest flow.

Summary The APQ-39 radar has proved to be an excellent indicator

of cloud and vertical moisture distributions. It has demonstrated the ability to detect 87 percent of low clouds, 75 percent of middle clouds, and 54 percent of high clouds experienced in the

-34-

Page 39: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

FIG. II RADAR AND SYNOPTIC COMPOSITE 0600 CST 27 FEBRUARY 1958

Page 40: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

FIG. 12 RADAR AND SYNOPTIC COMPOSITE 2100 CST 27 FEBRUARY 1958

Page 41: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

FIG. 13 RADAR AND SYNOPTIC COMPOSITE 1200 CST 28 FEBRUARY 1958

Page 42: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

-38-

Midwest during the winter months. The detectability is superior to that of 1.25 cm wavelength radar as investigated by Plank in which the percentages were found to be 55, 52, and 28 respectively. Echo characteristics are generally analogous to cloud genera and therefore permit good reliability in data interpretation. Limited data indicate the radar is superior to balloon observations for the measurement of cloud bases. The APQ-39 radar is extremely sensitive to the detection of surface precipitation. The effect of rain attenuation on cloud return

becomes noticeable when the rate exceeds approximately five mm/hr.

The radar exhibits great potential as an aid in subjective air mass and frontal analysis. The echo presentation can be incorporated with 3 cm radar RHI and PPI presentations to provide useful 3-dimensional radar perspectives of cloud and precipitation.

The author believes that further investigation will establish the cloud detection radar as an integral part in research and operational radar monitoring of meteorological phenomena.

Page 43: Evaluation of the AN/APQ-39(XA-3) Cloud Detector RadarEVALUATION OF THE AN/APQ-39(XA-3) CLOUD DETECTOR RADAR by K. E. Wilk FINAL REPORT on Contract No. AF 19(604)-]395 March 16, 1955

-39-

REFERENCES

1. Plank, V. G., Atlas, Do, and Paulsen, W. H., "The Nature and Detectibility of Clouds and Precipitation as Deter­mined by 1.25 cm Radar," J. Meteorology. 12, 1958, 358-378.

2. Leasure, R. B., and Thompson, G. A., CWO, Flight Test Results of Cloud Base and Top Indicator. AN/APQ-39(XA-1), WADC Technical Report 53-398, Wright Air Development Center, Dayton, Ohio, September, 1953, 64pp.

3. Changnon, S. A., Jr., and Huff, F. A., Cloud Distribution and Correlation with Precipitation in Illinois, Report of Investigation 33, Illinois State Water Survey, Urbana, Illinois, 1957, 83pp.

4. Robbiani, R. L., and Swingle, D. M., Effect of Collimation on a Two-Antenna Radar System. Engineering Report 1177, Signal Corps Engineering Laboratories, 15 May 1956, 12pp.

5. Austin, P. M., Note on Comparison of Ranges of Radio Set SCR-615-B and Radar Set AN/TPS-10A for Storm Detection, Technical Report 5, Weather Radar Research, Cambridge, Massachusetts Institute of Technology, 1947, 12pp.

6. Willett, Hurd C, American Air Mass Properties, Massachusetts Institute of Technology and Woods Hole Oceanographic Institution, Cambridge, Massachusetts, 1933, ll6pp.

7. Ryde, J. W., The Attenuation and Radar Echoes Produced at Centimeter Wavelengths by Various Meteorological Phenomena; Meteorological Factors in Radio Wave Propa­gation, The Physical Society, London, 1946.

8. Haddock, F. J., Scattering and Attenuation of Microwave Radiation Through Rain, Naval Research Laboratory, Washington, D. C., (unpublished manuscript), 1948.

9. Bartnoff, S., and Atlas, Do, "Microwave Determination of Particle-size Distribution," J. Meteorology, 8, 1951, 130-131.

10. Boucher, R. J., Synoptic - Physical Implications of 1.25 cm Vertical Beam Radar Echoes, Meteorological Radar Studies No. 5, Blue Hill Meteorological Observatory, Harvard University, 1957, 24pp+24pp.