Espion Prof i le System Ganzfeld ERG•EOG•Flash VEP
Espion Profile GanzfeldThe Espion Profile Ganzfeld ERG system uses software and hardware developed entirely by Diagnosys LLC to provide a compact, transportable system for recording not only any ISCEV standard ERG, but any conceivable ERG, EOG, or flash VEP test. Based on the world’s most advanced Ganzfeld stimulator, the ColorDome, paired with the most advanced electrophysiology amplifiers available, the precision-machined cart-mounted system is a powerful addition to any clinic or research facility. The ColorDome’s monocular companion, the ColorBurst, is optimized for pediatric or research use and may be mounted on the cart in addition to or in place of the ColorDome.
Stimulus featuresColorDome*
• Flash luminance range from 10-9 to 3000cd.s/m2
• Anywhere within this > trillion to one range luminance can be changed with no less than 1% resolution
• Any color background from 10-9 to 500 cd/m2
• Supports any Ganzfeld ERG test ever devised*see brochures for more detailed specs
Hardware features• 5 Channel headbox with industry-first 32-bit amplifiers• Drives either or both ColorDome and ColorBurst Stimulators• Integrated Power Supply Isolation Unit• 3 system configurations: Stand-alone, cart-mounted with power lift table (shown right) Stand-alone, bench-mounted (overleaf) or integrated into the Espion System
l e a d i n g t h e w a v e
ColorBurst*
• Flash luminance range from 10-9 to 40 cd.s/m2
• Anywhere within this > 10 billion to one range luminance can be changed with no less than 1% resolution
• Any color background from 10-9 to 500 cd/m2
• Easily supports any ISCEV standard or ancillary ERG test
DIA_Brochures 2011 CMYK.indd 3 4/27/11 1:37:40 PM
Copyright© 2011, Diagnosys LLC. The specification and features listed herein are accurate as of the date of printing. However Diagnosys reserves the right to change or modify specifications and features at any time without notice.
DOC # REV # DATE 13603 A 04/11/2011
Diagnosys LLC USA175 Cabot Street, Suite 500Lowell, MA 01854 USA
Phone 978.458.1600 Fax [email protected]
Diagnosys UK54 Impington LaneImpington,Cambridge CB249NJEngland
Phone +44(0) 1223 520699 Fax +44(0) 1223 [email protected]
www.diagnosysllc.com
Espion Prof i le System Ganzfeld ERG•EOG•Flash VEP
colorburst™ colordome™
Software features• Full integration with Espion software• Easy setup of standard and custom
protocols• Print preview and customizable print
layout templates• Export data
Run-time features• Real time trace display during testing• 50Hz/60Hz line noise removal option• Noise level monitoring• Lights on headbox guide lead
insertion and indicate quality of electrode contact
• Impedances can be checked at any time
• Adjustable real-time blink rejection• Smart data averaging• Patient fixation monitoring through
built in camera application• DC-coupled 32-bit amplifiers produce
"textbook" waveforms and cannot be overdriven
Analysis features• Show one to five channels onscreen
at one time• View, scale and average traces +
restore to original trace values• Edit sweeps in average after
acquisition
Database features• Username and password
security option• Different user authorization levels• Industry standard SQL relational
database• Client/Server facility to link across the
net or through a local network• Automatically generated and time-
stamped audit trail, necessary for clinical trials, makes it virtually impossible to lose data
• Automatic database backup option• Built-in password security with support
for CFR 21 part 11• Normative database available
ISCEV Standard Tests• Rod Response (Dark-adapted 0.01
ERG)• Maximal Rod-Cone Response
(Dark-adapted 3.0 ERG)• Oscillatory Potentials (Dark-adapted
3.0 oscillatory potentials )• Single Flash Cone Response
(Light-adapted 3.0 ERG)• 30 Hz Flicker (Light-adapted 3.0
flicker ERG)• Dark Adapted 10.0 ERG • Dark Adapted 30.0 ERG
A few of the many additional protocols not covered by ISCEV standards
• on–off responses• S-cone ERG• Photopic negative response• Scotopic threshold response (STR)• Double-flash ERG
Show one to five channels onscreen Show one to five channels onscreen
Client/Server facility to link across the
Automatically generated and time-
Automatic database backup optionBuilt-in password security with support
Desktop OptionThe Diagnosys Profile Ganzfeld ERG
system is also available in a desktop
mounted version.
TMalso compatible with:
espionespionpro�le
Scan here with Smart Device for our website
DIA_Brochures 2011 CMYK.indd 4 4/27/11 1:37:44 PM