EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of...

94
K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode EPIC Calibration Meeting, MPE Properties of EPIC Properties of EPIC pn at low energies: pn at low energies: LW mode LW mode
  • date post

    19-Dec-2015
  • Category

    Documents

  • view

    215
  • download

    1

Transcript of EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of...

Page 1: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

Properties of EPIC pn Properties of EPIC pn

at low energies: LW at low energies: LW

modemode

Page 2: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

Long closed LW exposures

revolution data setexposure

[s]date

0546 0134522001_PNS005 18 760 2002-12-03

0553 0125320901_PNS010 12 267 2002-12-17

0730 0154150601_PNS002 12 198 2003-12-05

0790 0134720801_PNS005 61 096 2004-04-01

09740305920301_PNU02

712 534 2005-04-04

116 855

Page 3: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn at low energies (LW)

Correction of offset shifts

Page 4: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

rev 546 rev 553

offset correction: before

Page 5: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

rev 546 rev 553

offset correction: after

Page 6: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

rev 546

EPIC-pn detector noise (LW)

residual offset map

Page 7: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 553

residual offset map

Page 8: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 730

residual offset map

Page 9: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 790

residual offset map

Page 10: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 974

residual offset map

Page 11: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

individual offset maps median of offset maps

Page 12: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

residual offset mapsrev 546 rev 790

Page 13: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

rev 546 rev 790cleaned residual offset maps

Page 14: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

rev 546 cleaned residual offset map residual offset map

Page 15: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

20 adu image, rev 546

Page 16: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

20 adu image, rev 546

before the correction of offset shifts

Page 17: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

20 adu image, rev 546

after the correction of offset shifts

Page 18: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

rev 546 rev 553

offset correction: before

Page 19: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

rev 546 rev 553

offset correction: after

Page 20: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

rev 730 rev 974

offset correction: before

Page 21: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

rev 730 rev 974

offset correction: after

Page 22: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

Low energy noise

EPIC-pn at low energies (LW)

Page 23: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

adu

Quadrant 0

50.7 ks closed FF (rev 059)

EPIC-pn detector noise

Page 24: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

2.16 : 12.16 : 1

Detector noise in FF and eFF mode

FF

eFF

detector noise per per secsec is ~2.16 times

higher in FF than in eFF

Page 25: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

2.16 : 12.16 : 1

Detector noise in FF and eFF mode

FF

eFF

detector noise per per secsec is ~2.16 times

higher in FF than in eFF

LW

Page 26: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

Detector noise in LW and FF mode

1.71 : 11.71 : 1

LW

FF

detector noise per per secsec is ~1.7 times higher in LW than

in FF**in the LW area

Page 27: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

Page 28: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

closed FF rev 462 23.2 ks

0 50 0 3

Page 29: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

Vela SNR FF rev 534 38.8 ks

without epreject

Page 30: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

with epreject

Vela SNR FF rev 534 38.8 ks

Page 31: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise in LW mode

Page 32: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

100% suppressio

n

Original method (“method 1”):

Suppression of events below 50 adu according to a reference noise

distribution deduced from long calclosed

exposures

Shown in blue is the result of this method applied to a (long)

calclosed exposure, to illustrate the effect on

a very faint X-ray source

20 50 adu

Page 33: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

20 50 adu

99% suppressio

n

Original method (“method 1”):

Suppression of events below 50 adu according to a reference noise

distribution deduced from long calclosed

exposures

Shown in blue is the result of this method applied to a (long)

calclosed exposure, to illustrate the effect on

a very faint X-ray source

Page 34: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

20 50 adu

90% suppressio

n

Original method (“method 1”):

Suppression of events below 50 adu according to a reference noise

distribution deduced from long calclosed

exposures

Shown in blue is the result of this method applied to a (long)

calclosed exposure, to illustrate the effect on

a very faint X-ray source

Page 35: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

20 50 adu

70% suppressio

n

Original method (“method 1”):

Suppression of events below 50 adu according to a reference noise

distribution deduced from long calclosed

exposures

Shown in blue is the result of this method applied to a (long)

calclosed exposure, to illustrate the effect on

a very faint X-ray source

Page 36: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

20 50 adu

50% suppressio

n

Original method (“method 1”):

Suppression of events below 50 adu according to a reference noise

distribution deduced from long calclosed

exposures

Shown in blue is the result of this method applied to a (long)

calclosed exposure, to illustrate the effect on

a very faint X-ray source

Page 37: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

This method works well for bright sources,

but may cause an apparent deficit of photons between ~150 and 250 eV for faint sources

Other problem: inhomogeneous residual noise at

~20 adu

EPIC-pn detector noise (LW)

20 50 adu

Page 38: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

Alternative method (“method 2”):

Try to determine the noise properties at higher

energies and extrapolate them down to 20 adu

EPIC-pn detector noise (LW)

Page 39: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

range used for fit: 40-100 adu

range shown: 20-400 adu

log-log

EPIC-pn detector noise (LW)

rev 790, all events

Page 40: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

range used for fit: 40-100 adu

range shown: 20-400 adu

x lin y log

rev 790, all events

Page 41: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

range used for fit: 40-100 adu

range shown: 20-100 adu

x lin y log

rev 790, all events

Page 42: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

range used for fit: 40-100 adu

range shown: 20-400 adu

x lin y log

rev 790, all events

Page 43: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

range used for fit: 40-100 adu

range shown: 20-400 adu

x log y log

rev 790, all events

Page 44: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

range used for fit: 40-100 adu

range shown: 20-400 adu

x log y log

rev 790, all events

Page 45: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

for comparison: method 1

EPIC-pn detector noise (LW)

range used for fit: 40-100 adu

range shown: 20-400 adu

x log y log

rev 790, all events

Page 46: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

method 2, rev 790

EPIC-pn detector noise (LW)

Page 47: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

method 2, rev 546

Page 48: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

method 2, rev 553

Page 49: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

method 2, rev 730

Page 50: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

method 2, rev 974

Page 51: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

Method 2 would work well if there were no temporal changes of the low-

energy noise

Whenever the noise properties of the sample and template are (slightly) different, the subtraction leads to

residuals in the spectra and the images

Other problem: assumption of constant noise properties along CCD rows (required in order to get enough

statistics) not valid for raw events:

EPIC-pn detector noise (LW)

Page 52: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 790

61 ks closed

after offset corrections

all events

20 adu

Page 53: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 790

61 ks closed

after offset corrections

all events

21 adu

Page 54: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 790

61 ks closed

after offset corrections

all events

22 adu

Page 55: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 790

61 ks closed

after offset corrections

all events

23 adu

Page 56: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 790

61 ks closed

after offset corrections

all events

24 adu

Page 57: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 790

61 ks closed

after offset corrections

all events

25 adu

Page 58: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 790

61 ks closed

after offset corrections

all events

26 adu

Page 59: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 790

61 ks closed

after offset corrections

all events

27 adu

Page 60: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 790

61 ks closed

after offset corrections

all events

28 adu

Page 61: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 790

61 ks closed

after offset corrections

all events

29 adu

Page 62: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 790

61 ks closed

after offset corrections

all events

30 adu

Page 63: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 790

61 ks closed

after offset corrections

all events

31-35 adu

Page 64: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 790

61 ks closed

after offset corrections

all events

36-40 adu

Page 65: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 790

61 ks closed

after offset corrections

all events

41-45 adu

Page 66: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 790

61 ks closed

after offset corrections

all events

46-50 adu

Page 67: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 790

61 ks closed

after offset corrections

all events

51-55 adu

Page 68: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 790

61 ks closed

after offset corrections

all events

56-60 adu

Page 69: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 790

61 ks closed

after offset corrections

all events

61-70 adu

Page 70: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 790

61 ks closed

after offset corrections

all events

71-80 adu

Page 71: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 790

61 ks closed

after offset corrections

all events

81-90 adu

Page 72: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 790

61 ks closed

after offset corrections

all events

91-100 adu

Page 73: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 790

61 ks closed

after offset corrections

all events

101-120 adu

Page 74: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 790

61 ks closed

after offset corrections

all events

121-140 adu

Page 75: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 790

61 ks closed

after offset corrections

all events

141-180 adu

Page 76: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 790

61 ks closed

after offset corrections

all events

181-250 adu

Page 77: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 790

61 ks closed

after offset corrections

all events

251-400 adu

Page 78: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

At which stage of the processing should the noise be suppressed ?

Comparison of raw data with processed data (after pattern

recognition and recombination):

Page 79: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 790

61 ks closed

no offset corrections

all events

20 adu

raw dataraw data

Page 80: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 790

61 ks closed

no offset corrections

singles

20 adu

processed dataprocessed data

Page 81: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 790

61 ks closed

no offset corrections

all events

30-50 adu

raw dataraw data

Page 82: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 790

61 ks closed

no offset corrections

all “entries”

30-50 adu

processed dataprocessed data

Page 83: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 790

61 ks closed

no offset corrections

non-singles

30-50 adu

processed dataprocessed data

Page 84: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (LW)

rev 790

61 ks closed

no offset corrections

singles

30-50 adu

processed dataprocessed data

Page 85: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

Comparison with FF:

EPIC-pn detector noise

Page 86: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (FF)

20 adu

raw dataraw data

rev 462

23 ks closed

no offset corrections

all events

Page 87: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (FF)

rev 462

23 ks closed

no offset corrections

singles

20 adu

processed dataprocessed data

Page 88: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (FF)

rev 462

23 ks closed

no offset corrections

all events

30-50 adu

raw dataraw data

Page 89: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (FF)

rev 462

23 ks closed

no offset corrections

all “entries”

30-50 adu

processed dataprocessed data

Page 90: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (FF)

rev 462

23 ks closed

no offset corrections

non-singles

30-50 adu

processed dataprocessed data

Page 91: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise (FF)

rev 462

23 ks closed

no offset corrections

singles

30-50 adu

processed dataprocessed data

Page 92: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

EPIC-pn detector noise

At which stage of the processing should the noise be suppressed ?

Noise near the CCD edges can be most efficiently suppressed by

rejecting invalid patterns

Noise suppression should be done after the pattern

recognition

Page 93: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE

raw_y

20-65 adu

nois

e

EPIC-pn detector noise properties (FF)

SAS CCF:

Also for FF, epreject may need some refinement..

Page 94: EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.

K. Dennerl, 2006 May 04EPIC pn at low energies: LW modeEPIC Calibration Meeting, MPE