Evaluation of Particle sensors for indoor air quality monitoring ......TSI Aerodynamic Particle...
Transcript of Evaluation of Particle sensors for indoor air quality monitoring ......TSI Aerodynamic Particle...
Evaluation of Particle Sensors for Indoor Air Quality
Monitoring and Smart Building Systems – KPI
Justification (ASHRAE 1756-RP)
Matthew Young, Graduate Research Assistant
Yangyang Zou, Graduate Research Assistant Dr. Andrew May, PI
Dr. Jordan Clark, Co-PI
Submitted to: ASHRAE RP-1756 Project Evaluation Subcommittee
Submitted on: December 20th, 2018
1
EVALUATION OF PARTICLE SENSORS FOR INDOOR AIR QUALITY MONITORING AND SMART BUILDING SYSTEMS
Introduction Because of the difficulty in predicting and controlling indoor PM based on an a priori
understanding of sources and transport, a data-driven approach is an exciting prospect and one
that could have tangible positive consequences for human health. The quickly growing fields of
sensor science, data analytics, and intelligent control of buildings promise numerous paradigm-
shifting advantages over traditional means of predicting and controlling PM in both residential
and commercial buildings. Until recently, monitoring indoor PM concentrations was more
expensive than other less direct means of quantifying indoor air quality.
The low-cost sensors we are testing are exclusively optical sensors. Optical sensors operate by
detecting light originating from light emitting diode (LED) or laser that is scattered as it passes
through a stream of particles passing between the LED/Laser and a photocell. Briefly, light
scattering by particles is a function of particle size distributions, particle concentrations, particle
refractive index, particle shape, environmental variables, wavelength of light source, and other
variables. To convert light signals sensed at the photocell to mass concentration, additional
information, such as the density of the particles, is required.
However, the ability to observe events and produce signals is only the first step needed for a
sensor to be operational. The second step is a calibration/correlation process that relates sensor
outputs (voltages, pulse occupancy ratios, or digital signals) and particle concentrations.
Kulkarni, Baron, & Willeke (2011) suggest that reliable and accurate calibration requires:
“sufficient knowledge of the capabilities and limitations of the instrument, adequate information
on the where the information will be used, appropriate test facilities, proper selection of a desired
test aerosol, a thorough investigation of relevant parameters, and a quality assurance program
that is followed throughout the test.” Once calibration is conducted, in most cases a linear
equation is developed and integrated into sensors software or provided by manufacturers so that
bare sensor outputs can be converted to real concentrations or densities.
However, while PM sensors have become less expensive, the reliability of these inexpensive
sensors is very much in question. After consulting with the Indoor Environment Group at
Lawrence Berkeley National Laboratory (LBNL) and reviewing relevant literature and standards
including IEEE_2700-2014 and the US EPA’s Air Sensor Guidebook (2014), we developed a list
of quantitative and qualitative key performance indicators (KPI’s) for assessing the ability of a
particle sensing device to be used in building applications. The list was further refined with close
consultation with the ASHRAE Project Monitoring Subcommittee, and our final KPI’s include:
Functional Size-specific Range, Accuracy and Precision within this range, Peak-catching Ability,
Meteorological Sensitivity (temperature and humidity), Size Effects, Stability Metrics
(hysteresis, long-term degradation, response to extreme conditions), Manufacturing Consistency,
and Ease of Use. The following sections describe each of these KPI’s.
2
EVALUATION OF PARTICLE SENSORS FOR INDOOR AIR QUALITY MONITORING AND SMART BUILDING SYSTEMS
Functional Size-specific Range Both the minimum and maximum limits of quantification that can be accurately reported by a
device may prove invaluable in determining a sensor’s usefulness in building applications. In
continuous monitoring applications, in which there is a cumulative effect of relatively low levels
of exposure, a sensor must be able to report low readings. In other cases, such as event detection
applications, it may be necessary to measure large concentrations for brief periods. The ideal
sensor will operate reliably across all conditions, but these limits must be evaluated, and a
functional range must be defined.
Limit of detection (LOD) for continuous PM sensors is one such indicator. Establishment of
the lower bound of a sensor’s functional range (which we are using interchangeably with the
term LOD) is important because under relatively clean indoor environments under normal
operating conditions, overall sensor performance may be a strong function of their ability to
accurately measure low concentrations. However, there is no standardized method for defining
the LOD for continuous PM monitoring by low-cost sensors (Wallace et al. 2011). LOD has
been previously defined as what can be detected with reasonable certainty using an analytical
method (Wallace et al. 2011). Wang et al (2015) used the LOD equation LOD = 3σblk/k, where
σblk is the standard deviation of a sensor’s response under “blank” conditions maintained in a
cleaned air chamber, and k is the slope of linear regression slope using incense as the particle
source. Zikova et al. (2017) define LOD as the least sensor concentration when the ratio of the
mean to the standard deviation exceeds 3. Sun et al. (2016) define LOD for gas sensors as LOD
=3σblk/k, similar to Wang et al. (2015). Similarly, Northcross et al. (2013) defined LOD as three
times the standard deviation when a sensor is exposed to a near-zero particle environment
(Northcross et al., 2013).
While the lower LOD has been studied extensively, there is no consensus method for
determining the upper limit of detection. In some cases, sensors can fail to read particle
concentrations far below the manufactures reported range, and in other cases the sensor may
outperform the manufactures reported range. Wang et al. (2016) observed a Syhitech DSM
sensor showed a saturation level at 5,000 µg/m3 even though the manufacture lists an operational
range of 0-1000 µg/m3. Austin et al. (2015) defined the upper limit of detection, or saturation
point, as the concentration at which a 10 µg/m3 increase in the reference monitor (TSI APS
3321) resulted in a less than 0.2 change in the Lo Pulse occupancy of the low-cost sensor
(Shinyei PPD42NS).
3
EVALUATION OF PARTICLE SENSORS FOR INDOOR AIR QUALITY MONITORING AND SMART BUILDING SYSTEMS
Precision and Accuracy
Sensor precision will be assessed on both size-resolved and non-size-resolved bases. Factors
contributing to lack of precision include: different wavelengths of light sources, orientation of
the light source and detector, mode of particle transfer from inlet to sensors, and air flow rate of
different instruments (Manikonda et al., 2016).
Researchers use a variety of metrics to define a sensors accuracy, such as: R2, slope,
intercepts, linearity, or an integrated approach. Even if sensor accuracy is poor, there is often a
linear relationship between low-cost sensors and research-grade equipment. Thus, low accuracy
can be accounted for by using a linear model to develop a calibration curve for the sensor. The
calibration curve is created by comparing a low-cost sensor to a reference sensor and exposing
both sensors to known concentrations. Comparing the sensor’s outputs gives the ability to
generate a best-fit curve, which may be either linear or non-linear equation, that allows the user
to obtain more accurate concentrations from the low-cost sensor outputs. However, in most cases
a linear regression analysis has proven to be the most used method for developing a calibration
equation (Budde et al. 2013, Sousan et al. 2016a, Sousan et al. 2016b, Taylor 2016, Sousan et al.
2017). Sousan et al. (2016b) suggests that any sensor can be used to accurately estimate mass
concentrations once calibrated for a specific aerosol type.
One influential factor in determining a sensor’s accuracy is the time-step used during
analysis. As shown below (Figure 1), Budde et al. (2013) tested a Sharp GP2Y1010 (GP2Y)
against a high-accuracy reference device (TSI Dustrak DRX 8533) and found no correlation
when initially comparing raw-sensor output from the Sharp GP2Y to the TSI DustTrak.
However, when samples were averaged over 1 second intervals the correlation improved
dramatically. And, when samples were averaged over 60 second intervals, there was an even
higher correlation between the sensors.
Zheng et al. (2018) tested five uncalibrated Plantower PMS50003 over various time scales.
The results showed that 1 hour aggregated data had a mean R2 of 0.4, 6 hour aggregated data had
a mean R2 of 0.80, 12 hour aggregated data had a mean R2 of 0.84, and 24 hour aggregated data
had a mean R2 of 0.93. This shows that there are immense improvements in correlations from <1
hour windows up to 6 hour windows, and then gradual improvements for larger windows sizes.
While increasing the time averaging interval may increase agreement in most cases, it may not
be suitable for some end user applications. For example, using a 24 hour time average might be
adequate for gauging outdoor air pollution, but it may fail to be used for specific event detection
applications.
Figure 1: Time Averaging Effect (Budde et al., 2013)
4
EVALUATION OF PARTICLE SENSORS FOR INDOOR AIR QUALITY MONITORING AND SMART BUILDING SYSTEMS
Peak-catching Ability Peak-catching is a sensor’s ability to measure and
react to a short-lived increase in concentration. To
increase applicability, sensors should be able to
accurately detect events that create a rapid increase in
indoor PM concentrations. They should also avoid
reading “false positives” which could inhibit an end-
user’s intended application of the sensor. A sensor’s
peak-catching ability can be measured by how it
responds to predetermined concentrations greater
than its baseline measurement, as well as the time it
takes to reach its peak concentration. Conversely,
sensors should be able to recover after a peak event
and return to accurately reading lower concentrations.
Peak-catching may also differ based on particle
composition. The return to steady-steady is also an
important observation because a sensor should be
able to differentiate whether peaks are linked
events or independent events (e.g., whether there
one event or two).
Several researchers have explored the ability of
low-cost sensors to detect events and defined peak-
catching ability in different ways. Singer & Delp
(2018) showed that peak-catching ability is an area
where low-cost sensors differ dramatically. Some
sensors are able to track quick changes in PM
concentrations with accuracy close to that of a
research-grade instrument, while some miss peaks
completely.
To address peak-catching for personal
exposure in a residential setting, Wallace et al. (2006)
defined an event as a 7 µg/m3 increase above the
baseline concentration. Then, the event was counted as
a single event until the running average fell below 10
µg/m3 and the slope changed from negative to
positive. Chan et al. (2017) identified events as a 10
μg/m3 increase above the baseline concentration.
Event start time was defined as the time when there are three consecutive measurements with
zero or positive change from the previous measurement. The decay period from the event was
defined as three consecutive measurements with zero or negative measurements following the
peak. Events were considered independent if the concentration either (a) dropped more than 50
μg/m3 below the peak concentration, or (b) the concentration dropped by more than 50% of the
peak concentration minus the baseline.
Figure 2: Peak-catching Test of Residential Activities (Wallace et al.,
2006)
Independent
Linked
Figure 3: Linked vs. Independent Events (Chan et al., 2017)
5
EVALUATION OF PARTICLE SENSORS FOR INDOOR AIR QUALITY MONITORING AND SMART BUILDING SYSTEMS
In other studies, researchers have tested
sensor peak catching ability using
controlled events. Manikonda et al. (2016)
found that during a 6-hour test with Arizona
Test Dust (ATD), 3 expected peaks were
observed. The peaks, as recorded by the
TSI Aerodynamic Particle Sizer (APS)
3321, were approximately 475, 200, and
700 µg/m3. When testing ATD, the APS
3321 reported overall higher concentrations
than the other sensors (Speck, Dylos,
AirAssure, and Grimm PAS-1.109). One of
the Speck monitors reported 30% higher
peak concentrations when compared to the other Speck monitor. On the contrary, all three
AirAssure sensors reported peak concentrations 50% lower in comparison to the APS 3321.
Furthermore, there may be differences in peak-catching quantification based on the concentration
of the peak.
Zikova, Hopke, & Ferro (2017) tested 66 Speck sensors in comparison to a Grimm PAS-
1.109, in an indoor residential setting for 3 days. The Grimm 1.109 identified two peaks
exceeding 20 µg/m3, and in both instances the average of the 66 Specks was greater than the
concentrations reported by the Grimm PAS-1.109. In one instance, the Speck average was five
times greater than the reported concentration from the Grimm PAS-1.109 (20 µg/m3 compared to
100 µg/m3). In the other instance, the Speck average was only 10 µg/m3 greater than the
concentration reported by the Grimm PAS-1.109 (45 µg/m3 compare to 55 µg/m3).
Figure 4: Peak-catching Test with ATD (Manikonda et al., 2016)
6
EVALUATION OF PARTICLE SENSORS FOR INDOOR AIR QUALITY MONITORING AND SMART BUILDING SYSTEMS
Meteorological Sensitivity For a variety of reasons, sensors
behave differently in different
environmental conditions, namely
temperature and relative humidity (RH).
Most studies have found that
temperature has a negligible effect on
the accuracy of low-cost sensors,
within a reasonable range of
temperatures. However, Budde et al.
(2013) noted a correlation between a
Sharp GP2Y sensor output and
temperature when compared at a
constant concentration (Figure 5).
The primary concern for RH effect is the hygroscopic growth that will occur in particles,
leading to inaccurate sensor readings. Additionally, water absorbs infrared radiation which
reduces light intensity received by the phototransistor and thus underestimates true
concentrations. High concentrations of water can also cause circuits to fail (Wang et al., 2016).
Different sensors and testing conditions have led to a range of suggested operational RH %, with
some studies indicating 80% RH as the upper limit (McMurry & Stolzenburg, 1989) and others
indicating 60% RH as an upper limit (Han, Symanski, & Stock, 2016).
Jayaratne et al. (2018) tested the Plantower PMS1003 in a chamber setting with a constant
exposure of 10±1 µg/m3. The PMS1003 maintained a steady reading around 9 µg/m3 until the
chamber reached a RH of 78%. At the maximum RH of 89%, the PMS1003 saw an increase to
about 16 µg/m3. Wang et al. (2016) conducted a lab evaluation of low-cost particle sensors,
including Shinyei PPD, Samyoung DSM, and a Sharp GP2Y, compared to the SidePak,
AirAssure, and a scanning mobility particle sizers (SMPS). It was found that as RH increased,
from 20% to 67% to 75% to 90%, sensor outputs would first increase, then begin to decrease
(Figure 6). These results were corroborated with the findings in the review of low-cost sensors
conducted by Rai et al. (2017), which confirmed that sensor outputs would first increase, and
then decrease as RH increased from 20% to 90%.
Figure 6: RH Effects (Wang et al., 2016)
Figure 5: Temperature Effect on Sensor Output (Budde et al., 2013)
7
EVALUATION OF PARTICLE SENSORS FOR INDOOR AIR QUALITY MONITORING AND SMART BUILDING SYSTEMS
Size Effects Sensors must be able to accurately
detect and register particles of different
sizes to improve their applicability in
various environments. Singer & Delp
(2018) found that in some cases neither
consumer nor research grade sensors
responded accurately to emissions
composed entirely of particles smaller than
0.3µm in diameter because they do not
scatter enough light. Additionally,
different sensors may respond more
accurately to smaller or larger particles
depending on their detection mechanisms
(e.g., lower sensor light source wavelength
results in better ability to detect smaller
particles).
Liu at al. (2017) tested four low-cost sensors (Sharp GP2Y, Shinyei PPD42NS (PPD),
Samyoung DSM501A (DSM), and Oneair CP-15-A4 (CP) compared to a Tapered Element
Oscillating Microbalance, and it was found that the slopes of calibration lines for all given
sensors in the cases excluding particles larger than 2.5 µm were higher than those in the cases
including particles larger than 2.5 µm. The author attributes this unlikely outcome to the fact that
low-cost sensors rely on temperature gradients or small fans to drive intake, and larger particles
require higher flowrates.
Wang et al. (2016) exposed a Shinyei PPD, Samyoung DSM, and a Sharp GP2Y sensor to
particles with sizes of 300, 600, and 900 nm. Saturation points for the PPD and DSM sensors for
900 nm particles can be seen at 100 µg/m3, while GP2Y remains linear. While the GP2Y
performed better in the presence of large particles, the PPD and DSM sensors performed better
with smaller particles. This indicates certain sensors perform better depending on particle size.
Figure 7: r2 Values for Different Particle Diameters (Taylor, 2016)
Figure 8: Sensor Performance for Different Sized Particles (Wang et al., 2016)
8
EVALUATION OF PARTICLE SENSORS FOR INDOOR AIR QUALITY MONITORING AND SMART BUILDING SYSTEMS
Stability Metrics A sensor must show both stability in general, meaning the precision and accuracy remain
relatively constant over a certain time period, and resilience in the environments in which they
will operate. These environments may be somewhat harsh, as within an air duct, an air intake, or
a kitchen environment. Preliminary evidence shows that fouling of optical chambers is one
mechanism by which sensors may lose accuracy over time (Austin et al., 2015). Poor stability
and poor resilience can both be addressed through frequent calibrations and maintenance
practices, respectively. But, if the frequency of either is too high, it can be burdensome.
Long-term degradation (drift) quantification Because low-cost sensors will be expected to operate continuously over long periods of time
without frequent calibration or maintenance, the ability of a sensor to give repeatable
measurements over its lifetime is of great importance for building applications. Drift refers to a
gradual change in performance over an extended period of time. To measure drift, it is best to
compare sensor readings at ambient levels over extended time to an initial baseline measurement.
However, this requires leaving a sensor in one location for periods of time ranging from months
to years. Causes of drift could include changes in weather, malfunctioning mechanical parts,
sensor damage from chemical exposure, or a change in intensity of the sensors light source
(Polidori, Papapostolou, & Zhang, 2016).
To evaluate sensor drift, Crilley et al. (2018) left an Alphasense OPC-N2 co-located with a
Grimm PAS-1.108 at an outdoor location for four months. It was found that after applying a RH
corrective factor, there was little observed drift for either sensor. However, drift may be unique
to specific sensors and uses. A study of 29 Speck sensors that were used in different experiments
over a year and a half showed an average monthly drift of -3.1% (Taylor, 2016). The Speck
sensors were obtained from the lending library for the Carnegie Library of Pittsburgh and thus
their specific uses were not known.
Response to extreme conditions For a sensor to successfully be used in occupational settings, they will need to be able to
endure extreme conditions, meaning consistent exposure to high concentrations and peaks. Most
low-cost sensors list their functional range up to concentrations much greater than what would be
expected in ambient conditions. To test the true functional range of a sensor it must be exposed
to increasing concentrations until it reaches its point of failure. It is unproven as to whether
manufacture listed accuracy is consistent at extreme concentrations.
9
EVALUATION OF PARTICLE SENSORS FOR INDOOR AIR QUALITY MONITORING AND SMART BUILDING SYSTEMS
Ease of Use/Communication Protocol/Output “Ease of use” is a broad term that we are using to capture the level of effort required to
implement each PM sensor in practice. For example, we are interested in the ease of sensor
integration with building control systems and existing communication protocols; ease of data
transmission to local or cloud-based archives; and ease of setup. Related to ease of use is the
relative autonomy of the sensors (i.e. can they collect, store, and communicate data without an
excessive frequency of maintenance?). There are essentially two varieties of low-cost sensors we
are testing: integrated devices and bare sensors. Bare sensors require the most steps to get from
sensor reading to applicable outputs. Many of the bare sensors output a voltage or other
electronic signal that must be recorded, and the user is required to use manufacturer-reported
calibration curves to determine particle concentrations. Conversely, integrated devices require
limited user input since the manufacturers include both the sensor and an algorithm to convert
the sensor output signal to particle concentrations in an “off-the-shelf” package. Many integrated
devices also offer built in features such Bluetooth or Wi-Fi connectivity, SD card storage,
meteorological sensors, or compatibility with other smart devices like thermostats. However,
with each feature there are pros and cons, for example Wi-Fi connected device may not work
with networks depending on their security.
10
EVALUATION OF PARTICLE SENSORS FOR INDOOR AIR QUALITY MONITORING AND SMART BUILDING SYSTEMS
Manufacturing Consistency Cost of particle sensors is
often reduced by using less
precise and less consistent
manufacturing methods. For
example, the cheapest sensors are
often die-cast with plastic, leading
to appreciable differences in
critical components such as the
optical chamber. To test a
manufacturers consistency,
multiple sensors (typically three
or more) must be tested at the
same location, and their outputs are then compared. The precision in manufacturing could affect
how sensors perform when used within a connected network.
For example, some studies (Budde et al., 2013; Taylor & Noubakhsh, 2015) suggest that
sensor accuracy can be improved through network average calibration practices. At The Air
Quality Sensor Performance Evaluation Center (AQ-SPEC), Polidori, Papapostolou, & Zhang
(2016) tested sensors in triplicate to determine their intra-modal variability, which is analogous
to our description of manufacturing consistency. The intra-modal variability was calculated
through a set of descriptive statistical parameters such as mean, median, and standard deviation.
If the intra-model variability was less than 20% (better agreement) then the concentrations were
averaged and used for the parameters for which the sensors were being tested (i.e., precision,
accuracy, detection limit, climate susceptibility), but if the intra-modal variability was greater
than 20% (worse agreement) then the sensor concentrations were treated independently for the
parametric evaluation. This methodology for testing manufacturing consistency shows how
consistent manufacturing could be implied to networks containing multiple sensors.
Figure 10: Uncalibrated Performance of Dylos Sensors (Budde et al., 2013)
Figure 11: Comparison of 66 Co-located Speck Sensors (Zikova et al., 2017)
11
EVALUATION OF PARTICLE SENSORS FOR INDOOR AIR QUALITY MONITORING AND SMART BUILDING SYSTEMS
References Austin, E., Novosselov, I., Seto, E., & Yost, M. (2015). Laboratory Evaluation of the Shinyei PPD42NS
Low-Cost Particulate Matter Sensor. PLoS One, 1-17.
Budde, M., Masri, R., Riedel, T., & Beigl, M. (2013). Enabling Low-Cost Particulate Matter
Measurement for Participatory Sensing Scenarios. 12th International Conference on Mobile and
Ubiquitous Multimedia.
Chan, W., Logue, J., Wu, X., Klepeis, N., Fisk, W., Noris, F., & Singer, B. (2018). Quantifying Fine
Particle Emission Events From Time-Resolved Measurements: Method Description and
Application to 18 California Low-Income Apartments. Indoor Air, 89-101.
Han, I., Symanski, E., & Thomas, S. H. (2017). Feasibility of Using Low-cost Portable Particle Monitors
for Measurement of Fine and Coarse Particulate Matter in Urban Ambient Air. Journal of the Air
& Waste Management Association, 330-340.
Holstius, D., Pillarisetti, A., Smith, K. R., & Seto, E. (2014). Field Calibrations of a Low-cost Aerosol
Sensor at a Regulatory Monitoring Site in California. Atmospheric Measurement Techniques,
1121-1131.
Jayaratne, R., Liu, X., Thai, P., Dunbabin, M., & Morawska, L. (2018). The Influence of Humidity on the
Performance of a Low-cost Air Particle Mass Sensor and the Effect on Atmospheric Fog.
Atmospheric Measurement Techniques, 4883-4890.
Jiang, R.-T., Acevedo-Bolton, V., Cheng, K.-C., Klepeis, N., Ott, W., & Hildemann, L. (2011).
Determination of Response of Real-Time SidePak AM510 Monitor to Secondhand Smoke, Other
Common Indoor Aerosols, and Outdoor Aerosol. Journal of Environmental Monitoring, 1695-
1702.
Kulkarni, P., Baron, P., & Willeke, K. (2011). Aerosol Measurement: Principles, Techniques, and
Applications, 3rd Edition. Hoboken: John Wiley and Sons.
Li, J., & Biswas, P. (2017). Optical Characterization Studies of a Low-cost Particle Sensor. Aerosol and
Air Quality Research, 1691-1704.
Liu, D., Zhang, Q., Jiang, J., & Chen, D.-R. (2017). Performance Calibration of Low-cost and Portable
Particular Matter (PM) Sensors. Journal of Aerosol Science, 1-10.
Manikonda, A., Zikova, N., Hopke, P., & Ferro, A. (2016). Laboratory Assessment of Low-Cost PM
Monitors. Journal of Aerosol Science, 29-40.
McMurry, P. H., & Stolzenburg, M. R. (1989). On the Sensitivity of Particle Size to Relative Humidity
for Los Angeles Aerosols. Atmospheric Environment, 497-705.
Northcross, A., Edwards, R., Johnson, M., Wang, Z.-M., Zhu, K., Allen, T., & Smith, K. (2013). A Low-
Cost Particle Counter as a Realtime Fine-Particle Mass Monitor. Environmental Science Process
and Impacts, 433-439.
Polidori, A., Papapostolou, V., & Zhang, H. (2016). Laboratory Evaluation of Low-Cost Air Quality
Sensors. Air Quality Sensor Performance Evaluation Center.
12
EVALUATION OF PARTICLE SENSORS FOR INDOOR AIR QUALITY MONITORING AND SMART BUILDING SYSTEMS
Rivas, I., Mazaheri, M., Viana, M., Moreno, T., Clifford, S., He, C., . . . Querol, X. (2017). Identification
of Technical Problems Affecting Performance of DustTrak DRX Aerosol Monitors. Science of
the Total Environment, 849-855.
Santi, E., Belosi, F., Santachiara, G., Prodi, F., & Berico, M. (2010). Real-Time Aerosol Photmeter and
Optical Particle Counter Comparison. Il Nuovo Cimento, 969-981.
Singer, B. and W.W. Delp (2018). Response of Consumer and Research Grade Indoor Air Quality
Monitors to Residential Sources of Fine Particles. Indoor Air, 1-38.
Sorensen, C., Gebhart, J., O'Hern, T., & Rader, D. (2011). Aerosol Measurement: Principles, Techniques,
and Applications. Hoboken, NJ: John Wiley & Sons, Inc.
Sousan, S., Koehler, K., Hallett, L., & Peters, T. (2017). Evaluation of Consumer Monitors to Measure
Particulate Matter. Journal of Aerosol Science, 123-133.
Sousan, S., Koehler, K., Hallett, L., & Peters, T. M. (2016a). Evaluation of the Alphasense Optical
Particle Counter (OPC-N2) and the Grimm Portable Aerosol Spectrometer (PAS-1.108). Aerosol
Science and Technology, 1352-1356.
Sousan, S., Koehler, K., Thomas, G., Park, J.-H., Hillman, M., Halterman, A., & Peters, T. (2016b). Inter-
Comparison of Low-Cost Sensors for Measuring the Mass Concentration of Occupational
Aerosols. Aerosol Science and Technology, 462-473.
Taylor, M., & Nourbakhsh, I. (2015). A Low-Cost Particle Counter and Signal Processing Method for
Indoor Air Pollution. WIT Transations on Ecology and The Environment, 337-348.
Taylor, M. (2016). Calibration and Characterization of Low-Cost Fine Particulate Monitors and their
Effect on Individual Empowerment (Doctoral dissertation). Retrieved from
https://www.ri.cmu.edu/publications/calibration-and-characterization-of-low-cost-fine-
particulate-monitors-and-their-effect-on-individual-empowerment/
Wallace, L., Wheeler, A., Kearney, J., Van Ryswyk, K., You, H., Kulka, R., . . . Xu, X. (2011).
Validation of Continuous Particle Monitors for Personal, Indoor, and Outdoor Exposures. Journal
of Exposure Science and Environmental Epidemiology, 49-64.
Wallace, L., Williams, R., Rea, A., & Croghan, C. (2006). Continuous Weeklong Measurements of
Personal Exposure and Indoor Concentrations of Fine Particles for 37 Health-Impaired North
Carolina Residents for Up to Four Seasons. Atmospheric Environment, 399-414.
Wang, Y., Li, J., Zhang, Q., Jiang, J., & Biswas, P. (2016). Laboratory Evaluation and Calibration of
Three Low-Cost Particle Sensors for Particulate Matter Measurement. Aerosol Science and
Technology, 1063-1077.
Weekly, K., Rim, D., Zhang, L., Bayen, A. M., Nazaroff, W. W., & Spanos, C. J. (2013). Low-cost
Coarse Airborne Particulate Matter for Sensing Indoor Occupancy Detection. IEEE International
Conference on Automation Science and Engineering (CASE).
Wilson, W., Chow, J. C., Claiborn, C., Fusheng, W., Engelbrecht, J., & Watson, J. G. (2002). Monitoring
of Particulate Matters Outdoors. Chemosphere, 1009-1043.
13
EVALUATION OF PARTICLE SENSORS FOR INDOOR AIR QUALITY MONITORING AND SMART BUILDING SYSTEMS
Zheng, T., Bergin, M. H., Johnson, K. K., Tripathi, S. N., Shilpa, S., Landis, M. S., . . . Carlson, D. E.
(2018). Field Evaluation of Low-cost Particulate Matter Sensors in High and Low Concentration
Environments. Atmospheric Measurement Techniques, 1-27.
Zikova, N., Hopke, P. K., & Ferro, A. R. (2017). Evaluation of New Low-cost Particle Monitors for
PM2.5 Concentrations Measurements. Journal of Aerosol Science, 24-34.