Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength,...

67
Ethicon Edge Competitive Performance Testing Wound Closure 2018 The third-party trademarks used herein are trademarks of their respective owners.

Transcript of Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength,...

Page 1: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon EdgeCompetitive PerformanceTesting

Wound Closure2018

The third-party trademarks used herein are trademarks of their respective owners.

Page 2: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Table of Contents

Sutures

ETHIBOND EXCEL® Polyester Suture . . . . . . 26-29

MONOCRYL® (poliglecaprone 25) Suture. . 30-31

PDS® (polydioxanone) Suture . . . . . . . . . . . . . . . . XX

PROLENE® Polypropylene Suture . . . . . . . . 32-40

Coated VICRYL® (polyglactin 910) Suture and Coated VICRYL® Plus Antibacterial (polyglactin 910) Suture . . . . . . . . . . . . . . . . . . 41-55

VICRYL RAPIDE™ (polyglactin 910) Suture . . XX

STRATAFIX™ Symmetric PDS™ Plus Knotless Tissue Control Device . . . . . . . . . . . . . . 56

STRATAFIX™ Spiral PDS™ Plus Knotless Tissue Control Device . . . . . . . . . . . . 57-58

STRATAFIX™ Spiral MONOCRYL™ PlusKnotless Tissue Control Device . . . . . . . . . . . . . . 58

Regional Sutures

ETHIBOND EXCEL Suture (Asia-Pacific) . . . . 59

Suture Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . 60

Topical Skin Adhesives

DERMABOND ADVANCED® Topical Skin Adhesive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61

DERMABOND® PRINEO® SkinClosure System (22 cm) . . . . . . . . . . . . . . . . . . . . . . 62

Advanced Suturing Systems

PROXISURE™ Suturing Device . . . . . . . . . . . . . . 63

Vascular Access Device Protection

BIOPATCH® Protective Disk with CHG . . . . 64-67

Needles

BV-1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . XX

BV130-5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5

BV175-6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-9

BV175-8 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

C-1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-12

CC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . XX

CC-1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . XX

CT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-14

CT-1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15-16

FS-2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . XX

P-3 PRIME . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . XX

PC-1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . XX

PS-1 PRIME. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . XX

PS-2 PRIME . . . . . . . . . . . . . . . . . . . . . . . . . . . . XX

RB-1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

SH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18-19

SH-1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . XX

V-5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . XX

V-7 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

V-34 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

STRATAFIX™ Knotless Tissue Control Device Needles . . . . . . . . . . . . . . 21-24

Needle Glossary . . . . . . . . . . . . . . . . . . . . 25

Page 3: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Surgical Needles

Competitive Performance Testing

*Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape, and Stiffness characteristics.

Reference: Technical Memo CT17-001. BV130-5 Competitive Assessment. February 10, 2017. Ethicon, Inc. 3

PENETRATION FORCE COMPARISON• BV130-5 EVERPOINT (EVP)

required less penetration

force than MV-135-5 from

1 to 30 passes

• BV130-5 EVERPOINT (EVP)

showed less variability in

penetration force than

MV-135-5 from 1 to 30

passes

• BV130-5 EVERPOINT (EVP)

displayed greater strength

than MV-135-5 regardless

of the angle at which force

was applied

Testing Outcomes

Ethicon Needle Alloy Code

EVERPOINT (EVP) BV130-5

Competitor Code

Medtronic MV-135-5

PENETRATION FORCE STANDARD DEVIATION

Pe

ne

tra

tio

n F

orc

e (

gra

ms)

Pass

70

60

50

40

30

20

10

0

MEDTRONIC, MV-135-5

ETHICON, BV130-5 (EVP)

1 10 20 30

GOOD

MEDTRONIC, MV-135-5

ETHICON, BV130-5 (EVP)

GOOD

Sta

nd

ard

De

via

tio

n

Pass1 10 20 30

Less penetration force required

Less variability in penetration force

Greater strength regardless of angle*

✓✓✓

1.2

3.0

8.0

6.6

5.0

1.3 1.4 1.5

STRENGTH COMPARISON

Be

nd

Mo

me

nt

(gra

m •

cm

)

Angle (degrees)

MEDTRONIC, MV-135-5

ETHICON, BV130-5 (EVP)

30

25

20

15

10

5

0 GOOD

0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90

Page 4: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Surgical Needles

Competitive Performance Testing

*Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape, and Stiffness characteristics.

Reference: Technical Memo CT17-001. BV130-5 Competitive Assessment. February 10, 2017. Ethicon, Inc. 4

PENETRATION FORCE COMPARISON

STRENGTH COMPARISON

• BV130-5 EVERPOINT (EVP)

required less penetration

force than 2xDR6 from

1 to 30 passes

• BV130-5 EVERPOINT (EVP)

showed less variability

in penetration force than

2xDR6 from 1 to 30 passes

• BV130-5 EVERPOINT (EVP)

displayed greater (0-40º)

or statistically equivalent

(45-90º) strength

compared to 2xDR6

Testing Outcomes

Ethicon Needle Alloy Code

EVERPOINT (EVP) BV130-5

Competitor Code

B. Braun 2xDR6

PENETRATION FORCE STANDARD DEVIATION

Pe

ne

tra

tio

n F

orc

e (

gra

ms)

Pass

60

50

40

30

20

10

0

B. BRAUN, 2xDR6

ETHICON, BV130-5 (EVP)

1 10 20 30

GOOD

B. BRAUN, 2xDR6

ETHICON, BV130-5 (EVP)

GOOD

Sta

nd

ard

De

via

tio

n

Pass1 10 20 30

Less penetration force required

Less variability in penetration force

Greater or statistically equivalent strength*

✓✓✓

1.2

3.3

8.5

6.66.0

1.3 1.4 1.5

Be

nd

Mo

me

nt

(gra

m •

cm

)

Angle (degrees)

B. BRAUN, 2xDR6

ETHICON, BV130-5 (EVP)

30

25

20

15

10

5

0 GOOD

0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90

Page 5: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Surgical Needles

Competitive Performance Testing

Reference: Technical Memo CT17-001. BV130-5 Competitive Assessment. February 10, 2017. Ethicon, Inc. 5

PENETRATION FORCE COMPARISON• BV130-5 required less

penetration force than

MV-135-5 from 1 to 30

passes

Testing Outcomes

Ethicon Needle Alloy Code

ETHALLOY BV130-5

Competitor Code

Medtronic MV-135-5

Pe

ne

tra

tio

n F

orc

e (

gra

ms)

Pass

60

50

40

30

20

10

0

MEDTRONIC, MV-135-5

ETHICON, BV130-5

1 10 20 30

BETTER

Less penetration force required✓

PENETRATION FORCE COMPARISON• BV130-5 required less

penetration force than

2xDR6 from 1 to 30 passes

Testing Outcomes

Ethicon Needle Alloy Code

ETHALLOY BV130-5

Competitor Code

B. Braun 2xDR6

Pe

ne

tra

tio

n F

orc

e (

gra

ms)

Pass

60

50

40

30

20

10

0

B. BRAUN, 2xDR6

ETHICON, BV130-5

1 10 20 30

BETTER

Less penetration force required✓

Page 6: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Surgical Needles

Competitive Performance Testing

*Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape, and Stiffness characteristics.

Reference: Technical Memo CT16-001. BV175-6 Competitive Assessment. July 19, 2016. Ethicon, Inc. 6

STRENGTH COMPARISON

PENETRATION FORCE COMPARISON• BV175-6 EVERPOINT (EVP)

required less penetration

force than MV-175-8 from

1 to 30 passes

• BV175-6 EVERPOINT (EVP)

showed less variability

in penetration force than

MV-175-8 from 1 to 30

passes

• BV175-6 EVERPOINT (EVP)

displayed greater strength

than MV-175-8 regardless

of the angle at which force

was applied

Testing Outcomes

Ethicon Needle Alloy Code

EVERPOINT (EVP) BV175-6

Competitor Code

Medtronic MV-175-8

PENETRATION FORCE STANDARD DEVIATION

Pe

ne

tra

tio

n F

orc

e (

gra

ms)

Pass

90

80

70

60

50

40

30

20

10

0

MEDTRONIC, MV-175-8, 8 mil

ETHICON, BV175-6 (EVP), 8 mil

1 10 20 30

GOOD

Be

nd

Mo

me

nt

(gra

m •

cm

)

Angle (degrees)

MEDTRONIC, MV-175-8, 8 mil

ETHICON, BV175-6 (EVP), 8 mil

70

60

50

40

30

20

10

00 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90

GOOD

MEDTRONIC, MV-175-8, 8 mil

ETHICON, BV175-6 (EVP), 8 milGOOD

Sta

nd

ard

De

via

tio

n

Pass1 10 20 30

Less penetration force required

Less variability in penetration force

Greater strength regardless of angle*

✓✓✓

2.5

4.9

8.57.6 7.4

2.3 2.2 2.2

Page 7: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Surgical Needles

Competitive Performance Testing

Reference: Technical Memo CT16-001. BV175-6 Competitive Assessment. July 19, 2016. Ethicon, Inc. 7

Pe

ne

tra

tio

n F

orc

e (

gra

ms)

Pass

90

80

70

60

50

40

30

20

10

0

MEDTRONIC, MV-175-8, 8 mil

ETHICON, BV175-6, 8 mil

1 10 20 30

• BV175-6 required less

penetration force than

MV-175-8 from 1 to 30

passes

• BV175-6 showed less

variability in penetration

force than MV-175-8 from

1 to 30 passes

Testing Outcomes

GOOD

PENETRATION FORCE COMPARISON

MEDTRONIC, MV-175-8, 8 mil

ETHICON, BV175-6, 8 mil

GOOD

Sta

nd

ard

De

via

tio

n

Pass

1 10 20 30

PENETRATION FORCE STANDARD DEVIATION

Ethicon Needle Alloy Code

ETHALLOY BV175-6

Competitor Code

Medtronic MV-175-8

Less penetration force required

Less variability in penetration force

✓✓

2.3

4.9

8.57.6 7.4

4.05.1

5.9

Page 8: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

*Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape, and Stiffness characteristics.

Reference: Technical Memo CT16-004. BV175-6 Competitive Assessment. August 23, 2016. Ethicon, Inc. 8

STRENGTH COMPARISON

• BV175-6 EVERPOINT (EVP)

showed less variability in

penetration force than

DR8f from 1 to 30 passes

• BV175-6 EVERPOINT (EVP)

displayed greater strength

than DR8f regardless of

the angle at which force

was applied

PENETRATION FORCE STANDARD DEVIATION

Be

nd

Mo

me

nt

(gra

m •

cm

)

Angle (degrees)

B. BRAUN, DR8f, 8 mil

ETHICON, BV175-6 (EVP), 8 mil

70

60

50

40

30

20

10

00 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90

GOOD

B. BRAUN, DR8f, 8 mil

ETHICON, BV175-6 (EVP), 8 milGOOD

Sta

nd

ard

De

via

tio

n

Pass1 10 20 30

2.5

7.3

9.0 9.4 9.3

2.3 2.3 2.2

PENETRATION FORCE COMPARISON• BV175-6 EVERPOINT (EVP)

required less penetration

force than DR8f from 1 to

30 passes

Testing Outcomes

Competitor Code

B. Braun DR8f

Pe

ne

tra

tio

n F

orc

e (

gra

ms)

Pass

80

70

60

50

40

30

20

10

0

B. BRAUN, DR8f, 8 mil

ETHICON, BV175-6 (EVP), 8 mil

1 10 20 30

GOOD

Less penetration force required

Less variability in penetration force

Greater strength regardless of angle*

✓✓✓

Ethicon Surgical Needles

Competitive Performance TestingEthicon Needle Alloy Code

EVERPOINT (EVP) BV175-6

Page 9: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Reference: Technical Memo CT16-004. BV175-6 Competitive Assessment. August 23, 2016. Ethicon, Inc. 9

• BV175-6 showed less

variability in penetration

force than DR8f from 1 to

30 passes

PENETRATION FORCE STANDARD DEVIATION

B. BRAUN, DR8f, 8 mil

ETHICON, BV175-6, 8 mil

GOOD

Sta

nd

ard

De

via

tio

n

Pass1 10 20 30

2.3

7.3

9.0 9.4 9.3

4.05.1

5.9

PENETRATION FORCE COMPARISON• BV175-6 required less

penetration force than

DR8f from 1 to 30 passes

Testing Outcomes

Competitor Code

B. Braun DR8f

Pe

ne

tra

tio

n F

orc

e (

gra

ms)

Pass

80

70

60

50

40

30

20

10

0

B. BRAUN, DR8f, 8 mil

ETHICON, BV175-6, 8 mil

1 10 20 30

GOOD

Less penetration force required

Less variability in penetration force

✓✓

Ethicon Surgical Needles

Competitive Performance TestingEthicon Needle Alloy Code

ETHALLOY BV175-6

Page 10: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

10

STRENGTH COMPARISON

Be

nd

Mo

me

nt

(gra

m •

cm

)

Angle (degrees)

70

60

50

40

30

20

10

00 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90

GOOD

*Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape, and Stiffness characteristics.

Reference: Technical Memo CT16-002. BV175-8 Competitive Assessment. July 19, 2016. Ethicon, Inc.

PENETRATION FORCE COMPARISON

Pe

ne

tra

tio

n F

orc

e (

gra

ms)

Pass

100

90

80

70

60

50

40

30

20

MEDTRONIC, MV-175-9, 8 mil

ETHICON, BV175-8 (EVP), 8 mil

1 10 20 30

• BV175-8 EVERPOINT (EVP)

required less penetration

force than MV-175-9 from

1 to 30 passes

• BV175-8 EVERPOINT (EVP)

showed less variability in

penetration force than

MV-175-9 from 1 to 30

passes

• BV175-8 EVERPOINT (EVP)

displayed greater strength

than MV-175-9 regardless

of the angle at which force

was applied

Testing Outcomes

GOOD

MEDTRONIC, MV-175-9, 8 mil

ETHICON, BV175-8 (EVP), 8 mil

MEDTRONIC, MV-175-9, 8 mil

ETHICON, BV175-8 (EVP), 8 mil

GOOD

Sta

nd

ard

De

via

tio

n

Pass1 10 20 30

PENETRATION FORCE STANDARD DEVIATION

Ethicon Needle Alloy Code

EVERPOINT (EVP) BV175-8

Competitor Code

Medtronic MV-175-9

Less penetration force required

Less variability in penetration force

Greater strength regardless of angle*

✓✓✓

Ethicon Surgical Needles

Competitive Performance Testing

2.6

5.0

8.2

6.8 6.5

2.5 2.6 2.7

Page 11: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

11

• C-1 displayed greater

(35-90º) or statistically

equivalent (0-30º)

strength compared

to CV-11

Testing Outcomes

STRENGTH COMPARISON

Be

nd

Mo

me

nt

(gra

m •

cm

)

Angle (degrees)

250

200

150

100

50

00 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90

GOOD

MEDTRONIC, CV-11

ETHICON, C-1

Ethicon Needle Alloy Code

ETHALLOY C-1

Competitor Code

Medtronic CV-11

Greater or statistically equivalent strength*✓

Ethicon Surgical Needles

Competitive Performance Testing

*Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape, and Stiffness characteristics.

Reference: Technical Memo CT17-006. C-1 Competitive Assessment. May 20, 2017. Ethicon, Inc.

Page 12: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

12

PENETRATION FORCE COMPARISON

Pe

ne

tra

tio

n F

orc

e (

gra

ms)

Pass

110

105

100

95

90

85

80

75

70

65

60

55

50

45

40

B. BRAUN, DR12

ETHICON, C-1

1 10 20 30

• C-1 required less

penetration force

than DR12 from

1 to 30 passes

• C-1 displayed greater

strength than DR12

regardless of the angle at

which force was applied

Testing Outcomes

GOOD

Ethicon Needle Alloy Code

ETHALLOY C-1

Competitor Code

B. Braun DR12

Less penetration force required✓Greater strength regardless of angle*✓

Ethicon Surgical Needles

Competitive Performance Testing

*Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape, and Stiffness characteristics.

Reference: Technical Memo CT17-006. C-1 Competitive Assessment. May 20, 2017. Ethicon, Inc.

STRENGTH COMPARISON

Be

nd

Mo

me

nt

(gra

m •

cm

)

Angle (degrees)

250

200

150

100

50

00 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90

GOOD

B. BRAUN, DR12

ETHICON, C-1

Page 13: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Surgical Needles

Competitive Performance Testing

*Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape, and Stiffness characteristics.

Reference: Technical Memo CT16-005. CT Competitive Assessment. October 17, 2016. Ethicon, Inc. 13

STRENGTH COMPARISON

PENETRATION FORCE COMPARISON• CT required less

penetration force

than GS-24 from

1 to 30 passes

• CT showed less variability

in penetration force than

GS-24 from 1 to 30 passes

• CT displayed greater

strength than GS-24

regardless of the angle at

which force was applied

Testing Outcomes

Competitor Code

Medtronic GS-24

PENETRATION FORCE STANDARD DEVIATION

Pe

ne

tra

tio

n F

orc

e (

gra

ms)

Pass

250

200

150

100

50

0

ETHICON, CT

MEDTRONIC, GS-24

1 10 20 30

GOOD

Be

nd

Mo

me

nt

(gra

m •

cm

)

Angle (degrees)

MEDTRONIC, GS-24

ETHICON, CT

6000

5000

4000

3000

2000

1000

00 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90

GOOD

MEDTRONIC, GS-24

ETHICON, CT

GOOD

Sta

nd

ard

De

via

tio

n

Pass1 10 20 30

Less penetration force required

Less variability in penetration force

Greater strength regardless of angle*

✓✓✓

8.9

23.7

32.7 34.1 34.8

16.019.0

22.0

6.5

Ethicon Needle Alloy Code

Stainless Steel CT

Page 14: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Surgical Needles

Competitive Performance Testing

*Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape, and Stiffness characteristics.

Reference: Technical Memo CT16-005. CT Competitive Assessment. October 17, 2016. Ethicon, Inc. 14

STRENGTH COMPARISON

PENETRATION FORCE COMPARISON• CT required less

penetration force

than HR40s from

1 to 30 passes

• CT showed less variability

in penetration force than

HR40s from 1 to 30 passes

• CT displayed greater

strength than HR40s

regardless of the angle at

which force was applied

Testing Outcomes

Competitor Code

B. Braun HR40s

PENETRATION FORCE STANDARD DEVIATION

Pe

ne

tra

tio

n F

orc

e (

gra

ms)

Pass

250

200

150

100

50

0

ETHICON, CT

B.BRAUN, HR40s

1 10 20 30

GOOD

Be

nd

Mo

me

nt

(gra

m •

cm

)

Angle (degrees)

B.BRAUN, HR40s

ETHICON, CT

6000

5000

4000

3000

2000

1000

00 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90

GOOD

B.BRAUN, HR40s

ETHICON, CTGOOD

Sta

nd

ard

De

via

tio

n

Pass1 10 20 30

Less penetration force required

Less variability in penetration force

Greater strength regardless of angle*

✓✓✓

8.910.5

19.020.1

22.3

16.019.0

22.0

Ethicon Needle Alloy Code

Stainless Steel CT

Page 15: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Surgical Needles

Competitive Performance Testing

*Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape, and Stiffness characteristics.

Reference: Technical Memo CT16-006. CT-1 Competitive Assessment. December 13, 2016. Ethicon, Inc. 15

STRENGTH COMPARISON

PENETRATION FORCE COMPARISON• CT-1 required less

penetration force

than GS-21 from

1 to 30 passes

• CT-1 showed less variability

in penetration force than

GS-21 from 1 to 30 passes

• CT-1 displayed greater

strength than GS-21

regardless of the angle at

which force was applied

Testing Outcomes

Competitor Code

Medtronic GS-21

PENETRATION FORCE STANDARD DEVIATION

Pe

ne

tra

tio

n F

orc

e (

gra

ms)

Pass

300

250

200

150

100

50

0

MEDTRONIC, GS-21

ETHICON, CT-1

1 10 20 30

GOOD

Be

nd

Mo

me

nt

(gra

m •

cm

)

Angle (degrees)

MEDTRONIC, GS-21

ETHICON, CT-13500

3000

2500

2000

1500

1000

500

00 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90

GOOD

MEDTRONIC, GS-21

ETHICON, CT-1

GOOD

Sta

nd

ard

De

via

tio

n

Pass1 10 20 30

Less penetration force required

Less variability in penetration force

Greater strength regardless of angle*

✓✓✓

11.3

27.230.0

28.7 28.4

17.9 20.924.8

Ethicon Needle Alloy Code

Stainless Steel CT-1

Page 16: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Surgical Needles

Competitive Performance Testing

Reference: Technical Memo CT16-006. CT-1 Competitive Assessment. December 13, 2016. Ethicon, Inc. 16

PENETRATION FORCE COMPARISON• CT-1 required less

penetration force

than HR37s from

1 to 30 passes

Testing Outcomes

Competitor Code

B. Braun HR37s

Pe

ne

tra

tio

n F

orc

e (

gra

ms)

Pass

300

250

200

150

100

50

0

B. BRAUN, HR37s

ETHICON, CT-1

1 10 20 30

GOOD

Less penetration force required✓

Ethicon Needle Alloy Code

Stainless Steel CT-1

Page 17: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Surgical Needles

Competitive Performance Testing

*Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape, and Stiffness characteristics.

Reference: Technical Memo CT17-005. RB-1 Competitive Assessment. April 20, 2017. Ethicon, Inc. 17

STRENGTH COMPARISON• RB-1 displayed greater

strength than CV-23

regardless of the angle at

which force was applied

Testing Outcomes

Competitor Code

Medtronic CV-23

Be

nd

Mo

me

nt

(gra

m •

cm

)

Angle (degrees)

MEDTRONIC, CV-23

ETHICON, RB-1300

250

200

150

100

50

0

0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90

GOOD

Greater strength regardless of angle*✓

Ethicon Needle Alloy Code

ETHALLOY RB-1

Page 18: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Surgical Needles

Competitive Performance Testing

18

STRENGTH COMPARISON

PENETRATION FORCE COMPARISON• SH required less

penetration force

than V-20 from

1 to 30 passes

• SH displayed greater

strength than V-20

regardless of the angle

at which force was applied

Testing Outcomes

Competitor Code

Medtronic V-20

Pe

ne

tra

tio

n F

orc

e (

gra

ms)

Pass

160

140

120

100

80

60

40

20

0

MEDTRONIC, V-20

ETHICON, SH

1 10 20 30

GOOD

Be

nd

Mo

me

nt

(gra

m •

cm

)

Angle (degrees)

MEDTRONIC, V-20

ETHICON, SH

1200

1000

800

600

400

200

0

0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90

GOOD

Less penetration force required

Greater strength regardless of angle*

✓✓

*Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape, and Stiffness characteristics.

Reference: Technical Memo CT17-003. SH Needle Competitive Assessment. March 20, 2017. Ethicon, Inc.

Ethicon Needle Alloy Code

Stainless Steel SH

Page 19: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Surgical Needles

Competitive Performance Testing

Reference: Technical Memo CT17-009. SH Competitive Assessment. September 26, 2017. Ethicon, Inc. 19

• SH showed less variability

in penetration force than

HR26 from 1 to 30 passes

Competitor Code

B. Braun HR26

Less penetration force required

Less variability in penetration force

✓✓

Ethicon Needle Alloy Code

Stainless Steel SH

• SH required less

penetration force than

HR26 from 1 to 10 passes

Testing OutcomesPENETRATION FORCE COMPARISON

Pe

ne

tra

tio

n F

orc

e (

gra

ms)

Pass

145

140

135

130

125

120

115

110

105

100

95

90

85

80

75

B. BRAUN, HR26

ETHICON, SH

1 10 20 30

GOOD

PENETRATION FORCE STANDARD DEVIATION

B. BRAUN, HR26

ETHICON, SHGOOD

Sta

nd

ard

De

via

tio

n

Pass1 10 20 30

5.4

7.6

11.110.6

10.210.4 9.89.5

Page 20: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Surgical Needles

Competitive Performance Testing

Reference: Technical Memo CT18-003. V-34 Competitive Assessment. April 3, 2018. Ethicon, Inc.

Reference: Technical Memo CT18-001. V-7 Competitive Assessment. February 15, 2018. Ethicon, Inc.

20

Competitor Code

Medtronic KV-7

Less penetration force required✓

Ethicon Needle Alloy Code

Stainless Steel V-7

• V-7 required less

penetration force than

KV-7 from 1 to 30 passes

Testing Outcomes

PENETRATION FORCE COMPARISONP

en

etr

ati

on

Fo

rce

(g

ram

s)

Pass

150

145

140

135

130

125

120

115

110

105

100

95

90

85

80

75

MEDTRONIC, KV-7

ETHICON, V-7

1 10 20 30

GOOD

Less penetration force required✓

• V-34 required less penetration

force than Meril Endo-Surgery 40

mm from 1 to 30 passes

Testing Outcomes

Competitor Code

Meril Endo-Surgery 40 mm

Ethicon Needle Alloy Code

Stainless Steel V-34

PENETRATION FORCE COMPARISON

Pe

ne

tra

tio

n F

orc

e (

gra

ms)

Pass

380

360

340

320

300

280

260

240

220

200

MERIL, ENDO-SURGERY 40 MM

ETHICON, V-34

1 10 20 30

GOOD

Page 21: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Surgical Needles | STRATAFIX

Competitive Performance Testing

*Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape, and Stiffness characteristics.

Reference: Technical Memo CT12-009. December 12, 2012. Ethicon, Inc.

• STRATAFIX CT-1 displayed

greater strength than

V-Loc™ GS-21 regardless of

the angle at which force

was applied

Testing Outcomes

Competitor Code

Medtronic V-Loc™ GS-21

Greater strength regardless of angle*✓

STRENGTH COMPARISON

To

rqu

e (g

ram

• c

m)

Angle (degrees)

MEDTRONIC, V-Loc™ GS-21, 39 mil

ETHICON, STRATAFIX CT-1, 40 mil

4000

3500

3000

2500

2000

1500

1000

500

00 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90

GOOD

Ethicon Needle Alloy Code

Stainless Steel CT-1

Page 22: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Surgical Needles | STRATAFIX

Competitive Performance Testing

Reference: Technical Memo CT12-009. December 12, 2012. Ethicon, Inc.

Competitor Code

Medtronic V-Loc™ GS-25

Less penetration force required

Less variability in penetration force

✓✓

PENETRATION FORCE COMPARISON• STRATAFIX CTX required

less penetration force than

V-Loc™ GS-25 from 1 to

30 passes

• STRATAFIX CTX

showed less variability

in penetration force

than V-Loc™ GS-25 from

1 to 30 passes

Testing Outcomes

PENETRATION FORCE STANDARD DEVIATION

Pe

ne

tra

tio

n F

orc

e (

gra

ms)

Pass

ETHICON, STRATAFIX CTX, 50 mil

MEDTRONIC, V-Loc™ GS-25, 50 mil

1 10 20 30

GOOD

MEDTRONIC, V-Loc™ GS-25, 50 mil

ETHICON, STRATAFIX CTX, 50 milGOOD

Sta

nd

ard

De

via

tio

n

Pass1 10 20 30

12.0

27.7

44.349.1

55.0

18.522.2 24.1

350

300

250

200

150

100

50

0

Ethicon Needle Alloy Code

Stainless Steel CTX

Page 23: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Surgical Needles | STRATAFIX

Competitive Performance Testing

Reference: Technical Memo CT12-009. December 12, 2012. Ethicon, Inc.

Competitor Code

Quill DE-14

Less penetration force required

Less variability in penetration force

✓✓

PENETRATION FORCE COMPARISON• STRATAFIX FS required

less penetration force than

Quill DE-14 from 1 to 30

passes

• STRATAFIX FS showed less

variability in penetration

force than Quill DE-14 from

1 to 30 passes

Testing Outcomes

PENETRATION FORCE STANDARD DEVIATION

Pe

ne

tra

tio

n F

orc

e (

gra

ms)

Pass

ETHICON, STRATAFIX FS, 29 mil

QUILL, DE-14, 28 mil

1 10 20 30

GOOD

QUILL, DE-14, 28 mil

ETHICON, STRATAFIX FS, 29 mil

GOOD

Sta

nd

ard

De

via

tio

n

Pass

1 10 20 30

10.0

20.6 21.5 21.7 22.5

10.9 10.4 11.1

250

200

150

100

50

0

Ethicon Needle Alloy Code

Stainless Steel FS

Page 24: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Surgical Needles | STRATAFIX

Competitive Performance Testing

*Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape, and Stiffness characteristics.

Reference: Technical Memo CT09-022. January 20, 2010. Ethicon, Inc.

STRENGTH COMPARISON

PENETRATION FORCE COMPARISON• STRATAFIX FS-2

required less penetration

force than Quill DE-12

from 1 to 30 passes

• STRATAFIX FS-2

showed less variability

in penetration force

than Quill DE-12 from

1 to 30 passes

• STRATAFIX FS-2

displayed greater strength

than Quill DE-12 regardless

of the angle at which force

was applied

Testing Outcomes

Competitor Code

Quill DE-12

PENETRATION FORCE STANDARD DEVIATION

Pe

ne

tra

tio

n F

orc

e (

gra

ms)

Pass

160

140

120

100

80

60

40

20

0

QUILL, DE-12, 22 mil

ETHICON, STRATAFIX FS-2, 24 mil

1 10 20 30

GOOD

To

rqu

e (g

ram

• c

m)

Angle (degrees)

QUILL, DE-12, 22 mil

ETHICON, STRATAFIX FS-2, 24 mil

1200

1000

800

600

400

200

00 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90

GOOD

QUILL, DE-12, 22 mil

ETHICON, STRATAFIX FS-2, 24 mil

GOOD

Sta

nd

ard

De

via

tio

n

Pass1 10 20 30

Less penetration force required

Less variability in penetration force

Greater strength regardless of angle*

✓✓✓

8.0

13.5

11.0 11.4 10.7

5.4

7.68.3

Ethicon Needle Alloy Code

Stainless Steel FS-2

Page 25: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Needle Glossary

Surgical Yield

Needles will flex under pressure or load. When the pressure

is released, the needle will ‘spring back’ to its original shape.

When the pressure on the needle becomes too great, the

needle will fail to return to its original shape. This is known

as Surgical Yield. Needle alloy, wire diameter, needle-body

geometry, and heat treatment during manufacturing are

all factors that can affect a needle's ability to resist pressure.

In more scientific terms, Surgical Yield may be defined

as the amount of Moment required to initiate plastic

deformation during bend tests. (ASTM standard F-1840-98a)

Refer to Graph.

Ultimate Moment

In lab tests, a curved needle is subjected to an increasing

pressure until the needle is bent 90º. The amount of

pressure the needle can resist until it bends 90º is referred

to as the Ultimate Moment. In more scientific terms,

Ultimate Moment is defined as the maximum Moment

applied during bend tests. (ASTM standard F-1840-98a)

Refer to Graph.

Width

I-BEAM SQUARE BODY

Height

NEEDLE BODY GEOMETRY

CROSS-SECTIONS

Be

nd

Mo

me

nt

(ne

wto

ns

• cm

)

Angular Deflection (degrees)

ElasticRegion

Surgical Yield

UltimateMoment

0 10 20 30 40 50 60 70 80 90

Graph

REPRESENTATION OF

NEEDLE STRENGTH PROFILES

I-Beam Body

This refers to the cross-section of the needle body.

In general terms, the height of the body is greater than the

width of the body, as shown in the illustration. If the corners

are rounded, then the body is referred to as a round corner

I-beam.

Square Body

This refers to the cross-section of the needle body.

In general terms, the height of the body is equal to the

width of the body, as shown in the illustration.

Ductility

Ductility is the ability of a needle to resist breaking while being bent under pressure. If a needle lacks ductility, it may break

when pressure is exerted upon it. If a needle is ductile, the needle will bend under pressure before it will break. Needle

designs aim to make a needle as strong as possible to resist bending, while also imparting ductility so it can bend when the

force applied becomes too strong. A test known as a Reshape Test determines the ductility of a needle. In this test, a needle

is bent 90º and then bent back to its original shape (or as close as the needle will allow). The more times a needle can be

bent in this manner, the greater the needle’s ability to resist breaking. Refer to Graph.

7

6

5

4

3

2

1

0

Page 26: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Sutures

Competitive Performance Testing

Reference: Technical Report PE-2017-0012. July 14, 2017. Ethicon, Inc.26

• 2-0 ETHIBOND EXCEL Suture displayed greater straight

tensile strength than 2-0 Ti-Cron™ Suture

• 2-0 ETHIBOND EXCEL Suture displayed less package

memory than 2-0 Ti-Cron™ Suture

Testing Outcomes

Competitor

Medtronic 2-0 Ti-Cron™ Suture

Greater straight tensile strength✓

Less package memory✓Greater controlled linear elongation✓Greater knot security✓

ETHICON, 2-0 ETHIBOND

MEDTRONIC, 2-0 Ti-Cron

14.0013.32

ETHICON, 2-0 ETHIBOND

MEDTRONIC, 2-0 Ti-Cron 6.8710.20

Straight Tensile Strength (lbs)

Package Memory - Short Pack (%)

Controlled Linear Elongation (%)

• 2-0 ETHIBOND EXCEL Suture displayed greater

controlled linear elongation than 2-0 Ti-Cron™ Suture

ETHICON, 2-0 ETHIBOND

MEDTRONIC, 2-0 Ti-Cron

13.9412.79

Knot Security Occurrences (%)

• 2-0 ETHIBOND EXCEL Suture displayed greater knot

security compared to 2-0 Ti-Cron™ SutureETHICON, 2-0 ETHIBOND

MEDTRONIC, 2-0 Ti-Cron

80

72

Ethicon Suture

2-0 ETHIBOND EXCEL® Polyester Suture

Page 27: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Sutures

Competitive Performance Testing

Reference: Technical Report PE-2017-0012. July 14, 2017. Ethicon, Inc.27

• 2-0 ETHIBOND EXCEL Suture displayed greater straight

tensile strength than 2-0 PremiCron® Suture

Testing Outcomes

Ethicon Suture

2-0 ETHIBOND EXCEL® Polyester Suture

Competitor

B. Braun 2-0 PremiCron® Suture

Greater straight tensile strength✓

Greater intrinsic straight tensile strength✓

Less package memory✓

Greater controlled linear elongation✓

More reliable knot slide✓

ETHICON, 2-0 ETHIBOND

B. BRAUN, 2-0 Premicron

14.0012.40

Straight Tensile Strength (lbs)

• 2-0 ETHIBOND EXCEL Suture displayed less package

memory than 2-0 PremiCron® Suture

ETHICON, 2-0 ETHIBOND

B. BRAUN, 2-0 Premicron

6.87

13.63

Package Memory - Short Pack (%)

• 2-0 ETHIBOND EXCEL Suture displayed greater intrinsic

straight tensile strength than 2-0 PremiCron® Suture

ETHICON, 2-0 ETHIBOND

B. BRAUN, 2-0 Premicron

103.6

94.7

Intrinsic Straight Tensile Strength (kpsi)

Controlled Linear Elongation (%)

• 2-0 ETHIBOND EXCEL Suture displayed greater

controlled linear elongation than 2-0 PremiCron® Suture

ETHICON, 2-0 ETHIBOND

B. BRAUN, 2-0 Premicron

13.9412.89

Knot Slide Occurrences (%)

• 2-0 ETHIBOND EXCEL Suture displayed more reliable

knot slide compared to 2-0 PremiCron® SutureETHICON, 2-0 ETHIBOND

B. BRAUN, 2-0 Premicron

100

67

Page 28: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Reference: Technical Report PE-2017-0023. October 2, 2017. Ethicon, Inc. 28

Ethicon Sutures

Competitive Performance TestingEthicon Suture

4-0 ETHIBOND EXCEL® Polyester Suture

Competitor

Medtronic 4-0 Ti-Cron™ Suture

Less package memory✓

• 4-0 ETHIBOND EXCEL Suture

displayed less package memory

than 4-0 Ti-Cron™ Suture

Testing Outcomes

5.307.43

Package Memory – Short Pack (%)

ETHICON, 4-0 ETHIBOND

MEDTRONIC, 4-0 Ti-Cron

Page 29: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Reference: Technical Report PE-2017-0023. October 2, 2017. Ethicon, Inc. 29

Ethicon Sutures

Competitive Performance Testing

119.91 115.86

Intrinsic Straight Tensile Strength (kpsi)

Ethicon Suture

4-0 ETHIBOND EXCEL® Polyester Suture

Competitor

B. Braun 4-0 PremiCron® Suture

Greater intrinsic straight tensile strength✓Less package memory✓Greater controlled linear elongation✓More reliable knot slide✓

• 4-0 ETHIBOND EXCEL Suture

displayed greater intrinsic

straight tensile strength than

4-0 PremiCron® Suture

Testing Outcomes ETHICON, 4-0 ETHIBOND

B. BRAUN, 4-0 Premicron

5.30

10.57

10.96 10.09

Package Memory – Short Pack (%)

Controlled Linear Elongation (%)

• 4-0 ETHIBOND EXCEL Suture

displayed less package memory

than 4-0 PremiCron® Suture

• 4-0 ETHIBOND EXCEL Suture

displayed greater controlled

linear elongation than 4-0

PremiCron® Suture

ETHICON, 4-0 ETHIBOND

B. BRAUN, 4-0 Premicron

ETHICON, 4-0 ETHIBOND

B. BRAUN, 4-0 Premicron

10087

Knot Slide Occurrences (%)

• 4-0 ETHIBOND EXCEL Suture

displayed more reliable knot

slide compared to 4-0

PremiCron® Suture

ETHICON, 4-0 ETHIBOND

B. BRAUN, 4-0 Premicron

Page 30: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

4.763.00

136.46

87.55

• 5-0 MONOCRYL Suture displayed

greater straight tensile strength

than 5-0 Caprosyn™ Suture

• 5-0 MONOCRYL Suture displayed

greater intrinsic straight tensile

strength than 5-0 Caprosyn™ Suture

Testing Outcomes

Greater intrinsic straight tensile strength✓Greater straight tensile strength✓

Greater knot tensile strength✓

ETHICON, 5-0 MONOCRYL

MEDTRONIC, 5-0 Caprosyn

ETHICON, 5-0 MONOCRYL

MEDTRONIC, 5-0 Caprosyn

Straight Tensile Strength (lbs)

Intrinsic Straight Tensile Strength (kpsi)

Reference: Technical Report PE-2017-0031. October 2, 2017. Ethicon, Inc.

Ethicon Sutures

Competitive Performance Testing

30

Ethicon Suture

5-0 MONOCRYL® (poliglecaprone 25) Suture

Competitor

Medtronic 5-0 Caprosyn™ Suture

Less package memory✓

More reliable knot slide✓

11.90

51.37

2.25 1.80

• 5-0 MONOCRYL Suture displayed

greater knot tensile strength than

5-0 Caprosyn™ Suture

ETHICON, 5-0 MONOCRYL

MEDTRONIC, 5-0 Caprosyn

Knot Tensile Strength (lbs)

• 5-0 MONOCRYL Suture

displayed less package memory

than 5-0 Caprosyn™ Suture

ETHICON, 5-0 MONOCRYL

MEDTRONIC, 5-0 Caprosyn

Package Memory - Short Pack (%)

100

83• 5-0 MONOCRYL Suture

displayed more reliable

knot slide compared to

5-0 Caprosyn™ Suture

ETHICON, 5-0 MONOCRYL

MEDTRONIC, 5-0 Caprosyn

Knot Slide Occurrences (%)

Page 31: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

4.763.71

• 5-0 MONOCRYL Suture displayed

greater straight tensile strength

than 5-0 Monosyn® Suture

• 5-0 MONOCRYL Suture

displayed greater intrinsic

straight tensile strength than

5-0 Monosyn® Suture

Testing Outcomes

Greater straight tensile strength✓Greater intrinsic straight tensile strength✓Less package memory ✓

ETHICON, 5-0 MONOCRYL

B. BRAUN, 5-0 Monosyn

ETHICON, 5-0 MONOCRYL

B. BRAUN, 5-0 Monosyn

136.46

91.80

Straight Tensile Strength (lbs)

Intrinsic Straight Tensile Strength (kpsi)

Reference: Technical Report PE-2017-0031. October 2, 2017. Ethicon, Inc. 31

Ethicon Sutures

Competitive Performance TestingEthicon Suture

5-0 MONOCRYL® (poliglecaprone 25) Suture

Competitor

B. Braun 5-0 Monosyn® Suture

• 5-0 MONOCRYL Suture

displayed less package memory

than 5-0 Monosyn® Suture

ETHICON, 5-0 MONOCRYL

B. BRAUN, 5-0 Monosyn

11.90

39.07

Package Memory – Short Pack (%)

Page 32: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Sutures

Competitive Performance Testing

Reference: Technical Report PE-2016-0020. July 21, 2017. Ethicon, Inc. 32

• 5-0 PROLENE Suture displayed less package memory

than 5-0 Optilene™ Suture

• 5-0 PROLENE Suture displayed greater controlled linear

elongation than 5-0 Optilene™ Suture

Testing Outcomes

Ethicon Suture

5-0 PROLENE® Polypropylene Suture

Competitor

B. Braun 5-0 Optilene™ Suture (Non-Absorbable)

Less package memory✓

Greater controlled linear elongation✓

ETHICON, 5-0 PROLENE

B. BRAUN, 5-0 Optilene

40.3

47.0

ETHICON, 5-0 PROLENE

B. BRAUN, 5-0 Optilene

35.231.6

Package Memory - Short Pack (%)

Controlled Linear Elongation (%)

• 5-0 PROLENE Suture displayed greater controlled linear

elongation than 5-0 Surgipro™ II Suture

Testing Outcomes

Ethicon Suture

5-0 PROLENE® Polypropylene Suture

Competitor

Medtronic 5-0 Surgipro™ II Polypropylene Monofilament Suture

Greater controlled linear elongation✓

ETHICON, 5-0 PROLENE

MEDTRONIC 5-0 Surgipro II

35.230.7

Package Memory - Short Pack (%)

Page 33: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Sutures

Competitive Performance Testing

Reference: Technical Report PE-2016-0020. July 21, 2017. Ethicon, Inc. 33

• 5-0 PROLENE Suture displayed greater straight tensile

strength than 5-0 Premilene® Suture

Testing Outcomes

Ethicon Suture

5-0 PROLENE® Polypropylene Suture

Competitor

B. Braun 5-0 Premilene® Suture (Non-Absorbable)

ETHICON, 5-0 PROLENE

B. BRAUN, 5-0 Premilene

2.21 1.86

Straight Tensile Strength (lbs)

Greater straight tensile strength✓

Greater knot tensile strength✓Greater intrinsic straight tensile strength✓

Less package memory✓

• 5-0 PROLENE Suture displayed greater knot tensile

strength than 5-0 Premilene® SutureETHICON, 5-0 PROLENE

B. BRAUN, 5-0 Premilene 1.75 1.53Knot Tissue Strength (lbs)

• 5-0 PROLENE Suture displayed greater intrinsic straight

tensile strength than 5-0 Premilene® Suture ETHICON, 5-0 PROLENE

B. BRAUN, 5-0 Premilene

84.672.7

Intrinsic Straight Tensile Strength (kpsi)

Package Memory - Short Pack (%)

• 5-0 PROLENE Suture displayed less package memory

than 5-0 Premilene® Suture

ETHICON, 5-0 PROLENE

B. BRAUN, 5-0 Premilene 40.3 43.4

Page 34: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Sutures

Competitive Performance Testing

Reference: Technical Report PE-2016-0020. October 19, 2016. Ethicon, Inc.34

• 6-0 PROLENE Suture displayed less package memory

than 6-0 Optilene™ Non-Absorbable Suture

• 6-0 PROLENE Suture displayed greater controlled linear

elongation than 6-0 Optilene™ Non-Absorbable Suture

Testing Outcomes

Ethicon Suture

6-0 PROLENE® Polypropylene Suture

Competitor

B. Braun 6-0 Optilene™ Non-Absorbable Suture

ETHICON, 6-0 PROLENE

B. BRAUN, 6-0 Optilene25.2

37.6

Package Memory – Short Pack (%)

ETHICON, 6-0 PROLENE

B. BRAUN, 6-0 Optilene

36.2

28.9

Controlled Linear Elongation (%)

Less package memory

Greater controlled linear elongation

✓✓

• 6-0 PROLENE Suture displayed greater controlled linear

elongation than 6-0 Surgipro™ II Suture

Testing Outcomes

Ethicon Suture

6-0 PROLENE® Polypropylene Suture

Competitor

Medtronic 6-0 Surgipro™ II Polypropylene Monofilament Suture

Greater controlled linear elongation✓

ETHICON, 6-0 PROLENE

MEDTRONIC, 6-0 Surgipro II

36.231.8

Controlled Linear Elongation (%)

Page 35: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Sutures

Competitive Performance Testing

Reference: Technical Report PE-2016-0020. October 19, 2016. Ethicon, Inc. 35

Ethicon Suture

6-0 PROLENE® Polypropylene Suture

Competitor

B. Braun 6-0 Premilene® Suture (Non-Absorbable)

• 6-0 PROLENE Suture displayed greater straight tensile

strength than 6-0 Premilene® Suture

Testing OutcomesETHICON, 6-0 PROLENE

B. BRAUN, 6-0 Premilene

1.060.79

• 6-0 PROLENE Suture displayed greater knot tensile

strength than 6-0 Premilene® Suture

ETHICON, 6-0 PROLENE

B. BRAUN, 6-0 Premilene

0.86 0.78

• 6-0 PROLENE Suture displayed less package memory

than 6-0 Premilene® Suture

Greater straight tensile strength✓

Greater knot tensile strength✓Greater intrinsic straight tensile strength✓

Greater controlled linear elongation✓Less package memory ✓

Straight Tensile Strength (lbs)

Knot Tensile Strength (lbs)

• 6-0 PROLENE Suture displayed greater intrinsic straight

tensile strength than 6-0 Premilene® Suture

ETHICON, 6-0 PROLENE

B. BRAUN, 6-0 Premilene

86.270.7

Intrinsic Straight Tensile Strength (kpsi)

ETHICON, 6-0 PROLENE

B. BRAUN, 6-0 Premilene25.234.1

Package Memory – Short Pack (%)

• 6-0 PROLENE Suture displayed greater controlled linear

elongation than 6-0 Premilene® SutureETHICON, 6-0 PROLENE

B. BRAUN, 6-0 Premilene

36.2

26.7

Controlled Linear Elongation (%)

Page 36: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Sutures

Competitive Performance Testing

Reference: Technical Report PE-2016-0020. November 8, 2016. Ethicon, Inc.36

• 7-0 PROLENE Suture displayed greater straight tensile

strength than 7-0 Premilene® Suture

• 7-0 PROLENE Suture displayed greater knot tensile

strength than 7-0 Premilene® Suture

Testing Outcomes

Ethicon Suture

7-0 PROLENE® Polypropylene Suture

Competitor

B. Braun 7-0 Premilene® Suture (Non-Absorbable)

Greater straight tensile strength✓

Greater knot tensile strength✓Greater intrinsic straight tensile strength✓

Less package memory✓Greater controlled linear elongation✓

ETHICON, 7-0 PROLENE

B. BRAUN, 7-0 Premilene

0.620.46

ETHICON, 7-0 PROLENE

B. BRAUN, 7-0 Premilene0.50 0.44

• 7-0 PROLENE Suture displayed less package memory

than 7-0 Premilene® Suture

ETHICON, 7-0 PROLENE

B. BRAUN, 7-0 Premilene7.5

9.7

• 7-0 PROLENE Suture displayed greater controlled linear

elongation than 7-0 Premilene® Suture

Straight Tensile Strength (lbs)

Knot Tensile Strength (lbs)

Package Memory - Short Pack (%)

• 7-0 PROLENE Suture displayed greater intrinsic straight

tensile strength than 7-0 Premilene® Suture

ETHICON, 7-0 PROLENE

B. BRAUN, 7-0 Premilene

99.881.6

Intrinsic Straight Tensile Strength (kpsi)

ETHICON, 7-0 PROLENE

B. BRAUN, 7-0 Premilene

29.325.2

Controlled Linear Elongation (%)

Page 37: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Sutures

Competitive Performance Testing

Reference: Technical Report PE-2016-0020. February 1, 2017. Ethicon, Inc.37

• 8-0 PROLENE Suture displayed greater straight tensile

strength than 8-0 Premilene® Suture

• 8-0 PROLENE Suture displayed greater intrinsic straight

tensile strength than 8-0 Premilene® Suture

Testing Outcomes

Ethicon Suture

8-0 PROLENE® Polypropylene Suture

Competitor

B. Braun 8-0 Premilene® Suture (Non-Absorbable)

ETHICON, 8-0 PROLENE

B. BRAUN, 8-0 Premilene

132.9123.6

ETHICON, 8-0 PROLENE

B. BRAUN, 8-0 Premilene

122.489.3

Straight Tensile Strength (gram-force)

Intrinsic Straight Tensile Strength (kpsi)

• 8-0 PROLENE Suture displayed greater knot tensile

strength than 8-0 Premilene® SutureETHICON, 8-0 PROLENE

B. BRAUN, 8-0 Premilene

107.7 101.6

Knot Tensile Strength (gram-force)

Greater straight tensile strength✓Greater intrinsic straight tensile strength✓Greater knot tensile strength✓

Page 38: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Sutures

Competitive Performance Testing

Reference: Technical Report PE-2016-0020. February 1, 2017. Ethicon, Inc.38

• 8-0 OS PROLENE Suture displayed greater straight tensile

strength than 8-0 Optilene™ Suture

• 8-0 OS PROLENE Suture displayed greater knot tensile

strength than 8-0 Optilene™ Suture

Testing Outcomes

Ethicon Suture

8-0 Oversized (OS) PROLENE® Polypropylene Suture

Competitor

B. Braun 8-0 Optilene™ Non-Absorbable Suture

ETHICON, 8-0 OS PROLENE

B. BRAUN, 8-0 Optilene

217.5161.1

ETHICON, 8-0 OS PROLENE

B. BRAUN, 8-0 Optilene

167.4112.9

Straight Tensile Strength (gram-force)

Knot Tensile Strength (gram-force)

• 8-0 OS PROLENE Suture displayed greater controlled

linear elongation than 8-0 Optilene™ SutureETHICON, 8-0 OS PROLENE

B. BRAUN, 8-0 Optilene

35.6824.42

Controlled Linear Elongation (%)

Greater straight tensile strength✓Greater knot tensile strength✓Greater controlled linear elongation✓

Page 39: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Sutures

Competitive Performance Testing

Reference: Technical Report PE-2016-0020. February 1, 2017. Ethicon, Inc.39

• 8-0 OS PROLENE Suture displayed greater straight tensile

strength than 8-0 Premilene® Suture

• 8-0 OS PROLENE Suture displayed greater intrinsic

straight tensile strength than 8-0 Premilene® Suture

Testing Outcomes

Ethicon Suture

8-0 Oversized (OS) PROLENE® Polypropylene Suture

Competitor

B. Braun 8-0 Premilene® Suture (Non-Absorbable)

Greater straight tensile strength✓

Greater intrinsic straight tensile strength✓

Greater knot tensile strength✓

ETHICON, 8-0 OS PROLENE

B. BRAUN, 8-0 Premilene

217.5

123.6

ETHICON, 8-0 OS PROLENE

B. BRAUN, 8-0 Premilene

109.289.3

Straight Tensile Strength (gram-force)

Intrinsic Straight Tensile Strength (kpsi)

• 8-0 OS PROLENE Suture displayed greater knot

strength than 8-0 Premilene® Suture

ETHICON, 8-0 OS PROLENE

B. BRAUN, 8-0 Premilene

167.4

101.6

Knot Tensile Strength (gram-force)

Page 40: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Sutures

Competitive Performance Testing

Reference: Technical Report PE-2016-0020. February 1, 2017. Ethicon, Inc.40

Greater straight tensile strength✓

Greater intrinsic straight tensile strength✓

Greater knot tensile strength✓

• 8-0 OS PROLENE Suture displayed greater straight tensile

strength than 8-0 Surgipro™ II Suture

• 8-0 OS PROLENE Suture displayed greater intrinsic

straight tensile strength than 8-0 Surgipro™ II Suture

Testing Outcomes

Ethicon Suture

8-0 Oversized (OS) PROLENE® Polypropylene Suture

Competitor

Medtronic 8-0 Surgipro™ II Polypropylene Monofilament Suture

ETHICON, 8-0 OS PROLENE

MEDTRONIC, 8-0 Surgipro II

217.5

173.8

ETHICON, 8-0 OS PROLENE

MEDTRONIC, 8-0 Surgipro II109.2 104.0

Straight Tensile Strength (gram-force)

Intrinsic Straight Tensile Strength (kpsi)

• 8-0 OS PROLENE Suture displayed greater knot tensile

strength than 8-0 Surgipro™ II Suture

ETHICON, 8-0 OS PROLENE

MEDTRONIC, 8-0 Surgipro II

167.4134.8

Knot Tensile Strength (gram-force)

Page 41: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

• Size 1 VICRYL Suture

displayed greater straight

tensile strength than

size 1 Novosyn® Suture

• Size 1 VICRYL Suture

displayed greater knot

tensile strength than

size 1 Novosyn® Suture

Testing Outcomes

Greater straight tensile strength✓

Greater knot tensile strength✓Greater intrinsic straight tensile strength✓

Less package memory✓

ETHICON, Size 1 VICRYL

B. BRAUN, Size 1 Novosyn

28.74 27.61

ETHICON, Size 1 VICRYL

B. BRAUN, Size 1 Novosyn

15.39 13.91

Straight Tensile Strength (lbs)

Knot Tensile Strength (lbs)

• Size 1 VICRYL Suture

displayed greater intrinsic

straight tensile strength

than size 1 Novosyn® Suture

ETHICON, Size 1 VICRYL

B. BRAUN, Size 1 Novosyn

94.9

84.5

Intrinsic Straight Tensile Strength (kpsi)

• Size 1 VICRYL Suture

displayed less package

memory than size 1

Novosyn® Suture

ETHICON, Size 1 VICRYL

B. BRAUN, Size 1 Novosyn

19.87

30.87

Package Memory - Short Pack (%)

Reference: Technical Report PE-2017-0019. September 25, 2017. Ethicon, Inc.

Ethicon Sutures

Competitive Performance Testing

41

Ethicon Suture

Size 1 Coated VICRYL® (polyglactin 910) Suture

Competitor

B. Braun Size 1 Novosyn® Suture

Page 42: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

28.75 27.61

ETHICON, Size 1 VICRYL Plus

B. BRAUN, Size 1 Novosyn

Reference: Technical Report PE-2017-0019. September 25, 2017. Ethicon, Inc. 42

Ethicon Sutures

Competitive Performance TestingEthicon Suture

Size 1 Coated VICRYL® Plus Antibacterial

(polyglactin 910) Suture

Competitor

B. Braun Size 1 Novosyn® Suture

• Size 1 VICRYL Plus Suture

displayed greater straight

tensile strength than

size 1 Novosyn® Suture

• Size 1 VICRYL Plus Suture

displayed greater intrinsic

straight tensile strength than

size 1 Novosyn® Suture

Testing Outcomes

Greater straight tensile strength✓Greater intrinsic straight tensile strength✓

Greater knot tensile strength ✓

ETHICON, Size 1 VICRYL Plus

B. BRAUN, Size 1 Novosyn

Straight Tensile Strength (lbs)

Intrinsic Straight Tensile Strength (kpsi)

88.084.5

ETHICON, Size 1 VICRYL Plus

B. BRAUN, Size 1 Novosyn

• Size 1 VICRYL Plus Suture

displayed greater knot

tensile strength than

size 1 Novosyn® Suture

Knot Tensile Strength (lbs)

15.13 13.91

Page 43: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

• Size 1 VICRYL Suture displayed

greater intrinsic straight tensile

strength than size 1 MITSU AB™

Suture

• Size 1 VICRYL Suture

displayed greater in vivo

breaking strength retention

(BSR) at 14, 21, and 28 days

compared to size 1 MITSU

AB™ Suture

Testing Outcomes

Greater intrinsic straight tensile strength✓Greater in vivo breaking strength retention✓Greater controlled linear elongation✓More reliable knot slide✓

ETHICON, Size 1 VICRYL

MERIL, Size 1 MITSU AB

ETHICON, Size 1 VICRYL

MERIL, Size 1 MITSU AB

94.989.4

Intrinsic Straight Tensile Strength (kpsi)

• Size 1 VICRYL Suture

displayed greater controlled

linear elongation than size 1

MITSU AB™ Suture

ETHICON, Size 1 VICRYL

MERIL, Size 1 MITSU AB

23.7

18.0

Controlled Linear Elongation (%)

ETHICON, Size 1 VICRYL

MERIL, Size 1 MITSU AB

97

0

Knot Slide Occurrences (%)

• Size 1 VICRYL Suture

displayed more reliable

knot slide compared to

size 1 MITSU AB™ Suture

Reference: Technical Report PE-2017-0039. April 17, 2018. Ethicon, Inc.

Ethicon Sutures

Competitive Performance Testing

43

Ethicon Suture

Size 1 Coated VICRYL® (polyglactin 910) Suture

Competitor

Meril Size 1 MITSU AB™ Suture

14 Days 21 Days 28 Days

22.615.7

9.9

18.2

0.0

10.5

In Vivo BSR Load (lbs)

Page 44: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

• Size 1 VICRYL Plus Suture

displayed greater in vivo

breaking strength retention

(BSR) at 14, 21, and 28 days

compared to size 1 MITSU

AB™ Suture

• Size 1 VICRYL Plus Suture

displayed greater controlled

linear elongation than size 1

MITSU AB™ Suture

Testing Outcomes

Greater in vivo breaking strength retention✓Greater controlled linear elongation✓More reliable knot slide✓

• Size 1 VICRYL Plus Suture

displayed more reliable

knot slide compared to

size 1 MITSU AB™ Suture

Reference: Technical Report PE-2017-0039. April 17, 2018. Ethicon, Inc.

Ethicon Sutures

Competitive Performance Testing

44

Ethicon Suture

Size 1 Coated VICRYL® Plus Antibacterial

(polyglactin 910) Suture

Competitor

Meril Size 1 MITSU AB™ Suture

ETHICON, Size 1 VICRYL Plus

MERIL, Size 1 MITSU AB

14 Days 21 Days 28 Days

20.615.7

9.9

14.6

0.05.2

In Vivo BSR Load (lbs)

ETHICON, Size 1 VICRYL Plus

MERIL, Size 1 MITSU AB

100

0

Knot Slide Occurrences (%)

ETHICON, Size 1 VICRYL Plus

MERIL, Size 1 MITSU AB23.918.0

Controlled Linear Elongation (%)

Page 45: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Sutures

Competitive Performance Testing

Reference: Technical Report PE-2016-0025. February 28, 2017. Ethicon, Inc.45

• Size 0 VICRYL Suture displayed greater

intrinsic straight tensile strength than

size 0 Polysorb™ Suture

Testing Outcomes

Ethicon Suture

Size 0 Coated VICRYL® (polyglactin 910) Suture

Competitor

Medtronic Size 0 Polysorb™ Suture

Greater intrinsic straight tensile strength✓Less package memory✓More reliable knot slide✓

ETHICON, Size 0 VICRYL

MEDTRONIC, Size 0 Polysorb

109.8102.7

Intrinsic Straight Tensile Strength (kpsi)

• Size 0 VICRYL Suture displayed

less package memory than

size 0 Polysorb™ Suture

ETHICON, Size 0 VICRYL

MEDTRONIC, Size 0 Polysorb 11.80

18.93

Package Memory - Short Pack (%)

• Size 0 VICRYL Suture displayed

more reliable knot slide compared

to size 0 Polysorb™ SutureETHICON, Size 0 VICRYL

MEDTRONIC, Size 0 Polysorb

100

53

Knot Slide Occurrences (%)

Page 46: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Sutures

Competitive Performance Testing

Reference: Technical Report PE-2016-0025. February 28, 2017. Ethicon, Inc.46

• Size 0 VICRYL Suture displayed greater intrinsic straight

tensile strength than size 0 Novosyn® Suture

Testing Outcomes

Ethicon Suture

Size 0 Coated VICRYL® (polyglactin 910) Suture

Competitor

B. Braun Size 0 Novosyn® Suture

Greater knot tensile strength✓Less package memory✓

Greater intrinsic straight tensile strength✓

Greater knot security✓

ETHICON, Size 0 VICRYL

B. BRAUN, Size 0 Novosyn

109.8103.0

Intrinsic Straight Tensile Strength (kpsi)

• Size 0 VICRYL Suture displayed greater knot tensile

strength than size 0 Novosyn® SutureETHICON, Size 0 VICRYL

B. BRAUN, Size 0 Novosyn11.53 11.04

Knot Tensile Strength (lbs)

• Size 0 VICRYL Suture displayed less package memory

than size 0 Novosyn® Suture

ETHICON, Size 0 VICRYL

B. BRAUN, Size 0 Novosyn

11.80

28.80

Package Memory - Short Pack (%)

• Size 0 VICRYL Suture displayed greater knot security

compared to size 0 Novosyn® Suture

ETHICON, Size 0 VICRYL

B. BRAUN, Size 0 Novosyn

8273

Knot Security Occurrences (%)

Page 47: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Sutures

Competitive Performance Testing

Reference: Technical Report PE-2016-0025. February 28, 2017. Ethicon, Inc.47

Greater intrinsic straight tensile strength✓More reliable knot slide✓

• Size 0 VICRYL Plus Suture displayed more reliable

knot slide compared to size 0 Polysorb™ Suture

ETHICON, Size 0 VICRYL Plus

MEDTRONIC, Size 0 Polysorb53

Knot Slide Occurrences (%)

• Size 0 VICRYL Plus Suture displayed greater intrinsic

straight tensile strength than size 0 Polysorb™ Suture

Testing Outcomes

Ethicon Suture

Size 0 Coated VICRYL® Plus Antibacterial (polyglactin 910) Suture

Competitor

Medtronic Size 0 Polysorb™ Suture

ETHICON, Size 0 VICRYL Plus

MEDTRONIC, Size 0 Polysorb

107.3 102.7

Intrinsic Straight Tensile Strength (kpsi)

100

Page 48: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Sutures

Competitive Performance Testing

Reference: Technical Report PE-2016-0025. February 28, 2017. Ethicon, Inc.48

Greater straight tensile strength✓Greater intrinsic straight tensile strength✓Greater knot tensile strength✓

• Size 0 VICRYL Plus Suture displayed greater intrinsic

straight tensile strength than size 0 Novosyn® SutureETHICON, Size 0 VICRYL Plus

B. BRAUN, Size 0 Novosyn

107.3 103.0

Intrinsic Straight Tensile Strength (kpsi)

• Size 0 VICRYL Plus Suture displayed greater

knot tensile strength than size 0 Novosyn® SutureETHICON, Size 0 VICRYL Plus

B. BRAUN, Size 0 Novosyn11.58 11.04

Knot Tensile Strength (lbs)

• Size 0 VICRYL Plus Suture displayed greater straight

tensile strength than size 0 Novosyn® Suture

Testing Outcomes

Ethicon Suture

Size 0 Coated VICRYL® Plus Antibacterial (polyglactin 910) Suture

Competitor

B. Braun Size 0 Novosyn® Suture

ETHICON, Size 0 VICRYL Plus

B. BRAUN, Size 0 Novosyn

23.22 22.32

Straight Tensile Strength (lbs)

Page 49: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Sutures

Competitive Performance Testing

Reference: Technical Report PE-2016-0025. December 15, 2016. Ethicon, Inc.49

Testing Outcomes

Ethicon Suture

2-0 Coated VICRYL® (polyglactin 910) Suture

Competitor

Medtronic 2-0 Polysorb™ Suture

Greater intrinsic straight tensile strength✓Less package memory✓More reliable knot slide✓

• 2-0 VICRYL Suture displayed less package memory

than 2-0 Polysorb™ Suture

ETHICON, 2-0 VICRYL

MEDTRONIC, 2-0 Polysorb 14.00

24.20

Package Memory - Short Pack (%)

• 2-0 VICRYL Suture displayed greater intrinsic straight

tensile strength than 2-0 Polysorb™ Suture

ETHICON, 2-0 VICRYL

MEDTRONIC, 2-0 Polysorb

125.3 119.0

Intrinsic Straight Tensile Strength (kpsi)

• 2-0 VICRYL Suture displayed more reliable knot

slide compared to 2-0 Polysorb™ SutureETHICON, 2-0 VICRYL

MEDTRONIC, 2-0 Polysorb

100

63

Knot Slide Occurrences (%)

Page 50: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Sutures

Competitive Performance Testing

Reference: Technical Report PE-2016-0025. December 15, 2016. Ethicon, Inc.50

25.10

• 2-0 VICRYL Plus Suture displayed less package memory

than 2-0 Polysorb™ Suture

Testing Outcomes

Ethicon Suture

2-0 Coated VICRYL® Plus Antibacterial (polyglactin 910) Suture

Competitor

Medtronic 2-0 Polysorb™ Suture

Less package memory✓More reliable knot slide✓

ETHICON, 2-0 VICRYL Plus

MEDTRONIC, 2-0 Polysorb16.3324.20

Package Memory - Short Pack (%)

• 2-0 VICRYL Plus Suture displayed more reliable

knot slide compared to 2-0 Polysorb™ SutureETHICON, 2-0 VICRYL Plus

MEDTRONIC, 2-0 Polysorb

100

63

Knot Slide Occurrences (%)

Page 51: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Sutures

Competitive Performance Testing

Reference: Technical Report PE-2016-0025. December 15, 2016. Ethicon, Inc.51

• 2-0 VICRYL Suture displayed greater straight tensile

strength than 2-0 Novosyn® Suture

• 2-0 VICRYL Suture displayed greater intrinsic straight

tensile strength than 2-0 Novosyn® Suture

Testing Outcomes

Ethicon Suture

2-0 Coated VICRYL® (polyglactin 910) Suture

Competitor

B. Braun 2-0 Novosyn® Suture

Greater straight tensile strength✓Greater intrinsic straight tensile strength✓Less package memory✓Greater knot security✓

ETHICON, 2-0 VICRYL

B. BRAUN, 2-0 Novosyn

16.48 16.05

ETHICON, 2-0 VICRYL

B. BRAUN, 2-0 Novosyn

125.3 120.4

Straight Tensile Strength (lbs)

Intrinsic Straight Tensile Strength (kpsi)

• 2-0 VICRYL Suture displayed less package memory than

2-0 Novosyn® Suture

ETHICON, 2-0 VICRYL

B. BRAUN, 2-0 Novosyn

14.00

25.10

Package Memory - Short Pack (%)

• 2-0 VICRYL Suture displayed greater knot security

compared to 2-0 Novosyn® SutureETHICON, 2-0 VICRYL

B. BRAUN, 2-0 Novosyn

92

72

Knot Security Occurrences (%)

Page 52: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Sutures

Competitive Performance Testing

Reference: Technical Report PE-2016-0025. December 15, 2016. Ethicon, Inc.52

• 2-0 VICRYL Plus Suture displayed less package memory

than 2-0 Novosyn® Suture

Testing Outcomes

Ethicon Suture

2-0 Coated VICRYL® Plus Antibacterial (polyglactin 910) Suture

Competitor

B. Braun 2-0 Novosyn® Suture

Less package memory✓Greater knot security✓

ETHICON, 2-0 VICRYL Plus

B. BRAUN, 2-0 Novosyn

16.33

25.10

Package Memory - Short Pack (%)

ETHICON, 2-0 VICRYL Plus

B. BRAUN, 2-0 Novosyn

87

72

Knot Security Occurrences (%)

• 2-0 VICRYL Plus Suture displayed greater knot security

compared to 2-0 Novosyn® Suture

Page 53: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Sutures

Competitive Performance Testing

Reference: Technical Report PE-2016-0025. January 11, 2017. Ethicon, Inc.53

• 3-0 VICRYL Suture displayed less package

memory than 3-0 Polysorb™ Suture

Testing Outcomes

Ethicon Suture

3-0 Coated VICRYL® (polyglactin 910) Suture

Competitor

Medtronic 3-0 Polysorb™ Suture

Less package memory✓More reliable knot slide✓

ETHICON, 3-0 VICRYL

MEDTRONIC, 3-0 Polysorb

12.17

22.03

Package Memory - Short Pack (%)

• 3-0 VICRYL Plus Suture displayed less package

memory than 3-0 Polysorb™ Suture

Testing Outcomes

Ethicon Suture

3-0 Coated VICRYL® Plus Antibacterial (polyglactin 910) Suture

Competitor

Medtronic 3-0 Polysorb™ Suture

Less package memory✓Superior knot slide characteristics✓

ETHICON, 3-0 VICRYL Plus

MEDTRONIC, 3-0 Polysorb

17.70

22.03

Package Memory - Short Pack (%)

ETHICON, 3-0 VICRYL

MEDTRONIC, 3-0 Polysorb

100

66

Knot Security Occurrences (%)

Knot Security Occurrences (%)

• 3-0 VICRYL Suture displayed more reliable

knot slide compared to 3-0 Polysorb™ Suture

ETHICON, 3-0 VICRYL Plus

MEDTRONIC, 3-0 Polysorb

100

66

• 3-0 VICRYL Plus Suture displayed more reliable

knot slide compared to 3-0 Polysorb™ Suture

Page 54: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Sutures

Competitive Performance Testing

Reference: Technical Report PE-2016-0025. January 11, 2017. Ethicon, Inc.54

• 3-0 VICRYL Suture displayed greater straight

tensile strength than 3-0 Novosyn® Suture

• 3-0 VICRYL Suture displayed greater knot tensile

strength than 3-0 Novosyn® Suture

Testing Outcomes

Ethicon Suture

3-0 Coated VICRYL® (polyglactin 910) Suture

Competitor

B. Braun 3-0 Novosyn® Suture

Greater straight tensile strength✓Greater intrinsic straight tensile strength✓Greater knot tensile strength✓Less package memory✓

ETHICON, 3-0 VICRYL

B. BRAUN, 3-0 Novosyn

10.01 9.45

ETHICON, 3-0 VICRYL

B. BRAUN, 3-0 Novosyn

5.24 4.91

Straight Tensile Strength (lbs)

• 3-0 VICRYL Suture displayed greater intrinsic

straight tensile strength than 3-0 Novosyn® Suture

ETHICON, 3-0 VICRYL

B. BRAUN, 3-0 Novosyn

121.1111.4

Intrinsic Straight Tensile Strength (kpsi)

Knot Tensile Strength (lbs)

• 3-0 VICRYL Suture displayed less package memory

than 3-0 Novosyn® Suture

ETHICON, 3-0 VICRYL

B. BRAUN, 3-0 Novosyn

12.17

23.87

Package Memory - Short Pack (%)

Page 55: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Sutures

Competitive Performance Testing

Reference: Technical Report PE-2016-0025. January 11, 2017. Ethicon, Inc.55

• 3-0 VICRYL Plus Suture displayed greater

straight tensile strength than 3-0 Novosyn® Suture

• 3-0 VICRYL Plus Suture displayed greater

knot tensile strength than 3-0 Novosyn® Suture

Testing Outcomes

Ethicon Suture

3-0 Coated VICRYL® Plus Antibacterial (polyglactin 910) Suture

Competitor

B. Braun 3-0 Novosyn® Suture

Greater straight tensile strength✓Greater knot tensile strength✓Less package memory✓

ETHICON, 3-0 VICRYL Plus

B. BRAUN, 3-0 Novosyn

9.76 9.45

ETHICON, 3-0 VICRYL Plus

B. BRAUN, 3-0 Novosyn

5.314.91

Straight Tensile Strength (lbs)

Knot Tensile Strength (lbs)

• 3-0 VICRYL Plus Suture displayed less

package memory than 3-0 Novosyn® Suture

ETHICON, 3-0 VICRYL Plus

B. BRAUN, 3-0 Novosyn

17.70

23.87

Package Memory - Short Pack (%)

Page 56: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Sutures

Competitive Performance Testing

References: 1. Tissue-holding comparisons, 100326296. May 26, 2015. Ethicon, Inc. 2. STRATAFIX™ Symmetric PDS™ Plus Knotless Tissue Control Device,Instructions for Use. Ethicon, Inc. Somerville, NJ. 3. V-Loc™ 180 Absorbable Wound Closure Device, Instructions for Use, Medtronic. 56

Ethicon Suture

STRATAFIX™ Symmetric PDS™ Plus

Knotless Tissue Control Device

Competitor

Medtronic V-Loc™ 180 Absorbable

Wound Closure Device

Greater tissue-holding strength*

Greater extended tissue support

✓✓

• STRATAFIX Symmetric PDS Plus Device

displayed greater tissue-holding

strength than V-Loc™ 180 Device*

• STRATAFIX Symmetric PDS Plus Device

provides 6 weeks of tissue support while

V-Loc™ 180 Device provides support for

only 3 weeks

Testing Outcomes

*In porcine fascia

MEDTRONIC, V-Loc 180 Device

90.06

78.38ETHICON, STRATAFIX Symmetric PDS Plus Device

Days Implantation

% S

tre

ng

th R

ete

nti

on

90

80

70

60

50

7 14 21 28 35 42

ETHICON, STRATAFIX Symmetric PDS Plus Device

MEDTRONIC, V-Loc 180 Device

Tissue-Holding Strength (lbf)

Page 57: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Sutures

Competitive Performance Testing

References: 1. Report for Assessment 100552029. July 14, 2017. Ethicon, Inc. 2. Report for Assessment 100559286. July 19, 2017. Ethicon, Inc. 3. Performance Testing Report 100491281. November 17, 2016. 57

Ethicon Suture

STRATAFIX™ Spiral PDS™ Plus

Knotless Tissue Control Device

Competitor

Medtronic V-Loc™ 180 Absorbable

Wound Closure Device

Greater extended tissue support

Statistically equivalent tissue-holding strength

✓=

• STRATAFIX Spiral PDS Plus Device displayed greater extended tissue support than V-Loc™ 180 Device over a range of sizes

Testing Outcomes

Day 0 Day 28 In Vitro Day 42 In Vitro

100 100

55

81

1

50

% S

tre

ng

th R

ete

nti

on

MEDTRONIC, Size 0 V-Loc 180 Device

ETHICON, Size 0 STRATAFIXSpiral PDS Plus Device

MEDTRONIC, Size 2-0 V-Loc 180 Device

ETHICON, Size 2-0 STRATAFIXSpiral PDS Plus Device

Day 0 Day 28 In Vitro Day 42 In Vitro

100 100

72

88

3

45

% S

tre

ng

th R

ete

nti

on

Day 0 Day 28 In Vitro Day 42 In Vitro

100 100

86

95

17

60

% S

tre

ng

th R

ete

nti

on

MEDTRONIC, Size 3-0 V-Loc 180 Device

ETHICON, Size 3-0 STRATAFIXSpiral PDS Plus Device

MEDTRONIC, Size 4-0 V-Loc 180 Device

ETHICON, Size 4-0 STRATAFIXSpiral PDS Plus Device

Day 0 Day 28 In Vitro Day 42 In Vitro

100 100

7168

14

61

% S

tre

ng

th R

ete

nti

on

• STRATAFIX Spiral PDS Plus Device displayed

statistically equivalent tissue-holding strength

compared to V-Loc™ 180 Device over a range

of sizes

MEDTRONIC, 3-0 V-Loc 180 Device

50.961.4

ETHICON, 3-0 STRATAFIX Spiral PDS Plus Device

Tissue-Holding Strength (lbf)

MEDTRONIC, 4-0 V-Loc 180 Device

26.923.4

ETHICON, 4-0 STRATAFIX Spiral PDS Plus Device

Tissue-Holding Strength (lbf)

Page 58: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Ethicon Sutures

Competitive Performance Testing

References: 1. Report for Assessment 100559286. July 19, 2017. Ethicon, Inc. 2. Performance Testing Report 100491281. November 17, 2016.

58

Ethicon Suture

STRATAFIX™ Spiral PDS™ Plus Knotless

Tissue Control Device

Competitor

Ethicon PDS® Plus Antibacterial

(polydioxanone) Suture

Greater tissue-holding strength✓

• STRATAFIX Spiral PDS Plus Device displayed

greater tissue-holding strength than PDS Plus

Suture over a range of sizes

Testing Outcomes ETHICON, Size 1 PDS Plus Device

93.3

69.0 ETHICON, Size 1 STRATAFIX Spiral PDS Plus Device

Tissue-Holding Strength (lbf)

ETHICON, 4-0 PDS Plus Device

26.9

15.9

ETHICON 4-0 STRATAFIX Spiral PDS Plus Device

Tissue-Holding Strength (lbf)

• STRATAFIX Spiral MONOCRYL Plus Device displayed

statistically equivalent tissue-holding strength compared

to V-Loc™ 90 Device over a range of sizes

Testing Outcomes

Ethicon Suture

STRATAFIX™ Spiral MONOCRYL™ Plus Knotless Tissue Control Device

Competitor

Medtronic V-Loc™ 90 Absorbable Wound Closure Device

Statistically equivalent tissue-holding strength=

MEDTRONIC, 2-0 V-Loc 90 Device56.5

45.8

ETHICON, 2-0 STRATAFIX Spiral MONOCRYL Plus Device

Tissue-Holding Strength (lbf)

MEDTRONIC, 4-0 V-Loc 90 Device

22.0 18.6 ETHICON, 4-0 STRATAFIX Spiral MONOCRYL Plus Device

Tissue-Holding Strength (lbf)

Page 59: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

14.009.33

Ethicon Regional Sutures

Competitive Performance Testing

Reference: Technical Report PE-2017-0028. September 22, 2017. Ethicon, Inc. 59

Ethicon Suture

2-0 ETHIBOND EXCEL® Polyester Suture

Asia-Pacific Competitor

Alfresa 2-0 Nescosuture® Suture

Greater straight tensile strength✓Greater intrinsic straight tensile strength✓Greater knot security✓

• 2-0 ETHIBOND EXCEL Suture

displayed greater straight

tensile strength than 2-0

Nescosuture® Suture

• 2-0 ETHIBOND EXCEL Suture

displayed greater intrinsic

straight tensile strength than

2-0 Nescosuture® Suture

Testing Outcomes

Straight Tensile Strength (lbs)

ETHICON, 2-0 ETHIBOND

ALFRESA, 2-0 Nescosuture

103.6

60.5

Intrinsic Straight Tensile Strength (kpsi)

ETHICON, 2-0 ETHIBOND

ALFRESA, 2-0 Nescosuture

80

2

Knot Security Occurrences (%)

ETHICON, 2-0 ETHIBOND

ALFRESA, 2-0 Nescosuture

• 2-0 ETHIBOND EXCEL Suture

displayed greater knot

security compared to

2-0 Nescosuture® Suture

Page 60: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

Suture Glossary

Straight Tensile Strength vs. Intrinsic Straight Tensile Strength

Straight tensile strength is a measure of the maximum load a suture

can resist before breaking. However, in a head-to-head matchup

when the diameter of one of the sutures exceeds the USP size listed

on the box—known as oversizing—this measurement alone does not

accurately describe which suture material is stronger. To correct for

oversizing, Ethicon Edge factors in suture diameter as well as

straight tensile strength to calculate intrinsic straight tensile strength.

When a particular matchup yields an advantage for the Ethicon

suture for intrinsic straight tensile strength, but not one for straight

tensile strength, the competitive suture may have been oversized.

Controlled Linear Elongation

Linear elongation is a key factor in how a suture handles and feels,

and refers to how much a suture stretches under tension. Ideally, a

suture should elongate to a point under tension to allow for wound

edema, but then return to its original length when edema subsides

to maintain the integrity of the closure. A suture that stretches more

and is able to return to its original length afterwards is said to exhibit

greater controlled linear elongation.

Knot Slide vs. Knot Security

Knot slide refers to a knot’s ability to slide down the suture during

tying, and it is essential to proper function. Ideally, the knot will slide

completely closed and then hold firm as pressure is applied to the

suture ends, eventually breaking at the knot. Failure of the knot to

slide properly may indicate a poorly constructed suture or an

inadequate coating.

The knot security test verifies that the knot will hold as pressure is

exerted on the ends of the suture. A passing knot exhibits no

movement or slippage before breaking, as indicated by a broken

suture with pigtails of equal length at the conclusion of the test.

60

Straight Tensile Strength

Controlled linear elongation

Knot Slide

Knot Security

Page 61: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

ADVANCED MEDICAL SOLUTIONS, Skin Affix

10.13

3.54

ETHICON, DERMABOND ADVANCED Adhesive

Tensile Strength (lbf)

ADVANCED MEDICAL SOLUTIONS, Skin Affix

95

ETHICON, DERMABOND ADVANCED Adhesive

Drying Time (sec)

Ethicon Topical Skin Adhesives

Performance Testing

References: 1. Protocol Report 09PD070. January 15, 2010. Ethicon, Inc. 2. Performance Evaluation AST-2013-0548. September 30, 2014. Ethicon, Inc. 3. Technical Report 11TR053. August 16, 2011. Ethicon, Inc. 4. Performance Evaluation AST-2015-0256. August 13, 2015. Ethicon, Inc.

• In separate studies, DERMABOND ADVANCED

Adhesive displayed greater tensile strength (2-D)

than Skin Affix™1,2*

• In separate studies, DERMABOND ADVANCED

Adhesive displayed faster drying time than

Skin Affix™2,3†

Testing Outcomes

Ethicon Topical Skin Adhesive

DERMABOND ADVANCED® Topical Skin Adhesive

Competitor

Advanced Medical Solutions Skin Affix™ Topical Skin Adhesive

Faster drying time2,3†✓Greater tensile strength1,2*✓

171

MCKESSON, Liquiband Exceed

10.13

6.40

ETHICON, DERMABOND ADVANCED Adhesive

Tensile Strength (lbf)

• In separate studies, DERMABOND ADVANCED

Adhesive displayed greater tensile strength (2-D)

than Liquiband® Exceed™1,4‡

Testing Outcomes

Ethicon Topical Skin Adhesive

DERMABOND ADVANCED® Topical Skin Adhesive

Competitor

McKesson Liquiband® Exceed™ Topical Skin Adhesive

Greater tensile strength1,4‡✓

61

*DERMABOND ADVANCED Adhesive value based on benchtop testing (09PD070) of porcine skin using the end-to-end skin tensile method RD-6-245 (N=50). Skin Affix™ value based on benchtop testing (AST-2013-0548) of porcine skin using end-to-end skin tensile method RD-6-245 (N=20).

†DERMABOND ADVANCED Adhesive value based on benchtop testing (11TR053) of porcine skin heated to 33ºC; product considered fully polymerized when no material transferred to cotton swab from any area of the application site (N=12). Skin Affix™ value based on benchtop testing (AST-2013-0548) of porcine skin heated to 33ºC; product considered set when no adhesive transferred to cotton swab (N=20).

‡DERMABOND ADVANCED Adhesive value based on benchtop testing (09PD070) of porcine skin using the end-to-end skin tensile method RD-6-245 (N=50). Liquiband® Exceed™ value based on benchtop testing (AST-2015-0256) of porcine skin using end-to-end skin tensile method RD-6-245 (N=23).

Page 62: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

ETHICON, DERMABOND ADVANCED Adhesive

with subcuticular 4-0 MONOCRYL Sutures

20.59

13.92

9.30 ETHICON, DERMABOND PRINEO System

ETHICON, DERMABOND ADVANCED Adhesive

without subcuticular sutures

Tissue-Holding Strength (lbs)

Ethicon Topical Skin Adhesives

Competitive Performance Testing

Reference: Performance Evaluation 100253930. September 4, 2014. Ethicon, Inc.62

• DERMABOND PRINEO System displayed

greater tissue-holding strength than DERMABOND

ADVANCED Adhesive with or without subcuticular

4-0 MONOCRYL® (poliglecaprone 25) Sutures

Testing Outcomes

Ethicon Topical Skin Adhesive

DERMABOND® PRINEO® Skin Closure System (22 cm)

Competitor

ETHICON DERMABOND ADVANCED® Topical Skin Adhesive

with and without subcuticular sutures

Greater tissue-holding strength✓

Page 63: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

MEDTRONIC, Endo Stitch with Surgidac

14.2812.96

ETHICON, PROXISURE with ETHIBOND

Knot Strength (lbf)

PENETRATION FORCE REQUIRED

MEDTRONIC, Endo Stitch Straight Needle

ETHICON, PROXISURE Curved Needle

GOOD

Pe

ne

tra

tio

n F

orc

e (

gra

ms)

Pass

1 35

56

173 168

85

Ethicon Advanced Suturing Systems

Competitive Performance Testing

References: 1. Technical Memo CT16-001. May 11, 2016. Ethicon, Inc. 2. Engineering Study PRC073011. December 24, 2015. Ethicon, Inc.63

• PROXISURE Suturing Device’s curved

needle required 3x less penetration force

than the straight Endo Stitch™ needle1*

• PROXISURE Suturing Device with

ETHIBOND EXCEL Suture achieved

greater knot strength than Endo

Stitch™ with Surgidac™ suture2†

Testing Outcomes

Ethicon Advanced Suturing System

PROXISURE™ Suturing Device

with ETHIBOND EXCEL® Polyester Suture

Less penetration force required1*✓Greater knot strength2†✓

Competitor

Medtronic Endo Stitch™ Suturing Device with Surgidac™

Polyester Suture

†Size 2-0 sutures were tested using a benchtop method and equipment and material as described in test method DOC021468 (ETHIBOND N=19, Surgidac N=17).

*Needles tested through 35 passes in 0.4 mm Permair in accordance with TM406-080 (N=10). All measurements taken using an optical comparator or dial indicator.

Page 64: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

360º coverage around catheter site✓

Easier removal designed to lower the risk of catheter dislodgment

64

Ethicon Vascular Access Device Protection

Competitive Performance TestingEthicon IV Protection

BIOPATCH® Protective Disk with CHG

Competitor

3M Tegaderm™ CHG Chlorhexidine Gluconate IV Securement Dressing

*BIOCLUSIVE is the securement dressing used with BIOPATCH Disk.References: 1. Westergom C, Mistry P, Bhende S, Aickin S. Comparative In Vitro/Ex Vivo Analysis of Two Catheter-Site Insertion Dressings. Presented at the Association for Vascular Access: September 11-14, 2008. Ethicon, Inc. 2. CDC Guidelines for the Prevention of Intravascular Catheter-Related Infections, 2011. 3. Approved claim 062483-161028. Ethicon, Inc. 4. Maki DG, Mermel LA, Kluger D, et al. The Efficacy of a Chlorhexidine-Impregnated Sponge (Biopatch) for the Prevention of Intravascular Catheter-Related Infection – A Prospective, Randomized, Controlled Multicenter Study. Abstracts of the 40th Interscience Conference on Antimicrobial Agents and Chemotherapy, September 2000, p. 422. 5. Ruschulte H, Franke M, Gastmeier P, et al. Prevention of central venous catheter related infections with chlorhexidine gluconate impregnated wound dressings: a randomized controlled trial. Ann Hematol. 2009;88:267-272. 6. Timsit JF, Schwebel C, Bouadma L, et al. Chlorhexidine-Impregnated Sponges and Less Frequent Dressing Changes for Prevention of Catheter-Related Infections in Critically Ill Adults. JAMA. 2009;301(12):1231-124. 7. Document BP-470-10-12/12. In Vitro Comparative Analysis of a Chlorhexidine Gluconate (CHG) Impregnated Sponge Dressing and a CHG-containing Hydrogel Dressing, 2010. Ethicon, Inc.

Only IV dressing with CHG proven in multiple randomized controlled trials (RCTs) to reduce incidence of catheter-related bloodstream infections (CRBSIs) in central venous and arterial catheters

• BIOPATCH Disk allows 360º delivery of CHG

around the catheter insertion site; Tegaderm™

CHG does not1

• BIOPATCH Disk is the only

IV dressing with CHG proven

in multiple RCTs to reduce

incidence of CRBSIs in central

venous and arterial catheters3

Testing Outcomes

• BIOPATCH Disk allows for easier

removal than Tegaderm™ CHG

which may lower the risk of

catheter dislodgment7

The density of skin flora at the catheter

insertion site is a major risk factor

for CRBSI2

3M, Tegaderm

ETHICON, BIOCLUSIVE*

3M, Tegaderm

0 5 73

Ne

wto

ns

16

12

8

4

0

Time (days)

FORCE REQUIRED TO REMOVE DRESSING

ETHICON, BIOPATCH

Maki, et al. 20004 687 BIOPATCH Disk significantly reduced the risk of life-threatening

CRBSI (1.2% in BIOPATCH group vs. 3.3% in control group, p=0.026)

Ruschulte, et al. 20095 601 BIOPATCH Disk significantly reduced CRBSI in immunocompromised

patients (6.3% in the BIOPATCH group vs. 11.3% in control group, p=0.016)

Study N Conclusion

Timsit, et al. 20096 1,636 Use of BIOPATCH Disk led to 69% decrease in CRBSI rate versus

non-chlorhexidine dressings (p=0.03)

Visuals depict CHG delivery zones for each product.

Page 65: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

65

Ethicon Vascular Access Device Protection

Competitive Performance TestingEthicon IV Protection

BIOPATCH® Protective Disk with CHG

Competitor

Medline Aegis™ CHG-Impregnated Foam Disc

• Unlike Aegis, BIOPATCH Disk

meets the updated CDC 1A

guideline as stated in the “2017

Updated Recommendations on

the Use of Chlorhexidine-Impregnated

Dressings for Prevention of

Intravascular Catheter-Related

Infections”

Testing Outcomes

ETHICON, BIOPATCH

MEDLINE, Aegis

7

6

5

4

3

2

1

0

BIOPATCH Disk is the only IV dressing with CHG proven in multiple randomized controlled trials (RCTs) to reduce incidence of catheter-related bloodstream infections (CRBSIs) in central venous and arterial catheters

✓BIOPATCH Disk meets the updated CDC 1A guideline as stated in the “2017 Updated Recommendations on the Use of Chlorhexidine-Impregnated Dressings for Prevention of Intravascular Catheter-Related Infections”

*Swab test resultsReferences: 1. Approved claim 062483-161028. Ethicon, Inc. 2. Maki DG, Mermel LA, Kluger D, et al. The Efficacy of a Chlorhexidine-Impregnated Sponge (Biopatch) for the Prevention of Intravascular Catheter-Related Infection – A Prospective, Randomized, Controlled Multicenter Study. Abstracts of the 40th Interscience Conference on Antimicrobial Agents and Chemothera-py, September 2000, p. 422. 3. Ruschulte H, Franke M, Gastmeier P, et al. Prevention of central venous catheter related infections with chlorhexidine gluconate impregnated wound dressings: a randomized controlled trial. Ann Hematol. 2009;88:267-272. 4. Timsit JF, Schwebel C, Bouadma L, et al. Chlorhexidine-Impregnated Sponges and Less Frequent Dressing Changes for Prevention of Catheter-Related Infections in Critically Ill Adults. JAMA. 2009;301(12):1231124. 5. BIOPATCH, Guardiva, Aegis. Study report for in vitro microbiological evaluation of Antimicrobial Barrier Dressings using zone of inhibition assay. June 28, 2016. Ethicon, Inc.

MRSA MRSE VRE P. aeruginosa A. baumanii K. pneumoniae C. albicans

DAYS WITH SUSTAINED MICROBICIDAL ACTIVITY UNDER THE DRESSING5*

• BIOPATCH Disk is the only

IV dressing with CHG proven

in multiple RCTs to reduce

incidence of CRBSIs in central

venous and arterial catheters1

Maki, et al. 20002 687 BIOPATCH Disk significantly reduced the risk of life-threatening

CRBSI (1.2% in BIOPATCH group vs. 3.3% in control group, p=0.026)

Ruschulte, et al. 20093 601 BIOPATCH Disk significantly reduced CRBSI in immunocompromised

patients (6.3% in the BIOPATCH group vs. 11.3% in control group, p=0.016)

Study N Conclusion

Timsit, et al. 20094 1,636 Use of BIOPATCH Disk led to 69% decrease in CRBSI rate versus

non-chlorhexidine dressings (p=0.03)

Page 66: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

66

Ethicon Vascular Access Device Protection

Competitive Performance Testing

*Swab test resultsReferences: 1. Approved claim 062483-161028. Ethicon, Inc. 2. Maki DG, Mermel LA, Kluger D, et al. The Efficacy of a Chlorhexidine-Impregnated Sponge (Biopatch) for the Prevention of Intravascular Catheter-Related Infection – A Prospective, Randomized, Controlled Multicenter Study. Abstracts of the 40th Interscience Conference on Antimicrobial Agents and Chemothera-py, September 2000, p. 422. 3. Ruschulte H, Franke M, Gastmeier P, et al. Prevention of central venous catheter related infections with chlorhexidine gluconate impregnated wound dressings: a randomized controlled trial. Ann Hematol. 2009;88:267-272. 4. Timsit JF, Schwebel C, Bouadma L, et al. Chlorhexidine-Impregnated Sponges and Less Frequent Dressing Changes for Prevention of Catheter-Related Infections in Critically Ill Adults. JAMA. 2009;301(12):1231124. 5. BIOPATCH, Guardiva, Aegis. Study report for in vitro microbiological evaluation of Antimicrobial Barrier Dressings using zone of inhibition assay. June 28, 2016. Ethicon, Inc.

ETHICON, BIOPATCH

BARD, Guardiva

7

6

5

4

3

2

1

0

DAYS WITH SUSTAINED MICROBICIDAL ACTIVITY UNDER THE DRESSING5*

MRSA MRSE VRE P. aeruginosa A. baumanii K. pneumoniae C. albicans

Ethicon IV Protection

BIOPATCH® Protective Disk with CHG

Competitor

Bard GuardIVa® Antimicrobial Hemostatic IV Dressing

Testing Outcomes

• BIOPATCH Disk is the only

IV dressing with CHG proven

in multiple RCTs to reduce

incidence of CRBSIs in central

venous and arterial catheters1

• Unlike GuardIVa®, BIOPATCH Disk

meets the updated CDC 1A

guideline as stated in the “2017

Updated Recommendations on

the Use of Chlorhexidine-Impregnated

Dressings for Prevention of

Intravascular Catheter-Related

Infections”

Maki, et al. 20002 687 BIOPATCH Disk significantly reduced the risk of life-threatening

CRBSI (1.2% in BIOPATCH group vs. 3.3% in control group, p=0.026)

Ruschulte, et al. 20093 601 BIOPATCH Disk significantly reduced CRBSI in immunocompromised

patients (6.3% in the BIOPATCH group vs. 11.3% in control group, p=0.016)

Study N Conclusion

Timsit, et al. 20094 1,636 Use of BIOPATCH Disk led to 69% decrease in CRBSI rate versus

non-chlorhexidine dressings (p=0.03)

✓BIOPATCH Disk is the only IV dressing with CHG proven in multiple randomized controlled trials (RCTs) to reduce incidence of catheter-related bloodstream infections (CRBSIs) in central venous and arterial catheters

✓BIOPATCH Disk meets the updated CDC 1A guideline as stated in the “2017 Updated Recommendations on the Use of Chlorhexidine-Impregnated Dressings for Prevention of Intravascular Catheter-Related Infections”

Page 67: Ethicon Edge Competitive Performance Testing · Competitive Performance Testing *Overall Strength, measured as Bend Moment, is a product of Surgical Yield, Ultimate Moment, Reshape,

©2018 Ethicon US, LLC. All rights reserved. 063555-180509

67

Ethicon Vascular Access Device Protection

Competitive Performance TestingEthicon IV Protection

BIOPATCH® Protective Disk with CHG

Competitor

Medtronic Kendall™ AMD Antimicrobial Foam Disc Dressing

• Unlike Kendall™, BIOPATCH Disk

meets the updated CDC 1A

guideline as stated in the “2017

Updated Recommendations on

the Use of Chlorhexidine-Impregnated

Dressings for Prevention of

Intravascular Catheter-Related

Infections”

*Swab test resultsReferences: 1. Approved claim 062483-161028. Ethicon, Inc. 2. Maki DG, Mermel LA, Kluger D, et al. The Efficacy of a Chlorhexidine-Impregnated Sponge (Biopatch) for the Prevention of Intravascular Catheter-Related Infection – A Prospective, Randomized, Controlled Multicenter Study. Abstracts of the 40th Interscience Conference on Antimicrobial Agents and Chemotherapy, September 2000, p. 422. 3. Ruschulte H, Franke M, Gastmeier P, et al. Prevention of central venous catheter related infections with chlorhexidine gluconate impregnated wound dressings: a randomized controlled trial. Ann Hematol. 2009;88:267-272. 4. Timsit JF, Schwebel C, Bouadma L, et al. Chlorhexidine-Impregnated Sponges and Less Frequent Dressing Changes for Prevention of Catheter-Related Infections in Critically Ill Adults. JAMA. 2009;301(12):1231124. 5. BIOPATCH-PHMB. Study report for in vitro microbiological evaluation of Antimicrobial Barrier Dressings using zone of inhibition assay. June 28, 2016. Ethicon, Inc.

ETHICON, BIOPATCH

MEDTRONIC, Kendall

7

6

5

4

3

2

1

0 MRSA MRSE VRE P. aeruginosa A. baumanii K. pneumoniae C. albicans

DAYS WITH SUSTAINED MICROBICIDAL ACTIVITY UNDER THE DRESSING5*

Testing Outcomes

• BIOPATCH Disk is the only

IV dressing with CHG proven

in multiple RCTs to reduce

incidence of CRBSIs in central

venous and arterial catheters1

Maki, et al. 20002 687 BIOPATCH Disk significantly reduced the risk of life-threatening

CRBSI (1.2% in BIOPATCH group vs. 3.3% in control group, p=0.026)

Ruschulte, et al. 20093 601 BIOPATCH Disk significantly reduced CRBSI in immunocompromised

patients (6.3% in the BIOPATCH group vs. 11.3% in control group, p=0.016)

Study N Conclusion

Timsit, et al. 20094 1,636 Use of BIOPATCH Disk led to 69% decrease in CRBSI rate versus

non-chlorhexidine dressings (p=0.03)

✓BIOPATCH Disk is the only IV dressing with CHG proven in multiple randomized controlled trials (RCTs) to reduce incidence of catheter-related bloodstream infections (CRBSIs) in central venous and arterial catheters

✓BIOPATCH Disk meets the updated CDC 1A guideline as stated in the “2017 Updated Recommendations on the Use of Chlorhexidine-Impregnated Dressings for Prevention of Intravascular Catheter-Related Infections”

For complete indications, contraindications, warnings, precautions, and adverse reactions, please reference full package insert.