ET – 71 / ET – 91 Comparisons: General Data ET 71 A / ET 71 V / ETM 4 ET 91 Construction Generator, Receiver, Bridge separate All of them are in one instrument Application Up to 120 FDM channels Up to 600 FDM channels Dimensions ET 70 A: 350x220x200 mm ET 70 V: 350x220x200 mm ETM 4: 120x120x40 mm 224 x 160 x 44 mm Weight ET 70 A: ~9 kg ET 70 V: ~9 kg ETM 4: ~1 kg ~1.5 kg Temperature Nominal: +5..+45°C Operating: 0..55°C Storage, transport: –40 to +70°C Nominal: 0..+45°C Operating: -5..50°C Storage, transport: –40 to +70°C Power Mains From internal battery From car battery Mains From internal NIMH battery From car battery Display Level display: Galvanometer Frequency display: digital Level and frequency display is digital with 320x240 pixel graphic LCD display Memory Not available Memory locations for test results Memory locations for setups USB Interfaces PC connection is not possible USB 1.1 host port for USB stick USB 1.1 device port for PC connection PC Program PC connection is not possible PC program for result and setup transfer Operation Modes ET 71 A / ET 71 V / ETM 4 ET 91 Transmitting Provided Provided Selective Receiving Provided Provided Wide band Receiving Provided Provided Impedance Measurement With external adapter Provided Return Loss Measurement With external adapter Provided Balance Measurement With external adapter Provided Wide band Noise Measurement Not provided Additional Service Impulse Noise Measurement Not provided Additional Service Spectrum Analyzer Not provided Additional Service Master Slave Measurement Not provided Additional Service Multi Tone Measurement Not provided Additional Service NEXT/LOSS Measurement Not provided Additional Service Battery Quick Charge Not provided Additional Service Battery Regenerative Charge Not provided Additional Service Parameter ET 71 A ET 91 Frequency Range: 200 Hz...620 kHz Accuracy: 2x10-5 ±1 digit Resolution: ~1 Hz Range: 100 Hz... 2.4 MHz Accuracy: 2x10-6 ±1 digit Resolution: 1 Hz Input Impedance Symmetrical and Coax: 0, 75, 135 (125), 150, 600 Ω Symmetrical and Coax: 0, 75, 135(125), 150, 600 Ω Output Level Balanced: +10 to 50 dBm, dB Coax: +10 to –50 dBm, dB Balanced: +10 ... 50 dBm, dB Coax: +10 ... –50 dBm Level accuracy Balanced: ~ ± 0,5 dB Coax: ~ ± 0,6 dB Balanced: ~ ± 0,3 dB Coax: ~ ± 0,3 dB

description

.

Transcript of ET 91 comparision.pdf

  • ET 71 / ET 91 Comparisons:

    General Data ET 71 A / ET 71 V / ETM 4 ET 91

    Construction Generator, Receiver, Bridge separate All of them are in one instrument

    Application Up to 120 FDM channels Up to 600 FDM channels

    Dimensions

    ET 70 A: 350x220x200 mm

    ET 70 V: 350x220x200 mm

    ETM 4: 120x120x40 mm

    224 x 160 x 44 mm

    Weight

    ET 70 A: ~9 kg

    ET 70 V: ~9 kg

    ETM 4: ~1 kg

    ~1.5 kg

    Temperature

    Nominal: +5..+45C

    Operating: 0..55C

    Storage, transport: 40 to +70C

    Nominal: 0..+45C

    Operating: -5..50C

    Storage, transport: 40 to +70C

    Power

    Mains

    From internal battery

    From car battery

    Mains

    From internal NIMH battery

    From car battery

    Display Level display: Galvanometer

    Frequency display: digital

    Level and frequency display is digital with

    320x240 pixel graphic LCD display

    Memory Not available Memory locations for test results

    Memory locations for setups

    USB Interfaces PC connection is not possible USB 1.1 host port for USB stick

    USB 1.1 device port for PC connection

    PC Program PC connection is not possible PC program for result and setup transfer

    Operation Modes ET 71 A / ET 71 V / ETM 4 ET 91

    Transmitting Provided Provided

    Selective Receiving Provided Provided

    Wide band Receiving Provided Provided

    Impedance Measurement With external adapter Provided

    Return Loss Measurement With external adapter Provided

    Balance Measurement With external adapter Provided

    Wide band Noise Measurement Not provided Additional Service

    Impulse Noise Measurement Not provided Additional Service

    Spectrum Analyzer Not provided Additional Service

    Master Slave Measurement Not provided Additional Service

    Multi Tone Measurement Not provided Additional Service

    NEXT/LOSS Measurement Not provided Additional Service

    Battery Quick Charge Not provided Additional Service

    Battery Regenerative Charge Not provided Additional Service

    Parameter ET 71 A ET 91

    Frequency

    Range: 200 Hz...620 kHz

    Accuracy: 2x10-5 1 digit

    Resolution: ~1 Hz

    Range: 100 Hz... 2.4 MHz

    Accuracy: 2x10-6 1 digit

    Resolution: 1 Hz

    Input Impedance Symmetrical and Coax:

    0, 75, 135 (125), 150, 600

    Symmetrical and Coax:

    0, 75, 135(125), 150, 600

    Output Level Balanced: +10 to 50 dBm, dB

    Coax: +10 to 50 dBm, dB

    Balanced: +10 ... 50 dBm, dB

    Coax: +10 ... 50 dBm

    Level accuracy Balanced: ~ 0,5 dB

    Coax: ~ 0,6 dB

    Balanced: ~ 0,3 dB

    Coax: ~ 0,3 dB

  • Parameter ET 71 V ET 91

    Frequency Range Wideband: 50 Hz... 620 kHz

    Selective: 200 Hz...620 kHz

    Wideband: 100 Hz... 2.4 MHz

    Selective: 100 Hz...2.4 MHz

    Band Widths 20 Hz, 3.1 kHz 20, 200 Hz, 1.74, 1.95, 3.1 kHz

    Terminated Inputs 75, 135 (125), 150, 600 5% 75, 135 (125), 150, 600 2%

    Unterminated Inputs

    Balanced:

    300 Hz620 kHz 5 k

    10 kHz..300 kHz 8 k

    Coax:

    3 k 50 pF

    Balanced and Coax:

    100 Hz... 2,4 MHz >20 k 30 pF

    Measuring Range Wideband:-60 to +40 dB (With attenuator)

    Selective: -90 to +40 dB (With attenuator)

    Wideband:-60 to +40 dB (With attenuator)

    Selective: -90 to +40 dB (With attenuator)

    Minimum Readable Level

    Selective: -120 dB

    Wideband: -90 dB

    Selective: -120 dB

    Wideband

    With 600 kHz filter: -80 dB

    With 4 kHz filter: -100 dB

    Measuring Accuracy

    Balanced Input:

    0 dB 20 kHz: 0,3 dB

    Coax Input:

    0 dB 20 kHz: 0,5 dB

    Balanced Input:

    0 dB 100 kHz: 0,2 dB

    Coax Input:

    0 dB 100 kHz: 0,2 dB

    Auto Ranging Not provided In all measuring modes

    Wideband Noise Measurement Not provided With 9 selectable filters

    Impulse Noise Measurement Not provided Maximum 72 hours

    Spectrum Analyzer Not provided

    Minimum Readable Level: -140 dBm/Hz

    Ranges

    2.4 MHz

    1.5 MHz

    600 kHz

    300 kHz

    20 kHz

    4 kHz

    Parameter ETM - 4 ET - 91

    Impedance Measurement

    With external adapter

    Frequency range:

    300 Hz620 kHz

    Measuring range:

    50 .3000

    Accuracy: 10% With built in bridge

    Return Loss Measurement

    With external adapter and

    external normal resistance

    Frequency range:

    300 Hz620 kHz

    Measuring range:

    60 .1200

    Accuracy: 1dB

    With built in bridge and

    Internal normal resistance

    Frequency range:

    200 Hz... 2,4 MHz

    Measuring range:

    60 .1200

    Accuracy: 1dB

    LCL Balance Measurement

    With external adapter

    Frequency range:

    300 Hz620 kHz

    Measuring range:

    0.40 dB

    Accuracy: 1dB

    With built in bridge

    Frequency range:

    200 Hz... 2,4 MHz

    Measuring range:

    0.40 dB

    Accuracy: 1dB