ENVISION - odev.com · technique according to the surgeon’s judgment and experience. ... plate is...
Transcript of ENVISION - odev.com · technique according to the surgeon’s judgment and experience. ... plate is...
![Page 1: ENVISION - odev.com · technique according to the surgeon’s judgment and experience. ... plate is positioned on the anterior surface of the verte-bral body. Mediolateral and caudocranial](https://reader030.fdocuments.us/reader030/viewer/2022021612/5b9ba48009d3f22d2a8d7b7f/html5/thumbnails/1.jpg)
E N V I S I O N®
2
A N T E R I O R C E R V I C A L P L A T E
S U R G I C A L T E C H N I Q U E
O R T H O D E V E L O P M E N T S P I N E
![Page 2: ENVISION - odev.com · technique according to the surgeon’s judgment and experience. ... plate is positioned on the anterior surface of the verte-bral body. Mediolateral and caudocranial](https://reader030.fdocuments.us/reader030/viewer/2022021612/5b9ba48009d3f22d2a8d7b7f/html5/thumbnails/2.jpg)
E N V I S I O N®
2 A N T E R I O R C E R V I C A L P L A T E
C O N T E N T S
l . I N T R O D U C T I O N . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
l l . S U R G I C A L E X P O S U R E A N D A P P R O A C H . . . . . . . . . . . . . 2
l l l . P L AT E S E L E C T I O N , C O N T O U R I N G A N D P O S I T I O N I N G . . . . . . 3
l V. T E M P O R A R Y P L A T E F I X A T I O N . . . . . . . . . . . . . . . . . . . 4
V. S C R E W H O L E P R E P A R A T I O N . . . . . . . . . . . . . . . . . . . . 4
V I . S C R E W I N S E R T I O N A N D L O C K I N G . . . . . . . . . . . . . . . . 6
V I l . I M P R O P E R S C R E W S E A T I N G . . . . . . . . . . . . . . . . . . . . . 8
V I l l . S C R E W R E M O V A L . . . . . . . . . . . . . . . . . . . . . . . . . . 1 0
l X . C O L L E T R E P L A C E M E N T . . . . . . . . . . . . . . . . . . . . . . . 1 1
X . C L O S U R E A N D P O S T O P E R A T I V E C A R E . . . . . . . . . . . . . 1 3
E N V I S I O N I N S T R U M E N T C A S E . . . . . . . . . . . . . . . . . 1 4
E N V I S I O N K I T. . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 6
A P P E N D I X . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 8
O R D E R I N G I N F O R M A T I O N . . . . . . . . . . . . . . . . . . . . . 1 8
D E S I G N I N G S U R G E O N S :
T h o m a s M . S w e e n e y , M . D . , P h . D . , F. A . A . O . S .Southeastern Spine Center
M a r k B . L o n s t e i n , M . D . , F. A . A . O . S .Sarasota Spine Specialists
The following technique is a general guide for instrumentation of the Envision®2 Anterior Cer-
vical Plate. It is expected that the surgeon is already familiar with the fundamentals of cervical
spinal fusion. Each patient represents an individual case that may require modification of the
technique according to the surgeon’s judgment and experience.
Please see the package insert for intended uses/indications, device description, contraindications, pre-
cautions, warnings and potential risks associated with the Envision®2 Anterior Cervical Plate.
U.S. Federal Law restricts this device to sale by or on the order of a physician.
I l l u s t r a t i o n s Jill Rhead, M.A.
![Page 3: ENVISION - odev.com · technique according to the surgeon’s judgment and experience. ... plate is positioned on the anterior surface of the verte-bral body. Mediolateral and caudocranial](https://reader030.fdocuments.us/reader030/viewer/2022021612/5b9ba48009d3f22d2a8d7b7f/html5/thumbnails/3.jpg)
l . I N T R O D U C T I O N
The Envision®2 Anterior Cervical Plate offers a unique blend of features engineered for ease-of-use and superior
clinical results. The rotationally-dynamic design enables load-sharing between implant and graft for a more reliable
fusion. It has a seamless, “no-step” self-locking mechanism while maintaining a low profile (2.0mm) that virtu-
ally eliminates soft tissue irritation. And, with large central fenestrations and a narrow footprint, the Envision®2 Anterior
Cervical Plate provides superior visualization of the interbody space.
T H E E N V I S I O N®
2 A N T E R I O R C E R V I C A L P L AT E I N C O R P O R AT E S T H E L AT E S T
I N C E R V I C A L P L AT E T E C H N O LO GY F O R S P I N A L F U S I O N .
1
S T R E N GT H A N D F L E X I B I L I T Y
The Envision®2 Anterior Cervical Plate offers a unique
blend of flexibility and strength, which increases load-
sharing and decreases the potential for implant fracture.
S E L F - LO C K I N G S Y S T E M
The Envision®2 Anterior Cervical Plate patented self-
locking mechanism automatically locks screws to the plate
during insertion. The mechanism employs both a rim lock
and a taper lock within the collet to minimize the potential
for screw backout and to provide audible, visual and tactile
indicators of locking mechanism engagement.
T H I N A N D N A R R O W
The Envision®2 Anterior Cervical Plate has a thin
(2mm) and narrow (16mm) design to minimize soft
tissue irritation and accommodate patient anatomy.
L A R G E G R A F T W I N D O W S
The Envision®2 Anterior Cervical Plate provides excel-
lent visualization of the graft and vertebral endplates
during placement of the plate.
1
2
3
4
1
2
3
4
![Page 4: ENVISION - odev.com · technique according to the surgeon’s judgment and experience. ... plate is positioned on the anterior surface of the verte-bral body. Mediolateral and caudocranial](https://reader030.fdocuments.us/reader030/viewer/2022021612/5b9ba48009d3f22d2a8d7b7f/html5/thumbnails/4.jpg)
l l . S U R G I C A L E X P O S U R E A N D A P P R O A C H
The patient is placed on the operating room table in the supine position with the head in slight extension and slight rotation
opposite the side of incision. The laryngeal nerve that controls the voicebox runs on the right side of the neck. The incision
is typically left sided so as to avoid potential post-op speech problems. Following appropriate positioning and padding, the
patient is prepped and draped in a sterile manner. A horizontal incision is made in the skin within a skin fold, if possible. The
incision is divided sharply. Electro-cautery can be used to control bleeding. Dissection then continues down to the level of the
subcutaneous tissue and platysma. The platysma is either divided or incised and retracted, allowing access to the fascial plane
between the sternocleidomastoid and “strap muscles.” The dissection is carried down to the level of the anterior aspect of the
spine, where the prevertebral fascia is incised using scissors and then dissected along the disc space using standard dissection
techniques. A spinal needle is used to mark the disc space; placement is verified via an intraoperative X-ray.
Having confirmed the appropriate level, a cervical dis-
traction set can be used with distracting pins in the up-
per and lower vertebral body. Gross dissection is then
performed using a #11 blade and pituitary rongeur. If
desired, an operating microscope can be used. Using mi-
cro curettes and rongeurs, completion of the discectomy
and/or corpectomy can be performed to the level of the
posterior longitudinal ligament. Removal of the poste-
rior longitudinal ligament can be performed to expose
the dura. (Figure 1) Having completed the discectomy,
the end plates are decorticated and posterior osteophytes
are removed, if necessary.
Interbody bone grafting is then performed for each disc
space or motion segment to be fused. Prior to placing
plates, all anterior osteophytes are removed to allow
placement of the plate on a smooth surface.
2
F I G U R E 1
![Page 5: ENVISION - odev.com · technique according to the surgeon’s judgment and experience. ... plate is positioned on the anterior surface of the verte-bral body. Mediolateral and caudocranial](https://reader030.fdocuments.us/reader030/viewer/2022021612/5b9ba48009d3f22d2a8d7b7f/html5/thumbnails/5.jpg)
l I I . P L A T E S E L E C T I O N , C O N T O U R I N G A N D P O S I T I O N I N G
When selecting the appropriate plate, it is important to know that the nominal length is based on the distance between screw
hole centers. The overall plate length is 8mm longer than the stated nominal length (see Appendix, Figure A).
The Plate Holder may be used to hold the plate against the cervical vertebrae for plate size selection. The Plate
Holder mates with the plate by grasping the section between any two screw hole pairs (Figure 2). To avoid en-
croachment into adjacent levels, the appropriate plate length is determined by placing the superior and inferior
screw holes close to the disc or corpectomy space being fused.
After an appropriate plate size has been selected, the
plate is positioned on the anterior surface of the verte-
bral body. Mediolateral and caudocranial position is then
confirmed.
The Envision®2 Anterior Cervical Plate is pre-contoured
for an average lordotic curve and vertebral body diameter.
In most cases, this standard geometry will be adequate. If
needed, the lordotic curvature of the plate may be modified
using the Plate Bender* (Figure 3).
It is recommended that the Plate Bender be adequately
lubricated with surgical instrument lubricant. To modify
the lordotic curve, carefully attach the cervical plate to the
Plate Bender and bend the plate to the desired curvature
within the “bending zones.” *
Excessive and/or reversed bending of the plate should be
avoided as this can significantly weaken the plate.
*Note: A ‘bending zone’ is the area between the screw-hole pairs. If
a plate is bent through the screw holes, the locking mechanism will
be damaged.
3
F I G U R E 2
F I G U R E 3
P L AT E H O L D E R
C E R V I C A L P L AT E
P L AT E B E N D E R
![Page 6: ENVISION - odev.com · technique according to the surgeon’s judgment and experience. ... plate is positioned on the anterior surface of the verte-bral body. Mediolateral and caudocranial](https://reader030.fdocuments.us/reader030/viewer/2022021612/5b9ba48009d3f22d2a8d7b7f/html5/thumbnails/6.jpg)
I V . T E M P O R A R Y P L A T E F I X A T I O N
After correct positioning of the plate has been established, a Fixation Pin may be used to provide stability during screw
placement. The Fixation Pin may be inserted through any of the screw holes using the Hex Driver (Figures 4 and 5).
Note: The Fixation Pin is designed for single use.
4
F I G U R E 4F I G U R E 5
V. S C R E W H O L E P R E P A R A T I O N
S E L F - D R I L L I N G
In preparation for screw insertion, the Punch may be used
to create a pilot hole. The Punch will center the pilot hole
in the screw hole of the plate; otherwise, the screw might
not lock to the plate. The Punch is placed in the desired
screw hole, angled up to 15°, and gently pressed into the
bone while rotating the handle about its axis (Figure 6).
The depth stop on the Punch provides for a pilot hole that
is 10mm deep.
S E L F - T A P I N G
Alternately, the pilot hole can be drilled. Both Hand
Drills and Power Drills are provided (Figure 7). A Single
Barrel Guide is provided for use with the two drill types
to center the drilled pilot holes in the screw holes.
F I G U R E 6
H E X D R I V E R
F I X AT I O N P I N
P U N C H
P U N C H
P L AT E H O L D E R
![Page 7: ENVISION - odev.com · technique according to the surgeon’s judgment and experience. ... plate is positioned on the anterior surface of the verte-bral body. Mediolateral and caudocranial](https://reader030.fdocuments.us/reader030/viewer/2022021612/5b9ba48009d3f22d2a8d7b7f/html5/thumbnails/7.jpg)
F I G U R E 7
The tip of the Single Barrel Guide is placed inside the collet of the plate
(Figure 8). This mate allows the surgeon to vary the angle for the pilot hole
in any direction. It is important to take care and limit the angle to 15° in any
one direction or the subsequently placed screw might not lock to the collet.
The main function of the Single Barrel Guide is to insure the pilot hole is
centered in the screw hole of the plate; otherwise, the screw might not lock to
the plate. The Single Barrel Guide also functions with the Drills to produce
a known pilot hole depth (the created pilot hole will be approximately 1mm
deeper than the corresponding screw).
Select the appropriate Drill for the corresponding screw length. Screws and
Drills are color matched for easy length identification (12mm: Green; 14mm:
Blue; 16mm: Magenta). Place the Drill through the Single Barrel Guide and
advance the Drill into the vertebral body to the depth allowed by the stop
on the Drill.
The Tap is provided for use in those rare cases where dense bone might
make screw advancement difficult. The Tap can be placed through the Single
Barrel Guide and advanced as far as the stop allows. This creates a tapped-
hole depth of 10mm.
5
F I G U R E 8
S I N G L E B A R R E LG U I D E
![Page 8: ENVISION - odev.com · technique according to the surgeon’s judgment and experience. ... plate is positioned on the anterior surface of the verte-bral body. Mediolateral and caudocranial](https://reader030.fdocuments.us/reader030/viewer/2022021612/5b9ba48009d3f22d2a8d7b7f/html5/thumbnails/8.jpg)
6
F I G U R E 1 0 F I G U R E 1 1
C O L L E T 1 - 2 M M
P L AT E
S C R E W
F I G U R E 9
H E X D R I V E R
V l . S C R E W I N S E R T I O N A N D L O C K I N G
Prior to screw insertion, it is good practice to clean any debris from the
screw hole. Debris may prevent the screw from fully seating and thus the
screw might not lock to the plate. The selected screw is inserted using the
Hex Driver. The mate between the Hex Driver and cervical screw is self-
retaining. This is achieved by gently pressing the Hex Driver into the hex
of a screw within the screw caddy. The Hex Driver is then used to drive
the screw into position (Figure 9).
P R O P E R I N S E R T I O N
The correct way to insert a screw into the collet and secure the Envision®2
Anterior Cervical Plate is demonstrated in the figures below:
Figure 10 - Illustrates a screw fully seated and properly locked in the collet.
Note: Fully seated, the head of the screw seats 1-2mm below the rim of the collet.
Figure 11 - Cutaway view that conveys the design intent of the screw
properly seated in the collet locking mechanism.
![Page 9: ENVISION - odev.com · technique according to the surgeon’s judgment and experience. ... plate is positioned on the anterior surface of the verte-bral body. Mediolateral and caudocranial](https://reader030.fdocuments.us/reader030/viewer/2022021612/5b9ba48009d3f22d2a8d7b7f/html5/thumbnails/9.jpg)
7
K E Y P O I N T S
• Listen for an audible click as the screw passes the anterior collet rim, followed by taper engagement. Look for the
position shown in Figures 10 and 11.
• Tightening the screw completely will lock the screw to the plate.
Two features are employed to lock the screw to the plate. The first feature is a rim on the collet that first expands
to allow the entrance of the screw and then retracts to keep the screw in place. The second feature is a Morse taper
that cold welds the screw to the collet (Figure 12). The rim will likely click as the screw passes into the collet. Then,
an additional 1 to 2 revolutions of the Hex Driver should suffice to engage the Morse taper. Resistence increases as
the Morse taper engages.
Note: It is good practice to visually inspect the screw after final tightening to ensure that the screw is seated fully in the collet,
below the locking rim, and that the collet is properly aligned within the plate screw hole.
F I G U R E 1 2
![Page 10: ENVISION - odev.com · technique according to the surgeon’s judgment and experience. ... plate is positioned on the anterior surface of the verte-bral body. Mediolateral and caudocranial](https://reader030.fdocuments.us/reader030/viewer/2022021612/5b9ba48009d3f22d2a8d7b7f/html5/thumbnails/10.jpg)
V l l . I M P R O P E R S C R E W S E A T I N G
While the Envision®2 Cervical Plate locking mechanism is one of the easiest to employ, it is still important to
distinguish between proper and improper screw seating. Therefore, once the screw is completely seated visual and
tactile verification will ensure that the locking mechanism will function as designed.
I M P R O P E R I N S E R T I O N
The figures below illustrate a screw that has been inserted and is flush with the collet, but is not fully seated in the collet.
Figure 13 - Illustrates a screw that is flush but not fully seated or properly locked in the collet.
Figure 14 - Cutaway view that conveys an improperly seated screw.
K E Y P O I N T S
• Although a screw may appear locked, a surgeon may falsely perceive that a screw is fully seated. A screw in this
position may present the appearance and tactile feel of a locked screw.
Note: If a screw is not fully seated or properly locked to the collet it will have an increased chance of back-out. Therefore it
is important to be certain that components are locked. Ambient noise may conceal the “click” and individual sensitivities
may cause false perception. Blood and tissue may impair visual verification. Therefore tactile feel may verify seating.
F I G U R E 1 3 F I G U R E 1 4
C O L L E T
P L AT E
S C R E W
8
![Page 11: ENVISION - odev.com · technique according to the surgeon’s judgment and experience. ... plate is positioned on the anterior surface of the verte-bral body. Mediolateral and caudocranial](https://reader030.fdocuments.us/reader030/viewer/2022021612/5b9ba48009d3f22d2a8d7b7f/html5/thumbnails/11.jpg)
Screw angle variability is achieved via the spherical collet in the plate.
This collet/plate interface accommodates up to 15° of angulation in
any direction (30º total) (Figure 15).
The locking mechanism of the Envision®2 Anterior Cervical Plate
has a working limit of 30º in screw angle variability. If the 30º
variability in screw angle is exceeded, the locking mechanism will
not function, which will greatly increase the likelihood of screw
back-out, and associated complications.
9
F I G U R E 1 5
S C R E W P A S S T H R O U G H
On occasion, an audible “click” is not heard or felt. If the “click” is not heard or felt, the user may then have
an inclination to continue tightening. Normally, an obvious tactile sensation of taper engagement is felt. A stopping
point is typically felt shortly (1-2 revolutions) after the audible click is heard or felt. The taper locking
mechanism is engaged at this point. Screw threading must cease 1-2 revolutions after the screw head passes the
anterior rim of the collet. Otherwise, a screw may be drawn completely through the collet and void the taper
locking mechanism.
15º15º15º15º
![Page 12: ENVISION - odev.com · technique according to the surgeon’s judgment and experience. ... plate is positioned on the anterior surface of the verte-bral body. Mediolateral and caudocranial](https://reader030.fdocuments.us/reader030/viewer/2022021612/5b9ba48009d3f22d2a8d7b7f/html5/thumbnails/12.jpg)
V l I I . S C R E W R E M O V A L
The cervical screws can be removed by simply reversing the
direction of driving, the removal of the screw will per-
manently deform the screw and collet, rendering them
both unusable. Note that repetitive threading of the
screw into the bone can decrease bone quality and screw
purchase.
If needed, the Screw Remover (Figure 16) may be used
to back a locked screw out of the cervical plate. This
will remove the collet from the plate. To do this, thread
the tip of the Screw Remover into the internal hex of the
screw by twisting the handle counter-clockwise. Then
twist the outer sleeve of the Screw Remover clockwise
until it contacts the plate. While holding the outer sleeve
of the Screw Remover against the plate, continue to
twist the handle of the Screw Remover counter-clock-
wise to thread the cervical screw out of the bone.
F I G U R E 1 6
10
S C R E W R E M O V E R
H A N D L E
O U T E R S L E E V E
![Page 13: ENVISION - odev.com · technique according to the surgeon’s judgment and experience. ... plate is positioned on the anterior surface of the verte-bral body. Mediolateral and caudocranial](https://reader030.fdocuments.us/reader030/viewer/2022021612/5b9ba48009d3f22d2a8d7b7f/html5/thumbnails/13.jpg)
11
F I G U R E 1 7
F I G U R E 1 8
K N O B
C O L L E T R E V I S I O N T O O L
C O R R E C T I N C O R R E C T
C O L L E T M A G A Z I N E
I X . C O L L E T R E P L A C E M E N T
Cervical plates are provided with the collets already installed. Nevertheless, in instances when a locked screw is re-
moved, the corresponding collet must be replaced. This is accomplished with the Collet Revision Tool and the
Collet Magazine (Figure 17).
The Collet Revision Tool is used to remove used collets from the plate and to insert replacement collets into the
plate. To engage a collet with the Collet Revision Tool, press the tip of the Collet Revision Tool into the collet and
lock its tip to the collet by tightening the knob at the top of the tool handle.
![Page 14: ENVISION - odev.com · technique according to the surgeon’s judgment and experience. ... plate is positioned on the anterior surface of the verte-bral body. Mediolateral and caudocranial](https://reader030.fdocuments.us/reader030/viewer/2022021612/5b9ba48009d3f22d2a8d7b7f/html5/thumbnails/14.jpg)
12
Improper orientation of the collet (i.e., upside down) would pre-
vent the screw from locking to the collet and could result in
screw back-out. It is good practice to retrieve new collets di-
rectly from the Collet Magazine with the Collet Revision Tool
to insure that the collet is oriented correctly. (Figure 18).
To remove a new collet from the cervical plate first, insert the
Collet Revision Tool into the collet. Next, tighten the knob on
the Tool. Using a rocking and swiveling motion with the Collet
Revision Tool the collet will release. (Figure 19).
Note: The collet cannot be directly pulled from the plate.
To install a collet, first retrieve a collet with the Collet Revision
Tool from the Collet Magazine; the knob in the Collet Revision
Tool must be loose. Next, tighten the knob to secure the collet.
The new collet is then installed in the plate by rocking and swiv-
eling the collet into the appropriate screw hole (Figure 20).
The collet cannot be impacted into place. The feeling of seating
the collet into proper position in the screw hole of the plate is
apparent. Ensure that the collet is fully seated in the plate by
spinning the collet about its axis to ensure the collet rotates free-
ly in the screw hole. The knob of the Collet Revision Tool is
then loosened to release the replaced collet.
F I G U R E 1 9
F I G U R E 2 0
![Page 15: ENVISION - odev.com · technique according to the surgeon’s judgment and experience. ... plate is positioned on the anterior surface of the verte-bral body. Mediolateral and caudocranial](https://reader030.fdocuments.us/reader030/viewer/2022021612/5b9ba48009d3f22d2a8d7b7f/html5/thumbnails/15.jpg)
13
X . C L O S U R E A N D P O S T O P E R A T I V E C A R E
After visual inspection and radiographic confirmation of optimal placement of the plate, screws, and bone graft, the wound
is irrigated with copious fluids. The wound is then closed over a drain, if desired. Closure technique is per
surgeon preference.
Collars may be used as indicated for patient comfort or for individual cases of decreased bone density after assessment of
quality. Always take routine postoperative radiographs.
![Page 16: ENVISION - odev.com · technique according to the surgeon’s judgment and experience. ... plate is positioned on the anterior surface of the verte-bral body. Mediolateral and caudocranial](https://reader030.fdocuments.us/reader030/viewer/2022021612/5b9ba48009d3f22d2a8d7b7f/html5/thumbnails/16.jpg)
14
E N V I S I O N®
2 A N T E R I O R C E R V I C A L P L A T E I N S T R U M E N T S
Cervical Plate
Single Barrel Guide
Punch
Power Drill
Collet
Hand Drill
Hex Driver 2.5mm
Bone Tap
![Page 17: ENVISION - odev.com · technique according to the surgeon’s judgment and experience. ... plate is positioned on the anterior surface of the verte-bral body. Mediolateral and caudocranial](https://reader030.fdocuments.us/reader030/viewer/2022021612/5b9ba48009d3f22d2a8d7b7f/html5/thumbnails/17.jpg)
15
Plate Holder
Collet Magazine
Fixation Pins
Cervical Screws
Screw Remover
Plate Bender
Caliper
Collet Revision Tool
![Page 18: ENVISION - odev.com · technique according to the surgeon’s judgment and experience. ... plate is positioned on the anterior surface of the verte-bral body. Mediolateral and caudocranial](https://reader030.fdocuments.us/reader030/viewer/2022021612/5b9ba48009d3f22d2a8d7b7f/html5/thumbnails/18.jpg)
16
E N V I S I O N®
2 A N T E R I O R C E R V I C A L P L A T E K I T
Envision Kit 254-9003
MIDDLE BOTTOM
PART NO. DESCRIPTION QTY
254-6002 Cervical Plate Instrument Case
254-0007 Punch 1
254-1003 Bone Tap 1
254-1004 Fixation Pin 4
254-1005 Single Barrel Drill Guide 1
254-1006 Plate Bender 1
254-1008 Internal Hex Driver 2.5mm 2
254-1009 Screw Remover 1
254-1010 Plate Holder 1
PART NO. DESCRIPTION QTY
254-1015 Collet Revision Tool 1
254-1016 Collet Magazine 1
254-1312 Drill 12mm 1
254-1314 Drill 14mm 1
254-1316 Drill 16mm 1
254-2312 Power Drill 12mm 1
254-2314 Power Drill 14mm 1
254-2316 Power Drill 16mm 1
254-0011 Caliper 1
154-1000 Collet 10
CASE TOP TOP
![Page 19: ENVISION - odev.com · technique according to the surgeon’s judgment and experience. ... plate is positioned on the anterior surface of the verte-bral body. Mediolateral and caudocranial](https://reader030.fdocuments.us/reader030/viewer/2022021612/5b9ba48009d3f22d2a8d7b7f/html5/thumbnails/19.jpg)
17
N O T E S
![Page 20: ENVISION - odev.com · technique according to the surgeon’s judgment and experience. ... plate is positioned on the anterior surface of the verte-bral body. Mediolateral and caudocranial](https://reader030.fdocuments.us/reader030/viewer/2022021612/5b9ba48009d3f22d2a8d7b7f/html5/thumbnails/20.jpg)
12187 South Business Park Drive
Draper, Utah 84020
Telephone: 801.553.9991
Fax: 801.553.9993
For more information visit
www.orthodevelopment.com
Ortho Development and all other registered trademarks and trademarks are the property of Ortho Development Corporation. Patent(s) pending. Part No. 351-1-10739 REV 04/2015
A P P E N D I X O R D E R I N F O R M A T I O N
PRODUCT # LEVEL NOMINAL ACTUAL LENGTH LENGTH
154-5114 One 14mm 22mm
154-5116 One 16mm 24mm
154-5118 One 18mm 26mm
154-5120 One 20mm 28mm
154-5122 One 22mm 30mm
154-5124 One 24mm 32mm
154-5226 Two 26mm 34mm
154-5228 Two 28mm 36mm
154-5230 Two 30mm 38mm
154-5232 Two 32mm 40mm
154-5235 Two 35mm 43mm
154-5238 Two 38mm 46mm
154-5241 Two 41mm 49mm
154-5244 Two 44mm 52mm
154-5247 Two 47mm 55mm
154-5346 Three 46mm 54mm
154-5349 Three 49mm 57mm
154-5352 Three 52mm 60mm
154-5355 Three 55mm 63mm
154-5358 Three 58mm 66mm
154-5361 Three 61mm 69mm
154-5364 Three 64mm 72mm
154-5367 Three 67mm 75mm
154-5460 Four 60mm 68mm
154-5465 Four 65mm 73mm
154-5470 Four 70mm 78mm
154-5475 Four 75mm 83mm
154-5480 Four 80mm 88mm
154-5485 Four 85mm 93mm
154-5490 Four 90mm 98mm
154-5495 Four 95mm 103mm
C E R V I C A L P L A T E( F I G U R E A )
S E L F - D R I L L I N G S C R E W( F I G U R E B )
GREEN 12mm
BLUE 14mm
MAGENTA 16mm
D R I L L C O L O R M A T C H I N G
S E L F - T A P P I N G S C R E W ( F I G U R E C )
F I G U R E A
16 m m
NO
MIN
AL
AC
TU
AL
2 m m
LE
NG
TH
F I G U R E B
LE
NG
TH
F I G U R E C
154-6012 4.0mm 12mm Light Green
154-6014 4.0mm 14mm Light Blue
154-6016 4.0mm 16mm Light Magenta
154-6312 4.35mm 12mm Dark Green
154-6314 4.35mm 14mm Dark Blue
154-6316 4.35mm 16mm Dark Magenta
PRODUCT # SIzE ACTUAL COLOR (DIAMETER) LENGTH
PRODUCT # SIzE ACTUAL COLOR (DIAMETER) LENGTH
154-6240 4.0mm 12mm Light Green
154-6440 4.0mm 14mm Light Blue
154-6640 4.0mm 16mm Light Magenta
154-6243 4.35mm 12mm Dark Green
154-6443 4.35mm 14mm Dark Blue
154-6643 4.35mm 16mm Dark Magenta